calibration method for mmw imaging using inexpensive miniature neon indicator lamp detectors
TRANSCRIPT
Sensors-8495-2013.R1 Accepted 1
Abstract-We have developed focal plan array (FPA) imagers based
on very inexpensive miniature neon indicator lamp detectors called
Glow Discharged Detectors (GDD's). The novelty of this paper is a
technique of calibration which permits reliable imaging with such
an inexpensive FPA. GDDs are commercial lamps which lack
uniformity as detectors. The technique developed in this study
cancels out differences in response between the FPA lamps caused
by both non-uniformity of the MMW collimated beam, and non-
uniformity of the GDDs and other components in the electronic
circuit. Utilizing this method in the new imaging system unifies
detector response. Results, shown with two types of sensor
configurations ( and scanner array), indeed show
better performances and improvement in visibility of details.
Index Terms— Millimeter waves, Imaging, Calibration
technique, Focal plane array, Systems embedded
I. INTRODUCTION
MMW/THz imaging systems in the electromagnetic spectrum
between and are required for applications in
homeland security, communications, medicine, and space
technology mainly because there is no known ionization
hazard for biological tissue, and atmospheric attenuation of
MMW/THz radiation at 100 GHz frequency relatively low for
practical imaging distances of even several kilometers (less
The paper was submitted for reviw in 23-Jun-2013. The authors are
grateful for the support of the Office of Naval Research and the US Army Night Vision and Electronic Sensors Directorate.
A. L. Author is a Ph.D. candidate in Electrical and Computer Engineering
Department, Ben-Gurion University of the Negev, Beer-sheva. (e-mail: [email protected])
N. S. Kopeika Author is a professor in Electrical and Computer
Engineering Department, Ben-Gurion University of the Negev, Beer-Sheva, Israel, (e-mail: [email protected])
Y. Y. Author is a Ph.D. Researcher and a member of Staff with the
Electro-optic Engineering Department, Ben-Gurion University of the Negev, Beer-Sheva, Israel. (e-mail: [email protected] )
D. R. Author is a Ph.D. candidate in the Electro-Optical Engineering
Department, Ben-Gurion University of the Negev, (e-mail: [email protected])
A. A. Author is a Ph.D. Researcher and a member of Staff with the
Department of Electrical and Electronic Engineering, Ariel University Center of Samaria, Ariel, Israel (e-mail: [email protected])
than 1dB/km) except for heavy rain [1], let alone tens of
meters that are needed for many security applications. New
techniques of distinguishing between objects and absorbing
background in order to recognize concealed objects through an
interfering media are very popular.
We use the known and simple fundamental phenomena of
glow discharge plasma indicator lamps as THz detectors. The
lack of inexpensive room temperature high speed detectors
makes focal plane arrays (FPAs) very expensive.
Consequently, scanning is often employed instead of real time
FPAs.
Room temperature detectors such as pyroelectric, Golay
cell, and bolometer devices, feature relatively long response
times on the order of milliseconds or more which makes real
time imaging difficult. Direct detection GDD NEP at
is about ⁄ as measured using
modulation frequency. [2, 10]. For lower modulation
frequencies such as it is closer to
⁄ . The difference stems from circuit electronics.
Microbolometers coupled with antennas are much more
sensitive. However, GDDs are orders of magnitude both faster
and cheaper, with response time on the order of a microsecond
[3] as limited by circuit electronics. Recently, using faster
electronics, this has been reduced to less than 100 ns. Speed is
important for video frame rates, especially for imaging
moving targets. Cost is important for fast focal plane arrays
and possible wide spread implementation in crowded areas.
