English Owner´s Manual Gnatus Intra Micromotor 32 MM with TB Spray, TB Spray, TM Spray without with 0499 Gnatus Intra Micromotor SL 30 TB Spray, TB Spray, TM Spray / TM SI / TB SI with without with 01 - Gnatus Intra Micromotor 02 - Coupling lining 03 - O-ring. GENERAL DATA Dimensions (mm) 01 01 01 02 02 03 BORDEN TYPE W/ EXTERNAL SPRAY BORDEN TYPE W/O SPRAY MIDWEST TYPE W/ EXTERNAL SPRAY TYPE BORDEN WITH SPRAY TYPE MIDWEST WITH SPRAY BORDEN TYPE W/ INTERNAL SPRAY MIDWEST TYPE W/ INTERNAL SPRAY 01 01 (TB) (TM) Compact handpiece, ergonomic design, body made of metal with surface treatment. Rotor supported on bearings providing a low noise level, speed adjustment in clockwise and counterclockwise directions, adapts to any contra angle or straight piece handpiece in the INTRA system. The Gnatus handpieces have been planned to offer maximum performance in the most severe clinical procedures. With a design which caters for ergonomics, torque, sensitivity and lightness. Produced with technology (MADE IN BRAZIL) and material of the utmost durability. Quality system ISO 9001:2000, guaranteeing that the products are manufactured within standardized procedures. The Gnatus handpieces are intended to remove caries, remove restorations and perform tooth sectioning , as an aid in extracting teeth. Products manufactured as per resolution RDC 59/00 – National Agency of Sanitary Surveillance - ANVISA. IMPORTANT: This equipment is for dental use use only. It must be operated and utilized by specialized professional (certified professional, according to the legislation of the country) and following the instructions of the manual. The operation of the equipment required, for the professional, the utilization of correct instruments and it should to be in perfect conditions of the use, and to protect the professional, the patients and others, in the eventual danger situation. • This equipment can’t be used in the presence of inflammable anesthetics or products that may cause explosion. • After finish life-cycle of the equipment, this equipment is throw way, so it should to be destroyed in appropriate area (according to the legislation of the country). • To guarantee the safe functioning of your equipment, use only the assemble configurations (Dental Chair, Dental and Water Units and Dental Light) supplied by Gnatus authorized Dealers / Technical Assistance. PRODUCT DESCRIPTION Congratulations. You have made a good choice when you decided to buy a GNATUS QUALITY product comparable to the best products available in the World. This manual is a general presentation of your product and it will give you important details to help you to solve possible problems. Please, read it and keep this with you. DEAR CUSTOMER This equipment is covered by the warranty against any manufacturing defect for the period and standards contained in the Warranty Certificate. The Warranty Certificate comes with the product and must be dilled in upon the delivery date of the product. Queries and information: GNATUS Help Desk : + 55 (16) 2102-5000. WARRANTY OF EQUIPMENT TECHNICAL SPECIFICATIONS • Work pressure: Maximum 275.79 kpa = 40 psi = 2.75 bar • Air consumption: 54 l/min. • Speed: 5,000 thru 20,000 rpm • Fitting terminal: Borden Type 02 holes Midwest Type 04 holes • Coupling: via the INTRA system Straight Piece Contra Angle • Sterilization: Autoclavable up to 135ºC Net Weight 32 MM: Type Borden: 77g Type Midwest: 80g Gross Weight 32 MM: Type Borden: 137g Type Midwest: 148g Net Weight SL 30: Type Borden: 80g Type Midwest: 83g Gross Weight SL 30: Type Borden: 140g Type Midwest: 143g Standards ISO 780:1997 - Packaging - pictorial marking for handling goods; NBR ISO 7785-1:1999- Dental handpieces NBR ISO 7785-2:2004- Dental handpieces ISO 1797.1.1992 - Dental rotary instrument - shanks NBR ISO 9687:2005 – Dental equipment – Graphic symbols; NBR ISO 13402:1997 – Dental and surgical instruments; NBR ISO 9168:2005 – Dental handpieces – Connection of hose; ISO 3964:1982 - Dental handpieces - Coupling dimensions EN 980:2008 (Ed. 2) - Graphical symbols for use in the labeling of medical devices; EN ISO 17664:2004 - Sterilization of medical devices; EN ISO 10993-1:2003 - Biogicol evaluction of medical devices; EN ISO 13485:2003 - Sistema de qualidade - Equipamento médico; Content of acessible and non-acessible demarcations Do not discard the packaging ATTENTION: - Don’t try to repair or assemble defective or inoperative parts, nor replace original parts by other’s manufacturer parts. Gnatus doesn’t supply other manufacturer’s electrical and mechanical parts. Only Gnatus or its authorized agents can repair this equipment using original spare parts. The operation safety and the technical specifications shall be valid only if original spare parts were used. - According to law, the manufacturer shall be held responsible for the equipment’s safety only if the maintenance, repairs and changes were carried out by himself or by and authorized agent, provided that the latter followed the directions specified by Gnatus. - The use of pneumatic or electric tools to perform any type of recording on handpieces, could lead to damage to the surface treatment of the product, causing the loss of warranty. - Because of the chemical properties of the material, there may be little variation in the color tone of handpieces as production batch. Transportation and storage - His equipment must be transported and stored observing the following directions: - Avoid falls and impacts; - Keep it dry, do not expose it to rain, water drops or wet floor; - Keep it away from water and direct sunlight, and in it original wrapping; - Upon transporting, respect the maximm stacking indicated on the outside of the packaging. - Transportation and storage temperature range: -12°C to 55°C; - Transportation and storage relative humidity range: 0% to 90%; - Atmospheric pressure range: 500hPa to 1060hPa (375 mmHg to 795 mmHg). Packing ambient conditions (between operations) - Ambient temperature range: +5°C to +45°C; - Packing relative humidity range: 30% to 75% (non condensing); Operation ambient conditions - Operation ambient temperature range: +10°C to +35°C; - Operation relative humidity range: 30% to 75% (non condensing); SYMBOLS Warning Indicates the ideal temperature for sterilization in autoclave. Recyclable. ABBREVIATIONS AR - High Speed MM - Micromotor PR - Straight Piece CA - Contra Angle TM - Midwest Type Packing symbols Packing to be transported and / or stored with care (should not suffer drop and neither receive impact). Temperature limit for the packing to be stored or transported. Packing to be transported and / or stored avoiding humidity, rains and wet floor. Packing to be transported and / or stored with the harrows up. It determines the maximum quantity of boxes which can be stacked during transportation and storage “as per packaging”. The packing must be stored and transported away from direct sun light exposure. TB - Borden Type PB - Push Button FG - Friction Grip S - Spray Warning – See the manual.

