discordance in scar detection between electroanatomical
TRANSCRIPT
J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y VO L . 6 , N O . 1 1 , 2 0 2 0
ª 2 0 2 0 B Y T H E A M E R I C A N C O L L E G E O F C A R D I O L O G Y F O U N D A T I O N
P U B L I S H E D B Y E L S E V I E R
Discordance in Scar Detection BetweenElectroanatomical Mapping and CardiacMRI in an Infarct Swine Model
Selcuk Kucukseymen, MD,a Hagai Yavin, MD,b Michael Barkagan, MD,a Jihye Jang, PHD,aAyelet Shapira-Daniels, MD,a Jennifer Rodriguez, BA,a David Shim, MD,a Farhad Pashakhanloo, PHD,a
Patrick Pierce, RT,a Lior Botzer,d Warren J. Manning, MD,a,c Elad Anter, MD,b Reza Nezafat, PHDa
ABSTRACT
ISS
Fro
Bo
OhdB
01
su
He
Sci
sys
dis
Th
ins
vis
Ma
OBJECTIVES This study sought to investigate the sensitivity of electroanatomical mapping (EAM) to detect scar as
identified by late gadolinium enhancement (LGE) cardiac magnetic resonance (CMR).
BACKGROUND Previous studies have shown correlation between low voltage electrogram amplitude and myocardial
scar. However, voltage amplitude is influenced by the distance between the scar and the mapping surface and its extent.
The aim of this study is to examine the reliability of low voltage EAM as a surrogate for myocardial scar using LGE-derived
scar as the reference.
METHODS Twelve swine underwent anterior wall infarction by occlusion of the left anterior descending artery (LAD)
(n ¼ 6) or inferior wall infarction by occlusion of the left circumflex artery (LCx) (n ¼ 6). Subsequently, animals un-
derwent CMR and EAM using a multielectrode mapping catheter. CMR characteristics, including wall thickness, LGE
location and extent, and EAM maps, were independently analyzed, and concordance between voltage maps and CMR
characteristics was assessed.
RESULTS LGE volume was similar between the LCx and LAD groups (8.5 � 2.2 ml vs. 8.3 � 2.5 ml, respectively;
p ¼ 0.852). LGE scarring in the LAD group was more subendocardial, affected a larger surface area, and resulted in
significant wall thinning (4.88 � 0.43 mm). LGE scarring in the LCx group extended from the endocardium to the
epicardium with minimal reduction in wall thickness (scarred: 5.4 � 0.67 mm vs. remote: 6.75 � 0.38 mm). In all the
animals in the LAD group, areas of low voltage corresponded with LGE and wall thinning, whereas only 2 of 6 animals in
the LCx group had low voltage areas on EAM. Bipolar and unipolar voltage amplitudes were higher in thick inferior walls
in the LCx group than in thin anterior walls in the LAD group, despite a similar LGE volume.
CONCLUSIONS Discordances between LGE-detected scar areas and low voltage areas by EAM highlighted the limi-
tations of the current EAM system to detect scar in thick myocardial wall regions. (J Am Coll Cardiol EP 2020;6:1452–64)
© 2020 by the American College of Cardiology Foundation.
N 2405-500X/$36.00 https://doi.org/10.1016/j.jacep.2020.08.033
m the aDepartment of Medicine (Cardiovascular Division), Beth Israel Deaconess Medical Center and Harvard Medical School,
ston, Massachusetts; bCardiac Electrophysiology Section, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland,
io; dDepartment of Radiology, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts; and
iosense Webster, Yokneam, Israel. This study was supported by the National Institute of Health (1R01HL129185, 1R01HL129185-
, 1R01HL129157, and 1R01HL127015) and the American Heart Association (15EIA22710040) to Dr. Nezafat. Dr. Pashakhanloo was
pported by the American Heart Association (18POST33990040). Dr. Shapira-Daniels was supported by the National Institute of
alth T32 training grant. Dr. Anter has received research grants from Biosense Webster; and has received honoraria from Boston
entific. Dr. Nezafat has a research agreement with Siemens Healthineers, the manufacturer of the cardiac magnetic resonance
tem used in this study. All other authors have reported that they have no relationships relevant to the contents of this paper to
close.
e authors attest they are in compliance with human studies committees and animal welfare regulations of the authors’
titutions and Food and Drug Administration guidelines, including patient consent where appropriate. For more information,
it the JACC: Clinical Electrophysiology author instructions page.
nuscript received June 23, 2020; revised manuscript received July 29, 2020, accepted August 11, 2020.
AB BR E V I A T I O N S
AND ACRONYM S
CMR = cardiac magnetic
resonance
CI = confidence interval
EAM = electroanatomical
mapping
EGM = electrogram
LAD = left anterior descending
LCx = left circumflex
LGE = late gadolinium
enhancement
LV = left ventricle
MI = myocardial infarction
RV = right ventricle
VT = ventricular tachycardia
J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0 Kucukseymen et al.O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4 Limitation of Voltage Mapping to Detect Scar
1453
V entricular tachycardia (VT) activation map-ping is the gold standard for mapping scar-related re-entry circuits, yet it is rarely
achieved due to long mapping times and/or hemody-namic intolerance (1). Substrate mapping during sinusrhythm has evolved as an alternative to activationand/or entrainment mapping to manage hemody-namically unstable VTs (2) by reducing the need formapping during prolonged periods of VT. However,substrate mapping techniques during sinus rhythmare often associated with substantial recurrence rates(3,4). Higher rates are partially due to assumptionsand misconceptions related to substrate mappingtechniques, such as the assumption that low voltageequals scar or the misconception that mapping ofthe ventricular surface provides accurate informationof the much larger unmapped intramural volume(5,6).
The relationship between low voltage amplitudeand the VT isthmus was initially established inarrhythmia surgeries of patients with large anteriorwall infarctions and aneurysms (1). Bipolar voltageamplitude <1.5 mV recorded with a quadripolar non-steerable woven catheter with a 2- mm tip, 1-mmring, and 5-mm interelectrode spacing was associ-ated with the location of the VT isthmus (7). Thisthreshold was later validated for a standard 4-mm tip,1-mm ring, and 2-mm interelectrode spacing mappingcatheter, and, more recently, for multielectrodemapping catheters in animal studies (8,9). Similar tothe original patients who underwent arrhythmiasurgery, these animal models had anterior infarctionsby left anterior descending artery (LAD) occlusion inwhich the anterior wall underwent remodeling withtransmural scarring, significant wall thinning, andpreservation of a thin endocardial layer (8–13). In-farctions at other territories, particularly at the leftcircumflex artery (LCx), often produced a more het-erogeneous scar without significant wall thinning(14,15). In humans, such infarctions often result inpatchy patterns of bipolar voltage abnormalities,which may suggest overall smaller infarct sizes, lowercollagen content relative to cardiomyocytes, largerareas of preserved sub-endocardium, and relativedifferences in scar-to-wall thickness ratios (16,17).