Ability of plasma to detect electromagnetic (EM) radiation
has been studied since the 1950's at least. Investigations
involved rare gas and rare gas mixtures enclosed in a glass
envelope containing at least two electrodes. The plasma is
formed by breaking down the gas via a DC voltage applied
between two electrodes. The plasma is then illuminated with
EM radiation of frequency greater than the plasma frequency
so that the incident radiation can be absorbed. At least two
mechanisms of detection have been identified, enhanced
ionization collision rate [4-9] and enhanced diffusion [5, 8, 10-
12]. The former increases the bias current, while the latter
decreases it. Change in current is proportional to incident EM
Calibration Method for MMW Imaging using
Inexpensive Miniature Neon Indicator Lamp
Detectors
Assaf Levanon1, Natan S. Kopeika
1,2, Yitzhak Yitzhaky
2, Daniel Rozban
2, Amir Abramovich
3
1Department of Electrical and Computer Engineering, Ben Gurion University of the Negev, Beer Sheva, Israel
2Department of Electro-Optical Engineering, Ben Gurion University of the Negev, Beer Sheva, Israel
3Department of Electrical and Electronic Engineering, Ariel University, Israel
This is the author's version of an article that has been published in this journal. Changes were made to this version by the publisher prior to publication.The final version of record is available at http://dx.doi.org/10.1109/JSEN.2014.2301933
Copyright (c) 2014 IEEE. Personal use is permitted. For any other purposes, permission must be obtained from the IEEE by emailing [email protected].
Sensors-8495-2013.R1 Accepted 2
wave power. The dominant detection mechanism in such
devices was found to be the enhanced ionization leading to
increase of lamp current [8-12].
GDD plasma properties have been studied in three-
electrode lamps in which the third electrode is used as a
Langmuir probe, and such internal avalanche amplification
has been calculated to be on the order of six orders of
magnitude [9]. Response by such devices is linear with
incident electromagnetic wave power [9]. In order to separate
the detected signal from the large dc bias, it is desirable to
modulate the intensity of the radiation, in which case the GDD
acts as an envelope detector.
Imaging experiments with a single GDD in a scanning
system indicate essentially no difference in image quality from
that obtained using a Schottky diode detector instead [13].
Indeed, GDDs have been shown to have potentially higher
responsivity and greater speed than Schottky diodes [14].
However, success in developing less expensive real-time
millimeter wave and terahertz imaging systems depends
largely on the ability to develop inexpensive focal plane
arrays, especially using inexpensive detectors such as GDDs.
Indeed, small GDD arrays such as and scanner
have been developed and tested and found to yield very good
image quality despite the large size and dimensions of the
lamps [15-19]. Recent results include using oversampling sub-
pixel shifting to obtain high resolution images [17-19] on the
order of 0.9 milliradians. If diameter lamps are used,
0.5 milliradian resolution should be achievable. The
oversampling increases signal-to-noise ratio and allows
thresholding at the noise level. This removes larger angle
diffraction blur from the image since such blur is buried in the
noise. The blur size reduction improves resolution beyond the
diffraction limit. Such performance in real time requires
simultaneous exposure of all detectors to the incident image.
The oversampling would require large numbers of detectors.
Hence, low cost of detectors becomes a significant issue.
The detector operation, architecture, and direct detection
imaging have been described previously. It was shown that it
is possible to detect metallic materials, and interpolate over
the image with the assistance of DSP algorithms. A quasi
optical system for the detector FPA was also designed, and
various kinds of imaging were demonstrated [16-19]. Lab
VIEW controlling software permits graphical algorithm
display and image processing. Those capabilities permit all
post processing in real time. The GDDs used here are in
diameter. There also exist commercial GDDs of
diameter. Both types of lamps exhibit similar sensitivity, as
quantified above. The lamps have shown excellent imaging
results when used as image scanning single detectors [13, 14]..
II. IMAGE SETUP
Two separate systems were designed, based on two kind of
FPA which utilize miniature neon lamp GDD devices as
detectors with a small parabolic mirror behind each one to
focus the incoming radiation back into the sensor.
Fig.1 depicts the experimental MMW imaging system
featuring the FPA size of . Each line
turns on for reading operation. The software system images all
8 rows to obtain a 64 pixel image. The source is
based on GaAs multipliers of Virginia Diodes, Inc.
