Page 1: FDA Update: Inspections, Observations and Metrics - OMTEC 2017
Page 2: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

OMTEC 2017 FDA Update

FDA’s focus as reflected in FDA-483 Observations with trends over time (historical view) On the horizon - FDA’s Case for Quality Initiative - Medical Device Quality Metrics with practical application (future view)

- Quality and Regulatory Solutions - FDA Compliance - Quality System Management & Readiness - Supplier Quality Management - Metrics & Data

Eduard Toerek, President - QUARA Innovations [email protected] Reinhold Toerek, Vice-President - QUARA Innovations [email protected] www.quarainnovations.com

© 2017 Quara Innovations LLC . For use permission, contact [email protected]

Page 3: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

2

Source: FDA’s Field Accomplishment and Compliance Tracking System (FACTS)

Key Finding CY2016 o Slight increase in overall number of

QS Surveillance Inspections in CY16 vs. CY15

o Increase in No Action Indicated (NAI) inspection outcomes (both domestic and foreign)

o Fewer 483s were issued to firms in CY16

o All QS subsystems saw a drop in the number of 483 observations

o Increase in Foreign Inspections and decrease in Domestic Inspections (consistent with the increase in foreign firms actively registered and listed)

o Number of WLs issued dropped from 121 (CY15) to 57 (CY16)

Page 4: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

3

FDA Establishment Inspections

Form FDA-483 Inspectional Observations • Issued at the conclusion of an inspection when an investigator

has observed any conditions that my constitute violations of the Food Drug and Cosmetic (FD&C) Act and related Acts

• Used to document concerns discovered during FDA Inspections • Multiple observations may be listed on the FDA-483 Form • Note: FDA-483s do NOT cover every possible deviation from

law and regulation

Page 5: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

4

FDA Establishment Inspections

Inspection Outcomes / Classification • Official Action Indicated (OAI)

– An OAI inspection classification occurs when significant objectionable conditions or practices were found and regulatory action is warranted to address the establishment's lack of compliance with statute(s) or regulation(s).

• Voluntary Action Indicated (VAI) – A VAI inspection classification occurs when objectionable conditions or practices were

found that do not meet the threshold of regulatory significance. Inspections classified with VAI violations are typically more technical violations of the FDCA.

• No Action Indicated (NAI) – An NAI inspection classification occurs when no objectionable conditions or practices were found during the

inspection or the significance of the documented objectionable conditions found does not justify further actions.

• If no enforcement action is contemplated, or after enforcement action is concluded, FDA provides inspected establishments with a final inspection report, called an Establishment Inspection Report (EIR)

Source: www.fda.gov

Page 6: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

5

Top 10 Foreign Inspection Location (CY15 – CY16)

Country Name CY 2015 # of Inspections

Country Name CY 2016 # of Inspections

China 126 China 179 Germany 90 Germany 71

Japan 44 Japan 60 Canada 42 United Kingdom 50

United Kingdom 35 Taiwan 35 Taiwan 35 France 29 France 30 Switzerland 29

Italy 26 Italy 27 S. Korea 22 Canada 26 Ireland 19 Ireland 25

Source: FDA

Page 7: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

6

QS Medical Device Inspections Outcomes, FY 2016

Domestic Inspection Outcomes % Foreign Inspection Outcomes % NAI 779 54% NAI 351 48% VAI 567 39% VAI 288 40% OAI 104 7% OAI 86 12%

Total 1,450 725

Source: FDA

Page 8: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

7 Source: www.fda.gov

CAPA and P&PC continue to be the most frequently cited QS subsystems.

Page 9: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

8

21 CFR 820 QS Regulation Subsystems

P&PC Description CAPA Description 820.50 Purchasing Controls 820.90 Nonconforming product 820.60 Identification 820.100 Corrective and preventive action 820.65 Traceability 820.198 Complaint files 820.70 Production and process controls MGMT Description 820.72 Inspection, measuring, and test equipment 820.5 Quality system 820.75 Process validation 820.20 Management responsibility 820.80 Receiving, in-process, and finished device acceptance 820.22 Quality audit 820.86 Acceptance status 820.25 Personnel 820.120 Device labeling DES Description 820.130 Device packaging 820.30 Design controls 820.140 Handling DOC Description 820.150 Storage 820.40 Document controls 820.160 Distribution 820.180 General records requirements 820.170 Installation 820.181 Device Master Record 820.200 Servicing 820.184 Device History Record 820.250 Statistical techniques 820.186 Quality System Record

