fabrication and characterization of thermal conductivity detectors (tcds) of different flow channel...
TRANSCRIPT
Fabrication and characterization of thermal conductivity detectors(TCDs) of different flow channel and heater designs
Y.E. Wua, K. Chenb,*, C.W. Chena, K.H. Hsua
aDepartment of Mechanical Engineering, National Taiwan University of Science and Technology, Taipei, TaiwanbDepartment of Mechanical Engineering, University of Utah, 50 S. Central Campus Dr., Rm. 2202, 84112 Salt Lake City, UT, USA
Received 19 October 2001; received in revised form 8 April 2002; accepted 14 April 2002
Abstract
Different flow channel designs and heaters made from different materials were tested for improving the performances of silicon-based
thermal conductivity detectors. One of the designs involved an electric heater sandwiched between two identical flow channels for high heat
transfer rates. The heater of the other design was suspended over a slot to reduce heat losses. The flow channels were etched in silicon wafers
and nickel heating elements were deposited on Pyrex glass, polyimide, and silicon nitride membranes. The transient behaviors of the heaters
and the wafer temperatures were measured and analyzed for different voltages. The effects of flow channel design and membrane material on
the heat transfer characteristics and sensitivities of the detectors were examined. Simple heat transfer models were developed to aid in
understanding and diagnosing detector behaviors and performances. The polyimide heater had the best signal conditions. The warm-up times
of the TCDs were found to be primarily dependent upon the package dimensions and properties. The double-channel TCD exhibited 20%
higher heat transfer rate compared to the single-channel design, but the sensitivities of these two designs differed only slightly.
# 2002 Published by Elsevier Science B.V.
Keywords: Micro GC; Response time; Sensitivity; Thermal conductivity detector
1. Introduction
A variety of thermal sensors [1] have been developed for
the measurements of heat flux, mass flow rate, velocity, as
well as physical and chemical properties (e.g., thermal
conductivity, chemical composition, etc.) of a fluid flow.
Examples of thermal sensors include hot-wire and hot-film
anemometers, thermal-pulse and other convective-type
flowmeters [2], TCDs, and so on. These sensors generally
involve a fluid flow over electrically heated filaments or thin
films. Various fluid and flow properties can be determined by
measuring the heat transfer rate or temperature change of the
heating elements. For most thermal sensors, the sensitivity
can be improved by increasing the heat transfer rate between
the heater and the fluid. A short response time is often
desired for reduced warm-up period and better temporal
resolution. Although attention in the present investigation
was limited to TCDs, results of this study are also useful to
other sensors that utilize heat transfer in a fluid flow for
property measurement or identification.
TCDs have been used in gas analysis for more than a
century [3]. In a TCD the change in gas thermal conductivity
due to the injection of a gas sample into the carrier gas
stream can be detected by measuring the change in heat
transfer rate of an electric heater in the flow. The flow
channel of a TCD is very long and lateral heat transfer is
almost exclusively by conduction. In recent years silicon
micromachined TCDs have been widely used in micro gas
chromatography (GC) systems. These micro GCs have the
size of a briefcase [4,5] and are capable of performing most
of the tasks a large chemical analyzer can do. The compact-
ness and portability of micro GCs make them extremely
suitable for environmental testing such as leakage detection
of hazardous chemicals, monitoring of various chemical
processes, and gas emission analysis of power plants,
Although TCDs may not be as sensitive as ionization
detectors for GC applications, they are concentration sensi-
tive devices while ionization detectors are mass sensitive. As
the detector size reduces, so does the fluid mass, but the
concentration remains unchanged. The TCD is therefore,
more advantageous to use than other types of detectors for
micro GCs [6,7].
A review of TCD heat transfer was given in Chen and
Wu’s paper [7]. Extensive studies of conventional TCDs
Sensors and Actuators A 100 (2002) 37–45
* Corresponding author. Tel.: þ1-801-581-4150; fax: þ1-801-585-9826.