(Charlottesville, VA, USA) that multiply a low frequency
source of . Fig.2 shows the second
detector system, the scanner detector array. The new
scanner contains 18 GDDs in 2 columns, with a horizontal
movement motor. The two off-axis columns of GDDs are
displaced horizontally to produce interlaced images.
Compared to the FPA, the scanner detector array
has increased field of view (FOV) in both directions. The
image size could reach lengthwise (vertically) and
widthwise (horizontally) which is dramatically larger
compared to the previous FPA size of . This is made possible by the horizontal scan motion
extent [width] and the relatively long vertical scanner
dimension [length]. A user-defined scan step size determines
the number of pixels within field-of-view. Each vertically
scanned location causes the lamps rows to be turned on for
reading operation. The imaging system includes a MMW
source of based on GaAs multipliers from Virginia
Diodes, Inc. (Charlottesville, VA, USA).
Both systems sources are amplitude modulated sinusoidal at
. The transmitted signal was coupled to free space by
a rectangular horn antenna, producing an approximately
fundamental mode Gaussian beam with waist on the order of
[20]. Scene illumination is therefore non-uniform in
intensity. The imaging quasi optical set-ups in both system are
the same, and include an off- axis collimating parabolic mirror
900 (OPM) of focal length with aperture of .
An on-axis spherical mirror of focal length with
aperture of diameter was used to generate the image.
A diagonal mirror was used in order to lay objects in
horizontal positions on a table. Imaging is therefore performed
without the need to move the optical setup for different
objects. The system components are described in Figs.1, 2.
The quasi optic systems were designed to image according to
geometrical optics principles. ZEMAX optical development
software was used to optimize the system for minimum
geometrical aberrations. Simulation of an “F” shape object
was employed in order to predict diffraction affects.
The dimensions obey the Gaussian imaging equation , where and are object and image distances,
and is focal length [ ].
Fig.1: Setup design of the MMW FPA system. The MMW source [lower
left] transmits the beam to the target [lower right]. A parabolic imaging
mirror [upper right] reflects the returned radiation from the target onto
the FPA [upper right].
This is the author's version of an article that has been published in this journal. Changes were made to this version by the publisher prior to publication.The final version of record is available at http://dx.doi.org/10.1109/JSEN.2014.2301933
Copyright (c) 2014 IEEE. Personal use is permitted. For any other purposes, permission must be obtained from the IEEE by emailing [email protected].
Sensors-8495-2013.R1 Accepted 3
Fig.2: Setup design of the MMW scanner detector array system. The
MMW source [lower left] transmits the beam to the target [lower right].
A parabolic imaging mirror [upper right] reflects the returned radiation
from the target onto the scanner detector array [upper-right].
In system 1, the set-up was designed to perform
to ) optical magnification of the object,
which was located from the imaging spherical
mirror . The GDD array was placed in the image plane
which was from the imaging spherical mirror .
In system 2 objects were placed distance from
imaging spherical mirror , so that the image plane appears
from there , yielding approximately unity
optical magnification.
III. CALIBRATION METHODS
It is worthwhile to develop methods for resolution and
identification improvement of details in the image. Such
operations open the possibility to perform successful imaging
by inexpensive commercial lamp FPAs as sensors
Developing a calibration technique in order to cancel out
differences in signal strength between FPA lamp signals is of
utmost importance. Differences and inconsistencies in the
image formation process need to be canceled. Lack of
uniformity is caused by non-uniformity of the collimated
beam, inconsistency in electrical components at the electrical
circuit, and GDD inconsistent power detection.
These characteristics are not unexpected, since GDD devices
are manufactured as inexpensive indicator lamps rather than
MMW radiation detectors.
In order to limit the non-uniformity across the GDD array, a
look-up-table (LUT) was created to yield effectively the same
response for each lamp. This LUT formation is based on the
linear response of each lamp when incident radiation is
detected [6]. The method procedure algorithm was
implemented in both and FPA systems through
novel system-based Lab-VIEW software (see Fig.3). This
system prevents the differences between lamps automatically,
by calculating the transform LUT. Fig.3 demonstrates the
graphical user interface (GUI) of the novel calibration system
of both FPA systems. The principle was similar, but each
system needs to implement the method in a different way
before data is transformed to the display unit.