Page 10: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

9 Source: www.fda.gov

Page 11: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

10 Source: www.fda.gov, Analysis by QUARA

Page 12: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

11

Most Frequently Cited 483 Observations (FY06 – FY16)

1) Procedures for Corrective and Preventive Action have not been [adequately] established. Specifically, ***

2) Procedures for receiving, reviewing, and evaluating Complaints by a formally designated unit have not been [adequately] established. Specifically,***

3) Written MDR procedures have not been [developed] [maintained] [implemented]. Specifically, ***

4) Corrective and Preventive Action activities and/or results have not been [adequately] documented. Specifically, ***

5) A Process whose results cannot be fully verified by subsequent inspection and test has not been [adequately] Validated according to established procedures. Specifically, ***

Source: FY06-16 Inspectional Observations

Page 13: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

12

Corrective and Preventive Action (CAPA)

• Triggers • Problem Statement • Risk Assessment • Containment • Corrections • Investigations • Root Cause • Verification • Corrective Action • Effectiveness Checks • Documentation

Page 14: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

13

Metrics

Looking ahead: FDA Case for Quality

Page 15: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

14

FDA Case for Quality

• Launched in 2011 and is part of CDRH’s 2016-2017 Strategic Priority to promote a culture of quality and organizational excellence.

– Core Components • Focus on Quality • Enhanced Data Transparency • Stakeholder Engagement

• MDIC has sponsored four Case for Quality Working Groups since 2014

– Maturity Model • Enable an organization to assess the capability of its quality system

to reliably develop and manufacture high quality medical devices. – Metrics

• Create well-defined, stakeholder-verified (FDA and industry) product quality metrics to predictively assess product quality.

– Advanced analytics • Offer hospital providers information and analysis techniques to evaluate

medical device quality and subsequent patient value. – Competencies

• Construct techniques that improve competency across systems and individual functions.

www.fda.gov www.mdic.org

Page 16: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

15

FDA – On the Horizon: Case for Quality

What is this Case for Quality Initiative? • …shift historical focus from compliance & enforcement action to device quality • ...launched after device quality data demonstrated lack of improvement in

risk to patient safety & the number of enforcement actions taken by FDA year after year

• ...goal is to proactively and predictively measure risk to product quality • ...Right The First Time mentality shifting as close to initial days of development

as possible • ...manageable system of metrics across the Total Product Lifecycle • ...ultimate goal is continual improvement with root cause of failure taken

back to earliest stages of development as possible

Source: MDIC Quality Metrics Best Practices

Page 17: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

16

Where is the FDA Case for Quality Going?

Source: www.fda.gov MDIC CfQ Metrics Workshop

Page 18: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

17

What Does This Mean?

FDA Regulatory Paradigm Shift • What does a focus on quality and organizational

excellence mean for FDA and innovation? – Increased manufacturing and product confidence – Faster time to markets, better information to drive

regulatory decisions, improved resource allocation – A focus on what is most important to patients

Remove participants from the agency work plan for routine inspections Waive pre-approval inspections where appropriate Engagement and meetings on issue resolution Modifying submission requirements and faster FDA response Accelerated approval path Competitive market around product excellence

Source: www.fda.gov

Page 19: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

18

Maturity Model Workstream - Goal Statement

Develop a program which leverages CMMI (Capability Maturity Model Integration) as the standard maturity model by which medical device organizations may measure their capability to produce high quality devices. FDA will adjust their engagement activities and submission requirements as a recognition of this independent assessment of quality maturity.

Source: MDIC CfQ

Page 20: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

19

Maturity Model

Page 21: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

20

Total Product Lifecycle

Page 22: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

21

Pilot Study Metrics

Total # of changes (product & process across projects)

total # of projects

Total # of changes (product & process for each project)

and/or

# of units mfg. Right First Time within or across lots

# of units started

Post-Production Metric

Production Revised Metric

Pre-Production Revised Metric

Aggregation of weighted (risk based) post market metrics: Service Records, Installation Failures, Complaints, MDR’s, Recalls Rates, Recalls Total….

Page 23: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

22

MDIC “Best Practices”

• Metric output can be used to understand root causes • Combine metric output with other metrics to understand

a more holistic picture and analyze trends • Goal is to provide a feedback loop to improve systems from the

earliest point possible that allowed the failure to occur originally

Purpose: To help organizations understand how best to use the output from the metrics to inform decisions and trigger actions

Page 24: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

23

Total Product Lifecycle

• This process [approach] can be used to identify ways to measure previously untracked areas of quality and/or risk

• Assess which critical requirements the metric is correlated to, in order to be sure it has the potential to be effective

• Be sure to assess the usefulness of the metric over time – Is it a flat-line result over time? – Are any decisions or actions ever taken as a result of tracking? – Has the metric demonstrated acceptable improvement and steady-state? – Is unacceptable quality and/or risk experienced even though

this metric is consistently acceptable?