Upload: dangkiet

Post on 02-Nov-2018

238 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: English Gnatus Intra Micromotor 32 MM TECHNICAL ...gnatus.com.br/2005/images/online/manuais_494539_Gnatus Intra... · Gnatus Intra Micromotor 32 MM with TB Spray, without TB Spray,

English

Owner´sManual

Gnatus Intra Micromotor 32 MMwith TB Spray, TB Spray, TM Spraywithout with

0499

Gnatus Intra Micromotor SL 30TB Spray, TB Spray, TM Spray / TM SI / TB SIwith without with

01 - Gnatus Intra Micromotor02 - Coupling lining03 - O-ring.

GENERAL DATA

Dimensions (mm)

010101

02 0203

BORDEN TYPE W/ EXTERNAL SPRAY BORDEN TYPE W/O SPRAY MIDWEST TYPE W/ EXTERNAL SPRAY

TYPE BORDEN WITH SPRAY

TYPE MIDWEST WITH SPRAY

BORDEN TYPE W/ INTERNAL SPRAY

MIDWEST TYPE W/INTERNAL SPRAY

0101(TB) (TM)

Compact handpiece, ergonomic design, body made of metal with surface treatment. Rotor supported on bearings providing a low noise level, speed adjustment in clockwise and counterclockwise directions, adapts to any contra angle or straight piece handpiece in the INTRA system.

The Gnatus handpieces have been planned to offer maximum performance in the most severe clinical procedures. With a design which caters for ergonomics, torque, sensitivity and lightness. Produced with technology (MADE IN BRAZIL) and material of the utmost durability.

