IDS-iSYS Intact PINP Instructions For Use IS-4000 IS-4000 v06 01 February 2017 English Page 1/6 1. Intended Use For In Vitro Diagnostic Use. The IDS-iSYS Intact PINP assay (intact amino-terminal propeptide of type I procollagen) is intended for the quantitative determination of intact PINP in human serum or plasma on the IDS-iSYS Multi-Discipline Automated System. The test is intended to be used as an aid in the management of postmenopausal osteoporosis. 2. Summary and Explanation An important step in the bone formation process is synthesis of type I collagen, which is the major organic component in bone matrix. During collagen synthesis, propeptides are released from both the amino- and the carboxyterminal parts of the procollagen molecule 1 . These propeptides are secreted into the blood circulation. The amino-terminal propeptide of type I procollagen (PINP) is probably the most specific and sensitive marker of bone formation 2 . PINP is a particularly useful marker for monitoring the efficacy of osteoporosis therapy with anabolic agents 3,4 , but it is also one of the best bone turnover markers for monitoring the efficacy of antiresorptive therapy 4,5 . 3. Method Description The assay is based on chemiluminescence technology. 20 μL of patient samples or calibrators are diluted in a diluent and a portion of this is incubated with a biotinylated anti-PINP monoclonal antibody, an acridinium labelled monoclonal antibody, streptavidin labelled magnetic particles and an assay buffer. The magnetic particles are captured using a magnet and a wash step performed to remove any unbound analyte. Trigger reagents are added; the resulting light emitted by the acridinium label is directly proportional to the concentration of intact PINP in the original sample. 4. Warnings and Precautions The IDS-iSYS Intact PINP is for in vitro diagnostic use only and is not for internal use in humans or animals. This product must be used strictly in accordance with the instructions set out in these Instructions For Use (IFU). Immunodiagnostic Systems Limited (IDS) will not be held responsible for any loss or damage (except as required by statute), howsoever caused, arising out of non-compliance with the instructions provided. CAUTION: This kit contains material of animal origin. Handle kit reagents as if capable of transmitting an infectious agent. Appropriate precautions and good laboratory practice must be used in the storage, handling and disposal of the kit reagents. Disposal of kit reagents should be in accordance with local regulations. Human materials Human material used in the preparation of this product has been tested by FDA recommended assays for the presence of antibody to Human Immunodeficiency Virus (HIV I and II), Hepatitis B surface antigen, antibody to Hepatitis C, and found negative. As no test can offer complete assurance that infectious agents are absent, the reagents should be handled according to Biosafety Level 2. Reagents containing Sodium Azide Calibrators CAL contain sodium azide (NaN3) >0.1% (w/w) (<1%). Classification under CLP: Acute Tox. 4 Aquatic Chronic 3 Hazard statements: EUH032 Contact with acids liberates very toxic gas H302 Harmful if swallowed H412 Harmful to aquatic life with long lasting effects Precautionary statements: P264 Wash hands thoroughly after handling. P270 Do not eat, drink or smoke when using this product. P273 Avoid release to the environment. P301/310 IF SWALLOWED: Immediately call a POISON CENTER or doctor. P330 Rinse mouth. 5. Handling Precautions Apart from the calibrators, which are lyophilised, the reagents provided in the kit are ready to use. Refer to the Calibrator section of the procedure for reconstitution methodology. Before a new cartridge is loaded onto the IDS-iSYS Multi-Discipline Automated System (System), the magnetic particle container requires mixing by the operator with a brisk rotation motion. This will re-suspend the magnetic particles that have settled during shipment. Care must be taken to avoid foam formation in the reagent cartridge.

Upload: vodung

Post on 23-Mar-2018

243 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Instructions For Use

IS-4000

IS-4000 v06 01 February 2017 English Page 1/6

1. Intended Use For In Vitro Diagnostic Use. The IDS-iSYS Intact PINP assay (intact amino-terminal propeptide of type I procollagen) is intended for the quantitative determination of intact PINP in human serum or plasma on the IDS-iSYS Multi-Discipline Automated System. The test is intended to be used as an aid in the management of postmenopausal osteoporosis.

2. Summary and Explanation An important step in the bone formation process is synthesis of type I collagen, which is the major organic component in bone matrix. During collagen synthesis, propeptides are released from both the amino- and the carboxyterminal parts of the procollagen molecule

1. These propeptides are secreted into the blood circulation. The amino-terminal propeptide of type I

procollagen (PINP) is probably the most specific and sensitive marker of bone formation2. PINP is a particularly useful marker

for monitoring the efficacy of osteoporosis therapy with anabolic agents3,4

, but it is also one of the best bone turnover markers for monitoring the efficacy of antiresorptive therapy

4,5.

3. Method Description The assay is based on chemiluminescence technology. 20 µL of patient samples or calibrators are diluted in a diluent and a portion of this is incubated with a biotinylated anti-PINP monoclonal antibody, an acridinium labelled monoclonal antibody, streptavidin labelled magnetic particles and an assay buffer. The magnetic particles are captured using a magnet and a wash step performed to remove any unbound analyte. Trigger reagents are added; the resulting light emitted by the acridinium label is directly proportional to the concentration of intact PINP in the original sample.

4. Warnings and Precautions The IDS-iSYS Intact PINP is for in vitro diagnostic use only and is not for internal use in humans or animals. This product must be used strictly in accordance with the instructions set out in these Instructions For Use (IFU). Immunodiagnostic Systems Limited (IDS) will not be held responsible for any loss or damage (except as required by statute), howsoever caused, arising out of non-compliance with the instructions provided. CAUTION: This kit contains material of animal origin. Handle kit reagents as if capable of transmitting an infectious agent. Appropriate precautions and good laboratory practice must be used in the storage, handling and disposal of the kit reagents. Disposal of kit reagents should be in accordance with local regulations.

Human materials Human material used in the preparation of this product has been tested by FDA recommended assays for the presence of antibody to Human Immunodeficiency Virus (HIV I and II), Hepatitis B surface antigen, antibody to Hepatitis C, and found negative. As no test can offer complete assurance that infectious agents are absent, the reagents should be handled according to Biosafety Level 2. Reagents containing Sodium Azide

Calibrators CAL contain sodium azide (NaN3) >0.1% (w/w) (<1%).

Classification under CLP: Acute Tox. 4 Aquatic Chronic 3 Hazard statements: EUH032 Contact with acids liberates very toxic gas H302 Harmful if swallowed H412 Harmful to aquatic life with long lasting effects

Precautionary statements: P264 Wash hands thoroughly after handling. P270 Do not eat, drink or smoke when using this product. P273 Avoid release to the environment. P301/310 IF SWALLOWED: Immediately call a POISON CENTER or doctor. P330 Rinse mouth.

5. Handling Precautions Apart from the calibrators, which are lyophilised, the reagents provided in the kit are ready to use. Refer to the Calibrator section of the procedure for reconstitution methodology. Before a new cartridge is loaded onto the IDS-iSYS Multi-Discipline Automated System (System), the magnetic particle container requires mixing by the operator with a brisk rotation motion. This will re-suspend the magnetic particles that have settled during shipment. Care must be taken to avoid foam formation in the reagent cartridge.

Page 2: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 February 2017 English Page 2/6

Shelf Life and Storage of Reagents Store the cartridge and calibrators in an upright position in the dark at 2 - 8 °C. Do not freeze the cartridge or the calibrators.

Reagent shelf life Cartridge Calibrators

Before opening at 2 - 8 °C To the expiry date

Cartridge after opening at 2 - 8 °C

28 Days N/A

Calibrators after reconstitution at 2 - 8 °C

N/A 28 Days

On board the System * 7 Days 3 Hours

* Continuous on board stability

6. Sample Collection and Storage The assay should be performed using serum (standard sampling tubes or tubes containing serum separating gel) or plasma (lithium heparin, sodium heparin, ammonium heparin or potassium (K2) EDTA) samples. Samples should be separated as soon as possible after collection. Store samples at -20 °C or below. Avoid repeated freeze - thaw of samples. Note:

- Samples containing particulate matter must be centrifuged before performing the assay. Do not use heat-inactivated samples.

- To minimise possible evaporation effects, samples, calibrators and controls should be measured within 3 hours of being placed on the System.

- Each laboratory should follow the guidelines or requirements of local, state, and/or federal regulations or accrediting organizations to establish its own specimens handling and storage stability. For guidance on appropriate practices, please refer to the CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Before performing assays, make sure that samples, calibrators and controls are at room temperature (18 - 22 °C). - Some commercially available sample collection tubes may affect the results of testing in particular cases. - It is recommended to follow the instructions of the tube manufacturer especially when processing samples in primary

tubes. - Centrifuged samples with a lipid layer on the top must be transferred to a sample cup or secondary tube. Care must

be taken to transfer only the clarified samples without the lipemic material.

7. Materials Materials Provided Reagent Cartridge

MPS3 Magnetic particles coated with streptavidin in phosphate BSA buffer with sodium azide as preservative (<0.1 %),

1 bottle, 2.3 mL

CONJ Anti-PINP murine monoclonal antibody labelled with an acridinium ester derivative, in buffer containing bovine and

mouse proteins with sodium azide as preservative (<0.1 %), 1 bottle, 4.95 mL

Ab-BIO Anti-PINP murine monoclonal antibody labelled with biotin, in buffer containing mouse and bovine proteins with

sodium azide as preservative (<0.1 %) 1 bottle, 16.8 mL

BUF Buffer containing mouse and bovine proteins, with sodium azide as preservative (<0.1 %), 1 bottle, 4.95 mL

DIL Buffer containing sodium azide as preservative (<0.1 %), 1 bottle, 23.7 mL

Calibrators

CALA Lyophilised buffer containing BSA and purified human PINP with <1.0 % sodium azide as preservative (<0.05 %

CALB reconstituted). 2 each of 2 concentration levels, 1.0 mL per bottle.

The Calibrators are prepared at the following nominal ranges:

Calibrator Rangea (ng/mL)

CALA 3.0 – 7.0 ng/mL

CALB 130.0 – 165.0 ng/mL

a Ranges are typical for the product and are intended as a guide only. Values may vary from lot to lot. Refer to the Certificate of

Analysis (CofA) on the mini CD for batch specific concentrations. The calibrator ranges are established by testing the calibrator samples in a minimum of 20 assay runs on one system. Mini CD Contains IFU for IDS-iSYS reagents, control ranges, CRY files and Certificate of Analysis.

Page 3: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 February 2017 English Page 3/6

Materials Required But Not Provided System Optional equipment / materials IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Assay Procedure Reagent Cartridge The reagents provided in the cartridge are ready to use, with the exception of the calibrators. Before a new cartridge is loaded on board the System, mix the magnetic particles container with a brisk rotation motion. Avoid foam formation. The barcode is read when the cartridge is loaded on the reagent tray. If the label cannot be read by the System barcode reader, a manual procedure exists to enter the barcode data (see the IDS-iSYS User Manual). The System automatically performs mixing of the magnetic particles for all reagent cartridges loaded on-board to maintain homogeneity. The cartridge should be loaded into the reagent tray at least 40 minutes prior to starting the assay. If the cartridge is removed from the reagent tray, store the cartridge upright at 2 - 8 °C in the dark. Calibrators The calibrators of this kit are traceable to the manufacturer’s working calibrators which are prepared from highly purified human PINP whose value is traceable to a commercially available method. The IDS-iSYS Intact PINP assay calibrators are lyophilised. Reconstitute immediately before use. Add 1.0 mL of distilled or deionised water to each bottle. Replace the stopper. Leave for 10 minutes to reconstitute with occasional gentle mixing by hand. Care should be taken to avoid the formation of foam. Calibrators can be loaded onto the System using either the sample cups or calibrator/control racks (30/31/32). Calibrator/control racks: place the calibrator vials into the appropriate rack and insert onto the System and proceed according to the instructions of the System User Manual. Discard the material in the calibrator vials after use. DO NOT re-use calibrator vials that have been inserted onto the System. Sample cups: Pipette approximately 200 μL of calibrators into sample cups and place onto the System. Proceed according to the instructions of the System User Manual. Discard the material in the sample cups after use. DO NOT return material to the calibrator vial. If calibrators are to be used more than once, they should be stored at 2-8 °C within 3 hours following reconstitution. When re-using the stored calibrator vials, bring the calibrators to room temperature (18 - 22°C) and invert gently before use. Calibration Frequency A new calibration is required:

- Upon loading each new lot of cartridges. - When the control values do not fall within the defined ranges. - When the calibration interval of 3 days has expired. - After System service.

Verification of the calibration is automatic and managed by the System. System Calibration The two IDS-iSYS Intact PINP assay calibrators are required to perform the adjustment of the master curve. The calibrators are supplied with the kit and calibrators from another lot must not be used. All levels of IDS-iSYS Intact PINP Control Set (IS-4030) MUST be measured at the same time as the calibrators to validate the master curve adjustment. All data required for the calibration of the cartridge can be found on the mini CD. Use calibrator levels A and B to adjust the master curve to the reagents on board the System. Check for the presence of an IDS-iSYS Intact PINP assay cartridge on the reagent tray and the availability of the cartridge master curve in the database. If the data for the lot of calibrators is not available on board the System, load the data using the mini CD provided with the reagent kit. Start the immunoassay calibration on the System according to the System User Manual. The calibration is carried out using 3 replicates of each level. One replicate may be removed to meet the calibration requirements. All control levels must also be measured in duplicate to calibrate the assay. Verify and approve the calibration according to the calibration status displayed in the calibration window and discard the remaining volume of calibrators and controls from the sample tray after use.

Page 4: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 February 2017 English Page 4/6

9. Quality Control Use the IDS-iSYS Intact PINP Control Set for quality control. To ensure validity of results at least three controls with varying levels of intact PINP should be measured. Other suitable control material can be used in addition to the IDS-iSYS Intact PINP Control Set. Controls should be tested at (or near) the beginning of every run containing patient samples and also during every calibration in accordance with local, state and/or federal regulations or accreditation requirements and your laboratory’s quality procedure. It is recommended that the controls be routinely run in duplicate. Laboratories should test controls at least once per shift. Refer to the IDS-iSYS Intact PINP Control Set IFU for preparation and handling instructions. Determination of Sample intact PINP levels Process samples according to the System User Manual.

10. Calculation of Results The concentration of intact PINP within each sample is calculated automatically. The display of the concentrations (screen or printed) is produced upon user request. The IDS-iSYS Intact PINP assay uses a 4-parameter logistic curve fit (4PL) to calculate the intact PINP concentrations.

11. Measurement Range (Reportable Range) The reportable range of the assay is 2 -230 ng/mL. Any value that reads below 2 ng/mL should be reported as “< 2 ng/mL”.

12. Dilution Samples with intact PINP concentrations above the reportable range should be diluted manually with a low concentration

human serum sample in a ratio of 1 in 2. The results for diluted samples must be multiplied by the dilution factor 2 and corrected for the concentration of the low sample.

13. Limitations of Use - As in the case of any diagnostic procedure, results must be interpreted in conjunction with the patient’s clinical

presentation and other information available to the physician. - The performance characteristics of this assay have not been established in a paediatric population. - The hook effect was tested using concentrations of analyte up to 2350 ng/mL. No hook effect was observed. - Heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro

immunoassays6. Patients routinely exposed to animals or to animal serum products can be prone to this interference

and anomalous values may be observed. - The following substances do not interfere in the IDS-iSYS intact PINP assay when the concentrations presented in

the following table are below the stated threshold.

Potentially Interfering Agent

Threshold Concentration

Lipid 2803 mg/dL

Bilirubin 200 mg/dL

Haemoglobin 500 mg/dL

Biotin 300 nmol/L

14. Expected Values The following ranges were determined using the IDS-iSYS Intact PINP assay and are provided for information only. The 95 % reference interval for apparently healthy adults were calculated by a non-parametric method following guidance from CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Normal Adults 27.7 – 127.6 ng/mL (n = 150) The above ranges should be considered as guidelines only; it is recommended that each laboratory establish its own expected range based upon its own patient population.

Page 5: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 February 2017 English Page 5/6

15. Performance Data Representative performance data are shown. Results obtained at individual laboratories may vary. 15.1 Sensitivity The limit of blank (LoB), limit of detection (LoD) and limit of quantitation (LoQ) were determined with guidance from CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation” using 60 blanks and 120 low level samples.

Sensitivity Concentration

LoB 0.5 ng/mL

LoD < 1.0 ng/mL

LoQ < 1.0 ng/mL

15.2 Precision Precision was evaluated in accordance with a modified protocol based on CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”. A total of 3 serum controls were assayed using one lot of reagents in quadruplicate once per day for 20 days on two systems.

Control n Mean

(ng/mL)

Within run Total

SD CV SD CV%

1 80 14.3 0.4 2.6% 0.6 4.2%

2 80 44.6 1.3 3.0% 2.4 5.3%

3 80 100.9 3.0 3.0% 4.5 4.4%

15.3 Recovery Recovery was assessed by adding intact PINP to samples prior to assay.

Sample Conc. ng/mL

PINP added ng/mL

Measured ng/mL

Recovery ng/mL

Recovery %

44.3 44.3 33.5 33.5 37.3 37.3

35.3 74.8 35.3 74.8 35.3 74.5

77.1 115.1 66.4 101.5 70.8 105.2

32.8 70.9 32.9 68.0 33.5 67.9

Mean

93 95 93 91 95 91 93

15.4 Linearity Linearity was evaluated based on CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”. Samples containing varying concentrations of intact PINP were assayed in duplicate. The resulting mean concentrations were compared to predicted concentrations. Samples were prepared by diluting a high patient sample with a low patient sample prior to assay.

Predicted Concentration (ng/mL)

Measured Concentration (ng/mL)

Variation

ng/mL %

5.8 30.5 55.3 80.0 104.8 129.6 154.3 179.1 203.8

7.6 31.5 55.0 78.4 101.9 127.2 158.0 179.0 204.6

1.8 0.9 -0.2 -1.6 -2.9 -2.4 3.7 0.0 0.7

- 3 0 -2 -3 -2 2 0 0

15.5 Method Comparison The IDS-iSYS Intact PINP assay was compared against a commercially available quantitative radioimmunoassay, following CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. A total of 363 samples, selected to represent a wide range of intact PINP concentrations [8.7 - 230 ng/mL], was assayed by each method. Linear regression analysis was performed on the comparative data:

n Slope 95% CI Intercept (ng/mL) 95% CI Correlation Coefficient (r)

363 0.86 0.84 to 0.88 -1.9 -3.4 to -0.3 0.98

Page 6: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 February 2017 English Page 6/6

15.6 Specificity The specificity was assessed with the following analytes.

Analyte Cross-Reactivity

Intact PINP 100%

Rat / mouse PINP < 0.01%

PIINP < 0.03%

N-Mid Osteocalcin < 0.02%

16. Symbols used

REF Catalogue Number

IVD In Vitro Diagnostic Device

Manufacturer

Applied in accordance with directive 98/79/EC

Rx Only Caution: Federal law restricts this device to sale by or on the order of a physician (Only for US)

17. Bibliography 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., ‘Heterophilic antibodies: a problem for all immunoassays’. Clin Chem, 34, 1988, pp 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tel.: +44 191 519 0660 • Fax: +44 191 519 0760 e-mail: [email protected] • www.idsplc.com Immunodiagnostic Systems USA Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tel.: +1 877 852 6210 • Fax: +1 301 990 4236 e-mail: [email protected]

Germany Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tel.: +49 692 6019 0960 • Fax: +49 692 6019 0949 e-mail: [email protected]

France Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tel.: +33 140 770 450 • Fax: +33 140 770 455 e-mail: [email protected]

Scandinavia Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tel: +45 448 400 91 e-mail: [email protected]

Belgium Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tel.: +32 425 226 36 • Fax: +32 425 251 96 e-mail: [email protected]

Brazil IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasil. CEP: 05422 001 Tel: +55 113 740 6100 • Fax: + 55 113 740 6105 email: [email protected]

Page 7: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Mode d’emploi

IS-4000

IS-4000 v06 01 février 2017 Français Page 1/6

1. Utilisation prévue Pour utilisation diagnostique In vitro. Le dosage IDS-iSYS Intact PINP (propeptide amino-terminal du procollagène de type I total) est destiné à la détermination quantitative du PINP total dans le sérum ou le plasma humain dans l’IDS-iSYS Multi-Discipline Automated System. Le test doit être une aide dans la prise en charge de l’ostéoporose post-ménopausique.

2. Summary and Explanation La synthèse du collagène de type I, principal constituant organique de la matrice osseuse, constitue une étape importante du processus de formation de l’os. Lors de la synthèse du collagène, les propeptides sont libérés à la fois de la partie amino-terminale et de la partie carboxy-terminale de la molécule de procollagène

1. Ces propeptides sont ensuite libérés dans la

circulation sanguine. Le propeptide amino-terminal du procollagène de type I (PINP) est probablement le marqueur le plus spécifique et le plus sensible de la formation osseuse

2. Le PINP est un marqueur particulièrement utile dans la surveillance de

l’efficacité du traitement de l’ostéoporose avec des agents anaboliques3,4

, mais c’est également l’un des meilleurs marqueurs du turnover osseux dans la surveillance de l’efficacité des traitements aux agents anti-résorption

4,5.

3. Description de la méthode Le dosage repose sur la technologie de la chimiluminescence. 20 µL d’échantillon patient ou de calibrateurs sont dilués dans un diluant et une partie est incubée avec un anticorps monoclonal anti-PINP biotinylé, un anticorps monoclonal marqué à l’acridinium, des particules magnétiques marquées à la streptavidine et un tampon de dosage. Les particules magnétiques sont capturées à l’aide d’un aimant et un lavage est effectué pour évacuer tout analyte non lié. Des réactifs déclencheurs sont ajoutés ; la lumière résultante émise par le marquage à l’acridinium est directement proportionnelle à la concentration de PINP total dans l’échantillon d’origine.

4. Avertissements et précautions L’IDS-iSYS Intact PINP est destiné exclusivement au diagnostic in vitro et ne convient pas à un usage interne chez l’homme ou l’animal. Ce produit doit être utilisé en stricte conformité avec les instructions contenues dans ce Mode d’emploi (MdE). Immunodiagnostic Systems Limited (IDS) ne pourra être tenu responsable de quelque perte ou dommage que ce soit (sauf dans les cas prévus par la loi) découlant du non-respect des instructions fournies. ATTENTION : ce kit contient une substance d’origine animale. Manipulez les réactifs du kit en gardant à l’esprit qu’ils peuvent transmettre un agent infectieux. Vous devez appliquer les précautions et les bonnes pratiques de laboratoire qui s’imposent pour le stockage, la manipulation et l’élimination des réactifs du kit. L’élimination des réactifs du kit doit être effectuée dans le respect des réglementations locales.

Matériaux d’origine humaine La substance humaine utilisée dans la préparation de ce produit a été soumise aux analyses recommandées par la FDA pour la détection des anticorps du virus de l’immunodéficience humaine (VIH I et II), de l’antigène de surface de l’hépatite B, des anticorps de l’hépatite C, et s’est révélée négative. Aucun test ne pouvant garantir de manière absolue l’absence d’agents infectieux, les réactifs doivent être manipulés au niveau de biosécurité 2. Réactifs contenant de l’azoture de sodium

Les calibrateurs CAL contiennent de l’azoture de sodium (NaN3) > 0,1 % (pourcentage massique) (< 1%).

Règlement CLP (classification, étiquetage et emballage) : Acute Tox. 4 Aquatic Chronic 3 Mentions de danger: EUH032: Au contact d'un acide, dégage un gaz très

toxique. H302: Nocif en cas d'ingestion. H412: Nocif pour les organismes aquatiques, entraîne

des effets néfastes à long terme.

Conseils de prudence: P264: Se laver mains soigneusement après

manipulation. P270: Ne pas manger, boire ou fumer en manipulant ce

produit. P273: Éviter le rejet dans l'environnement. P301+310: EN CAS D'INGESTION: appeler immédiatement

un CENTRE ANTIPOISON ou un médecin. P330: Rincer la bouche.

5. Précautions de manipulation À l’exception des calibrateurs, qui sont lyophilisés, les réactifs fournis dans le kit sont prêts à l’emploi. Se reporter à la section relative aux calibrateurs pour connaître la méthode de reconstitution.

Page 8: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 février 2017 Français Page 2/6

Avant de charger une nouvelle cartouche dans l’IDS-iSYS Multi-Discipline Automated System (le Système), l’opérateur doit mélanger le récipient à particules magnétiques en lui appliquant un mouvement de rotation rapide. Cela remet en suspension les particules magnétiques qui se sont déposées pendant le transport. Prenez soin d’éviter la formation de mousse dans la cartouche de réactif.

Durée de conservation et stockage des réactifs Stockez les cartouches et les calibrateurs en position verticale dans l’obscurité à 2–8 °C. Ne congelez pas les cartouches et les calibrateurs.

Durée de conservation des réactifs Cartouche Calibrateurs

Avant ouverture à 2–8 °C Jusqu’à la date d’expiration

Cartouche après ouverture entre 2 et 8 °C

28 jours N/A

Étalons après reconstitution entre 2 et 8 °C

N/A 28 jours

Après chargement dans le Système * 7 jours 3 heures

* Stabilité continue après chargement

6. Prélèvement et stockage des échantillons Le dosage doit être effectué avec des échantillons de sérum (tubes de prélèvement standard ou tubes contenant du gel séparateur de sérum) ou de plasma (héparine de lithium, héparine de sodium, héparine d’ammonium ou EDTA de potassium [K2]). Les échantillons doivent être séparés dès que possible après prélèvement. Stockez les échantillons à une température inférieure ou égale à -20 °C. Évitez de congeler et décongeler les échantillons à plusieurs reprises. Remarque :

- Les échantillons contenant des particules doivent être centrifugés avant l’analyse. N’utilisez pas d’échantillons inactivés par la chaleur.

- Pour minimiser tout effet d’évaporation possible, les échantillons, les calibrateurs et les contrôles doivent être mesurés dans les 3 heures qui suivent leur placement dans le Système.

- Il incombe à chaque laboratoire de respecter les consignes ou réglementations locales et/ou nationales ou celles des organismes agréés en la matière afin de définir ses propres normes de manipulation des échantillons et de stabilité lors du stockage. Pour des conseils sur les pratiques appropriées, reportez-vous à CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Avant d’effectuer des dosages, assurez-vous que les échantillons, les calibrateurs et les contrôles sont à température ambiante (18–22 °C).

- Certains tubes de prélèvement disponibles dans le commerce sont susceptibles d’affecter les résultats des tests dans certains cas particuliers.

- Il est recommandé de respecter les consignes du fabricant du tube, notamment pour le traitement d’échantillons en tubes primaires.

- Les échantillons centrifugés qui sont recouverts d’une couche lipidique devront être transférés dans une coupelle ou un tube secondaire. Prenez soin de ne transférer que les échantillons clarifiés, sans la substance lipémique.

