g7 acetabular system -...
TRANSCRIPT
G7 acetabular SYSteMDesign rationale
G7 Des iGn Rat ionale
Surgeon DeveloperS
Simplicity. efficiency. performance.
the G7 acetabular System unites the latest technological developments in bearing design with a simple, elegant and
highly flexible instrumentation system designed to optimize the operating room experience. the system’s wide range of
acetabular shell, fixation and bearing options provides the power to personalize implant selection for each within a single
platform that delivers:
› Modern shell design features for
maximized performance
› Multiple clinically proven bearing
and fixation options
› Personalized instrumentation
system for increased
operating room efficiency
› Patent pending color-coded
implant and instrumentation
system for streamlined
surgical flow
› Dr. Mike berend center for Hip and Knee Surgery Mooresville, IN, uSa
› Professor thorsten Gehrke eNDO-Klinik Hamburg, Germany
› Professor Guido Grappiolo Humanitas clinical Institute rozzano Milano, Italy
› Dr. David liu Gold coast centre for bone and Joint Surgery Gold coast, australia
› Dr. William long the computer Surgical Institute los angeles, ca, uSa
› Professor eric Masterson university of limerick Medical School limerick, Ireland
› Dr. Michael Miller university Orthopedic Specialists tucson, aZ, uSa
› Dr. rafael Sierra rochester, MN, uSa
› Dr. Matthew Squire university of Wisconsin Orthopedics Madison, WI, uSa
› Professor Nobuhiko Sugano Osaka university Osaka, Japan
Patient-SPecific mini tRaySinStrument Wave trayS
boneMaster limited Hole
Freedom constrained liner
biolox® delta ceramic Head
biolox® delta Option ceramic Head
cocr Head Freedom constrained cocr Head
Sizes 28-36 mm ce marked
Sizes 28-40 mm ce marked
Neutral liner
High Wall liner
10 Degree liner
Neutral liner
High Wall liner
10 Degree liner
liner
acetabular ShellS
ceramic linerS
femoral heaDS
e1 antioxidant arcomXlpolyethylene linerS
biolox® delta
efficiency doesn’t demand the elimination of options. the G7 System allows surgeons and hospitals
to provide the optimal implants for each patient while benefiting from the efficiency of a
streamlined delivery system.
SyStem overvieW
Sizes28-44 mm
Size32 mm and 36 mm
0102
PPS limited Hole
DesiGneD foR PeRfoRmance
Polyethylene and Hard bearing locking Mechanism accommodates multiple polyethylene configuration options and hard bearing surfaces where available
Pre-Plugged Screw Holes provide the operative flexibility to choose between a solid shell construct or adding screws for additional stability
the G7 dome loading design provides maximum liner support to minimize locking mechanism fatigue.
hemiSpherical DeSign the hemispherical design of the G7 acetabular
shell closely matches the shape of the natural
acetabulum, allowing for minimal removal of
bone while maximizing contact area between
the implant and bone.
50 mm
180°
DesiGneD foR PeRfoRmance
* Primary identification should be made using size and letter designations.
aDvanceD fixation oPtions
ppS porouS plaSma Spray: PPS Porous Plasma Spray has over
25 years of clinical history providing
primary implant fixation for the
biomet cementless portfolio.
using biocompatible titanium alloy
powder and proven plasma spray processing, PPS offers:
› High fatigue strength achieved by never heating
the substrate1
› Short- and long-term fixation gained from variable
pore size distribution2
› roughened scratch fit due to irregularly shaped particles2,3
bonemaSter: boneMaster is an advanced coating
technology designed to improve the
performance of cementless implants
by mimicking natural bone physiology.
With the biological benefits of
hydroxyapatite (Ha) and an enhanced
nano-structure based on apatite crystals
found in bone,4 boneMaster offers:
› enhanced implant stability by creating a favorable
environment for osteoblast adhesion
› roughness of PPS Porous Plasma Spray with 1/10th the
thickness of traditional Ha
› reduced fibrous ingrowth due to unique nano-structure
which mimics apatite found in bone, promoting
osteoconductivity and faster bone integration5-7
› Significantly greater bone density postoperatively compared
to identical plasma-sprayed Ha stems8,9
Shell configuration G7 shells are available in a
limited hole design with three
holes clustered near the apex of
the shell to allow for placement
of screws in the strongest part of
the pelvic bone. Screw holes are
pre-plugged to limit the migration
of polyethylene debris that could
potentially result in osteolysis and
provide an uninterrupted support
structure to reduce the possibility
creep of the polyethylene into open
screw holes. the plugs can be
removed if supplemental fixation
is desired.
