Download - Manual Ph Metru
Edition 10/10
Manual de Operare
Transmițători de măsurare inteligenți
pentru pH, redox (ORP), ISE, conductivitate, oxigen, clor
MV 4010
MV 4015
MV 4016
MV 4020
MV 4025
MV 4030
MV 4060
Sensortechnik Meinsberg GmbH Quality System certified to DIN EN ISO
Kurt-Schwabe-Straße 6 D-04720 Ziegra-Knobelsdorf/Gemany
Internet: www.meinsberg.de Tel.: +49 34327 623-0 Fax: +49 34327 623-79
1.
2.
3.
PREZENTARE......................................................................................................................3
Siguranta..............................................................................................................................3
Conexiunile Terminalului, Instructiuni de Operare pe scurt, Ajustarea di Fabrica.......4
Diagrame de conectare a senzorilor..................................................................................6
Senzor de Temperatura................................................................................................................. 6
pH/redox (ORP) si Electrozi Ion-Selectivi (ISE) ........................................................................ 7
Celula de masurare Conductivitate, Senzori de Oxigen si Clor................................................. 8
CALIBRARE DIN BUTOANE ...............................................................................................9
Calibrare Transmitator de pH (Transmitator ISE) ................................................................... 9
Calibrare Transmitator de Conductivitate .................................................................................11
Calibrare Transmitator Oxigen ...................................................................................................12
Calibrare Transmitator Redox (ORP).........................................................................................12
Calibrare Transmitator Clor........................................................................................................12
SOFTWARE PC „DINMODULE“ ........................................................................................13
Instalarea și punerea în lucru........................................................................................................13
Masurare On-line...........................................................................................................................14
Configurare.....................................................................................................................................14
Calibrare.........................................................................................................................................16
Conectarea mai multor emițătoare la o rețea (INSTRUMENTATIE MULTI-PARAMETRU) .........................................................................................................18
SOFTWARE / FIRMWARE UPDATE...................................................................................18
ÎNTREȚINERE, ELIMINAREA .............................................................................................19
ÎNTREȚINERE Sensor / Electrode .............................................................................................19
Disposal............................................................................................................................................19
4.
4.1
4.2
4.3
5.
5.1
5.2
5.3
5.4
5.5
6.
6.1
6.2
6.3
6.4
7.
8.
9.
9.1
9.2
10. SPECIFICATII..........................................................................................................................19
11. ACCESORII ............................................................................................................................19
2/14
1. Prezentare
Emițătoarele inteligente sunt adăpostite într-o cutie de plastic modular față de 22,5 mm pentru montare pe șine profil omega.
Acestea sunt excelent potrivite pentru măsurarea continuă a mai multor parametri, în legătură cu procesul de control
central PLC. Separate de auto-suficiente emițătoare conlucreaza cu senzori electrochimici și transforma semnale in analogic
sau interfețe de ieșire digitale. Datorită tehnologiei integrate microprocesor emițătoarele oferă configurare ușoară de bază
(un fel de semnal de ieșire de curent sau tensiune, scalarea semnalului de ieșire, limita, etc), în conformitate cu sarcina de
măsurare aplicație specifică prin intermediul USB-interfata si programul software "DinModule ". Producătorul are deja o
ajustare de bază din fabrică. Utilizatorul însuși poate realiza modificări și adaptări ulterioare personalizate ale ajustării de
bază înainte de a începe sau în timpul măsurării. Odată ce emițătorul este pre-reglat funcționează cu butoane de calibrare
rapidă pe partea frontală a emițătorului. Cele 3 butoane, un LED de stare și LC-Display permit calibrarea ușoară cu puncte
de pre-reglate fără PC. În cazul în care senzorul corespunzător este influențată de (de exemplu, soluție tampon pH, soluție
redox (ORP) tampon, soluție de calibrare a conductivității, aerul ambiant etc), standardul pre-reglat ca punct de calibrare
corespunzător, transmițătorul determină datele caracteristice senzorului automat și regleazăcaracteristica de transmisie în
consecință. În plus, o calibrare mai exacta a senzorului și ajustarea corespunzătoare a emițătorului pot fi realizate cu ajutorul
unui PC conectat și programul software "DinModule". Fiecare transmițător dispune de măsurarea parametrului
corespunzător și temperatura, dacă se utilizează un senzor de temperatură integrat cu sondă sau un senzor de temperatură
separat, două ieșiri izolate pentru curent 0 (4) ... 20 mA sau tensiune 0 ... 5 V DC, unul limita plutitoare sau ieșire de releu de
alarmă, precum și o interfață USB pentru scopuri de configurare. pH, potențial redox (ORP) și ISE (electrozi ion-selectivi)
emițătoare sunt potrivite pentru conectare directă cu electrozi combinați și separate de măsurare lanțuri. Celulele de
conductivitate și senzori de oxigen amperometrici pentru conectare la transmițătorul pentru conductivitate și oxigen ar trebui
să fie echipate cu un senzor de temperatură integrat. Emițătoarele accepta senzori de temperatură Pt 1000. Transmițătorul de
clor este potrivit pentru conexiune directă la senzori pentru măsurarea dezinfectante (clor liber și total, dioxid de clor, ozon),
în cazul în care acești senzori dispun de un curent de ieșire cu 2 fire analog.
