comparison of input values of shielding design parameters for medical linac with iaea safety report...
TRANSCRIPT
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
1/22
Presented by Supervised by
M. Anwarul Islam
Medical Physicist
SQUARE Hospitals Ltd
Dhaka-Bangladesh
Email: [email protected]
Golam Abu Zakaria
Professor & Head
Gummersbach Hospital, AcademicTeaching Hospital of the University ofCologne Germany
Comparison of input values of shielding design parameters formedical linac with IAEA Safety Report Series 47 and calculated
values of shielding parameters with treatment delivered datafrom Square Hospitals.
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
2/22
Objectives
Optimal protection to the radiation worker& public
Easy handling to regulatory compliance
Feasibility check to IMRT up gradation
Optimal RT service delivery with 3DCRT &
IMRT
Avoidance the unnecessary shielding cost
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
3/22
Method & Materials
More than 800 radiotherapy patients data
analyzed from SQUARE hospitals ltd
Analyzed field sizes, gantry angles, photonenergy, delivered monitor units and dose
Analyzed average treatment time for3DCRT and IMRT
Calculated maximum treatment capacitiesper dayfor all modalities
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
4/22
Calculated possible maximum beam ONtime for 1 year of treatments
Calculated Instantaneous Dose Rate (IDR)
to follow the IAEA protocol
(1mSv/y for public and 20mSv/y for
occupational dose limit)
Method & Materials
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
5/22
IAEA Shielding CalculationParameters
Yearly radiation dose limit for radiationworker (20mSv/year) & public (1mSv/year)
Maximum workload of the machine /week
- depends on the number of patients treating / week
- treatment modalities (3DCRT/IMRT)
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
6/22
Use factor of the barrier (primary -1/4 &secondary -1)
Occupancy factor (occupational nature, outside of the barrier. Example: office, reception,shop-1, corridors-1/4, toilets, stairways-1/16 etc)
Energy of the machine
IAEA Shielding CalculationParameters
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
7/22
Why more shielding required for IMRT ??
Shielding features Comments
Delivered dose tothe tumor perfraction
Similar for DCRT & IMRT. Workload isequal for both modalities for primaryshielding thickness calculation
Scattering radiationby patient
Similar for 3DCRT & IMRT
Scattered radiation
by primary barrier
Similar for both modalities
Linac head leakage 3 - 5 times more for IMRT than 3DCRT
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
8/22
Findings
Analyzed 1608 treatment fields
Average field size: 13.9cm 16.6 cm
Maximum field size: 22.6cm 27.4cm
IAEA recommendation: 40cm 40cm
Field Size (3DCRT/2D treatment with 6MV):
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
9/22
Cont.
Treatment with 10MV beam
Average field size: 13.6cm 13.8 cm
Maximum field size: 24.8cm 24.8cm
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
10/22
IAEA recommendation: 0.25 for primary beam ofall angle (0, 90, 180, 270)
Barrier Use Factors
Calculated dataModality Gantry Angle,
degPrimary barrier Calculated Use
factorIAEA
3DCRT
With
6MV and
10 MV
46-135 Vertical wall 0.33 0.25
136-225 Roof 0.15 0.25
226-315 Vertical wall 0.33 0.25
316-45 Floor 0.19 0.25
(Use factors calculated from angle wise MU delivery data)
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
11/22
Cont.
VerticalWall
VerticalWall
Roof
Floor
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
12/22
Workload depends on
- working hour per day
- no. patients treat per day
- working day per week
Workload
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
13/22
The calculated maximum workload was found tobe 740 Gy / week and 500 Gy / week for 3DCRTwith 6MV and 10MV respectively
IAEA recommendation
Workload usually 1000Gy/week used with 6MV
for 3DCRT and 600Gy/week used for higherenerg
Calculated Workload
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
14/22
Calculated IMRT factor, C1 = 4.36
Calculated workload = 1800 Gy/Week
The workload for IMRT is applicable to calculatethe shielding thickness against the head leakage
radiation only.Leakage radiation is generally (0.1-0.2)% of
useful beam
Cont.
Workload for IMRT center
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
15/22
Treatment time basis workload
Average treatment time/patient was 9.64 Min
Average MU delivered /treatment was 297
2D/3DCRT treatment capacity per day (8h) = 50
Workload = 148Gy/Day i.e 740Gy/week
(3DCRT/2D treatments)
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
16/22
Cont.
Average treatment time/patient was 16.92 Min
Average MU delivered /treatment was 1285
IMRT treatment capacity per day (8h) = 28
Workload = 360Gy/day i.e 1800Gy/week
IMRT treatments
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
17/22
Considering Maximum dose rate = 600MU/Min1 MU 1cGy at isocenter of the machine
Max. Beam ON time for 3DCRT/2D = 102 hour/year
(when machine fully occupied with 3DCRT/2D treatment only)
Max. Beam ON time for IMRT = 253 hour/year(when machine fully occupied with IMRT treatment only)
The proposed design limit for IDR is following an IAEArecommendation with 1 mSv/year for public and 20
mSv/year for occupational workers
IDR calculation for design limit
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
18/22
Calculated IDR for design limit
TreatmentModalities
Calculated IDR (in Sv/h) fordesign limit
Public Area Occupational Area
3DCRT center1 IDR
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
19/22
Permissible design limit
TreatmentModalities
OccupancyArea Permisible Design Limit (in Sv/h)
IAEA UK USA Bangladesh
N/A
Public
---
IDR
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
20/22
Conclusion
The IDR Survey (by NSRC) results in front of thedoor & linac control room was < 4 Sv/h
(FS= 40 40 cm2, E = 10MV, DR = 600 MU/Min,G.A = 0, 90, 180, 270)
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
21/22
Cont.
The calculated radiation dose at the neardoor & control room of SQUARE hospital wasfound to be 0.2 mSv/year
(Data analyzed within the year of 2010 and 2011 inthe basis of the total MU delivered within this time
and NSRC survey data)
Finally, the study reveals that the radiotherapydepartment in Square hospital is sufficiently
radiation protected for radiation workers and public.
-
8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli
22/22