Cardiac magnetic resonance (CMR) imaging hasbecome an integral evaluation component in patientsundergoing VT ablation (18–20). Late gadoliniumenhancement (LGE) identifies scar regions associatedwith the ventricular substrate. In addition to the scar,wall thickness and strain can be accurately measuredvia cine images to better assess left ventricular (LV)remodeling and anatomical functional effects of thescar (10,21,22). Numerous studies have correlated scar
regions on LGE images to areas of low voltageon substrate mapping (6,9,21,23–28), and it iswell accepted that EAM can detect scarringobserved on LGE. However, voltage mapsonly represent tissue characteristics within afew millimeters of the catheter tip, limitingtheir sensitivity to detect mid-myocardialscar (29). In contrast, LGE is not capable ofidentifying scar within 1 to 2 mm (i.e., 1 to 2pixels) of the blood pool due toblood�myocardium partial voluming (30,31).These limitations have raised concernsregarding the sensitivity of current EAMtechnologies to detect myocardial scar.
We sought to investigate the concordancebetween myocardial scar detection by LGEand voltage amplitude in a swine model of
anterior and inferior healed myocardial infarction.METHODS
EXPERIMENTAL STUDY DESIGN. This study included12 Yorkshire swine of either sex weighing 35 to 40 kg.The animals were divided into 2 groups and under-went infarction in either the LAD (n ¼ 6) or the LCx(n ¼ 6) using balloon occlusion as previouslydescribed (32). Differences between the proceduresincluded 1) balloon occlusion time (180 min for LAD,60 to 120 min for LCx); and 2) targeted coronary ar-teries (occlusion of the LAD was performed after thesecond diagonal branch and occlusion of the LCx wasperformed after the second obtuse marginal branch toreduce mortality and allow a similar territory ofischemia). Animals then lived for 8 to 10 weeks beforeCMR and electrophysiology studies. CMR and elec-trophysiology studies were performed under generalanesthesia with mechanical ventilation. The studywas conducted at Beth Israel Deaconess MedicalCenter and approved by the Institutional Animal Careand Use Committee.
CMR IMAGING. CMR was performed using a 3-T car-diac magnetic resonance scanner (Siemens Vida,Erlangen, Germany) with an 18-channel body receivercoil. Animals underwent in vivo CMR 1 week beforeterminal mapping and ex vivo CMR immediately afterbeing killed. The CMR protocol included evaluation offunction using short- and long-axis slices withbalanced steady-state free-precession imaging(TR¼ 3.2ms; TE¼ 1.4ms; flip angle¼40 degrees; voxelsize ¼ 0.9 � 0.9 � 8.0 mm3; field of view ¼ 360 �270 mm2). Ten to 15 min after administration of0.2 mmol/kg of intravenous gadobutrol (Gadavist,Bayer HealthCare, Berlin, Germany), 3-dimensionalLGE imaging was performed using an inversion-
TABLE 1 Cardiac Magnetic Resonance Imaging Parameters
LV FunctionsLAD
(n ¼ 6)LCx
(n ¼ 6) p Value
EDV, ml 209 � 17 172 � 21 0.009
ESV, ml 116 � 10 91 � 16 0.011
SV, ml 92 � 12 81 � 6 0.074
LVEF, % 44 � 3.7 47 � 3.0 0.160
CO, I/ml 7.9 � 0.9 6.9 � 0.8 0.081
CI, I/ml/m2 3.8 � 0.4 3.6 � 0.3 0.252
HR, min 85 � 10 84 � 5 0.788
Total mass, g 98 � 10 99 � 13 0.967
Tissue tracking (global systolic peak strain)
Radial, % 15 � 4.2 15 � 4.0 0.799
Circumferential, % 10 � 2.2 11 � 2.2 0.821
Longitudinal, % 12.0 � 3.1 12.5 � 2.5 0.783
LGE (in vivo)
Total volume, ml 8.6 � 2.2 8.3 � 2.5 0.852
Total mass, g 9.0 � 2.3 8.7 � 2.6 0.851
LGE burden, % 8.2 � 2 8.9 � 2 0.540
Area of subendocardial LGE bordering LV blood pool, cm2 12.5 � 3.1 4.1 � 1.2 <0.001
LGE (ex-vivo)
Total volume, ml 9.0 � 1.8 8.5 � 2.1 0.711
Total mass, g 9.5 � 1.9 9.1 � 2.2 0.688
LGE burden, % 7.9 � 1.6 8.3 � 1.8 0.802
Wall thickness
Mean wall thickness of remote myocardium, mm 6.6 � 0.4 6.7 � 0.4 0.703
Mean wall thickness of scarred segments, mm 1.8 � 0.3 5.4 � 0.7 <0.001
Wall thinning percentage on scarred segments, % 73 � 40 20 � 10 <0.001
Mean amount of thinning, mm 4.9 � 0.4 1.4 � 0.8 <0.001
Values are mean � SD.
CI ¼ cardiac index; CO ¼ cardiac output; EDV ¼ end-diastolic volume; ESV ¼ end-systolic volume; HR ¼ heartrate; LAD ¼ left anterior descending; LCx ¼ left circumflex; LGE ¼ late gadolinium enhanced; LGE burden ¼ LGEvolume to LV volume ratio; LV ¼ left ventricular; LVEF ¼ left ventricle ejection fraction; SV ¼ stroke volume;
Kucukseymen et al. J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0
Limitation of Voltage Mapping to Detect Scar O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4
1454
recovery sequence with gradient-echo imaging(TR ¼ 3.85 ms; TE ¼ 1.7 ms; flip angle ¼ 18 degrees;voxel size ¼ 1.5 � 1.5 � 1.5 mm3; field of view ¼ 385 �288 � 96 mm3).
For ex vivo CMR imaging, 0.2mmol/kg gadobutrol(Gadavist, Bayer HealthCare) was administeredintravenously 15min before the animals were killedduring the mapping procedure. Imaging commencedwithin 1 h of heart removal using an isotropic 3-dimensional T1-weighted spoiled gradient-echosequence with the following parameters:TR ¼ 27 ms; TE ¼ 4.9 ms; flip angle ¼ 25 degrees;voxel size ¼ 0.4 � 0.4 � 0.5 mm3.