Source: MDIC Case for Quality, Metrics

Page 25: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

24

Pre-Production: Design Robustness

• Goal of the Pre-Production Metric: to drive the Right-First-Time (RTF) mindset in the research and development phase such that post-design transfer changes due to inadequate product/process development are not needed.

Only include changes required due to inadequate product or process development (harmonize definition across organization)

Page 26: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

25

ECO

ECR

Manufacturing Field Design

Change ECR

Mfg or Design

Cost of: • Verification • Validation • Process Validation • Ext. Certification • Registration/Clearance • Cost/FA • Cost/CAPA • Cost/Design Change

Impacted by changes • IEC (Elec. Safety…) • Toxicology/Bio-compatibility • Environmental (WEEE, RoHS…) • Declaration of Conformity • Supplier Qualification • Cleaning/Sterilization • Risk Management • Labeling • Training • Disposition (Inventory, Scrap, Rework…)

Summing Junction for Post-RfDT product and process changes due to inadequate product/process development

Change CR

Pre-Production / Design Robustness

Data Sources • NCRs • SCARs • Deviations • Improvement Projects

Data Sources • Complaints • Adverse Events (MDR, MDV...) • Recalls / Field Actions • Installation Reports • Service Events • Product Returns

MDIC Pre-Production Metric

Design Xfr

Release for Design Transfer (RfDT)

Design Phases/Processes • Requirements • Specifications • Verification • Validation • Process Validation • Clinical Trials • Prototyping • Usability • Bio-compatibility • Risk Management • Design Review • Lifecycle testing • Compliance (i.e. to standards) • Labeling

Manufacturing Processes • Inspection • Process Validation • Inventory Control • Environmental Control • Process Capability • Control / Reaction Plans • FMEA • Preventive Maintenance • Calibration

Y

N

Design

Mfg

Y

N

Total # of changes (product & process across projects)

total # of projects

and/or

Total # of changes (product & process for each project)

Note: Each organization must determine what constitutes a project

Page 27: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

26

Issues within 12 months of Design Transfer (RfDT)

Page 28: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

27

How well do you really know your systems? • Can you metrics be interpreted accurately? • What questions should you be asking your organization

in regards to data? • Are you, your suppliers and your customers all singing

from the same hymnal? Decide what data you need How will this data be used? • Capability? • Defects? • Yield? • etc.

Data Collection / Metrics / Dashboards

Develop data collection plan • What will we measure? • How will we measure it? • Where will we measure it? • How often will we measure it?

Ensure data integrity (MSA) • Accurate • Precise • Repeatable • Reproducible Determine how data will be presented Determine Reaction to data • Containment • Correction • Improvement • Design Change

Page 29: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

28

What do we need to know about the process?

Page 30: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

29

Data Collection Concepts / Preparation

Do we know what defective means? • Have we defined the acceptable

and unacceptable aspects of the process?

• For example: If you were asked to count the number of defective M&Ms in a bag, we’d get a wide range of answers. Why?

Page 31: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

30

Data Collection Concepts / Preparation

Do we know what acceptable means? • Not everyone sees the same things. What is

acceptable to some may be unacceptable to others.

That’s not a manufacturing defect, It’s supposed to have a hole in the middle

Page 32: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

31

Data Collection Plan

Develop Data Collection Plan

• A well-prepared Data Collection Plan helps ensure successful analysis of the problem. Data Collection Plans should answer the following set of questions: – What data will be collected (including data type) – Attribute data — qualitative (Yes/No, Pass/Fail, Damage/No Damage) – Variable data — quantitative (Time, Dimensions, Percentage) – Why the data is needed – Where data will be collected – How the data will be collected – Who will collect the data

Key Concept Collecting an appropriate amount of the right data. Too much data can add complexity to the data review and analysis. Too little data may force the team to engage in a secondary data collection effort. Likewise, correctly specifying what data is to be collected (enough to get a complete picture of the process) will help the team avoid unnecessarily repeating initial collection activities.

Page 33: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

32

Can you trust your data?

