electrokinetic supercharging-electrospray ionisation-mass spectrometry for separation and on-line...
TRANSCRIPT
Research Article
Electrokinetic supercharging-electrosprayionisation-mass spectrometry forseparation and on-line preconcentrationof hypolipidaemic drugs in water samples
Electrokinetic supercharging, a powerful on-line preconcentration technique in CE, was
for the first time hyphenated with ESI-MS for the on-line concentration and separation of
five hypolipidaemic drugs. The electrophoretic separation was performed in a co-EOF
mode using the EOF reversal agent, hexadimethrine bromide, in ammonium bicarbo-
nate electrolyte, pH 9.00. The ionic strength and the amount of methanol in the buffer
were optimised in a multivariate manner using artificial neural networks, with the
optimal conditions being 60 mM ammonium bicarbonate containing 60% methanol,
providing baseline resolution of the five hypolipidaemics within 20 min. Using electro-
kinetic supercharging, the sensitivity of the method was improved 1000-fold over a
conventional injection under field-amplified sample stacking conditions with LODs of
180 ng/L. This is the first report of the separation of hypolipidaemics by CE. The
developed method was validated and then applied to the determination of the target
drugs in water samples from Hobart city.
Keywords:
CE / ESI-MS / Electrokinetic supercharging / Hypolipidaemic drugs / Watersamples DOI 10.1002/elps.200900420
1 Introduction
Hypolipidaemic drugs have been prescribed widely for
several years for the treatment of different forms of
hyperlipidaemia. Among the different classes of hypolipi-
daemic drugs, statins and fibrates are most commonly used
as they possess high effectiveness in reducing cholesterol
and triglyceride levels in human plasma, thereby reducing
mortality associated with coronary heart disease [1]. The
daily dose of these drugs ranges between 10 and 1200 mg,
depending on the severity of the coronary heart disease and
the type of drug prescribed. Because of their acidic nature as
well as their physicochemical properties, hypolipidaemic
drugs are not fully eliminated in municipal sewage
treatment plants (STPs) and as such, small amounts are
discharged as contaminants into the receiving waters [2].
They have been detected in the aquatic environment at
concentrations ranging from ng/L to mg/L in STP effluent
and surface waters [3, 4]. Regular and continuous input of
these drugs into the environment has led to various adverse
effects on the aquatic and terrestrial organisms, for example
exposure to gemfibrozil adversely affects the immune
function of the bivalve mollusc, Mytilus galloprovincialis [5],
and it also reduces testosterone in the goldfish Carassiusauratus [6]. Atorvastatin elicits phytotoxic effects on the
aquatic macrophytes Lemna gibbaand Myriophyllum sibiri-cum [7]. The development of sensitive, rapid, and simple
methods for the analysis of such drugs in water samples is
therefore of great importance from the chemical, environ-
mental and toxicological points of view.
The hyphenation of CE with MS is nowadays considered
as a competitive approach to classical MS-hyphenated
separation techniques such as LC-MS and GC-MS [8]. The
on-line combination of CE and MS is a powerful analytical
technique that separates analytes according to the differ-
ences in their electrophoretic mobilities, and provides
information on molecular masses and/or fragmentation of
the analysed substances [9]. Despite the fact that CE-MS has
been applied routinely in drug analysis [10], proteomic
analysis [11], metabolomics [12], organic food contaminants
[13], and analysis of carbohydrates [14], only a few CE-MS
Mohamed DawodMichael C. BreadmoreRosanne M. GuijtPaul R. Haddad
Australian Centre for Research onSeparation Science (ACROSS),School of Chemistry, Universityof Tasmania, Hobart, Tasmania,Australia
Received July 12, 2009Revised August 4, 2009Accepted August 9, 2009
Abbreviations: ANN, artificial neutral network; CF-EKS, EKSand counter-flow; DCM, dichloromethane; EKS,
electrokinetic supercharging; FASS, field-amplified samplestacking; HDMB, hexadimethrine bromide; LE, leadingelectrolyte; LLE, liquid–liquid extraction; STP, sewagetreatment plant; TE, terminating electrolyte
Correspondence: Dr. Michael C. Breadmore, Australian Centrefor Research on Separation Science (ACROSS), School ofChemistry, University of Tasmania, Hobart, Tasmania, AustraliaE-mail: [email protected]: 161-3-6226-2858
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
Electrophoresis 2010, 31, 1184–11931184
methods have been reported for the analysis of pharma-
ceuticals in environmental samples [15]. Ahrer et al. [16]
applied CE-ESI-MS for the determination of drug residues
in surface water samples and obtained LODs between 27
and 93 mg/L, which were decreased to 4.8 and 19 ng/L after
combination of SPE and liquid-liquid extraction (LLE).
