Page 1: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Extending the Limits of Detection into the PPT Range

with a QMS

Jonathan Leslie Mass Spectrometry Solutions

MKS Instruments, Inc. RGA 7 October, 2016

Page 2: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Introduction

Quadrupole Mass Spectrometers (QMS) are used in many process applications from fuel cells to semiconductor manufacturing and from environmental monitoring to high energy physics.

In recent years, QMS has increasingly been used for process monitoring in a wide range of vacuum processing techniques requiring detection of low-level contaminants.

Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the background signal from the RGA itself and interference from bulk gases can make the detection of ppm levels of some contaminants difficult in reality. 2

Page 3: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Introduction (cont’d) For applications operating at high pressure above10-4 mbar,

background contribution can be substantially reduced with the use of a closed ion source (CIS).

The closed ion source extends the detection level in to the ppb range and combined with MKS V-lensTM technology in to the sub 10 ppb range.

To increase the limit of detection further the sampling of the analysis gas needs to be enriched. Such techniques are now available allowing detection in to the parts per trillion (ppt) range.

3

Page 4: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Open Ion Source

The open ion source RGA is typically placed inside the vacuum chamber to be monitored.

At higher pressure, above 10-5 mbar the source becomes non-linear, above 10-4 mbar the ion count reduces.

To operate at higher process pressure, the RGA must be fitted in a separate vacuum chamber and sample gas leaked in.

4 aetsepee

Souce C

tact outg

Page 5: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Closed Ion Source

The closed ion source (CIS) mass spectrometer samples gas from an external chamber.

The pressure in the source is much higher than a standard open ion source RGA, typically 1 x 10-3 mbar.

The higher partial pressures of trace gases increases the sensitivity compared to RGA background.

5

Gas Inlet

Page 6: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Closed Ion Source Sampling

6

Background: H2 , H2O, CO / N2, CO2

Orifice

10 mbar 10 -3 mbar

10 -5 mbar

10 -5 mbar 10 -9 mbar

Process Chamber

Filament

Mass Analyzer Closed Ion Source

Orifice

Total Pressure

Impurity Partial Pressure

Page 7: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Example Sample Inlet

7

Page 8: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Effect of Source Pressure

8

0.00E+00

1.00E-03

2.00E-03

3.00E-03

4.00E-03

5.00E-03

6.00E-03

0.00E+00 4.00E-03 8.00E-03 1.20E-02 1.60E-02

Rep

orte

d Pr

essu

re

Actual Pressure

As source pressure is increased, response becomes non-linear and eventually starts to drop, this is known as roll over.

Page 9: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

0

50

100

150

200

250

0.00 20.00 40.00 60.00 80.00 100.00

Sign

al (p

pb)

m/z

No Deflector

0

50

100

150

200

250

0.00 20.00 40.00 60.00 80.00 100.00

Sign

al (p

pb)

m/z

No DeflectorOff Axis

Reduction of low mass noise using V-lens

9

The high pressure source allows the production of metastable neutrals, this is especially bad with Argon.

Page 10: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Removal of Line of Sight

10

Removal of line of sight reduced the effect of metastable neutrals completely giving a non-gas dependant baseline.

The improved ion optics also increased system sensitivity

Page 11: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Best Sensitivity of System

Sampling from a chamber at 1e-3 mbar of air with a Vision 2000P-XDTM xenon concentration of 90 ppb, dwell time 1.2 seconds per data point, multiplier gain 3000.

11

24 ppb

4 ppb

19 ppb

24 ppb

9 ppb 8 ppb

Page 12: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Further Reductions in Limit of Detection by Sample Enrichment

A membrane inlet system was built for sample enrichment.

Sample gas was passed over a polydimethylsiloxane membrane at about 50 sccm.

Various thicknesses and numbers of membranes were tested, a single 0.2mm membrane was chosen.

The response time when switching the sample was around 10 s to reach 90% of final response.

12

Gas in

Gas out

Membrane

Connection to RGA

Page 13: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

A Universal Solution?

