high speed mass spectrometer as a gas chromatography detector

Upload: anks7993

Post on 03-Jun-2018

222 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/12/2019 High Speed Mass Spectrometer as a Gas Chromatography Detector

    1/16

    Section

    V

    Paper 16

    u. c

    A

    HIGH SPEED M SS SPECTROMETER

    S G S

    CHROM TOGR PHY DETECTOR

    IN THE DETERMIN TION

    O

    CRUDE OIL COMPOSITION

    V. F Gaylor

    -

    C.

    N.

    Jones

    **

    -

    J.H.Landerl

    ***

    -

    E.

    C. Hughes

    ****

    Abstract,

    A

    rapid crude oi l analysis procedure providing

    boiling range data to 725 F and naphtha composition

    in less than two hours has been devised. Whole crude is

    charged to a gas chromatography unit comprising a strip -

    per column and an analytical column.

    A

    t ime-of-f l ight mass spectrometer is used as an auxi l iary

    detector for a conventional gas chromatography appa-

    ratus. Continuous spectral monitoring of the column ef-

    f luent permits identi f ication of single chromatographic

    peaks. The spectrometers

    1

    microsecond scan rate is

    particu larly valuable for studying distribu tion of indiv idual

    compounds in non-homogeneous multicom pone nt chrom-

    atographic peaks.

    The mass spectrometer detector is used to characterize

    chromatographic separations of virgin naphthas boi l ing

    to 400 . Detailed spectra l analyses pro ve that the poor-

    ly resolved peaks and shoulders obtained fro m a packed

    chromatographic column could be related to hydrocarbon

    type composition.

    A

    simple analyt ical procedure for estimating cycloparaf-

    f in content from chromatograms of crude oi l and reform-

    ing feed results in a method that is suitable for routine

    refine ry use.

    Rsum.

    Larticle dcrit une nouvelle mthode danalyse

    rapide du ptrole brut, indiquant les points dbullition

    jusqu 725

    F

    et la composit ion des naphtas en moins

    de deux heures. On introduit le brut entier dans une

    unit de chromatographie gazeuse comprenant une co-

    lonne de sparation et une colonne danalyse.

    On monte un spectromtre de masse

    temps de rponse

    ultra rapide uti l is comme dtecteur auxi l iaire, sur un

    appareil classique de chromatographie gazeuse, Le con-

    trle spectral continu de l eff luent de la colonne permet

    diden tifier chaque pic chroma tographique simple et, grce

    l

    vitesse de balayage du spectromtre

    100

    micro-

    secondes),

    l

    est possible de dist inguer la rp art i t ion des

    diffrents individus dans les pics complexes non homo-

    gnes.

    On uti l ise le dtecteur de spectromtrie de masse pour

    caractriser les fractions de naphtas vierges ayant jus

    qu

    400

    o F 205 C) de po int dbu llition initialemen t

    spares par chromatographie. Des analyses spectrales

    dtai l les montrent qu i l est possible de rel ier les pics

    mdiocrement

    rsolus

    et les sai l l ies peu diffrencies

    fournis par une colonne chromatographique garnissage

    une composition type dhydrocarbure.

    Une mthode danalyse sim ple perm ettant dvaluer la

    teneur en cycloparaffines des bruts et des charges de

    reforming, partir de leurs chromatogrammes, offre un

    outi l appropri aux besoins usuels des raff ineries.

    Introduction

    The numerous applications of gas chromatography

    to analysis of petroleum streams are well known.

    Widespread, routine us s have, however, been

    gener ally limited to the rel atively simple Ci through

    C hydrocarbons. Outstanding successes in analyz-

    ing more complex mixtures have been achieved,

    using specialized techniques and equipment, but

    often involve increased costs and analysis time.

