functional phase-correlated micro-ct imaging of small rodents with low dose

7
Functional Phase–Correlated Micro–CT Imaging of Small Rodents with Low Dose Stefan Sawall a , Frank Bergner a , Andreas Hess b , Robert Lapp c , Markus Mronz c , Marek Karolczak a , and Marc Kachelrieß a a Institute of Medical Physics (IMP), Friedrich-Alexander-University of Erlangen-N¨ urnberg, Henkestr. 91, Erlangen, Germany b Institute of Experimental and Clinical Pharmacology and Toxicology, Friedrich-Alexander-University of Erlangen-N¨ urnberg, Fahrstraße 17, Erlangen, Germany c CT Imaging GmbH, Henkestr. 91, Erlangen, Germany ABSTRACT Functional imaging of an animals thoracic region requires cardiac and respiratory gating. The information on respiratory motion and ECG required for double-gating are extracted from the rawdata and used to select the projections appropriate for a given motion phase. A conventional phase-correlated reconstruction (PC) therefore uses only a small amount of the total projections acquired. Thus the resulting images comprise a high noise level unless acquired with very high dose, and streak artifacts may occur due to the sparse angular sampling. Here, we are aiming at getting high fidelity images even for relatively low dose values. To overcome these issues we implemented an iterative reconstruction method encompassing a five-dimensional (spatial, cardiac-temporal, respiratory-temporal) edge-preserving filter. This new phase-correlated low-dose (LDPC) reconstruction method is evaluated using retrospectively-gated, contrast-enhanced micro CT data of mice. The scans performed com- prise 7200 projections within 10 rotations over 5 minutes. A tube voltage of 65 kV was used resulting in an administered dose of about 500 mGy. 20 respiratory phases and 10 cardiac phases are reconstructed. Using LDPC reconstruction the image noise is typically reduced by a factor of about six and artifacts are almost removed. Reducing the number of projections available for reconstruction shows that we can get comparable image quality with only 200 mGy. LDPC enables high fidelity low-dose double-gated imaging of free breathing rodents without compromises in image quality. Compared to PC image noise is significantly reduced with LDPC and the administered dose can be reduced accordingly. 1. INTRODUCTION Double–gated in–vivo small animal cone–beam micro–CT scans provide five–dimensional information about the object: the three volume dimensions plus the temporal dimensions of the respiratory motion and the heart motion, respectively. Double gating is typically performed to separate respiratory from cardiac motion when imaging the animal’s lung or heart. When ignoring the fact that cardiac motion also induces motion in the lung itself imaging the lung can be done using respiratory gating only. 1 However, to image the heart of small animals one cannot neglect the translations and deformations of the heart induced by the respiratory motion. Hence cardiac imaging requires double gating. 2–6 To obtain functional information of the heart or lung it may be important to cover the complete motion cycles and thereby to reconstruct four– or five–dimensional representations of the object. While this paper focusses on the five–dimensional reconstructions the low–dose phase–correlated image reconstruction method we propose can also be applied to four–dimensional data. It is our aim to significantly improve the image quality achievable in low–dose single– or double–gated micro–CT scans of small specimen. On the one hand we want to reduce streak artifacts that result from sparse angular sampling and on the other hand we want to reduce image noise resulting from the small amount of photons available in a given combination of respiratory and cardiac phase. Send correspondence to Stefan Sawall ([email protected])

Upload: dkfz

Post on 05-Apr-2023

0 views

Category:

Documents


0 download

TRANSCRIPT

Functional Phase–Correlated Micro–CT Imagingof Small Rodents with Low Dose

Stefan Sawalla, Frank Bergnera, Andreas Hessb, Robert Lappc, Markus Mronzc, MarekKarolczaka, and Marc Kachelrießa

aInstitute of Medical Physics (IMP), Friedrich-Alexander-University of Erlangen-Nurnberg,Henkestr. 91, Erlangen, Germany

bInstitute of Experimental and Clinical Pharmacology and Toxicology,Friedrich-Alexander-University of Erlangen-Nurnberg, Fahrstraße 17, Erlangen, Germany