7. Matériaux Matériaux fournis Cartouche de réactif

MPS3 Particules magnétiques recouvertes de streptavidine dans un tampon BSA phosphaté avec de l’azoture de sodium

comme conservateur (< 0,1 %), 1 flacon, 2,3 mL

CONJ Anticorps monoclonal anti-PINP de souris marqué avec un dérivé d’ester d’acridinium dans un tampon contenant des

protéines bovines et de souris avec de l’azoture de sodium comme conservateur (< 0,1 %), 1 flacon, 4,95 mL

Ab-BIO Anticorps monoclonal anti-PINP de souris marqué à la biotine dans un tampon contenant des protéines bovines et de

souris avec de l’azoture de sodium comme conservateur (< 0,1 %) 1 flacon, 16,8 mL

BUF Tampon contenant des protéines bovines et de souris avec de l’azoture de sodium comme conservateur (< 0,1 %),

1 flacon, 4,95 mL

DIL Tampon contenant de l’azoture de sodium comme conservateur (< 0,1 %), 1 flacon, 23,7 mL

Calibrateurs

CALA Tampon lyophilisé contenant du BSA et du PINP humain purifié avec < 1,0 % avec de l’azoture de sodium comme

CALB conservateur (< 0,05 % reconstitué). 2 pour chacun des 2 niveaux de concentration, 1,0 mL par flacon.

Les calibrateurs sont préparés dans les plages nominales suivantes :

Calibrateur Plagea (ng/mL)

CALA 3,0–7,0 ng/mL

CALB 130,0–165,0 ng/mL

a Ce sont les plages types du produit et elles ne sont données qu’à titre indicatif. Les valeurs peuvent varier d’un lot à l’autre.

Se référer au Certificat d’analyse (Cd’A) sur le mini CD pour les concentrations spécifiques du lot. Les plages de calibrateurs sont définies en testant les échantillons de calibrateurs avec au moins 20 analyses dans un système.

Page 9: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 février 2017 Français Page 3/6

Mini CD Contient le mode d’emploi pour les réactifs IDS-iSYS, les plages de contrôle, les fichiers CRY et les Certificats d’analyse. Matériaux requis mais non fournis Système Matériaux/équipement optionnels IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) Kit de déclencheur IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Procédure d’analyse Cartouche de réactif Les réactifs fournis dans la cartouche sont prêts à l’emploi, à l’exception des calibrateurs. Avant de charger une nouvelle cartouche dans le Système, mélangez le récipient contenant les particules magnétiques en lui appliquant un mouvement de rotation rapide. Évitez la formation de mousse. Le code-barres est lu lorsque la cartouche est chargée dans le plateau de réactifs. Si l’étiquette ne peut pas être lue par le lecteur de code-barres du Système, une procédure manuelle permet de saisir les données du code-barres (voir le Manuel d’utilisation IDS-iSYS). Le Système effectue automatiquement le mélange des particules magnétiques pour toutes les cartouches de réactifs qui ont été chargées afin d’en maintenir l’homogénéité. La cartouche doit être chargée dans le plateau de réactifs 40 minutes au moins avant le début du test. Si la cartouche est retirée du plateau de réactifs, il faudra la stocker en position verticale, dans l’obscurité, à 2–8 °C. Calibrateurs Les calibrateurs de ce kit peuvent être tracés grâce aux calibrateurs du fabricant qui sont préparés à partir d’un PINP humain hautement purifié dont la valeur peut être tracée grâce à une méthode disponible dans le commerce. Les calibrateurs du dosage IDS-iSYS Intact PINP sont lyophilisés. Reconstituer immédiatement avant utilisation. Ajoutez 1,0 mL d’eau distillée ou désionisée dans chaque flacon. Remplacer le bouchon. Laissez le flacon pendant 10 minutes pour la reconstitution en le mélangeant délicatement de temps à autre. Prenez soin d’éviter la formation de mousse. Les étalons peuvent être chargés sur le système soit à l’aide de cupules à échantillons soit de portoirs étalons/contrôles (30/31/32). Portoirs calibrateurs/contrôles : placez les flacons des calibrateurs sur le portoir approprié et insérez ce dernier dans le Système puis procédez conformément aux instructions du Manuel d’utilisation du système. Après utilisation, éliminez le contenu des flacons de calibrateur. NE PAS réutiliser des flacons de calibrateur qui ont été insérés dans le système. Cupules à échantillons : pipetez environ 200 μL des calibrateurs dans des cupules à échantillons puis placez celles-ci dans le Système. Procédez conformément aux instructions du Manuel d’utilisation du Système. Éliminez le contenu des coupelles après utilisation. NE PAS remettre de produit dans le flacon de calibrateur. Si un calibrateur doit être utilisé plusieurs fois, il faut le conserver à 2–8 °C dans les 3 heures suivant la reconstitution. Pour réutiliser des flacons de calibrateur ayant été stockés, amenez-les à température ambiante (18–22 °C) et retournez-les doucement avant utilisation. Fréquence de calibration Une nouvelle calibration est nécessaire :

- À chaque chargement d’un nouveau lot de cartouches. - Lorsque les valeurs de contrôle ne se situent pas dans les plages définies. - Quand l’intervalle de calibration de 3 jours est dépassé. - Après une maintenance du Système.

La vérification de la calibration est automatique et gérée par le Système. Calibration du Système Les deux calibrateurs du dosage IDS-iSYS Intact PINP sont nécessaires pour le réglage de la courbe de référence. Les calibrateurs sont fournis avec le kit et il ne faut pas utiliser de calibrateurs provenant d’un autre lot. Tous les niveaux de l’IDS-iSYS Intact PINP Control Set (IS-4030) DOIVENT être analysés en même temps que les calibrateurs afin de valider le réglage de la courbe de référence. Toutes les données nécessaires au calibrage de la cartouche peuvent être trouvées sur le mini CD. Utilisez les niveaux A et B du calibrateur pour régler la courbe de référence par rapport aux réactifs à bord du Système. Vérifiez la présence d’une cartouche de dosage IDS-iSYS Intact PINP dans le plateau de réactifs et la disponibilité de la courbe de référence de la cartouche dans la base de données. Si les données du lot de calibrateurs ne sont pas disponibles dans le Système, chargez les données à partir du mini CD fourni avec le kit de réactifs. Commencez la calibration de l’immunodosage sur le Système conformément au Manuel d’utilisation du Système. La calibration est effectuée à l’aide de 3 exemplaires de chaque niveau. Un de ces exemplaires pourra être retiré afin de répondre aux exigences de la calibration. Tous les niveaux de contrôle doivent également être mesurés en double pour étalonner le dosage.

Page 10: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 février 2017 Français Page 4/6

Vérifiez et approuvez la calibration selon le statut affiché dans la fenêtre de calibration et éliminez après utilisation ce qui reste des calibrateurs et des contrôles sur le plateau d’échantillon.

9. Contrôle qualité Utilisez l’IDS-iSYS Intact PINP Control Set pour le contrôle qualité. Pour garantir la validité des résultats, il faut analyser au moins trois contrôles à des taux de PINP total différents. Vous pouvez utiliser d’autres matériaux de contrôle que l’IDS-iSYS Intact PINP Control Set. Il faut tester des contrôles au début ou presque de chaque session de travail sur des échantillons patient ainsi que lors de chaque calibration effectuée en conformité avec les règlements ou les exigences d’accréditation des autorités locales, de l’état et/ou fédérales, et avec les procédures de qualité de votre laboratoire. Il est recommandé de toujours exécuter les contrôles en double. Les laboratoires se devront de tester les contrôles à chaque changement d’équipe. Reportez-vous au mode d’emploi de l’IDS-iSYS Intact PINP Control Set pour connaître les instructions de préparation et de manipulation. Détermination des niveaux de PINP total des échantillons Traitez les échantillons conformément au Manuel d’utilisation du Système.

10. Calcul des résultats La concentration de PINP total dans chaque échantillon est automatiquement calculée. L’affichage des concentrations (écran ou impression) est effectuée sur demande de l’utilisateur. Le dosage IDS-iSYS Intact PINP utilise un ajustement à la courbe logistique à 4 paramètres (4PL) pour calculer les concentrations de PINP total.

11. Plage de mesure (plage à rapporter) La plage rapportable du dosage est 2–230 ng/mL. Toute valeur inférieure à 2 ng/mL doit être rapportée comme « < 2 ng/mL ».

12. Dilution Les échantillons dont les concentrations de PINP total sont supérieures à la plage rapportable doivent être dilués manuellement avec un échantillon de sérum humain basse concentration à un taux de 1 pour 2. Les résultats obtenus avec des échantillons dilués doivent être multipliés par le facteur de dilution 2 pour obtenir la concentration de l’échantillon original.

13. Limites d’utilisation - Comme pour toute procédure diagnostique, les résultats doivent être interprétés eu égard au tableau clinique du

patient et aux autres informations à la disposition du médecin. - Les caractéristiques de performance de cette analyse n’ont pas été démontrées en pédiatrie. - L’effet « crochet » a été testé avec des concentrations d’analyte allant jusqu’à 2350 ng/mL. Aucun effet crochet

n’a été observé. - Des anticorps hétérophiles du sérum humain peuvent réagir avec les immunoglobulines du réactif et interférer avec

les immunodosages in vitro6. Les patients quotidiennement exposés aux animaux ou aux produits sériques d’origine

animale peuvent être sujets à ce type d’interférence, il convient alors d’examiner toute valeur anormale. - Les substances suivantes ne perturbent pas le dosage IDS-iSYS intact PINP quand les concentrations indiquées

dans le tableau suivant sont inférieures au seuil défini.

Agent pouvant interférer

Limite de Concentration

Lipide 2 803 mg/dL

Bilirubine 200 mg/dL

Hémoglobine 500 mg/dL

Biotine 300 nmol/L

14. Valeurs attendues Les plages suivantes ont été déterminées à l’aide du dosage IDS-iSYS Intact PINP et sont fournies à titre informatif uniquement. L’intervalle de référence de 95 % pour les adultes apparemment en bonne santé a été calculé à l’aide d’une méthode non paramétrique conformément à la directive CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Adultes sains 27,7–127,6 ng/mL (n = 150) Les plages ci-dessus doivent être considérées uniquement comme des directives, il est recommandé à chaque laboratoire d’établir ses propres plages attendues sur la base de sa propre population de patients.

Page 11: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 février 2017 Français Page 5/6

15. Données de performances Les données de performances représentatives s’affichent. Les résultats obtenus dans chaque laboratoire peuvent varier. 15.1 Sensibilité La limite de blanc (LB), la limite de détection (LD) et la limite de quantification (LQ) ont été déterminées d’après CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation” au moyen de 60 blancs et de 120 échantillons de faible niveau.

Sensibilité Concentration

LB 0,5 ng/mL

LD < 1,0 ng/mL

LQ < 1,0 ng/mL

15.2 Précision La précision a été évaluée suivant un protocole modifié basé sur CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”. Au total, 3 contrôles de sérum ont été testés au moyen d’un lot de réactifs en quatre exemplaires, une fois par jour pendant 20 jours, sur deux systèmes.

Contrôle n Moyenne (ng/mL)

Dans la session Total

Écart-type Coefficient de variation

Écart-type % coeff. variation

1 80 14,3 0,4 2,6 % 0,6 4,2 %

2 80 44,6 1,3 3,0 % 2,4 5,3 %

3 80 100,9 3,0 3,0 % 4,5 4,4 %

15.3 Régénération La régénération a été évaluée par l’ajout de PINP total aux échantillons avant le dosage.

Conc. échantillon

ng/mL

PINP ajouté ng/mL

Mesurée ng/mL

Régénération ng/mL

Régénération %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Moyenne

93 95 93 91 95 91 93

15.4 Linéarité La linéarité a été évaluée sur la base de CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”. Les échantillons contenant diverses concentrations de PINP total ont été analysés en double. Les concentrations moyennes issues de cette analyse ont été comparées aux concentrations prévues. Les échantillons ont été préparés en diluant un échantillon patient élevé avec un échantillon patient faible avant l’analyse.

Concentration prévue (ng/mL)

Concentration mesurée (ng/mL)

Variation

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9 -0,2 -1,6 -2,9 -2,4 3,7 0,0 0,7

- 3 0 -2 -3 -2 2 0 0

15.5 Comparaison des méthodes Le dosage IDS-iSYS assay name a été comparé à un radioimmunodosage quantitatif disponible dans le commerce, conformément à CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. Au total, 363 échantillons, sélectionnés pour représenter une large plage de concentrations de PINP total [8,7–230 ng/mL], ont été analysés avec chacune des méthodes. Une analyse de régression linéaire a été réalisée sur les données comparatives :

n Pente 95 % IC Point d’intersection (ng/mL) 95 % IC Coefficient de corrélation (r)

363 0,86 0,84 à 0,88 -1,9 -3,4 à -0,3 0,98

Page 12: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 février 2017 Français Page 6/6

15.6 Spécificité La spécificité a été évaluée avec les analytes suivants.

Analyte Réactivité croisée

PINP total 100 %

PINP de rat/souris < 0,01 %

PIINP < 0,03 %

Ostéocalcine N-Mid < 0,02 %

16. Symboles utilisés

REF Numéro de référence

IVD Dispositif de diagnostic In vitro

Fabricant

Appliqué conformément à la directive 98/79/CE

17. Bibliographie 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., ‘Heterophilic antibodies: a problem for all immunoassays’. Clin Chem, 34, 1988, pp 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tél. : +44 191 519 0660 • Fax : +44 191 519 0760 courriel : [email protected] • www.idsplc.com Immunodiagnostic Systems États-Unis Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tél. : +1 877 852 6210 • Fax : +1 301 990 4236 courriel : [email protected]

Allemagne Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tél. : +49 692 6019 0960 • Fax : +49 692 6019 0949 courriel : [email protected]

France Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tél. : +33 140 770 450 • Fax : +33 140 770 455 courriel : [email protected]

Scandinavie Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tél. : +45 448 400 91 courriel : [email protected]

Belgique Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tél. : +32 425 226 36 • Fax : +32 425 251 96 courriel : [email protected]

Brésil IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasil. CEP : 05422 001 Tél. : +55 113 740 6100 • Fax : + 55 113 740 6105 courriel : [email protected]

Page 13: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Gebrauchsanweisung

IS-4000

IS-4000 v06 01 Februar 2017 Deutsch Seite 1/6

1. Verwendungszweck Für die diagnostische In-vitro-Verwendung. Der IDS-iSYS Intact PINP-Assay (intaktes amino-terminales Propeptid des Typ-I-Prokollagens) dient zur quantitativen Bestimmung von intaktem PINP in Humanserum oder -plasma mit dem IDS-iSYS Multi-Discipline Automated System. Dieser Test ist als Hilfsmittel bei der Behandlung von postmenopausaler Osteoporose zu verwenden.

2. Summary and Explanation Ein wichtiger Abschnitt bei der Knochenbildung ist die Synthese des Typ I Kollagens, das die organische Hauptkomponente der Knochengrundsubstanz darstellt. Während der Kollagensynthese werden von dem amino- als auch dem carboxyterminalen Teil des Prokollagenmoleküls Propeptide freigesetzt

1. Diese Propeptide werden in den Blutkreislauf abgegeben. Das

aminoterminale Propeptid des Typ I Prokollagens (PINP) ist wahrscheinlich der Marker, der für die Knochenbildung am meisten spezifisch und empfindlich ist

2. PINP ist ein besonders nützlicher Marker für die Kontrolle der Wirksamkeit einer

Osteoporosetherapie mit Anabolika3,4

, doch auch einer der besten Knochenumbaumarker für die Kontrolle der Wirksamkeit einer anti-resorptiven Therapie

4,5.

3. Methodenbeschreibung Der Test basiert auf der Chemilumineszenz-Technologie. 20 µL der Patientenproben oder Kalibratoren werden in einem Verdünnungsmittel verdünnt und dann wird ein Teil davon mit einem biotinylierten monoklonalen Anti-PINP-Antikörper, einem mit Akridinium markierten Antikörper und mit Streptavidin markierten magnetischen Partikeln sowie einem Assay-Puffer inkubiert. Die magnetischen Partikel werden dann mithilfe eines Magneten eingefangen und es wird ein Waschschritt durchgeführt, um ungebundene Analyten zu entfernen. Anschließend werden Trigger-Reagenzien hinzugefügt. Das durch die Akridinium-Markierung abgegebene Licht ist direkt proportional zu der Konzentration an intaktem PINP in der ursprünglichen Probe.

4. Warnungen und Vorsichtsmaßnahmen Der IDS-iSYS Intact PINP-Assay ist ausschließlich für die In-vitro-Diagnose vorgesehen und nicht für die innere Anwendung bei Mensch und Tier geeignet. Das Produkt muss streng gemäß der in dieser Gebrauchsanweisung beschriebenen Anleitungen verwendet werden. Immunodiagnostic Systems Limited (IDS) ist nicht haftbar für Verluste oder Schäden aufgrund der Nichtbeachtung dieser Anweisungen (außer in Fällen, in denen dies durch Gesetze vorgeschrieben ist), unabhängig davon, wie diese entstanden sind. ACHTUNG: Dieses Kit enthält Material tierischen Ursprungs. Behandeln Sie Kit-Reagenzien so, als ob diese Infektionserreger übertragen könnten. Bei der Lagerung, Handhabung und Entsorgung von Kit-Reagenzien müssen geeignete Vorsichtsmaßnahmen und gute Laborpraktiken Anwendung finden. Die Entsorgung der Kit-Reagenzien nach den örtlichen Vorschriften durchführen.

Humanmaterial Humanmaterial, das für die Vorbereitung dieses Produkts verwendet wurde, wurde anhand von FDA-empfohlenen Assays auf das Vorhandensein von Antikörpern für den humanen Immundefizienz-Virus (HIV I und II), das Hepatitis B-Oberflächenantigen sowie Antikörper für Hepatitis C geprüft und als negativ befunden. Da bei keinem Test vollständig ausgeschlossen werden kann, dass infektiöse Stoffe vorhanden sind, müssen die Reagenzien gemäß der biologischen Schutzstufe 2 gehandhabt werden. Natriumazid enthaltende Reagenzien

Kalibratoren (CAL) enthalten Natriumazid (NaN3) > 0,1 % (Gew.-%) (< 1 %).

Einstufung (CLP): Acute Tox. 4 Aquatic Chronic 3 Gefahrenhinweise: EUH032: Entwickelt bei Berührung mit Säure sehr giftige

Gase. H302: Gesundheitsschädlich bei Verschlucken. H412: Schädlich für Wasserorganismen, mit langfristiger

Wirkung.

Sicherheitshinweise: P264: Nach Gebrauch Hände gründlich waschen. P270: Bei Gebrauch nicht essen, trinken oder rauchen. P273: Freisetzung in die Umwelt vermeiden. P301+310: BEI VERSCHLUCKEN: Sofort

GIFTINFORMATIONSZENTRUM oder Arzt anrufen.

P330: Mund ausspülen.

Page 14: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 Februar 2017 Deutsch Seite 2/6

5. Vorsichtsmaßnahmen bei der Handhabung Mit Ausnahme der lyophilisierten Kalibratoren sind die in diesem Kit enthaltenen Reagenzien gebrauchsfertig. Die Rekonstitution der Kalibratoren wird im Abschnitt zu den Kalibratoren beschrieben. Bevor eine neue Kartusche in das IDS-iSYS Multi-Discipline Automated System (System) geladen wird, muss der Behälter mit magnetischen Partikeln durch den Bediener mit einer kräftigen Drehbewegung gemischt werden. Dadurch werden die magnetischen Partikel, die sich während dem Versand abgesetzt haben, wieder aufgewirbelt. Achten Sie darauf, dass sich in der Reagenzienkartusche kein Schaum bildet. Haltbarkeit und Lagerung von Reagenzien Lagern Sie die Kartusche und Kalibratoren in einer senkrechten Position bei Dunkelheit und 2–8 °C. Kartusche oder Kalibratoren nicht einfrieren.

Reagenzien-Haltbarkeit Kartusche Kalibratoren

Vor dem Öffnen bei 2–8 °C Bis zum Verfallsdatum

Kartusche nach dem Öffnen bei 2–8 °C

28 Tage N. Z.

Kalibratoren nach der Rekonstitution bei 2–8 °C

N. Z. 28 Tage

Auf dem System * 7 Tage 3 Stunden

* Kontinuierliche Stabilität auf dem System

6. Probenentnahme und -lagerung Der Test sollte unter Verwendung von Serumproben (Standard-Probenentnahmeröhrchen oder Röhrchen mit Trenngel) oder Plasmaproben (Lithium-Heparin, Natrium-Heparin, Ammoniumheparin oder Kalium-EDTA (K2-EDTA)) durchgeführt werden. Die Proben sollten so bald wie möglich nach der Entnahme getrennt werden. Proben bei –20 °C oder darunter aufbewahren. Wiederholtes Einfrieren und Auftauen der Proben ist zu vermeiden. Hinweis:

- Proben, welche Partikelmasse enthalten, müssen vor dem Durchführen des Assays zentrifugiert werden. Keine hitzeinaktivierten Proben verwenden.

- Um die Auswirkungen einer Verdampfung so gering wie möglich zu halten, die Proben, Kalibratoren und Kontrollen innerhalb von 3 Stunden nach Einsetzen in das System messen.

- Jedes Labor hat bei der Sicherstellung seiner eigenen Probenhandhabungs- und Lagerstabilität die Richtlinien oder Anforderung lokaler, staatlicher und/oder bundesstaatlicher Regulierungs- oder Akkreditierungsbehörden zu beachten. Für Informationen zu geeigneten Vorgehensweisen die Richtlinien CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”, lesen.

- Stellen Sie vor dem Durchführen von Assays sicher, dass sich die Proben, Kalibratoren und Kontrollen auf Raumtemperatur (18–22 °C) befinden.

- Manche handelsübliche Probenentnahmeröhrchen können in bestimmten Fällen die Testergebnisse beeinträchtigen. - Es wird empfohlen, die Anweisungen des Röhrchenherstellers zu beachten, insbesondere, wenn Proben in

Primärröhrchen verarbeitet werden. - Zentrifugierte Proben mit einer Lipidschicht auf ihrer Oberfläche müssen in ein Probengefäß oder Sekundär-

Röhrchen gegeben werden. Es ist darauf zu achten, dass nur die geklärten Proben ohne lipemisches Material transferiert werden.

7. Materialien Bereitgestellte Materialien Reagenzienkartusche

MPS3 Magnetische, mit Streptavidin beschichtete Partikel in einem Phosphat-BSA-Puffer und mit Natriumazid als

Konservierungsmittel (< 0,1 %), 1 Flasche, 2,3 mL

CONJ Monoklonaler mausartiger Anti-PINP-Antikörper, markiert mit einem Akridiniumesterderivat, in einem Puffer mit

Rinder- und Mausproteinen und mit Natriumazid als Konservierungsmittel (< 0,1 %), 1 Flasche, 4,95 mL

Ab-BIO Monoklonaler mausartiger Anti-PINP-Antikörper, markiert mit Biotin, in einem Puffer mit Rinder- und Mausproteinen

und mit Natriumazid als Konservierungsmittel (< 0,1 %), 1 Flasche, 16,8 mL

BUF Puffer mit Maus- und Rinderproteinen und mit Natriumazid als Konservierungsmittel (< 0,1 %), 1 Flasche, 4,95 mL

DIL Puffer mit Natriumazid als Konservierungsmittel (< 0,1 %), 1 Flasche, 23,7 mL

Kalibratoren

CALA Lyophilisierter Puffer mit BSA und gereinigtem Human-PINP und mit < 1,0 % Natriumazid als Konservierungsmittel

CALB (< 0,05 % rekonstituiert). Je 2 für 2 Konzentrationen, 1,0 mL pro Flasche.

Die Kalibratoren werden mit folgenden nominalen Bereichen vorbereitet:

Kalibrator Bereicha (ng/mL)

CALA 3,0 bis 7,0 ng/mL

CALB 130,0 bis 165,0 ng/mL

a Bereiche sind typisch für das Produkt und dienen lediglich als Anhaltspunkt. Die Werte können von Charge zu Charge

variieren. Chargen-spezifische Konzentrationen sind dem Analysezertifikat (CoA) auf der Mini-CD zu entnehmen.

Page 15: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 Februar 2017 Deutsch Seite 3/6

Die Kalibratorbereiche werden ermittelt, indem die Kalibratorproben in mindestens 20 Assay-Durchläufen auf einem System getestet werden. Mini-CD Enthält die Gebrauchsanweisung für die IDS-iSYS-Reagenzien, Kontrollbereiche, CRY-Dateien und das Analysezertifikat. Erforderliche, aber nicht bereitgestellte Materialien System Optionale Geräte / Materialien IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Assay-Verfahren Reagenzienkartusche Die in der Kartusche bereitgestellten Reagenzien sind gebrauchsfertig, mit Ausnahme der Kalibratoren. Bevor eine neue Kartusche auf das System geladen wird, mischen Sie den Behälter mit den magnetischen Partikeln mit einer schnellen Rotationsbewegung. Schaumbildung vermeiden. Der Barcode wird eingelesen, sobald die Kartusche auf das Reagenzienfach geladen wird. Wenn das Etikett nicht vom Barcode-Lesegerät des Systems eingelesen werden kann, steht ein manuelles Verfahren zur Verfügung, mit dem die Barcode-Daten eingegeben werden können (siehe IDS-iSYS-Benutzerhandbuch). Das System führt den Mischvorgang der Partikel für alle geladenen Reagenzienkartuschen, die auf das System geladen wurden, automatisch aus, damit Homogenität sichergestellt wird. Die Kartusche muss mindestens 40 Minuten vor Beginn des Assays auf das Reagenzienfach geladen werden. Wird die Kartusche aus dem Reagenzienfach entfernt, lagern Sie die Kartusche senkrecht bei 2–8 °C bei Dunkelheit.

Kalibratoren Die Kalibratoren dieses Kits sind auf die Arbeitskalibratoren des Herstellers rückverfolgbar, die aus hochgereinigtem Human-PINP angesetzt wurden, dessen Wert auf eine handelsübliche Methode rückverfolgbar ist.

Die Kalibratoren des IDS-iSYS Intact PINP-Assays sind lyophilisiert. Erst unmittelbar vor Gebrauch rekonstituieren. Geben Sie 1,0 mL destilliertes oder deionisiertes Wasser in jede Flasche. Setzen Sie den Stopfen wieder auf. 10 Minuten zum Rekonstituieren stehen lassen und gelegentlich per Hand vorsichtig mischen. Achten Sie darauf, dass sich kein Schaum bildet.

Die Kalibratoren können auf das System entweder mithilfe der Probenbecher oder der Kalibrator-/Kontrollgestelle (30/31/32) geladen werden.

Kalibrator-/Kontrollgestelle: Setzen Sie die Kalibratoren in das entsprechende Gestell ein und laden Sie dieses auf das System. Fahren Sie anschließend gemäß den Anweisungen im Benutzerhandbuch des Systems fort. Verwerfen Sie das Material in den Kalibratorfläschchen nach der Nutzung. Die Kalibratorfläschchen, die auf das System geladen wurden, dürfen NICHT wiederverwendet werden.

Probenbecher: Pipettieren Sie circa 200 μL Kalibratorflüssigkeit in die Probenbecher und laden Sie diese auf das System. Fahren Sie gemäß den Anweisungen im Benutzerhandbuch des Systems fort. Verwerfen Sie das Material in den Probenbechern nach der Verwendung. Geben Sie das Material NICHT wieder in das Kalibratorfläschchen.