Ahrer and Buchberger [17] also investigated the use of
aqueous and non-aqueous CE-ESI-MS for the analysis of the
same drugs in river water samples and LODs of 5–19 ng/L
were obtained. Macia et al. [18] used both aqueous and non-
aqueous BGEs in CE-ESI-MS for the analysis of acidic drugs
in surface and STP effluent water samples. The capillary
was coated with hexadimethrine bromide (HDMB) to
permanently reverse the EOF. LODs down to 100 ng/L were
achieved after SPE when the method was applied to surface
water samples, and when the method was applied for STP
water, LLE had to be included before the SPE to
preconcentrate and clean the water samples even further.
CE-ESI-TOF-MS has been reported for the determination of
antidepressant drugs in surface waters and STP effluents.
Again, SPE was used to obtain an enrichment factor of up to
1000 to obtain the necessary detection limits [19]. Sulfona-
mides in ground water samples were determined by CE-MS
and CE-MS/MS, LODs down to 5 mg/L were obtained [20].
The major drawback of CE is its high concentration
LOD, arising predominantly because of the very small inner
diameter (typically 50–100 mm) which limits the volume of
sample that can be injected and hence the absolute amount
of analyte that can be detected. The severity of this drawback
is manifested strongly when dealing with difficult samples
as in the case of analysis of pharmaceuticals in environ-
mental samples. Such samples bring further obstacles to
any CE method for two main reasons. First, pharmaceutical
compounds are typically present at very low concentrations.
In order to detect these low concentrations, injection of a
large volume of sample into the capillary is required. This is
not feasible in CE, because the amount of sample that could
be introduced into the capillary is rather limited (typically
o1 mL). Second, the matrix is not easily handled and usually
requires the use of off-line chromatographic sample clean-
up, such as SPE and/or LLE, as evidenced in all of the
currently published CE-MS methods discussed above, which
brings an additional effort to the method. These two
limitations account for the obvious lack of CE-MS methods
for analysis of pharmaceuticals in environmental samples.
In order to overcome the first limitation, the amount of
analytes injected into the capillary needs to be increased.
However, this often results in inferior separation due to the
reduction in efficiency, unless the analytes are maintained
as sharp bands as the injected volume is increased, a process
called stacking [21–23]. The most elegant solution to over-
come the second limitation is to use a very selective stacking
approach, such that only the analytes of interest are
concentrated while other components of the sample matrix
are not; this phenomenon is a characteristic outcome for
certain on-line preconcentration approaches, such as ITP
[21]. The recently developed method of electrokinetic
supercharging (EKS) is capable of achieving the above goals
and therefore presents an interesting possibility for the
enrichment of trace levels of pharmaceuticals in environ-
mental samples. EKS is considered as one of the most
powerful on-line preconcentration approaches, as it
combines the highly selective nature of ITP with the high
enrichment levels attainable by field-amplified sample
stacking. The general steps for any EKS system are as
follows: (i) filling the capillary with the BGE; (ii) introduc-
tion of a plug of leading electrolyte (LE) by hydrodynamic
injection; (iii) electrokinetic injection of the sample solution;
(iv) hydrodynamic injection of a plug of terminating elec-
trolyte (TE); (v) application of separation voltage, and start-
ing separation by ITP-CZE.
In the current work, we report the first hyphenation of
EKS with ESI-MS as a powerful combination for the
separation, on-line preconcentration and detection of hypo-
lipidaemic drugs. The electrophoretic separation was
performed in a co-EOF mode using an EOF reversal agent,
and the separation was optimised by variation of the ionic
strength of the separation buffer as well as the percentage of
the organic modifier added to the buffer. After optimisation
of separation parameters, the developed EKS-ESI-MS
method was validated, and then applied for the separation
and online preconcentration of five hypolipidaemic drugs in
drinking water and wastewater samples from Hobart city.
The presented work shows the first method for electro-
phoretic separation and on-line preconcentration of a
mixture of hypolipidaemic drugs.
2 Materials and methods
2.1 Chemicals, reagents and standards
The compounds studied, atorvastatin, fluvastatin, gemfibro-
zil, pravastatin and rosuvastatin (all 498%) were purchased
from Sequoia (Oxford, UK). 3-(Cyclohexylamino)1-propane-
sulphonic acid (CAPS) (99%), dichloromethane (DCM)
(HPLC-Grade), HDMB, and fuming hydrochloric acid
(37%) were from Sigma–Aldrich (St. Louis, MO, USA).