c CT Imaging GmbH, Henkestr. 91, Erlangen, Germany

ABSTRACT

Functional imaging of an animals thoracic region requires cardiac and respiratory gating. The information onrespiratory motion and ECG required for double-gating are extracted from the rawdata and used to select theprojections appropriate for a given motion phase. A conventional phase-correlated reconstruction (PC) thereforeuses only a small amount of the total projections acquired. Thus the resulting images comprise a high noiselevel unless acquired with very high dose, and streak artifacts may occur due to the sparse angular sampling.Here, we are aiming at getting high fidelity images even for relatively low dose values. To overcome these issueswe implemented an iterative reconstruction method encompassing a five-dimensional (spatial, cardiac-temporal,respiratory-temporal) edge-preserving filter. This new phase-correlated low-dose (LDPC) reconstruction methodis evaluated using retrospectively-gated, contrast-enhanced micro CT data of mice. The scans performed com-prise 7200 projections within 10 rotations over 5 minutes. A tube voltage of 65 kV was used resulting in anadministered dose of about 500 mGy. 20 respiratory phases and 10 cardiac phases are reconstructed. UsingLDPC reconstruction the image noise is typically reduced by a factor of about six and artifacts are almostremoved. Reducing the number of projections available for reconstruction shows that we can get comparableimage quality with only 200 mGy. LDPC enables high fidelity low-dose double-gated imaging of free breathingrodents without compromises in image quality. Compared to PC image noise is significantly reduced with LDPCand the administered dose can be reduced accordingly.

1. INTRODUCTION

Double–gated in–vivo small animal cone–beam micro–CT scans provide five–dimensional information about theobject: the three volume dimensions plus the temporal dimensions of the respiratorymotion and the heart motion,respectively. Double gating is typically performed to separate respiratory from cardiac motion when imagingthe animal’s lung or heart. When ignoring the fact that cardiac motion also induces motion in the lung itselfimaging the lung can be done using respiratory gating only.1 However, to image the heart of small animals onecannot neglect the translations and deformations of the heart induced by the respiratory motion. Hence cardiacimaging requires double gating.2–6 To obtain functional information of the heart or lung it may be importantto cover the complete motion cycles and thereby to reconstruct four– or five–dimensional representations of theobject. While this paper focusses on the five–dimensional reconstructions the low–dose phase–correlated imagereconstruction method we propose can also be applied to four–dimensional data. It is our aim to significantlyimprove the image quality achievable in low–dose single– or double–gated micro–CT scans of small specimen.On the one hand we want to reduce streak artifacts that result from sparse angular sampling and on the otherhand we want to reduce image noise resulting from the small amount of photons available in a given combinationof respiratory and cardiac phase.

Send correspondence to Stefan Sawall ([email protected])

Figure 1. Illustration of the dual source concept and a picture of the scanner used for our study. Mouse 1 was scannedwith a single source version while mouse 2 was scanned in the dual source version.

2. METHOD

2.1 Scan Mode

For our scans we chose a scan mode that acquires 7200 projections during 10 contiguous rotations. In case ofunexpected incidences, such as object motion, we may use only the subset of data that is not corrupted. Inaddition this scan mode allows us to study the dose reduction potential by doing reconstructions from only asubset of the data, e.g. by using only the first five rotations for image reconstruction.

2.2 Intrinsic Gating

To correlate our reconstruction with the motion phases of the animal heart and lung we detect correspondingsynchronization information directly from the rawdata by evaluating the center of mass within certain ROIs inthe projection images.6–9 In the case of slowly rotating micro–CT scanners a motion period is much shorter thanthe time needed for a half or a full rotation as in contrast to fast rotating clinical CT scanners. The extractionof the synchronization signal may therefore neglect the gantry rotation and such rather simple methods aresufficient to detect the periodic motion.