    * Authors Biographies v ide las t page

    Methods which retain the speed and simplicity

    features of the original chromatographic technique

    may require a new approach to interpretation of

    complex chromatograms. This paper describes such

    an approach t o crude oil analysis, based

    on

    research

    with a mass spectrometer used as a chromatographic

    detector.

    Previously reported applications of gas chromato-

    graphy to quantitative crude oil analyses were

    limited in scope. WebbI3 injected whole cr ude oil

    directly into an analytical column and measured

    Ci

    through S hydr,ocarbons. Martin and WintersQ

    201

  • 8/12/2019 High Speed Mass Spectrometer as a Gas Chromatography Detector

    2/16

    and Rysse1berget2) used a short precolumn to retain

    heavy ends and quantitatively determined Cz

    through C7 and C2 through C5 hydrocarbons, respec-

    tively. A temperature-programmed gas chromato-

    graphy column was used by Barras and Boyle') for

    predicting jet fuel freezing point from crude oil

    analysis.

    The temperature programming technique vastly

    extends the useful molecular weight range of a single

    chromatographic analysis. Calculation of crude oil

    boiling range distribution from a temperature-pro-

    grammed chromatogram seemed feasible. Eggertsen,

    Groennings and Holst5) conver ted programmed

    chromatographic analyses of petrole um fractions to

    distillation-type curves which closely approximated

    a 60-plate distillation; Barras and Boyle1) used

    analysis on an uncoated Chromosorb column to

    predict yield points on a l-plate distillation column.

    While distillation-type curves we re of obvious

    value, determination of chemical composition from

    crude oil chromatograms was equally desirable.

    Chromatographic procedures were designed to

    optimize both molecular weight range and s epara tion

    efficiency in a single analy sis. Normal tricosane was

    eluted in abou t 35 minutes while cons iderab le resolu-

    tion of low molecular weight hydrocarbons was

    retained.

    As

    many as 50 distinct peaks and shoulders

    were obtained in th e C3 through Cii range, constit ut-

    ing a fingerprint profile qualitatively useful for

    characterizing crude oil. Quantitative interpretation

    in terms of single compounds or over-all hydrocar-

    bon type distribution require d some knowleld,ge of

    composition of each single peak.

    Identificationof the C3 through C7 peaks presen ted

    no particular problems. Identification and determina-

    tion of single compounds in the Cs-Cii range is more

    difficult because of increasing complexity of composi-

    tion. Desty, Goldup and Swanton3) identified more

    than half of 122 peaks resolved from the C3-Cn

    portion of the American Petroleum Institute 's Ponca

    City crude in a 20-hour run on a coated capillary

    column. Polgar, Holst and Groennings ) accom-

    plished complete analys is Of C7 and CS alkanes, cyclo-

    pentanes and cyclohexanes with two successive an-

    alyses on a 300-feet capillary column totalling about

    3 hours;

    80

    O o

    of the possible compounds were quanti-

    tative ly accounted for a s individuals. Lindeman and

    Anni@) use'd a magne tic-type mass spe ctrometer

    for complete qualitati ve and quanti tative analys is of

    chromatographic pea ks of a

    440

    F

    end-point naphtha

    on a packed column; over 60 single compounds were

    identified and general structure assignments wer

    report ed for many Ci0 and Cii compounds.

    It was unrealistic to expec t that single compoun

    ana lyses could be achieved chromatographically from

    the

    30

    poorly separated peaks representing th

    hundreds of

    c

    through Cii compounds likely presen

    in crude oil. Consistency of the peak patt ern, how

    ever, and large variations in pea k size distribution

    from sample to sample suggested that gross hydro

    carbon -type distribution a nalyses might be compute

    direc tly from the chromatogram. Peak height measu

    rements were preferred for this purpose since appa

    rent separation even between discrete peaks wa

    usually poor. Qualitative composition analysis a

    each peak maximum was thus desired. Time-O

    Flight mass spectromete r, which produces 10,00

    spectra per second, had the required speed an

    sensitivity for monitoring chromatographic colum

    effluent )

    4

    6 .