2. Siguranta
Acest manual de utilizare conține informații fundamentale care trebuie respectate în conexiune cu instalarea, pornirea,
funcționarea și întreținerea emițătoarelor. Prin urmare, este absolut vital pentru utilizator să citiți acest manual înainte de a
lucra cu el.
Calificarea utilizatorului Emițătoarele și întregul sistem de măsurare au fost concepute pentru multi-parametru măsurători analitice. Se presupune că
utilizatorul / operatorul și personalul de întreținere au abilitățile profesionale corespunzătoare și experiență pentru a ști
proprietățile specifice ale sistemelor de măsurare analitică, sa stăpânească manipularea în condiții de siguranță a
substanțelor chimice, pentru exemplu, la întreținerea de electrozi / senzori, și să poată evalua pericolele și riscurile care
decurg din aceasta. Utilizatorul trebuie să se asigure că sunt respectate legislația și procedurile cu privire la protecția muncii,
prevenirea accidentelor și manipularea de produse chimice. Lucrări de instalații electrice
Emițătoarele vin gata de funcționare. Deschiderea aparatului poate expune piese sub tensiune. Conectarea la sursa de
alimentare și ieșirea de releu trebuie să fie făcut fără conexiune la alimentare și cu atenție la instrucțiunile privind
siguranța. O persoană de specialitate care este familiarizat cu pericolele asociate cu acestea ar trebui să facă un astfel de
lucru. Intervenție în aparat va duce la expirarea garanției.
Instalare și pornirea Instalați unitatea și modulele externe, astfel încât condițiile specificate în "Specificații vor fi ținute în orice circumstanțe. Șina
necesar pentru montarea trebuie să fie conectat în mod corect la impamantre - PE. Emițătoarele nu pot fi utilizate în condiții
exterioare, fără carcasă de protecție suplimentară. Nu utilizați cabluri de senzori și de interfață altele decât cele recomandate
de către producător. Pentru senzori și accesorii, se aplica instrucțiunile și reglementările din manualele de utilizare de operare
respective și fișele descriptive
Utilizare corectă Emițătoarele sunt destinate pentru măsurare, control și reglare parametri de analiză. Având în vedere "Specificații" paragraful
din capitolul 8, operare și folosind aparatul pentru această aplicație este utilizarea corectă. Orice aplicație dincolo de acest
lucru și modificări individuale sau extensii sunt necorespunzătoare și vor duce la pierderea dreptului la garanție. Când
conectați aparatul la senzori electrochimici, se ia întotdeauna în considerare viața lor și de uzura normală, deoarece aceasta
poate duce la funcționarea defectuoasă a sistemului de măsurare și reglementarea sau controlul asociate cu acestea.
Utilizatorul trebuie să ia măsuri adecvate pentru a limita efectele nocive ale astfel de funcționări defectuoase.
Instructiuni generale de siguranta Emițătoarele au fost fabricate și testate în conformitate cu indicațiile și standardele relevante pentru echipamente electronice de
măsurare. Acestea au părăsit fabrica în stare inofensiv tehnic. Buna funcționare și fiabilitatea operațională a emițătoarelor va fi
garantată numai dacă se respectă, în general, măsurile de siguranță și instrucțiunile specifice de siguranță prezentate în acest
manual de utilizare, și în instrucțiunile de utilizare ale componentelor. În cazul în care se presupune că funcționarea în
siguranță a aparatului sau a componentelor sale nu mai este posibilă, scoateți aparatul și componentele sale din serviciu și
protejati-l împotriva acționării accidentale. Funcționarea în condiții de siguranță nu va mai fi posibil în cazul în care unitatea
sau componentele sale
3/14
Prezinta deteriorari de la transport a / au fost depozitate în condiții improprii pentru un timp mai lung. Prezinta deteriorari vizibile nu mai funcționează așa cum este descris în acest manual de utilizare.
. Dacă aveți îndoieli, vă rugăm să contactați
furnizorul.