E-mail address: [email protected] (K. Chen).
0924-4247/02/$ – see front matter # 2002 Published by Elsevier Science B.V.
PII: S 0 9 2 4 - 4 2 4 7 ( 0 2 ) 0 0 1 4 4 - 9
have been carried out and published in various books (e.g.,
[3]) and technical journals (e.g., [8–11]). On the other hand
few research results of TCDs that were fabricated using the
lithography technique were published despite the fact that
silicon-based TCDs have become commercially available
for years. Early research of silicon-based GCs and TCDs
was performed at Stanford University [6,12]. Shallow chan-
nels were etched on a silicon wafer and thin metallic films
were used as heaters. Convective heat transfer was neglected
and a lumped thermal model was developed in Jerman’s
thesis [6] for heat transfer analysis of their silicon-based
TCD. The TCD, Jerman and co-workers developed was
found to be very sensitive to pressure difference due to
the shallow channels they used. The entrance effect of TCDs
of planar structure was found to be confined to a distance
about four times the thermal entry length of a circular tube
subject to a uniform thermal boundary condition [7].
The present investigation was motivated by the lack of
detailed experimental data for flow channel and heater
designs and their effects on the performances of silicon-
based TCDs. In this paper nickel (Ni) heating elements were
deposited on different membranes and their transient beha-
viors and noise levels were recorded and studied. A one-
dimensional heat transfer analysis was applied to different
flow channel designs: one with the gas flow above the heater
only and the other comprised of an upper and a lower flow
channel. Electricity consumption rates of the two designs
were measured and their sensitivities compared. The signal
conditions of our TCDs were influenced by the supplied
voltage and the membrane material. The warm-up time
depended primarily on the silicon package dimensions.
The double-channel design had a higher heat transfer rate,
but the sensitivities of the double- and single-channel
designs were very close to each other.
2. Design and fabrication of silicon-based TCDs
The simplest TCD design that can be batch-fabricated
using the lithography technique involves a thin film depos-
ited on a substrate as a resistor to generate heat. Above the
heater is an etched flow channel bonded or glued to the
substrate surface. The major drawback of this single-channel
design is the conductive losses through the substrate. To
reduce the conductive losses, a deep channel, evacuated or
opened to the atmosphere, is often fabricated underneath the
heater. Fabrication and test results of the single-channel
TCD can be found in Jerman’s thesis [6].
From the heat transfer point of view, a better TCD design
is the one with the electric heater sandwiched between
two flow channels identical in dimensions, material, and
configuration. The heat generated by the heater will pass
through the gas volumes in the upper and lower flow channels
equally. The conductive losses beneath the heater can be
almost totally eliminated. However, since most metallic thin
films do not have very good mechanical properties, the thin
heating element suspended over the lower flow channel may
not have sufficient strength to prevent it from being deflected.
Although deflection of the heating element can be alleviated
by reducing the channel width or by increasing the film
thickness, the former will increase the heat losses to the
channel sidewalls while the latter will result in a lower
electric resistance of the heater. The remedy we devised
was to deposit the heating element onto a membrane of better
mechanical properties. The membrane must be made of an
electrically insulating material. It must have sufficient
strength to remain un-deflected when suspended over a wide
flow channel. A low thermal conductivity is also desired for
reducing the conductive loss from the membrane to the
channel sidewalls. Due to the presence of a membrane
beneath the heating element, heat transfer to the lower flow
channel is less than that to the upper channel and the TCD
performance is somewhat compromised.
In the present investigation both single- and double-
channel designs were tested. In the single-channel design
a silicon wafer with the flow channel etched in was bonded
to a membrane that forms the channel floor. A thin film of Ni
was deposited on the top surface of the membrane. Below
the membrane was another silicon channel open to the
atmosphere. The air in the open channel reduced the heat
losses from the bottom of the heater. The double-channel
design had the same membrane, heating element, and upper
flow channel as the single-channel one. The bottom of the
membrane, however, was bonded to another silicon wafer
with an identical flow channel etched in. The equivalent
thermal circuits of these two designs are depicted in Fig. 1.