    Photographic recording of spe ctr

    permits qualita tive composition determination at an

    single instant during elution of the chromatographi

    peak. Mass spectra obtained in this way we re use

    to estimate rela tive hydrocarbon-type distribution a

    each peak maximum.

    quipment and Procedures

    A commercial gas chromatography unit (Mod

    K-2 Kromo-Tog) employing a filament-type therma

    conductivity detector,and equipped with temperatur

    programmer, flow controller and precolumn assembly

    was used (Burrell Corporation, Pittsburgh, Pennsyl

    vania) . Primary elements of t he chromatographi

    equipment ar e adequately described in the manufac

    turer's literature. Details of the conventional flow

    system are depicted in Figure 1 and are largely self

    explanatory. The hairpin-shaped analytical colum

    was 250 cm. long and 5 cm. I.D. The V4-inch I.D. pre

    column was one foot long and contained packin

    identical

    to

    the analytical column. Primary helium

    carrier was directed through the precolumn durin

    sample injection and maintained until desired por

    tions of crude oil had been flushed into the analytic a

    column. Carr ier flow was th en switched to precolum

    by-pass position. Crude oil heav y ends trapped in th

    precolumn were backflushed and vented to atmos

    phere at the flash vaporizer inlet port.

    202 VI1

  • 8/12/2019 High Speed Mass Spectrometer as a Gas Chromatography Detector

    3/16

    FIGUR

    FLOW

    S Y S T M

    FIGUR

    FLOW

    S Y S T M

    Two di fferent column packings an'd corresponding

    experim ental parameters wer e used, Table I. Highest

    molecular weight range w as achieved with Procedure

    II while Procedure I was p refe rred for resolu tion of

    low molecular weight hydrocarbons. Resolution of

    Ci-Cs hydrocarbons was optimized in either proce-

    dure by holding column temperature a t 25' C until

    n-C5 was eluted.

    Column packings w ere tempera ture stabilized be-

    fore use b y heating for 24 hours under helium flow.

    Columns I and II were programmed to 250'C and

    300' C, respec tively, without detectable liquid phase

    bleeding. Maximum molecular weight determined

    was Cis for Procedure I and Ce3 for Procedure II.

    Chromatograms were recorde d on a 1.0 millivolt

    Potentiometer recoilder (Minneapolis-Honeywell Reg.

    Co., Philadelphia, Pennsylvania), equipped with Disc

    integrator (Disc Instrument Company, Santa Ana,

    California) for peak area dete rmination .

    The Time-Of-Flight mass ~pectrometer'~)

    5)

    Model 12-100 (The Bendix Corporation, Cincinnati,

    Ohio), was attached to the chromatographic equip-

    TABLE I

    CHROMATOGRAPHIC OPERATING PAR-TERS

    Procedure I Procedure

    II

    Column Packing

    Analytical Column Temperature:

    Initial Temperature.

    .........................

    Initial Temperature Hold.

    Heating Ra te .. .............................

    Final Temperature ..........................

    Precolumn Temperature.

    Precolumn

    Flush

    Time

    .........................

    Flow

    Rate.. .................................

    Sample Size

    5.0 wt. yoApiezon L

    on

    30-60

    mesh ChromosorbP

    25

    C

    3.0

    minutes

    7 C p r

    min

    250

    C

    200 c

    3 minutes

    5 OB per

    min

    .o12 ml.

    1.0 wt. yo D.C. 710 Fluid

    and 3.0 wt. Apiezon

    L

    on 40 50 mesh

    Chromosorb

    P

    25

    C

    2.0 minutes

    11

    C

    per min.

    300 c

    300

    c

    1

    minute

    230

    cc

    per min.

    .O24 ml.

    ment at the detector vent (Figure

    1).