Wenn die Kalibratoren mehr als einmal verwendet werden sollen, müssen sie innerhalb von 3 Stunden nach der Rekonstitution bei 2–8 °C gelagert werden. Wenn Sie gelagerte Kalibratorfläschchen erneut verwenden, warten Sie, bis die Kalibratoren auf Raumtemperatur gebracht wurden (18–22 °C) und drehen Sie sie vor der Verwendung vorsichtig auf den Kopf.

Kalibrationshäufigkeit Eine neue Kalibration ist erforderlich:

- wenn eine neue Charge an Kartuschen geladen wird. - wenn sich die Kontrollwerte nicht innerhalb der festgelegten Bereiche befinden. - wenn das Kalibrierungsintervall von 3 Tagen überschritten wurde. - nach einer Instandhaltung des Systems.

Die Verifizierung der Kalibration geschieht automatisch und wird vom System verwaltet.

Systemkalibration Die zwei IDS-iSYS Intact PINP-Assay-Kalibratoren sind erforderlich, um die Anpassung der Masterkurve vornehmen zu können. Die Kalibratoren werden mit dem Kit geliefert. Die Kalibratoren aus einer anderen Charge dürfen nicht verwendet werden. Alle Konzentrationen des IDS-iSYS Intact PINP Control Set (IS-4030) MÜSSEN zur selben Zeit wie die Kalibratoren gemessen werden, um die Anpassung der Masterkurve zu validieren.

Alle für die Kalibration der Kartusche erforderlichen Daten befinden sich auf der Mini-CD. Verwenden Sie die Kalibrator-Konzentrationen A und B, um die Masterkurve an die Reagenzien auf dem System anzupassen. Überprüfen Sie, ob im Reagenzienfach eine Kartusche des IDS-iSYS Intact PINP-Assays vorhanden ist und ob sich die Kartuschen-Masterkurve in

Page 16: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 Februar 2017 Deutsch Seite 4/6

der Datenbank befindet. Wenn die Daten der Kalibratoren-Charge auf dem System nicht vorhanden sind, laden Sie die Daten von der Mini-CD, die zusammen mit dem Reagenzien-Kit bereitgestellt wurde.

Beginnen Sie die Immunassay-Kalibration auf dem System wie im Benutzerhandbuch des Systems beschrieben. Die Kalibration wird mithilfe von drei Replikaten jeder Konzentration durchgeführt. Ein Replikat kann entfernt werden, damit die Kalibrationsanforderungen erfüllt werden. Alle Kontrollkonzentrationen müssen zudem doppelt gemessen werden, um den Assay zu kalibrieren. Verifizieren und bestätigen Sie die Kalibration anhand des Kalibrationsstatus im Kalibrationsfenster und entfernen Sie das verbleibende Volumen der Kalibratoren und Kontrollen nach dem Gebrauch vom Probentablett.

9. Qualitätskontrolle Verwenden Sie das IDS-iSYS Intact PINP Control Set zur Qualitätskontrolle. Um die Richtigkeit der Ergebnisse sicherzustellen, müssen mindestens drei Kontrollen mit unterschiedlichen Konzentrationen an intaktem PINP gemessen werden. Neben dem IDS-iSYS Intact PINP Control Set können auch andere geeignete Kontrollmaterialien verwendet werden. Kontrollen sollten am (oder nahezu am) Anfang jedes Laufs mit Patientenproben und auch während jeder Kalibrierung getestet werden. Dies erfolgt nach den geltenden Bestimmungen oder Akkreditierungsbedingungen und dem Verfahren zur Qualitätskontrolle Ihres Labors. Es wird empfohlen, die Kontrollen routinemäßig doppelt zu testen. Labore sollten die Kontrollen mindestens einmal pro Schicht testen. Anweisungen zur Vorbereitung und Handhabung finden Sie in der Gebrauchsanweisung des IDS-iSYS Intact PINP Control Set. Bestimmung der Probenkonzentrationen von intaktem PINP Verarbeiten Sie die Proben entsprechend dem Benutzerhandbuch des Systems.

10. Berechnung der Ergebnisse Die Konzentration des intakten PINP in jeder Probe wird automatisch berechnet. Die Anzeige der Konzentrationen (auf dem Bildschirm oder als Ausdruck) wird auf Anforderung erstellt. Der IDS-iSYS Intact PINP-Assay nutzt eine logistische Kurvenanpassung mit 4 Parametern (4PL), um die Konzentrationen des intakten PINP zu berechnen.

11. Messbereich Der Messbereich des Assays ist 2–230 ng/mL. Jeder Messwert kleiner als 2 ng/mL muss als „< 2 ng/mL“ angegeben werden.

12. Verdünnung Proben mit Konzentrationen des intakten PINP über dem zulässigen Messbereich sind manuell mit einer Humanserumprobe

niedriger Konzentration im Verhältnis von 1:2 zu verdünnen. Das Ergebnis von verdünnten Proben muss mit dem

Verdünnungsfaktor 2 multipliziert und für die Konzentration der niedrigen Probe korrigiert werden.

13. Nutzungsbeschränkungen - Wie bei jedem Diagnoseverfahren müssen die Ergebnisse in Verbindung mit dem Krankheitsbild des Patienten sowie

anderen dem Arzt vorliegenden Informationen interpretiert werden. - Die Leistungsmerkmale dieses Assays wurden nicht in einer pädiatrischen Population etabliert. - Der Hook-Effekt wurde mit Analytkonzentrationen von bis zu 2350 ng/mL getestet. Es wurde kein Hook-Effekt

beobachtet. - Heterophile Antikörper in Humanseren können mit Immunglobulinen des Reagenz reagieren und somit In-vitro-

Immunassays stören6. Patienten, die regelmäßig Tieren oder Tierserumprodukten ausgesetzt sind, können

gegenüber dieser Interferenz anfällig sein, wodurch unter Umständen anormale Werte zu beobachten sind. - Die folgenden Substanzen interferieren nicht mit dem IDS-iSYS Intact PINP-Assay, wenn die in der nachfolgenden

Tabelle enthaltenen Konzentrationen unter dem angegebenen Grenzwert liegen.

Potenziell interferierende Stoffe

Grenzwert Konzentration

Lipide 2803 mg/dL

Bilirubin 200 mg/dL

Hämoglobin 500 mg/dL

Biotin 300 nmol/L

14. Erwartete Werte Die nachfolgenden Bereiche wurden mithilfe des IDS-iSYS Intact PINP-Assays festgelegt und dienen nur zu Informationszwecken. Die 95%-Referenzintervalle für scheinbar gesunde Erwachsene wurden anhand einer nicht parametrischen Methode gemäß den Richtlinien aus CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory” berechnet.

Normale Erwachsene 27,7–127,6 ng/mL (n = 150) Die oben genannten Bereiche gelten nur als Richtlinien. Es wird empfohlen, dass jedes Labor basierend auf seiner eigenen Patientenpopulation seine eigenen erwarteten Bereiche festlegt.

Page 17: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 Februar 2017 Deutsch Seite 5/6

15. Leistungsdaten Repräsentative Leistungsdaten werden angezeigt. Die in individuellen Labors erhaltenen Ergebnisse können sich unterscheiden. 15.1 Sensitivität Die Leerwertgrenze (LoB), Nachweisgrenze (LoD) und Quantifizierungsgrenze (LoQ) wurden anhand der Richtlinien CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation”, mit 60 Kontrollproben und 120 Proben mit geringen Konzentrationsstufen bestimmt.

Sensitivität Konzentration

LoB 0,5 ng/mL

LoD < 1,0 ng/mL

LoQ < 1,0 ng/mL

15.2 Präzision Die Präzision wurde mithilfe eines modifizierten Protokolls auf Basis von CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”, evaluiert. Insgesamt wurden 3 Serumkontrollen aus einer Reagenziencharge 20 Tage lang einmal am Tag vierfach auf zwei Systemen getestet.

Kontrolle n Mittelwert (ng/mL)

Innerhalb Lauf Gesamt

SD CV SD CV %

1 80 14,3 0,4 2,6 % 0,6 4,2 %

2 80 44,6 1,3 3,0 % 2,4 5,3 %

3 80 100,9 3,0 3,0 % 4,5 4,4 %

15.3 Wiederfindung Die Wiederfindung wurde vor dem Testen durch das Hinzufügen von intaktem PINP zu den Patientenproben bewertet.

Probenkonz. ng/mL

Hinzugef. PINP

ng/mL

Gemess. ng/mL

Wiederfindung ng/mL

Wiederfindung %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Durchschnitt

93 95 93 91 95 91 93

15.4 Linearität Die Linearität wurde gemäß CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”, evaluiert. Proben mit unterschiedlichen Konzentrationen von intaktem PINP wurden in zweifacher Ausführung (Duplikat) getestet. Die so erhaltene mittlere Konzentration wurde mit den erwarteten Konzentrationen verglichen. Die Proben wurden vor dem Assay durch Verdünnen einer Patientenprobe mit hoher Konzentration in einer Patientenprobe mit geringer Konzentration vorbereitet.

Erwartete Konzentration (ng/mL)

Gemessene Konzentration (ng/mL)

Variation

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9

–0,2 –1,6 –2,9 –2,4 3,7 0,0 0,7

– 3 0

–2 –3 –2 2 0 0

15.5 Methodenvergleich Der IDS-iSYS Intact PINP-Assay wurde mit einem handelsüblichen quantitativen Radioimmunassay gemäß CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples” verglichen. Mit jeder Methode wurden insgesamt 363 Proben getestet, die so ausgewählt waren, dass eine große Bandbreite an Konzentrationen von intaktem PINP [8,7–230 ng/mL] repräsentiert war. Mit den Vergleichsdaten wurde eine lineare Regressionsanalyse durchgeführt:

n Steigung 95%-KI Schnittpunkt (ng/mL) 95%-KI Korrelationskoeffizient (r)

363 0,86 0,84 bis 0,88 –1,9 –3,4 bis –0,3 0,98

Page 18: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 Februar 2017 Deutsch Seite 6/6

15.6 Spezifität Die Spezifität wurde anhand der folgenden Analyten bewertet.

Analyt Kreuzreaktivität

Intakt. PINP 100 %

Ratten-/Maus-PINP < 0,01 %

PIINP < 0,03 %

N-Mid Osteocalcin < 0,02 %

16. Verwendete Symbole

REF Katalognummer

IVD In-vitro-Diagnostikum

Hersteller

Gemäß Richtlinie 98/79/EG angewendet

17. Literatur 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., ‘Heterophilic antibodies: a problem for all immunoassays’. Clin Chem, 34, 1988, pp 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tel.: +44 191 519 0660 • Fax: +44 191 519 0760 E-Mail: [email protected] • www.idsplc.com Immunodiagnostic Systems USA Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tel.: +1 877 852 6210 • Fax: +1 301 990 4236 E-Mail: [email protected]

Deutschland Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tel.: +49 692 6019 0960 • Fax: +49 692 6019 0949 E-Mail: [email protected]

Frankreich Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, Frankreich Tel.: +33 140 770 450 • Fax: +33 140 770 455 E-Mail: [email protected]

Skandinavien Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Dänemark Tel.: +45 448 400 91 E-Mail: [email protected]

Belgien Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgien Tel.: +32 425 226 36 • Fax: +32 425 251 96 E-Mail: [email protected]

Brasilien IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasilien CEP: 05422 001 Tel.: +55 113 740 6100 • Fax: + 55 113 740 6105 E-Mail: [email protected]

Page 19: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Istruzioni per l’uso

IS-4000

IS-4000 v06 01 febbraio 2017 Italiano Pagina 1/6

1. Uso previsto Per uso diagnostico in vitro. Il dosaggio IDS-iSYS Intact PINP (propeptide amino-terminale intatto di procollagene di tipo I) è destinato alla determinazione quantitativa del PINP intatto nel siero o nel plasma umano tramite il sistema IDS-iSYS Multi-Discipline Automated System. Il test è destinato a essere utilizzato come strumento ausiliario nella gestione dell’osteoporosi postmenopausale.

2. Summary and Explanation Un passo importante nel processo della formazione ossea è la sintesi del collagene di tipo I, il componente organico maggiormente presente nella matrice ossea. Durante la sintesi del collagene i propeptidi vengono rilasciati da entrambe le porzioni amminoterminali e carbossiterminali della molecola procollagene

1. Tali propeptidi vengono secreti nella circolazione

sanguigna. Il propeptide amminoterminale del procollagene di tipo I (PINP) è forse il marker più specifico e sensibile di formazione ossea

2. Il PINP è un marker particolarmente utile per controllare l'efficacia della terapia per l'osteoporosi con agenti

anabolici3,4

, ma è anche uno dei migliori marker di turnover osseo per il controllo dell'efficacia della terapia anti-riassorbimento

4,5.

3. Descrizione del metodo Il dosaggio è basato sulla tecnologia della chemiluminescenza. 20 µL di campioni del paziente o calibratori vengono disciolti in un diluente, incubandone una parte con un anticorpo monoclonale anti-PINP biotinilato, un anticorpo monoclonale marcato con acridinio, particelle magnetiche marcate con streptavidina e un tampone del dosaggio. Le particelle magnetiche vengono catturate tramite un magnete e un lavaggio eseguito per rimuovere qualsiasi analita non legato. Dopo l’aggiunta dei reattivi trigger, si legge la luminescenza emessa dalla marcatura con acridinio, che è direttamente proporzionale alla concentrazione del PINP intatto nel campione originale.

4. Avvertenze e precauzioni Il kit IDS-iSYS Intact PINP è realizzato esclusivamente per uso diagnostico in vitro e non per uso interno in persone o animali. Questo prodotto deve essere utilizzato rigorosamente in ottemperanza alle indicazioni riportate nelle presenti istruzioni per l’uso. Immunodiagnostic Systems Limited (IDS) non è responsabile per eventuali danni o perdite (tranne quanto stabilito dalla legge) di qualsiasi origine, provocate dalla mancata aderenza alle istruzioni fornite. ATTENZIONE: il kit contiene materiale di origine animale. Manipolare i reagenti del kit come potenziali vettori di agenti infettivi. È necessario seguire le opportune precauzioni e le corrette prassi di laboratorio nella conservazione, nella manipolazione e nello smaltimento dei reagenti del kit. Lo smaltimento dei reagenti del kit deve essere eseguito nel rispetto delle normative locali vigenti. Materiali di origine umana Il materiale di origine umana usato nella preparazione del kit è stato dosato e trovato negativo con metodi raccomandati da FDA per la presenza di anticorpi diretti contro i virus dell’Immunodeficienza Umana (HIV I e HIV II), l’antigene di superficie del virus dell’epatite B e gli anticorpi diretti contro il virus dell’epatite C. Poiché nessun metodo può garantire l’assenza di questi o altri agenti infettivi, manipolare tutti i reattivi secondo le norme di Biosicurezza di livello 2. Reattivi contenenti sodio azide

I calibratori CAL contengono azoturo di sodio (NaN3) > 0,1% (p/p) (< 1%).

Classificazione CLP: Acute Tox. 4 Aquatic Chronic 3 Indicazioni di pericolo: EUH032: A contatto con acidi libera gas molto tossici. H302: Nocivo se ingerito. H412: Nocivo per gli organismi acquatici con effetti di

lunga durata.

Consigli di prudenza: P264: Lavare accuratamente mani dopo l'uso. P270: Non mangiare, né bere, né fumare durante l'uso. P273: Non disperdere nell'ambiente. P301+310: IN CASO DI INGESTIONE: contattare

immediatamente un CENTRO ANTIVELENI o un medico

P330: Sciacquare la bocca.

5. Precauzioni per la manipolazione dei reagenti Fatta eccezione per i calibratori, che sono liofilizzati, i reagenti contenuti nel kit sono pronti all’uso. Per la metodologia di ricostituzione, vedere la sezione della procedura relativa ai Calibratori. Prima di caricare una nuova cartuccia nel sistema IDS-iSYS Multi-Discipline Automated System (sistema), l’operatore deve agitare manualmente per rotazione il contenitore delle particelle magnetiche, per risospendere le particelle che si sono depositate sul fondo del contenitore durante la conservazione. Fare attenzione a evitare la formazione di schiuma nella cartuccia dei reattivi.

Page 20: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febbraio 2017 Italiano Pagina 2/6

Durata e conservazione dei reagenti Conservare la cartuccia e i calibratori in posizione verticale a 2–8 °C al buio. Non congelare la cartuccia e i calibratori.

Durata del reagente Cartuccia Calibratori

Prima dell’apertura a 2–8 °C Fino alla data di scadenza

Cartuccia Dopo l’apertura a 2–8 °C

28 giorni NA

Calibratori Dopo la ricostituzione a 2–8 °C

NA 28 giorni

A bordo dell’analizzatore* 7 giorni 3 ore

* Stabilità per conservazione continuativa a bordo

6. Raccolta e conservazione dei campioni Il dosaggio deve essere effettuato su campioni di siero (provette di campionamento standard o provette contenenti gel per la separazione del siero) o plasma (litio eparina, sodio eparina, ammonio eparina o EDTA di potassio (K2)). Separare il plasma dalla parte corpuscolata nel più breve tempo possibile. Conservare i campioni a -20 °C o a temperature inferiori. Evitare cicli ripetuti di congelamento-scongelamento dei campioni. Note:

- I campioni contenenti particolato devono essere centrifugati prima dell’esecuzione del dosaggio. Non utilizzare campioni inattivati con il calore.

- Per ridurre al minimo i possibili effetti di evaporazione, i campioni, i calibratori e i controlli devono essere misurati entro 3 ore dal caricamento nel sistema.

- Ogni laboratorio deve seguire le linee guida o le disposizioni delle normative locali, regionali e/o statali o delle organizzazioni di accreditamento per definire la stabilità di manipolazione e di conservazione dei propri campioni. Per informazioni sulle prassi più appropriate, fare riferimento alla procedura CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Prima di effettuare l’analisi, assicurarsi che campioni, calibratori e controlli siano a temperatura ambiente (18–22 °C). - In determinati casi, alcune provette per la raccolta dei campioni disponibili in commercio potrebbero influire sui

risultati dei test. - Si raccomanda di seguire le istruzioni del produttore delle provette, in particolare quando si analizzano i campioni in

provette primarie. - I campioni centrifugati che presentano uno strato lipidico sulla superficie devono essere travasati in una coppetta

portacampione o in una provetta secondaria. Fare attenzione a travasare solo i campioni chiarificati, senza la componente lipemica.

7. Materiali Materiali forniti Cartuccia reattivi

MPS3 Particelle magnetiche rivestite con streptavidina in tampone BSA fosfato con azoturo di sodio come conservante

(< 0,1%). 1 flacone, 2,3 mL

CONJ L’anticorpo monoclonale murino anti-PINP marcato con un derivato di estere di acridinio, in tampone contenente

proteine bovine e murine con azoturo di sodio come conservante (< 0,1%), 1 flacone, 4,95 mL.

Ab-BIO Anticorpo monoclonale murino anti-PINP marcato con biotina, in tampone contenente proteine bovine e murine con

azoturo di sodio come conservante (< 0,1%), 1 flacone, 16,8 mL.

BUF Tampone contenente proteine bovine e murine, con azoturo di sodio come conservante (< 0,1%), 1 flacone,

4,95 mL.

DIL Tampone contenente azoturo di sodio come conservante (< 0,1%), 1 flacone, 23,7 mL.

Calibratori

CALA Tampone liofilizzato contenente BSA e PINP umano purificato con < 1,0% di azoturo di sodio come conservante

CALB (< 0,05% ricostituito). 2 flaconi per 2 livelli di concentrazione, 1,0 mL ciascuno.

I calibratori sono preparati con i seguenti intervalli nominali:

Calibratore Intervalloa (ng/mL)

CALA 3,0–7,0 ng/mL

CALB 130,0–165,0 ng/mL

a Gli intervalli sono relativi al prodotto e sono forniti solo a titolo indicativo. I valori possono variare in base al lotto. Fare

riferimento al Certificato di analisi (CoA) sul mini CD per le concentrazioni specifiche per i lotti. Gli intervalli del calibratore sono stabiliti mediante analisi dei campioni del calibratore in almeno 20 dosaggi su un unico sistema. Mini CD Contiene le istruzioni per l’uso per i reagenti IDS-iSYS, l’intervallo di valori dei controlli, i file CRY e il Certificato di analisi.

Page 21: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febbraio 2017 Italiano Pagina 3/6

Materiali richiesti ma non forniti Analizzatori Attrezzature/materiali opzionali IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Procedura di dosaggio Cartuccia reattivi I reagenti forniti nella cartuccia sono pronti all’uso, a eccezione dei calibratori. Prima di introdurre una nuova cartuccia nel sistema, agitare manualmente per rotazione il contenitore delle particelle magnetiche. Evitare la formazione di schiuma. Non appena la cartuccia viene inserita nel vassoio dei reattivi, viene letto il barcode. Se il lettore di codici a barre del sistema non riesce a leggere l’etichetta, esiste una procedura manuale per inserire i dati del codice a barre (vedere il manuale per l’uso del sistema IDS-iSYS). Il sistema esegue automaticamente la miscelazione delle particelle magnetiche per tutte le cartucce di reagenti caricate a bordo dello strumento per preservarne l’omogeneità. La cartuccia deve essere caricata nel vassoio dei reattivi almeno 40 minuti prima dell’avvio del dosaggio. Se si rimuove cartuccia dal vassoio dei reattivi, conservarla in posizione verticale a 2–8 °C al buio. Calibratori I calibratori di questo kit sono tracciabili sui calibratori in funzione del produttore preparati con PINP umano altamente purificato, il cui valore è tracciabile con un metodo disponibile in commercio. I calibratori del dosaggio IDS-iSYS Intact PINP sono liofilizzati. Ricostituire immediatamente prima dell’uso. Aggiungere 1,0 mL di acqua distillata o deionizzata a ogni flacone. Riposizionare il tappo. Lasciare ricostituire per 10 minuti miscelando di tanto in tanto manualmente con delicatezza. Fare attenzione a evitare la formazione di schiuma. I calibratori possono essere caricati nell’analizzatore utilizzando le coppette portacampione o i rack per calibratori/controlli (30/31/32). Rack per calibratori/controlli: posizionare i flaconi dei calibratori nell’apposito rack, inserire quest’ultimo nel sistema e procedere secondo le istruzioni del manuale per l’uso del sistema. Eliminare il materiale dei flaconi dopo l’uso. NON riutilizzare i flaconi dei calibratori inseriti nell’analizzatore. Coppette portacampione: Pipettare circa 200 μL di calibratore nelle coppette portacampione e inserirle nell’analizzatore. Seguire le istruzioni del manuale per l’uso del sistema. Eliminare il materiale contenuto nelle coppette portacampione dopo l’uso. Non ri-inserire i calibratori avanzati nei flaconi originali. Se i calibratori devono essere utilizzati più di una volta, conservarli a 2–8 °C entro 3 ore dalla ricostituzione. Per riutilizzarli, portare i flaconi dei calibratori conservati a temperatura ambiente (18–22 °C) e agitarli delicatamente per inversione prima dell’uso. Frequenza di calibrazione Viene richiesta una nuova calibrazione:

- Quando si inserisce un nuovo lotto di cartucce. - Quando i valori dei controlli non rientrano negli intervalli di riferimento. - Alla scadenza dell’intervallo di calibrazione di 3 giorni. - Dopo l’assistenza al sistema.

La verifica della calibrazione è automatica ed è gestita dal sistema. Calibrazione dell’analizzatore I due calibratori del dosaggio IDS-iSYS Intact PINP sono necessari per effettuare l’adattamento della curva master. Utilizzare sempre i calibratori forniti con il kit; non è possibile utilizzare calibratori di un lotto diverso. Tutti i livelli del sistema IDS-iSYS Intact PINP Control Set (IS-4030) DEVONO essere misurati contemporaneamente ai calibratori perché l’adattamento della curva master venga ritenuto valido. Tutti i dati richiesti per la calibrazione della cartuccia sono presenti sul mini CD. Utilizzare i calibratori A e B per allineare la curva master ai reagenti presenti a bordo del sistema. Verificare la presenza di una cartuccia del dosaggio IDS-iSYS Intact PINP nel contenitore dei reagenti e la disponibilità della curva master relativa alla cartuccia nel database. Se i dati riferiti al lotto dei calibratori non sono presenti a bordo del sistema, caricare i dati dal mini CD fornito con il kit di reagenti. Avviare la calibrazione degli immunodosaggi sul sistema in base al manuale per l’utente del sistema. La calibrazione viene eseguita utilizzando 3 replicati di ciascun livello. Un valore può essere rimosso per soddisfare i requisiti di calibrazione. Per calibrare il dosaggio, tutti i livelli dei controlli devono inoltre essere misurati in duplicato. Verificare e approvare la calibrazione in base allo stato visualizzato nella finestra di calibrazione e rimuovere il volume rimanente di calibratori e controlli dal vassoio dei campioni dopo l’utilizzo.

Page 22: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febbraio 2017 Italiano Pagina 4/6

9. Controllo di qualità Usare il set IDS-iSYS Intact PINP Control Set per il controllo di qualità. Per garantire la validità dei risultati, devono essere misurati almeno tre controlli a diversi livelli di PINP intatto. Oltre al set IDS-iSYS Intact PINP Control Set possono essere utilizzati altri materiali di controllo idonei. I controlli devono essere analizzati all’inizio (o poco prima) di ciascun test contenente campioni di pazienti e durante ogni calibrazione, in conformità alle normative locali, regionali e/o statali o ai requisiti di accreditamento e alla procedura di qualità del proprio laboratorio. Si raccomanda di dosare i controlli sempre in duplicato. I laboratori devono testare i controlli almeno una volta per ogni turno. Per la preparazione e la manipolazione, fare riferimento alle istruzioni per l’uso del set IDS-iSYS Intact PINP Control Set. Determinazione dei livelli di PINP intatto nel campione Dosare i campioni secondo le istruzioni contenute nel manuale per l’uso del sistema.

10. Calcolo dei risultati La concentrazione di PINP intatto di ogni campione viene calcolata automaticamente. La visualizzazione delle concentrazioni (su schermo o stampata) è disponibile su richiesta dell’operatore. Il dosaggio IDS-iSYS Intact PINP utilizza una curva logistica a 4 parametri (4PL) per calcolare le concentrazioni di PINP intatto.

11. Range della misurazione (range refertabile) L’intervallo di misurazione refertabile del dosaggio è 2–230 ng/mL. Qualsiasi lettura di valori inferiori a 2 ng/mL deve essere

indicata come “< 2 ng/mL”.

12. Diluizione I campioni con concentrazioni di PINP intatto superiori all’intervallo refertabile devono essere diluiti manualmente insieme a un campione di siero umano a bassa concentrazione in un rapporto di 1 a 2. I risultati dei campioni diluiti devono essere moltiplicati per il fattore di diluizione 2 e corretti per la concentrazione del campione a bassi livelli.