Notice that the double-channel design has an additional
conduction resistance for the heat flow path from the mem-
brane bottom to the surroundings. However the total heat
flux to the gas is qu þ ql in comparison with just qu in the
single-channel TCD.
Fabrication and packaging of our TCDs are now briefly
described. A silicon wafer was first thoroughly cleaned
using a procedure similar to RCA 1 and 2 [4]. A positive
photo resist (Hoechst AZ 4903) was then spin-coated onto
the wafer. Hexamethyldisilazane (HMDS) was used as an
adhesion promoter to improve photo resist (PR) adhesion.
The aligner we used was Karl Suss JB3 UV 300/400 with
the g-line (wavelength ¼ 436 nm) light source. The flow
channels were wet-etched [4,13] in the silicon wafer using
a mixture of KOH and de-ionized water. The gas entry and
exit ports were etched using ethylene diamine-phrocatechol
water (EDP). The flow channels were only 20 mm deep and
smooth surfaces were desired. On the contrary the gas entry
and exit ports were much deeper (175–250 mm) for the
insertion of commercial capillaries and surface roughness
was not important. Therefore, EDP was used as the etchant
to speed up the wet etching process. The open channel
beneath the membrane of the single-channel TCD was
fabricated in a similar fashion. Since the lower channel
was etched through the silicon wafer, no etch stop was
needed and the faster etchant EDP was used.
38 Y.E. Wu et al. / Sensors and Actuators A 100 (2002) 37–45
The same PR (Hoechst AZ 4903) and pattern transfer
process were used for the deposition of the Ni heating
element on different membrane materials. After spin coat-
ing, softbaking, exposuring, and developing of the PR, a
chrome (Cr) layer of 15–25 nm was coated before Ni film
deposition for better adhesion. The heating element pattern
was transferred onto the Ni film by liftoff instead of etching
since lift-off has better line-width control of the transferred
pattern. Removal of the unwanted Ni film was accomplished
by immersing the membrane in acetone for 2 h, followed by
1–2 min of ultrasonic agitation. Shown in Fig. 2 is the laser
scanning microscope (LSM) image of the snaking heating
element. The width of the heating element was approxi-
mately 14 mm and the average thickness, measured by a
Tencor a-step profile meter, was around 1.5 mm. After the
heating element was fabricated, gold leads were deposited
onto the membrane in a similar process.
Three different materials were tested for the membrane on
which the Ni heating element was deposited using the
thermal evaporation method. They are polyimide, Pyrex
glass, and silicon nitride. The solidified polyimide sheet
(Kapton-FN by DuPont) was 50.8 mm thick and coated with
12.7 mm Teflon on both sides. The polyimide sheet was soft
and deposition of a uniform metallic film on it proved to be
challenging. To avoid deformation and over-softening, the
rinse temperature and time of the polyimide sheet cannot
exceed 80 8C and 5 min. A gray film was observed (Fig. 3)
when thin PR (Hoechst AZ 1500) was used. We suspected
the gray film was Cr that diffused through the PR layer
and stuck to the polyimide surface. After switching the
PR to AZ 4903 and reducing the spin coating speed from
5000 to 4000 rpm, the gray film no longer appeared. Four
identical heating elements on the polyimide surface can be
seen in Fig. 3. These heating elements form the four resistors
in a Wheatstone-bridge circuit [14] that can detect slight
imbalances between the two resistors in the testing channel
and the other two resistors in the reference flow channel of
a TCD.
