experimental - inflibnetshodhganga.inflibnet.ac.in/bitstream/10603/763/8/08_chapter 2.pdf · matec...
TRANSCRIPT
Chapter 11 Eperimen~dlechniques.. .
EXPERIMENTAL TECHNIQUES
The experimental techniques employed in the present study to determine the
ultrasonic veloc~ty, absorption, density and viscosity are described in this chapter All
the chemicals used in this present work are of AR/BDH quality and are used as such
without further purification.
2.1. ULTRASONIC LIQUID CELL:
The ultrasonic velocity and absorption measurements are done with a ultrasonic
liquid cell and the cell is designed and fabricated in this laboratory.
The cell fabricated in this laboratory is a double walled stainless steel cell with a
central uniform bore of 24.5mm diameter and a depth of 23.89mm with a perfectly
plane, highly polished bottom surface(f1oor). A stainless steel circular d ~ s c of 65mm
diameter with inner hole of 22mm diameter and 5mm length is designed to fix the PEZ
crystal. It has two 8mm Allen bolts and springs to moves up and down the PEZ crystal
in the bore. Through the fine adjustment of the Allen bolts with an Allen key,
parallelism can be achieved between the floor of the cell, which acts as a reflector and
the PEZ crystal. The cell has an outer shell through which water can be circulated. It
has an inlet and outlet pipes through which the water is circulated in order to malntain
the temperature of the solution constant. The cell diagram is shown in figure 2.1. The
temperature of the circulating water is maintained at a fixed temperature by a
thermostatically controlled water bath. The temperature is maintained constant with in
M.01K. A two terminal digital thermometer is designed and fabricated in this
laboratory for measuring the temperature of the measuring liquid and the water
circulated through the outer shell of the cell. Two small provisions are provided in the
cell for this purpose. The thermometer has an accuracy of 10.01K.
Figure 2.1 Ultrasonic Liquid Cell Diagram
Chapter I I ~ ~ e r i m e n ~ d T e c h n i ~ u e s . . .
The ultrasonic velocity and absorption are measured using pulse echo overlap
and pulse echo technique respectively. The experimental setup is shown in figure 2.2 &
figure 2.3 and the echo pattern is shown in figure 2.4.
2.2, ULTRASONIC ABSORPTION MEASUREMENT:
2.2.1. PULSED POWER OSCILLATOR:
A sharp RF electrical pulse with variable pulse duration from 2ps to 20ps is
applied to the crystal with a suitable repetition frequency. The reflected wave returns to
the transducer and a part of its energy is converted into electrical signal. This signal is
amplified and displayed on the oscilloscope. The trace of the oscilloscope is a series of
echoes with decreasing amplitude. The decrease in the amplitude is a measure of
absorption.
The schematic block diagram of the PULSED POWER OSCILLATOR (EL - 550)
is shown in the figure 2.5. It is basically a 100 watts pulsed RF power source designed to
generate low distortion fast rising / falling pulsed RF signals. It generates RF pulses
from 1 to 50 MHz with use of six plug - in coils. The pulse repetition rate of the pulsed
power oscillator is variable from 50 to 2500 Hz. The accurate RF signal output from PPO
is measured using a frequency counter (APLAB - 1116 UNIVERSAL TIME AND
FREQUENCY COUNTER). In present the study, the ultrasonic absorption
measurements are carried out using t h s pulsed power oscillator in the frequency range
3 - 11MHz.
The RF signal is fed to the cell through a matchng network (Pre Amplifier)
designed in this laboratory. The matching net work (Pre Amplifier) circuit has been
incorporated from the ULTRASONIC PULSE ECHO INTERFEROMETER
(SD UI - 003). The diagram of the matching network is shown in the figure 2.6. It is
acting as a transmitter and receiver for the cell through 50 Ohm coaxial cable, i.e., it
transmits the RF signal from the PPO to the cell and receives the output signal from the
Figure 2.2 Experimental setup for Pulsed Power Oscillator and Ultrasonic Time Intervalormeter
Figure 2.3 Experimental setup for MATEC 7700 model Ultrasonic System
Figure 2.4 Echo pattern
Matching Network
Trig in 1 Trig Out
Figure 2.5 Pulsed Power Oscillator Block Diagram
1 2 . 6 ~ AC
Power Supplies
+300 v DC
Prf & Trig Int - Ext
v v V
1 2 . 6 ~ AC
+700v DC
_, Pulse Width . -
+70v DC I
Power Oscillator
Buffer Amplifer
Pufse CW / Pulse Amplifer
Pulsed Power Oscillator
Figure 2.6 Matching Network (Pre - Amplifier) Circuit
Chapter 11 EperimentdTechniques.. .
cell. Then this matching network output is fed to the triggered oscilloscope
[BPL - INDIA DTO 31500 OSCILLOSCOPE]. All the power supplies of the system are
electrically regulated.
