Electric Micromotor System

Instruction for Use

Maxso® E600P

Page 2: Electric Micromotor System - Beyes Dental Handpiece and

2

Operation Manual

•EquipmentClassification- Class II equipment: - Type B shock protection: - Recommended sterilization: See 7. Sterilization-Donotuseinthepresenceofflammableliquidsorgases.

Cautions for safe handling and operation

Readthesesafetyprecautionsfirstbeforeusingthisproduct.These instructions will show you how to operate this product safely and prevent injury to yourself orothers.Theyareclassifiedbythedegreeand/orseverityofdanger.Allcautionsregardingsafetyshouldbeobserved.

Classification Degree and severity of danger

Warning Provides an instruction where personal injury may occur.

Caution Provides an instruction where minor to medium injury may occur.

Notice Providesaninstructionthatshouldbeobservedforsafetyreasons.

Warning

-ToPreventElectricShock:Donotpluginthepowercordwithwethands.-ToPreventElectricShock:Preventwaterfromgettingontothecontrolbox.-ToPreventElectricShock:Pluginandremovethepowerbyholdingontothebaseoftheplugand

not the cord.-ToPreventElectricShock:Useanelectricaloutletwhichisgrounded.-Ifyoufeelanyabnormalitysuchasexcessivevibration,heat,ornoise,stopusingtheunit immediately.

Caution

-Whenoperatingthissystem,alwaysconsiderthesafetyofthepatient.-Theproductisdesignedforclinicaldentalusebyqualifiedpersonnel-Everyconnectionmustbefirm.Ifitisnotfirmlyconnected,itmaycauseanairorwaterleak,ortheLEDmightnotturnon.-Donotdroptheproductorcauseittobeimpactedinanyway.Thiscouldcauseamalfunctionanddamage.-ThissystemistobeusedwithISO9168,typeBortypeChoseconnector.-Thissystemismeanttofunctionintemperaturesof0-40°C(32-104°F),humidityat10-85%RH,atmosphericpressure700-1060hPaandnomoistureorcondensationintheControlUnit.Operatingoutside of these parameters can cause a malfunction.-WheninstallingtheControlBoxorMotor,donotbendortwistthetubingorthecord.Noexcessive

force is need to do the installation.-Donotusecontaminatedair(bydustormoisture).Iftheaircontainswaterordust,itmightcauseamalfunctionoroverheating.

Page 3: Electric Micromotor System - Beyes Dental Handpiece and

3

Caution

Unit-Graspcordbytheplugtoremovefromtheoutlet.Donotpulloryankthecorditself.Itmaycause

the wire to malfunction. -CareshouldbetakennottoplacetheACcordnearagasburner.Neverattempttorepairaburnedmotorcord.Alwaysreplaceitwithanewcord.-Priortouse,alwayscheckforvibration,noiseandoverheating.Ifanyabnormalitiesaredetected,stopusingimmediatelyandcontactyourdealer.-Becarefulnottospillwaterontotheunitasitmayresultinafireoranelectricshockduetoashort

circuit.

Motor, Hand piece -Donotusethisproductunderstrongstressforextendedperiodsoftime.Itcancauseoverheating.- Do not connect or disconnect the cord until the drive motor has completely stopped.-Donotconnect/disconnectthehandpieceduringoperation.-VerifythattheSpeedControlSwitchisadjustedwithintheallowablespeedbeforeuse.-Connectonly4-hole,5-holeor6-pintubing.-AirRequirements:Dry,freefromcontaminationandoil.Useacompressorwithadryairsystem/Installanairfilterifnecessary/Blowoutthelinesbeforeinstallation.-Donotautoclave(oranyotherhightemperaturesterilization)ControlBox,ACAdaptor,MotorCode.-Theusershallberesponsibleforoperationandmaintenanceandinspection.

NoticeDuringrotation,themotorandmotorcordmayaffectcomputerandLANcables.Noisecouldbeheardduringoperationnearradioreceivers.Aftertreatment,immediatelyturnthepowerandairsupplyoff.Removethepowercordiftheunitisnottobeusedforalongtime.

1. Specifications

Model Maxso E600P

Speed 2,000~40,000rpm

Torque 3.4Ncm

Rated Input AC24V/DC36

NoiseLevel <60dB

Dimensions W127xD149xH54(mm)

Page 4: Electric Micromotor System - Beyes Dental Handpiece and

4

2. Component Names

1. ControlBox 2. DigitalSpeedDisplay 3. Gear Ratio Lamp 4. Gear Ration Select Key 5. FWD/REVSelectKey 6. Program1Button 7. Program2Button 8. SpeedAdjusters

19

102

3

4

6

5

7

8

12

11 13 14

16

15

9. Motor10. MotorCord11. PowerSwitch12. ACAdapterConnector13. MotorCordConnector14. TurbineHoseConnector15. ACCord16. ACAdaptor

Page 5: Electric Micromotor System - Beyes Dental Handpiece and

5

3. Control Box Operation

Caution

Makesurethereisnoresidualairorwaterintheturbinehosewhenconnectingittothecontrolbox.Handtightenthenutproperlywithoutexcessiveforcewhenconnectingtheturbinehoseandthemotor cord connector.