Pyrex glass has good mechanical properties and a low
thermal conductivity. It was therefore, also tested as a
membrane material of the heater. Slots or shallow channels
were first etched in a silicon wafer. Corning # 7740 Pyrex
glass 500 mm in thickness was then anodically bonded
[15,16] to the wafer. The Pyrex glass/silicon wafer was
immersed in 49% HF with ultrasonic agitation for 40 min
to reduce its thickness to about 33 mm. The rough Pyrex
glass surface after etching was polished to 20 mm using
aluminum-oxide powders. Cr and Ni layers were then
deposited on the Pyrex glass surface, and the heating ele-
ment pattern was transferred by lift-off. The same pattern
transfer process was repeated for the deposition of gold leads
onto the Pyrex glass surface.
Another membrane material we tested was silicon nitride
(Si3N4). This material is very hard and strong, and is easier
to work with for metallic pattern transfer. Prior to the
deposition of the Ni heating elements, a layer of 0.5 mm
Si3N4was deposited on the silicon wafer using the LPCVD
(low-pressure chemical vapor deposition) technique. A gold
film about 1000 A thick was then fabricated by lift-off to
protect the desired silicon nitride regions. The undesired
silicon nitride was washed out in 85% phosphoric acid
(H3PO4) at 140 8C. Due to its high stiffness, silicon nitride
had a higher success rate than polyimide for the fabrication
of the heater assembly (heatingþ element þ membrane).
This extremely thin membrane, however, lasted only a
few experiments in our tests.
The upper half of the TCD, which was comprised of the
upper flow channel and the heater, and the lower half which
consisted of the lower flow channel or a slot in a silicon
wafer, were glued together using EPO-TEK 301-2 by Epoxy
Technology. During the packaging process cautions must be
Fig. 1. Cross sections of the (a) double- and (b) single-channel TCDs, and
the corresponding thermal circuits. Conduction resistances ¼ L/k and
convection resistances ¼ 1/h, where L is the thickness; k the thermal
conductivity; h the heat transfer coefficient.
Y.E. Wu et al. / Sensors and Actuators A 100 (2002) 37–45 39
taken to prevent the adhesive from seeping into the flow
channels, which may block the gas flow and/or shorten the
heating elements.
Before packaging, the heater was placed in a furnace to
measure its electric resistances at different temperatures.
The temperature–electric resistance relationships of the
heaters we fabricated were found to be fairly linear for
temperatures ranging from 20 to 200 8C, with the tempera-
ture coefficient of resistance very close to 0.003 O cm/8C.
3. Test results and discussion
3.1. Transit-time measurements
The goal of this test is to determine the time required for
the TCD output signal to reach a steady state. The tested
TCDs were the single-channel type (Fig. 1(b)), but their
heaters were made from different materials. The TCDs were
placed on a metallic rack in an evacuated stainless-steel
Fig. 2. LSM image of the deposited heating element.
Fig. 3. Gray film on the surface of the polyimide membrane during pattern transfer.
40 Y.E. Wu et al. / Sensors and Actuators A 100 (2002) 37–45
chamber during the test. Attention was focused on heat
conduction from the heating element to the membrane
and the silicon wafers. Results of this test can be used to
determine the warm-up time of a TCD. The information is
also useful to TCDs and TCD-like structures using a voltage
excitation to measure fluid properties or to improve the
sensor sensitivity. For instance, in Lacey’s patent [17], a
sinusoidally alternating voltage was applied to the Wheat-
stone bridge, or the sensor resistor was alternatively
operated at two different temperatures to eliminate the
influence of wall temperatures. Bonne and co-workers mea-
sured the time responses of DC excited filaments in their
flowmeter and pressure sensor [18,19] for fluid composition
correction.