13

From: THIN SILICONE MEMBRANES-THEIR PERMEATION PROPERTIES AND SOME APPLICATIONS Annals of the New York Academy of Sciences, vol. 146, issue 1 Materials in , pp. 119-137 W. L. Robb

GAS NAME FORMULA SILICONE PERMEABILITY COEFFICIENT (Barrer)

Nitrogen N2 280

Oxygen O2 600

Argon Ar 600

Hydrogen H2 650

Helium He 350

Methane CH4 950

Ethane C2H6 2500

Carbon dioxide CO2 3250

Toluene C7H8 9130

Hydrogen Sulphide H2S 10000

Benzene C6H6 10800

Sulphur Dioxide S2O 15000

Water H2O 36000

Page 14: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

BTX in Nitrogen

14 toge

50

yee

50

ouee

50

eeeaace C C austG

Oe at

00

C

Cus

ebae0

00

65

0

00

0

00

0

63

The gas mix was automatically changed every 2 hours, the sequence being

Page 15: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Results of 7 Day Run

15

Over the 7 days the drift in sensitivity was approximately 20% but the reproducibility was good and all steps are clearly visible for all masses even when going from 0 to 630 PPT.

Page 16: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Data Analysis 1

16

The data from the last 3 cycles taking mass 78 for Benzene (black) 92 for Toluene (green) 106 for P-Xylene (blue)

Page 17: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Data Analysis 2

17

Using the average of the peak heights and comparing to the actual concentrations (based on the dilution of the test gas) the system shows good linearity.

0.00E+001.00E-042.00E-043.00E-044.00E-045.00E-046.00E-047.00E-048.00E-049.00E-041.00E-03

0 1 2 3 4 5

Peak

Hei

ght (

mB

ar)

Actual Concentration (PPB)

Peak Height vs Actual Concentration

BenzeneTolueneP-Xylene

Page 18: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

BTX Results

Benzene PPB Toluene PPB P-Xylene PPB Background 0.59 0.36 0.39 Standard Deviation 0.04 0.04 0.06 3 x Standard Deviation 0.12 0.12 0.18

The membrane inlet gives good stability and linearity for BTX at low concentrations with a detection limit in the 40-60 parts per trillion range.

The most intense peak for Toluene and Xylene is m/z 91, if there was no need to differentiate these 2 compounds the limit of detection would be lower.

18

Page 19: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Coping with Interfering Peaks

In the BTX example the interferences between the sample peaks raised the limit of detection.

Interferences between compounds of interest and residual or matrix gases can also cause limitations in detectability.

There are many instances of this, for example it is common to want to monitor CO in air. – CO has a mass of 28 – Nitrogen also has a mass of 28 – CO2 with a mass of 44 also gives a fragment ion at mass 28

19

Page 20: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

H2S Detection Using CirrusTM

Analog spectrum of 10 ppm H2S in a balance of N2

20

Gives a detection limit for H2S of 50 ppb and a precision on the 10 ppm reading of +/- 0.5 ppm with a sample time of 2 seconds

Page 21: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Air Spectrum Using Cirrus Scanning the same peaks in air but with no H2S

21

Gives an equivalent H2S background due to O2 isotope at 34 amu of 2000 ppm.

Using mass 32 and a subtraction for mass 34 would possibly improve detection limit to 50 ppm.

Page 22: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Effect of Electron Energy Measuring the peaks at lower electron energies can be

effective, if monitoring NH3, water interference can be removed by scanning 17 eV or below.

22

Since oxygen is easier to ionise electron energy cannot be used to remove it’s interference.

The interference of H2S at mass 34 can be removed by running below 25 eV

Page 23: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Universal Calibration Software

Interfering peaks can be subtracted automatically with the Universal Calibration Software.

This software allows a spectrum of each pure compound to be saved as a calibration standard.

If all the major components are known, the overall composition can be de-convolved using simultaneous linear equations.

23

Page 24: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

H2S in Synthetic Air Measuring a 7.7, 3.3, 0.7 and 0.2 ppm H2S in Air mix

24

Results were not perfectly linear Implies a detection limit of around 0.5 ppm

Page 25: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Effect of wet and dry air

25

Water(18) O2 (32) N2 (29)

H2S (34)

Gas mixer switched between dry N2 and N2 bubbled through water.

The effect on H2S of several hundred times increase in water vapour is small.

Page 26: Extending the Limits of Detection into the PPT Range with ...Many modern designed residual gas analysers do have the sensitivity to detect sub-ppm level contaminants. However, the

Summary

The membrane inlet allows fast enrichment of a wide range of sample molecules.

The response time is equivalent to the standard capillary inlet. BTX with no interfering peaks showed an LOD as low as 40ppt. H2S in air has the LOD reduced from 2000 ppm to 30 ppm using

the membrane inlet, this can be lowered to less than 1 ppm with the subtraction of interfering peaks.

The system ran reliably for 8 months without significant loss in sensitivity.

26