3. Conexiune la terminal, instrucțiuni de operare, reglare din fabrică
Conexiunele terminalului
22,5
mS/µS
99 Dimensiuni in mm
114
Terminal pH / redox(ORP) / ISE* conductivitate 2-electrode-cell
A1 guard ecran
A2 Electrod de masura Fir de semnal
A3 guard ecran
A4 Ecran/electrod de referinta
A5
A6
A7
A8
temp. senzor Pt 1000-1
temp. senzor Pt 1000-1
temp. senzor Pt 1000-2
temp. senzor Pt 1000-2
Pt 1000-1
Pt 1000-1
Pt 1000-2
Pt 1000-2
conductivitate 4-electrode-cell
Electrod de masura
supplying electrode
supplying electrode
electrod de masura
Pt 1000-1
Pt 1000-1
Pt 1000-2
Pt 1000-2
oxygen
anod
catod
anod
chlorine
intrare semnal +
intrare semnal -
intrare semnal +
Pt 1000-1
Pt 1000-1
Pt 1000-2
Pt 1000-2
Pt 1000-1
Pt 1000-1
Pt 1000-2
Pt 1000-2
* * ISE: măsurarea concentrației de ioni cu un electrod ion selectiv
Terminal
B1
B2
B3
B4
B5
B6
B7
B8
Description
power 15…24 V AC/DC
power GND
power GND
power 15…24 V AC/DC
output 1: 0…5 V or 0/4…20mA refer to GND*
output 0: 0…5 V or 0/4…20mA refer to GND*
relay closing contact (30 V DC/1 A, 125 V AC/0.5A)
relay closing contact (30 V DC/1 A, 125 V AC/0.5A)
Nota!
Este recomandat să instalați o unitate de alimentare
separată pentru alimentarea emițătoarelor. Cu aceasta
stabilitatea sistemului de măsurare electrochimică
poate fi asigurată.
O unitate de alimentare potrivita 24V DC / 1A pentru
montare pe sina DIN (pentru alimentarea a max. 10
emițătoare) pot fi livrate la cerere (a se vedea 11. piese
accesorii).
1)
4/14
Afisaj
Emițătoarele din seria MV40xx au un mic ecran în panoul frontal. Acesta
afișează valoarea categoria măsurare 1 și categoria 2 și valori de măsurare a
temperaturii. Este afișat suplimentar starea releelor.
Display-ul se schimbă după 20 de minute automat în modul de economisire energie. Apăsând orice buton al
transmițătorului respectiv afișajul este pornit timp de 20 minute din nou.
Instrucțiuni de operare pentru calibrare
Schimba la modul de calibrare (LED-ul va clipi portocaliu / galben), apăsând pe MODE. Frecvență intermitent al
LED-ul este utilizată ca un indicator de măsurare a stabilității valoare. Intermitent rapid și non-ritmic arată o
valoare de schimbare. Intermitent lent și regulat indică o valoare de măsurare de stabilizare la intrare.
Pentru calibrarea senzorului / electrodului legat cu transmițătorul de însoțire trebuie să fie scufundate sau
influențate de soluțiile standard sau medii de calibrare convenite pentru aceste puncte de calibrare. Apăsarea
butonului C1 calibrează la punctul de calibrare 1 (senzor / electrod într-o soluție standard de calibrare sau punct
de calibrare conform medie 1) și C2 calibrează la punctul de calibrare 2 (senzor / electrod într-o soluție standard
de calibrare sau punct de calibrare conformitate mediu 2). Butoanele de la punctele de calibrare și standardele
corespunzătoare pot fi adaptate cu ajutorul programului de software-ul de configurare.
- pH si ISE O calibrare în două puncte cu două standarde diferite (inclusiv valorile de măsurare așteptate) este
necesară. Din acest motiv fiecare buton trebuie să fi reglat la unul dintre aceste standarde. Pentru aceasta
calibrarea este terminată după calibrare a emițătorului cu senzor conectat la ambele soluții standard și
apăsând a fiecăruia dintre butoanele corespunzătoare
- Potențial redox (ORP), oxigen și conductivitate transmițător
Un singur punct de calibrare într-un standard sau o soluție de testat sau orice alt mediu de calibrare adecvat
este suficientă. Astfel selectabil, ambele butoane pot fi ajustate la identice sau diferite standarde și doar un
singur buton trebuie apăsat pentru calibrare.Redox potential (ORP), oxygen and conductivity transmitter
- Transmiterul de clor Calibrarea trebuie să se facă printr-o comparație sau măsurare de referință (a se vedea punctul 5.5),
deoarece nu există nici un standard de calibrare adecvat disponibil. Concentrația dezinfectanții din mediul
măsurat este determinată de o măsurare sau de laborator metodă independent (adică metoda fotometrică).
Semnalul de ieșire a transmițătorului sau un ecran conectat cu semnalul de ieșire trebuie să fie ajustată cu
ajutorul butoanelor de pe transmițător la valoarea de referință. Majorarea Butonul C1 (0,01 mg / l per
apăsare de buton) și scăderea butonul C2 (0,01 mg / l per apăsare de buton), semnalul de ieșire sau
valoarea afișată. Deci, este foarte ușor să se adapteze ieșire transmițătorului la valoarea curentă măsurată,
determinată prin compararea sau metodă de referință, din mediul în care senzorul este cufunda.
Părăsi modul de calibrare apăsând butonul MODE (LED-ul clipește verde din nou, corespunzător la modul de
măsurare).