Ammonium hydrogen carbonate (98%) was from Ajax
Chemicals (Sydney, Australia). Sodium hydroxide (98%)
was from BDH (Kilsyth, Australia). Ammonium hydroxide
solution (28%) was from Fluka (Buchs, Switzerland).
Methanol (HPLC-Grade) was from Ajax Finechem (Seven
Hills, Australia). Water was treated with a Millipore (North
Ryde, Australia) Milli-Q water purification system. A stock
standard solution of 1 mg/mL of each drug was prepared in
methanol. A mixed standard solution of the five hypolipi-
daemic drugs was prepared at a concentration of 0.1 mg/mL
in methanol. The working standard solutions were prepared
daily by diluting the stock standard solution with Milli-Q
water. All solutions were stored in dark containers at 41C.
The working BGE solution had a concentration of 60 mM of
ammonium hydrogen carbonate (pH 9.00) containing 60%
(v/v) methanol unless otherwise stated. The BGE solutions
Electrophoresis 2010, 31, 1184–1193 CE and CEC 1185
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
were prepared freshly daily, sonicated for 5 min and filtered
through a 0.45 mm membrane filter.
2.2 Instrumentation
Electrophoretic separations were performed using an
Agilent 3D capillary electrophoresis system and Agilent
ChemStation software (Waldbronn, Germany). Electrophor-
esis and preconcentration experiments were carried out
using fused silica capillaries 50 mm id and 375 mm od
obtained from Polymicro Technologies (Phoenix, AZ, USA)
of 88 cm total length (88 cm effective length). The capillary
temperature was set at 251C. New capillaries were flushed
with 1 M NaOH for 120 min, with Milli-Q water for 10 min,
with HDMB (1% w/v) for 120 min, and with the BGE for
10 min.
ESI-MS detection was performed with an Agilent 6320
Ion Trap LC/MS, 6300 Ion Trap Control software, and
Agilent Pump 1200 Series.
2.3 Field-amplified sample stacking (FASS)
The hypolipidaemic drugs dissolved in Milli-Q water were
injected by applying hydrodynamic pressure at 50 mbar for
40 s; this sample plug represents 2.26% of capillary volume.
As shown in Fig. 1, all the drugs are weak acids. They have
pKa values ranging between 4.25 and 4.75 [24], and under
the experimental conditions used, they are negatively
charged and migrate towards the anode by means of a
combination of their own electrophoretic migration in
addition to the reversed EOF. Table 1 shows the pKa values
and the electrophoretic mobilities of the studied analytes, as
well as those of the major components of the BGE.
2.4 EKS
The capillary was filled with the BGE, then a short plug of
Milli-Q water was introduced by hydrodynamic injection at
40 mbar for 5 s, sample was then injected electrokinetically
by a negative voltage (�10 kV) for 170 s, and finally a small
volume of TE (1 mM CAPS) was injected hydrodynamically
at 50 mbar for 10 s. A separation voltage of �25 kV was
applied to re-stack the injected analytes and for their
separation. A positive pressure of 5 mbar was applied
during the separation process.
2.5 Preparation of water samples
Wastewater was obtained from the effluent of Self Point
STP (Hobart, Australia), while drinking water was obtained
from the tap in Hobart, Australia. Prior to analysis, the
samples were filtered through a 0.45 mm nylon membrane
syringe filter (Phenomenex, Australia) in order to eliminate
particulate matter. The samples were stored in dark glass
containers and kept in the refrigerator at �41C. The
conductivity of all the solutions used was measured using
a benchtop conductivity meter (lab CHEM-c) from TPS Pty.
(Springwater, Australia).
2.6 LLE
One millilitre of wastewater sample spiked with the
hypolipidaemic drugs (50 ppb) was acidified with 0.1 mL
hydrochloric acid, shaken, and then 1 mL of DCM was
added. The sample was then centrifuged (Eppendorf 5425
centrifuge, Hamburg, Germany) at 3000 rpm for 3 min. The
DCM layer was then separated and evaporated to dryness
under a gentle stream of air, and the dry residue was
reconstituted with 1 mL of Milli-Q water and used directly
for EKS-ESI-MS.