Proprietary color-coded Shell corresponds with instrumentation and labels to create an efficient and accurate Or environment*
clinically Proven PPS Fixation has over 25 years of successful clinical history providing initial scratch-fit stability and excellent long-term fixation1-4
0304
beaR inG oPt ions
polyethylene materialSMeeting the modern demands of bearing surfaces means achieving the optimal balance of maximized strength, wear
resistance and oxidation resistance. biomet’s polyethylene bearings achieve this by pairing the latest technological
advancements with the strong heritage of clinically proven arcom polyethylene. the G7 system offers two
polyethylene materials for use with biolox® delta ceramic or cocr femoral heads.
e1 antioxidant Infused technology
e1 is the only antioxidant infused bearing technology
to use a proprietary diffusion process to maximize
strength, wear resistance and oxidative resistance via
the oxidative protection of vitamin e. Vitamin e actively
protects the bearing surface from the damaging effects
of oxidation over time.10 In addition, e1 acetabular liners
offer wear rates similar to hard-on-hard bearings when
paired with a ceramic head.1,11
arcomXl Highly crosslinked Polyethylene
arcomXl acetabular liners utilize a solid state
deformation process to obtain optimal wear resistance
without sacrificing mechanical strength or increasing
oxidation levels. arcomXl shows a 47% decrease in
volumetric wear rate when compared to a moderately
crosslinked polyethylene.1
CoCr on E1 Biolox® deltaon E1
55.0
50.0
45.0
40.0
35.0
30.0
25.0
20.0
15.0
10.0
5.0
0 3.36 2.82
Polyethylene WeaR9
5 million cycles on a hip simulator
Volu
met
ric W
ear
(mm
3 /m
illio
n C
ycle
s)
Biolox® deltaon ArComXL
12.8910.04
CoCr onArComXL
* Gamma Sterilized polyethylene test performed with 28 mm bearings. april, 2007
all other tests were performed with 62 mm acetabular shells and 44 mm bearings. Sept., 2012
*Gamma Sterilized Polyethylene 53.3 mm3
polyethylene profileS
10 Degree
Face changing
Freedom constrained
High Wall
Neutral
Designed to increase range of
motion, while decreasing the
possibility of impingement, by
means of an optimized chamfer
around the inner diameter of
the liner.
Offers 5mm material build
up (typically positioned in the
posterior superior quadrant) for
additional stability.
used to restore the center
of rotation and achieve more
joint stability when acetabular
components have been
vertically placed.
Designed to actively counter
the distractive forces that can
lead to recurrent hip dislocation.
utilizes a cobalt chrome head
with circumferential flats for
easier head reduction and a
constraining ring to supplement
the liner’s stability by increasing
resistance to lever-out forces.