Dacă după apăsarea butonului C1 sau C2 (pentru pH după apăsarea butonului de-al doilea) se modifică LED-ul roșu o eroare
de calibrare este produs (defecțiune senzor sau date caracteristice senzori în afara limitelor stabilite, cablu defect, eșec în
timpul calibrarii, etc). Un LED verde indică o calibrare corectă și validă, înseamnă date senzori caracteristice sunt în limitele
definite și caracteristica de transmisie a emițătorului a fost adaptat la datele de senzori. De asemenea, pentru acest
transmițător MV 4060 este o excepție. LED-ul indică o calibrare corectă și valabilă numai după apăsarea butonului mod (de
finisare de calibrare).
Pre-adjustments of the calibration points buttons C1 and C2 at delivery (factory adjustment)
Bu- ton
C1
pH
Solutie Tampon pH acc. NBS
pH = 6.87 at 25 °C
Solutie Tampon pH acc. NBS
pH = 4.0 at 25 °C
conductivitate electrod 2 celule Conductivity Standard
Solution 0.1 N KCl (12.9 mS/cm at 25 °C)
Standard Conductivitate Solutie 0.01 N KCl
(1413 µS/cm at 25 °C)
conductivitate electrod 4 celule Conductivity Standard
Solution 0.1 N KCl (12.9 mS/cm at 25 °C)
Standard Conductivitate Solutie 0.01 N KCl
(1413 µS/cm at 25 °C)
Oxigen
Aer ambient = 100 % air saturatie de oxigen
Masurare in gaze
Aer ambient = 102 % air saturatie de oxigen dizolvat masurat in apa
Redox (ORP)
Solutie tampon Redox UH = 427 mV
(Pt-Ag/AgCl: 220 mV at 25 °C)
Solutie tampon Redox UH = 675 mV
(Pt-Ag/AgCl: 468 mV at 25 °C)
C2
5/14
Categorii de masurare
Pe lângă categoria principală de măsurare (de exemplu pH-ul, conductivitatea, etc) fiecare emițător oferă mai multe așa-numitele
categorii de măsură secundare determinate sau calculate din categoria de măsurare principală și temperatura în categoria de
măsurare. Programul de configurare face posibilă afișarea următoarelor categorii de măsurare la ieșirile a emițătoarelor
relevante.
Categoria de
masura1
Valoare pH
potential redox (ORP) ca valoare
apsoluta voltaj in mV
ISE voltage in mV
conductivity in mS/cm or µS/cm
Categoria de masura2
Electrod de voltaj in mV
Categoria de
masura3 Categoria de
masura4 Categoria de
masura5
MV 4010
MV 4015
MV 4016
MV 4020 MV 4025
MV 4030
MV 4060
*
potential redox (ORP) referred to the standard hydrogen electrode in mV
ion concentration in concentration units
resistance in ohms salinitate in g/kg*
sensor current in nA oxygen partial pres- sure in kPa
air pressure ** oxygen saturation index oxygen concentration in in % mg/l or ppm
concentration in mg/l sensor output current in mA
Salinitatea este un parametru sumă special pentru apa de mare. Valorile de masă de la Institutul National de Oceanografie al Marii Britanii și UNESCO definesc o
gamă de salinitate de 2-42 g / kg. Aceste valori sunt baza de calcul a salinitatii referire la conductivitatea măsurată. Din cauza unei game limitate de 2 ... 42 g / kg
un calcul de salinitate este posibil numai în domenii de măsurare de 0 ... 20 mS / cm și 0 ... 100 mS / cm conductivitatea. ** Numai modelul MV 4030-LK măsoară presiunea aerului actuală.
4.
4.1
Diagrame de conetare senzori
Senzor de temperatura
Circuit cu 4 fire
A5 PT1000-1
A5 A6 A7 A8 Cablu de masura K43-PT/...
cable core
A6 PT1000-1
cable core
A7 PT1000-2
shield
A8 PT1000-2
shield
Pt 1000
Circuit de 3 fire Circuit de 2 fire
A5 A6 A7 A8 A5 A6 A7 A8
Pt 1000 Pt 1000
6/14
4.2 pH/redox (ORP) and Ion-Selective (ISE) Electrode
Electrod Combinat
A1 guard
A1 A2 A3 A4 measuring cable K43/...
-
A2 pH-signal
cable core
A3 guard
-
A4 reference
shield
Electrod de referinta si de masura separati
A1 A2 A3 A4
A1 guard
pH-sensor reverence-electrode
- -
A2 Semnal pH
cable core -
A3 guard
- -
A4 refeerinta
shield cable core
eletrod de pH combinat u senzzor de
temperatura integrat (K 54 cablu triaxial)
eletrod de pH combinat u senzzor de temperatura
integrat (K 19 cablu multifilar)
A5 A6 A7 A8 A5 A6 A7 A8
A1 A2 A3 A4 A1 A2 A3 A4 PE
A.1 guard
measuring cable K54/...
measuring cable K19/...