The electric current of the heating element was monitored
during the transient test for a supplied voltage. Test results of
three membrane materials are shown in Figs. 4 and 5. In the
transient period the heating element temperature rose and its
resistance increased. The current of the heating element
therefore, decreased and eventually approached a constant
value when the heater reached thermal equilibrium with its
surroundings. The output currents were very noisy when the
supplied voltages were low, as shown in Fig. 4. High-voltage
results (Fig. 5) showed much better signal conditions. This is
because the background temperature fluctuations had less
effect on the heater temperature when it was heated to a
higher temperature. These current plots also reveal that the
times required for a TCD to reach a steady state were about
the same for different voltages. The transit times of the three
membranes we tested were very close. The output current of
the polyimide membrane had the lowest noise level, espe-
cially at low voltages. It was therefore, selected in the heat
transfer test of different flow channel designs.
A lumped-system analysis was first applied to the mem-
brane beneath the heating element. The thermal time con-
stant [20] (the time required for the temperature of a lumped
system to come within 37% of its steady-state value) of the
polyimide membrane was found to be in the millisecond
range. Thermal time constants of the other two membranes
were even smaller due to their small thicknesses and high
Fig. 4. Time variation of the heater current for a supplied voltage of 4 V. (Membrane material: polyimide.)
Fig. 5. Time variations of the heater currents for a supplied voltage of
12 V. (Membrane materials: (a) Si3N4; (b) Pyrex glass; (c) polyimide.)
Y.E. Wu et al. / Sensors and Actuators A 100 (2002) 37–45 41
thermal diffusivities. The slow warm-up periods observed
in the test therefore, must be due to the large thermal
capacitance of the silicon package. This explains why the
required times for the three TCDs to reach a steady state were
very close to one another. Although the membranes in these
TCDs were different, the silicon packages were identical.
Time variation of the average temperature of the silicon
package can be estimated from the following equation for a
lumped system:
rVcdT
dt¼ Qmembrane�wafer � h Awafer�airðT � TairÞ
� Qwafer�rack (1)
where r, V, c, T are the density, volume, specific heat, and
average temperature of the silicon wafers; h is the heat
transfer coefficient; and t the time variable. The three terms
on the RHS represent the heat input to the silicon package
from the heater, the convective heat loss from package
surfaces, and the conductive heat loss to the metallic rack
beneath the TCDs during the transient heat transfer test. The
conductive loss term can be expressed as:
Qwafer�rack ¼ ðT � TrackÞRcond
(2)
The conduction resistance, Rcond, is proportional to the length
and inversely proportional to the thickness and thermal
conductivity of the silicon package. The rack temperature
remained nearly constant during the test.
Since the test chamber was evacuated, there was no
convective heat loss during the transient test. The complete
solution to the heat equation with h ¼ 0 consists of a steady-
state and a transient solution:
T�Track ¼ ys þ yt (3)
where
ys ¼ Qmembrane�waferRcond (4)
yt ¼ ðTinitial � Track � ysÞ exp�t
RcondrVc
� �(5)
The thermal time constant of the transient process is there-
fore, equal to RcondrVc. The package was made up of two
4 in:� 4 in. silicon wafers glued together. There are two sets
of flow channels and heaters fabricated at the center of the
silicon wafers. As a result only one half of the package was
considered in the transient conduction analysis.
The thermal capacitance was computed from:
ðrcÞðVÞ ¼ ð1:6� 106 Jm�3 K�1Þð2� 350mm� 4 in:� 2in:Þ¼ 5:78JK�1 (6)
The conduction resistance was estimated from:
The thermal time constant calculated from the above
estimation is about 28 s for the three TCDs. This estimate
is in good agreement with the transient current plots in
Figs. 4 and 5. The noise levels in these current measurements
seemed to be dependent primarily upon the thermal capa-
citances of the heaters. The good signal conditions of the
polyimide membrane are probably attributed to its large
volume.