0506
polyethylene liner thickneSS matrix
45°
apex
Minimum Poly Liner Thickness at 45° (mm)
Shell SizeHead Size
28 32 36 40 44
42 - A4.3
44 - A
46 - B 6.3 4.3
48 - C 7.3 5.3
50 - D 8.3 6.3 4.3
52 - E 9.3 7.3 5.3
54 - F10.3 8.3 6.3 4.3
56 - F
58 - G11.3 9.3 7.3 5.3
60 - G
62 - H11.3 9.3 7.3 5.3
64 - H
66 - I
11.3 9.3 7.368 - I
*70 - I
*72 - I
*74 - J
14.3 12.3 10.3*76 - J
*78 - J
*80 - J
*Not available in limited hole configuration
Minimum Poly Liner Thickness at Apex (mm)
Shell SizeHead Size
28 32 36 40 44
42 - A4.7
44 - A
46 - B 6.7 4.7
48 - C 7.7 5.7
50 - D 8.7 6.7 4.7
52 - E 9.7 7.7 5.7
54 - F10.7 8.7 6.7 4.7
56 - F
58 - G11.7 9.7 7.7 5.7
60 - G
62 - H11.7 9.7 7.7 5.7
64 - H
66 - I
11.7 9.7 7.768 - I
*70 - I
*72 - I
*74 - J
14.7 12.7 10.7*76 - J
*78 - J
*80 - J
*Not available in limited hole configuration
the G7 acetabular System offers biolox® delta ceramic
heads for ceramic on polyethylene articulations in 28, 32,
36, and 40 mm diameters. biolox® delta ceramic is a highly
biocompatible mixed oxide ceramic with high fracture
resistance and ultra low wear. constructed of 82% alumina,
17% zirconia and other trace elements, this third generation
ceramic is an alumina matrix composite that provides
increased fracture toughness and strength when compared
to traditional alumina ceramic.10
With over 1.6 million implanted worldwide, biolox® delta ball
heads offer a clinically successful high performance option.10
ceramic mater ial
1.6 millionover
implanteD WorlDWiDe10
polyethylene locking Strengththe G7 polyethylene locking feature maximizes
strength and minimizes micromotion by means
of a tightly toleranced tongue and groove fit
between the shell and the liner. While offering
high locking strength, the unique geometry of
the locking barb also requires minimal impaction
force to engage the liner into the shell, reducing
risk of shell malalignment.
Strength teSting:acetabular systems must be strong enough to withstand the loading conditions commonly seen in the hip,
maximizing strength and minimizing micromotion. liner push out, lever out, torque out and rim impingement are
industry accepted testing methods which measure the strength of both the locking mechanism and the liner.
G7 polyethylene and ceramic liners were subjected to push out, lever out and torque out testing under worst
case conditions.
lockinG mechanism stRenGth
harD bearing locking Strengththe hard bearing locking feature utilizes
a 10 degree conical taper as well as a
roughened finish to achieve maximum
surface contact with the liner, resulting in
a secure taper fit. an apical button on the
liner aids in alignment upon insertion but is
not a part of the locking mechanism.
0708
lever out Strength
60
50
40
30
20
10
0
2843% lower
50
2353% lower
Forc
e (N
m)
Strykertrident
X3 (10 Mrad)
DePuy SynthesPinnacle
altrX (7.5 Mrad)
biomet G7
e1 (10 Mrad)
51Statisticallyequivalent
Zimmercontinuum
longevity (10 Mrad)
Lever Out Strength11
lever Out: Disassembly force required to dislodge a polyethylene liner from an acetabular shell
all tests were performed with 54 mm acetabular shells and 32 mm acetabular liners
puSh out Strength
3000
2500
2000
1500
1000
500
0
22452061
8% lower
84762% lower
142036% lower
biomet G7
e1 (10 Mrad)
Forc
e (N
)
Strykertrident
X3 (10 Mrad)
DePuy SynthesPinnacle
altrX (7.5 Mrad)
Zimmercontinuum
longevity (10 Mrad)
Push Out Strength11
Push Out: axial force required to dislodge a polyethylene liner from an acetabular shell.
all tests were performed with 54 mm acetabular shells and 32 mm acetabular liners
Polyethylene results:
the average push out force of a G7 polyethylene acetabular liner is 2245 N.11
60 almoSt
timeS the minimum acceptance criteria11
liner puSh out: Push out testing measures the axial force required
to dislodge a polyethylene liner from an acetabular
shell. conducted per aStM F1820-97, an assembled
shell and liner was attached to a push-out fixture that
supported the cup without distortion. an axial load of
2 inches per minute was applied to the liner through
the apical hole of the shell.11
Similar to the G7 polyethylene testing regimen,
G7 ceramic liners were subjected to push out,
lever out and torque out testing. additional tests
were also conducted, specific to the ceramic liner.
Hard bearing results: the average push out force of a G7 acetabular ceramic liner is
11,590 N,1 almost 60 times the minimum acceptance criteria of 200 N, set by ceramtec,
the manufacturer of biolox® delta ceramic.1,12
liner lever out teSting:Measures the disassembly force required to dislodge a polyethylene liner from an acetabular shell. conducted per a published testing method,13 a metal lever arm was inserted into a slot cut into the liner (polyethylene) or a metal lever arm was inserted into a washer epoxyed into the liner (hard bearing) and loaded at 2 inches per minute until liner separation.