-
-
A2 Semnal pH
cable core (bl)
cable core
A3 guard
-
-
A4 Referinta
inner shield (rt)
shield
A5 PT1000-1
bridge circuit to A.6
green
A6 PT1000-1
bridge circuit to A.5
brown
A7 PT1000-2
bridge circuit to A.8
yellow
A8 PT1000-2
outer shield (grey)
white
7/14
În special pentru conectarea cablului de măsurare și cablul triaxial pentru electrozii de masurare pH, redox și ISE la intrare
rezistivă a transmiterului trebuie sa acorde atenție firului de semnal intern sa fie cât mai scurtă posibil și sa rămâna acoperit, cat
mai departe de ecran.
Cabu de semnal
intern
ecran
4.3 Celula de masurare Conductivitate, Oxigen si Senzor de Clor
conductivitate electrod cu 2 celule
A1 A2 A3 A4
A1 -
Cablu de
masura K43
Cablu de
masura K18
conductivitate electrod cu 4 celule
-
-
A2 measuring electrode
Miez
alb
A3 measuring electrode
ecran
ecran
A4 -
-
-
A5 PT1000-1
-
gri
A6 PT1000-1
-
Verde
A7 PT1000-2
-
maro
A8 PT1000-2
-
roz
A1 A2 A3 A4
measuring electrode
measuring electrode
supplying electrode
supplying electrode
A1 Electrod de
masura
Cablu de
masura K17 roz
A2 Electrod
furnizor
maro
A3 Electrod
furnizor
verde
A4 Electrod de
masura
gri
8/14
A5 PT1000-1
shield
A6 PT1000-1
alb
A7 PT1000-2
galben
A8 PT1000-2
albastru
senzor de oxigen
A1 A2 A3 A4
catod anod
A1 -
Cablu de
masura K39 -
A2 Cathode
alb
A3 Anode
ecran
A4 -
-
A5 PT1000-1
gri
A6 PT1000-1
maro
A7 PT1000-2
verde
A8 PT1000-2
roz
Senzor pentru dezinfectanti (clor, Dioxid de clor, ozon)
A1 A2 A3 A4
- +
An integrated or separated temperature sensor connected as per para. 4.1.
5.
5.1
Calibrarea din Butoane
Calibrare transmițător pH (transmițător ISE)
Pentru o calibrare ușoara și simpla "în timpul funcționării" a unui emițător conectat cu electrodul de însoțire folosi cele trei
butoane de pe partea frontală a emițătorului. Butoane C1 și C2 definesc punctele de calibrare, ce corespund soluțiilor
standard inclusiv estimate de măsură, gamă. În cazul în care electrodul conectat este cufundat în (de exemplu, soluție tampon
de pH) in conformitate cu punctele de calibrare standard, transmițătorul calculează automat datele caracteristice ale
senzorului și reglează funcția de transmitere în conformitate cu aceste date.
Ca un exemplu calibrarea unui transmițător pH-ului cu reglare fabrică, alin. 3 este descrisa. Programul software de
configurare face posibilă să se schimbe aceste valori la alte soluții tampon. Această caracteristică utilizează în mod automat,
valoarea corectă pH-ului în funcție de temperatura reală pentru măsurători și etalonări, cu senzor de temperatura conectat
(acorde o atenție la echilibrul de temperatura de electrodul și senzor de temperatură). Este egal ce soluție tampon este
utilizata pentru prima oară pentru calibrare; doar este necesară utilizarea butonului corect pentru soluția tampon conexă de
calibrare corectă.
Operare: Transmițătorul este în modul de măsurare normal, indicat prin clipirea LED-ului verde.
C2 MODE C1
green
9/14
Schimba la modul de calibrare apăsând butonul MODE. LED-ul isi schimbă culoarea în portocaliu.
C2 MODE C1
orange
Scufundați electrodul / senzorul în prima soluție tampon pH. LED-ul clipeste rapid aritmic portocaliu indicand
valoarea de măsură in schimbare. În cazul în care valoarea măsurată se stabilizează, clipirile vor fi lente și ritmice.
Acum, apăsați pe butonul de C1 în conformitate cu prima soluție tampon (punctul prima calibrare de exemplu, pH
6,87)
C2 MODE
orange
Clătiți electrodul / senzorul cu apă distilată / apă deionizată și se scufundă-l în cea de a doua soluție tampon.
LED-ul portocaliu indica unui valoare de măsură de stabilizare, de intermitent încet și ritmic.
Acum, apăsați butonul C2 în conformitate cu ajustate soluția tampon a doua (al doilea punct de calibrare de exemplu pH 4.01).