2.3 Image Reconstruction

Let f denote the object, X the x–ray transform operator, and p be the projection data, such that p = Xf .Our standard image reconstruction is the Feldkamp algorithm which we denote with X−1

Std and which resultsin the standard image fStd = X−1

Stdp.10 The standard Feldkamp algorithm is not phase–correlated. We make

us of it below to define the McKinnon–Bates algorithm. To perform respiratory and cardiac–correlated imagereconstruction we use a phase–correlated Feldkamp algorithm X−1

PC that filters and backprojects only thoseprojections that lie in the desired temporal window. The temporal window itself is defined by specifying therespiratory phase r and the cardiac phase c, both are values between 0 and 1 and count relative to one motionperiod, and by specifying the widths Δr and Δc of these temporal windows. The respiratory and cardiac phase–correlated image is denoted as fPC = X−1

PCp. Since only few projections fall into the desired temporal window,streak artifacts may occur unless a very large number of projections at very fine angular increments is acquired.The McKinnon–Bates (MKB) algorithm can be used to address this issue.11, 12 It works as follows. First,a standard reconstruction is performed to obtain a prior image. This prior image is blurry in those regionswhere motion is present, and it is of high image quality elsewhere. Then, a forward projection of the priorimage is performed and subtracted from the measured rawdata. These subtracted data are then used for aphase–correlated reconstruction which is added to the prior image. Mathematically:

fMKB = fStd + X−1PC(p− XfStd).

Since the respiratory motion dominates the cardiac motion we extend the MKB approach to a two stepalgorithm. In a first step we apply MKB only to respiratory gating and in the second step we use the respiratory–gated MKB image as a prior for the additional cardiac gating. This procedure ensures the best possible priorfor cardiac gating.

To reduce noise we apply edge preserving denoising bilateral filters13 in up to five dimensions. To definethe bilateral filter let us restrict to one dimension, for convenience. Bilateral filtering of a function f(x) is thendefined as

Bf(x) =

∫dtD(x, t)R(x, t)f(t)∫dtD(x, t)R(x, t)

with B denoting the bilateral filtering operator and

D(x, t) = e−(x− t

σx

)2

R(x, t) = e−(f(x)− f(t)

σf

)2

being the domain and the range filter, respectively. The parameters σx and σf are the widths of the Gaussiandomain and range filters, respectively. Since respiratory and cardiac gating yields five–dimensional volumesf(x, y, z, r, c), with r denoting the respiratory and c denoting the cardiac phase, we can apply bilateral filteringin up to five dimensions. The corresponding domain filter parameters are denoted as σx, σy, σz, σr and σc,respectively.

The final volume, obtained by bilateral filtering the MKB volume, is the low–dose phase–correlated volume

fLDPC = BfMKB

which we want to compare against the conventional (and more simple) phase–correlated volume fPC in thefollowing.

3. MEASUREMENTS

We currently have several data sets available for double gating that were measured in-house. Two of them shallbe presented here. The data of both mice were acquired with a dedicated in–vivo cone–beam micro–CT scanner(TomoScope Synergy Twin, CT Imaging GmbH, Erlangen, Germany) in single source mode installed at theInstitute of Medical Physics, Erlangen, Germany. The system consists of a micro focus x–ray source mountedat a distance of 170 mm to the isocenter and a 1024× 1024 flat panel detector mounted at a distance of 39 mmto the isocenter. The size of the quadratic detector pixels was 50 µm. To increase the detector’s readout rateto 25 frames per second a two–by–two binning of the detector pixels was performed. The scans were conductedat 65 kV tube voltage with a tube current of 0.3 mA. 7200 projections were acquired per scan in a circulartrajectory over an angular range of 3600◦. The scan time for these ten rotations was 288 s. The tube currenttime product was 87 mAs, the absorbed dose was measured as 500 mGy in each case.

Both mice were anesthesized with a combination of ketamine and rompun. A bloodpool contrast agent(Fenestra VC, Advanced Research Technologies, Saint Laurent for mouse 1 and Binition, Binitio BiomedicalInc., Ottawa for mouse 2) was used for contrast enhancement. The mean respiratory rate of mouse 1 was150 rpm (respirations per minute) and the mean heart rate was 300 bpm (beats per minute) whereas mouse2 showed a mean respiratory rate of 120 rpm and a mean heart rate of 260 bpm. With 40 ms exposure timeper projection, about 11 projections were acquired during one respiration cycle and about 6 projections wereacquired during one cardiac cycle of each mouse. This, in turn, implies that the respiratory temporal resolutionis limited to about 10% of the respiratory cycle and that the cardiac temporal resolution is limited to about 20%of the cardiac cycle. To avoid a further decrease in temporal resolution the RC–correlated image reconstructionswere carried out with a respiratory window width of Δr = 10% and a cardiac window width of Δc = 20%. Thesize of each reconstructed volume was set to 512× 512× 512 voxels with a voxel size of 40 µm. Volumes were

