high blood pressure 8

For more 20 years, noninvasive blood pressure level (NIBP) monitors happen to be widely used inoperating rooms and critical care units to closely monitor blood pressure level in patients spanningvarious ages. Regardless of the widespread use of automated blood pressure level monitors,clinicians still deliberate within the accuracy and longevity of automated NIBP devices compared toother ways of blood pressure level determination. These answers to commonly asked questionsabout the use of NIBP monitoring derive from clinical research.

Page 2: high blood pressure 8

Q: How can hypertension measurements obtained with automatic noninvasive blood pressure level(NIBP) devices compare to direct arterial measurement of hypertension?

Page 3: high blood pressure 8

Clinical studies have demonstratedthat when blood pressures diastolicand systolic, and mean arterial)dependant upon NIBP monitors fromvarious manufacturers are incomparison to direct arterialpressures, both values are, onaverage, within 5 mm Hg of eachother. (1-9) Factors such as theanatomical location of measurementplay a role in the differences whichexist between direct and indirectmethods. In comparing brachialarterial pressure obtained with theNIBP monitor to radial arterialpressure obtained by direct arterialcannulation, radial arterial pressureis usually higher since the radialartery is a smaller vessel and helps tocreate greater potential to deal withflow, which in turn results in a higherblood pressure levels reading thanthat determined via the brachialartery for example.

These studies also indicate thatoccasionally, somebody NIBPhypertension determination value willbe different by as much as 37 mm Hgfrom the direct arterial value. (1-9)This large discrepancy signifies thattreatment should not be made based on a single NIBP determination without comparison for anauscultatory blood pressure determination or several consecutive measurements done by an NIBPmonitor.

Q: Is it essential to use the http://www.cdc.gov/bloodpressure/ correct cuff size when utilizingautomatic NIBP devices?

Page 4: high blood pressure 8

Yes. Using a cuff that is certainly 1oo small will cause falsely high readings, and making use of a cuffwhich is too large will bring about falsely low readings. The cuff width selected should equal 40% inthe arm circumference (see Figure). The American Heart Association ideas for appropriate cuff sizesbased upon upper-arm circumference ought to be followed when you use NIBP monitors (see Table).(10)

Q: What are the patient-related complications connected with using NIBP devices?

Skin and tissue compression from NIBP monitors, which can cause skin irritation and bruising, arehigh blood pressure symptoms in men the most frequently occurring complications. Prolonged useand frequent blood pressure levels determinations can result in venous pooling and congestion.Excessive venous pressures can cause tissue ischemia and nerve damage. (11,12)

Q: What factors can affect obtaining accurate NIBP measurements?

Page 5: high blood pressure 8

Several circumstances can prevent accurate determination of blood pressure with NIBP devices.Highly irregular or rapid cardiac rhythms make it tough to accurately determine blood pressurelevels using NIBP devices due to the great beat to overcome variability. Most NIBP devices employoscillometric technology that may be influenced by fairly regular cardiac rhythms to figure out bloodpressure. Excessive patient movement including shivering, restlessness, or external movement likethat from your helicopter, ambulance transport, or perhaps a rapid-cycling ventilator can obstructdetection of cardiac oscillations by the NIBP monitor. This may lead to erroneous blood pressurelevel measurements.