• What does this mean? • Stability • Accurate (Bias) • Linearity • Repeatability • Reproducibility

• Are we measuring the “right” thing? – For example, does our data match with what our

customers are saying? – Liftgate Example

Page 34: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

33

Measurement System Analysis (MSA)

• The purpose of performing a Measurement System Analysis is to ensure the information collected is a true representation of what is occurring in the process.

• It is important to remember that Total Variation is the sum of Process Variation and Measurement System Variation. Therefore, minimizing measurement variation ensures that only process variation is reflected by the data.

At the conclusion of the Measurement System Analysis, you should know:

• Whether the measurement system is “capable” of gathering data that accurately reflect variation in the process

• Whether there is measurement error, how big it is and a method of accounting for it

• What confidence level can be attached to the measurements collected

• Whether or not measurement increments are small enough to show variation

• Sources of measurement error • Whether the measurement system will be stable over time

Measurement System: The thing being measured + the device(s) used to take the measurement + the person doing the measuring

Page 35: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

34

Components of Measurement System Error

• Resolution/Discrimination • Accuracy (bias effects) • Linearity • Stability (consistency) • Repeatability-test-retest (Precision) • Reproducibility (Precision)

Each component of measurement error contributes to variation, causing wrong decisions to be made.

Page 36: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

35

We did the MSA; so what?

• Stable: Capacity of measurement system to obtain the same result when measuring the same part over a significant period of time. (Stability implies only common cause variations.)

• Accurate (Bias): The closeness of a measured value to a standard or known value

• Linear: A measure of bias over the range of the measurement device.

• Repeatable: Can the same person measure the same part multiple times with the same measurement device and get the same value?

• Reproducible: Can different people measure the same part with the same measurement device and get the same value?

The analysis answers the questions, is the Measurement System:

Measurement system corrections resulting from MSA lead to precise and accurate data

Page 37: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

36

We did the MSA, so what? The analysis answers the questions; is the Measurement System: • What is the measurement error? • What do I do with this measurement error? • Capable of measuring the process? • If I improve my process, will the measurement system still be

ok? – The output of an MSA provides indices that represent the

measurement systems ability to measure within the process spread and also within the tolerance band. If process improvements are made or the tolerance limits are tightened beyond the measurement system capability, the changes to the measurement system are necessary.

• Will the MSA ever need to be repeated?

Page 38: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

37

Let’s Start Measuring…oh wait...

What about a sampling plan?

Sampling Approaches:

Random SystematicSampling Sampling

Each unit has the Sample every nth same chance of one (ie: every 3rd)being selected.

Stratified SubgroupRandom SamplingSampling

Randomly sample Sample n unitsa proportionate every nth time number from (ie: 3 units everyeach group. Hour); then

calculate the mean(proportion) foreach subgroup.

SamplePopulation

SamplePopulation

AABBBBCDDD

A

AA

A

BB B

BBBB

C

C

D D DD D D

B

SamplePopulation or Process

Preserve time order

SamplePopulation or Process

Preserve time order

SampleProcess

9:00 9:30 10:3010:00

Preserve time order

SampleProcess

9:00 9:30 10:3010:00

Preserve time order

Sampling Considerations

• Where o Location in the process where process steps directly

affect outputs (strong relationship). o Maximize opportunity for problem

identification (cause data). • Frequency

o Dependent on volume of transactions and/or activity

o Unstable process – more frequent sampling o Stable process – less frequent sampling o Dependent on how precise the measurement must

be to make a meaningful business decision • Considerations

o Is the sample representative of the process or population?

o Is the process stable? o Is the sample random? o Is there an equal probability of selecting any data

point?

Page 39: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

38

What’s the Point? (Take-aways)

Do we / Have we: • Identified the appropriate data that we need to collect? • Determined how to collect it? • Determined where to collect it? • Determined when to collect it? • Determined the sampling plan? • Studied the measurement system? • Determined how to manage the data? • Understand the sources of variation in our processes? • Developed reaction plans? • Know how to form a proper problem statement? • Know how to effectively conduct an investigation? • Understand interim and permanent corrective action? • Verified root cause? • Can we eliminate unnecessary measurements?

Page 40: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

Thank You!!! Eduard Toerek, President - QUARA Innovations [email protected] Reinhold Toerek, Vice-President - QUARA Innovations [email protected] www.quarainnovations.com © 2017 Quara Innovations LLC . For use permission, contact [email protected]

Page 41: FDA Update: Inspections, Observations and Metrics - OMTEC 2017

40

Page 42: FDA Update: Inspections, Observations and Metrics - OMTEC 2017