    The variable

    leak (Model 9101-M, Granville-Phillips Company,

    Pullman, Washington) was adjus ted to raise internal

    spectrometer pressure to 2X 10 mm. mercury with

    pure helium effluent. Temperature of the

    2 5 VE

    I.D.

    line connecting the leak to the spectrometer inlet

    port was maintained at 170' C and the spectrom eter

    source chamber was hea ted to about 175OC. The

    ionizing source was ope rated at

    2.5

    amperes filament

    current, .125 microamperes trap current and O e.v.

    energy level.

    Spectra were viewed on th e screen of a Type 541

    A

    oscilloscope, equipped with Type CA Preamplifier

    (Tektronix, Incorporated, Cleveland, Ohio) and pho-

    tographe8dwith a Polaroid camera. Relative mass line

    intensities were

    estimated by use of a Microcard

    Reader (Microcard Reader Corporation, Wes t Salem,

    Wisconsin).

    V/16 2 3

  • 8/12/2019 High Speed Mass Spectrometer as a Gas Chromatography Detector

    4/16

    Discussion

    boilin,g range yield point calibration. The columns

    described separated paraffins essentially by boiling

    point and exhibited only slight selectivity for naph-

    thenes and aromatics. Naphthenes were retarded,

    Normal paraffin peaks in the crude oil chromato- relative to paraffins, by about one peak width and

    grams were readily identified by either relative size aromatic selectivity was only slightly greater.

    or elution time (Figure 2) and were employed for

    Chromatographic analyses of carefully fractionated

    Boiling Range Distribution Analysis

    FIGURE

    CR UDE

    O I L

    C H R O M A T O G R A M P R O C E D U R E

    distillates of both highly aromatic and naphthenic

    crude oils showe d that error due to column selectivi ty

    was negligible. Consequently, norm al paraffin boiling

    point-elution time plots were used for yield point

    calibration.

    Liquid volume per cent distillation yields were

    oalculated .directly from summed peak areas. Mess-

    nerlO)showed that thermal conductivity molar area

    response for hydrocarbons is a function of both

    molecular weight effects. Normal paraffin area

    for single compounds thus requires accurate know-

    ledge of re lat ive response of each compound. Aver-

    aging structure effects seemed permissible, however,

    when summing >areas of peaks composed of many

    hydrocarbons likely including most possible struc-

    tures.

    Quantitative oalibration with normal paraffin

    standa rds was a rea dy means of compensating for

    molecular weight and structure. Quantitative analysis

    response per un it liquid volume sampled also tended

    to average out structure effects at each carbon

    number level (Figure 3 . Liquid volume per cent

    yields were therefore calculated from arealvolume

    constants experime ntally de termined for each C5-

    C 3 normal paraffin. Cons tants for C4 and lighter

    NORMAL PARAFFINS

    OIS

    FWUFFINS

    xNAPHTHENES

    AROMATICS

    FIGURE

    EFFECT OF H Y D R O C A R B O N T Y P E O N

    LIQUID

    V O L U M E R ES P O N S E

    paraffins were obtained by extrapolating linear

    molar response curves.

    2 4

    Vi

    6

  • 8/12/2019 High Speed Mass Spectrometer as a Gas Chromatography Detector

    5/16

    Composition o f

    L i gh t N a p h t h a a nd O f f - G a s

    Rresolution of low boiling hydrocarbons was suffi-

    cient to obtain considerable composition information

    from the crude oil chromatogram (Figure

    4 .

    Complete

    separa tion of propane, isobutane and normal butane

    was not achieved; resolution was, howeve r, adequate

    for estimating concentration s with reasonable

    degree of accuracy.

    Octane number of light naphtha boiling 55-220

    F

    was estimated from relative concentratio ns of normal

    paraffins and summed isoparaffins, naphthenes and

    benzene. Weighted averag e blending octane numbers

    of 45.0 for normal paraffins and 79.5 for isoparaffins,

    naphthenes and benzene, calculated from pure

    compound octane numbers determined by the Ameri-

    can Petroleum Institute s Projec t 45, were used for

    octane numbers prediction. Values estimated in this

    way were at least

    as

    accurate a s values calculated

    similarly from single compound analyse s.