2.7 ESI-MS
For online hyphenation with EKS, different MS parameters
were optimised; these include sheath liquid composition,
sheath liquid flow rate, nebuliser pressure, drying gas
temperature and flow rate, and electrospray voltage. The
sheath liquid was 2 mM ammonium hydroxide in metha-
nol/Milli-Q water (50:50 v/v). A good trade-off between
electrospray stability and MS sensitivity was obtained with a
flow rate of 0.45 ml/min for the sheath liquid. Nebuliser
pressure was set to a minimum (7.00 psi) to provide stable
electrospray operation and to maximise the separation
efficiency. Additional spray chamber parameters were set
as follows: dry gas flow rate 5.00 L/min, drying gas
temperature 2501C, electrospray voltage 13.00 kV (end plate
voltage �3.00 kV). Total ion scan was performed at 200–650
m/z.
2.8 Validation study
BGE and standard solutions were freshly prepared daily for
assessing the reproducibility of the method. Precision was
expressed as percentage relative standard deviation and was
calculated by dividing the standard deviation by the mean
value of five replicate determinations for the within-day and
between-day measurements.
3 Results and discussion
The lack of CE-MS methods for the determination of
pharmaceuticals in environmental samples has inspired us
to develop a method for the determination of hypolipidae-
mic drugs in environmental samples. In previous studies
[25, 26], we reported the use of EKS and counter-flow
(CF-EKS) for the determination of non-steroidal
Electrophoresis 2010, 31, 1184–11931186 M. Dawod et al.
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
anti-inflammatory drugs in water samples. We obtained
sensitivity enhancements of 2400-fold with EKS, and while
the LODs with CF-EKS were even lower, the system we
developed cannot be transferred to the Agilent CE-MS
instrument as it is not possible to apply a negative pressure
on the inlet to establish a suitable counter-flow. Never-
theless, the enhancement with EKS alone was substantial
and when combined with MS this approach could be the
most appropriate solution for analysis of pharmaceuticals in
environmental samples.
3.1 Optimisation of separation selectivity
Highly efficient and rapid electrophoretic separations are
typically obtained when analytes migrate in the same
direction as the EOF (i.e. co-EOF separation). In previous
studies [25, 26], we reversed the EOF using the cationic
polymer HDMB, and found stable and reproducible results
when a small concentration of the reversing agent (0.1%
w/v) was added to the BGE. In the case of hyphenation
with ESI-MS, the addition of the surfactant to the BGE
is not desirable, because it will impair analyte detection as a
result of serious suppression of ionisation [27]. In the
present work, our experiments showed that EOF reversal
could be obtained without addition of HDMB to the running
buffer, as regular flushing with HDMB between runs
produced a reproducible reversed EOF. After establishing
a stable system with a reversed EOF, the possibility of
Figure 1. Chemical structures and molecular masses of the studied hypolipidaemic drugs.
Table 1. Electrophoretic mobility and pKa values for the
analytes and ion species included in the EKS-ESI-MS
study
Compound Electrophoretic
mobility (� 10�9 m2 v�1 s�1)
pKa value
Atorvastatin �7.90 4.20
Fluvastatin �9.60 4.10
Gemfibrozil �10.90 4.75
Pravastatin �8.40 4.60
Rosuvastatin �9.00 4.40
HCO3- �46.10 (�39.96)a) 6.35
CAPS �25.00 (�0.91)a) 10.40
a) Values between brackets are the effective mobility values for
the LE and TE calculated from PeakMaster 5.2 software (http://
www.natur.cuni.cz/�gas/pm52.exe).
Electrophoresis 2010, 31, 1184–1193 CE and CEC 1187
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
separating the hypolipidaemics in a MS-compatible BGE
was examined. We tested the separation of the drugs with
ammonium bicarbonate 30 mM, pH 9.00. In this BGE, the
separation was accomplished in less than 15 min, but
baseline resolution was not obtained for all analytes. In
order to improve resolution, we undertook a multivariate
optimisation of the ionic strength and methanol, as
both are well known to have a significant influence on
resolution and both are also MS-compatible [28]. The
experimental space was defined between 10–70 mM ammo-
nium bicarbonate and 0–70% v/v methanol. Higher
concentrations of the buffer increased electric current, and
this was undesirable both from the loss in efficiency due to
Joule heating effect [29] as well as the current limitations
due to ESI hyphenation (the system current limit is set to
50 mA by Agilent software). Higher percentages of methanol
caused a substantial reduction in the magnitude of the
reversed EOF, and resulted in severe prolongation of the
separation time. Over this experimental space, separations
were performed at nine training points using the four
corner points, the centre point, and four mid points.
Additional separations were performed at four random
points selected within the space, with two used in
the verification points of the artificial neutral network
(ANN) training, with the remaining two being used to
check the accuracy of the trained ANN. From these data an
ANN with an (2-11-3) architecture was trained with
very good correlation between the observed and predicted
values for separation time, current, and minimum resolu-
tion (r2 values of 0.9977, 0.9983, and 0.9998, respectively).
