electrocardiography artifact: what you do not know, you do not recognize

2
EDITORIALS Electrocardiography Artifact: What You Do Not Know, You Do Not Recognize William G. Stevenson, MD, William H. Maisel, MD “W at je niet kent, herken je niet” (1). (“What you do not know, you do not recog- nize.”) Reminder to students of electro- cardiography inscribed in the lecture hall of Professor Hein J.J. Wellens, MD. Measurement of the heart’s electrical activity by the surface electrocardiogram (ECG) was a breakthrough in the evaluation of cardiac arrhythmias. The ECG is critical to the safety and care of hospitalized patients with heart disease. Accurate and timely recognition of life-threaten- ing cardiac arrhythmias is required of all physicians who work in acute care settings. Of equal importance is the distinction of arrhythmias from artifact due to dis- connected leads, motion, electromagnetic interference, or equipment failure. The ECG records electrical activity from cardiac depo- larization and repolarization detected by electrodes on the skin. The resulting electrical potentials are small (sev- eral millivolts, at most) and occur over a frequency range of 0.05 to 100 Hz (cycles per second) (2,3). Signals of similar frequency and of similar or larger amplitude gen- erated by noncardiac sources cause ECG artifact. These signals can be classified as nonphysiologic or physiologic (2). Nonphysiologic artifact is usually caused by poor electrode contact, electromagnetic interference, or defec- tive ECG equipment. Physiologic artifact can be pro- duced by muscle activitity (electromyographic poten- tials) or skin. Because muscle electrical signals are high frequency and dissimilar from cardiac electrical activity, they can usually be filtered from the ECG. Skin electrical signals are produced by movement and stretching of the skin and are the primary source of ECG motion artifact. Because the potential for ECG artifact is widely recog- nized, the results of the survey conducted by Knight and coworkers (4) published in this issue of the American Journal of Medicine are surprising. Of 766 physicians who reviewed an ECG rhythm strip of artifact that mimicked a wide QRS tachycardia, only 6% of internists, 42% of car- diologists, and 62% of cardiac electrophysiologists recog- nized the ECG abnormality. These concerning misdiag- noses occurred despite visible sinus rhythm QRS com- plexes that marched independently through the tracing and were clearly dissociated from the artifact. The clinical implications of such a misdiagnosis are important. Electrophysiologic testing is appropriate for patients with nonsustained ventricular tachycardia and depressed left ventricular function after myocardial in- farction because implantable defibrillators (ICD) im- prove survival for those patients found to have inducible ventricular tachycardia. Thus a misdiagnosis of nonsus- tained ventricular tachycardia may extend the hospital stay or lead to unnecessary cardiac testing. In fact, a ret- rospective review of patients referred for arrhythmia management documented these misdiagnoses (5). The distinction of artifact from fact is not always easy. Artifact may mimic tachycardia, bradyarrhythmia, or pacemaker malfunction. The reliable hallmarks of either undisturbed true QRS complexes marching unimpeded through the tracing or nonphysiologic rapid or “prema- ture” electrical activity may be absent. When simulta- neous recordings of blood pressure or pulmonary artery pressure are available, as is common in intensive care units, dissociation of systolic pressure waveforms from the surface electrocardiogram may provide clues to the diagnosis of artifact. Occasionally artifact cannot be un- equivocally distinguished from an arrhythmia, and fur- ther assessment guided by concern for the patient’s safety is appropriate. Although the findings of this study are concerning, the incidence of clinically relevant misdiagnoses is unknown and may be less than implied by the present study. Sur- veys have important limitations and, by nature, invite bias. The response rate was low; only 16% of internists and fewer than half of the cardiologists and electrophysi- ologists responded. Perhaps the nonrespondents did not answer because they felt the ECG was obviously artifact. The inability of the respondents to interview the patient to clarify the clinical context may result in an overestima- tion of the incidence of the problem. Finding that an al- leged 19-beat run of rapid heart action was asymptomatic would hopefully prompt a more careful review of the tracing with attention to the possibility that artifact is the cause. Despite these limitations, the findings are provoc- ative. Even if the rate of misdiagnosis were reduced by Am J Med. 2001;110:402– 403. From the Cardiovascular Division, Brigham and Women’s Hospital, Boston, Massachusetts. Requests for reprints should be addressed to William G. Stevenson, MD, Cardiovascular Division, Brigham and Women’s Hospital, 75 Francis Street, Boston, Massachusetts 02115. 402 q2001 by Excerpta Medica, Inc. 0002-9343/01/$–see front matter All rights reserved. PII S0002-9343(00)00637-4

Upload: william-g-stevenson

Post on 31-Oct-2016

215 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Electrocardiography artifact: what you do not know, you do not recognize

EDITORIALS

Electrocardiography Artifact: What You Do NotKnow, You Do Not Recognize

William G. Stevenson, MD, William H. Maisel, MD

“Wat je niet kent, herken je niet” (1). (“Whatyou do not know, you do not recog-nize.”) Reminder to students of electro-

cardiography inscribed in the lecture hall of ProfessorHein J.J. Wellens, MD.

Measurement of the heart’s electrical activity by thesurface electrocardiogram (ECG) was a breakthrough inthe evaluation of cardiac arrhythmias. The ECG is criticalto the safety and care of hospitalized patients with heartdisease. Accurate and timely recognition of life-threaten-ing cardiac arrhythmias is required of all physicianswho work in acute care settings. Of equal importance isthe distinction of arrhythmias from artifact due to dis-connected leads, motion, electromagnetic interference,or equipment failure.