Figure 2. Comparison between conventional phase–correlated reconstruction (respiratory and cardiac gating) of mouse1 and our new iterative low–dose phase–correlated method that is a combination of an extended MKB approach withbilateral filtering. The difference image shows no object structures which indicates that the new method does not resultin a reduction of temporal resolution. Noise was reduced from 175 HU to 30 HU. Respiratory phase r = 0%, cardiacphase c = 80%. (0 HU / 400 HU)

generated with a step size of 5% in the respiratory cycle and a step size of 10% in the cardiac cycle. Therefore,each 5D dataset consisted of a total of 200 volumes resulting in a memory requirement of 50 GB. The data wereprocessed on a standard PC equipped with 64 GB of memory using a commercially available software package(RayConStruct-IR, RayConStruct GmbH, Nurnberg, Germany). As the data were collected retrospectively, novote from the ethical committee was required.

4. RESULTS

Although the contrast enhancement in the heart between blood and myocardium of mouse 1 is only about110 HU our reconstruction and postprocessing approach yields images of surprisingly high quality. Figure 2compares the results of our new RC–gated approach (LDPC) with a conventional RC–gated approach (PC)that simply feeds only those projections corresponding to the desired motion phases into a modified Feldkampreconstruction algorithm. With our gating parameters of using a respiratory window of 10% width and a cardiacwindow of 20% width only 2% of the projections fall into desired motion phases. Hence the conventional simplegating approach suffers from significantly increased image noise and from streak artifacts due to sparse viewsampling. With the new and improved algorithm we can reduce image noise from 175 HU to 30 HU and wecan remove all streak artifacts while maintaining the full temporal resolution (as far as one can tell from thedifference images). It should be noted that image noise was measured in a homogeneous volume of interest (VOI)in the liver region. The same VOI was used for all reconstructions and it is not necessarily intersecting thoseslices that are shown in the images. Mouse 2 shows an enhancement between blood and myocardium of about450 HU. Figure 3 contains a comparison of the PC reconstruction method against the proposed LDPC methodat different dose levels. A decrease in dose was simulated by limiting the number of gantry rotations and thusthe number of projections available for reconstruction. Compared to the PC reconstruction the usage of LDPCreduces the voxel noise from 125 HU to 26 HU at a dose level of 500 mGy. Furthermore the streak artifactsdue to the sparse angular sampling of projections have been reduced significantly. A decrease in dose down to250 mGy shows an increase of streak artifacts in the images reconstructed with PC which are nearly removed

Figure 3. Phase–correlated reconstructions of mouse 2 centered at r = 60% and c = 0% with window widths Δr = 10% andΔc = 20%. The left panel shows the conventional phase–correlated reconstructions while the right panel is the proposedlow dose approach. Dose levels ranging from 60 mGy to 500 mGy were obtained by using only fractions of the dataavailable. (400 HU / 800 HU)

within the LDPC reconstruction. The voxel noise is raised to 198 HU in the PC images and 57 HU in the LDPCimages respectively. Administering a dose of only 60 mGy reduces image quality more significantly. The imagesoriginating from a standard phase–correlated reconstruction suffer from a high noise level of about 345 HU andstreak artifacts prevent the anatomical structures of interest, e.g. the heart, from being identified as such. Incontrast the LDPC reconstruction is still capable of recovering anatomical details.

4.1 Comparison with other approaches

The experimental data available to us are very limited. Further on, mouse 1 shows only a very low enhancementbetween blood and myocardium, which is probably due to misadministered contrast agent. Regarding otherreferences, such as references,3–5 for example, it can be seen that a typical enhancement should be in the orderof 400 to 500 HU in the mouse heart. Since the said references also state important image quality parameterstogether with estimates of the dose we want to compare our results to those published in the literature. We havetherefore prepared table 1 to see the most important differences at a glance. It should be noted that the datashown in this table only allow for a very rough comparison between the methods, mainly because importantdetails of how certain parameters were measured or calculated are not disclosed in the publications.