Quality system ISO 9001:2000, guaranteeing that the products are manufactured within standardized procedures. The Gnatus handpieces are intended to remove caries, remove restorations and perform tooth sectioning , as an aid in extracting teeth. Products manufactured as per resolution RDC 59/00 – National Agency of Sanitary Surveillance - ANVISA.

IMPORTANT: This equipment is for dental use use only. It must be operated and utilized by specialized professional (certified professional,

according to the legislation of the country) and following the instructions of the manual. The operation of the equipment required, for the professional, the utilization of correct instruments and it should to be in perfect conditions of the use, and to protect the professional, the patients and others, in the eventual danger situation.

• This equipment can’t be used in the presence of inflammable anesthetics or products that may cause explosion. • After finish life-cycle of the equipment, this equipment is throw way, so it should to be destroyed in appropriate area (according to

the legislation of the country).• To guarantee the safe functioning of your equipment, use only the assemble configurations (Dental Chair, Dental and Water Units

and Dental Light) supplied by Gnatus authorized Dealers / Technical Assistance.

PRODUCT DESCRIPTION

Congratulations. You have made a good choice when you decided to buy a GNATUS QUALITY product comparable to the best products available in the World.

This manual is a general presentation of your product and it will give you important details to help you to solve possible problems. Please, read it and keep this with you.

DEAR CUSTOMER

This equipment is covered by the warranty against any manufacturing defect for the period and standards contained in the Warranty Certificate. The Warranty Certificate comes with the product and must be dilled in upon the delivery date of the product.

Queries and information: GNATUS Help Desk : + 55 (16) 2102-5000.

WARRANTY OF EQUIPMENT

TECHNICAL SPECIFICATIONS

• Workpressure: Maximum 275.79 kpa = 40 psi = 2.75 bar

• Airconsumption: 54 l/min.

•Speed: 5,000 thru 20,000 rpm

• Fittingterminal: Borden Type 02 holes Midwest Type 04 holes

• Coupling: via the INTRA system Straight Piece Contra Angle

• Sterilization: Autoclavable up to 135ºC

• NetWeight32MM: Type Borden: 77g Type Midwest: 80g

• GrossWeight32MM: Type Borden: 137g Type Midwest: 148g

• NetWeightSL30: Type Borden: 80g Type Midwest: 83g

• GrossWeightSL30: Type Borden: 140g Type Midwest: 143g

Standards ISO 780:1997 - Packaging - pictorial marking for handling goods; NBR ISO 7785-1:1999- Dental handpieces NBR ISO 7785-2:2004- Dental handpieces ISO 1797.1.1992 - Dental rotary instrument - shanks NBR ISO 9687:2005 – Dental equipment – Graphic symbols; NBR ISO 13402:1997 – Dental and surgical instruments; NBR ISO 9168:2005 – Dental handpieces – Connection of hose; ISO 3964:1982 - Dental handpieces - Coupling dimensions EN 980:2008 (Ed. 2) - Graphical symbols for use in the labeling of medical devices;EN ISO 17664:2004 - Sterilization of medical devices;EN ISO 10993-1:2003 - Biogicol evaluction of medical devices;EN ISO 13485:2003 - Sistema de qualidade - Equipamento médico;

Content of acessible and non-acessible demarcations

Do not discard the packaging

ATTENTION:- Don’t try to repair or assemble defective or inoperative parts, nor replace original parts by other’s manufacturer parts. Gnatus doesn’t

supply other manufacturer’s electrical and mechanical parts. Only Gnatus or its authorized agents can repair this equipment using original spare parts. The operation safety and the technical specifications shall be valid only if original spare parts were used.

- According to law, the manufacturer shall be held responsible for the equipment’s safety only if the maintenance, repairs and changes were carried out by himself or by and authorized agent, provided that the latter followed the directions specified by Gnatus.

- The use of pneumatic or electric tools to perform any type of recording on handpieces, could lead to damage to the surface treatment of the product, causing the loss of warranty.

- Because of the chemical properties of the material, there may be little variation in the color tone of handpieces as production batch.