Image analysis was performed using CVI42 appli-cation (version 5.11, Circle Cardiovascular ImagingInc., Calgary, Ontario, Canada). LV and right ven-tricular (RV) volumes were quantified by manuallytracing the end-diastolic and end-systolic endocardialcontours. For global strain measurement, LV endo-cardial and epicardial borders were manually tracedat end-diastole in electrocardiographically gatedbalanced steady-state free precession 4-, 3-, and 2-
chamber long-axis sequences using a point-and-clickapproach. The automatic border tracking algorithmwas applied to track image features throughout thecardiac cycle. Tracking was visually reviewed andmanually corrected. For LGE quantification, endo-cardial and epicardial borders were first manuallycontoured. LGE scar was defined as areas with signal>3 SDs above the healthy remote region. LGE volume,mass, and burden (i.e., percentage of LV scar to LVvolume) were calculated. The surface area of the LGEbordering the LV blood pool was measured by calcu-lating LGE scar length in contact with the LV bloodpool, multiplied by slice thickness (SupplementalFigure 1). The regional wall thickness was measuredby dividing the balanced steady-state free precessioncine short-axis images of the myocardium into 100equally distributed segments using CVI42 (CircleCardiovascular Imaging Inc.) (SupplementalFigure 2). Wall thinning was calculated as the per-centage—thinning rate of the scarred region versusthe healthy remote myocardium (SupplementalFigure 2). All analyses were performed by S.K. (with5 years of experience in CMR); an additional reviewer(F.P. with 10 years of experience) repeated the mea-surements of 4 variables to assess interobserverreproducibility: 1) remote wall thickness; 2) scarredwall thickness; 3) LGE volume; and 4) subendocardialLGE area.
EAM PROTOCOL. All animals underwent EAMfollowing a 10-week survival period and #1 week afterCMR by an experienced electrophysiologist with >15years of experience treating patients with ventriculararrhythmias and 7 years of experience with large an-imal models of arrhythmia. A 6-F pentapolar catheter(Bard EP, Lowell, Massachusetts) was placed in theRV apex for pacing and as an intracardiac activationreference. An intracardiac echocardiography catheterwas positioned in the right atrium or RV and routinelyused for each procedure. EAM was performed usingCarto3 (Biosense Webster Inc, Diamond Bar, Califor-nia) and a multielectrode bidirectional mappingcatheter (Octaray; Biosense Webster Inc,) thatincluding 8 splines and 48 electrodes (26,33). Theperformance of this catheter, including its bipolar andunipolar voltage cutoffs for scar tissue, were recentlyreported. Specifically, the fifth percentile of the bi-polar voltage amplitude recorded in healthy ventri-cles with this catheter was 1.34 mV, which was similarto the cutoff of other catheters (w1.5 mV). Similarly,its mean unipolar voltage amplitude in healthy ven-tricles was relatively similar to that of the PentaRaycatheter (Biosense Webster Inc; 5.2 � 1.9 mV [median:5.2 mV] vs. 5.6 � 2.1 mV [median: 5.5 mV]) (33).
print&web4C=F
PO
FIGURE 1 Tissue Characterization of Post-Infarct Scars Using 3D LGE
Reformatted 3-dimensional (3D) late gadolinium enhancement (LGE) images from 2 animals with (A and B) left anterior descending (LAD) and
(C and D) left circumflex (LCx) occlusion. Arrows show areas of scar in LGE images. Scar in LAD animals was mainly localized in the sub-
endocardial region in areas with significant wall thinning on cine images and in areas with low unipolar and bipolar voltages on electro-
anatomical mapping (EAM) as well, but scar extended from the endocardium to epicardium in LCx animals with preserved wall thickness and
showed normal unipolar and bipolar voltages on EAM.
J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0 Kucukseymen et al.O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4 Limitation of Voltage Mapping to Detect Scar
1455
EAM of the LV was performed via a retrogradeaortic approach during RV pacing at a cycle length of450 to 550 ms using a fill threshold of <5 mm. Thehigh-pass filter was 0.5 Hz for unipolar and 30 Hz forbipolar electrograms (EGMs). The low-pass filter was500 Hz for both. Data acquisition was automated us-ing the following inclusion criteria: 1) QRSmorphology stability defined as $95% morphologystability compared with the paced QRS configuration;2) activation time stability of #5 ms between 2consecutive beats; and 3) maximal distance #3 mmfrom the pre-made anatomical shell. In 2 LAD and 4LCx cases, the epicardium was mapped by accessingthe pericardial space using the modified Sosa tech-nique (34).
EGM analysis was performed off-line at a sweepspeed of 200 to 400 mm/s on either Carto3 or theLabSystem Pro EP recording system (Bard, BostonScientific, Lowell, Massachusetts). The minimum bi-polar voltage amplitude of collected EGMs was30 mV, which is twice the voltage level in our labo-ratories. All EGMs were manually reviewed to excludeectopic beats and artifacts. Activation time was an-notated to the near-field potential as determined bypresence of high-frequency potentials exhibitingspatiotemporal propagation across multiple elec-trodes at a similar acquired beat (1).
Low voltage areas were measured using the stan-dard surface area measurement tool on the Carto3system. The border of the low voltage area on each
print&web4C=F
PO
FIGURE 2 Comparison of Myocardial Wall Thinning
Representative short-axis cine images and extracted left ventricular (LV) and right ventricular (RV) segmentation in diastole from base to apex in both groups show
differences in wall thinning on scar-related walls. Wall thickness is represented in millimeters.
Kucukseymen et al. J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0
Limitation of Voltage Mapping to Detect Scar O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4
1456
map was manually traced using Carto3 area mea-surement software. The low voltage burden wascalculated using the mesh feature on the Carto3software (endocardial low voltage area/total endo-cardial voltage area percentage). Low voltage cutoffswere set at 1.5 mV for bipolar amplitude.
HISTOLOGY DATA. In 1 animal from each group, theheart was harvested and placed in a 10% bufferedformalin solution for >1 week. After tissue fixation, 1heart from each group was serially sectioned parallelto the atrioventricular groove into 1-cm-thick slicesstarting from the apex. Tissue samples were paraffinembedded and sectioned at 5-mm thickness perpen-dicular to the epicardial surface so each sectionshowed the full thickness of the ventricular wall from
the epicardium to the endocardium. Slides werestained with Masson’s trichrome for collagen detec-tion and subsequently scanned and digitized. Thehistology technique was not systematically exam-ined; it was only used to validate the scar features.
