Page 1: Kit Contents Equipment Assay Procedure – Heat Lysis

95°CLysis Block

WorkflowTray

TransferRack

Kit Contents Equipment

GASReaction

Tubes

C/GReaction

Tubes

DilutionBufferTube

LysisBufferTube

PCM305100EN00 (04/17)

Amplification and Detection

SolanaInstrument

Transfer 50 µL to each of the GAS and C/G

Reaction Tubes.

1

Rehydrate each Reaction Tube by pipetting up

and down 5 times.

2 43

Close lids tightlyand proceedto next step.

65 8

Close the lid and selectappropriate assay protocol.

Run complete in 25 minutes.

Review results.Lower the Reaction Tubesinto Solana.

Release the Reaction Tubesfrom the Transfer Rack bypulling the Transfer Rack

toward your body.

Procedure Card

Strep Complete AssayUsing ESwabTM Collection Device

Study the Package Insert thoroughly before using this Procedure Card. This is not a complete Package Insert.

7

Use the Transfer Rack to liftReaction Tubes from the

Workflow Tray.

ESwab

Assay Procedure – Heat Lysis

VORTEX5 seconds

VORTEX5 seconds

Heat for 5 minutesat 95°C.

Add 50 µL to Dilution

Buffer Tube.

1 3

Add 50 µL to

Lysis Tube.

42 5 6

Setup the Workflow Tray withpatient samples, Lysis BufferTubes, Dilution Buffer Tubes,

and Reaction Tubes.

7 8

VORTEX5 seconds