The last row of the table further shows a quality parameter Q given as

Q ∝ 1

σ√DΔ2

where σ is the image noise observed in unenhanced regions of the mouse tissue, D is the dose reported,and Δ is the sampling distance of the rays scaled to the isocenter (which would typically be the diameter ofthe field of measurement divided by the number of detector rows or columns). The definition of the qualityparameter reflects the fact that image noise variance is proportional to one over dose and to one over thefourth power of spatial resolution. The constant of proportionality is chosen to obtain a quality value of 100%

Reference3 Reference4 Reference5 Mouse 1 Mouse 2

Type single source single source single source single source single sourceGating 6×prospective retrospective retrospective retrospective retrospective

R–synchronization ventilation pressure chamber pneumatic cushion kymogram kymogramC–synchronization ECG ECG ECG kymogram kymogramSource to isocenter 480 mm 540 mm 573 mm 170 mm 170 mm

Scan time 9 min 50 s 80 s 5 min 5 minTube voltage 80 kV 80 kV 80 kV 65 kV 65 kVTube current 150 mA 50 mA 50 mA 0.3 mA 0.3 mA

Integration time 10 ms 12 ms 10 ms 40 ms 40 msProjections 252 4160 8000 7200 7200

Current time product 2268 mAs 2500 mAs 4000 mAs 86.4 mAs 86.4 mAs

Scan range 189◦ 3600◦ 5760◦ 3600◦ 3600 ◦View increment 0.75◦ 0.86◦ 0.72◦ 0.5◦ 0.5◦

Spatial sampling 90 µm 150 µm 242 µm 80 µm 80 µmFull cycle dose 1840 mGy 280 mGy 248 mGy 500 mGy 500 mGy

Respiratory rate 90 rpm 51 rpm 30 rpm 150 rpm 120 rpmHeart rate 500 bpm 405 bpm 330 bpm 300 bpm 260 bpm

Contrast agent Fenestra VC Fenestra VC Fenestra VC Fenestra VC BinitioEnhancement 500 HU 360 HU 540 HU 110 HU 450 HUImage noise 50 HU 36 HU 95 HU 175HU / 30HU 125HU / 26 HU

Quality parameter 21% 27% 4% 15% / 87% 21% / 100%

Table 1. Parameters for two different scan and reconstruction approaches in comparison. Note that reference3 does sampleonly 50% of the cardiac cycle (six 10 ms wide samples within a 120 ms cardiac cylce). Hence the actual dose reported as920 mGy is only half of the dose that would be required to sample the complete cycle. Respiratory and heart rates arethe averages reported in each reference and correspond to mouse scans. The spatial sampling corresponds to the detectorpixel size after binning, if applicable, at the isocenter.

for the LDPC reconstructions of mouse 2. The other quality value stated for our experiment corresponds tothe conventional phase–correlated reconstruction and therefore is far lower than 100%. It must be emphasizedthat the quality parameter is nothing but a rough estimate due to the following reasons. First of all, spatialresolution was assumed to be proportional to the sampling distance and the effect of the reconstruction process,that potentially lowers the spatial resolution, could not be taken into account. In addition the dose valuesreported are measured using different methods and different phantoms and therefore are not highly accurate.And finally the image noise quoted is neither measured in difference images, and therefore is likely to containeffects such as streak artifacts or non–uniform background, nor was it measured in the same region of the mousein all cases. Nevertheless, there are two things to be learned regarding the evaluation of the quality parameter:Without advanced image reconstruction and signal processing techniques, our results lie well within the range ofwhat has been published in the literature so far. And switching from standard phase–correlated reconstructionto low–dose phase–correlated reconstruction has the potential to increase the image quality by a factor of aboutfive. A decrease in dose was simulated by limiting the number of rotations and thus the number of projectionsavailable for reconstruction. The image acquisition was conducted at a dose of about 250 mGy

5. CONCLUSION AND DISCUSSION

LDPC is a new technique to significantly improve phase–correlated imaging from highly undersampled data. Wedemonstrate that dose savings of about an order of magnitude are possible compared to standard phase–correlatedreconstruction approaches. With sophisticated reconstruction techniques such as the LDPC algorithm it ispossible to perform 4D or 5D phase–correlated imaging at about the same dose level as required for conventional3D studies. Furthermore the method is not limited to a certain type of scanner or geometry as the images shownthroughout this paper originate from different imaging systems of different vendors. Using LDPC thereforeappears to allow for longitudinal in–vivo studies of the rodent heart as well as for long term studies in preclinicalresearch.