C2 MODE
orange
Calibrarea este corectă dacă după apăsarea butonului pentru al doilea tampon LED, se modifică la verde. Toate datele
caracteristice ale electrodului / senzorului sunt în limite acceptabile și caracteristica de transmisie a emițătorului este
ajustată în conformitate cu aceste date. Dacă LED-ul, se schimbă o perioadă scurtă de timp la roșu o eroare de
calibrare a avut loc sau datele senzorului caracteristice sunt limite definite in afara. Datele de calibrare stabilite în
această calibrare nu sunt acceptate și nu sunt stocate în acest caz. Acordați atenție că o eroare de calibrare este afișată
după utilizarea celei de a doua soluții standard.
C2 MODE
green
C1 C2 MODE
red
C1
C1
C1
Pentru a termina calibrarea și schimba la modul normal de măsurare apăsați butonul MODE. Acum, LED-ul se aprinde
intermitent din nou verde.
C2 MODE
orange
C1 C2 MODE
green
C1
Calibrarea unui lanț de măsurare-ion selectiv (ISE electrod combinat sau separat electrod de măsurare-ion selectiv și electrod de
referință), în legătură cu o ISE transmițător este aproape identic cu calibrarea pH-ului. Este o calibrare în două puncte, cu soluții
standard corespunzătoare pentru tipul de ioni care urmează să fie stabilite.. Soluțiile standard folosite cu unitatea lor de
concentrare (de exemplu, mg / l, ppm etc) determina domeniul de măsurare și unitatea concentrației pentru ieșirea de măsurare.
Utilizați butoanele C1 și C2 pentru o simpla "în timpul funcționării" calibrare cu standardele definite de asemenea. Reglați
butoanele în conformitate cu standardele, care ar trebui să includă gama estimată de măsurare. Datorită valorii diferită a ionilor o
verificare a datelor caracteristice ale senzorului nu este posibilă. O calibrare corectă are nevoie întotdeauna un punct de calibrare
(standard de însoțire), ar trebui să fie amplasat în partea a treia deasupra și un punct de calibrare este situat în sub a treia parte a
domeniului de măsurare. Acorde o atenție la dependența de temperatură a soluțiilor standard. Calibrarea are nevoie de experiențe
speciale și prezență precum toate aplicatiile cu electrozi ion-selectivi (ISE).
10/14
5.2 Callibrare transmiter de Conductivitate
Pentru măsurarea conductivității unui singur punct de calibrare este suficientă. Prin intermediul celor trei butoane și definirea
unui punct de calibrare corect calibrarea transmițătorului cu celula de măsurare de însoțire "în timpul funcționării" este foarte
usoara și simplă. Utilizați punctele de calibrare definite de către conform reglare fabrică. alin. 3 în conformitate cu domeniul
de măsurare estimat. Dependența de temperatură a conductivității a acestor standarde sunt stocate în emițător și se calculează în
conformitate cu temperatura reală dată de senzorul de temperatură integrat în celula de măsurare (atentie la timpul necesar
pentru echilibru de temperatură a senzorului cu temperatură mediului și). Alte puncte de calibrare pot fi reglate cu ajutorul
programului de software-ul de configurare. Întotdeauna unui calibrare corect bazată pe utilizarea stocate soluțiile standard de
calibrare în conformitate cu butoanele de însoțire
Operare: Transmiterul se afla in modul de masurare, indicat de LED-ul verde intermitent
C2 MODE C1
green
Schimbati la modul Calibrare apasand butonul MODE. LED-ul isi schimba culoarea in portocaliu
C2 MODE C1
orange
Acum, scufundați celula de conductivitate gata să măsoare (înainte de clătire de calibrare cu apă distilată /
deionizată) în soluția standard de conductivitate, în conformitate cu butonul C1 sau C2 programat. Un clipit rapid
intermitent non-ritmic a LED-ului portocaliu indică valoarea de măsurare in schimbare. În cazul în care valoarea
măsurată se stabilizează, Indicatorul clipeste lent și ritmic.
Acum, apăsați butonul în conformitate cu soluția standard de conductivitate utilizată (punct de calibrare).
C2 MODE
orange
Calibrarea s-a terminata corect dacă după apăsarea butonului se schimbă LED de culoare verde. Datele caracteristice
ale celulei (constanta celulei) este în limite acceptabile și caracteristica de transmisie a emițătorului este ajustată în
conformitate cu aceste date. În cazul în care se schimbă LED pentru o perioadă scurtă de timp la roșu o eroare de
calibrare s-a produs sau datele celulei sunt în afara limitelor definite. Datele de calibrare determinate nu sunt
acceptate și nu sunt stocate în acest caz.
C2 MODE
green
C1 C2 MODE
red
C1
C1
Pentru a finaliza schimbarea de calibrare pentru modul de măsurare normal prin apăsarea butonului MODE. Acum,
LED-ul se aprinde intermitent din nou verde.