    FIGURE

    P A R T I A L C R UD E O I L C H R O M A T O G R A M

    P R O C ED U R

    E

    LE G E N D : 1 M E T H A N E N O N - C O N D . G A S E S

    2.

    E T H A N E

    3

    P R O P A N E

    4

    I S O - B U T A N E

    5 N - B U T A N E

    6 . I S O - P E N T A N E

    7. N - P E N T A N E

    Peak Label l ing

    As many as 30 major peak s and shoulders were

    obtained i n the n-CI to n-Ci1 portion of the chromato-

    gram. Peak pattern could be considerably altered by

    relatively small changes in column packing, as, for

    example, Column I

    vs

    Column II (Figures 2 and 4 .

    However, with identical ana lytical procedures, the

    over-all peak patte rn was consisten t for many differ-

    ent crude oil types. Most of the composition analyses

    wer e obtained from Column though the same pro-

    cedure s and principles w ere also effective for Column

    II.

    Chromatographic peaks eluted between n-Cs and

    n-Ci1 w ere labelled for da ta hand lingpurposes. Major

    group numbers were assigned to each series of pe aks

    eluted between successive normal paraffins. Group

    8

    peaks included all peaks eluted between n-C7 and

    n-CS, Group 9 peaks were eluted between n-Cs and

    n-Cg, etc. (Figure

    4 .

    ingle peaks in each group were

    VA

    205

  • 8/12/2019 High Speed Mass Spectrometer as a Gas Chromatography Detector

    6/16

    further identified by elution order with in the group.

    A partia l relative retention time PRRT wa s calculat-

    ed for each peak:

    I

    where Peak X is eluted between n-Pi and n-Pn.

    The peak number identification included both group

    number and PRRT anmdhus .defined exact location on

    the chromatogram. For example Peak 8-50 was

    eluted exact ly halfway between n-C7 and n-Ca.

    Group number does not necessarily define carbon

    number. The peaks immediately preceding a normal

    paraffin of ca rbon number N usually contained both

    and N 1 naphthenes and aromatics and isoparaf-

    fins largely of N carbon number.

    i l l

    a s s

    Spectrometer

    nalyses

    1

    Useful mass spectr a

    of

    single peaks were achieved

    only by minimizing lag time between chromatogra-

    phic and spectrometer units. Careful attention to

    length and temperature of

    low

    pressure inlet lines

    produced virtually simultaneous response from both

    ddtection systems. Over-all spectrometer speed and

    sensitivity were sufficient to permit qualitative

    characterization of small shoulders immediately fol-

    lowing or preceding major peaks. For example

    spectra of Peak 9-.69 show ed the peak like ly

    contained a large amount of 3-methyl octane while

    the 9-.75 shou lder wa s largely composed

    of

    ethyl

    benzene and one or more Cg naphthenes Figure 5.

    m

    98

    m e

    n

    I

    t

    z

    39

    4

    43 55 57 67 69 7 81

    8

    9 97 1 6 126

    mie

    m

    e

    FIGURE

    M A S S S P EC T RA L C H A R A C T E R I Z A T I O N

    OF

    A D J A C E N T

    P E A K S

    Contribution of 3-methyl octane to spectra of t he

    9-.75 shou lder was small. Similarly the naphthe nic

    character of the 10 .45 shoulder was not obscured

    by paraffincontribution from the larger

    10 .61

    peak.

    Spectral comparisons of most poorly separated

    peaks showed chromatographic column efficiency to

    be a good deal better than was superficially apparent.

    Expected non-homogenity was detected and often

    explained occasional anomalous peaks resulting from

    large variations in single compound concentration.