The echo heights are fitted to an exponentially decaying ~7ave of the form Ae-'"
and the value of a is evaluated and expressed in nepersjm. The distance between crystal
and floor is measured and the absorption per unit length is calculated and expressed in
nepers/m. It is assumed that the floor reflects the entire sound energy incident upon it.
The accuracy in the measurement of absorption is 3%.
2.2.2. MATEC 7700 MODEL ULTRASONIC SYSTEM:
The experimental setup used for making the ultrasonic absorption measurements
in the frequency 25 - 89 MHz consists of mainly the PEO system, the temperature
controlling system, the cryostat for low temperature measurements and the water bath
for h g h temperature measurements.
The PEO system consists of MATEC model 7700 pulse modulator and receiver
together with model 760 V rf plug-in, model 110 high resolution frequency source,
model 1128 decade divider and dual delay generator, model 2470B attenuation
recorder, model 70 impedance matching network etc. The frequency counter used as
HIL (Inch) model 2722 and the oscilloscope was a 100 MHz one with z-axis input
(HIL model 5022).
The block diagram of the experimental setup used is shown in figure 2.7. The
tunable cw soruce (model 110) has a highly stable internal high frequency oscillator
(12-50 MHz) from whch the required low frequency cw signal for PEO is generated by
selectable frequency division. This signal is available at terminal 2, while the high
frequency is available at terminal 1 for accurate counting by the frequency counter
(model 2722). The dual delay and divider unit (model 122B) has dividers selectable as
10, 100 and 100. The division factor 100 is quite acceptable for most measurements,
Frequency Tunable CW Counter Source ( I 10)
1 CW out -- -
1 2 Trigger Pulse (DirectIDivided - SW2)
3 Intensifying Pulse
I
Dual Delay & Divider ( I 22B)
Divided
Trig.
Attenuation Recorder Out -
(2470B)
Trig. In Video In
Intense. -mG--j, SWI*.
Trig. In
(7700) Receiver Out yt Recert- 100 MHz
Oscilloscope (5022)
Pulse Modulator & video+ CH2 Ext. Trig. Receiver 10-90 MHz Out
1 I
-
Impedance
.+
I ultrasonic I 7
Figure 2.7 Block Diagram of the MATEC 7700 Model
I Liquid Cell
Ultrasonic System
1 Digital Thermometer I
Chapter I I 'E,perimenid%chniques. . .
which means that the next rf pulse is sent to the sample only after a time interval
corresponding to 100 number of echoes. The terminal 2 of this unit gives trigger pulses
for the CIIO. The CRO is always operated in the external sweep trlgger mode. By using
switch SW2 in 1228 the trigger at terminal 2 can be made direct trigger or dilrided
trigger for observing the overlapped echoes or the full echo pattern respectively. The
dual delay generators in 122B can be adjusted for delay and width of the intensifying
pulses for selecting the two echoes of interest for overlap. These pulses are available at
terminal 3 and are connected to the z-input of the CRO through the selector switch
SW1.
The divided trigger from 1228 goes to the pulse modulator and receiver unit
(model 7700 with rf-plug-in in model 760 V). This is the most important unit in the
setup. An rf pulse packet of peak power 1 kW is obtained at the output when the unitis
triggered at the input. The rf frequency of the triggered power oscillator can be adjusted
in the range of 10 to 90 MHz. the width and amplitude of the pulse are adjustable. For
good pulse shape the unit is usually operated at full amplitude and the amplitude
reduction is achieved by using an rf attenuator (Alan attenuator) at the output as shown
in the block diagram. The unit has a sensitive tunable superhetrodyne receiver with a
maximum gain of 100 dB for amplifying the echoes. The amplified echoes are available
through the receiver out terminal whch is connected to the CRO channel 1 input. The
amplified echoes are also detected and the detected output (envelope of the echoes) is
available at the video out terminal which is connected to the attenuation recorder and
CRO channel 2 input. For optimum signal to noise ratio in the amplification of weak
echoes, an impedance matching network (model 70) is connected before the PEZ
transducer.