STATISTICAL ANALYSIS. Continuous variables withnormal distribution are expressed as mean � SD.Categorical variables are expressed as percentages.The independent sample Student’s t-test and Pear-son’s chi-square test were used to compare normallydistributed continuous and dichotomous variables,respectively. For multiple group comparisons, 1-wayanalysis of variance was used with post hoc Bonfer-roni’s correction. The intraclass correlation coeffi-cient was used to assess intra- and interobserver
TABLE 2 EAM Bipolar Voltage Map Parameters Collected During RV Pacing
Scar-related LV Wall* LAD (n ¼ 6) LCx (n ¼ 6) p Value
Density of EGMs 1,533 � 543 618 � 192 0.010
Mean unipolar voltage, mV 3.8 � 0.9 9.3 � 2.1 <0.001
Mean bipolar voltage, mV 1.8 � 0.3 2.8 � 0.2 <0.001
EAM-derived scar
Mean unipolar voltage, mV 2.3 � 0.3 4.6 � 0.6 0.001
Mean bipolar voltage, mV 0.7 � 0.1 1.1 � 0.2 0.001
Endocardial low voltage area,† cm2 13.8 � 1.7 3.0 � 1.6 <0.001
Endocardial low voltage burden,‡ % 7.9 � 0.7 2.1 � 0.2 <0.001
Values are mean � SD. *Scar related wall: anterior for LAD group, inferior for LCx group. †Low voltage: bipolarvoltage <1.5 mV. ‡The Mesh feature on the Carto3 software was used to calculate.
EAM ¼ electroanatomical mapping; EGMs ¼ electrograms; other abbreviations as in Table 1.
J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0 Kucukseymen et al.O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4 Limitation of Voltage Mapping to Detect Scar
1457
reliability. Analyses were performed using SPSSsoftware v.25.0 (IBM Corporation, Armonk, NewYork) and GraphPad Prism software v.8.0 (GraphPadSoftware, San Diego, California). p Values <0.05 wereconsidered statistically significant.
RESULTS
CMR ANALYSIS. All animals underwent in vivo CMRat a mean day of 60 � 3.6 following myocardialinfarction and at a mean day of 6 � 1.6 before the EAMprocedure.
Table 1 summarizes the comprehensive CMR data.LV end-diastolic and end-systolic volumes weregreater in the LAD group than those in the LCx group.LV mass and LV ejection fraction were similar in bothgroups. The global radial, circumferential, and lon-gitudinal systolic peak strain values were also similarin both of the LAD and LCx groups.
In LAD animals, LGE was present mostly in theanterior myocardial segments, extending from themid-ventricle to the apex (Figures 1A and 1B). In LCxanimals, LGE was observed on the inferior myocardialsegments, extending from the base to the apex(Figures 1C and 1D). LGE scar volume, mass, andburden in in vivo and ex vivo 3-dimensional LGEimages were similar for both groups (Table 1). Toavoid false-positive image results, we also analyzedex vivo images thoroughly and compared in vivo andex vivo images head-to-head (Supplemental Figure S3and Supplemental Table 1).
Areas of LGE scar bordering the blood pool were signif-icantly different between the LAD (12.5� 3.1 cm2) and LCx(4.1 � 1.2 cm2) groups (p < 0.0001) (Table 1 andSupplemental Figure 4). Although LGE scar was moresubendocardial in the LAD group, LCx animals had LGEscars that extended from the endocardium to the epicar-dium with a significant presence in the mid-myocardium(Supplemental Figure 5). There were significant changesinwall thickness between LCx and LAD animals (Figure 2).Although averagewall thickness in healthy remote regionswas similar between groups (LCx: 6.75� 0.38mmvs. LAD:6.66 � 0.42 mm; p ¼ 0.703), wall thickness in scarredsegments was significantly affected in the LAD group (1.78� 0.26 mm), but the scarred region was similar to healthyremote regions in the LCx animals (5.38 � 0.67 mm). Theextent of wall thinning was significant in the LAD group,whereas 73 � 40% of the wall had reduced thickness,consistent with our previous experience (10).
Intra- and interobserver reliability correlation co-efficients for each group were 0.97 (95% confidenceinterval: [CI]: 0.95 to 0.98) and 0.94 (95% CI: 0.8 to0.99) for remote wall thickness measurements, 0.97(95% CI: 0.95 to 0.98) and 0.87 (95% CI: 0.80 to 0.91)
for scarred wall thickness, 0.95 (95% CI: 0.91 to 0.97)and 0.86 (95% CI: 0.78 to 0.93) for LGE volume, and0.98 (95% CI: 0.97 to 0.99) and 0.95 (95% CI: 0.94 to0.99) for the subendocardial LGE area, respectively(Supplemental Table 2).
EAM ANALYSIS. Animals underwent EAM at a meanday of 59.5 (range 56 to 68 days) after infarct induc-tion, with the same survival period for LAD and LCxanimals. We analyzed EGMs on the scar-related walls:anterior wall for the LAD group and inferior wall forthe LCx group (1,533 � 543 vs. 618 � 192 EGMs,respectively; p ¼ 0.01) during RV pacing.
Table 2 shows EGM data for the anterior wall,inferior wall, and EAM-derived scar parameters. Forscar-related walls, the mean bipolar (1.77 � 0.3 mV vs.2.8 � 0.2 mV; p < 0.0001) and unipolar voltages (3.83� 0.9 mV vs. 9.25 � 2.1 mV; p < 0.0001) were lower inthe LAD group. The mean voltage within the EAM-derived scar was higher in the LCx group than thatin the LAD group: 1.12 � 0.2 mV versus 0.7 � 0.1 mV(p ¼ 0.001) for bipolar voltage; and 4.63 � 0.6 mVversus 2.3 � 0.3 mV (p ¼ 0.001) for unipolar voltage,respectively.
The low bipolar voltage (<1.5 mV) area and lowvoltage burden were 13.8 � 1.7 cm2 and 7 � 0.7% ofthe LV in LAD animals compared with 3.8 � 2.9 cm2
and 1.8 � 0.2% of the LV in LCx animals (p < 0.0001for both).
Epicardial EAM was performed in 6 of 12 animals (2in LAD and 4 in LCx). Small areas of low bipolarvoltage corresponding to LGE were seen in the LADgroup, indicating LGE presence on the endocardialanteroseptal location in both swine. However, therewere no areas of low bipolar or unipolar voltage in theLCx group corresponding to LGE scar (SupplementalFigure 6).
CONCORDANCE BETWEEN SCAR DETECTION USING
CMR AND EAM. In an all-swine analysis, wall
print&web4C=F
PO
FIGURE 3 Distribution of Bipolar and Unipolar Voltage Amplitudes
Bar plots show the (A) inverse association between mean bipolar (red bars) and unipolar (blue bars) voltage amplitudes and wall thinning
grade in all animals. (B) Comparison of mean bipolar (red bars) and unipolar (blue bars) voltage amplitudes between LV walls with LGE for both
groups. The LAD group with an anterior scar shows lower bipolar voltage amplitude than that in the LCx group with an inferior scar. Abbre-
viations as in Figures 1 and 2. * and " indicates p < 0.0001.