13. Limitazioni d’uso - Come nel caso di qualsiasi procedura diagnostica, i risultati devono essere interpretati unitamente ai dati clinici del

paziente e ad altre informazioni a disposizione del medico. - Non sono state stabilite le caratteristiche di azione di questo dosaggio nella popolazione pediatrica. - L’effetto hook è stato analizzato usando concentrazioni di analita fino a 2350 ng/mL. Non è stato osservato alcun

effetto uncino. - Gli anticorpi eterofili nel siero umano possono reagire con le immunoglobuline dei reagenti, interferendo con gli

immunodosaggi in vitro6. I pazienti regolarmente esposti agli animali o a prodotti derivati da siero animali possono

essere soggetti a questa interferenza, quindi si potrebbero osservare valori anomali. - Le seguenti sostanze non interferiscono con il dosaggio IDS-iSYS intact PINP quando le concentrazioni riportate

nella seguente tabella sono inferiori alla soglia indicata.

Agente potenzialmente interferente

Concentrazione di soglia

Lipide 2.803 mg/dL

Bilirubina 200 mg/dL

Emoglobina 500 mg/dL

Biotina 300 nmol/L

14. Valori attesi I seguenti intervalli sono stati stabiliti utilizzando il dosaggio IDS-iSYS Intact PINP e vengono forniti esclusivamente a scopo informativo. L’intervallo di confidenza al 95% per gli adulti apparentemente sani è stato calcolato con un metodo non parametrico secondo le linee guida CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Adulti normali 27,7–127,6 ng/mL (n = 150) Gli intervalli sopra indicati devono essere considerati solo come linee guida; si raccomanda a ogni laboratorio di stabilire gli intervalli attesi sulla base della propria popolazione di pazienti.

Page 23: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febbraio 2017 Italiano Pagina 5/6

15. Dati di prestazione Sono mostrati dati di prestazione rappresentativi. I risultati ottenuti nei singoli laboratori possono variare. 15.1 Sensibilità La sensibilità analitica (limit of blank, LoB), la dose minima rilevabile (limit of detection, LoD) e il limite di quantificazione (limit of quantitation, LoQ) sono stati determinati secondo le linee guida CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation” utilizzando 60 replicati del blank e 120 campioni a bassi livelli.

Sensibilità Concentrazione

LoB 0,5 ng/mL

LoD < 1,0 ng/mL

LoQ < 1,0 ng/mL

15.2 Precisione La precisione è stata valutata secondo un protocollo modificato basato sulle linee guida CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”. Un totale di 3 controlli di siero sono stati dosati utilizzando un lotto di reagenti in quadruplicato, uno al giorno per 20 giorni in due sistemi.

Controllo n Media

(ng/mL) Intra-test Totale

DS CV DS CV%

1 80 14,3 0,4 2,6% 0,6 4,2%

2 80 44,6 1,3 3,0% 2,4 5,3%

3 80 100,9 3,0 3,0% 4,5 4,4%

15.3 Recupero Il recupero è stato valutato aggiungendo PINP intatto ai campioni prima di eseguire il dosaggio.

Conc. campione

ng/mL

PINP aggiunto ng/mL

Misurata ng/mL

Recupero ng/mL

Recupero %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Media

93 95 93 91 95 91 93

15.4 Linearità La linearità è stata valutata secondo le linee guida CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”. Sono stati analizzati in duplicato campioni contenenti varie concentrazioni di PINP intatto. Le concentrazioni medie risultanti sono state confrontate con le concentrazioni previste. I campioni sono stati preparati diluendo un campione di paziente alto con un campione di paziente basso prima del dosaggio.

Concentrazione prevista (ng/mL)

Concentrazione misurata (ng/mL)

Variazione

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9 -0,2 -1,6 -2,9 -2,4 3,7 0,0 0,7

- 3 0 -2 -3 -2 2 0 0

15.5 Confronto tra metodi Il dosaggio IDS-iSYS nome dosaggio è stato confrontato con un radioimmunodosaggio quantitativo disponibile in commercio, secondo CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. Un totale di 363 campioni, selezionati per rappresentare un vasto intervallo di concentrazioni di PINP intatto [8,7–230 ng/mL], è stato dosato con entrambi i metodi. L’analisi di regressione lineare è stata effettuata su dati comparativi:

n Pendenza I. C. 95% Intercetta (ng/mL)

I. C. 95% Coefficiente di correlazione (r)

363 0,86 Da 0,84 a 0,88 -1,9 Da -3,4 a -0,3 0,98

Page 24: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febbraio 2017 Italiano Pagina 6/6

15.6 Specificità La specificità è stata valutata con i seguenti analiti.

Analita Reattività crociata

PINP intatto 100%

PINP di ratto/topo < 0,01%

PINP < 0,03%

Osteocalcina N-Mid < 0,02%

16. Simboli utilizzati

REF Numero di catalogo

IVD Dispositivo diagnostico in vitro

Produttore

Applicato in conformità alla Direttiva 98/79/CE

17. Bibliografia 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., ‘Heterophilic antibodies: a problem for all immunoassays’. Clin Chem, 34, 1988, pp 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tel.: +44 191 519 0660 • Fax: +44 191 519 0760 e-mail: [email protected] • www.idsplc.com Immunodiagnostic Systems USA Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tel.: +1 877 852 6210 • Fax: +1 301 990 4236 e-mail: [email protected]

Germania Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tel.: +49 692 6019 0960 • Fax: +49 692 6019 0949 e-mail: [email protected]

Francia Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tel.: +33 140 770 450 • Fax: +33 140 770 455 e-mail: [email protected]

Scandinavia Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tel.: +45 448 400 91 e-mail: [email protected]

Belgio Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tel.: +32 425 226 36 • Fax: +32 425 251 96 e-mail: [email protected]

Brasile IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasil. CEP: 05422 001 Tel.: +55 113 740 6100 • Fax: + 55 113 740 6105 e-mail: [email protected]

Page 25: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Instrucciones de uso

IS-4000

IS-4000 v06 01 febrero de 2017 Español Página 1/6

1. Uso previsto Para uso en diagnóstico in vitro. El ensayo IDS-iSYS Intact PINP (propéptido aminoterminal intacto del procolágeno de tipo I) está destinado a la determinación cuantitativa del PINP en suero humano o plasma en IDS-iSYS Multi-Discipline Automated System. La prueba se ha diseñado como complemento para el tratamiento de la osteoporosis posmenopáusica.

2. Summary and Explanation Un paso importante en el proceso de formación de hueso es la síntesis del colágeno de tipo I, que es el principal componente orgánico de la matriz ósea. Durante la síntesis de colágeno, se liberan propéptidos de las partes amino y carboxiterminales de la molécula de procolágeno

1. Estos propéptidos se secretan en la circulación sanguínea. El propéptido amino-terminal del

procolágeno de tipo I (PINP) es probablemente el marcador más específico y sensible de la formación ósea2. El PINP es un

marcador especialmente útil para monitorizar la eficacia del tratamiento de la osteoporosis con agentes anabólicos3,4

, pero también es uno de los marcadores de recambio óseo para vigilar la eficacia del tratamiento anti-rresortivo

4,5.

3. Descripción del método Este ensayo se basa en la tecnología de la quimioluminiscencia. Se diluyen 20 µL de las muestras del paciente o los calibradores en un diluyente y una parte de ellos se incuba en un anticuerpo anti-PINP monoclonal biotinilado, un anticuerpo monoclonal marcado con acridinio, partículas magnéticas marcadas con estreptavidina y un tampón de ensayo. Las partículas magnéticas se capturan usando un imán y un paso de lavado realizado para eliminar cualquier analito no ligado. Se añaden reactivos desencadenantes, la luz resultante emitida por la etiqueta de acridinio es directamente proporcional a la concentración del PINP intacto en la muestra original.

4. Advertencias y precauciones El IDS-iSYS Intact PINP está diseñado únicamente para diagnóstico in vitro y no está destinado al uso interno en humanos o animales. Este producto debe emplearse siguiendo rigurosamente las instrucciones descritas en estas instrucciones de uso. Immunodiagnostic Systems Limited (IDS) no se responsabilizará de ninguna pérdida o daño (salvo en los casos requeridos por la ley), que se origine por incumplimiento de las instrucciones indicadas. PRECAUCIÓN: Este kit contiene material de origen animal. Manipule los reactivos del kit como potenciales transmisores de agentes infecciosos. Se deben seguir las precauciones pertinentes y emplear las prácticas de laboratorio recomendadas durante el almacenamiento, la manipulación y la eliminación de los reactivos del kit. La eliminación de los reactivos del kit debe realizarse siguiendo las normativas locales.

Materiales humanos El material utilizado a la hora de preparar este producto se ha analizado mediante pruebas recomendadas por la FDA para detectar la posible presencia de anticuerpos del virus de la inmunodeficiencia humana (VIH I y II), del antígeno de superficie del virus de la hepatitis B y de anticuerpos del virus de la hepatitis C, con resultados negativos. Dado que ninguna prueba puede ofrecer garantías plenas de la ausencia de agentes infecciosos, los reactivos deben manipularse de acuerdo al Nivel de bioseguridad 2. Reactivos que contienen azida sódica

Los calibradores CAL contienen azida de sodio (NaN3) > 0,1 % (p/p) (< 1 %).

Clasificación según el reglamento CLP: Acute Tox. 4 Aquatic Chronic 3 Indicaciones de peligro: EUH032: En contacto con ácidos libera gases muy tóxicos. H302: Nocivo en caso de ingestión. H412: Nocivo para los organismos acuáticos, con

efectos nocivos duraderos.

Consejos de prudencia: P264: Lavarse manos concienzudamente tras la

manipulación. P270: No comer, beber ni fumar durante su utilización. P273: Evitar su liberación al medio ambiente. P301+310: EN CASO DE INGESTIÓN: Llamar

inmediatamente a un CENTRO DE INFORMACIÓN TOXICOLÓGICA o a un médico.

P330: Enjuagarse la boca.

5. Precauciones para la manipulación Además de los calibradores liofilizados, los reactivos que se suministran en el kit están listos para su uso. Consulte la sección del calibrador del procedimiento para obtener la metodología de reconstitución.

Page 26: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febrero de 2017 Español Página 2/6

Antes de cargar un nuevo cartucho en el IDS-iSYS Multi-Discipline Automated System (sistema), el operario debe mezclar el recipiente de las partículas magnéticas con un movimiento enérgico de rotación. Esto resuspenderá las partículas magnéticas que se hayan depositado durante el transporte. Se debe tener precaución para evitar formar espuma en el cartucho de reactivos. Caducidad y almacenamiento de los reactivos Almacene el cartucho y los calibradores en posición vertical en un lugar oscuro a 2 °C a 8 °C. No congele el cartucho ni los calibradores.

Caducidad de reactivos Cartucho Calibradores

Antes de abrir a 2 °C a 8 °C Hasta la fecha de vencimiento

Cartucho después de abrir a entre 2 y 8 °C

28 días ND

Calibradores después de la reconstitución a entre 2 y 8 °C

ND 28 días

Introducido en el sistema * 7 días 3 horas

* Estabilidad continua introducido en el sistema

6. Recogida y almacenamiento de muestras El ensayo debe realizarse usando muestras de suero (tubos de muestras estándar o tubos que contienen gel para la separación de suero) o plasma (heparina de litio, heparina sódica, heparina de amoníaco o EDTA [K2] de potasio). Las muestras deberán separarse lo antes posible después de su recogida. Conserve las muestras a una temperatura de –20 °C o inferior. Evite congelar y descongelar repetidamente las muestras. Nota:

- Las muestras que contienen partículas se deben centrifugar antes de realizar el análisis. No utilice muestras inactivadas por calor.

- Para reducir al mínimo los posibles efectos de evaporación, las muestras, los calibradores y los controles deben medirse en el plazo de 3 horas desde el momento en el que se colocan en el sistema.

- Cada laboratorio debe seguir las directrices o requisitos de las normativas locales, estatales y federales u organizaciones de acreditación, con el fin de establecer su propia manipulación de muestras y estabilidad de almacenamiento. Para obtener orientación sobre prácticas adecuadas, se debe consultar el documento CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Antes de llevar a cabo los ensayos, asegúrese de que las muestras, los calibradores y los controles estén a temperatura ambiente (de 18 °C a 22 °C).

- Algunos tubos de recogida de muestras disponibles en el mercado pueden repercutir en los resultados de las pruebas en casos particulares.

- Se recomienda seguir las instrucciones del fabricante del tubo, especialmente para el procesamiento de muestras en tubos primarios.

- Las muestras centrifugadas con una capa de lípidos en la parte superior deben transferirse a un recipiente para muestras o tubo secundario. Se debe tener precaución para transferir únicamente las muestras clarificadas sin el material lipémico.

7. Materiales Materiales suministrados Cartucho de reactivos

MPS3 Partículas magnéticas recubiertas con estreptavidina en un tampón de fosfato con albúmina sérica bovina con azida

de sodio como conservante (< 0,1 %), 1 frasco, 2,3 mL

CONJ Anticuerpo monoclonal murino anti-PINP marcado con un derivado del éster de acridinio, en tampón que contiene

proteínas bovinas y de ratón, con azida sódica como conservante (< 0,1 %), 1 frasco, 4,95 mL

Ab-BIO Anticuerpo monoclonal murino anti-PINP marcado con biotina, en tampón que contiene proteínas bovinas y de ratón,

con azida sódica como conservante (< 0,1 %) 1 frasco, 16,8 mL.

BUF Tampón que contiene proteínas bovinas y de ratón, con azida de sodio como conservante (< 0,1 %), 1 frasco,

4,95 mL

DIL Tampón con azida de sodio como conservante (< 0,1 %), 1 frasco, 23,7 mL

Calibradores

CALA Tampón liofilizado que contiene BSA y PINP humano purificado con < 1,0 % de azida de sodio como conservante

CALB (< 0,05 % reconstituido). 2 de cada 2 niveles de concentración, 1,0 mL por frasco.

Los calibradores se preparan con los siguientes rangos nominales:

Calibrador Rangoa (ng/mL)

CALA Entre 3,0 y 7,0 ng/mL

CALB Entre 130,0 y 165,0 ng/mL

a Los rangos son habituales para el producto y solo se proporcionan como guía. Los valores pueden variar de un lote a otro.

Consulte el Certificado de análisis (CoA) en el mini CD para conocer las concentraciones específicas del lote.

Page 27: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febrero de 2017 Español Página 3/6

Los rangos del calibrador se establecen mediante pruebas realizadas en las muestras del calibrador en un mínimo de 20 análisis de muestras en un mismo sistema. Mini CD Incluye las instrucciones de uso para los reactivos IDS-iSYS, rangos de control, archivos CRY y certificado de análisis. Materiales necesarios, pero no suministrados Sistema Equipo/materiales opcionales IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Procedimiento del análisis Cartucho de reactivos Los reactivos proporcionados en el cartucho están listos para el uso,salvo los calibradores. Antes de que se cargue un nuevo cartucho en el sistema, mezcle el recipiente de las partículas magnéticas con un movimiento enérgico de rotación. Evite la formación de espuma. El código de barras se lee cuando el cartucho se carga en la bandeja de reactivos. Si el lector de códigos de barras del sistema no puede leer la etiqueta, existe un procedimiento manual para introducir los datos del código de barras (consulte el manual del usuario del IDS-iSYS). El sistema realiza automáticamente la mezcla de las partículas magnéticas de todos los cartuchos de reactivos introducidos para mantener la homogeneidad. El cartucho debe cargarse en la bandeja de reactivos al menos 40 minutos antes de comenzar el ensayo. Si se retira el cartucho de la bandeja de reactivos, conserve el cartucho verticalmente a entre 2 y 8 °C en la oscuridad. Calibradores Los calibradores de este kit se pueden trazar hasta los calibradores activos del fabricante que se preparan con el PINP humano altamente purificado, cuyo valor se puede trazar hasta el método disponible en el mercado. Los calibradores del ensayo IDS-iSYS Intact PINP están liofilizados. Reconstituya inmediatamente después de su uso. Añada 1,0 mL de agua destilada o desionizada en un frasco de cada nivel. Sustituya el llenador. Deje durante 10 minutos para reconstituir mezclando manualmente por momentos de manera suave. Se debe tener precaución para evitar la formación de espuma. Los calibradores pueden cargarse en el Sistema utilizando recipientes para muestras o gradillas para calibradores/controles (30/31/32). Gradillas para calibradores/controles: coloque los viales del calibrador en la gradilla adecuada e insértela en el sistema y continúe de acuerdo con las instrucciones del Manual del usuario del sistema. Deseche el material en los viales del calibrador después de su uso. NO reutilice los viales del calibrador que se han insertado en el sistema. Recipientes para muestra: Pipetee aproximadamente 200 μL de calibradores en los recipientes para muestras y colóquelos en el sistema. Continúe de acuerdo con las instrucciones incluidas en el Manual del usuario del sistema. Deseche el material en los recipientes para muestras después de su uso. NO devuelva el material al vial del calibrador. Si los calibradores deben usarse más de una vez, se deben almacenar a entre 2 y 8 °C en las 3 horas siguientes a la reconstitución. Al reutilizar los viales del calibrador almacenados, ponga los calibradores a temperatura ambiente (entre 18 °C y 22 °C) e inviértalos suavemente antes de su uso. Frecuencia de calibración Se necesita una nueva calibración:

- Tras cargar cada nuevo lote de cartuchos. - Cuando los valores control no estén dentro de los intervalos definidos. - Cuando ha transcurrido el intervalo de calibración de 3 días. - Después del mantenimiento del sistema.

La comprobación de la calibración es automática y está dirigida por el sistema. Calibración del sistema Se necesitan los dos calibradores del ensayo IDS-iSYS Intact PINP para realizar el ajuste de la curva principal. Los calibradores se suministran con el kit y no deben usarse calibradores de otro lote. Todos los niveles del IDS-iSYS Intact PINP Control Set (IS-4030) SE DEBEN medir al mismo tiempo que los calibradores para validar el ajuste de la curva principal. Todos los datos necesarios para la calibración del cartucho pueden encontrarse en el mini CD. Use los niveles A y B de calibradores para ajustar la curva principal en los reactivos integrados en el sistema. Compruebe la presencia de un cartucho del ensayo IDS-iSYS Intact PINP en la bandeja de reactivos y la disponibilidad de la curva maestra del cartucho en la base de

Page 28: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febrero de 2017 Español Página 4/6

datos. Si no se dispone de los datos del lote de los calibradores dentro del sistema, cargue los datos usando el mini CD facilitado con el kit de reactivos. Comience la calibración del inmunoensayo en el sistema de acuerdo con el Manual del usuario del sistema. La calibración se lleva a cabo usando 3 réplicas de cada nivel. Se puede eliminar una réplica para cumplir los requisitos de calibración. También deben medirse todos los niveles de control por duplicado para calibrar el ensayo. Verifique y apruebe la calibración de acuerdo con el estado de calibración que se muestra en la ventana de calibración y deseche el volumen restante de calibradores y controles de la bandeja de muestras después de su uso.

9. Control de calidad Utilice el IDS-iSYS Intact PINP Control Set para el control de calidad. Para garantizar la validez de los resultados, se deben medir al menos tres controles con diferentes niveles de PINP intacto. Se puede usar otro material de control adecuado además del IDS-iSYS Intact PINP Control Set. Los controles deben probarse al comienzo de cada análisis, o prácticamente en cada análisis, que incluya muestras de pacientes y también durante cada calibración de acuerdo con las regulaciones locales, estatales o federales o los requisitos de acreditación y el procedimiento de calidad de su laboratorio. Se recomienda que los controles se evalúen regularmente por duplicado. Los laboratorios deben probar los controles al menos una vez por cada turno. Consulte las instrucciones de preparación y manipulación en el manual del usuario del IDS-iSYS Intact PINP Control Set. Determinación de los niveles de PINP intacto de las muestras Procese las muestras de acuerdo con el Manual del usuario del sistema.

10. Cálculo de los resultados La concentración de PINP intacto de cada muestra se calcula de forma automática. La visualización de las concentraciones (en pantalla o impresas) se produce si el usuario la solicita. El ensayo IDS-iSYS Intact PINP emplea un ajuste de curva logística de 4 parámetros (4PL) para calcular las concentraciones de PINP intacto.

11. Rango de medición (rango notificable) El rango notificable del ensayo es de 2 a 230 ng/mL. Cualquier valor que esté por debajo de 2 ng/mL debe notificarse como

“< 2 ng/mL”.

12. Dilución Las muestras con concentraciones de PINP intacto superiores al rango notificable deben disolverse de forma manual con una muestra de suero humano de baja concentración en una relación de 1 a 2. Los resultados de las muestras diluidas deben multiplicarse por el factor de dilución 2 y corregirse para la concentración de la muestra baja.

13. Límites de uso - Como en cualquier procedimiento diagnóstico, los resultados se deben interpretar junto con los hallazgos clínicos del

paciente y otra información de la que el médico disponga. - Las características de rendimiento de este análisis no se han establecido para una población pediátrica. - El efecto gancho se analizó utilizando concentraciones de analito de hasta 2350 ng/mL. No se observó efecto

gancho. - Los anticuerpos heterofílicos en el suero humano pueden presentar reacciones con las inmunoglobulinas reactivas,

que interfieren con los inmunoensayos in vitro6. Los pacientes que se exponen habitualmente a animales o a

productos de suero animal, pueden ser propensos a esta interferencia y puede que se observen valores anómalos. - Las siguientes sustancias no interfieren con el ensayo IDS-iSYS intact PINP si las concentraciones que se presentan

en la siguiente tabla son inferiores al umbral determinado.

Agentes que pueden interferir

Umbral de Concentración

Lípido 2803 mg/dL

Bilirrubina 200 mg/dL

Hemoglobina 500 mg/dL

Biotina 300 nmol/L

14. Valores esperados Los siguientes rangos se determinaron utilizando el ensayo IDS-iSYS Intact PINP y se proporcionan solo con fines informativos. El intervalo de referencia del 95 % para adultos aparentemente sanos se calculó usando un método no paramétrico conforme a las directrices de CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Adultos normales 27,7-127,6 ng/mL (n = 150) Los rangos anteriores deben considerarse directrices solamente; se recomienda que cada laboratorio establezca su propio rango previsto en función de su propia población de pacientes.

Page 29: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febrero de 2017 Español Página 5/6

15. Datos de rendimiento Se muestran los datos de rendimiento representativos. Los resultados obtenidos en diferentes laboratorios pueden diferir. 15.1 Sensibilidad El límite del blanco (LoB), límite de detección (LoD) y el límite de cuantificación (LoQ) se determinaron siguiendo las directrices de CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation” usando 60 blancos y 120 muestras de bajo nivel.

Sensibilidad Concentración

LoB 0,5 ng/mL

LoD < 1,0 ng/mL

LoQ < 1,0 ng/mL

15.2 Precisión La precisión se evalúo de acuerdo con el protocolo modificado basado en CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”. Se sometieron a ensayos numerosos 3 controles de suero usando un lote de reactivos por cuadriplicado una vez por día durante 20 días en dos sistemas.

Control n Media

(ng/mL) Intraprueba Total

SD CV SD CV%

1 80 14,3 0,4 2,6 % 0,6 4,2 %

2 80 44,6 1,3 3,0 % 2,4 5,3 %

3 80 100,9 3,0 3,0 % 4,5 4,4 %

15.3 Recuperación La recuperación se analizó añadiendo PINP intacto a las muestras antes del análisis.

Conc. de muestra ng/mL

PINP añadido ng/mL

Medido ng/mL

Recuperación ng/mL

Recuperación %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Media

93 95 93 91 95 91 93

15.4 Linealidad La linealidad se evaluó de acuerdo con CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”. Las muestras con concentraciones cambiantes de PINP intacto se analizaron por duplicado. Las concentraciones medias resultantes se compararon con las concentraciones predichas. Las muestras se prepararon mediante la dilución de una muestra alta de paciente con una baja antes del análisis.

Concentración prevista (ng/mL)

Concentración medida (ng/mL)

Variación

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9

–0,2 –1,6 –2,9 –2,4 3,7 0,0 0,7

– 3 0

–2 –3 –2 2 0 0

15.5 Comparación de métodos El ensayo IDS-iSYS se comparó con un radioinmunoanálisis cuantitativo disponible en el mercado, según CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. Se analizaron un total de 363 muestras, seleccionadas para representar un amplio rango de concentraciones dePINP intacto [8,7-230 ng/mL], usando cada uno de los métodos. Se realizó un análisis de regresión lineal sobre los datos comparativos:

n Pendiente IC del 95 % Intersección

(ng/mL) IC del 95 %

Coeficiente de correlación (r)

363 0,86 Entre 0,84 y 0,88 –1,9 Entre –3,4 y –0,3 0,98

Page 30: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 febrero de 2017 Español Página 6/6

15.6 Especificidad La especificidad se evaluó con los siguientes analitos.

Analito Reactividad cruzada

PINP intacto 100 %

PINP de rata/ratón < 0,01 %

PIINP < 0,03 %

N-Mid Osteocalcin < 0,02 %

16. Símbolos empleados

REF Número del catálogo

IVD Dispositivo para diagnóstico in vitro

Fabricante

Aplicado de acuerdo con la directiva 98/79/CE

17. Bibliografía 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., ‘Heterophilic antibodies: a problem for all immunoassays’. Clin Chem, 34, 1988, pp 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tel.: +44 191 519 0660 • Fax: +44 191 519 0760 Correo electrónico: [email protected] • www.idsplc.com Immunodiagnostic Systems Estados Unidos Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tel.: +1 877 852 6210 • Fax: +1 301 990 4236 Correo electrónico: [email protected]

Alemania Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tel.: +49 692 6019 0960 • Fax: +49 692 6019 0949 Correo electrónico: [email protected]

Francia Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tel.: +33 140 770 450 • Fax: +33 140 770 455 Correo electrónico: [email protected]

Escandinavia Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tel.: +45 448 400 91 Correo electrónico: [email protected]

Bélgica Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tel.: +32 425 226 36 • Fax: +32 425 251 96 Correo electrónico: [email protected]

Brasil IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasil. CEP: 05422 001 Tel.: +55 113 740 6100 • Fax: + 55 113 740 6105 Correo electrónico: [email protected]

Page 31: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Brugsanvisning

IS-4000

IS-4000 v06 01 februar 2017 Dansk Side 1/6

1. Beregnet brug Til in vitro-diagnostisk brug. IDS-iSYS Intact PINP-analysen (intakt amino-terminal propeptid af type I procollagen) er beregnet til kvantitativ bestemmelse af intakt PINP i humant serum eller plasma på IDS-iSYS Multi-Discipline Automated System. Testen er beregnet til brug som en hjælp i forbindelse med behandlingen af postklimakteriel osteoporose.

2. Summary and Explanation Et vigtigt trin i knogledannelsesprocessen er syntese af type I collagen, der er den største organiske komponent i knoglematrix. Under collagensyntesen frigives propeptider fra både amino- og carboxyterminaldele af procollagen molekylet

1.

Disse propeptider afgives til blodbanen. Amino-terminal propeptid af type I procollagen (PINP) er sandsynligvis den mest specifikke og sensitive markør for knogledannelse

2. PINP er en særlig nyttig markør til monitorering af effektiviteten af

osteoporosebehandling med anaboliske agenser3,4

, men er også en af de bedste knogleomsætningsmarkører til monitorering af effektiviteten af anti-resorptiv behandling

4,5.