3.2. Heat transfer in TCDs of different flow-channel
designs
Heat transfer characteristics of single- and double-chan-
nel TCDs were compared and analyzed in this test. The
polyimide membrane was selected since it exhibited the best
signal conditions in the transit time test. A single-channel
TCD and a double-channel TCD made from the same
materials were placed in the stainless-steel chamber. The
chamber was filled with helium gas—a carrier gas com-
monly used in GC applications. The stagnant gas in the flow
channels simulated the conduction-dominant heat transfer
process in TCD operations. A thermocouple was attached to
the lower channel surface of the double-channel TCD for
measuring the silicon wafer temperature. The measured
surface temperature of the double-channel TCD together
with the electric currents of both TCDs is presented in Fig. 6.
The average electric resistances of the heating elements
were 822 and 773 O for the single- and double-channel
TCDs at a supplied voltage of 14 V. Deduced from the
temperature coefficient of resistance, the heating element
temperatures of the single- and double-channel TCDs were
193.5 and 176.3 8C, respectively at this supplied voltage.
Assuming the outer surface temperatures of the upper and
lower flow channels were close to each other, the heat fluxes
through the upper and lower flow channels can be calculated
from:
qu ¼ ðTNi � TSiÞsum of ðthickness=thermal conductivityÞ
¼ ðTNi � TSiÞðL=kÞhelium in upper channel þ ðL=kÞupper Si wafer
¼ 1:33 MW m�2 (8)
Rcond ¼ wafer length=2
silicon thermal conductivity � 2 � wafer thickness � wafer width
¼ 4 in:=2
148 W m�1 K�1 � 2 � 350 mm � 4 in:Þ¼ 4:83 K W�1 (7)
42 Y.E. Wu et al. / Sensors and Actuators A 100 (2002) 37–45
ql ¼ðTNi �TSiÞ
ðL=kÞmembraneþðL=kÞheliumin lowerchannelþðL=kÞlowerSiwafer
¼ 0:237MWm�2 (9)
And the total heat transfer rate of the double-channel TCD
is the sum of qu and ql multiplied by the channel floor area.
The calculated heat transfer rate of 0.283 W is 12% higher
than the power consumption of the electric heater, which
was determined from the product of the heater current and
voltage.
A couple of causes may contribute to the discrepancy
between the heat transfer model and experimental measure-
ments. For one thing, the back of the thermocouple bead was
exposed to the cool surroundings and the bead diameter was
larger than the flow channel width. As a result the thermo-
couple temperature was slightly lower than the temperature
of the channel surface it measured. The major reason is
probably the overestimation of the membrane surface tem-
perature. Since the heating element did not cover the entire
floor of the flow channel, the average floor temperature
should be lower than the heating element temperature.
The 12% higher heat transfer rate of the theoretical solution
indicates the average floor temperature should be about
18 8C lower than the heating element temperature TNi.
It is very difficult to accurately measure the gas tempera-
ture in the channel under the polyimide membrane or the
membrane surface temperature for the single-channel TCD.
It was assumed in our analysis that the outer surface
temperature of the open silicon channel and the gas tem-
perature at the channel opening (point o in Fig. 1(b)) were
close to the lower channel surface temperature of the double-
channel TCD. The calculated heat fluxes through the closed
and open channels of the single-channel TCD are:
qu ¼ ðTNi � TSiÞðL=kÞhelium in upper channel þ ðL=kÞupper Si wafer
¼ 1:49 MW m�2 (10)
ql ¼ðTNi � TSiÞ
½ðL=kÞmembrane þ ðL=kÞhelium in open channel�¼ 0:0118 MW m�2 (11)
The power consumption is : Q ¼ ðqu þ qlÞA ¼ 0:271 W
(12)
Results of the above heat transfer analysis changed only
slightly if the gas temperature at the channel opening was
assumed to be the surroundings’ temperature of 24 8C.
When the single-channel TCD is operated in atmosphere,
the open channel will be filled with air. If the thermal
conductivity of air were used in Eq. (11), ql would change
to 0.0582 MW m�2, but the TCD power consumption, Q,
would increase by 3% only.