Kucukseymen et al. J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0
Limitation of Voltage Mapping to Detect Scar O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4
1458
thinning percentage (graded: <25% [mild], 25% to75% [moderate], >75% [severe]) was inversely asso-ciated with mean bipolar voltage (mild vs. moderate:2.70 � 1.4 mV vs. 2.26 � 1.4 mV; p < 0.0001; mild vs.severe: 2.70 � 1.4 mV vs. 1.14 � 1.2 mV; p < 0.0001;moderate vs. severe: 2.26 � 1.4 mV vs. 1.14 � 1.2 mV;p < 0.0001) (Figure 3A) and with mean unipolarvoltage (between mild and moderate: 7.14 � 2.9 mVvs. 6.23 � 3.5 mV; p < 0.0001; between mild and se-vere: 7.14 � 2.9 mV vs. 3.33 � 2.3 mV; p < 0.0001;between moderate and severe: 6.23 � 3.5 mV vs. 3.33� 2.3 mV; p < 0.0001) (Figure 3B). There were signif-icant differences in mean bipolar (0.63 � 0.69 mV vs.2.52 � 1.35 mV; p < 0.0001) (Figure 3A) and unipolarvoltages (2.28 � 0.84 mV vs. 7.23 � 2.94 mV;p < 0.0001) (Figure 3B) between scar-related LV wallsfor both groups (anterior wall for LAD, inferior wallfor LCx).
On EAM, all LAD animals had a large low bipolarvoltage area compatible with LGE-derived scar fea-tures. However, only 2 of 6 animals in the LCx grouphad low bipolar voltage areas, despite having asimilar large LGE scar on in vivo and ex vivo CMR,and histopathological gross as well (Figure 4,Supplemental Figure 6-8 and Supplemental Table 3).LGE scar in the LAD group corresponded with lowvoltage areas on EAM, whereas LGE scar in the LCx
group did not correspond with low voltage EAM areas(Central Illustration). Full dataset for each pig is alsoavailable in the Supplemental Tables 4 and 5.
HISTOPATHOLOGICAL ASSESSMENT. LGE-derivedscar was validated histologically in both groups. Inaddition, 1 swine from each group showed visualcorrelation between histology samples and CMR im-ages in both in vivo and ex vivo scans with respect toscar extension into the myocardial segment from 1swine for each group (Supplemental Figures 4 and 9).Preserved myocardial tissue was more notable in aslice from the LCx animal.
DISCUSSION
In this preclinical study, we demonstrated the limi-tations of EAM in identifying LGE scar. The discor-dance between low voltage EGMs on EAM and LGEscar highlighted the limitation of catheter mapping toidentify large scar extending from the endocardiumto epicardium within a normal wall thickness.
Despite similar LGE scar sizes between both animalgroups, there were differences in LGE scar charac-teristics. Infarction in the LAD group caused signifi-cant remodeling with greater wall thinning, whichresulted in better LGE scar detection by catheter
print&web4C=F
PO
FIGURE 4 Electroanatomic Characterization of Post-Infarct Scars
EAM bipolar and unipolar voltage maps for each animal (left column for LAD group and right column for LCx group) showing extensive low
voltage areas in the LAD group (anterior view) but limited low voltage areas in the LCx group (inferior view). Only 2 animals had EAM-derived
scar areas that were mostly mismatched with LGE-derived scar. Abbreviations as in Figure 1.
J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0 Kucukseymen et al.O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4 Limitation of Voltage Mapping to Detect Scar
1459
print&
web4C=FPO
CENTRAL ILLUSTRATION Mismatch in Scar Delineation Between Electroanatomical Bipolar/Unipolar Voltage Mapsand Late Gadolinium Enhancement
Kucukseymen, S. et al. J Am Coll Cardiol EP. 2020;6(11):1452–64.
Slices on electroanatomical mapping (EAM) bipolar and unipolar voltage maps (dotted lines) corresponding to short-axis 3-dimensional late gadolinium enhancement
(LGE) images showing low bipolar and unipolar voltage areas in the left anterior descending (LAD) group, but nearly normal voltage in the left circumflex (LCx) group
despite scar presence on LGE images. CMR ¼ cardiac magnetic resonance.
Kucukseymen et al. J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0
Limitation of Voltage Mapping to Detect Scar O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4
1460
mapping. However, wall thickness was considerablypreserved in the LCx group and in cases in which LGEscars extended from the subendocardial to epicar-dium regions. These differences contributed to thelimitations of catheter mapping in identifying LGEscar.
We and others have extensively used the anteriormyocardial infarction (MI) swine model for investi-gation of ventricular arrhythmia (25,35–38). However,
scars in this model were large and caused wall thin-ning and significant remodeling of the heart (10).With advances in acute MI guidelines and shorterdoor-to-balloon times (39), scars in patients with MIare often much smaller, with preserved wall thick-ness. Therefore, the LCx group might better representscar characteristics in our patients.
Scar identification mismatch between CMR andEAM mismatch might be more common than
J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0 Kucukseymen et al.O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4 Limitation of Voltage Mapping to Detect Scar
1461
currently anticipated. Catheter mapping sensitivity toidentify scars is highly dependent on the location anddistance of the scar with respect to the epicardium(35). Wolf et al. (35) investigated the correlation be-tween the endocardial potential and infarct trans-murality in an anterior infarction dog model. Theydemonstrated that the extent of infarct transmuralitycould be accurately assessed in vivo using electricalparameters derived from the endocardial surface.Codreanu et al. (19) investigated the correlation ofLGE- and EAM-derived scar in 10 patients who un-derwent post-infarct VT ablation and reported amismatch in infarct surfaces between voltage mapsand LGE in one-third of infarcted areas. Takigawaet al. (40) used a previous version of the mappingcatheter we used to evaluate EGM amplitudes withinthe scar, including near- and far-field componentsbased on bipolar spacing in thin and thick myocardialwalls. They discovered that far-field voltages had apositive correlation with wall thickness; in contrast,near-field voltages had no correlation with wallthickness. This annotation issue was reported as animportant limitation of current EAM systems toexplain why an important number of EGMs showednormal voltages even in transmural scars on CMR(40). Desjardins et al. (41) correlated LGE with EAM inpost-infarct patients and demonstrated a somewhatsignificant correlation between unipolar voltage andscar depth (R ¼ �0.61) but a weak correlation withbipolar voltage (R ¼ �0.47). In post-MI patients,Wijnmaalen et al. (16) showed strong correlation be-tween transmural infarcts and low voltage areas onEAM. Despite this correlation, visual analysis,particularly in patients with inferior MI, showed amismatch between EAM- and LGE-derived scar areas.In these patients with inferior MI, they reportedlarger LGE-derived scar areas than low voltage areason EAM and several scar areas that were not detectedby EAM (16). Unlike the bipolar EGM, the usefulnessof the unipolar EGM to localize scar is not welldefined (42). Unipolar signals have been suggested tohave a larger field of view than bipolar signals, whichallows detection of subepicardial scars during endo-cardial EAM (35,42,43). Hutchinson et al. (42)described endocardial unipolar signal characteristicsin patients with epicardial scars. They identifiedabnormal epicardium using an endocardial unipolarvoltage of <8.3 mV in patients with nonischemiccardiomyopathy and normal LV endocardial bipolarvoltage. Compared with these previous studies(35,42,43), our unipolar voltage cutoff of 5.5 mV waslower. This difference in unipolar voltage cutoffmight be related to the smaller electrode size of themultielectrode catheter or differences in the scar-
related wall thicknesses between LAD and LCx ani-mals. However, the mean unipolar voltage ampli-tudes on scar-related walls in the LAD group were>8.3 mV in our study.