Fig.3: are coordinates with respect to the source power
(source intensity levels), n and m are the line equation parameters
adjusted for each lamp sensor; (a) A calibration graphical interface of
FPA. (b) Calibration graphical interface of scanner detector
array
The procedure of the calibration is as follows:
In order to produce the LUT, the coordinates with respect to
source power of the system (intensity levels) and are
selected (see Fig. 4 – horizontal axis). These coordinates
reference to radiation illumination intensity irradiated from the
source. An MMW image is taken with intensity and another
with intensity. Each lamp (pixel) produces its own line
equation (where m and n are the slope and the
-intercept parameters of the line) using the volt root mean
square (VRMS) outputs and with respect to the intensity
of the GDD (Fig. 4 – vertical axis) in each detector test image.
These line equations are the function of intensity transform,
and parameters and are adjusted for each lamp so as to
yield a uniform response. New intensity value (new in
Fig.4) will transform to the appropriate by using the known
n and m, and this presents the calibrated value for
Pixel . Consequently, by choosing different gaps
between the values of and in the test image, we can
control the intensity resolution in the display unit.
As we present in the results, both systems perform such
calibration, so that the same absorbed radiation intensity will
effectively produce the same response for each lamp.
Fig 4: Calibration technique description: X’s are the coordinates which
refer to radiation illumination intensity irradiated from the source. Y’s
are the real VRMS intensity results for the detected sinusoidal modulated
wave.
This is the author's version of an article that has been published in this journal. Changes were made to this version by the publisher prior to publication.The final version of record is available at http://dx.doi.org/10.1109/JSEN.2014.2301933
Copyright (c) 2014 IEEE. Personal use is permitted. For any other purposes, permission must be obtained from the IEEE by emailing [email protected].
Sensors-8495-2013.R1 Accepted 4
The quality of the calibrated MMW images was improved
after the implementation of this technique and results are
presented in Figs.6-8. Each imaging result is presented over
the Lab View Graphical Interface User (GUI) and stored for
post-processing procedures. The resulting MMW images here
are presented in a spatially interpolated contour-view.
A contour line of a function of two variables is a curve along
which the function has a constant value (as show in
Fig.5). The space between each pair of adjacent elevation
lines is filled with respectively constant mapped gray level.
This technique produces a non-pixelated interpolated version
of the image, with a larger number of pixels (see MMW
photos of Figs.6-8). Here, the images are created with 8 bits
quantization, which means 256 gray levels.
Fig.5: Description of same-elevation (contour) lines.
Figs. 6 (b, c) and 7 (b, c) are product of an pixel
resolution while Figs. 8 (b, c) and Figs.8 (e, f) were created
with and pixels resolution, respectively.
Fig. 6(a) describes a resolution chart with
3 rectangular holes of size . As seen, in Fig.
6 (c) the resolution chart rectangular holes are much more
recognizable than in Fig.6 (b) where the holes almost cannot
be seen. Fig.7(c) also shows better results in comparison to
Fig.7 (b) after the lamps were calibrated over the image plane.
As mentioned above, the aim of this technique is to obtain a
reliable space-invariant linear-intensity response image.
Results indicate that it is indeed possible.
Fig 6: (a) A metal of resolution chart object, (b) 2D
contour-based interpolation MMW image before applying calibration, (c)
same image after applying calibration.
Fig 7: (a) A photograph of a metallic object with circular hole in plate, (b)
2D contour-based interpolation MMW image result before applying
calibration, (c) same image after applying calibration.
Fig.8 describes results with FPA scanner array
detector. As can be seen, the figures transformed by the
calibration technique are much better in terms of uniformity of
intensity.
In order to recombine the two off-axis column detectors of the
scanner and obtain their true spatial location, the
default step size is . Fig.8 (b, c) are made with this
default scan step in order for the interlace column to include
all 36 lamps. Images with that scan setup include
pixel resolution and are of size.