The attenuation can be most conveniently measured by an automatic procedure
for whch commercial equipment is available (Matec. hc . (USA) Model 2470). In the
automatic system, two gates with variable delay are set on the two echoes of interest to
Chapter I I EperirnentdTechniques. . .
sample them. The amplitude of the first echo 1s held constant by AVC circuitry, and the
amplitude of the second echo is sampled at its peak. A calibrated logarithmic amplifier
converts the sampled amplitude to decibels relative to the constant amplitude of the
first echo. The decibel level is recorded on a built-in strip chart that has several
calibrated scales and a variable baseline, so that small changes in attenuation can be
measured at various total loss levels. In this equipment the attenuation can also be
noted from panel mater calibrated in dB. The swritch SWl can be toggled to pole 2 for
connecting the intensifying pulse output from the attenuation recorder to the Z-input of
the CRO. By this way the two echoes of interest can be selected for attenuation
measurement. One important advantage of the PEO technique is the capability to
measure the velocity and attenuation at the same time. In present the study this
MATEC model 7700 ultrasonic system is used only for measuring attenuation in pulse
echo mode in the frequency range 25 - 90 MHz.
2.3. MEASUREMENT OF ULTRASONIC VELOCITY:
The ultrasonic velocity is measured using the pulse echo overlap technique. The
pulse echo overlap (PEO) method is very versatile and hghly accurate technique for
measuring the velocity of the ultrasonic waves in materials and structures. A high
absolute accuracy is capable of accurately measuring from any cycle of one echo to the
corresponding cycle of the next echo. The pulse-echo-overlap method is able to handle
diffraction (beam spreading) and phase corrections, properly, so that absolute accuracy
of the PEO method may exceed the accuracy of most of the other methods.
The block diagram of pulse echo overlap measurements with broad pulses is
shown in figure 2.8. The principle of measurement is to make the two signals of interest
overlap on the oscilloscope by driving the X-axis with a frequency whose period is the
travel time between the signals of interest. Then one signal appears on one sweep of the
oscilloscope, and the other signal appears on the next sweep. The X-axis sweep
r
CW - Pulse Pulse
oscillator . 5 Pulser - Limiter
- 'I
Counter Cell
Figure 2.8 Block diagram of the Pulse Echo Overlap method
Chapter 11 E~erimentdlec/migues . . .
frequency is supplied by a Cn7 oscillator. For jitter-free overlap, the signals of interest
must be synchronized with the phase of the CW voltage. This condition is achielred by
the repetition rate of the input pulse from the phase of the CW voltage by a
frequency divider. Division by a large number (e.g. 1000) allows all the echoes from one
pulse to be attenuated before the next pulse is applied. The output of the frequency
divider is a trigger signal synchronous, with the phase of the CW voltage. The trigger
signal triggers the main pulser, which pulse the transducer. A diode limiter circuit
keeps the input pulse from overloading the amplifier. The main pulser also triggers two
intensifying pulses, which are applied to the cathode ray tube to intensify the trace. This
feature is necessary to distinguish the two signals of interest from the rest of the echoes
in the trace.
The ULTRASONIC TIME INTERVALOMETER (UTI-101) is intended for
precise measurement of ultrasonic velocity in solids and liquids using pulse-echo-
overlap technique. The absolute accuracy of velocity measurement using this technique
may be as high as 2 parts in lo4, while the relative sensitivity (AV) can be as high as 10
parts in lo6. The instrument uses a broadband pulse to excite the transducer. All the
circuitry required for pulse-echo-overlap such as high voltage pulse to excite the
transducer, a continuous wave oscillator with high resolution and low phase jitter,
delayed strobe pulse generators to aid intensification of the trace, eight digit frequency
counter are built into this compact instrument. One of the important features of this
instrument is provision of extra facility, which enables one to achieve overlap with
scopes having no intensity modulation facility. The accuracy in the measurement of the
velocity is kO.OOlrn~-~.
Chapter I I E~enmrrtallechniques. . .
2.4. MEASUREMENT OF DENSITY:
The density of the solution is measured using a specific gravity bottle of capacity
10ml. It is filled with the solution carefully up to the fixed marks. The mass of the
solution is iound using a single pan balance [SARTORIUS] with an accuracy of
0.001gm. The density p is computed using the relation
where m is the mass of the solution and V is the volume of the specific gravity bottle.