3. Metodebeskrivelse Analysen er baseret på kemiluminescensteknologi. 20 µL patientprøve eller kalibrator fortyndes med et fortyndingsmiddel, hvorefter en del deraf inkuberes med et biotinyleret anti-PINP monoklonalt antistof, et acridiniummærket monoklonalt antistof, streptavidinmærkede magnetiske partikler og en analysebuffer. De magnetiske partikler opfanges ved at bruge en magnet og en vaskefase, der udføres for at fjerne eventuelle ubundne analytter. Der tilsættes trigger-reagenser. Det resulterende lys, der udsendes af acridinium-mærket er ligefremt proportionalt med koncentrationen af intakt PINP i den oprindelige prøve.

4. Advarsler og forholdsregler IDS-iSYS Intact PINP er udelukkende beregnet til in vitro diagnostisk brug og er ikke beregnet til indvortes brug hos hverken mennesker eller dyr. Dette produkt skal bruges i nøje overensstemmelse med anvisningerne i denne brugsanvisning. Immunodiagnostic Systems Limited (IDS) kan ikke holdes ansvarlig for noget tab eller nogen skader (medmindre dette er bestemt ved lov), uanset hvordan disse er opstået, hvis de skyldes manglende overholdelse af den udleverede brugervejledning. FORSIGTIG: Dette sæt indeholder materiale af animalsk oprindelse. Håndter sættets reagenser, som om de er i stand til at overføre et smitsomt stof. Passende forholdsregler og god laboratoriepraksis skal overholdes, hvad angår opbevaring, håndtering og bortskaffelse af sættets reagenser. Bortskaffelse af sættets reagenser skal foregå i overensstemmelse med lokale bestemmelser.

Humane materialer Humant materiale anvendt i fremstillingen af dette produkt er testet i analyser anbefalet af FDA i henseende til forekomst af antistoffer mod human immundefekt virus (HIV I og II), hepatitis B overfladeantigen samt antistof mod hepatitis C, og er fundet negativ. Eftersom ingen test kan give fuld sikkerhed for, at der ikke findes smitstoffer, skal reagenserne behandles i henhold til biosikkerhed niveau 2. Reagenser med indhold af natriumazid

Kalibratorer CAL indeholder natriumazid (NaN3) > 0,1 % (w/w) (< 1%).

Klassificering iht. CLP: Acute Tox. 4 Aquatic Chronic 3 Faresætninger: EUH032: Udvikler meget giftig gas ved kontakt med syre. H302: Farlig ved indtagelse. H412: Skadelig for vandlevende organismer, med

langvarige virkninger.

Sikkerhedssætninger: P264: Vask hænder grundigt efter brug. P270: Der må ikke spises, drikkes eller ryges under

brugen af dette produkt. P273: Undgå udledning til miljøet. P301+310: I TILFÆLDE AF INDTAGELSE: Ring omgående

til en GIFTINFORMATION eller en læge. P330: Skyl munden.

5. Forholdsregler ved håndtering Med undtagelse af kalibratorerne, som er frysetørrede, er sættets reagenser klar til brug. Se procedurens afsnit Kalibratorer vedrørende rekonstitutionsmetode. Før en ny patron placeres i IDS-iSYS Multi-Discipline Automated System (System), skal operatøren blande indholdet i beholderen med de magnetiske partikler med en rask, roterende bevægelse. Dette resuspenderer de magnetiske partikler, der har bundfældet sig under transporten. Det skal omhyggeligt undgås, at der dannes skum i reagenspatronen.

Page 32: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Dansk Side 2/6

Holdbarhed og opbevaring af reagenser Opbevar patronen og kalibratorerne i opretstående position på et mørkt sted ved 2-8 °C. Nedfrys ikke patronen eller kalibratorerne.

Holdbarhed af reagens Patron Kalibratorer

Før åbning ved 2-8 °C Til udløbsdato

Patron efter åbning ved 2-8 °C

28 dage Ikke relevant

Kalibratorer efter rekonstitution ved 2-8 °C

Ikke relevant 28 dage

På systemet* 7 dage 3 timer

*Kontinuerlig stabilitet

6. Prøveopsamling og opbevaring Analysen skal udføres ved hjælp af serum- (standardprøvetagningsrør eller rør, der indeholder serumseparationsgel) eller plasmaprøver (litiumheparin, natriumheparin, ammoniumheparin eller kalium- (K2) EDTA). Prøverne skal separeres så hurtigt som muligt efter opsamlingen. Opbevar prøver ved -20 °C eller derunder. Undgå gentagen nedfrysning/optøning af prøverne. Bemærk:

- Prøver, der indeholder partikler skal centrifugeres, inden analysen udføres. Brug ikke varmeinaktiverede prøver. - For at minimere eventuelle fordampningseffekter skal prøver, kalibratorer og kontroller måles inden for 3 timer efter

placering i Systemet. - Alle laboratorier skal følge de retningslinjer eller krav, der findes i lokale, statslige og/eller regionale bestemmelser

eller hos akkrediteringsorganisationerne, for at etablere deres egen prøvehåndtering og opbevaringsstabilitet. For vejledning i passende praksis henvises der til CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Inden analyserne udføres, skal der sikres, at prøver, kalibratorer og kontroller har stuetemperatur (18-22 °C). - Visse kommercielt tilgængelige rør til opsamling af prøver kan påvirke testresultaterne i særlige tilfælde. - Det anbefales at følge rørproducentens anvisninger, især når prøver analyseres i primærrør. - Centrifugerede prøver med lipidlag øverst skal overføres til en prøvekop eller et sekundærrør. Man skal være

omhyggelig med kun at overføre afklarede prøver uden lipæmisk materiale.

7. Materialer Leverede materialer Reagenspatron

MPS3 Magnetiske partikler belagt med streptavidin i fosfat-BSA med natriumazid som konserveringsmiddel (< 0,1 %),

1 flaske, 2,3 mL

CONJ Anti-PINP murint monoklonalt antistof mærket med et acridiniumester-derivat i buffer, der indeholder bovine og

museproteiner med natriumazid som konserveringsmiddel (< 0,1 %), 1 flaske, 4,95 mL

Ab-BIO Anti-PINP murint monoklonalt antistof mærket med biotin i buffer, der indeholder muse- og bovine proteiner med

natriumazid som konserveringsmiddel (< 0,1 %), 1 flaske 16,8 mL

BUF Buffer, der indeholder muse- og bovine proteiner, med natriumazid som konserveringsmiddel (< 0,1 %), 1 flaske,

4,95 mL

DIL Buffer, der indeholder natriumazid som konserveringsmiddel (< 0,1 %), 1 flaske, 23,7 mL

Kalibratorer

CALA Frysetørret buffer, der indeholder BSA og oprenset humant PINP med < 1,0 % natriumazid som konserveringsmiddel (< 0,05 %

CALB rekonstitueret). 2, hver med 2 koncentrationsniveauer, 1,0 mL pr. flaske.

Kalibratorerne klargøres i følgende nominelle intervaller:

Kalibrator Intervala (ng/mL)

CALA 3,0 til 7,0 ng/mL

CALB 130,0 til 165,0 ng/mL

a Intervallerne er typiske for produktet og er kun tænkt som vejledende. Værdierne kan variere fra parti til parti. Der henvises til

analysecertifikatet (CofA) på mini-CD’en, hvad angår partispecifikke koncentrationer. Kalibratorintervallerne etableres ved at teste kalibratorprøverne i mindst 20 analyser kørt på ét system. Mini-CD Indeholder brugsanvisning til IDS-iSYS-reagenser, kontrolintervaller, CRY-filer og analysecertifikat.

Page 33: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Dansk Side 3/6

Nødvendige materialer, som ikke følger med System Valgfrit udstyr/materialer IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Analyseprocedure Reagenspatron De reagenser, der leveres i patronen, er klar til brug, med undtagelse af kalibratorerne. Før en ny patron placeres i Systemet, skal indholdet i beholderen med de magnetiske partikler blandes med en rask, roterende bevægelse. Undgå skumdannelse. Stregkoden aflæses, når patronen sættes i reagensbakken. Hvis stregkodelæseren på Systemet ikke kan læse etiketten, er der en manuel fremgangsmåde til indkodning af stregkodedataene (se brugervejledningen til IDS-iSYS). Systemet blander automatisk alle de magnetiske partikler for alle reagenspatroner, der isættes, for at opretholde homogenitet. Patronen skal sættes i reagensbakken mindst 40 minutter før start af analysen. Hvis patronen fjernes fra reagensbakken, skal den opbevares opretstående ved 2-8 °C på et mørkt sted. Kalibratorer Kalibratorerne i dette kit er sporbare til producentens arbejdskalibratorer, som klargøres ud fra højt oprenset humant PINP, hvis værdi er sporbar til en kommercielt tilgængelig metode. IDS-iSYS Intact PINP-analysens kalibratorer er frysetørret. Rekonstitueres umiddelbart før brug. Tilsæt 1,0 mL destilleret eller afioniseret vand til hver flaske. Sæt proppen på igen. Lad kalibratorerne stå og rekonstituere i 10 min. med lejlighedsvis forsigtig manuel blanding. Pas på, der ikke dannes skum. Kalibratorer kan sættes i Systemet enten ved hjælp af prøvekopperne eller stativerne til kalibratorer/kontroller (30/31/32). Kalibrator-/kontrolstativer: Placer kalibratorhætteglassene i det relevante stativ, og indsæt det i Systemet, fortsæt i henhold til instruktionerne i brugervejledningen til systemet. Kassér materialet i kalibratorhætteglassene efter brug. Kalibratorhætteglas, som allerede har været i systemet, må IKKE genanvendes. Prøvekopper: Afpipetter cirka 200 μL kalibratorer i prøvekopperne, og anbring dem på systemet. Følg anvisningerne i betjeningsvejledningen til Systemet. Bortskaf materialet i prøvekopperne efter brug. Materialet må IKKE hældes tilbage i hætteglassene med kalibratorer. Hvis kalibratorerne skal bruges mere end en gang, skal de opbevares ved 2 til 8 °C inden for 3 timer efter rekonstitution. Når kalibratorerne i de opbevarede hætteglas skal anvendes igen, bringes de tilbage til stuetemperatur (18-22 °C) og vendes forsigtigt før brug. Kalibreringsfrekvens Der kræves ny kalibrering:

- Efter isætning af hvert nyt parti patroner. - Når kontrolværdierne ikke falder inden for de definerede intervaller. - Når kalibreringsintervallet på 3 dage er udløbet. - Efter service på Systemet.

Der sker automatisk verificering af kalibreringen. Dette styres af Systemet. Kalibrering af systemet De to IDS-iSYS Intact PINP analysekalibratorer skal foretage justeringen af master-kurven. Kalibratorerne leveres sammen med kittet, og der må ikke bruges kalibratorer fra et andet parti. Alle niveauer af IDS-iSYS Intact PINP Control Set (IS-4030) SKAL måles samtidigt med kalibratorerne for at validere masterkurvejusteringen. Alle de nødvendige data til kalibrering af patronen findes på mini-CD’en. Benyt kalibratorniveauerne A og B til at justere masterkurven til reagenserne på Systemet. Kontrollér, at der er en IDS-iSYS Intact PINP analysepatron på reagensbakken, og at der er afgang til masterkurven for patronen i databasen. Hvis dataene for partiet af kalibratorer ikke er tilgængelige i Systemet, skal dataene indlæses ved hjælp af den mini-CD, der følger med i reagenssættet. Start kalibreringen af immunanalysen på Systemet i henhold til brugervejledningen til Systemet. Kalibreringen udføres ved anvendelse af 3 replikater for hvert niveau. Der må fjernes et replikat for at overholde kalibreringskravene. Alle kontrolniveauer skal også måles to gange for at kalibrere analysen. Verificer og godkend kalibreringen ud fra den kalibreringsstatus, der vises i kalibreringsvinduet, og bortskaf resten af kalibratorerne og kontrollerne fra prøvebakken efter brug.

Page 34: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Dansk Side 4/6

9. Kvalitetskontrol Anvend IDS-iSYS Intact PINP Control Set til kvalitetskontrol. For at sikre resultaternes validitet skal der måles mindst tre kontroller med varierende niveauer af intakt PINP. Andet relevant kontrolmateriale kan anvendes ud over IDS-iSYS Intact PINP Control Set. Kontroller skal testes ved begyndelsen (eller tæt derpå) af hver kørsel med patientprøver og under hver kalibrering, i overensstemmelse med lokale og/eller statslige bestemmelser eller godkendelseskrav og laboratoriets kvalitetsprocedure. Det anbefales, at kontrollerne rutinemæssigt køres i duplikat. Laboratoriet bør teste kontrollerne mindst én gang pr. vagt. Der henvises til brugervejledningen til IDS-iSYS Intact PINP Control Set for instruktioner vedrørende klargøring og håndtering. Bestemmelse af prøvens niveauer af intakt PINP Prøverne analyseres i henhold til anvisningerne i brugervejledningen til Systemet.

10. Beregning af resultater Koncentrationen af intakt PINP i hver af prøverne beregnes automatisk. Koncentrationerne vises på skærmen eller som udskrift, når bruger anmoder herom. IDS-iSYS Intact PINP-analysen benytter en 4-parameters logistisk kurvetilpasning (4PL) til beregning af koncentrationen af intakt PINP.

11. Måleområde (rapporterbart interval) Det rapporterbare interval for analysen er 2-230 ng/mL. Enhver værdi under 2 ng/mL skal rapporteres som ”< 2 ng/mL”.

12. Fortynding Prøver med koncentrationer af intakt PINP over det rapporterbare interval skal fortyndes manuelt med en human serumprøve

med lav koncentration i et forhold på 1 til 2. Resultaterne for de fortyndede prøver skal multipliceres med fortyndingsfaktoren 2 og korrigeres for koncentrationen af den lave prøve.

13. Begrænsninger for brug - Som ved enhver diagnostisk procedure skal resultaterne fortolkes i sammenhæng med patientens kliniske tilstand og

andre oplysninger, som lægen har adgang til. - Ydeevnekarakteristikaene ved denne analyse er ikke etableret i en pædiatrisk population. - Hook-effekten blev testet ved brug af koncentrationer af analyt på op til 2350 ng/mL. Der blev ikke observeret en

hook-effekt. - Heterofile antistoffer i humant serum kan reagere med reagensets immunglobulin og interferere med in vitro-

immunanalyser6. Patienter, der jævnligt udsættes for dyr eller animalske serumprodukter, kan være disponerede for

denne interferens, hvorfor anormale værdier kan forekomme. - Følgende stoffer interfererer ikke med IDS-iSYS intact PINP analysen, når koncentrationerne i tabellen herunder

ligger under den angivne tærskel.

Potentielt interfererende stof

Tærskel Koncentration

Lipid 2803 mg/dL

Bilirubin 200 mg/dL

Hæmoglobin 500 mg/dL

Biotin 300 nmol/L

14. Forventede værdier Følgende intervaller blev bestemt ved anvendelse af IDS-iSYS Intact PINP-analysen og er kun til information. 95 % referenceintervallet for tilsyneladende raske voksne blev beregnet med en ikke-parametrisk metode efter vejledning fra CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Normale voksne 27,7-127,6 ng/mL (n = 150) Intervallerne herover skal betragtes som kun vejledende; det anbefales, at hvert laboratorium etablerer sit eget forventede interval på grundlag af laboratoriets egen patientpopulation.

Page 35: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Dansk Side 5/6

15. Ydeevnedata Der vises repræsentative ydeevnedata. Resultaterne, der opnås på de enkelte laboratorier, kan variere. 15.1 Sensitivitet Blindværdigrænsen (LoB), detektionsgrænsen (LoD) og kvantiseringsgrænsen (LoQ) blev bestemt med vejledning fra CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation” ved anvendelse af 60 blindprøver og 120 lavniveauprøver.

Sensitivitet Koncentration

LoB 0,5 ng/mL

LoD < 1,0 ng/mL

LoQ < 1,0 ng/mL

15.2 Præcision Præcision blev evalueret i henhold til en modificeret protokol baseret på CLSI EP-5A2, ”Evaluation of Precision Performance of Quantitative Measurement Methods”. I alt 3 serumkontroller blev analyseret ved hjælp af et parti reagenser i fire eksemplarer én gang om dagen i 20 dage på to systemer.

Kontrol n Middel (ng/mL)

I kørsel Total

SD CV SD CV%

1 80 14,3 0,4 2,6 % 0,6 4,2 %

2 80 44,6 1,3 3,0 % 2,4 5,3 %

3 80 100,9 3,0 3,0 % 4,5 4,4 %

15.3 Recovery Recovery blev vurderet ved tilsætning af intakt PINP til prøver før analyse.

Prøvekonc. ng/mL

PINP tilsat ng/mL

Målt ng/mL

Recovery ng/mL

Recovery %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Middelværdi

93 95 93 91 95 91 93

15.4 Linearitet Linearitet blev vurderet på grundlag af CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures”. A Statistical Approach”. Prøver indeholdende varierende koncentrationer af intakt PINP blev analyseret i duplikat. De resulterende middelkoncentrationer blev sammenlignet med de forventede koncentrationer. Prøverne blev klargjort ved fortynding af en patientprøve med højt niveau med en patientprøve med lavt indhold forud for analysen.

Forventet koncentration (ng/mL)

Målt koncentration (ng/mL)

Variation

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9 -0,2 -1,6 -2,9 -2,4 3,7 0,0 0,7

- 3 0 -2 -3 -2 2 0 0

15.5 Metodesammenligning IDS-iSYS Intact PINP-analysen blev sammenlignet med en kommercielt tilgængelig kvantitativ radioimmunanalyse, i henhold til CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. I alt 363 prøver, udvalgt med henblik på at repræsentere et bredt interval af koncentrationer af intakt PINP [8,7-230 ng/mL], blev analyseret med de to metoder. Lineær regressionsanalyse blev udført på de komparative data:

n Hældningskoefficient 95 % CI Intercept (ng/mL) 95 % CI Korrelationskoefficient (r)

363 0,86 0,84 til 0,88 -1,9 -3,4 til -0,3 0,98

Page 36: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Dansk Side 6/6

15.6 Specificitet Specificiteten blev vurderet med følgende analytter.

Analyt Krydsreaktivitet

Intakt PINP 100 %

Rotte-/muse-PINP < 0,01 %

PIINP < 0,03 %

N-Mid Osteocalcin < 0,02 %

16. Anvendte symboler

REF Katalognummer

IVD In vitro-diagnostisk udstyr

Fabrikant

Anvendes i overensstemmelse med direktiv 98/79/EF

17. Bibliografi 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., ‘Heterophilic antibodies: a problem for all immunoassays’. Clin Chem, 34, 1988, pp 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tlf.: +44 191 519 0660 • Fax: +44 191 519 0760 e-mail: [email protected] • www.idsplc.com Immunodiagnostic Systems USA Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tlf.: +1 877 852 6210 • Fax: +1 301 990 4236 e-mail: [email protected]

Tyskland Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tlf.: +49 692 6019 0960 • Fax: +49 692 6019 0949 e-mail: [email protected]

Frankrig Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tlf.: +33 140 770 450 • Fax: +33 140 770 455 e-mail: [email protected]

Skandinavien Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tlf.: +45 448 400 91 e-mail: [email protected]

Belgien Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tlf.: +32 425 226 36 • Fax: +32 425 251 96 e-mail: [email protected]

Brasilien IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP, Brasil. CEP: 05422 001 Tlf.: +55 113 740 6100 • Fax: + 55 113 740 6105 e-mail: [email protected]

Page 37: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Bruksanvisning

IS-4000

IS-4000 v06 01 februari 2017 Svenska Sida 1/6

1. Avsedd användning För in vitro-diagnostik. IDS-iSYS Intact PINP-analysen (intakt aminoterminal propeptid typ I prokollagen) är avsedd för kvanitativ bestämning av intakt PINP i humant serum eller plasma på IDS-iSYS Multi-Discipline Automated System. Testet är avsett att användas som hjälpmedel vid hantering av osteoporos hos kvinnor som nått klimakteriet.

2. Summary and Explanation Ett viktigt steg i benbildningsprocessen är syntes av typ I-kollagen, vilket är den främsta organiska komponenten i benmatrix. Under kollagensyntes frisätts propeptider både från de amino- och karboxiterminala ändarna av prokollagenmolekylen

1.

Propeptiderna utsöndras i blodomloppet. Den aminoterminala propeptiden av typ I-prokollagen (PINP) är troligen den mest specifika och känsliga markören för benbildning

2. PINP är en särskilt användbar markör för att kontrollera vilken effekt behandling

med anabola preparat har på osteoporos3,4

, men det är också en av de bästa omsättningsmarkörerna för att kontrollera effekten av behandling mot bennedbrytning

4,5.

3. Beskrivning av metod Analysen baseras på kemiluminescensteknik. 20 µL patientprover eller kalibratorer späds i ett spädningsmedel och en del av detta inkuberas med en biotinylerad anti-PINP monoklonal antikropp, en monoklonal antikropp märkt med akridinium, magnetiska partiklar märkta med streptavidin och en analysbuffert. De magnetiska partiklarna fångas med hjälp av en magnet och ett tvättsteg utförs för att ta bort eventuell obunden analyt. Utlösarreagenser tillsätts; det resulterande ljuset som utstrålas av akridiniumetiketten är direkt proportionerligt till koncentrationen av intakt PINP i det ursprungliga provet.

4. Varningar och säkerhetsåtgärder IDS-iSYS Intact PINP är endast avsett för in vitro-diagnostiskt bruk och inte för invändigt bruk i människor eller djur. Denna produkt måste användas strängt enligt de anvisningar som anges i denna bruksanvisning. Immunodiagnostic Systems Limited (IDS) påtar sig inget ansvar för eventuell förlust eller skada (utom där så krävs i lag) oavsett orsak, som uppkommer genom underlåtenhet att efterleva de angivna anvisningarna. VARNING: Detta kit innehåller material av animaliskt ursprung. Hantera kitreagenserna som om de skulle kunna överföra smittämne. Lämpliga försiktighetsmått och god laboratoriesed måste följas vid lagring, hantering och bortskaffande av kitreagenserna. Bortskaffande av kitreagenserna ska ske enligt lokala bestämmelser.

Humana material Humant material som använts vid beredningen av denna produkt har testats för förekomsten av antikropp mot humant immunbristvirus (HIV I och II), hepatit B-ytantigen, antikropp mot hepatit C och funnits vara negativa enligt analyser som rekommenderats av FDA. Eftersom inget test kan helt tillförsäkra att smittsamma ämnen saknas ska reagenserna hanteras enligt biologisk säkerhetsnivå 2. Reagenser som innehåller natriumazid

Kalibratorerna CAL innehåller natriumazid (NaN3) > 0,1 % (vikt/vikt) (< 1 %).

Klassificering enligt CLP: Acute Tox. 4 Aquatic Chronic 3 Faroangivelser: EUH032: Utvecklar mycket giftig gas vid kontakt med syra. H302: Skadligt vid förtäring. H412: Skadliga långtidseffekter för vattenlevande

organismer

Skyddsangivelser: P264: Tvätta händerna grundligt efter användning. P270: Ät inte, drick inte och rök inte när du använder

produkten. P273: Undvik utsläpp till miljön. P301+310: VID FÖRTÄRING: Kontakta genast

GIFTINFORMATIONSCENTRAL eller läkare. P330: Skölj munnen.

5. Försiktighetsåtgärder vid hantering Bortsett från kalibratorerna, som är lyofiliserade, är reagenserna som ingår i satsen klara att användas. Se kalibreringsdelen av proceduren för rekonstitutionsmetodik. Innan en ny kassett laddas på IDS-iSYS Multi-Discipline Automated System (systemet) måste behållaren för magnetiska partiklar blandas av användaren med en livlig roterande rörelse. På så sätt suspenderas de magnetiska partiklarna som har satt sig under transport igen. Försiktighet måste iakttas för att undvika skumbildning i reagenskassetten.

Page 38: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februari 2017 Svenska Sida 2/6

Hållbarhet och lagring av reagenser Förvara kassetten och kalibratorerna i upprättstående läge i mörker vid 2–8 °C. Frys inte kassetten eller kalibratorerna.

Hållbarhet för reagens Kassett Kalibratorer

Före öppning vid 2–8 °C Till utgångsdatumet

Kassett, öppnad vid 2–8 °C

28 dagar Ej tillämpligt

Kalibratorer, efter rekonstituering vid 2–8 °C

Ej tillämpligt 28 dagar

På systemet * 7 dagar 3 timmar

* kontinuerlig stabilitet på instrumentet

6. Provtagning och förvaring Analysen ska utföras med serum- (standardprovrör eller rör som innehåller serumseparationsgel) eller plasmaprover (litiumheparin, natriumheparin, ammoniumheparin eller kalium (K2) EDTA). Proverna ska separeras så fort som möjligt efter insamling. Förvara prover vid –20 °C eller lägre. Undvik att frysa och tina upp prover upprepade gånger. Observera:

- Prover som innehåller partiklar ska centrifugeras innan analysen utförs. Använd inte värmeinaktiverade prover. - Prover, kalibratorer och kontroller ska mätas inom 3 timmar efter att ha placerats i systemet för att minimera möjliga

avdunstningseffekter. - Varje laboratorium ska följa riktlinjerna eller kraven i lokala, statliga och/eller federala bestämmelser eller enligt

ackrediteringsorganisationer för att fastställa sin egen provhantering och provens lagringsstabilitet. Vägledning om lämpliga förfaranden finns i CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Innan analyser utförs, säkerställ att prover, kalibratorer och kontroller är vid rumstemperatur (18–22 ºC). - Vissa i handeln tillgängliga provinsamlingsrör kan påverka resultaten av testning i vissa fall. - Det rekommenderas att provrörstillverkarens anvisningar följs, särskilt vid bearbetning av prover i primära provrör. - Centrifugerade prover med ett lipidlager överst måste överföras till en provkopp eller ett sekundärt rör. Försiktighet

måste iakttas med att endast överföra de klara proverna utan det lipemiska materialet.

7. Material Levererat material Reagenskassett

MPS3 Magnetpartiklar överdragna med streptavidin i en fosfatbuffert med natriumazid som konserveringsmedel (< 0,1 %),

1 flaska, 2,3 mL

CONJ Anti-PINP monoklonal musantikropp märkt med ett akridiniumesterderivat, i en buffert innehållande proteiner från

nötkreatur och mus med natriumazid som konserveringsmedel (< 0,1 %), 1 flaska, 4,95 mL.

Ab-BIO Anti-PINP monoklonal musantikropp märkt med biotin, i en buffert innehållande proteiner från mus och nötkreatur

med natriumazid som konserveringsmedel (< 0,1 %), 1 flaska, 16,8 mL.

BUF Buffert som innehåller proteiner från mus och nötkreatur, med natriumazid som konserveringsmedel (< 0,1 %), 1 flaska,

4,95 mL

DIL Buffert som innehåller natriumazid som konserveringsmedel (< 0,1 %), 1 flaska, 23,7 mL

Kalibratorer

CALA Lyofiliserad buffert som innehåller BSA och renat humant PINP med < 1,0 % natriumazid som konserveringsmedel

CALB (< 0,05 % rekonstituerat). 2 av vardera av 2 koncentrationsnivåer, 1,0 mL per flaska.