Fig.8 (b) includes high intensity in the middle of the image,
which disappeared by use of the calibration technique (as seen
in Fig.8 (c))
Figs.8 (e, f) are produced by only one column of 18 lamps, in
scan steps of 1mm. It was made in order to obtain only
pixel resolution, with image size of . The over sampling technique is used together with
the calibration technique to obtain excellent results. Fig.8 (e)
presents illumination with the un-collimated beam (higher
intensity in the middle of the image plane).Moreover there are
some GDD lamps with noticeable differences in responsivity.
The outputs of both imaging structures seem to be improved
with the new calibration technique.
Figure 8: Results with FPA scanner array detector (a) A
photographs of a metallic “F” object, (b) 2D contour-based interpolation
MMW image result before applying calibration, (c) the same image after
applying the calibration process, (d) an object with circular holes in a
metal plate, (e) 2D contour-based interpolation MMW image result
before applying calibration, (f) image after applying calibration.
IV. DISCUSSION AND CONCLUSIONS
We succeeded in obtaining MMW images of metallic and
dialectical objects at , with two different systems.
One is an FPA, and the other is a scanning FPA
imager. Both imagers use very low cost commercial GDD
lamps as detectors. Indicator lamps as detectors can greatly
reduce the cost of millimeter wave imaging systems, while
obtaining both speed and good image quality. However, non-
uniform illumination and non-uniform GDD response limit
image quality noticeably.
This research shows that methods such as calibration solve
both problems and thereby improve the image, permitting
This is the author's version of an article that has been published in this journal. Changes were made to this version by the publisher prior to publication.The final version of record is available at http://dx.doi.org/10.1109/JSEN.2014.2301933
Copyright (c) 2014 IEEE. Personal use is permitted. For any other purposes, permission must be obtained from the IEEE by emailing [email protected].
Sensors-8495-2013.R1 Accepted 5
imaging with much better capability of identification and
accuracy especially when combined with oversampling...
In conclusion, the low cost of GDD FPAs can permit real
time relatively high image quality by illuminating all detectors
simultaneously, and using oversampling and calibration as
shown here.
ACKNOWLEDGMENT
The authors are grateful for the support of the Office of
Naval Research and the US Army Night Vision and Electronic
Sensors Directorate. We also appreciate the support of the
Institute for Future defense Technologies research named for
the Medvedi, Schwartzman, and Gensler Families.
REFERENCES
[1] A.H. Lettington, I.M. Blankson, M. Attia, and D. Dunn, “Review of imaging architecture”, Proc. SPIE, 4719, p. 327-340 (2002).
[2] A. Rogalski and F. Sizov, "Terahertz detectors and focal plane arrays",
Opto-Electron. Rev. 19, pp. 346-404, 2011. [3] A. Abramovich, N.S Kopeika, D. Rozban , E. Farber, "Inexpensive
detector for terahertz imaging", Applied Optics, 46, No 29, pp. 7207-
7211, October, 2007. [4] M.A. Lampert and A.D. White, "Microwave techniques for studying
discharges in gases", Electron. Commun., 30, pp. 124-128, June, 1953.
[5] B. J. Udelson, "Effect of microwave signals incident upon different regions of a dc hydrogen glow discharge",. J Appl. Phys., 28, pp. 380-
381, March, 1959.
[6] P.J.W. Severin and A.G. Van Nie, "A simple and rugged wide-band gas discharge detector for millimeter waves", IEEE Trans. Microwave
Theory Tech., MTT-14, pp. 431-436, Sept, 1966.
[7] N. H. Farhat, "Optimization of millimeter wave glow discharge detectors", Proc. IEEE, 62, pp. 279-281, Feb, 1974.
[8] N. S. Kopeika, "On the mechanism of glow discharge detection of
microwave and millimeter wave radiation", Proc. IEEE, 63, pp. 981-982, June, 1975.