Figure 2 shows the response surface for minimum
resolution from the trained ANN. It is clear that the optimal
conditions based on the highest minimum resolution
showed that the best results were obtained with 60 mM of
the buffer and 60% v/v of methanol. This combination gives
a reasonable separation time and the current profile was
within the system limit for CE-ESI. A representative
separation obtained under these conditions is shown in
Fig. 3. The five hypolipidaemic drugs were successfully
separated in less than 20 min. It is noteworthy that we tested
the same ANN experiments but using ammonium acetate as
a BGE. Similar results were obtained, yet inclusion of
methanol (450% v/v) with acetate buffer gave a highly
intense peak of HDMB with m/z of 265, which suppressed
the ionisation of the drugs.
3.2 EKS-ESI
One of the online preconcentration strategies that has not
been hyphenated with MS is EKS. This technique was first
described by the group of Hirokawa [30, 31] for the analysis
of rare-earth ions. Urbanek et al.[32] applied EKS for
determination of sub-ppb levels of Fe(II), Co(II), and Ni(II)
with LODs at the low ng/L level for standards, however, the
LODs were ten times higher for real samples due to a lower
loading rate during the field amplified stacking injection
step. Recently, Busnel and co-workers [33] applied EKS for
peptide analysis with 1000–10 000-fold improvement in
sensitivity. Interestingly a buffer system that is compatible
with MS was used, yet coupling their method with MS was
not applied.
It is well known that there is a limitation in the BGEs
that can be used in any on-line CE-MS analysis. This limited
choice of electrolytes brings considerable restrictions not
only to the CE separation itself but also to the potential
sample stacking [9]. For EKS, this has implications on the
choice of the BGE as well as the LE and the TE. This
limitation leads to further difficulties for the EKS-ESI-MS
hyphenation.
In our system, ammonium bicarbonate was used as a
separation buffer, as the high electrophoretic mobility of
bicarbonate allows it to also act as a LE for these drugs. The
choice of the TE is not straightforward. The terminator
should ideally have an electrophoretic mobility lower than
that of the analytes of interest, but for MS detection, it is also
important that it does not foul the MS. In our previous
systems [25, 26] we used CHES as a TE. Initial experiments
using CHES as the TE gave poor results as CHES co-
migrated with atorvastatin (as shown in Fig. 4A) and also
caused contamination of the electrospray, requiring tens of
minutes for the signal to reach baseline again. CAPS was
investigated as an alternative TE because it has a higher pKa
value than CHES and therefore will have a lower mobility,
thereby minimising the chance of co-migration with any of
the analytes. When this was tried, surprisingly CAPS was
also more compatible with ESI-MS than CHES as it did not
contaminate the electrospray kit, and the baseline returned
to its normal value after detection of CAPS peak, as shown
in Fig. 4B.
The optimum conditions involved filling the capillary
with the BGE, after which a short plug of Milli-Q water was
Figure 2. Response surface estimated from the minimumresolution of hypolipidaemic drugs with concentration ofammonium bicarbonate and percentage of methanol.
Electrophoresis 2010, 31, 1184–11931188 M. Dawod et al.
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
introduced at a pressure of 50 mbar for 6 s. This plug has
been reported to be important for improvement of sensi-
tivity and reproducibility before the field amplified stacking
injection process [34]. In our system, there was no further
improvement in sensitivity with the water plug, yet it was
very important for stabilisation of the electric current during
electrokinetic injection of the sample. The sample was then
introduced electrokinetically by application of �10 kV for
170 s. Systematic attempts to increase the injection time
and/or injection voltage resulted in inferior reproducibility
of the method due to instability of electric current; this is
due to the introduction of large amount of the low
conductivity sample matrix. Moreover, breakage of the
capillary approximately 2.5 cm from the capillary inlet was
observed after 20–30 runs. This is most likely a result of a
high localised electric field at the interface between the
Figure 3. Base peak and extracted ionchromatograms of five hypolipidaemicdrugs 10 ppm in Milli-Q water. CEconditions: fused silica capillary,88 cm�50 mm id; BGE 60 mM ammo-nium bicarbonate containing 60%methanol, pH 9.00. Hydrodynamic injec-tion of sample at 50 mbar for 40 s.Separation voltage: �25 kV. MS condi-tions as in the text.