C2 MODE
orange
C1 C2 MODE
green
C1
11/14
5.3 Calibrarea Transmiter de Oxigen
Calibrarea și reglarea unui emițător de oxigen împreună cu senzorul de însoțire sunt complet identice cu calibrarea unui
transmițător conductivitate pentru că este un singur punct de calibrare prea. Ajustarea din fabrică în conformitate cu Paragraf
3 pentru butoanele C1 și C2 definesc punctele de calibrare, care corespund pentru o calibrare usoara a senzorului in aer
pentru diferite aplicații. Pentru ambele aplicații, măsurarea conținutului de oxigen din aer (de exemplu, pentru tehnologia de
compostare sau supraveghere a aerului în interiorul clădirilor) și măsurarea oxigenului dizolvat în apă sau lichide,
întotdeauna ambiantul cu valoare medie 20,9 Vol. -.% Oxigen este utilizat ca mediu de calibrare (cum ar fi un standard).
Acordați atenție la timpul necesar pentru echilibru de temperatură de temperatură mediu și sonda integrate în senzor. Alte
puncte de calibrare pot fi reglate cu ajutorul programului de software-ul de configurare. Întotdeauna o calibrare corectă de
oxigen se bazează pe utilizarea punctelor de calibrare predefinite în conformitate cu butoanele de însoțire. În plus, să acorde
o atenție pentru a da senzorului timpul necesar pentru echilibru de temperatură în special pentru schimbări rapide de
temperatură prin punerea senzorului din apă în aer
5.4 Calibrarea transmiterului Redox (ORP)
Măsurarea Potențialului redox (ORP) este măsurarea potențialului între electrodul de măsură platină și electrodul de
referință, in cele mai multe din cazuri combinat ca o unitate comună, redox sau ORP electrodul combinat. Ieșirea din
emițătoare urmează acest potențial proporțional. De obicei o calibrare nu este necesară și in cele mai multe din cazuri nu e
solicitată. Oricum, funcționarea corectă a sistemului de de măsură redox poate fi verificată cu ajutorul soluțiilor de calibrare
redox (ORP). Astfel, la cerere și / sau necesară adaptarea caracteristicii transmițător la rezultatul de calibrare este posibilă ca
o transformare liniară, astfel că ieșirea de măsură este identică cu potențialul redox dat (ORP) soluție standard (de calibrare și
reglare). Calibrarea unui transmițător redox (ORP) este complet identică cu calibrarea unui transmițător conductivitate cu un
singur punct de calibrare. În conformitate cu domeniul de de măsură estimat sunt definite două soluții standard redox diferite
pentru butoanele 1 și 2 ca puncte de calibrare în conform ajustarea din fabrică. alin. 3. Dependența de temperatură a
potențialului redox a acestor standarde sunt stocate în transmițător, dar poate fi utilizat în cazul în care o senzor de
temperatură este conectat doar (atentie la timpul necesar pentru echilibru de temperatură a senzor de temperatură mediului
și). Alte puncte de calibrare pot fi reglate cu ajutorul programului de software-ul de configurare. Baza pentru o calibrare
corectă este întotdeauna de a folosi soluția standard de calibrare stocate în conformitate cu butonul de însoțire.
5.5 Calibrarea Transmiterului de Clor
Calibrarea transmițătorului clor este o calibrare, cu un singur punct. La această valoare curentă clor-ul mediului în care
senzorul este imersat trebuie să fie măsurată cu o trimitere sau măsurare comparație (de exemplu, metoda DPD-). Valoarea de
ieșire a emițătorului (semnal analogic Ieșirea, de afișare conectat cu ieșire analogică, afișare pe unitatea de afișare), trebuie să
fie ajustate la valoarea curentă astfel determinata dacă senzorul este cufundat în același mediu.
Operare: Transmițătorul este în modul de măsurare normală, indicat prin clipirea LED verde.
C2 MODE C1
verde
Schimba la modul de calibrare apăsând butonul MODE. LED-ul se schimbă culoarea de la portocaliu.
C2 MODE C1
portocaliu
Acum, reglați cu C1 (valoarea crește cu fiecare singură apăsare de buton, la 0,01 mg / l) și C2 (valoare scade cu
fiecare singură apăsare de buton, la 0,01 mg / l) valoarea exactă clor măsurată prin măsurare de referință. (de
exemplu DPD). Pentru o mare diferență butoanele pot fi apăsat un timp mai îndelungat. Important pentru acest
calibrare este verificarea vizuală a semnalului de ieșire dat de transmițător. Acest lucru poate fi realizat fie prin
intermediul software-ului "DinModule", valoarea afișată pe unitatea de afișare AZ 3000 sau prin semnalul analogic
de ieșire. Prin utilizarea unui semnal de ieșire analogic pentru verificarea ajustarea este important pentru a opri
funcția "HOLD" în timpul calibrării ("DinModule" software de configurare, buton). În caz contrar, semnalele de
ieșire analogice sunt înghețate în timpul calibrării (a se vedea, de asemenea, alin. "6.3 Configurare") și nu se poate
verifica nici o ajustare.