[9] N. S. Kopeika, "Glow discharge detection of long wavelength
electromagnetic radiation: Cascade ionization process internal signal gain and temporal and spectral response properties, IEEE trans. on
Plasma Science, PS-6, pp.139-157, June 1978.
[10] G. D. Lobov, "Gas discharge detector of microwave oscillations", Radiotekh Elektron., 5 pp. 152-165, 1960.
[11] D. Rozban, N.S. Kopeika, A. Abramovich, and E. Farber, "Terahertz
detection mechanism of inexpensive sensitive glow discharge detectors", J. Appl. Phys., 103, pp. 093306-1 – 093306-4, May, 2008
[12] A. Abramovich, N.S. Kopeika, D. Rozban, "THz Polarization effects on
detection responsivity of Glow Discharge Detectors (GDD)", IEEE Sensors J., 9, pp. 1181-1184, Oct., 2009.
[13] L. Hou, H. Park, and X.C. Zhang, "Terahertz wave imaging system
based on glow discharge detector", IEEE J. of Selected Topics in Quantum Electron. 17, 177-182, Jan.-Feb., 2011.
[14] L. Hou and W. Shi, "Fast Terahertz Continuous-wave Detector Based on Weakly Ionized Plasma", IEEE Electron Device Letters, vol. 33, pp.
1583-1585, 2012
[15] A. Abramovich, N.S. Kopeika, and D. Rozban, "Design of diffraction
limited focal plane arrays for mm wavelength and terahertz radiation using glow discharge detector pixels", J. Appl. Phys, 104, pp. 033302-1
– 033302-4, Aug, 2008.
[16] D. Rozban, A. Levanon, H. Joseph, A. Akram, A. Abramovich, Y.
Yitzhaky, N.S. Kopeika, A. Belenky, O. Yadid-Pecht, "Inexpensive THz
focal plane array imaging using miniature neon indicator lamps as detectors", IEEE Sensors J., 11, pp. 1962-1968, Sept. 2011.
[17] M. Shilemay, D. Rozban, A. Levanon, Y. Yitzhaky, N.S. Kopeika, O. Yadid-Pecht, A. Abramovich "Performance quantification of a
millimeter-wavelength imaging system based on inexpensive glow-
discharge-detector focal-plane array", Applied Optics, Vol. 52, Issue 7, pp. C43-C49 (2013).
[18] A. Abramovich N. S. Kopeika , D. Rozban , A. Levanon , Moshe Shilemay,; A. Akram , H. Joseph , O. Yadid-Pecht , A. Belenky, "Super
resolution and optical properties of THz double row array based on
inexpensive Glow Discharge Detector (GDD) pixels" Proc. SPIE vol. 8188, millimeter Wave and Terahertz Sensors and Technology IV,
81880A (October 13, 2011); doi:10.1117/12.898024
[19] A. Levanon, N. S. Kopeika, Y. Yitzhaky, A.Abramovich, D. Rozban, H.
Joseph, A. Aharon (Akram)1, A. Belenky, M. Gefen, O. Yadid–Pecht,
"Over-sampling advances in mm-wave scan imaging using inexpensive Ne indicator lamp detectors", Optical Engineering, 52 (6), 063202
(June 24, 2013); doi: 10.1117/1.OE.52.6.063202
[20] A. Murphy and R. Padman, "Phase centers of horn antennas using
Gaussian mode analysis", IEEE Tran. Antennas Propagation, Vol. 38,
No 8, pp 1306-1310, Aug., (1990).