A novel mapping method technique for character-izing and targeting the intramural substrate using aretractable needle within an endocardial catheter isemerging as an effective approach for intra-myocardial mapping and ablation (44–46). Infusionneedle catheters were developed by modifying acatheter originally designed for intramyocardial drugdelivery (47). Unipolar EGMs can thus be recordedfrom the needle, and bipolar EGMs can be recordedbetween the needle and the tip of the catheter. Thisapproach also allows unipolar pacing and radio-frequency delivery (44–49). There are some studiesthat evaluated the safety and effectiveness of intra-mural needle ablation (45,46,49). A larger multi-center study of patients with treatment-refractoryventricular arrhythmias resulted in arrhythmia elim-ination in 48% of patients and improvement inarrhythmia burden (49). Although these novel tech-niques for targeting intramural sources of VT arepromising, their long-term effectiveness, safety, andcardiac complications warrant additional study.
Another mapping challenge with respect to inferiorinfarcts is the presence of scar tissue adjacent to orjust behind papillary muscles. EAM and ablationprocedures in patients in this region may be difficultdue to the lack of contact between the catheter tipand the tissue (50,51). However, this can be poten-tially overcome by flexible bidirectional catheters andadjunct imaging guidance (e.g., intracardiac echo-cardiography), which was used for each animal in thisstudy to avoid false-negative voltage mapping resultsand to improve success rates (50–52).
Several studies reported that advanced cardiacimaging integration techniques resulted in accuratecharacterization of the arrhythmogenic substrate andenhanced ablation procedure efficiency and safety(18,53–55). A total of 157 consecutive patients withventricular arrhythmias were prospectively enrolledby Hennig et al. (53). Diagnosis before performingCMR was based on standard diagnostic tests (echo-cardiography, stress test, and invasive angiography),yet the implementation of CMR determined a newdiagnosis or a diagnostic change in 48 patients. Njeimet al. (54) observed similar findings in 20 patientswith a history of failed ablation who underwent CMRfor substrate identification before a repeat ablationprocedure. In these patients, CMR identified bothepicardial and mid-myocardial scars, indicating thatepicardial mapping might not be effective in patientswith only a mid-myocardial substrate. Similarly, our
PERSPECTIVES
COMPETENCY IN MEDICAL KNOWLEDGE: Sub-
strate mapping using EAM systems during sinus
rhythm is a common approach for identifying the VT
substrate. A bipolar signal threshold of 1.5 mV is often
used to differentiate healthy and scarred regions.
Because there are multiple scar types with different
biophysical features, repeating this standard method
might not be applicable to most scars. In certain cases,
post-infarction remodeling could result in walls with
preserved thickness, thus presenting as a more com-
plex structure with unique electroanatomical proper-
ties. Operators should be vigilant to validate low
voltage areas on EAM even when using multielectrode
mapping catheters. CMR can provide detailed scar
characteristics that might not be readily available with
current state-of-the-art EAM systems.
TRANSLATIONAL OUTLOOK: This preclinical
study examined the sensitivity of EAM unipolar and
bipolar voltage maps to analyze the correlation be-
tween areas of low voltage on EAM- and LGE-derived
scar areas. Our results showed that voltage maps
could not identify all scarred regions and had limited
sensitivity at detecting scar in a thick LV wall.
Kucukseymen et al. J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0
Limitation of Voltage Mapping to Detect Scar O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4
1462
LCx swine had preserved epicardium with only mid-myocardial LGE scar, which suggested a lack of cor-relation between epicardial EAM and LGE scar.
STUDY LIMITATIONS. Our study had several limita-tions. The number of animals was small, althoughconsistent with larger preclinical studies. LGE imageswere often collected at mid-diastole, which did notcorrespond to the timing of EAM data collection. Thismight have resulted in spatial registration errors be-tween LGE and EAM. Therefore, we did not performspatial registration between the 2 modalities andinstead analyzed LGE and EAM data independently.The inability to obtain consistently good contact at allpoints when using the Octaray catheter might havecontributed to the reported discordance between lowvoltage area and CMR findings. This was especiallyrelevant for the endocardial surface of the infarctedLCx territory, which was less flat and had moreintracavitary structures than the anterior wall andapex. We used a Thermocool catheter with contactforce in only 2 swine, which showed large normalamplitudes. Further studies using catheters are war-ranted to verify our findings. Comprehensive analysisof epicardial mapping data was not performed,because epicardial mapping was not performed in allanimals. EGM characteristics were not described inthis study and warrant further investigation to betterunderstand the role of EGM morphology among LGE-and voltage-derived scars in thick and thin myocar-dium. Our study was not designed to investigate VTinducibility in our models; therefore, we did notsystematically collect and analyze VT inducibilitydata in all animals. Finally, histological analysis wasnot performed on all animals, which precluded in-depth investigation.
CONCLUSIONS
Our preclinical study demonstrated the limitations ofcurrent state-of-the-art catheter mapping systems toidentify scarred regions extending from sub-endocardial to epicardial regions because of discor-dances between low voltage areas in EAM and
scarring on CMR. Furthermore, voltage maps couldvary significantly between areas of reduced or pre-served wall thickness, regardless of the presence orthe extent of scar.
ACKNOWLEDGMENTS The authors acknowledgecontributions from Biosense Webster for providingassistance in performing the mapping and extractingthe electrogram data.
ADDRESS FOR CORRESPONDENCE: Dr. Reza Neza-fat, Department of Medicine, Cardiovascular Division,Beth Israel Deaconess Medical Center, 330 BrooklineAvenue, Boston, Massachusetts 02215. E-mail:[email protected].