3-1 Connect the hand piece tubing from your delivery unit. Fittheturbinehoseplugfromthedeliveryunittotheturbinehoseconnectorandtightencompletely.

Caution

- Do not use any other hose than the ones provided with the unit- Do not use contaminated air-Blowoutthelinesbeforeinstallation

Caution

NeveruseanyACadapterotherthantheoneprovidedwiththisproduct.

3-4 Connecting the Power CordInsertthepowercordintotheinletoftheACadapter.

3-2 Connecting the Motor.Connectthemotorbyaligningthemotorcordplugandinsertfirmly.

3-3 Connecting the AC adapterInserttheACadapterplugintotheACadapterconnectorbeforepluggingitintothewalloutlet.

Page 6: Electric Micromotor System - Beyes Dental Handpiece and

6

4. Handling the Motor

4-1 Connecting/Disconnecting the motor and motor cord.Alignandinsertthemotorpinfirmlyintothepinholesofthemotor cord connector and fasten the motor cord nut securely.

Toremovethemotorcordfromthemotor,unscrewanddetachthemotorcordnut,andgentlypulloutthemotorcordconnector.

4-2 Connecting/ Disconnecting the motor and hand pieceToinserttheE-Typehandpieceintothemotor,alignthehandpieceandmotor,thenturnuntilyouhearaclick(positioningpinsarealigned)

Toremovethehandpiece,simplypullthehandpiecefromthemotor.

Caution

-Careshouldbetakenwhenusinganyhandpieceattachmentwhichrequireslubrication(E-Typemicromotor).-Followinglubrication,standthehandpieceattachmentverticallyuntiltheexcesssolutiondrains.

Wipe the hand piece then attach it to the motor.- Do not connect or disconnect the handpiece until the motor has completely stopped.

5. Operation

5-1InsertthepowercordplugintotheACoutletandturnthepowerswitchon.Whenthepowerswitchison,thelampsandindicator will come on.

5-2Beforeusingthissystem,makesuretooperatetheFootAirCalibrationfunction.

5-3PushtheGearRatioSelectKeytoindicatethespeedoftheattachment you will use.

*CheckTableBelowforSpeeds

HandPieceGearRatio Indicator Speed (min-1(rpm))

16:1 1-25 100–2,500

1:1 2–40 2,000–40,000

1:5 1–20 10,000–200,000

MotorMotor CordConnector

Motor CordNut

Motor Cord

Handpiece(Option)Positioning

Gear Ratio Lamp

Motor Insert

Push

Page 7: Electric Micromotor System - Beyes Dental Handpiece and

7

5-4SetthemaximumspeedusingtheRotationAdjustmentKey(+/-).

5-5SelectFWD/REVrotationusingtheFWD/REVSelectkey.

5-6Themicromotoroperationiscontrolledbytheairswitch/foot pedal of the delivery unit.

5-7ProgramFunction.Youcanmemorizethesetting:1)setthevaluesforgearratio,speed2)PressP1orP2keyformorethan3secondsuntilitbeeping,thesettingiscompleted.

Push

DigitalSpeedDisplay

Increase

Decrease

FWD/REVLamp

6. Sterilization

Sterilize the motor only.Forsterilization,werecommendtheautoclavesterilizationmethod.Sterilizationisrequiredthefirsttimeyouusetheunitaswellasaftereachpatientasnotedbelow.

Autoclaving1)Turnoffthepower.2)Detachthemotorfromthemotorcord.3)Cleanthesurfaceofthemotorwithabrush(Donotuseametalbrush),andwipeitwithcottonanddisinfectantalcohol.4)ScrewtheMotorCaptotheMotor.PuttheAutoclavePlugonto the motor insert.5)Insertintoautoclaveouch.Sealthepouch.6)Autoclavableuptoamaximumof135°C.7)KeeptheMotorintheautoclavablepouchuntilyouareready to use it to keep it clean.

Caution

-Donotautoclave(oranyotherhightemperaturesterilization)ControlBox,ACAdaptor,orMotorCord.-Donotlubricatethemotor.- Do not wipe or immerse the system in acidic water or acidic solutions.-Donotsterilizewithdirtonthesurface,itmaycauserust.