Acknowledgment

This work was supported by the Deutsche Forschungsgemeinschaft (DFG) under grant FOR 661. Our fastmulti–core server platforms were provided by the Intel Corporation. The reconstruction software RayConStruct-IR was provided by RayConStruct GmbH, Nurnberg, Germany. We thank Dr. Laura Konerth and SandraStrobelt, Lehrstuhl fur Pharmakologie und Toxikologie, University of Erlangen–Nurnberg, for carrying out themeasurements. We further would like to thank Binitio Biomedical Inc. for providing their latest contrast agent.

REFERENCES

[1] Ford, Nancy, L., Nikolov, H. N., Norley, C. J., Thornton, M. M., Foster, P. J., Drangova, M., and Holdsworth,D. W., “Prospective respiratory–gated micro–CT of free breathing rodents,” Med. Phys. 32, 2888–2898(Sept. 2005).

[2] Badea, C., Hedlund, L., and Johnson, G., “Micro–CT with respiratory and cardiac gating,” Med. Phys. 31,3324–3329 (Dec. 2004).

[3] Badea, C., Fubara, B., Hedlund, L., and Johnson, G., “4D micro–CT of the mouse heart,” MolecularImaging 4, 110–116 (Apr./June 2005).

[4] Drangova, M., Ford, N. L., Detombe, S. A., Wheatley, A. R., and Holdsworth, D. W., “Fast retrospectivelygated quantitative four–dimensional (4D) cardiac micro computed tomography imaging of free–breathingmice,” Investigative Radiology 42, 85–94 (Feb. 2007).

[5] Bartling, S. H., Stiller, W., Grasruck, M., Schmidt, B., Peschke, P., Semmler, W., and Kiessling, F.,“Retrospective motion gating in small animal CT of mice and rats,” Investigative Radiology 42, 704–714(oct 2007).

[6] Dinkel, J., Bartling, S. H., Kuntz, J., Grasruck, M., Kopp-Schneider, A., Iwasaki, M., Dimmeler, S., Gupta,R., Semmler, W., Kauczor, H.-U., and Kiessling, F., “Intrinsic gating for small–animal computed tomogra-phy: A robust ECG–less paradigm for deriving cardiac phase information and functional imaging,” Circu-lation: Cardiovascular Imaging 1(3), 235–243 (2008).

[7] Zijp, L., Sonke, J.-J., and van Herk, M., “Extraction of the respiratory signal from sequential thorax cone–beam x–ray images,” Proc. of the 14th ICCR, Seoul, Korea , 507–509 (2004).

[8] Sonke, J.-J., Zijp, L., Remeijer, P., and van Herk, M., “Respiratory correlated cone beam CT,” Med.Phys. 32, 1176–1186 (Apr. 2005).

[9] Bartling, S. H., Dinkel, J., Stiller, W., Grasruck, M., Madisch, I., Kauczor, H.-U., Semmler, W., Gupta,R., and Kiessling, F., “Intrinsic respiratory gating in small–animal CT,” European Radiology 18, 1375–1384(July 2008).

[10] Feldkamp, L., Davis, L., and Kress, J., “Practical cone–beam algorithm,” Journal of the Optical Society ofAmerica 1, 612–619 (June 1984).

[11] McKinnon, G. C. and Bates, R., “Towards imaging the beating heart usefully with a conventional CTscanner,” IEEE Transactions on Biomedical Engineering BME-28, 123–127 (Feb. 1981).

[12] Garden, K. L. and Robb, R. A., “3-D reconstruction of the heart from few projections: A practical im-plementation of the McKinnon–Bates algorithm,” IEEE Transactions on Medical Imaging MI-5, 233–239(Dec. 1986).

[13] Tomasi, C. and Manduchi, R., “Bilateral filtering for gray and color images,” Proc. 6th Int. Conf. ComputerVision , 839–846 (1998).