Electrophoresis 2010, 31, 1184–1193 CE and CEC 1189
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
low conductivity sample matrix and the high conductivity
BGE zone [35]. After injection, the TE (CAPS, 1 mM) was
introduced hydrodynamically by injection at 50 mbar for
10 s. This volume of TE was just sufficient for ITP step to
take place and longer injection plugs were judged to leave
smaller room for the capillary for subsequent separation of
the sample after destacking of the ITP step. Finally, the
separation voltage (�25 kV) was set for the ITP-CZE step.
Figure 4B shows the separation obtained with EKS-ESI-MS
using CAPS as TE.
Under optimum conditions, the EKS-ESI-MS method
yielded a 1000-fold improvement in detection sensitivity of
the hypolipidaemic drugs when compared with conven-
tional FASS. The LODs (at three times signal to noise) were
560, 180, 230, 200, and 260 ng/L for atorvastatin, fluvastatin,
gemfibrozil, pravastatin, and rosuvastatin, respectively.
Figure 4. Base peak chroma-tograms of five hypolipidae-mic drugs 50 ppb in Milli-Qwater. CE conditions: fusedsilica capillary, 88 cm� 50 mmid; BGE 60 mM ammoniumbicarbonate containing 60%methanol, pH 9.00. EKI ofsample at �10 kV for 170 s.Hydrodynamic injection of1 mM TE at 50 mbar for 10 s.(A) CHES and (B) CAPS.Separation voltage: �25 kV.MS conditions as in the text.
Table 2. Within-day and between-day reproducibilities (RSD%) of migration times, peak areas, and peak heights for EKS-ESI-MS
Compound Within-day RSD% (n 5 5) Between-day RSD% (n 5 5)
Migration time Peak area Peak height Migration time Peak area Peak height
Atorvastatin 0.75 2.90 8.12 6.38 7.96 9.36
Fluvastatin 0.41 2.89 3.44 4.49 4.06 6.32
Gemfibrozil 0.26 2.02 4.82 3.96 .65 5.70
Pravastatin 0.32 2.46 6.02 6.46 7.88 7.11
Rosuvastatin 0.21 2.04 2.88 3.24 4.51 5.50
Electrophoresis 2010, 31, 1184–11931190 M. Dawod et al.
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
Figure 5. Base peak chroma-togram obtained from EKS-ESI-MS of (A) drinking waterspiked with 50 mg/L of thedrugs and (B) drinking watersample spiked with 50 mg/L ofthe drugs, then diluted withMilli-Q water 1:4 and condi-tions as in Fig. 4B.
Figure 6. Base peak chroma-togram obtained from EKS-ESI-MS of (A) wastewaterspiked with 50 mg/L of thedrugs and (B) wastewatersample spiked with 50 mg/L ofthe hypolipidaemic drugsafter LLE. Conditions as inFig. 4B.
Electrophoresis 2010, 31, 1184–1193 CE and CEC 1191
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
3.3 Method validation
The reproducibility of the method was investigated by
determining the within-day and between-day precision values.
Within-day precision was evaluated by performing five
replicate separations of the five hypolipidaemic drugs at a
concentration of 50 mg/L. Between-day precision was evalu-
ated by performing the same separations for five different
days. Precision (expressed as percentage relative standard
deviation (RSD%)) was calculated for migration times, peak
areas, and peak heights. Table 2 shows the RSD% values for
within-day and between-day precision for the above para-
meters. These precision evaluation studies were performed on
the same capillary. The RSD% studies were conducted on
different capillary sections from the same batch as well as
different capillaries from different batches, and there was no
significant difference in RSD% values obtained.
3.4 Application to water samples
To demonstrate the potential of the developed EKS-ESI-MS
method for the analysis of hypolipidaemic drugs in different
water samples, the proposed method was applied for the
analysis of two water samples. Drinking water was collected
from the tap, while wastewater was collected from the
effluent of Self Point STP (Hobart, Australia).
Figure 5 shows the electropherograms of drinking water
spiked with 50 mg/L of the hypolipidaemic drugs and drinking
water spiked with 50 mg/L of each of the drugs after dilution of
1:4 with Milli-Q water. It is well known that the conductivity of
the sample plays an important role in any EKS system, as
matrix ions will be injected in the EKI step, and thereby
decrease the amounts of analytes injected. Measuring the
conductivity of the drinking water, it was found to be 51
mS/cm, as opposed to that of the standards which was 3.8 mS/
cm. A 1:4 dilution would reduce the conductivity of the water
sample to approximately 12 mS/cm and bring the conductivity
of drinking water close to the standard solutions and subse-
quently facilitate the applicability of EKS. An experimental
study confirmed that a 1:4 dilution was the lowest dilution
factor that decreases the conductivity of the sample enabling
online preconcentration by EKS.