    Xylene and Cg naphthe ne for example wer e often

    detec ted spectr ally in the leading edge of n-Cg peaks

    Figure 6 through not evident chromatographically

    A

    shoulder on the n-Cg peak sometimes observed in

    206 V 16

  • 8/12/2019 High Speed Mass Spectrometer as a Gas Chromatography Detector

    7/16

    consistent with boiling point relations between

    branched and normal paraffins of the sam e carbon

    Identification of single compounds, though possible

    in many instances, was not a primary aim. Mass

    spectra were used semi-quantitatively to estimate

    hydrocarbon-typedistribution-ateach pea k and shoul-

    number.

    h

    SAMPLE A n Cg

    PEAK

    FIGURE

    der maximum. Repeatability was about rf

    3

    O o and

    accuracy was largely unknown. Spectral analysis

    could not necessarily define absence

    of either naph-

    thenic or paraffinic hydrocarbon, as was the cace for

    aromatics. Apparent detection of small amounts of

    paraffin in a predominantly naphthenic peak, or con-

    versely, could have resulted from spectral inaccura-

    cy, contamination from an adjacent poorly separa ted

    peak, or could have been real. Hydrocarbon-type

    distribution est imates were used only for qualita-

    \

    J

    SAMPLE

    B

    n C9

    tive guidance an,d inaccu racy was of no particu lar

    concern.

    PEAK

    X Y L E N E D E T E C T I O N I N

    TWO

    CRUDE OI LS

    M J O R X Y L E N E ml LI NES =

    91

    92 105

    106 107

    chromatograms of more aromatic crude oils, was

    largely composed of xy lene wi th smaller amounts

    o normal nonane and g naphthene.

    Mass spect ra of compara ble chromatographic

    peaks were qualitatively identical from sample to

    sample. Comparative spectra contained the same

    mass lines though relative intensities varied some-

    what with differences in single compound or com-

    pound-type distribution within a peak. Spectra of

    the 9-.12 peak in paraffinic crude oils, for example,

    indicated approximately equal concentrations

    of

    paraffinic (isoalkanes) and naphthenic (cycloalkanes)

    compounds while naphthenes predominated in the

    same peak in naphthen ic crudes, Figure 7.

    Many

    of

    the chromatographic peaks were, how-

    eve r, consist ently pure in hydrocarbon-type com-

    position. Mass spectra of 9-.62 peaks were predo-

    minantly paraffinic and 10-.10 shoulder spectra

    were highly naphthenic in both paraffinic and na ph-

    thenic-type crude oils, Figure

    7.

    In ge neral, isoparaf-

    fin s were concentrated in the chromatographic PRRT

    ranges of

    .6

    to

    7

    and

    .2

    to

    3

    in each peak group.

    Conversely, in the PRRT ranges 1 to

    .2 .4

    to 5 and

    .8

    to

    .9,

    isoparaff in levels w ere minimum and naph-

    thenic spectra were usually observed. This PRRT

    distribution of isoparaffins in each peak group is

    Quantit ative Hydroc arbon-Type Distribution

    Accurate naphthene contents of peak group frac-

    tions, collected from a p repa rative scale ga s chroma-

    tography unit, were determined by conventional

    mass spect rometry. Direct comparisons of composit-

    ed a nalytica l ,data showed tha t th e chromatographic

    peak profile could be related to gross hydrocarbon-

    type distribution. Relative heights of predominantly

    naphthenic peaks were directly proportional to total

    naphthene content in each group, Figure 8. The major

    naphthenic peak was larger tha n the major isoparaf-

    fin peak in each group fraction of the Illinois crude

    oil; specifically, 7-.27

    >

    7-.74

    8 .30

    >

    6 . 6 4

    9-.47

    >

    9-42 10-.45 > 10-.61 and 11--48

    >

    11-62. The reverse was true for the more paraffinic

    East Texas crude;

    7-.27

    7-.74 8-.30