RE F E RENCE S
1. Anter E, Kleber AG, Rottmann M, et al. Infarct-related ventricular tachycardia: redefining the elec-trophysiological substrate of the isthmus during si-nus rhythm. J Am Coll Cardiol EP 2018;4:1033–48.
2. Marchlinski FE, Callans DJ, Gottlieb CD, Zado E.Linear ablation lesions for control of unmappableventricular tachycardia in patients with ischemicand nonischemic cardiomyopathy. Circulation2000;101:1288–96.
3. Cronin EM, Bogun FM, Maury P, et al. 2019HRS/EHRA/APHRS/LAHRS expert consensusstatement on catheter ablation of ventricular ar-rhythmias. Heart Rhythm 2020;17:e2–154.
4. Kuck KH, SchaumannA, Eckardt L, et al. Catheterablation of stable ventricular tachycardia beforedefibrillator implantation in patients with coronaryheart disease (VTACH): a multicentre randomisedcontrolled trial. Lancet 2010;375:31–40.
5. Josephson ME, Anter E. Substrate mappingfor ventricular tachycardia. Assumptions and mis-conceptions. J Am Coll Cardiol EP 2015;1:341–52.
6. Anter E. Electroanatomical mapping at acrossroads. J Am Coll Cardiol EP 2019;5:1168–71.
7. Fenoglio JJ Jr., Pham TD, Harken AH,Horowitz LN, Josephson ME, Wit AL. Recurrentsustained ventricular tachycardia: structure and
J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0 Kucukseymen et al.O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4 Limitation of Voltage Mapping to Detect Scar
1463
ultrastructure of subendocardial regions in whichtachycardia originates. Circulation 1983;68:518–33.
8. Callans DJ, Ren JF, Michele J, Marchlinski FE,Dillon SM. Electroanatomic left ventricular map-ping in the porcine model of healed anteriormyocardial infarction. Correlation with intracar-diac echocardiography and pathological analysis.Circulation 1999;100:1744–50.
9. Tschabrunn CM, Roujol S, Dorman NC,Nezafat R, Josephson ME, Anter E. High-resolu-tion mapping of ventricular scar: comparison be-tween single and multielectrode catheters. CircArrhythm Electrophysiol 2016;9:e003841.
10. Whitaker J, Tschabrunn CM, Jang J, et al.Cardiac MR characterization of left ventricularremodeling in a swine model of infarct followed byreperfusion. J Magn Reson Imaging 2018.
11. Ashikaga H, Sasano T, Dong J, et al. Magneticresonance-based anatomical analysis of scar-related ventricular tachycardia: implications forcatheter ablation. Circ Res 2007;101:939–47.
12. Pashakhanloo F, Herzka DA, Mori S, et al.Submillimeter diffusion tensor imaging and lategadolinium enhancement cardiovascular magneticresonance of chronic myocardial infarction.J Cardiovasc Magn Reson 2017;19:9.
13. Pashakhanloo F, Herzka DA, Halperin H,McVeigh ER, Trayanova NA. Role of 3-dimensionalarchitecture of scar and surviving tissue in ven-tricular tachycardia: insights from high-resolutionex vivo porcine models. Circ Arrhythm Electro-physiol 2018;11:e006131.
14. Pirolo JSHG, Moore GW. Infarct expansion:pathologic analysis of 204 patients with a singlemyocardial infarct. J Am Coll Cardiol 1986;7:349–54.
15. Davies M. The pathology of myocardialischaemia. J Clin Pathol Suppl 1977;11:45–52.
16. Wijnmaalen AP, van der Geest RJ, van Huls vanTaxis CF, et al. Head-to-head comparison ofcontrast-enhanced magnetic resonance imagingand electroanatomical voltage mapping to assesspost-infarct scar characteristics in patients withventricular tachycardias: real-time image integra-tion and reversed registration. Eur Heart J 2011;32:104–14.
17. Mountantonakis SE, Park RE, Frankel DS, et al.Relationship between voltage map “channels” andthe location of critical isthmus sites in patientswith post-infarction cardiomyopathy and ventric-ular tachycardia. J Am Coll Cardiol 2013;61:2088–95.
18. Andreu D, Berruezo A, Ortiz-Perez JT, et al.Integration of 3D electroanatomic maps andmagnetic resonance scar characterization into thenavigation system to guide ventricular tachycardiaablation. Circ Arrhythm Electrophysiol 2011;4:674–83.
19. Codreanu A, Odille F, Aliot E, et al. Electro-anatomic characterization of post-infarct scarscomparison with 3-dimensional myocardial scarreconstruction based on magnetic resonance im-aging. J Am Coll Cardiol 2008;52:839–42.
20. Mukherjee RK, Whitaker J, Williams SE,Razavi R, O’Neill MD. Magnetic resonance imaging
guidance for the optimization of ventriculartachycardia ablation. Europace 2018;20:1721–32.
21. Gupta S, Desjardins B, Baman T, et al. Delayed-enhanced MR scar imaging and intraproceduralregistration into an electroanatomical mappingsystem in post-infarction patients. J Am Coll Car-diol Img 2012;5:207–10.
22. Jang J, Whitaker J, Leshem E, et al. Localconduction velocity in the presence of late gado-linium enhancement and myocardial wall thinning.Circ Arrhythm Electrophysiol 2019;12:e007175.
23. Anter ETC, Josephson ME. High-resolutionmapping of scar-related atrial arrhythmias usingsmaller electrodes with closer interelectrodespacing. Circ Arrhythm Electrophysiol 2015;8:537–45.
24. Graham AJ, Orini M, Lambiase PD. Limitationsand challenges in mapping ventricular tachycardia:new technologies and future directions. ArrhythmElectrophysiol Rev 2017;6:118–24.
25. Leshem E, Tschabrunn CM, Jang J, et al. High-resolution mapping of ventricular scar: evaluationof a novel integrated multielectrode mapping andablation catheter. J Am Coll Cardiol EP 2017;3:220–31.
26. Sroubek J, Rottmann M, Barkagan M, et al.A novel octaray multielectrode catheter for high-resolution atrial mapping: electrogram character-ization and utility for mapping ablation gaps.J Cardiovasc Electrophysiol 2019;30:749–57.
27. Thajudeen A, Jackman WM, Stewart B, et al.Correlation of scar in cardiac MRI and high-resolution contact mapping of left ventricle in achronic infarct model. Pacing Clin Electrophysiol2015;38:663–74.
28. Oduneye SO, Pop M, Shurrab M, et al. Distri-bution of abnormal potentials in chronic myocar-dial infarction using a real time magneticresonance guided electrophysiology system.J Cardiovasc Magn Reson 2015;17:27.