Kalibratorerna förbereds enligt följande nominella intervaller:

Kalibrator Intervalla (ng/mL)

CALA 3,0 till 7,0 ng/mL

CALB 130,0 till 165,0 ng/mL

a Intervallen är typiska för produkten och är endast avsedda som en vägledning. Värdena kan variera från parti till parti. I

analyscertifikatet (CofA) på mini-cd:n anges de batchspecifika koncentrationerna. Kalibratorintervallerna etableras genom att testa kalibratorproverna med minst 20 analyser på ett system. Mini-CD-skiva Innehåller bruksanvisningen till IDS-iSYS-reagenser, kontrollområden, CRY-filer och analysintyg.

Page 39: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februari 2017 Svenska Sida 3/6

Material som krävs men inte medföljer System Tillvalsutrustning/material IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Analysförfarande Reagenskassett Reagenserna i kassetten är klara för användning, med undantag av kalibratorerna. Innan en ny kassett laddas på systemet, blanda behållaren med magnetiska partiklar med en livlig roterande rörelse. Undvik skumbildning. Streckkoden läses in när kassetten har laddats på reagensbrickan. Om etiketten inte kan läsas av systemets streckkodläsare finns det en manuell procedur för inmatning av streckkodsdata (se bruksanvisningen till IDS-iSYS). Systemet blandar automatiskt de magnetiska partiklarna för alla reagenskassetter som laddats på instrumentet för att bibehålla homogenitet. Kassetten ska laddas i reagensbrickan minst 40 minuter innan analysen inleds. Om kassetten tas bort från reagensbrickan, förvara kassetten i upprättstående läge vid 2–8 °C i mörker. Kalibratorer Kalibratorerna i denna sats kan spåras till tillverkarens arbetskalibratorer som förbereds från höggradigt renat humant PINP vars värde kan spåras till en kommersiellt tillgänglig metod. IDS-iSYS Intact PINP-analyskalibratorerna är lyofiliserade. Gör rekonstitutionen omedelbart innan användning. Tillsätt 1,0 mL destillerat eller avjoniserat vatten till varje flaska. Sätt tillbaka korken. Låt stå i 10 minuter för att rekonstitueras med försiktig manuell blandning då och då. Försiktighet ska iakttas för att förhindra skumbildning. Kalibratorerna kan laddas i systemet antingen med hjälp av provbägarna eller med kalibrator-/kontrollrack (30/31/32). Kalibrator-/kontrollrack: Placera kalibratorampullerna i avsett rack och sätt i dem i systemet och fortsätt sedan enligt instruktionerna i användarhandboken för systemet. Kassera materialet i kalibratorampullerna efter användning. Återanvänd INTE kalibratorampuller som har satts in på systemet. Provkoppar: Pipettera cirka 200 μL med kalibratorer i provkoppar och placera i systemet. Fortsätt enligt instruktionerna i systemets användarhandbok. Kassera materialet i provkopparna efter användning. Häll INTE tillbaka material i kalibratorampullen. Om kalibratorer ska användas mer än en gång ska kalibratorerna förvaras vid 2–8 °C inom 3 timmar efter rekonstruktion. När de förvarade kalibratorampullerna återanvänds ska kalibratorerna värmas till rumstemperatur (18–22 ºC) och vändas försiktigt före användning. Kalibreringsfrekvens En ny kalibrering krävs:

- Vid laddning av varje nytt parti med kassetter. - När kontrollvärdena inte hamnar inom de definierade områdena. - När kalibreringsintervallet på 3 dagar är slut. - Efter service på systemet.

Verifiering av kalibrering är automatisk och hanteras av systemet. Systemkalibrering De två IDS-iSYS Intact PINP-analyskalibratorerna är nödvändiga för att masterkurvan ska kunna justeras. Kalibratorerna levereras med kitet och kalibratorer från ett annat parti får inte användas. Alla nivåer av IDS-iSYS Intact PINP Control Set (IS-4030) MÅSTE mätas samtidigt som kalibratorerna för att validera justeringen av masterkurvan. Alla data som krävs för kalibrering av kassetten finns på mini-CD-skivan. Använd kalibratornivåerna A och B för att justera masterkurvan enligt reagenserna som är ombord på systemet. Kontrollera att det finns en IDS-iSYS Intact PINP-analyskassett på reagensbrickan och att kassettmasterkurvan är tillgänglig i databasen. Om data för partiet kalibratorer inte finns i systemet, ladda data med hjälp av mini-CD-skivan som medföljer reagenskitet. Starta kalibreringen av immunanalysen på systemet i enlighet med användarhandboken för systemet. Kalibreringen utförs med hjälp av 3 replikat på varje nivå. Ett replikat kan tas bort för att uppfylla kalibreringskraven. Alla kontrollnivåer måste också mätas dubbelt för att analysen ska kalibreras. Verifiera och godkänn kalibreringen i enlighet med kalibreringsstatus som visas i kalibreringsfönstret och kassera den återstående volymen med kalibratorer och kontroller på provbrickan efter användning.

Page 40: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februari 2017 Svenska Sida 4/6

9. Kvalitetskontroll (QC) Använd IDS-iSYS Intact PINP Control Set för kvalitetskontroll. För att säkerställa resultatens giltighet ska minst tre kontroller med olika koncentrationer av intakt PINP mätas. Annat lämpligt kontrollmaterial kan användas utöver IDS-iSYS Intact PINP Control Set. Kontrollerna bör testas i början (eller strax efter början) av varje körning som innehåller patientprover samt varje gång du genomför en kalibrering i enlighet med lokala eller nationella bestämmelser eller ackrediteringskrav och laboratoriets kvalitetsrutiner. Vi rekommenderar att kontrollerna rutinmässigt körs i duplikat. Laboratorier ska testa kontroller minst en gång per arbetspass. Anvisningar för beredning och hantering finns i bruksanvisningen till IDS-iSYS Intact PINP Control Set. Fastställande av nivåer av intakt PINP i prover Bearbeta proverna i enlighet med användarhandboken för systemet.

10. Beräkning av resultat Koncentrationen av PINP i varje prov beräknas automatiskt. Visningen av koncentrationerna (skärm eller utskrift) görs på begäran av användaren. Vid IDS-iSYS Intact PINP-analys används en logistikkurvanpassning med 4 parametrar (4PL) för att beräkna koncentrationerna av intakt PINP.

11. Mätintervall (rapporterbart intervall) Det rapporterbara området för analysen är 2–230 ng/mL. Varje värde under 2 ng/mL ska rapporteras som ”< 2 ng/mL”.

12. Spädning Prover med intakt PINP-koncentrationer över det rapporterbara intervallet ska spädas manuellt med låg koncentration i förhållandet 1 till 2. Resultaten för spädda prover måste multipliceras med spädningsfaktor 2 och korrigeras för koncentrationen i det låga provet.

13. Användningsbegränsningar - Som i fallet av alla diagnostiska förfaranden måste resultat tolkas tillsammans med patientens kliniska presentation

och annan information som är tillgänglig för läkaren. - Prestandakarakteristiken för denna analys har inte fastställts i en pediatrisk population. - Hook-effekten testades med användning av analytkoncentrationer upp till 2 350 ng/mL. Ingen hook-effekt

observerades. - Heterofila antikroppar i humant serum kan reagera med reagensimmunoglobuliner och störa in vitro-immunanalyser

6. Patienter som rutinmässigt exponeras för djur eller djurserumprodukter kan ha mer fallenhet för denna störning och

avvikande värden kan observeras. - Följande ämnen stör inte IDS-iSYS intact PINP-analysen när de koncentrationer som anges i följande tabell ligger

under det angivna tröskelvärdet.

Potentiellt störande medel

Tröskel- Koncentration

Lipid 2 803 mg/dL

Bilirubin 200 mg/dL

Hemoglobin 500 mg/dL

Biotin 300 nmol/L

14. Förväntade värden Följande intervall har fastställts med hjälp av IDS-iSYS Intact PINP -analysen och tillhandahålls endast i informationssyfte. Referensintervallet på 95 % för synbart friska vuxna beräknades med en icke parametrisk metod enligt vägledning i CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Normala vuxna 27,7–127,6 ng/mL (n = 150) Ovanstående intervall bör endast betraktas som riktlinjer. Vi rekommenderar att varje laboratorium fastställer sitt eget förväntade intervall baserat på den egna patientpopulationen.

Page 41: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februari 2017 Svenska Sida 5/6

15. Prestandauppgifter Representativa prestandauppgifter visas. De erhållna resultaten vid enskilda laboratorier kan variera. 15.1 Känslighet Blankgräns (LoB), detekteringsgräns (LoD) och kvantiteringsgräns (LoQ) fastställdes med vägledning från CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation” med hjälp av 60 blanka och 120 lågnivåprover.

Känslighet Koncentration

LoB 0,5 ng/mL

LoD < 1,0 ng/mL

LoQ < 1,0 ng/mL

15.2 Noggrannhet Precision utvärderades i enlighet med ett modifierat protokoll baserat på CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”. Sammanlagt 3 serumkontroller analyserades fyra gånger med ett parti med reagenser en gång per dag under 20 dagar på två system.

Kontroll n Medelvärde

(ng/mL)

Inom körning Totalt

SD CV SD CV %

1 80 14,3 0,4 2,6 % 0,6 4,2 %

2 80 44,6 1,3 3,0 % 2,4 5,3 %

3 80 100,9 3,0 3,0 % 4,5 4,4 %

15.3 Återhämtning Återhämtning utvärderades genom att tillsätta intakt PINP till prover under analysen.

Provkonc. ng/mL

PINP tillsatt ng/mL

Mätning ng/mL

Återhämtning ng/mL

Återhämtning %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Medel

93 95 93 91 95 91 93

15.4 Linjäritet Linjäriteten utvärderades baserat på CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”. Prover innehållande varierande koncentrationer av intakt PINP analyserades två gånger. De resulterande medelkoncentrationerna jämfördes med förväntade koncentrationer. Proverna bereddes genom spädning av ett högt patientprov med ett lågt patientprov före analysen.

Beräknad koncentration (ng/mL)

Uppmätt koncentration (ng/mL)

Variation

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9

–0,2 –1,6 –2,9 –2,4 3,7 0,0 0,7

- 3 0

–2 –3 –2 2 0 0

15.5 Metodjämförelse IDS-iSYS analysnamn-analysen jämfördes med en kommersiellt tillgänglig kvantitativ analys enligt CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. Totalt har 363 prover, utvalda för att representera ett brett intervall av intakt PINP -koncentrationer [8,7–230 ng/mL] analyserades med varje metod. Linjär regressionsanalys utfördes på jämförelsedata:

n Lutning 95 % CI Intercept (ng/mL) 95 % CI Korrelationskoefficient (r)

363 0,86 0,84 till 0,88 –1,9 –3,4 till –0,3 0,98

Page 42: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februari 2017 Svenska Sida 6/6

15.6 Specificitet Specificiteten bedömdes med följande analyter.

Analyt Korsreaktivitet

Intakt PINP 100 %

PINP från råtta/mus < 0,01 %

PIINP < 0,03 %

N-Mid Osteokalcin < 0,02 %

16. Använda symboler

REF Katalognummer

IVD Enhet för in vitro-diagnostik

Tillverkare

Tillämpas i enlighet med rådets direktiv 98/79/EG

17. Litteraturförteckning 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., ‘Heterophilic antibodies: a problem for all immunoassays’. Clin Chem, 34, 1988, sid 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tel.: +44 191 519 0660 • Fax: +44 191 519 0760 e-post: [email protected] • www.idsplc.com Immunodiagnostic Systems USA Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tel.: +1 877 852 6210 • Fax: +1 301 990 4236 e-post: [email protected]

Tyskland Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tel.: +49 692 6019 0960 • Fax: +49 692 6019 0949 e-post: [email protected]

Frankrike Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tel.: +33 140 770 450 • Fax: +33 140 770 455 e-post: [email protected]

Skandinavien Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tel: +45 448 400 91 e-post: [email protected]

Belgien Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tel.: +32 425 226 36 • Fax: +32 425 251 96 e-post: [email protected]

Brasilien IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasil. CEP: 05422 001 Tel: +55 113 740 6100 • Fax: + 55 113 740 6105 e-post: [email protected]

Page 43: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Bruksanvisning

IS-4000

IS-4000 v06 01 februar 2017 Norsk Side 1/6

1. Bruksområde Til in vitro-diagnostisk bruk. IDS-iSYS Intact PINP-assayet (intakt aminoterminal propeptid av prokollagen type I) er beregnet for kvantitativ måling av intakt PINP i humant serum eller plasma på IDS-iSYS Multi-Discipline Automated System. Testen skal brukes som et hjelpemiddel ved behandling av postmenopausal osteoporose.

2. Summary and Explanation Et viktig skritt i bendannelsesprosessen er syntesen av type I-kollagen, som er den viktigste organiske komponenten i benmatriks. Under kollagensyntese frigjøres propeptider både fra amino- og karboksyterminaldelene av prokollagen-molekylet

1.

Disse propeptidene skilles ut i blodsirkulasjonen. Amino-terminal propeptid av type I-prokollagen (PINP) er sannsynligvis den mest spesifikke og sensitive markøren for bendannelse

2. PINP er en særlig nyttig markør for overvåking av effektiviteten til

osteoporosebehandling med anabole midler3,4

, men det er også en av de beste benomsetningsmarkører overvåking av effektiviteten ved antiresorptiv behandling

4,5.

3. Metodebeskrivelse Assayet er basert på kjemisk luminescensteknologi. 20 µL av prøver eller kalibratorer blir fortynnet i et fortynningsmiddel, og en del av dette inkuberes med et biotinylert anti-PINP-monoklonalt antistoff, et akridinium-merket monoklonalt antistoff, streptavidin-merkede magnetiske partikler og et assaybuffer. De magnetiske partiklene fanges ved hjelp av en magnet, og det utføres et vasketrinn for å fjerne eventuelt ubundet analytt. Triggerreagenser tilsettes, lyset som derved emitteres fra akridinmerket, er omvendt proporsjonalt med konsentrasjonen av intakt PINP i den opprinnelige prøven.

4. Advarsler og forholdsregler The IDS-iSYS Intact PINP er kun til in vitro-diagnostisk bruk og ikke til innvortes bruk hos mennesker eller dyr. Dette produktet må brukes i strengt samsvar med instruksjonene som er angitt i denne bruksanvisningen. Immunodiagnostic Systems Limited (IDS) skal ikke holdes ansvarlig for eventuelle tap eller skader (unntatt det som kreves i medhold av lov), uansett årsak, som følge av manglende overholdelse av instruksjonene. FORSIKTIG: Dette settet inneholder et materiale av animalsk opprinnelse. Håndter settreagensene som om de kan overføre smittestoff. Passende forholdsregler og god laboratoriepraksis må følges ved oppbevaring, håndtering og avhending av settreagensene. Avhending av settreagensene skal skje i samsvar med lokale forskrifter.

Humane materialer Det humane materialet brukt i klargjøringen av dette produktet er testet i FDA-anbefalte assayer for påvisning av antistoff mot human immunsviktvirus (HIV I og II), hepatitt B-overflateantigen, antistoff mot hepatitt C, og funnet å være negative. Ettersom ingen test kan garantere fullstendig fravær av smittestoffer, skal reagenser behandles i henhold til biosikkerhetsnivå 2. Reagenser som inneholder natriumazid

Kalibratorene CAL inneholder natriumazid (NaN3) > 0,1 vekt-% (< 1 %).

Klassifisering i henhold til CLP: Acute Tox. 4 Aquatic Chronic 3 Fareutsagn: EUH032: Ved kontakt med syrer utvikles meget giftig gass. H302: Farlig ved svelging. H412: Skadelig, med langtidsvirkning, for liv i vann.

Forsiktighetsutsagn: P264: Vask hender grundig etter bruk. P270: Ikke spis, drikk eller røyk ved bruk av produktet. P273: Unngå utslipp til miljøet. P301+310: VED SVELGING: Kontakt umiddelbart et

GIFTINFORMASJONSSENTER eller lege. P330: Skyll munnen.

5. Forholdsregler for håndtering Bortsett fra kalibratorene som lyofiliseres, er reagensene som leveres i pakken, klare til bruk. Se Kalibrator-avsnittet av prosedyren for rekonstitueringsmetodikk. Før en ny kassett settes inn i IDS-iSYS Multi-Discipline Automated System (systemet), krever den magnetiske partikkelbeholderen at operatøren blander den med en rask rotasjonsbevegelse. Dette vil resuspendere de magnetiske partiklene som er bunnfelt under transport. Pass på å unngå skumdannelse i reagenskassetten.

Page 44: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Norsk Side 2/6

Holdbarhet og oppbevaring av reagenser Oppbevar kassett og kalibratorer i en oppreist stilling i mørke ved 2–8 °C. Ikke frys kassetten eller kalibratorene.

Reagensenes holdbarhet Kassett Kalibratorer

Før åpning ved 2–8 °C Til utløpsdatoen

Kassett etter åpning ved 2–8 °C

28 dager Ikke aktuelt

Kalibratorer etter rekonstituering ved 2–8 °C

Ikke aktuelt 28 dager

I systemet * 7 dager 3 timer

* Kontinuerlig stabilitet i systemet

6. Prøvetaking og prøveoppbevaring Assayet bør gjennomføres ved hjelp av serum (standard prøverør eller rør som inneholder serumsepareringsgel) eller plasmaprøver (litiumheparin, natriumheparin, ammoniumheparin eller kalium- (K2) EDTA). Prøvene skal separeres så snart som mulig etter innsamling. Oppbevar prøvene ved -20 °C eller lavere. Unngå gjentatt frysing og tining av prøver. Merk:

- Prøver som inneholder partikkelmateriale må sentrifugeres før analysen utføres. Varmeinaktiverte prøver skal ikke brukes.

- For å minimere mulig fordunsting bør prøver, kalibratorer og kontrolløsninger måles innen 3 timer etter at de er plassert på systemet.

- Hvert laboratorium skal følge retningslinjene eller kravene i lokale, statlige og/eller nasjonale forskrifter eller akkrediteringsorganisasjoners krav for å bestemme egen prøvehåndterings- og lagringsstabilitet. For veiledning om aktuell praksis henvises det til CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Før analyseringen utføres, må det sørges for at prøver, kalibratorer og kontrolløsninger holder romtemperatur (18–22 °C). - Noen kommersielt tilgjengelige prøvetakingsrør kan i særlige tilfeller påvirke resultatene av testingen. - Det anbefales å følge instruksene fra rørprodusenten, spesielt ved behandling av prøvene i primærrørene. - Sentrifugerte prøver med et lipidlag på toppen må overføres til et prøvebeger eller et sekundært rør. Vær nøye med å

overføre kun de klarede prøvene uten det lipemiske materialet.

7. Materialer Medfølgende materiale Reagenskassett

MPS3 Magnetiske partikler belagt med streptavidin i fosfat-BSA-buffer med natriumazid som konserveringsmiddel (< 0,1 %),

1 flaske, 2,3 mL

CONJ Anti-PINP monoklonalt antistoff fra mus/rotte merket med et akridinesterderivat, i buffer som inneholder bovine

proteiner og museproteiner med natriumazid som konserveringsmiddel (< 0,1 %), 1 flaske, 4,95 mL.

Ab-BIO Anti-PINP monoklonalt antistoff fra mus/rotte merket med biotin, i buffer som inneholder bovine proteiner og

museproteiner med natriumazid som konserveringsmiddel (< 0,1 %), 1 flaske, 16,8 mL.

BUF Buffer som inneholder bovine proteiner og museproteiner med natriumazid som konserveringsmiddel (< 0,1 %),

1 flaske, 4,95 mL

DIL Buffer som inneholder natriumazid som konserveringsmiddel (< 0,1 %), 1 flaske, 23,7 mL

Kalibratorer

CALA Lyofilisert buffer som inneholder BSA og renset humant PINP med < 1,0 % natriumazid som konserveringsmiddel (< 0,05 %

CALB rekonstituert). 2 hver av 2 konsentrasjonsnivåer, 1,0 mL per flaske.

Kalibratorene er klargjort på følgende nominelle områder:

Kalibrator Områdea (ng/mL)

CALA 3,0–7,0 ng/mL

CALB 130,0–165,0 ng/mL

a Områdene er typiske for produktet og er kun ment som en veiledning. Verdiene kan variere fra parti til parti. Se

analysesertifikatet (CofA) på mini-CD-en for batchspesifikke konsentrasjoner. Kalibratorområder etableres ved å teste kalibratorprøvene i minimum 20 analyser på ett system. Mini-CD Inneholder bruksanvisning for IDS-iSYS-reagenser, kontrollområder, CRY-filer og analysesertifikat.

Page 45: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Norsk Side 3/6

Påkrevd materiale som ikke medfølger System Valgfritt utstyr/materialer IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Assayprosedyre Reagenskassett Reagensene som følger med kassetten er klare til bruk, med unntak av kalibratorene. Før en ny kassett settes inn i systemet, må beholderen for magnetiske partikler blandes med en rask rotasjonsbevegelse. Unngå skumdannelse. Strekkoden avleses når kassetten er lastet på reagensbrettet. Hvis etiketten ikke kan leses av systemets strekkodeleser, finnes det en manuell prosedyre for å legge inn strekkodedataene (se brukerhåndboken for IDS-iSYS). Systemet utfører automatisk blanding av de magnetiske partiklene for alle innsatte reagenskassetter for å opprettholde homogenitet. Kassetten skal legges på reagensbrettet minst 40 minutter før du starter assayet. Hvis kassetten fjernes fra reagensbrettet, oppbevares kassetten oppreist ved 2–8 °C i mørke. Kalibratorer Kalibratorene i dette settet kan spores til produsentens arbeidskalibratorer som er fremstilt av godt renset humant PINP, der verdien er sporbar til en kommersielt tilgjengelig metode.

IDS-iSYS Intact PINP-assay kalibratorer er frysetørret. Rekonstituer umiddelbart før bruk. Tilsett 1,0 mL av destillert eller

avionisert vann til hver flaske. Bytt stopperen. La stå i 10 minutter for å rekonstituere med sporadisk, forsiktig blanding for hånd. Vær forsiktig så du unngår skumdannelse. Kalibratorene kan settes inn i systemet ved hjelp av enten prøvebegre eller kalibrator-/kontrollstativ (30/31/32). Kalibrator-/kontrollstativer: Plasser kalibratorflaskene i riktig stativ, sett dem inn i systemet, og fortsett i henhold til instruksjonene i brukerhåndboken for systemet. Kasser materialet i kalibratoren etter bruk. IKKE gjenbruk kalibratorglass som er blitt satt inn i systemet. Prøvebegre: Pipetter ca. 200 μL kalibrator i prøvebegrene, og sett dem inn i systemet. Fortsett i henhold til instruksjonene i brukerhåndboken for systemet. Kasser materialet i prøvebegrene etter bruk. IKKE hell materiale tilbake på kalibratorampullen. Hvis kalibratorer skal brukes mer enn én gang, skal kalibratorer oppbevares ved 2–8 °C i løpet av 3 timer etter rekonstituering. Ved gjenbruk av de lagrede kalibratorglassene bringes kalibratorene til romtemperatur (18–22 °C) og snus forsiktig før bruk. Kalibreringsfrekvens Ny kalibrering er nødvendig:

- Ved lasting av hvert nytt parti med kassetter. - Når kontrollverdiene ikke faller innenfor definerte områder. - Når kalibreringsintervallet på 3 dager er utløpt. - Etter service på systemet.

Verifisering av kalibreringen skjer automatisk og styres av systemet. Kalibrering av systemet De to IDS-iSYS Intact PINP -assaykalibratorene er påkrevd for å utføre justeringen av hovedkurven. Kalibratorene leveres med settet; kalibratorer fra et annet parti skal ikke brukes. Alle nivåer av IDS-iSYS Intact PINP Control Set (IS-4030) MÅ måles samtidig med kalibratorene for å validere hovedkurvejusteringen. Alle data som er nødvendig for kalibrering av kassetten finnes på mini-CD-en. Bruk kalibratornivåene A og B for å justere hovedkurven til reagensene på systemet. Kontroller om det finnes en IDS-iSYS Intact PINP-assaykassett på reagensbrettet og om det er en tilgjengelig kassetthovedkurve i databasen. Hvis dataene for kalibratorpartiet ikke er tilgjengelig i systemet, lastes dataene ved hjelp av mini-CD-en som følger med reagenssettet. Start immunassaykalibreringen på systemet i henhold til brukerhåndboken for systemet. Kalibreringen utføres ved anvendelse av 3 replikater av hvert nivå. Ett replikat kan fjernes for å innfri kalibreringskravene. Alle kontrollnivåer må også måles i duplikat for å kalibrere assayet. Verifiser og godkjenn kalibreringen i henhold til kalibreringsstatusen som vises i kalibreringsvinduet og kasser det gjenværende volumet av kalibratorer og kontroller fra prøvebrettet etter bruk.

Page 46: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Norsk Side 4/6

9. Kvalitetskontroll Bruk IDS-iSYS Intact PINP Control Set til kvalitativ kontroll. For å sikre gyldigheten av resultatene skal det måles minst tre kontroller med ulike nivåer av intakt PINP. Annet egnet kontrollmateriale kan brukes i tillegg til IDS-iSYS Intact PINP Control Set. Kontrolløsningene bør testes i (eller nær) begynnelsen av hver analyse som inneholder pasientprøver samt ved hver kalibrering i henhold til lokale og statlige forskrifter eller akkrediteringskrav og ditt laboratoriums kvalitetsrutiner. Det anbefales at kontrolløsningene rutinemessig kjøres i duplikat. Laboratoriet skal teste kontrollene minst én gang per skift. Se bruksanvisningen for IDS-iSYS Intact PINP Control Set for instruksjoner om klargjøring og håndtering. Bestemmelse av nivåer med intakt PINP i prøve Behandle prøvene i henhold til brukerhåndboken for systemet.

10. Beregning av resultater Konsentrasjonen av intakt PINP i hver prøve beregnes automatisk. Visningen av konsentrasjoner (på skjerm eller utskrift) genereres på forespørsel fra bruker. IDS-iSYS Intact PINP-assayet anvender en logistisk kurve med 4 parametre (4PL) til beregningen av konsentrasjoner med intakt PINP.

11. Måleområde (rapporterbart område) Det rapporterbare området for assayet er 2–230 ng/mL. Verdier som måles under 2 ng/mL, skal rapporteres som «< 2 ng/mL».

12. Fortynning Prøver med intakt PINP som er over det rapporterbare området, må fortynnes manuelt med lavkonsentrert human serumprøve i et forhold på 1 til 2. Resultatene for fortynnede prøver må multipliseres med en fortynningsfaktor på 2 og korrigeres iht. konsentrasjonen av den lave prøven.