First A. Author, Assaf Levanon was
born in Israel in 1975. He received the
M.Sc. degree in Electrical and
Computer Engineering from Ben-
Gurion University of the Negev, Beer-
sheva, Israel, in 2011, where he is
currently pursuing a Ph.D. His current
research interests include millimeter
and sub-millimeter wave systems, expertise in image
processing algorithms for enhancing THz images, new
detection methods, beyond limit of diffraction resolutions
issues, and hardware design configuration for Ne lamp FPAs
for mm wave security, military, and civilians applications
Second B. Author, Natan S. Kopeika
(SM’79–LSM’10) was born in
Baltimore, MD, in 1944. He received
the B.Sc., M.Sc., and Ph.D. degrees in
Electrical Engineering from the
University of Pennsylvania,
Philadelphia, in 1966, 1968, and 1972,
respectively. He joined the Ben-Gurion
University of the Negev, Beer-Sheva, Israel, in 1973, where he
was the Chair of the Department of Electrical and Computer
Engineering for two terms from 1989 to 1993, Reuven and
Francis Feinberg Professor of Electrooptics since 1994, and
the first Chairman of the new Department of Electro-Optical
Engineering. He has authored or co-authored over 180 papers
in internationally reviewed journals and well over 100 papers
at various conferences. He has authored the textbook entitled
A System Engineering Approach to Imaging (SPIE Press,
1998) [first printing 1998, second printing 2000]. He is the
Topical Editor of Marcel Dekker for the topic “Atmospheric
Optics” in their Encyclopedia of Optical Engineering. He has
co-authored a book entitled Applied Aspects of Optical
communication and LIDAR (CRC Press, 2010). His current
research interests include interactions of electromagnetic
waves with plasmas, optogalvanic effects, environmental
effects on optoelectronic devices, imaging system theory,
propagation of light through atmosphere, imaging through
This is the author's version of an article that has been published in this journal. Changes were made to this version by the publisher prior to publication.The final version of record is available at http://dx.doi.org/10.1109/JSEN.2014.2301933
Copyright (c) 2014 IEEE. Personal use is permitted. For any other purposes, permission must be obtained from the IEEE by emailing [email protected].
Sensors-8495-2013.R1 Accepted 6
atmosphere, image processing and restoration from blur,
imaging in presence of motion and vibration, light detection
and ranging (LIDAR), target acquisition, and image quality in
general. Kopeika was a recipient of the J. J. Thomson Prize
from the IEE in 1999. He is a fellow of SPIE.
Third C. Author, Yitzhak Yitzhaky
received his BS, MS, and PhD degrees in
electrical and computer engineering from
Ben Gurion University, Israel, in 1993,
1995, and 2000, respectively. From 2000
to 2002, he was a postdoctoral research
fellow at the Schepens Eye Research
Institute, Harvard Medical School,
Boston, Massachusetts. He is currently with the Electro-Optics
Unit at Ben Gurion University. His research is mainly in the
fields of image restoration and enhancement, image analysis,
and imaging systems.
Forth C. Author, Daniel Rozban
received the B.Sc. degree in Electronic
engineering from Ariel University of
Samaria, Ariel, Israel, in 2006, and the
M.Sc. degree in Electro-Optical
Engineering from Ben-Gurion University
of the Negev, Beer-Sheva, Israel, in 2009,
where he is currently pursuing the Ph.D.
degree. He is a Researcher with the
Submillimeter Wave Laboratory, Department of Electrical and
Electronic Engineering, Ariel University of Samaria.
Fifth E. Author, Amir Abramovich
received the B.Sc. and M.Sc. degrees in
Electrical and Computer Engineering
from Ben-Gurion University of the
Negev, Beer-Sheva, Israel, in 1989 and
1991, respectively, and the Ph.D. degree
from Tel-Aviv University, Tel-Aviv,
Israel, in 2001. He was with the Counter
Measures and Spectroscopic
Characterization Department, TAAS Israel Industries. He is a
Researcher and a member of Staff with the Department of
Electrical and Electronic Engineering, Ariel University Center
of Samaria, Ariel, Israel, where he heads the Millimeter and
Submillimeter Wave Laboratory. Dr. Abramovich was a
recipient of research grants to develop and construct THz
imaging systems for homeland security purposes and THz
spectroscopy for material recognition and identification.
This is the author's version of an article that has been published in this journal. Changes were made to this version by the publisher prior to publication.The final version of record is available at http://dx.doi.org/10.1109/JSEN.2014.2301933
Copyright (c) 2014 IEEE. Personal use is permitted. For any other purposes, permission must be obtained from the IEEE by emailing [email protected].