The ECG records electrical activity from cardiac depo-larization and repolarization detected by electrodes onthe skin. The resulting electrical potentials are small (sev-eral millivolts, at most) and occur over a frequency rangeof 0.05 to 100 Hz (cycles per second) (2,3). Signals ofsimilar frequency and of similar or larger amplitude gen-erated by noncardiac sources cause ECG artifact. Thesesignals can be classified as nonphysiologic or physiologic(2). Nonphysiologic artifact is usually caused by poorelectrode contact, electromagnetic interference, or defec-tive ECG equipment. Physiologic artifact can be pro-duced by muscle activitity (electromyographic poten-tials) or skin. Because muscle electrical signals are highfrequency and dissimilar from cardiac electrical activity,they can usually be filtered from the ECG. Skin electricalsignals are produced by movement and stretching of theskin and are the primary source of ECG motion artifact.

Because the potential for ECG artifact is widely recog-nized, the results of the survey conducted by Knight andcoworkers (4) published in this issue of the AmericanJournal of Medicine are surprising. Of 766 physicians whoreviewed an ECG rhythm strip of artifact that mimicked awide QRS tachycardia, only 6% of internists, 42% of car-diologists, and 62% of cardiac electrophysiologists recog-

nized the ECG abnormality. These concerning misdiag-noses occurred despite visible sinus rhythm QRS com-plexes that marched independently through the tracingand were clearly dissociated from the artifact.

The clinical implications of such a misdiagnosis areimportant. Electrophysiologic testing is appropriate forpatients with nonsustained ventricular tachycardia anddepressed left ventricular function after myocardial in-farction because implantable defibrillators (ICD) im-prove survival for those patients found to have inducibleventricular tachycardia. Thus a misdiagnosis of nonsus-tained ventricular tachycardia may extend the hospitalstay or lead to unnecessary cardiac testing. In fact, a ret-rospective review of patients referred for arrhythmiamanagement documented these misdiagnoses (5).

The distinction of artifact from fact is not always easy.Artifact may mimic tachycardia, bradyarrhythmia, orpacemaker malfunction. The reliable hallmarks of eitherundisturbed true QRS complexes marching unimpededthrough the tracing or nonphysiologic rapid or “prema-ture” electrical activity may be absent. When simulta-neous recordings of blood pressure or pulmonary arterypressure are available, as is common in intensive careunits, dissociation of systolic pressure waveforms fromthe surface electrocardiogram may provide clues to thediagnosis of artifact. Occasionally artifact cannot be un-equivocally distinguished from an arrhythmia, and fur-ther assessment guided by concern for the patient’s safetyis appropriate.

Although the findings of this study are concerning, theincidence of clinically relevant misdiagnoses is unknownand may be less than implied by the present study. Sur-veys have important limitations and, by nature, invitebias. The response rate was low; only 16% of internistsand fewer than half of the cardiologists and electrophysi-ologists responded. Perhaps the nonrespondents did notanswer because they felt the ECG was obviously artifact.The inability of the respondents to interview the patientto clarify the clinical context may result in an overestima-tion of the incidence of the problem. Finding that an al-leged 19-beat run of rapid heart action was asymptomaticwould hopefully prompt a more careful review of thetracing with attention to the possibility that artifact is thecause. Despite these limitations, the findings are provoc-ative. Even if the rate of misdiagnosis were reduced by

Am J Med. 2001;110:402– 403.From the Cardiovascular Division, Brigham and Women’s Hospital,Boston, Massachusetts.

Requests for reprints should be addressed to William G. Stevenson,MD, Cardiovascular Division, Brigham and Women’s Hospital, 75Francis Street, Boston, Massachusetts 02115.

402 q2001 by Excerpta Medica, Inc. 0002-9343/01/$–see front matterAll rights reserved. PII S0002-9343(00)00637-4

Page 2: Electrocardiography artifact: what you do not know, you do not recognize

half, the results of this study would be concerning. It isclear that a substantial number of physicians have diffi-culty recognizing ECG artifact.

The reasons for the high rate of misdiagnosis of ECGartifact are unclear. The ECG is a diagnostic test capableof yielding false negative or false positive results, but it isso routine that it is perhaps taken for granted. PerhapsECG artifact is such a common occurrence that it is nowlargely neglected in formal teaching. A review of six majorclinical cardiology and electrocardiography textbooksdiscloses no discussion of the topic. Apparently, manyphysicians have forgotten how closely ECG artifact canmimic an arrhythmia. We agree with the suggestion byKnight and colleagues that artifact should always beconsidered in the differential diagnosis of an ab-normal ECG. Dr. Wellens advises us that “what you donot know, you do not recognize.” Dr. Knight and col-

leagues remind us that “what you do not look for, you willnot recognize.”

REFERENCES1. Bar FWHM. Main stem stenosis. In: Smeets J, Doevendans P, Jo-

sephson M, et al, eds. Professor Hein J. J. Wellens. 33 Years of Cardi-ology and Arrhythmology. Norwell, Mass: Kluwer AcademicPublishers; 2000:495.

2. Smith M. Rx for ECG monitoring artifact. Crit Care Nurse. 1984;4:64 –56.

3. Bailey JJ, Berson AS, Garson A Jr, et al. Recommendations for stan-dardization and specifications in automated electrocardiography:bandwidth and digital signal processing. Circulation. 1990;81:730 –739.

4. Knight BP, Pelosi F, Michaud GF, et al. Physician interpretation ofelectrocardiographic artifact that mimics ventricular tachycardia.Am J Med. 2001:110:335–338.

5. Knight BP, Pelosi F, Michaud GF, et al. Clinical consequences ofelectrocardiographic artifact mimicking ventricular tachycardia.N Engl J Med. 1999;341:1270 –1274.

Electrocardiography Artifact/Stevenson and Maisel

April 1, 2001 THE AMERICAN JOURNAL OF MEDICINEt Volume 110 403