When unspiked wastewater was analysed, none of the
drugs was detected, however, an unknown negative ion peak
was obtained with m/z of 315. Upon spiking the sample
with the drugs, the LODs were found to be 20 times higher
than the standard solutions due to the higher conductivity of
the wastewater sample compared to standard solutions of
the studied drugs [26]. The measured conductivity
of wastewater was 521 mS/cm, which requires several folds
of dilution, which dramatically increases the LODs and
hinders the application of EKS. For this reason, we devel-
oped a simple LLE method to extract the target analytes and
to eliminate ions present in the high conductivity sample
matrix. Figure 6 shows the electropherogram of a waste-
water sample spiked with 50 mg/L of each of the hypolipi-
daemic drugs before and after LLE with DCM. LLE of spiked
wastewater samples showed values for recovery %7SD
(n 5 3) of 44.6472.98, 88.6373.21, 92.3373.89,
49.6672.17 and 88.1272.09 for atorvastatin, fluvastatin,
gemfibrozil, pravastatin and rosuvastatin, respectively.
LODs were 1260, 203, 250, 403 and 294 ng/L for atorvasta-
tin, fluvastatin, gemfibrozil, pravastatin and rosuvastatin,
respectively. Although the LODs for wastewater samples are
higher than those obtained for the standard solutions of the
drugs, they are lower than the levels reported for some of
these analytes in wastewater samples in some cities in
Europe and North America. Gemfibrozil was detected at
concentrations up to 478 ng/L in STP effluents (Vancouver,
Canada) [36], 497 ng/L in STP effluent (Near Ebro river,
North East Spain) [37] and 340 ng/L in ground water
(Ontario, Canada) [38]. It is also important to report that the
EKS-ESI-MS method presented is suitable for the analysis of
the target analytes in low conductivity water samples with-
out the need for sample pretreatment.
4 Concluding remarks
The current work demonstrates the development of a new,
powerful CE-MS hyphenation. The simple and sensitive
method was applied for the separation and on-line preconcen-
tration of hypolipidaemic drugs in water samples. The method
uses a combination of EKS as an on-line preconcentration
technique in CE with the highly precise ESI-MS detection
system. The proposed method allowed the separation of five
hypolipidaemic drugs in less than 20 min. To the best of our
knowledge, this is the first separation and/or on-line
preconcentration system for these drugs. The enhancement
in detection sensitivity was 1000-fold over FASS and allows
LODs to get down to 180 ng/L. When the method was applied
to drinking and wastewater samples, the LODs achieved with
drinking water spiked were about five times higher than
standards, while for wastewaters, the high conductivity of the
samples required LLE to obtain sufficiently low LODs with
EKS. Although no further improvement in preconcentration
was obtained with LLE, the extraction method could be
modified in the future to obtain additional enhancement in
sensitivity. The proposed method could be applied directly
without sample pretreatment for analysis of some of these
analytes in some European and North American cities, where
the expected concentrations of such drugs are slightly higher
than the LODs achieved with the proposed method.
M. D. thanks the Egyptian ministry for higher education forprovision of a scholarship. This work was supported by theAustralian Research Council through the award of an ARC QEIIFellowship to M. C. B. (DP0984745), ARC postdoctoral Fellowshipto R. M. G. (DP0557083) and a Federation Fellowship to P. R. H.(FF0668673). The authors also thank Mrs. Francis Smith fromHobart city council for supplying the STP effluent samples.
The authors have declared no conflict of interest.
Electrophoresis 2010, 31, 1184–11931192 M. Dawod et al.
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
5 References
[1] Novakova, L., Lopez, S. A., Solichova, D., Satınsky, D.,Kulichova, B., Horna, A., Solich, P., Talanta 2009, 78,834–839.
[2] Heberer, T., J. Hydrol. 2002, 266, 175–189.
[3] Daughton, C. G., Ternes, T. A., Environ. Health Perspect.1999, 107, 907–938.
[4] Petrovic, M., Hernando, M. D., Dıaz-Cruz, M. S., Barcelo,D., J. Chromatogr. A 2005, 1067, 1–14.
[5] Canesi, L., Lorusso, L. C., Ciacci, C., Betti, M., Regoli, F.,Poiana, G., Gallo, G., Marcomini, A., Chemosphere2007, 69, 994–1002.
[6] Mimeault, C., Woodhouse, A., Miao, X. S., Metcalfe,C. D., Moon, T. W., Trudeau, V. L., Aquat. Toxicol. 2005,73, 44–54.