C2 MODE
portocaliu
C1
12/14
Pentru a termina calibrarea, schimba modul de măsurare normal prin apăsarea butonului MODE.
C2 MODE C1
C2 MODE
green
C1 C2 MODE
red
C1
Acum, LED-ul se aprinde intermitent din nou verde. Calibrarea este corect terminata, toate datele caracteristice ale
electrodului / senzorului sunt în limite acceptabile și caracteristica de transmisie a emițătorului este ajustată în
conformitate cu aceste date. Transmițătorul lucrează în modul de măsurare din nou. În cazul în care se schimbă LED
pentru o perioadă scurtă de timp la roșu o eroare de calibrare s-a produs sau datele senzorului caracteristice sunt in afara
limitelor definite. Datele de calibrare stabilite în această calibrare nu sunt acceptate și nu sunt stocate în acest caz.
13/14
9. Maintenance, Disposal
Emițătoarele din seria MV 40xx sunt aproape fără întreținere. În cazul murdări este permisă numai să curățați exteriorul cu o
cârpă umedă. Pentru aceasta modulele trebuie să fie deconectat de la sursa de alimentare! Curățarea cu detergent agresiv,
CARE conține solvenți (de exemplu, acetona) este interzisă. Altfel deteriorarea carcasei ABS poate apărea.
9.1 Mentenanta Senzor / Electrode
Pentru întreținerea și depozitarea senzorilor / electrozilor vă rugăm să respectați manualele corespunzătoare ale senzorilor /
electrozilor.
9.2 Înlăturare
Va rugam sa ne trimiteti instrumentele de măsură vechi și senzorii pentru eliminare. Sensortechnik Meinsberg GmbH se duce
înapoi gratuit și reciclează / dispune resturi electronice într-un mod competent.
Atenție!
Nu aruncați instrumente de măsurare vechi, la deșeurile menajere, acest lucru este ilegal. Vă rugăm să evitati eliminarea la
punctele de colectare publice.
10. Specificatii
Specificatii generale pentru transmiterele de masurare:
Configuratie prin intermediul interfeței USB și program de configurare corespunzătoare
Scaling și determinarea semnalelor de ieșire analogice
meniul de calibrare, definiția din punctele de etalonare
Compensarea temperaturii
LCD grafic, 48 x 64 pixel, luminat, afisaj pentru 3 valori (valoare 1, valoare 2 si temperatura)
MODE: Schimbare la modul calibrare C1, C2: Valori de calibrare (configuratie predefinita) sau ajustre a semnalului
de iesire prin comparatie sau masuratoare de referinta (transmiter de clor) LED multicolor (semnalizare masurare stabila) LED portocaliu: Calibrare LED verde: Masurare LED rosu: Eroare de calibrare
2 x 0(4) ... 20 mA or 0 ... 5 V, Izolate incarcare 500 acuratete 0,2 %
incarcare 2 kΩ, acuratete 0,2 %
USB, conectabil bus la o retea prin RS-232
Maxim inchidere contact 125 V AC, 60 V DC; 30 VA
15 ... 24 V AC/DC, aprox. 1.5 VA
0 ... 50 °C
acc. EN 61326 class B
carcasă din plastic pentru montare pe șină DIN EN 50022-35, grad de protectie IP 40 terminale pentru cabluri Secțiune 0.2 ... 2.5 mm2; 3-pini soclu pentru RS-232 cable connector; optional sina pentru conector bus
Afisaj
Functiile Butoanelor
Indicator de Status
Semnale de Iesire
Iesiri de Curent
Iesiri de Voltaj Interfata
Iesire de releu
Sursa de alimentare
Temperatura ambientala
EMC
Carcasa Conexiuni electrice
Domeniile de măsurare, electrozi / senzori recomandat potriviti pentru conectarea la emițătoarele și desenele dimensionale de
incintă în conformitate cu fișele individuale de date tehnice detaliate.
11. Accesorii
Accesorii optionae si echipamente aditionale
Interface cable connection for the USB connection (3 pins phone plug / USB connector) and PC software program „DINModule“
Sina pentru conector bus Connector for direct mounting on the plastic enclosure for network connection of several measuring transmitters by the RS-232 interface
Unitate de alientare Power supply unit in an enclosure for mounting on rail tracks (input 85... 264 V AC; output 24 V)
Protectia carcasei IP 65 Carcasă de protecție pentru montaj pentru unul sau mai multe emițătoare de măsurare pe teren în aer liber
Unitate afisaj AZ 3000 Modul într-o incintă pentru montarea pe sine pentru afișarea a până la 8 valori măsurate ale sistemului de
logare de date inclusiv, definiție de a 2 conexiuni valori de măsurare virtuale, 2 ieșiri analogice și de
transmitere a datelor (convertor de interfață)
14/14
Interfata MV