Just like the double-channel results, the calculated heat
transfer rate of the single-channel TCD is 14% higher than
the measured one, indicating the average floor temperature
was again slightly lower than the heating element tempera-
ture. If the floor temperatures could be more accurately
determined, the one-dimensional heat transfer model should
yield better agreement with the experimental data. Never-
theless our model showed the heat transfer to the helium gas
in the double-channel TCD was about 20% higher than that
in the single-channel design at a supplied voltage of 14 V.
The difference in heat transfer rates agreed well with
experimental measurements.
Although more heat was transferred to the helium gas in
the double-channel TCD, the double-channel design con-
sumed more electricity for a desired heater temperature.
Theoretically heat transfer to the gas flow in an ideal double-
channel TCD is 100% more than that in an ideal single-
channel TCD, and the sensitivity is doubled too. In reality
the sensitivities of the two designs do not differ as much
as the heat transfer rates due to thermal resistances other
than the gas resistances in the heat flow paths. The TCD
sensitivity is a measure of the influence of the change in gas
Fig. 6. Measured wafer temperature and heater currents at different voltages.
Y.E. Wu et al. / Sensors and Actuators A 100 (2002) 37–45 43
thermal conductivity on the heater power. A comparison of
the sensitivities of the two flow channel designs is presented
in Fig. 7. This comparison was made for TNi ¼ 500 K and
TSi ¼ To ¼ 300 K. The abscissa of the plot is the gas
thermal conductivity normalized by helium thermal con-
ductivity at 400 K. Fig. 7 shows that while the heat flux of
the double-channel TCD maintains approximately 20%
higher when the gas thermal conductivity varies from 0.9
to 1.1 of khelium, the changes in heater power consumptions
of the two TCDs are very close. In this comparison dq00 of the
double-channel TCD is only about 3% higher than that of the
single-channel TCD. Better sensitivity improvement over
the single-channel design can be achieved if the double-
channel TCD uses thinner membrane and silicon wafers.
4. Conclusions
TCDs of different flow channel designs and heater mate-
rials were fabricated and their performances and transient
behaviors characterized. One-dimensional heat transfer and
lumped system models were developed and compared with
experimental observations. It was found in the transient test
that the TCD response time depended primarily on the
dimensions of the silicon package. The power input affected
only the signal’s noise level. The double-channel design
transferred more heat to the gas flow than the single-channel
design, but the increase in sensitivity due to the additional
flow channel was not as significant as the increase in heat
transfer rate. The average heater surface temperature was
slightly lower than the heating element temperature deduced
from the resistance measurement.
Acknowledgements
This research was supported by the Chung Shan Institute
of Science and Technology (CSIST). The advice and
information provided by Dr. S.W. Ko of CSIST is greatly
appreciated.
References
[1] G.C.M. Meijer, A.W. Herwaarder, Thermal Sensors, Institute of
Physics Publishing, Philadelphia, PA, 1994, pp. 7–13.
[2] U. Dillner, E. Kessler, S. Poser, V. Baier, J. Muller, Low power
consumption thermal gas-flow sensor based on thermopiles of highly
effective thermoelectric materials, Sens. Actuators A 60 (1997) 1–4.
[3] H.A. Daynes, Gas Analysis by Measurement of Thermal Conductiv-
ity, Cambridge University Press, Cambridge, Great Britain, 1933,
pp. 1–302.
[4] M. Madou, Fundamentals of Microfabrication, Boca Raton, NY,
1997, pp. 1–177.
[5] E.S. Kolesar Jr., Review and summary of a silicon micromachined
gas chromatography system , IEEE, Trans. Components, Packaging
Manufac. Technol. B 21 (4) (1998) 324–328.
[6] J.H. Jerman, A Miniature Thin-Film Thermal Conductivity Detector
for An Integrated Gas Chromatograph, Ph.D. Thesis, Stanford
University, Stanford, CA, 1981.