50 nm11
DiSaSSembly force requireD to DiSloDge a polyethylene liner from an acetabular Shell
Hard bearing results: the average lever out force of a G7 ceramic acetabular liner is 32 Nm.1
Polyethylene results: the average lever out force of a G7 polyethylene acetabular liner is 50 Nm.11
0910
liner torque out:Measures the disassembly force required to torque-out a liner from an acetabular shell. conducted per
documented test methods13,15 the shell and liner were inserted into a test fixture which applies torsional force
to the shell and liner. torsion was applied at a rate of 1 degree per second until the liner rotated 45 degrees
as noted by lines marked on the face of the shell and liner.
Hard bearing results: the average
torque-out force of a G7 ceramic acetabular
liner is 63 Nm. the maximum frictional torque
seen in-vivo at the femoral head to liner
interface is approximately 2.4 Nm.1,13,16
rim impingement: the ability of the acetabular liner to withstand possible rim impingement, or impaction of the rim of the liner
by the stem trunnion, is also key to the long term success of the implant. rim impingement loading can occur in-vivo
as a result of malalignment or patient movements that require a large range of motion.
conducted per aStM F2582, the trunnion of a stem was allowed to contact the rim of the liner at a joint reactive
force of 1000 N, simulating rim impingement loading conditions. Following the test, the liners were visually inspected
for signs of damage.
Polyethylene results:
Visual inspection of arcomXl and e1 liners showed
no cracking, pitting or gross damage to the inner and
outer region of the rim.1
Polyethylene results: the average torque-out
force of a G7 polyethylene acetabular liner is
29 Nm. the maximum frictional torque seen
in-vivo at the femoral head to liner interface is
approximately 2 Nm.11,13,16
1112
freeDom conStraineD liner lever out:liner lever out was also measured for the Freedom constrained acetabular liner. conducted per a
published testing method,15 load was applied to a femoral head with a lever arm attached at the rate of 2
inches per minute until the head or liner dislodged from the shell.
Freedom constrained liner results:
Shell/liner combinations, the average lever out force
of a G7 Freedom constrained acetabular liner is
18 Nm, although none of the constrained liners actually
levered out during the test. Per its designed intent, the
constrained femoral head dislodged before the liner in
all of the samples. this feature is designed to prevent
forces that might dislodge the fixation of the acetabular
shell in the case of a constrained liner.1
inSertion valiDation anD apical button burSt teSting Fracturing of a ceramic liner is the most
common complication in the rare instance
of ceramic failure. this fracture may
be caused by intraoperative chipping
that can occur when the ceramic liner
is not aligned properly in an acetabular
shell upon impaction or if the surgeon
forcibly tries to realign a malaligned liner.
For this reason, the G7 ceramic liner is
designed with an apical button to help
aid in alignment upon insertion. additional
testing validated this design feature.
Insertion Validation
a cadaveric validation study was
conducted to simulate intraoperative
insertion of a ceramic liner into an
acetabular shell. this demonstrates the
optimal performance of the G7 acetabular
System with regards to intraoperative liner
chipping. On separate occasions, two
practicing surgeons exposed the hip joint
while maintaining standard incision lengths
and soft tissue dissection and inserted 40
ceramic liners. In total 80 ceramic liners
were included in the validation study.1
results: During the cadaveric validation
study none of the ceramic liners chipped
when inserted into an acetabular shell that
had been impacted using a “line to line” or
“2 mm under” reaming technique.1
results: there was no significant difference between the
burst strength of the liners before and after fatigue testing.
examination of the fracture patterns before and after fatigue
of the liners showed that the liners failed with a spiraled
crack originating from the rim of the liner. the fracture pattern
of the liners does not show any interaction with the apical
feature, indicating that the novel geometry is not detrimental
to the strength or performance of the liner.1
80 ceramic linerS Were incluDeD in the valiDation StuDy.
apical button Fatigue burst
evaluates any impact of a novel apical button
feature on the burst strength of the ceramic
liner, under extreme loading conditions in
excess of any testing parameters prescribed
by national or international standards.