29. Rottmann M. Multiple Interacting Rotational,Focal, and Wavelet Activities at HeterogeneousTissue in Atrial Fibrillation. Karlsruhe, Germany:Karlsruher Institut Für Technologie (KIT), 2018.
30. Fahmy AS, Neisius U, Tsao CW, et al. Grayblood late gadolinium enhancement cardiovascu-lar magnetic resonance for improved detection ofmyocardial scar. J Cardiovasc Magn Reson 2018;20:22.
31. Basha TA, Tang MC, Tsao C, et al. Improveddark blood late gadolinium enhancement (DB-LGE) imaging using an optimized joint inversionpreparation and T2 magnetization preparation.Magn Reson Med 2018;79:351–60.
32. Tschabrunn CM, Roujol S, Nezafat R, et al.A swine model of infarct-related reentrant ven-tricular tachycardia: electroanatomic, magneticresonance, and histopathological characterization.Heart Rhythm 2016;13:262–73.
33. Barkagan M, Sroubek J, Shapira-Daniels A,et al. A novel multielectrode catheter for high-density ventricular mapping: electrogram charac-terization and utility for scar mapping. Europace2020;22:440–9.
34. Sosa ESM, d’Avila A, Pilleggi F. A new tech-nique to perform epicardial mapping in the
electrophysiology laboratory. J Cardiovasc Elec-trophysiol 1996;7:531–6.
35. Wolf T, Gepstein L, Dror U, et al. Detailedendocardial mapping accurately predicts thetransmural extent of myocardial infarction. J AmColl Cardiol 2001;37:1590–7.
36. Reddy VY, Wrobleski D, Houghtaling C,Josephson ME, Ruskin JN. Combined epicardialand endocardial electroanatomic mapping in aporcine model of healed myocardial infarction.Circulation 2003;107:3236–42.
37. Barkagan M, Leshem E, Shapira-Daniels A,et al. Histopathological characterization of radio-frequency ablation in ventricular scar tissue. J AmColl Cardiol EP 2019;5:920–31.
38. McCall FC, Telukuntla KS, Karantalis V, et al.Myocardial infarction and intramyocardial injectionmodels in swine. Nat Protoc 2012;7:1479–96.
39. Antman EM, Anbe DT, Armstrong PW, et al.ACC/AHA guidelines for the management of pa-tients with ST-elevation myocardial infarction–executive summary. A report of the AmericanCollege of Cardiology/American Heart AssociationTask Force on Practice Guidelines (Writing Com-mittee to revise the 1999 guidelines for themanagement of patients with acute myocardialinfarction). J Am Coll Cardiol 2004;44:671–719.
40. Takigawa M, Martin R, Cheniti G, et al.Detailed comparison between the wall thicknessand voltages in chronic myocardial infarction.J Cardiovasc Electrophysiol 2019;30:195–204.
41. Desjardins B, Crawford T, Good E, et al. Infarctarchitecture and characteristics on delayedenhanced magnetic resonance imaging and elec-troanatomic mapping in patients with post-infarction ventricular arrhythmia. Heart Rhythm2009;6:644–51.
42. Hutchinson MD, Gerstenfeld EP, Desjardins B,et al. Endocardial unipolar voltage mapping todetect epicardial ventricular tachycardia substratein patients with nonischemic left ventricular car-diomyopathy. Circ Arrhythm Electrophysiol 2011;4:49–55.
43. Spears DA, Suszko AM, Dalvi R, et al. Rela-tionship of bipolar and unipolar electrogramvoltage to scar transmurality and compositionderived by magnetic resonance imaging in patientswith nonischemic cardiomyopathy undergoing VTablation. Heart Rhythm 2012;9:1837–46.
44. Sapp JL, Cooper JM, Soejima K, et al. Deepmyocardial ablation lesions can be created with aretractable needle-tipped catheter. PACE 2004;27:594–9.
45. Sapp JL, Beeckler C, Pike R, et al. Initial humanfeasibility of infusion needle catheter ablation forrefractory ventricular tachycardia. Circulation2013;128:2289–95.
46. Abdel Wahab A, Stevenson W, Thompson K,et al. Intramural ventricular recording and pacingin patients with refractory ventricular tachycardia:initial findings and feasibility with a retractableneedle catheter. Circ Arrhythm Electrophysiol2015;8:1181–8.
47. Neira V, Santangeli P, Futyma P, et al. Ablationstrategies for intramural ventricular arrhythmias.Heart Rhythm 2020;17:1176–84.
Kucukseymen et al. J A C C : C L I N I C A L E L E C T R O P H Y S I O L O G Y V O L . 6 , N O . 1 1 , 2 0 2 0
Limitation of Voltage Mapping to Detect Scar O C T O B E R 2 6 , 2 0 2 0 : 1 4 5 2 – 6 4
1464
48. Markman TM, Nazarian S. Treatment of ven-tricular arrhythmias: what’s new? Trends Car-diovasc Med 2019;29:249–61.
49. Stevenson WG, Tedrow UB, Reddy V, et al.Infusion needle radiofrequency ablation for treat-ment of refractory ventricular arrhythmias. J AmColl Cardiol 2019;73:1413–25.
50. Doppalapudi H, Yamada T, McElderry HT,Plumb VJ, Epstein AE, Kay GN. Ventricular tachy-cardia originating from the posterior papillarymuscle in the left ventricle: a distinct clinicalsyndrome. Circ Arrhythm Electrophysiol 2008;1:23–9.
51. Kautzner J, Peichl P. Papillary muscle ventric-ular tachycardia or ectopy: diagnostics, catheter
ablation and the role of intracardiac echocardiog-raphy. Arrhythm Electrophysiol Rev 2019;8:65–9.
52. Enriquez A, Briceno D, Tapias C, et al.Ischemic ventricular tachycardia from below theposteromedial papillary muscle, a particular en-tity: substrate characterization and challengesfor catheter ablation. Heart Rhythm 2019;16:1174–81.
53. Hennig A, Salel M, Sacher F, et al. High-reso-lution three-dimensional late gadolinium-enhanced cardiac magnetic resonance imaging toidentify the underlying substrate of ventriculararrhythmia. Europace 2018;20:f179–91.
54. Njeim M, Yokokawa M, Frank L, et al. Value ofcardiac magnetic resonance imaging in patients
with failed ablation procedures for ventriculartachycardia. J Cardiovasc Electrophysiol 2016;27:183–9.
55. Njeim M, Desjardins B, Bogun F. Multimodalityimaging for guiding EP ablation procedures. J AmColl Cardiol Img 2016;9:873–86.
KEY WORDS cardiac cardiac magneticresonance, electroanatomical mapping,myocardial infarction model, wall thickness
APPENDIX For supplemental figures andtables, please see the online version of thispaper.