AutoclavePlug

Motor Cap

Tighten

Page 8: Electric Micromotor System - Beyes Dental Handpiece and

8

8. Error Codes

Error Code Issue Cause Solution

E0 Overcurrenterror(InCircuit)

Detected an overcurrent in the circuit

Contact your dealer

E1 Overcurrenterror.(InSoftware)

Detected an overcurrent in the software

Contact your dealer

E2 Retry Error Motor was overloaded and stopped and can not restart.

Contact your dealer

E3 ITRIP Error Faultymotororinternalcircuit Contact your Dealer

E5 OvervoltageInputError

Overvoltagetothecontrolbox Contact your dealer

E6 LEDvoltageerror Overvoltageintheinternalcir-cuit.(morethan3.7V)

Contact your dealer

E8 Overheatingerror Hightemperaturesafetysystemactivated due to extended use .

Wait for a while and letitcooldownbeforeusingagain.

7. Maintenance

IftheO-Ringsbecomeworn,themotormightnotconnectwelltothehandpiece,orwatermayleak.ReplacetheO-Rings.

Changing the O-RingRemovetheO-Ringatthemotorinsertwithapointedtool,andmountthenewO-ringsintothegroove.

Caution

Thereare4O-RingsontheMotorInsert.Theblueoneisthinnerthantheother3.Makesuretoinstall them in the proper spot.

Caution

IftheO-Ringsbecomeworn,itmaycause:-Air/Waterleak-Difficultyattaching/disconnectingthehandpiece-ExcessVibration

Page 9: Electric Micromotor System - Beyes Dental Handpiece and

9

9. Troubleshooting

Whenhavinganissuewiththeunit,checkthisguidebeforecontactingyourdealer.Thesearethemost common issues which may occur.

Issue Cause Solution

Pilot Lamp does not come on Power switch is off-ACAdaptorisnotconnectedproperly-Internalfuseisblown

Turn on the switch-Check the connection-Contact your dealer

Motor doesn’t not run Tubing,MotorCord,ACadapt-er is not connected properly-Airpressureisnotsufficient

Check the connection-Check the air pressure from the delivery unit-refertoErrorcode9

The rotation speed is too slow -Airpressurefromdeliveryunitis too low-Airpressureonthedeliveryunit is lower than the “Lower limit of air pressure”.

Check the air pressure on the delivery system.-Do“Foot/Aircalibration

Beepswhenturningontheswitch

Youaresteppingonthefootpedalwhenturningonthepower switch

Do not step on the foot pedal whileturningontheswitch

E9 Motor start error Motor didn’t reach the preset speedintheprescribedamountof time.Break down of the motor cord wire,orinternalcircuitmalfunc-tion

Contact your dealer

EA LEDvoltageerror VoltagefortheLEDisn’thighenough

Contact your dealer

Eb Airoffseterror Lower than preset ‘Lower limit of air pressure”

Displayedbyfoot/aircalibrationfunction.NotaMalfunction

Ec Airfullscaleerror Higherthanpreset“Upperlimitofair pressure”Airpressurehigherthan0.4MPa

Displayedbyfoot/aircalibrationfunction.Notamalfunction

EE EEPROMError Malfunctionofthesettingsmemory

Contact your dealer

Page 10: Electric Micromotor System - Beyes Dental Handpiece and

10

10.1 Scope of Warranty

BEYESDentalCanadaInc.(‘BEYES’)warrantstotheoriginalretailpurchaserthatitwillbeatBEYESoptiontorepairorreplacecomponentsofthedentalproductsmanufacturedbyBEYES(exceptforcomponentsnotwarrantedunder‘Exclusions’)thataredefectiveinmaterialorworkmanshipundernormaluseandservice.BEYES’obligationunderthislimitedwarrantyislimitedtotherepairorreplacementoftheapplicablecomponents.ThislimitedwarrantyshallonlyapplytodefectsthatarereportedtoBEYESwithintheapplicablewarrantyperiodandwhich,uponexaminationbyBeyes,provetobedefective.Thiswarrantyextendsonlytothefirstretailpurchaserofaproductandisnottransferableorassignable.Replacementcomponentsorproductsmaybeusedand/orrefurbishedcomponentsorproducts,providedtheyareoflikequalityandspecificationsasnewcomponentsorproducts.

10.2 Applicable Warranty PeriodTheapplicablewarrantyperiod,measuredfromthedateofinvoicetotheoriginaluser,shallbewarrantedforaperiodof25months.