13. Bruksbegrensninger - Som ved alle diagnostiske prosedyrer må resultatene tolkes i forbindelse med pasientens kliniske presentasjon og

annen informasjon som er tilgjengelig for legen. - Dette assayets ytelseskarakteristika i en pediatrisk populasjon er ikke ennå fastslått. - Hook-effekten ble testet ved bruk av analyttkonsentrasjoner på opptil 2350 ng/mL. Ingen Hook-effekt ble observert. - Heterofile antistoffer i humant serum kan reagere med reagensimmunglobuliner som interfererer med in vitro -

immunassayer6. Pasienter som rutinemessig eksponeres for animalske produkter eller produkter med animalsk

serum kan være utsatt for interferens, og avvikende verdier kan observeres. - Følgende substanser påvirker ikke IDS-iSYS Intact PINP-assayet når konsentrasjonene som vises i følgende tabell,

ligger under fastsatt terskel.

Potensielt interfererende stoff

Terskel Konsentrasjon

Lipid 2803 mg/dL

Bilirubin 200 mg/dL

Hemoglobin 500 mg/dL

Biotin 300 nmol/L

14. Forventede verdier Følgende områder ble bestemt ved hjelp av IDS-iSYS Intact PINP-assayet og er kun til informasjon. 95 %-referanseintervallet for tilsynelatende friske voksne ble beregnet med en ikke-parametrisk metode iht. veiledning fra CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Normale voksne 27,7–127,6 ng/mL (n = 150) Områdene over skal kun betraktes som veiledende; det anbefales at hvert laboratorium etablerer sitt eget forventede område basert på sin egen pasientpopulasjon.

Page 47: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Norsk Side 5/6

15. Ytelsesdata Her vises representative ytelsesdata. Resultater oppnådd på enkelte laboratorier kan variere. 15.1 Sensitivitet Blankgrense (LoB), deteksjonsgrense (LoD) og kvantifiseringsgrense (LoQ) bestemmes med veiledning fra CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation” ved bruk av 60 blank- og 120 lavnivåprøver.

Sensitivitet Konsentrasjon

LoB 0,5 ng/mL

LoD < 1,0 ng/mL

LoQ < 1,0 ng/mL

15.2 Presisjon Presisjon ble evaluert i henhold til en modifisert protokoll basert på CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”. Totalt 3 serumkontroller ble analysert ved hjelp av ett reagensparti i kvadruplikat én gang per dag i 20 dager på to systemer.

Kontroll n Gjennomsnitt

(ng/mL)

Innenfor kjøring Totalt

SD KV SD KV %

1 80 14,3 0,4 2,6 % 0,6 4,2 %

2 80 44,6 1,3 3,0 % 2,4 5,3 %

3 80 100,9 3,0 3,0 % 4,5 4,4 %

15.3 Gjenoppretting Gjenoppretting ble vurdert ved å tilsette intakt PINP til prøvene før analysering.

Prøvekonsentrasjon ng/mL

PINP tilsatt ng/mL

Målt ng/mL

Gjenoppretting ng/mL

Gjenoppretting %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Middelverdi

93 95 93 91 95 91 93

15.4 Linearitet Linearitet ble evaluert basert på CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”. Prøver som inneholdt varierende konsentrasjoner av intakt PINP ble analysert i duplikat. De resulterende middelkonsentrasjonene ble sammenlignet med forutsagte konsentrasjoner. Prøvene ble fremstilt ved å fortynne en høy pasientprøve med en lav pasientprøve før assayet.

Predikert konsentrasjon (ng/mL)

Målt konsentrasjon (ng/mL)

Variasjon

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9 -0,2 -1,6 -2,9 -2,4 3,7 0,0 0,7

- 3 0 -2 -3 -2 2 0 0

15.5 Sammenligning av metoder IDS-iSYS -assayet ble sammenlignet med et kommersielt tilgjengelig kvantitativ radioimmunassay, iht. CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. Totalt 363 prøver, utvalgt for å representere et bredt utvalg av intakt PINP -konsentrasjoner [8,7–230 ng/mL], ble analysert med hver metode. Lineær regresjonsanalyse ble utført på de komparative dataene:

n Kurve 95 % KI Krysning (ng/mL) 95 % KI Korrelasjonskoeffisient (r)

363 0,86 0,84 til 0,88 -1,9 -3,4 til -0,3 0,98

Page 48: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 februar 2017 Norsk Side 6/6

15.6 Spesifisitet Spesifisiteten ble vurdert med følgende analytter.

Analytt Kryssreaktivitet

Intakt PINP 100 %

PINP fra rotte/mus < 0,01 %

PIINP < 0,03 %

N-mid-osteokalsin < 0,02 %

16. Brukte symboler

REF Katalognummer

IVD In vitro-diagnostisk enhet

Produsent

Anvendes i samsvar med direktiv 98/79/EF

17. Bibliografi 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., ‘Heterophilic antibodies: a problem for all immunoassays’. Clin Chem, 34, 1988, pp 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tlf.: +44 191 519 0660 • Faks: +44 191 519 0760 e-post: [email protected] • www.idsplc.com Immunodiagnostic Systems USA Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tlf.: +1 877 852 6210 • Faks: +1 301 990 4236 e-post: [email protected]

Tyskland Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tlf.: +49 692 6019 0960 • Faks: +49 692 6019 0949 e-post: [email protected]

Frankrike Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tlf.: +33 140 770 450 • Faks: +33 140 770 455 e-post: [email protected]

Skandinavia Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tlf.: +45 448 400 91 e-post: [email protected]

Belgia Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tlf.: +32 425 226 36 • Faks: +32 425 251 96 e-post: [email protected]

Brasil IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasil. CEP: 05422 001 Tlf.: +55 113 740 6100 • Faks: + 55 113 740 6105 E-post: [email protected]

Page 49: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Návod k použití

IS-4000

IS-4000 v06 01 únor 2017 česky strana 1/6

1. Určené použití Pro diagnostické použití in vitro. Test IDS-iSYS Intact PINP (intaktní aminoterminální propeptid prokolagenu typu I) je určený ke kvantitativnímu stanovení intaktního PINP v lidském séru nebo plazmě na systému IDS-iSYS Multi-Discipline Automated System. Tento test se používá jako pomůcka při léčbě postmenopauzální osteoporózy.

2. Shrnutí a vysvětlení Významnou součástí procesu tvorby kosti je syntéza kolagenu typu I, který je základní organickou složkou kostní matrice. Během syntézy kolagenu se z amino- i karboxyterminální části molekuly prokolagenu uvolňují propeptidy.

1 Tyto propeptidy jsou

vylučovány do krevního oběhu. Aminoterminální propeptid prokolagenu typu I (PINP) je pravděpodobně nejspecifičtější a nejcitlivější marker tvorby kostí.

2 PINP jako marker je zejména užitečný při monitorování účinnosti anabolické léčby

osteoporózy3,4

, ale zároveň se jedná o jeden z nejlepších markerů přeměny kostní hmoty sledované při zjišťování účinnosti antiresorpční léčby.

4,5

3. Popis metody Tento test je založen na chemiluminiscenční technologii. 20 µL pacientských vzorků nebo kalibrátorů se naředí ředicí látkou a část objemu se inkubuje s biotinylovanou monoklonální protilátkou proti PINP, monoklonální protilátkou značenou akridiniem, magnetickými částicemi značenými streptavidinem a testovacím pufrem. Magnetické částice se zachytí pomocí magnetu a provede se promytí, jehož cílem je odstranit nenavázaný analyt. Poté se přidají startovací reagencie; množství vzniklého světla vyzářeného akridiniem použitým ke značení je přímo úměrné koncentraci intaktního PINP v původním vzorku.

4. Upozornění a opatření Test IDS-iSYS Intact PINP je určen pouze pro diagnostické použití in vitro, nikoli pro vnitřní použití u lidí či zvířat. Tento výrobek musí být používán přísně v souladu s pokyny uvedenými v návodu k použití. Společnost Immunodiagnostic Systems Limited (IDS) nebude zodpovídat za žádné ztráty či škody (kromě požadavků vyplývajících ze zákona) způsobené jakýmkoli způsobem, k nimž došlo v důsledku nedodržení poskytnutých pokynů. POZOR: Tato souprava obsahuje materiál živočišného původu. Zacházejte s reagenciemi z této soupravy jako s materiály potenciálně přenášejícími infekční choroby. Při skladování reagencií ze soupravy, manipulaci s nimi a jejich likvidaci musí být dodržována příslušná bezpečnostní opatření a zásady správné laboratorní praxe. Likvidace reagencií ze soupravy se musí provádět v souladu s místními předpisy.

Materiály lidského původu Materiál lidského původu použitý při přípravě tohoto výrobku byl testován pomocí stanovení schválených agenturou FDA na přítomnost protilátek proti viru lidské imunodeficience (HIV I a II), povrchového antigenu hepatitidy B a protilátek proti viru hepatitidy C a bylo zjištěno, že je negativní. Protože žádný test neumožňuje dosáhnout úplné jistoty, že infekční agens není přítomen, s reagenciemi by se mělo zacházet podle zásad úrovně biologické bezpečnosti 2. Reagencie obsahující azid sodný

Kalibrátory CAL obsahují azid sodný (NaN3) > 0,1 % (w/w) (< 1%).

Klasifikace podle CLP: Acute Tox. 4 Aquatic Chronic 3 Stand. vety o nebezpecnosti: EUH032 Uvolňuje vysoce toxický plyn při styku s kyselinami. H302 Zdraví škodlivý pri požití. H412 Škodlivý pro vodní organismy, s dlouhodobými

účinky.

Pokyny pro bezpecné zacházení: P264 Po manipulaci důkladně omyjte ruce. P270 Při používání tohoto výrobku nejezte, nepijte ani

nekuřte. P273 Zabraňte uvolnění do životního prostředí. P301/310 PRI POŽITÍ: Okamžite volejte TOXIKOLOGICKÉ

INFORMACNÍ STREDISKO/lékare. P330 Vypláchnete ústa.

5. Opatření při manipulaci Kromě kalibrátorů, které jsou lyofilizovány, jsou reagencie v soupravě připraveny k použití. Postup při rekonstituci je popsán v části Kalibrátor. Než vložíte do systému IDS-iSYS Multi-Discipline Automated System („systém“) novou kazetu, musíte rychlým promnutím mezi dlaněmi promíchat nádobku s magnetickými částicemi. Tím se resuspendují magnetické částice, které se během přepravy usadily. Dbejte, aby se v kazetě s reagenciemi nevytvořila pěna.

Page 50: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 únor 2017 česky strana 2/6

Doba použitelnosti a uchovávání reagencií Uchovávejte kazetu a kalibrátory ve svislé pozici v temnu při teplotě 2–8 °C. Nezmrazujte kazetu ani kalibrátory.

Doba použitelnosti reagencií Kazeta Kalibrátory

Před otevřením při teplotě 2–8 °C Do data použitelnosti

Kazeta po otevření při teplotě 2–8 °C

28 dní –

Kalibrátory po rekonstituci při teplotě 2–8 °C

– 28 dní

V přístroji* 7 dní 3 hodiny

*Stabilita při nepřetržitém umístění v přístroji

6. Odběr a uchovávání vzorků Test se provádí na vzorcích séra (standardní zkumavky na vzorky nebo zkumavky obsahující gel pro oddělení séra) nebo plazmy (heparin lithný, heparin sodný, heparin amonný nebo draselná sůl (K2) EDTA). Separaci vzorků je třeba provést co nejdříve po odběru. Uskladněte vzorky při teplotě -20 °C nebo nižší. Vzorky se nesmí opakovaně zmrazit a rozmrazit. Poznámka:

- Vzorky obsahující částice se musí před provedením testu odstředit. Nepoužívejte vzorky deaktivované teplem. - Aby byly případné účinky odpařování co nejmenší, je třeba provést měření vzorků, kalibrátorů a kontrolních vzorků

do 3 hodin po vložení do systému. - Každá laboratoř se musí řídit pokyny nebo požadavky místních, státních a/nebo federálních předpisů či akreditačních

organizací a stanovit vlastní pravidla ohledně stability při manipulaci se vzorky a skladování vzorků. Pokyny ohledně vhodných postupů viz dokument CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Před provedením testů zajistěte, aby vzorky, kalibrátory i kontroly měly pokojovou teplotu (18–22 °C). - Některé komerčně dostupné zkumavky pro odběr vzorků mohou mít v některých konkrétních případech vliv na

výsledky testování. - Doporučuje se postupovat podle pokynů výrobce zkumavek, zejména pokud jsou vzorky zpracovávány v primárních

zkumavkách. - Odstředěné vzorky s vrstvou lipidů na povrchu musí být přeneseny do nádobky na vzorky nebo do sekundární

zkumavky. Je třeba dbát, aby byly přeneseny pouze čiré vzorky bez lipemického materiálu.

7. Materiály Dodané materiály Kazeta s reagenciemi

MPS3 Magnetické částice potažené streptavidinem ve fosfátovém BSA pufru s azidem sodným jako konzervační látkou (< 0,1 %),

1 lahvička, 2,3 mL

CONJ Myší monoklonální protilátka proti PINP značená derivátem akridinium esteru, v pufru s obsahem hovězích a myších

proteinů a azidem sodným jako konzervační látkou (< 0,1 %), 1 lahvička, 4,95 mL

Ab-BIO Myší monoklonální protilátka proti PINP značená biotinem, v pufru s obsahem myších a hovězích proteinů a azidem

sodným jako konzervační látkou (< 0,1 %), 1 lahvička, 16,8 mL

BUF Pufr s obsahem myších a hovězích proteinů, s azidem sodným jako konzervační látkou (< 0,1 %), 1 lahvička, 4,95 mL

DIL Pufr obsahující azid sodný jako konzervační látku (< 0,1 %), 1 lahvička, 23,7 mL

Kalibrátory

CALA Lyofilizovaný pufr s obsahem BSA a purifikovaného lidského PINP s < 1,0% azidem sodným jako konzervační látkou (< 0,05 %

CALB po rekonstituci). Po 2 lahvičkách pro každou ze 2 hladin koncentrace, 1,0 mL na lahvičku.

Připravené kalibrátory mají následující jmenovitá rozmezí:

Kalibrátor Rozmezía (ng/mL)

CALA 3,0–7,0 ng/mL

CALB 130,0–165,0 ng/mL

a Tato rozmezí jsou pro výrobek typická a mají sloužit pouze jako vodítko. Hodnoty v jednotlivých šaržích se mohou lišit.

Koncentrace specifické pro konkrétní šarže jsou uvedeny v certifikátu analýzy (Certificate of Analysis, CoA) na mini disku CD. Rozmezí kalibrátoru byla stanovena testováním vzorků kalibrátoru nejméně ve 20 testovacích cyklech na jednom systému. Mini disk CD Obsahuje návod k použití reagencií IDS-iSYS, rozsahy kontrol, soubory CRY a certifikát analýzy.

Page 51: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 únor 2017 česky strana 3/6

Potřebné materiály, které nejsou součástí dodávky Systém Volitelné vybavení/materiály IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Postup testu Kazeta s reagenciemi Reagencie v této kazetě jsou připraveny k použití, s výjimkou kalibrátorů. Před vložením nové kazety do systému promíchejte nádobku s magnetickými částicemi rychlým promnutím mezi dlaněmi. Zabraňte tvorbě pěny. Po vložení kazety do zásobníku s reagenciemi se načte čárový kód. Nedokáže-li čtečka čárových kódů systému přečíst štítek, lze zadat data čárového kódu ručně (postup viz uživatelská příručka k systému IDS-iSYS). Systém automaticky zamíchá magnetické částice ve všech kazetách s reagenciemi vložených v přístroji, aby si reagencie zachovaly homogenitu. Kazetu je třeba vložit do zásobníku s reagenciemi nejméně 40 minut před zahájením testu. Po vytažení kazety ze zásobníku s reagenciemi kazetu uskladněte ve svislé poloze v temnu při teplotě 2–8 °C. Kalibrátory Kalibrátory v této soupravě mají návaznost na funkční kalibrátory výrobce připravené z vysoce purifikovaného lidského PINP, jehož hodnota má návaznost na komerčně dostupnou metodu. Kalibrátory IDS-iSYS Intact PINP jsou lyofilizované. Ihned před použitím proveďte rekonstituci. Do každé lahvičky přidejte 1,0 mL destilované nebo deionizované vody. Znovu nasaďte uzávěr. Nechte 10 minut probíhat rekonstituci a občas jemně ručně promíchejte. Dbejte, aby se nevytvořila pěna. Kalibrátory lze vložit do systému buď v nádobkách na vzorky, nebo ve stojáncích na kalibrátory/kontroly (30/31/32). Stojánky na kalibrátory/kontroly: vložte lahvičky s kalibrátory do příslušného stojánku a ten vložte do systému a poté pokračujte podle pokynů uvedených v uživatelské příručce k systému. Po použití materiál v lahvičkách s kalibrátory zlikvidujte. NEPOUŽÍVEJTE lahvičky s kalibrátory, které již byly vloženy do systému. Nádobky na vzorky: Pipetou přeneste přibližně 200 μL kalibrátorů do nádobek na vzorky a vložte je do systému. Pokračujte podle pokynů uvedených v uživatelské příručce k systému. Po použití materiál v nádobkách na vzorky zlikvidujte. NEVRACEJTE materiál do lahviček s kalibrátory. Mají-li být kalibrátory použity více než jednou, musíte je do 3 hodin od rekonstituce uložit při teplotě 2–8 °C. Při opětovném použití lahviček s uskladněným kalibrátorem nechte kalibrátory dosáhnout pokojové teploty (18–22 °C) a před použitím je mírně promíchejte převrácením. Frekvence provádění kalibrace Nová kalibrace je požadována za následujících podmínek:

- Vždy po vložení nové šarže kazet. - Když hodnoty kontrol neleží v definovaných rozmezích. - Po exspiraci kalibračního intervalu 3 dnů. - Po servisním zásahu na systému.

Ověření kalibrace je automatické a řídí je systém. Kalibrace systému K úpravě kalibrační křivky jsou potřeba dva kalibrátory IDS-iSYS Intact PINP. Tyto kalibrátory se dodávají spolu se soupravou. Je zakázáno používat kalibrátory z jiné šarže. Všechny hladiny kontrol ze soupravy IDS-iSYS Intact PINP Control Set (IS-4030) MUSÍ být měřeny současně s kalibrátory, aby byla validována úprava vzorové kalibrační křivky. Všechna data potřebná ke kalibraci kazety lze najít na mini disku CD. K úpravě nastavení vzorové křivky podle reagencií v systému použijte kalibrátory úrovně A a B. Zkontrolujte, že v zásobníku s reagenciemi je vložena kazeta IDS-iSYS Intact PINP a že v databázi je k dispozici vzorová kalibrační křivka. Pokud nejsou k dispozici údaje pro šarži kalibrátorů vloženou do systému, načtěte příslušná data z mini disku CD dodaného společně se soupravou reagencií. Zahajte kalibraci imunologického testu na systému podle pokynů uvedených v uživatelské příručce k systému. Kalibrace se provádí s použitím 3 replikátů od každé úrovně. Pro splnění požadavků na kalibraci lze jeden replikát odebrat. Také všechny úrovně kontrol musí být při kalibraci testu měřeny duplicitně. Ověřte a schvalte kalibraci podle stavu kalibrace, který se zobrazuje v kalibračním okně, a po použití zlikvidujte zbývající objem kalibrátorů a kontrol ze zásobníku na vzorky.

Page 52: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 únor 2017 česky strana 4/6

9. Kontrola kvality Ke kontrole kvality použijte soupravu kontrol IDS-iSYS Intact PINP Control Set. K zajištění validity výsledků je nutné změřit nejméně tři kontroly s různými hladinami intaktního PINP. Kromě soupravy kontrol IDS-iSYS Intact PINP Control Set lze současně použít také jiný vhodný kontrolní materiál. Kontroly by měly být testovány (minimálně) na začátku analýzy každé dávky vzorků pacienta a rovněž by měly být testovány během každé kalibrace v souladu s místními, státními a/nebo federálními předpisy či požadavky na akreditaci a postupy zajištění kvality používanými vaší laboratoří. Doporučuje se běžně analyzovat kontroly v duplikátech. Laboratoře by měly testovat kontroly alespoň jednou během každé směny. Příprava a pokyny k manipulaci jsou uvedeny v návodu k použití soupravy kontrol IDS-iSYS Intact PINP Control Set. Stanovení koncentrací intaktního PINP ve vzorcích Zpracujte vzorky podle pokynů uvedených v uživatelské příručce k systému.

10. Výpočet výsledků Koncentrace intaktního PINP v každém vzorku bude vypočítána automaticky. Na požádání uživatele se zobrazí přehled koncentrací (na obrazovce nebo se vytiskne). Test IDS-iSYS Intact PINP využívá k výpočtu koncentrací intaktního PINP 4parametrovou logistickou aproximaci křivky (4PL).

11. Rozsah měření (rozsah pro vydávání výsledků) Rozsah pro vydání výsledků testu je 2–230 ng/mL. Jakákoli hodnota nižší než 2 ng/mL bude uvedena jako „< 2 ng/mL“.

12. Ředění Vzorky s koncentracemi intaktního PINP převyšujícími rozsah vydávaných hodnot musí být ručně naředěny pomocí vzorku lidského séra s nízkou koncentrací PINP v poměru 1 ku 2. Výsledky pro naředěné vzorky musí být vynásobeny faktorem ředění 2 a korigovány na koncentraci v séru s nízkou koncentrací analytu.

13. Omezení použití - Stejně jako v případě jakýchkoli jiných diagnostických postupů musí být výsledky interpretovány společně s klinickým

obrazem pacienta a jinými informacemi, které má lékař k dispozici. - Charakteristiky tohoto testu v pediatrické populaci nebyly zjištěny. - Tzv. „hook effect“ byl testován pomocí koncentrací analytu až do 2 350 ng/mL. Nebyl pozorován žádný „hook effect“. - Heterofilní protilátky v lidském séru mohou reagovat s imunoglobuliny reagencií, které interferují s výsledky

imunologických testů in vitro.6 Pacienti, kteří se běžně setkávají se zvířaty nebo s výrobky z živočišných sér, mohou

být k této interferenci náchylní a mohou u nich být zjištěny anomální hodnoty. - Následující látky neinterferují s testem IDS-iSYS Intact PINP, pokud jsou koncentrace uvedené v následující tabulce

nižší než stanovené prahové hodnoty.

Potenciálně rušivá látka

Prahová koncentrace

Lipidy 2 803 mg/dL

Bilirubin 200 mg/dL

Hemoglobin 500 mg/dL

Biotin 300 nmol/L

14. Očekávané hodnoty Následující rozmezí byla stanovena pomocí testu IDS-iSYS Intact PINP a jsou uvedena pouze pro informaci. 95% referenční interval pro zjevně zdravé dospělé byl vypočítán pomocí neparametrické metody podle dokumentu CLSI C28-A3, “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Zdraví dospělí 27,7–127,6 ng/mL (n = 150) Výše zmíněná rozmezí lze považovat pouze za doporučené hodnoty; doporučuje se, aby si každá laboratoř stanovila vlastní rozmezí očekávaných hodnot na základě vlastní populace pacientů.

Page 53: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 únor 2017 česky strana 5/6

15. Charakteristiky Zde jsou uvedeny reprezentativní charakteristiky. Výsledky získané v jednotlivých laboratořích se mohou lišit. 15.1 Citlivost Mez slepého stanovení (LoB), mez detekce (LoD) a mez kvantifikace (LoQ) byly stanoveny podle pokynů uvedených ve směrnici CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation”, s použitím 60 slepých stanovení a 120 vzorků s nízkou koncentrací.

Citlivost Koncentrace

LoB 0,5 ng/mL

LoD < 1,0 ng/mL

LoQ < 1,0 ng/mL

15.2 Přesnost Přesnost byla hodnocena v souladu s upraveným protokolem podle dokumentu CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”. Byly analyzovány celkem 3 kontrolní vzorky séra pomocí jedné šarže reagencií ve čtyřech paralelách jednou denně po dobu 20 dnů na dvou systémech.

Kontrola n Průměr (ng/mL)

V rámci stanovení Celkem

SD VK SD VK %

1 80 14,3 0,4 2,6 % 0,6 4,2 %

2 80 44,6 1,3 3,0 % 2,4 5,3 %

3 80 100,9 3,0 3,0 % 4,5 4,4 %

15.3 Výtěžnost Výtěžnost byla vyhodnocena přidáním intaktního PINP ke vzorkům před testem.

Koncentrace ve vzorku

ng/mL

Přidáno PINP ng/mL

Naměřeno ng/mL

Výtěžnost ng/mL

Výtěžnost %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Střední hodnota

93 95 93 91 95 91 93

15.4 Linearita Linearita byla hodnocena podle pokynů uvedených v dokumentu CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”. Vzorky obsahující různé koncentrace intaktního PINP byly analyzovány ve dvou paralelách. Výsledné průměrné koncentrace byly porovnávány s očekávanými koncentracemi. Vzorky byly připraveny ředěním pacientského vzorku o vysoké koncentraci pacientským vzorkem o nízké koncentraci před testem.

Očekávaná koncentrace (ng/mL)

Naměřená koncentrace (ng/mL)

Variabilita

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9 -0,2 -1,6 -2,9 -2,4 3,7 0,0 0,7

– 3 0 -2 -3 -2 2 0 0

15.5 Srovnání metod Test IDS-iSYS název testu byl porovnáván s komerčně dostupným kvantitativním radioimunologickým testem v souladu s dokumentem CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. V rámci každé metody byl proveden test celkem 363 vzorků vybraných tak, aby reprezentovaly široké spektrum koncentrací intaktního PINP [8,7–230 ng/mL]. Na údajích ze srovnání byla provedena analýza lineární regresí:

n Sklon (směrnice) 95% CI Průsečík (ng/mL) 95% CI Korelační koeficient (r)

363 0,86 0,84 až 0,88 -1,9 -3,4 až -0,3 0,98

Page 54: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 únor 2017 česky strana 6/6

15.6 Specifičnost Specificita byla hodnocena s použitím následujících analytů.

Analyt Zkřížená reaktivita

Intaktní PINP 100 %

Krysí/myší PINP < 0,01 %

PIINP < 0,03 %

Osteokalcin N-Mid < 0,02 %

16. Použité symboly

REF Katalogové číslo

IVD Diagnostický zdravotnický prostředek in vitro

Výrobce

Označení použito v souladu se směrnicí 98/79/ES

17. Literatura 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., „Heterophilic antibodies: a problem for all immunoassays“. Clin Chem, 34, 1988, pp 27–33.

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tel.: +44 191 519 0660 • Fax: +44 191 519 0760 E-mail: [email protected] • www.idsplc.com Immunodiagnostic Systems USA Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tel.: +1 877 852 6210 • Fax: +1 301 990 4236 E-mail: [email protected]

Německo Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tel.: +49 692 6019 0960 • Fax: +49 692 6019 0949 E-mail: [email protected]

Francie Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tel.: +33 140 770 450 • Fax: +33 140 770 455 E-mail: [email protected]

Skandinávie Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tel.: +45 448 400 91 E-mail: [email protected]

Belgie Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tel.: +32 425 226 36 • Fax: +32 425 251 96 E-mail: [email protected]

Brazílie IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasil. CEP: 05422 001 Tel.: +55 113 740 6100 • Fax: + 55 113 740 6105 E-mail: [email protected]

Page 55: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IDS-iSYS Intact PINP Instrukcja użycia

IS-4000

IS-4000 v06 01 luty 2017 Polski Strona 1/6

1. Przeznaczenie Do użytku w diagnostyce in vitro. Oznaczenie IDS-iSYS Intact PINP (nienaruszony aminokońcowy propeptyd kolagenu typu I) jest przeznaczone do ilościowego oznaczania nienaruszonego PINP w surowicy lub osoczu ludzkim IDS-iSYS Multi-Discipline Automated System. Test jest przeznaczony do stosowania jako pomoc w leczeniu osteoporozy pomenopauzalnej.