Transportation and storage

- His equipment must be transported and stored observing the following directions:- Avoid falls and impacts;- Keep it dry, do not expose it to rain, water drops or wet floor;- Keep it away from water and direct sunlight, and in it original wrapping;- Upon transporting, respect the maximm stacking indicated on the outside of the packaging. - Transportation and storage temperature range: -12°C to 55°C;- Transportation and storage relative humidity range: 0% to 90%;- Atmospheric pressure range: 500hPa to 1060hPa (375 mmHg to 795 mmHg).

Packing ambient conditions (between operations)- Ambient temperature range: +5°C to +45°C;- Packing relative humidity range: 30% to 75% (non condensing);

Operation ambient conditions- Operation ambient temperature range: +10°C to +35°C;- Operation relative humidity range: 30% to 75% (non condensing);

SYMBOLS

Warning

Indicates the ideal temperature for sterilization in autoclave.

Recyclable.

ABBREVIATIONS AR - High Speed MM - Micromotor PR - Straight Piece CA - Contra Angle TM - Midwest Type

Packing symbols

Packing to be transported and / or stored with care (should not suffer drop and neither receive impact).

Temperature limit for the packing to be stored or transported.

Packing to be transported and / or stored avoiding humidity, rains and wet floor.

Packing to be transported and / or stored with the harrows up.

It determines the maximum quantity of boxes which can be stacked during transportation and storage “as per packaging”.

The packing must be stored and transported away from direct sun light exposure.

TB - Borden Type PB - Push Button FG - Friction GripS - Spray

Warning – See the manual.

Page 2: English Gnatus Intra Micromotor 32 MM TECHNICAL ...gnatus.com.br/2005/images/online/manuais_494539_Gnatus Intra... · Gnatus Intra Micromotor 32 MM with TB Spray, without TB Spray,

NUM. REG.:10229030039

d.:

30

00

51

96

7 R

ev.

05

Manufacturer/ Distribuitor:

Technical uties:DGilberto Henrique Canesin Nomelini – CREA-SP: 0600891412

EQUIPAMENTOS MÉDICO-ODONTOLÓGICOS LTDA.

Rod. Abrão Assed , Km 53+450m - Cx. Postal 782CEP 14097-500 - Ribeirão Preto - S.P. - BrasilFone (16) 2102-5000 - Fax (16) 2102-5001

C.N.P.J. 48.015.119/0001-64 - Insc. Est. 582.329.957.115www.gnatus.com.br - [email protected]

This product complies with the requirements of directive 93/42/EEC (with marking EC)

Maintenance of the BORDEN Type Lining:

fig.A fig.B fig.C

- Insert a screwdriver of 1.2mm in the face of the lining as per (fig. A), using it as a lever until removing it from the hole. - Remove the damaged lining (fig. B).- Place the new lining in the hole leaving the holes aligned, an with a screwdriver push the lining carefully downwards (fig. C).

MAINTENANCE

Speed adjustment In order to adjust the speed and direction of speed, turn the control ring as per the fig. below.

clockwisedirectionandspeed

CONTROLRING

TERMINALNUT

MIDWESTtype

Maintenance of O-Ring:

01

fig.A fig.B fig.C

- Use the same procedure as the BORDEN type.

Maintenance of the MIDWEST Type Lining:

TERMINALNUTBORDENtype

01

02

OPERATION

Terminal Coupling- Ensure that the connection of the instrument is compatible with that of the equipment (Borden Type or Midwest Type as per the

figures). Connect and thread the nut of the terminal turning it in a clockwise direction until it is fastened.

Functioning- For the handpiece to function well it is necessary to have a driving pressure of 275.79 kpa = 40 psi = 2.75 bar. - Before using it for the first time, follow the lubrication procedures.

NOTE:ITISINDISPENSABLETHATTHEDRIVINGAIRISDRY,CLEANANDUNCONTAMINATED.DIRTYANDMOISTAIRLEADTOPREMATUREWEAROFTHEBEARINGS.

MECHANICALSTABILITY:OWINGTOITSDELICATECOMPONENTS,THISHANDPIECEMUSTNOTBEDROPPEDORUNDERGOIMPACTS.