10.3 ExclusionsThislimitedwarrantydoesnotcoverandBEYESshallnotbeliableforthefollowing;1)Defects,damageorotherconditionscaused,inwholeorinpart,bymisuse,abuse,negligence,alteration,accident,freightdamage,negligentstorage,tamperingorfailuretoseekandobtainrepairorreplacementinatimelymanner;2)Productswhicharenotinstalled,used,andproperlycleanedandmaintainedasrequiredorrecommendedintheBEYES‘Installation’and/or‘Installation/User’sManual’fortheapplicableproduct,includingthespecifiedstructuralandoperationalenvironmentconditionsandelectricalpowerrequirements;3)Productsconsideredtobeofaconsumableorsterilenature;4)AccessoriesorpartsnotmanufacturedbyBEYES;5)Chargesbyanyoneforadjustments,repairs,replacementparts,installationorotherworkperformeduponorinconnectionwithsuchproductswhicharenotexpresslyauthorizedinwritinginadvancebyBEYES;

LEDdoesnotlightup Reached the end of life expec-tancy

Contact your dealer

Motorheatsupabnormallyduringrotation

Coolantairisnotcomingoutofthe delivery unit

Check the air pressure of the delivery system

WaterLeakage Turbine,Motorcordisnotconnected correctly

Check the connection

Whenturnedon,thesettingvalues are different from what they were set to.

Turned off the motor while it wasrotating

Do not turn off the unit until afterthemotorstopsspinning

10. Beyes Limited Warranty Statement

Page 11: Electric Micromotor System - Beyes Dental Handpiece and

11

6)Costsandexpensesofroutinemaintenanceandcleaning;7)RepresentationsandwarrantiesmadebyanypersonorentityotherthanBEYES;8)Matchingofcolor,grainortextureexcepttocommerciallyacceptablestandards;9)Changesincolorcausedbynaturalorartificiallight;10)Custommanufacturedproducts;11)AlterationsormodificationstotheproductbyanypersonorentityotherthanBEYES;12)ProductsthatwouldotherwisebecoveredunderSections1and2ofthislimitedwarranty,butareacquired:(i)fromapersonorentitythatisnotBEYESoroneofitsauthorizeddealers;or(ii)fromaBEYESdealerthatisnotauthorizedtoselltheproductatissueinthegeographicterritorywherethepurchaserislocated,orisnotauthorizedtoselltheproductatissuewithinthemedical,animalhealthordentalmarket,asthecasemaybe,inwhichpurchaserintendstousetheproduct.

10.4 Exclusive Remedy; Consequential Damages DisclaimerBEYES’OBLIGATIONUNDERTHISLIMITEDWARRANTYISTHEREPAIRORREPLACEMENTOFDEFECTIVEPARTS.BEYESSHALLNOTBELIABLEFORANDHEREBYDISCLAIMSANYDIRECT,SPECIAL,INDIRECT,INCIDENTAL,EXEMPLARYORCONSEQUENTIALDAMAGESORDELAYS,INCLUDING,BUTNOTLIMITEDTO,DAMAGESFORLOSSOFPROFITSORINCOME,LOSSOFUSE,DOWNTIME,COVERANDEMPLOYEEORINDEPENDENTCONTRACTORWAGES,PAYMENTSANDBENEFITS.

10.5 Warranty DisclaimerTHISLIMITEDWARRANTYISBEYES'ONLYWARRANTYANDISINLIEUOFALLOTHERWARRANTIES,EXPRESSORIMPLIED.BEYESMAKESNOIMPLIEDWARRANTIESOFANYKINDINCLUDINGANYIMPLIEDWARRANTIESOFMERCHANTABILITYORFITNESSFORAPARTICULARPURPOSE.THISWARRANTYISLIMITEDTOTHEREPAIRORREPLACEMENTOFDEFECTIVEPARTS.

Page 12: Electric Micromotor System - Beyes Dental Handpiece and

Beyes Dental Canada Inc.23-595MiddlefieldRoadToronto,Ontario,P1V3S2Canada Tel:1-855-603-1888Fax: 1-855-720-1228Email: [email protected]:www.beyes.ca

Lotus NL B.V.KoninginJulianaplein10,IeVerd,2595AA,TheHague,NetherlandsTel: +31645171879(English) +31626669008(Dutch)Email: [email protected]:www.beyes.ca 0197

Printed in CanadaFile#:ENI019Rev.3/06.01.21

Federallawrestrictsthisdevicetosalebyorontheorderofadentist,physician,oranyotherpractitionerlicensedbythe law of the states in which he or she practices to use or order the use of this device. Beyes Dental Canada Inc. is not responsibleforanytypographicalerrors.Subjectstotechnicalchangeswithoutnotice.*Allbrandsareholdersoftheirrespective trademarks.