The accuracy of the measurement is 0.001x103kg/rn3.
2.5. MEASUREMENT OF VISCOSIT\I':
The shear viscosity of solutions is determined using an OSTWALD'S
VISCOMETER. Using the values of viscosity of water at 303K from literature, and by
measuring the time of flow of the solution from the viscometer, the shear viscosity is
where qs, pl, ti, are the viscosity, density, and t ~ m e of flow of solution in the viscometer
respectively and qo,po,to are the corresponding values for the distilled water. The
accuracy in the measurement of the measurement is M.ls.
2.6. JEOL - JES - TElOO ELECTRON SPIN RESONANCE SPECTROMETER:
The schematic diagram of the instrumentation of JEOL - JES - TElOO EPR
spectrometer is shown in figure 2.9. The main principal components of the instrument
comprises of Source, Resonator, Magnetic field and Detector.
Recorder
r - - - - - - - - -
- - . copel
I
I
1 I
- A
I I
L d
Figure 2.9 Block diagram of a typical X-band EPR spectrometer
Chapter I I E~erirrzentdlechnigues.. .
2.6.1. SOURCE:
The microwave source is ~~sual ly the klystron. It is a vacuum tube that is
characteristic for its low noise property. The valve of this klystron consists of a heater
filament, a cathode, a reflector filament and a resonant cavity in which the electrical
oscillations are maintained. The field is generated by oscillations within its own tunable
cavity.
2.6.2. RESONATOR:
This is a resonant cavity that allows the microwave through the hole called iris.
The frequency of the source is tuned to the appropriate resonate frequency of the cavity.
The corresponding w7avelength of the resonant are related to the cavity dimensions.
2.6.3. MAGNETIC FIELD:
The static magnetic field for the EPIi must be stable. An electromagnet of field
strength of about 10000 gauss is connected with the sweep generator and the subsidiary
coils to provide variable field strength. The sample is placed in the magnetic field of a
cavity resonator. The sample placed inside the cavity is such that the static magnetic
field is perpendicular to the microu7ave magnetic field.
2.6.4. DETECTOR:
All modern EPR spectrometers use the semiconducting crystal diode rectifiers as
the basis for the detecting system. The essential requirements for the detector is that the
signal leaving the crystal with the frequency and amplitude corresponds to the
modulation frequency applied the sample. The subsequent amplification of the signal
with or without phase selection leads the ultimate display of the spectrum either by an
oscilloscope or pen recorder.
Chapter 11 E ~ p e ~ m ~ n t d l e c h n q u e s . . .
The work carried out in this thesis is done on a jEOL JES-TE100 ESIt
spectrometer operating at X-band frequenc~es, ha~ring a 100 kHz field modulatior~ to
obtain a first derivative EIJR spectrum. DPIJH, with a g value of 2.0036, has been used
for g-factor calculations.
2.6.5. INTERPRETATION OF EPR SPECTRA:
As one can measure EIJR spectra for solution, powder and single crystal samples,
the procedure to obtain spin Hamiltonian parameters from these spectra must be
known. In order to calculate g and A values, the follomring expression is used,
g = ( g u r ~ B~IJI 'H )/ B
where,
B is the magnetic field position at the EPR peak
BDPP~: is the field position corresponding to DPPH
g~~1~1.i is the g-value of DI'PH (g = 2.0036)
One can as well calculate the g-value d~rectiy using the spectrometer frequency
at wluch resonance occurs. The expression is as follows,
g = (hv / PB)
where, v is the resonance frequency.
The hyperfine (hf) coupling constant 'A' is given by the field separation between
the hyperfine components. If the spacing is unequal, an average of them is taken to be
the value for A. For n number of hyperfine lines, the average hyperfine value is,
A = ( B n - B I ) / ( ~ - 1 )
where,
Bn is the field position for the ntk hyperfine line
BI is the first hyperfine line field position.
Chapter I I E;i-pe~trzentailech~ziques.. .
2.7. COMPUTATION:
From the measured parameters viz., density, viscosity, velocity and attenuation,
the other parameters viz., adiabatic compressibillty, free length, observed absorption,
classical absorption and excess absorption are computed using the standard relations.
The absorption coefficient (a) is computed from the measured attenuation values using
the straight line fitting method. The computation of parameters is done with the help of
a "C" program written for this purpose and be given In Appendix B.