2. Podsumowanie i informacje ogólne Ważnym etapem w procesie tworzenia kości jest synteza kolagenu typu I, który jest głównym składnikiem organicznym macierzy kostnej. Podczas syntezy kolagenu propeptydy są uwalniane zarówno z części amino-, jak i karboksykońcowej cząsteczki prokolagenu

1. Propeptydy te są wydzielane do krwiobiegu. Propeptyd aminokońcowy prokolagenu typu I (PINP) jest

przypuszczalnie najbardziej swoistym i czułem markerem tworzenia kości2. PINP to szczególnie użyteczny marker do

monitorowania skuteczności leczenia osteoporozy za pomocą anabolików3,4

, ale jest również jednym z najlepszych markerów obrotu kostnego do monitorowania skuteczności terapii antyresorpcyjnej

4,5.

3. Opis metody Zestaw opiera się na technice chemiluminescencji. 20 µL próbki pacjenta lub kalibratorów rozcieńcza się w rozcieńczalniku i część inkubuje się z biotynylowanym przeciwciałem monoklonalnym anty-PINP, przeciwciałem monoklonalnym znakowanym akrydyną, cząstkami magnetycznymi znakowanymi streptawidyną i buforem oznaczenia. Cząstki magnetyczne są wychwytywane za pomocą magnesu i przeprowadzany jest etap wymywania w celu usunięcia niezwiązanego analitu. Dodawane są odczynniki wyzwalające, a uzyskana emisja światła znacznika akrydynowego jest wprost proporcjonalna do stężenia nienaruszonego PINP w pierwotnej próbce.

4. Ostrzeżenia i środki ostrożności Oznaczenie IDS-iSYS Intact PINP jest przeznaczone wyłącznie do użytku w diagnostyce in vitro i nie jest przeznaczone do stosowania wewnętrznego u ludzi lub zwierząt. Produkt musi być stosowany ściśle według instrukcji podanych w niniejszej Instrukcji użycia. Firma Immunodiagnostic Systems Limited (IDS) nie będzie ponosić odpowiedzialności za jakiekolwiek straty lub szkody (z wyjątkiem przypadków określonych ustawowo), które powstały na skutek nieprzestrzegania dostarczonych instrukcji. PRZESTROGA: Niniejszy zestaw zawiera materiał pochodzenia zwierzęcego. Postępować z zestawem jak z materiałem potencjalnie zakaźnym. Należy przestrzegać środków ostrożności i dobrej praktyki laboratoryjnej podczas przechowywania odczynników z zestawu, obchodzenia się z nimi i utylizowania ich. Odczynniki z zestawu utylizować zgodnie z lokalnymi przepisami.

Materiały ludzkie Materiał pobrany od ludzi, który wykorzystano do przygotowania tego produktu, badano przy użyciu oznaczeń zalecanych przez amerykańską Agencję ds. Żywności i Leków (FDA) na obecność przeciwciał przeciw ludzkiemu wirusowi niedoboru odporności (HIV-1 i 2), antygenu powierzchniowego wirusa zapalenia wątroby typu B i przeciwciał przeciw WZW typu C, i uzyskano wyniki ujemne. Ponieważ żaden test nie gwarantuje całkowitej pewności braku czynników zakaźnych, z odczynnikami należy postępować zgodnie z poziomem 2. bezpieczeństwa biologicznego. Odczynniki zawierające azydek sodu

Kalibratory CAL zawierają azydek sodu (NaN3) > 0,1% (stężenie wagowe) (< 1%).

Klasyfikacja wg CLP: Acute Tox. 4 Aquatic Chronic 3 Rodzaj zagrozenia: EUH032 W kontakcie z kwasami uwalnia bardzo toksyczne gazy. H302 Dzia³a szkodliwie po po³knieciu. H412 Działa szkodliwie na organizmy wodne, powodując długotrwałe skutki.

Srodki ostroznosci: P264 Dokładnie umyć ręce po użyciu. P270 Nie jesć, nie pić i nie palić podczas używania produktu. P273 Unikać uwolnienia do srodowiska. P301/310 W PRZYPADKU POŁKNIECIA: Natychmiast

skontaktowac sie z OSRODKIEM ZATRUC/lekarzem.

P330 Wypłukac usta.

5. Ostrzeżenia dotyczące stosowania Z wyjątkiem kalibratorów, które są liofilizowane, odczynniki znajdujące się w zestawie są gotowe do użycia. Metodyka rozpuszczania została opisana w części procedury dotyczącej kalibratorów. Przed umieszczeniem nowej kasety w systemie IDS-iSYS Multi-Discipline Automated System (System) zawartość pojemnika z cząstkami magnetycznymi musi zostać wymieszana przez operatora przez wykonanie szybkiego ruchu obrotowego. Spowoduje to ponowne zawieszenie cząstek magnetycznych, które osiadły podczas wysyłki. Należy zachować ostrożność, aby uniknąć tworzenia piany w kasecie odczynnika.

Page 56: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 luty 2017 Polski Strona 2/6

Warunki i okres przechowywania odczynników Kasety i kalibratory należy przechowywać w pozycji pionowej w ciemności w temperaturze 2–8°C. Nie zamrażać kasety ani kalibratorów.

Okres przechowywania odczynników Kaseta Kalibratory

Przed otwarciem w temperaturze 2–8°C Do terminu ważności

Kaseta po otwarciu w temperaturze 2–8°C

28 dni ND.

Kalibratory po rozpuszczeniu w temperaturze 2–8°C

ND. 28 dni

W systemie* 7 dni 3 godziny

* Ciągła stabilność w systemie

6. Pobieranie i przechowywanie próbek W oznaczeniu należy wykorzystywać próbki surowicy (standardowe probówki do pobierania lub probówki zawierające żel do separacji surowicy) lub osocza (pobranego na heparynę litową, heparynę sodową, heparyną amonową lub EDTA potasowy (K2)). Po pobraniu próbki należy możliwie jak najszybciej oddzielić od elementów morfotycznych. Próbki należy przechowywać w temperaturze –20°C lub niższej. Unikać powtarzanych cykli zamrażania/rozmrażania próbek. Uwaga:

- Próbki z zawartością cząstek należy odwirować przed przeprowadzeniem oznaczenia. Nie wolno używać próbek dezaktywowanych termicznie.

- W celu zminimalizowania możliwego wpływu parowania pomiary próbek, kalibratorów i próbek kontrolnych należy przeprowadzić w ciągu 3 godzin po umieszczeniu w Systemie.

- Każde laboratorium powinno opracować własne zalecenia dotyczące stabilności przechowywanych próbek i postępowania z nimi, zgodne z wytycznymi akredytowanych organizacji oraz lokalnymi, krajowymi i unijnymi przepisami. Wytyczne dotyczące odpowiednich praktyk, patrz CLSI GP44-A4, “Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests”.

- Przed wykonaniem oznaczeń należy upewnić się, że próbki, kalibratory i kontrole mają temperaturę pokojową (18–22°C). - Niektóre dostępne w handlu probówki do pobierania próbek mogą w pewnych określonych przypadkach wpływać na

wyniki badania. - Zaleca się przestrzegać instrukcji producenta próbówek, zwłaszcza w przypadku przetwarzania próbki

w pierwotnej próbówce. - Odwirowane próbki z warstwą lipidową u góry muszą zostać przeniesione do naczynka na próbkę lub probówki

wtórnej. Należy zachować ostrożność, aby przenosić wyłącznie próbki klarowane bez materiału lipemicznego.

7. Materiały Dostarczane materiały Kaseta odczynnika

MPS3 Cząstki magnetyczne pokryte streptawidyną w buforze fosforanowym BSA zawierającym azydek sodu jako

substancję konserwującą (< 0,1%), 1 butelka, 2,3 mL

CONJ Mysie przeciwciało monoklonalne anty-PINP znakowane pochodną estru akrydyny w buforze zawierającym białka

bydlęce i mysie z azydkiem sodu jako substancją konserwującą (< 0,1%), 1 butelka, 4,95 mL

Ab-BIO Mysie przeciwciało monoklonalne anty-PINP znakowane biotyną w buforze zawierającym białka mysie i bydlęce

z azydkiem sodu jako substancją konserwującą (< 0,1%), 1 butelka, 16,8 mL

BUF Bufor zawierający białka mysie i bydlęce oraz azydek sodu jako substancję konserwującą (< 0,1%), 1 butelka, 4.95 mL

DIL Bufor zawierający azydek sodu jako substancję konserwującą (< 0,1%), 1 butelka, 23,7 mL

Kalibratory

CALA Liofilizowany bufor zawierający BSA i oczyszczony ludzki PINP z < 1,0% azydku sodu jako substancję konserwującą (< 0,05%

CALB po rozpuszczeniu). 2 szt. na każde 2 poziomy stężenia, 1,0 mL na butelkę.

Kalibratory są przygotowane przy następujących zakresach nominalnych:

Kalibrator Zakresa (ng/mL)

CALA 3,0–7,0 ng/mL

CALB 130,0–165,0 ng/mL

a Zakresy są typowe dla produktu i przeznaczone wyłącznie jako wskazówki. Wartości mogą się różnić między seriami.

Specyfikacje dotyczące stężeń danej serii można znaleźć w certyfikacie analizy (ang. Certificate of Analysis, CoA) na płycie mini CD. Zakresy kalibratora są ustalone przez badanie próbek kalibratorów podczas przynajmniej 20 przebiegów pracy w jednym systemie. Mini CD Zawiera instrukcję użycia odczynników IDS-iSYS, zakresy kontroli, pliki CRY oraz świadectwo analizy.

Page 57: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 luty 2017 Polski Strona 3/6

Materiały wymagane, lecz niedostarczane System Sprzęt / materiały opcjonalne IS-310400 IS-4030 IS-CC100 IS-CS100 IS-CW100 IS-CT100 IS-6010 IS-CSC105 IS-DS100

IDS-iSYS Multi-Discipline Automated System IDS-iSYS Intact PINP Control Set IDS-iSYS Cuvettes IDS-iSYS System Liquid (Syst. l) IDS-iSYS Wash Solution (Wash S) IDS-iSYS Triggers Set IDS-iSYS Cartridge Check System (CCS) Sample Cups (500 µL) D-SORB solution

IS-4035

IDS-iSYS Intact PINP Calibration Verifiers

8. Procedura wykonania oznaczenia Kaseta odczynnika Odczynniki dostarczane w kasecie są gotowe do użycia, z wyjątkiem kalibratorów. Przed załadowaniem nowej kasety do Systemu należy wymieszać pojemnik z cząstkami magnetycznymi przez wykonanie szybkiego ruchu obrotowego. Unikać tworzenia piany. Kod kreskowy jest odczytywany podczas ładowania kasety na tacę odczynnika. Jeżeli etykieta nie może być odczytana przez czytnik kodów kreskowych Systemu, istnieje procedura ręcznego wprowadzania danych kodu kreskowego (patrz podręcznik użytkownika systemu IDS-iSYS). System automatycznie przeprowadza mieszanie cząstek magnetycznych dla wszystkich załadowanych kaset odczynników, aby zachować jednorodność. Kaseta powinna być załadowana na tacę odczynników przynajmniej 40 minut przed rozpoczęciem oznaczenia. Jeżeli kaseta jest wyjęta z tacy z odczynnikami, należy przechowywać ją pionowo w temperaturze 2–8°C w ciemności. Kalibratory Kalibratory w tym zestawie są zgodne z kalibratorami roboczymi producenta, które są przygotowane z wysoce oczyszczonego ludzkiego PINP, którego wartość jest zgodna z metodą dostępną w handlu. Kalibratory oznaczenia IDS-iSYS Intact PINP są liofilizowane. Rozpuścić bezpośrednio przed użyciem. Do każdej butelki dodać 1,0 mL wody destylowanej lub dejonizowanej. Ponownie zamknąć korkiem. Pozostawić na 10 minut do rozpuszczenia, od czasu do czasu delikatnie mieszając ręką. Zachować ostrożność, aby uniknąć tworzenia piany. Kalibratory można umieścić w systemie, korzystając z pojemników na próbki lub statywów na kalibratory/próbki kontrolne (30/31/32). Statywy na kalibratory/kontrole: umieścić fiolki z kalibratorami w odpowiednim statywie, włożyć do systemu i kontynuować zgodnie z instrukcjami w podręczniku użytkownika Systemu. Odrzucić materiał we fiolkach kalibratora po użyciu. NIE UŻYWAĆ ponownie fiolek kalibratorów, które zostały umieszczone w systemie. Naczynka na próbkę: Pipetować około 200 μL kalibratorów do naczynek na próbkę i umieścić w Systemie. Kontynuować zgodnie z instrukcjami w podręczniku użytkownika Systemu. Odrzucić materiał w naczynkach na próbkę po użyciu. NIE zwracać materiału do fiolki kalibratora. Jeżeli kalibratory mają być używane więcej niż raz, po rozpuszczeniu należy je umieścić w temperaturze 2–8°C w ciągu 3 godzin od rozpuszczenia. Po ponownym użyciu przechowywanych fiolek kalibratorów, doprowadzić kalibratory do temperatury pokojowej (18–22°C) i delikatnie odwrócić przed użyciem. Częstotliwość kalibracji Nowa kalibracja jest wymagana:

- Po załadowaniu każdej nowej serii kaset. - Gdy wartości kontroli nie wypadają w zdefiniowanych zakresach. - Po upływie 3 dni od poprzedniej kalibracji. - Po serwisie Systemu.

Weryfikacja kalibracji jest automatyczna i zarządzana przez System. Kalibracja systemu Do wykonania dopasowania krzywej wzorcowej wymagane jest użycie dwóch kalibratorów oznaczenia IDS-iSYS Intact PINP. Kalibratory są dostarczane w zestawie i nie wolno używać kalibratorów z innej serii. Aby było możliwe dostosowanie krzywej wzorcowej, wszystkie poziomy z zestawu kontrolnego IDS-iSYS Intact PINP Control Set (IS-4030) MUSZĄ zostać oznaczone w tym samym czasie co kalibratory. Wszystkie dane wymagane do kalibracji kasety można znaleźć na dysku mini CD. Użyć kalibratorów poziomu A oraz B do dostosowania krzywej wzorcowej do odczynników znajdujących się w Systemie. Sprawdzić obecność kasety oznaczenia IDS-iSYS Intact PINP na tacy odczynników oraz dostępność krzywej kalibracyjnej kasety w bazie danych. Jeżeli dane dla serii kalibratorów nie są dostępne w Systemie, należy wprowadzić dane używając dostarczonego z zestawem odczynnika dysku mini CD. Rozpocząć kalibrację oznaczenia immunologicznego w Systemie zgodnie z podręcznikiem użytkownika Systemu. Kalibrację przeprowadza się z użyciem 3 powtórzeń na każdy poziom. W celu spełnienia wymogów kalibracji można usunąć jedno powtórzenie. W celu kalibracji oznaczenia należy również oznaczyć dwukrotnie wszystkie poziomy kontroli. Zweryfikować i zatwierdzić kalibrację zgodnie ze statusem kalibracji wyświetlanym w oknie kalibracji oraz odrzucić pozostałą objętość kalibratorów i kontroli z tacy próbki po użyciu.

Page 58: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 luty 2017 Polski Strona 4/6

9. Kontrola jakości Do kontroli jakości należy używać zestawu kontrolnego IDS-iSYS Intact PINP Control Set. W celu zapewnienia uzyskiwania prawidłowych wyników należy dokonać pomiaru co najmniej trzech próbek kontrolnych o różnych stężeniach nienaruszonego PINP. Jako uzupełnienie zestawu kontrolnego IDS-iSYS Intact PINP Control Set można wykorzystać inny stosowny materiał kontrolny. Kontrole należy testować na początku (w przybliżeniu) każdej serii zawierającej próbki od pacjenta, a także podczas każdej kalibracji zgodnie z wymogami przepisów lokalnych, stanowych i (lub) federalnych bądź akredytacji oraz procedurami zapewnienia jakości danego laboratorium. Zaleca się, aby kontrole były rutynowo oznaczane dwukrotnie. W laboratoriach oznaczenia próbek kontrolnych powinny być wykonywane co najmniej raz na każdej zmianie. Instrukcje dotyczące przygotowania oraz postępowania zawiera instrukcja użycia zestawu IDS-iSYS Intact PINP Control Set. Określenie stężeń nienaruszonego PINP w próbce Podczas wykonywania oznaczeń próbek należy postępować zgodnie z instrukcjami zawartymi w podręczniku użytkownika Systemu.

10. Obliczanie wyników Stężenie nienaruszonego PINP w każdej próbce jest obliczane automatycznie. Wyświetlanie stężeń (ekran lub wydruk) jest generowane na żądanie. W oznaczeniu IDS-iSYS Intact PINP do obliczenia stężeń nienaruszonego PINP wykorzystywane jest dopasowanie 4-parametrowej krzywej logistycznej (4PL).

11. Zakres pomiaru (zakres sprawozdawczy) Zakres miarodajny oznaczenia wynosi 2–230 ng/mL. Każdą wartość poniżej 2 ng/mL należy zgłaszać jako „< 2 ng/mL”.

12. Rozcieńczenie Próbki o stężeniu nienaruszonego PINP przekraczającym zakres pomiaru należy rozcieńczyć ręcznie w stosunku 1 do 2 próbką ludzkiej surowicy o niskim stężeniu. Wyniki próbek rozcieńczonych należy mnożyć razy 2 i korygować o stężenie dla próbki o niskim stężeniu.

13. Ograniczenia zastosowania - Podobnie jak w przypadku każdego testu diagnostycznego, wyniki zawsze należy interpretować w odniesieniu do

stanu klinicznego pacjenta i innych danych dostępnych lekarzowi. - Nie oceniano działania tego oznaczenia podczas badania próbek od dzieci. - Efekt wysokiej dawki zbadano używając stężeń analitu do 2350 ng/mL. Nie zaobserwowano efektu wysokiej dawki. - Przeciwciała heterofilne obecne w ludzkiej surowicy mogą reagować z immunoglobulinami znajdującymi się

w odczynnikach, powodując zakłócenia w immunologicznych oznaczeniach in vitro6. Wyniki mogą być zaburzone

zwłaszcza u pacjentów rutynowo narażonych na kontakt ze zwierzętami lub produktami na bazie surowicy zwierząt. Może to prowadzić do nietypowych wartości oznaczenia.

- Następujące substancje nie zakłócają działania oznaczenia IDS-iSYS intact PINP przy stężeniach niższych od wartości progowych podanych w tabeli.

Czynniki potencjalnie wpływające na wynik testu

Wartość progowa Stężenie

Lipid 2803 mg/dL

Bilirubina 200 mg/dL

Hemoglobina 500 mg/dL

Biotyna 300 nmol/L

14. Oczekiwane wartości Poniższe zakresy ustalono z wykorzystaniem oznaczenia IDS-iSYS Intact PINP. Mają one charakter wyłącznie pomocniczy. 95% przedział referencyjny dla zdrowych dorosłych obliczono metodą nieparametryczną przestrzegając wytycznych z dokumentu CLSI C28-A3 “Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory”.

Zdrowe osoby dorosłe 27,7–127,6 ng/mL (n = 150) Przedstawione powyżej zakresy należy traktować wyłącznie jako wytyczne. Zaleca się, aby każde laboratorium opracowało własny oczekiwany zakres na podstawie swojej populacji pacjentów.

Page 59: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 luty 2017 Polski Strona 5/6

15. Dane dotyczące działania Pokazano wyłącznie przykładowe dane dotyczące działania. Wyniki w danym laboratorium mogą być odmienne. 15.1 Czułość Zgodnie z normą CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation” określono granicę próbki ślepej (LoB), granicę detekcji (LoD) i granicę oznaczalności (LoQ) wykorzystując 60 próbek ślepych oraz 120 próbek o niskim stężeniu.

Czułość Stężenie

LoB 0,5 ng/mL

LoD < 1,0 ng/mL

LoQ < 1,0 ng/mL

15.2 Precyzja Precyzja została oceniona według zmodyfikowanego protokołu, opartego na wytycznej CLSI EP-5A2, “Evaluation of Precision Performance of Quantitative Measurement Methods”. Łącznie 3 kontrole surowicy oznaczano na dwóch systemach przez 20 dni w czterech powtórzeniach na dobę z wykorzystaniem jednej serii odczynników.

Kontrola n Średnia (ng/mL)

W obrębie cyklu Łącznie

SD CV SD CV%

1 80 14,3 0,4 2,6% 0,6 4,2%

2 80 44,6 1,3 3,0% 2,4 5,3%

3 80 100,9 3,0 3,0% 4,5 4,4%

15.3 Odzysk Odzysk oceniono dodając nienaruszony PINP do próbek przed oznaczeniem.

Stęż. próbki ng/mL

Dodano PINP ng/mL

Zmierzony ng/mL

Odzysk ng/mL

Odzysk %

44,3 44,3 33,5 33,5 37,3 37,3

35,3 74,8 35,3 74,8 35,3 74,5

77,1 115,1 66,4 101,5 70,8 105,2

32,8 70,9 32,9 68,0 33,5 67,9

Średnia

93 95 93 91 95 91 93

15.4 Liniowość Liniowość oceniono na podstawie wytycznych CLSI EP-6A, “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”. Próbki o różnych stężeniach PINP oznaczano w dwóch powtórzeniach. Uzyskane średnie stężenia porównywano ze stężeniami przewidywanymi. Przed oznaczeniem próbki przygotowano przez rozcieńczenie próbek od pacjentów o wysokim stężeniu próbkami od pacjentów o niskim stężeniu.

Stężenie prognozowane (ng/mL)

Stężenie zmierzone (ng/mL)

Zmienność

ng/mL %

5,8 30,5 55,3 80,0 104,8 129,6 154,3 179,1 203,8

7,6 31,5 55,0 78,4 101,9 127,2 158,0 179,0 204,6

1,8 0,9

–0,2 –1,6 –2,9 –2,4 3,7 0,0 0,7

- 3 0

–2 –3 –2 2 0 0

15.5 Porównanie metod Oznaczenie IDS-iSYS porównano z dostępnym w handlu ilościowym testem immunologicznym zgodnie z wytycznym CLSI EP-9A2, “Method Comparison and Bias Estimation Using Patient Samples”. Za pomocą każdej metody oznaczono 363 próbki dobrane tak, aby reprezentowały szeroki zakres stężeń nienaruszonego PINP [8,7–230 ng/mL]. Dane porównawcze zostały poddane analizie regresji liniowej:

n Nachylenie 95% CI Punkt przecięcia (ng/mL) 95% CI Współczynnik korelacji (r)

363 0,86 od 0,84 do 0,88 –1,9 od –3,4 do –0,3 0,98

Page 60: IDS-iSYS Intact PINP Instructions For Use IS-4000 v06 - IDS-iSYS...for monitoring the efficacy of osteoporosis therapy ... MPS3 Magnetic particles coated with streptavidin in phosphate

IS-4000 v06 01 luty 2017 Polski Strona 6/6

15.6 Swoistość Swoistość oznaczono z następującymi analitami.

Analit Reaktywność krzyżowa

Nienaruszony PINP 100%

Szczurzy/mysi PINP < 0,01%

PIINP < 0,03%

Osteokalcyna N-Mid < 0,02%

16. Użyte symbole

REF Numer katalogowy

IVD Wyrób do diagnostyki in vitro

Producent

Oznaczenie nadane zgodnie z dyrektywą 98/79/WE

17. Piśmiennictwo 1. Risteli J, Risteli L 2006 Products of bone collagen metabolism. In: Seibel MJ, Robins SP, Bilezikian JP, editors.

Dynamics of bone and cartilage metabolism. Principles and clinical applications, second edition. London: Academic Press p391-405.

2. Melkko J, Kauppila S, Niemi S, Risteli L, Haukipuro K, Jukkola A, Risteli J 1996 Immunoassay for intact amino-terminal propeptide of human type I procollagen. Clin Chem 42:947-954.

3. Chen P, Satterwhite JH, Licata AA, Lewiecki EM, Sipos AA, Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD response to teriparatide in postmenopausal women with osteoporosis. J Bone Miner Res 20:962-970.

4. Finkelstein JS, Leder BZ, Burnett SA, Wyland JJ, Lee H, De la Paz AV, Gibson K Neer RM 2006 Effects of teriparatide, alendronate, or both on bone turnover in osteoporotic men. J Clin Endocrinol Metab 91:2882-2887.

5. Nenonen A, Cheng S Ivaska KK, Alatalo SL, Lehtimaki T, Schmidt-Gayk H, Uusi-Rasi K, Heinonen A, Kannus P, Sievanen H, Vuori I, Vaananen HK, Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment: comparison with other markers of bone turnover. J Bone Miner Res 20: 1804-1812.

6. Boscato, LM. and Stuart, MC., „Heterophilic antibodies: a problem for all immunoassays”. Clin Chem, 34, 1988, pp 27–33

Immunodiagnostic Systems Limited, 10 Didcot Way, Boldon Business Park, Boldon, Tyne & Wear, NE35 9PD, UK Tel.: +44 191 519 0660 • Faks: +44 191 519 0760 e-mail: [email protected] • www.idsplc.com Immunodiagnostic Systems USA Immunodiagnostic Systems (IDS) Inc, 948 Clopper Road, Gaithersburg, MD 20878, USA. Tel.: +1 877 852 6210 • Faks: +1 301 990 4236 e-mail: [email protected]

Niemcy Immunodiagnostic Systems GmbH (IDS GmbH), Rahmhofstraße 2-4, 60313 Frankfurt am Main Tel.: +49 692 6019 0960 • Faks: +49 692 6019 0949 e-mail: [email protected]

Francja Immunodiagnostic Systems SAS, 153 Avenue D’Italie, 75013 Paris, France Tel.: +33 140 770 450 • Faks: +33 140 770 455 e-mail: [email protected]

Skandynawia Immunodiagnostic Systems Nordic a/s (IDS Nordic a/s), International House, Center Boulevard 5, 2300 København S, Denmark Tel.: +45 448 400 91 e-mail: [email protected]

Belgia Immunodiagnostic Systems S.A., Rue Ernest Solvay 101, 4000 Liège, Belgium Tel.: +32 425 226 36 • Faks: +32 425 251 96 e-mail: [email protected]

Brazylia IDS Brasil Diagnósticos Ltda. Rua dos Pinheiros, 610 - conj 41 Edificio Win Work Pinheiros São Paulo, SP Brasil. CEP: 05422 001 Tel.: +55 113 740 6100 • Faks: + 55 113 740 6105 e-mail: [email protected]