[7] K. Chen, Y.E. Wu, Thermal analysis and simulation of the
microchannel flow in miniature thermal conductivity detectors, Sens.
Actuators A79 (2000) 211–218.
[8] W.A. Wiseman, Katharometer behavior, Ann. N.Y. Acad. Sci. 72
(1959) 685–697.
[9] D.M. Rosie, Quantitation of thermal conductivity detectors, J.
Chromatogr. Sci. 11 (1973) 237–250.
[10] J. Sevcik, Detectors in gas chromatography, J. Chromatogr. Library 4
(1976) 39–58.
[11] Y. Huang, H.H. Bau, The effects of forced convection on the power
dissipation of constant-temperature thermal conductivity sensors,
ASME J. Heat Transfer 119 (1997) 30–37.
[12] S.C. Terry, J.H. Jerman, J.B. Angell, A gas chromatographic air
analyzer fabricated on a silicon wafer, IEEE Trans. Electr. Devices
ED-26 (12) (1979) 1880–1886.
[13] H. Robbins, B. Schwartz, Chemical etching of silicon; part I, J.
Electrochem. Soc. 106 (1959) 505–508.
[14] A.F.P. Van Putten, A constant voltage constant current Wheatstone
bridge configuration, Sens. Actuators 13 (1988) 103–109.
[15] M. Despont, H. Gross, F. Arrouy, C. Stebler, U. Staufer, Fabrication
of a silicon–Pyrex–silicon stack by a.c. anodic bonding, Sens.
Actuators A 55 (1996) 219–224.
Fig. 7. Heat fluxes and heat flux changes of single- and double-channel TCDs. [dq00 ¼ q00ðkÞ � q00ðk ¼ kheliumÞ].
44 Y.E. Wu et al. / Sensors and Actuators A 100 (2002) 37–45
[16] A. Cozma, B. Puers, Characterization of electrostatic bonding of
silicon and Pyrex glass, J. Micromech. Microeng. 5 (2) (1995) 98–
102.
[17] R.F. Lacey, Thermal Conductivity Detector, US Patent 5,379,630,
1995.
[18] U. Bonne, D. Kubisiak, Flowmeter Fluid Composition and
Temperature Correction, US Patent 5,237,523 (1993).
[19] U. Bonne, Absolute Pressure Sensor and Method, US Patent
5,303,167 (1994).
[20] F.P. Incropera, D.P. DeWitt, Fundamentals of Heat and Mass
Transfer, 5th Edition, Wiley, New York, NY, 2001, pp. 240–254.
Biographies
Ye-Ee Wu received the PhD degree from the Department of Chemical
Engineering and Material Science, Syracuse University, in 1983. He has
been an Associate Professor in the Mechanical Engineering Department,
National Taiwan University of Science and Technology, since 1985. Prior
to that he worked for the China Steel Corporation for 2 years as a research
scientist. Dr. Wu’s research interests include microfabrication technology,
non-destructive evaluation, and mechanical metallurgy.
Kuan Chen received his PhD degree from the University of Illinois at Urbana-
Champaign in 1981. Immediately thereafter he jointed the Faculty of
Mechanical Engineering of the University of Utah, and became an Associate
Professor in 1988. He was a professor in the Mechanical Engineering
Department, National Taiwan University of Science and Technology, from
1997 to 1999. His current research interests include microfabrication, thermal
plasmas, thermoelectrics, and microscale thermal systems and phenomena.
Chao-Wen Chen received his MS degree from the Department of
Mechanical Engineering, National Taiwan University of Science and
Technology, in 2000. He is now serving in the Army as a Lieutenant to
fulfill his military obligation.
Keng-Hao Hsu received his MS degree from the Department of
Mechanical Engineering, National Taiwan University of Science and
Technology, in 2000. He is now serving in the air force as a Sergeant to
fulfill his military obligation.
Y.E. Wu et al. / Sensors and Actuators A 100 (2002) 37–45 45