[7] Brain, R. A., Johnson, D. J., Richards, S. M., Hanson,M. L., Sanderson, H., Lam, M. W., Young, C., Mabury,S. A., Sibley, P. K., Solomon, K. R., Aquat. Toxicol. 2004,70, 23–40.
[8] Varenne, A., Descroix, S., Anal. Chim. Acta 2008, 628,9–23.
[9] Pantuckova, P., Gebauer, P., Bocek, P., Krivankova, L.,Electrophoresis 2009, 30, 203–214.
[10] Smyth, W. F., Electrophoresis 2006, 27, 2051–2062.
[11] Fonslow, B. R., Yates, J. R., J. Sep. Sci. 2009, 32,1175–1188.
[12] Ramautar, R., Somsen, G. W., de Jong, G. J., Electro-phoresis 2009, 30, 276–291.
[13] Font, G., Ruiz, M. J., Fernandez, M., Pico, Y., Electro-phoresis 2008, 29, 2059–2078.
[14] Campa, C., Coslovi, A., Flamigni, A., Rossi, M., Electro-phoresis 2006, 27, 2027–2050.
[15] Macia, A., Borrull, F., Calull, M., Aguilar, C., Trends.Anal. Chem. 2007, 26, 133–153.
[16] Ahrer, W., Scherwenk, E., Buchberger, W., J. Chroma-togr. A. 2001, 910, 69–78.
[17] Ahrer, W., Buchberger, W., Montash. Chem. 2001, 132,329–337.
[18] Macia, A., Borrull, F., Calull, M., Aguilar, C., Electro-phoresis 2004, 25, 3441–3449.
[19] Himmelsbach, M., Buchberger, W., Klampfl, C. W.,Electrophoresis 2006, 27, 1220–1226.
[20] Soto-Chinchilla, J. J., Garcıa-Campana, A. M., Gamiz-Gracia, L., Electrophoresis 2007, 28, 4164–4172.
[21] Breadmore, M. C., Thabano, J. R. E., Dawod, M.,Kazarian, A. A., Quirino, J. P., Guijt, R. M., Electro-phoresis 2009, 30, 230–248.
[22] Breadmore, M. C., Electrophoresis 2007, 28, 254–281.
[23] Breadmore, M. C., Haddad, P. R., Electrophoresis 2001,22, 2464–2489.
[24] Wang, S. R., Chen, Y., Wu, L. P., Miao, W. J., Xiong,M. J., Chen, C., Zhong, Z. R., Ye, L. M., J. Pharm.Biomed. Anal. 2008, 46, 243–249.
[25] Dawod, M., Breadmore, M. C., Guijt, R. M., Haddad,P. R., J. Chromatogr. A 2008, 1189, 278–284.
[26] Dawod, M., Breadmore, M. C., Guijt, R. M., Haddad,P. R., J. Chromatogr. A 2009, 1216, 3380–3386.
[27] Rundlett, K. L., Armstrong, D. W., Anal. Chem. 1996, 68,3493–3497.
[28] Altria, K. D., Goodall, D. M., Rogan, M. M., Chromato-graphia 1992, 34, 19–24.
[29] Evenhuis, C. J., Haddad, P. R., Electrophoresis 2009, 30,897–909.
[30] Hirokawa, T., Okamoto, H., Gas, B., Electrophoresis2003, 24, 498–504.
[31] Okamoto, H., Hirokawa, T., J. Chromatogr. A 2003, 990,335–341.
[32] Urbanek, M., Delaunay, N., Miche, R., Varenne, A.,Gareil, P., Electrophoresis 2007, 28, 3767–3776.
[33] Busnel, J. M., Lion, N., Girault, H. H., Electrophoresis2008, 29, 1565–1572.
[34] Chien, R. L., Burgi, D. S., J. Chromatogr. 1991, 559,141–152.
[35] Zhang, C. X., Thormann, W., Anal. Chem. 1996, 68,2523–2532.
[36] Verenitch, S. S., Lowe, C. J., Mazumder, A., J. Chro-matogr. A 2006, 1116, 193–203.
[37] Gros, M., Petrovic, M., Barcelo, D., Environ. Toxicol.Chem. 2007, 26, 1553–1562.
[38] Carrara, C., Ptacek, C. J., Robertson, W. D., Blowes,D. W., Moncur, M. C., Sverko, E., Backus, S., Environ.Sci. Technol. 2008, 42, 2805–2811.
Electrophoresis 2010, 31, 1184–1193 CE and CEC 1193
& 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com