G7 acetabular shells were potted at a
60 degree inclination angle to simulate extreme
cup placement. ceramic liners were then
inserted and subjected to a 10 million cycle
fatigue test. Post fatigue testing, load was
applied until the liner burst.1
1314
Over 1 million times per year, biomet helps one surgeon provide personalized care to one patient.
the science and art of medical care is to provide the right solution for each individual patient. this requires clinical mastery, a human connection between the surgeon and the patient, and the right tools for each situation.
at biomet, we strive to view our work through the eyes of one surgeon and one patient. We treat every solution we provide as if it’s meant for a family member.
Our approach to innovation creates real solutions that assist each surgeon in the delivery of durable personalized care to each patient, whether that solution requires a minimally invasive surgical technique, advanced biomaterials or a patient-matched implant.
When one surgeon connects with one patient to provide personalized care, the promise of medicine is fulfilled.
this publication and all content, is protected by copyright, trademarks and other intellectual property rights owned by or licensed to biomet Inc. or its affiliates unless otherwise indicated. It must not be used, copied or reproduced in whole or in part without the express written consent of biomet or its authorized representatives.
biolox® delta is a trademark of ceramtec.
this material is intended for the physicians and the biomet sales force only. the distribution to any other recipient is prohibited.
check for local product clearances and reference product specific instructions for use.
For product information see package insert and biomet’s website.
Material not approved for use or distribution in France.
One Surgeon. One Patient.
References
1. Data on file at biomet. bench test results not necessarily indicative of clinical performance.
2. Market, D.c. et al. Initial Scratch Fit Stability of acetabular cups: comparison of three Porous coating Systems. ISta. San Diego, ca. 1997.
3. andersson, l. et al. Immediate or late Weight bearing after uncemented total Hip arthroplasty. Journal of Arthroplasty. 16(8): 1063-65, 2001.
4. robler S,. et al. electrochemically assisted deposition of thin calcium phosphate coating at near-physiological pH and temperature. Journal of Biomedical Materials Research. 64a:655-663. 2003.
5. Mont Ma. et al. Proximally coated Ingrowth Prosthesis. a review. Clinical Orthopedics 344: 139-149. 1997.
6. emerson r., et al. effect of circumferential plasma-spray porous coating on the rate of femoral osteolysis after total hip arthroplasty. The Journal of Bone and Joint Surgery. 81: 1291-1298. 1999.
7. bourne r., et al. Ingrowth surfaces: Plasma spray coating to titanium alloy hip replacements. Clinical Orthopedics Related Research 298: 37-46. 1994.
8. Schliephake H., et al. biomimetic calcium phosphate composite coating of dental implants, International Journal of Oral Maxillofacial Implants: 21 (5):738-46. 2006.
9. Schliephake H., et al. biological performance of biomimetic calcium phosphate coating of titanium implants in the mandible a pilot study in dogs, Journal of Biomedical Material Research. 64a: 225-234. 2003.
10. uS FDa cleared claim. See biomet.com/e1 for complete claim language.
11. Study conducted by biomet with independent lab. laboratory testing august, 2013. test results are not necessarily indicative of clinical performance.
12. Data on file at ceramtec GmbH.
13. FDa draft guidance: testing of acetabular prosthesis
14. tradonsky, S., et al. Performance characteristics of two Piece acetabular cups Series II. Scientific exhibit. 62nd annual aaOS Meeting. 1996.
15. Haft, G., et al. a biomechanical analysis of Polyethylene liner cementation into a Fixed Metal acetabular Shell. The Journal of Bone and Joint Surgery, 85-a: Number 6, June 2003.
16. Marlowe, D., et al. Modularity of Orthopedic Implants. Issue 1301. aStM International, January 1997.
One Surgeon. One Patient.
ORTHOPEDICSSM
©2013 biomet Orthopedics • Form No. bMet0756.0-INt • reV0913
Legal Manufacturerbiomet Orthopedics P.O. box 58756 e. bell DriveWarsaw, Indiana 46581-0587 uSa
Biomet UK Ltd.Waterton Industrial estatebridgend, South WalescF31 3Xa uK
Note: For ceramic components contained within this Surgical technique biomet uK, ltd is the legal Manufacturer.
www.biomet.com
0086
European Representativebiomet uK, ltd.Waterton Industrial estatebridgend, South WalescF31 3Xa uK