ATTENTION:

CHANGE:- Insert a screwdriver of 1.2mm in the channel of the O-ring. 1, moving it outwards (fig. A), first removing the ring from channel 1 and then from channel 2, (fig. B). - In order to fasten the new ring, execute the opposite process without the screwdriver (fig. C).

fig.A fig.B fig.C

Cleaning and Disinfection:These processes should be used in the handpieces.- Uncouple the piece from the INTRA fit terminal.- Remove the diamond tip or bur.- Clean the external part using alcohol at 70% or another disinfectant which does not have an acid base, ammonia or chlorine.

LUBRIFICATION

TYPE BORDEN

Lubrication in spray- The lubrication should be executed by the hole using the adapter as

illustrated in the figure- Press the oil spray maintaining the bottle in the vertical position and

apply for 1 or 2 seconds.- Cover the piece with absorbent paper to remove the excess of oil.- Handpieces not submitted to the sterilization process must be lubricated

at least twice a day.

Note:- For the perfect functioning of the instrument, only use the lubricant

specified by Gnatus at the end of each work period.

TYPE MIDWEST

NEVERSUBMERGETHEINSTRUMENTINADISINFECTIONBATH.

NEVERUSEMICRO-ABRASIVEMATERIALORSTEELWOOLUPONCLEANING.

NEVERUSEPRODUCTPERACETICACIDFORCLEANING,SINCEITISCORROSIVE.

LubricantFor executing the lubrication of your equipment, we recommend using the product GTSGnatus.

ATTENTION:• The lubricant does not come with the product.• Not using the lubricant will lead to the loss of the warranty of the product.

Sterilization:

1 - Cleaning- Use the previous process.

2 - Lubrication- Use the previous process:

3 - Sterilization- Follow the instructions of the manufacturer of the appliance.- Only sterilize clean and lubricated handpieces.- Clean, grease and remove the overflow oil in the apparatus according to the previous indicate, put it in the surgical grade paper (not provided by Gnatus) and sterilize it in autoclave until 135°C (275°F) through a cycle.- The bit breaker can be placed into an autoclaved up to 135°C (275°F) during one cycle.

BEFORESTERILIZATION,CONNECTITTOTHEDELIVERYUNITANDACTIVATEITFORAFEWSECONDSTOELIMINATETHEEXCESSOFOIL.

NOTE:IFTHEEXCESSOFOILISNOTELIMINATED,ITMAYCAUSEBURNTRESIDUEINTHEBEARING.

STERILIZEBEFORE1STUSE,PRODUCTDONOTPROVIDEDSTERILIZED.

EXECUTETHISPROCESS,WHENEVERSTERILIZING.

NOTVALIDORRECOMMENDTOTHESTERILIZATIONBYETHILENEOXYDE.

DONOTSTERILIZEINANOVEN

- Turn the control ring in a clockwise or counterclockwise direction.

- Lubricate as per the indications.

- Tighten the connection.

- Send to the authorized service.

- The micromotor does not function.

- Insufficient speed.

- The control ring is in the intermediary position between the counterclockwise and clockwise directions of speed.

- Lack of lubrication or inadequatelubrication.

- The connection of the instrument / terminal is loose.

- Bearing damaged.

- Lack of air or low air pressure.

- Lack of lubrication or Inadequate lubrication.

- Check the air pressure in the network.

- Lubricate as per the indications.

- Failure and noise in the functioning.- Air leakage.

- Lack of lubrication.

- Examine the hoses and connections.

- Lubricate as indicated.

TROUBLESHOOTING

The most important aspect related to equipment care is that concerning spare parts.To guarantee the life span of your equipment, use only original Gnatus spare parts. They are sure to follow the technical specifications

and standards required by Gnatus.We must also point out to you our chain of authorized dealers. Only dealers that make part of this chain will be able to keep your

equipment constantly new for they count on technical assistants who have been trained and on spedific tools for the correct maintenance of your equipment.

Doubts and information: GNATUS call center + 55 (16) 2102-5000.

FINAL CONSIDERATIONS

ATTENTION:Any other problem that could happen with its equipment, should only be repaired by an authorized technician Gnatus.The bad use, negligence or maintenance not made by technician authorized by the maker, will imply in the loss of the warranty.

Troubleshooting: Probable Causes Solution: