comparison of input values of shielding design parameters for medical linac with iaea safety report...

Upload: anwar01726

Post on 06-Apr-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    1/22

    Presented by Supervised by

    M. Anwarul Islam

    Medical Physicist

    SQUARE Hospitals Ltd

    Dhaka-Bangladesh

    Email: [email protected]

    Golam Abu Zakaria

    Professor & Head

    Gummersbach Hospital, AcademicTeaching Hospital of the University ofCologne Germany

    Comparison of input values of shielding design parameters formedical linac with IAEA Safety Report Series 47 and calculated

    values of shielding parameters with treatment delivered datafrom Square Hospitals.

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    2/22

    Objectives

    Optimal protection to the radiation worker& public

    Easy handling to regulatory compliance

    Feasibility check to IMRT up gradation

    Optimal RT service delivery with 3DCRT &

    IMRT

    Avoidance the unnecessary shielding cost

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    3/22

    Method & Materials

    More than 800 radiotherapy patients data

    analyzed from SQUARE hospitals ltd

    Analyzed field sizes, gantry angles, photonenergy, delivered monitor units and dose

    Analyzed average treatment time for3DCRT and IMRT

    Calculated maximum treatment capacitiesper dayfor all modalities

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    4/22

    Calculated possible maximum beam ONtime for 1 year of treatments

    Calculated Instantaneous Dose Rate (IDR)

    to follow the IAEA protocol

    (1mSv/y for public and 20mSv/y for

    occupational dose limit)

    Method & Materials

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    5/22

    IAEA Shielding CalculationParameters

    Yearly radiation dose limit for radiationworker (20mSv/year) & public (1mSv/year)

    Maximum workload of the machine /week

    - depends on the number of patients treating / week

    - treatment modalities (3DCRT/IMRT)

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    6/22

    Use factor of the barrier (primary -1/4 &secondary -1)

    Occupancy factor (occupational nature, outside of the barrier. Example: office, reception,shop-1, corridors-1/4, toilets, stairways-1/16 etc)

    Energy of the machine

    IAEA Shielding CalculationParameters

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    7/22

    Why more shielding required for IMRT ??

    Shielding features Comments

    Delivered dose tothe tumor perfraction

    Similar for DCRT & IMRT. Workload isequal for both modalities for primaryshielding thickness calculation

    Scattering radiationby patient

    Similar for 3DCRT & IMRT

    Scattered radiation

    by primary barrier

    Similar for both modalities

    Linac head leakage 3 - 5 times more for IMRT than 3DCRT

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    8/22

    Findings

    Analyzed 1608 treatment fields

    Average field size: 13.9cm 16.6 cm

    Maximum field size: 22.6cm 27.4cm

    IAEA recommendation: 40cm 40cm

    Field Size (3DCRT/2D treatment with 6MV):

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    9/22

    Cont.

    Treatment with 10MV beam

    Average field size: 13.6cm 13.8 cm

    Maximum field size: 24.8cm 24.8cm

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    10/22

    IAEA recommendation: 0.25 for primary beam ofall angle (0, 90, 180, 270)

    Barrier Use Factors

    Calculated dataModality Gantry Angle,

    degPrimary barrier Calculated Use

    factorIAEA

    3DCRT

    With

    6MV and

    10 MV

    46-135 Vertical wall 0.33 0.25

    136-225 Roof 0.15 0.25

    226-315 Vertical wall 0.33 0.25

    316-45 Floor 0.19 0.25

    (Use factors calculated from angle wise MU delivery data)

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    11/22

    Cont.

    VerticalWall

    VerticalWall

    Roof

    Floor

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    12/22

    Workload depends on

    - working hour per day

    - no. patients treat per day

    - working day per week

    Workload

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    13/22

    The calculated maximum workload was found tobe 740 Gy / week and 500 Gy / week for 3DCRTwith 6MV and 10MV respectively

    IAEA recommendation

    Workload usually 1000Gy/week used with 6MV

    for 3DCRT and 600Gy/week used for higherenerg

    Calculated Workload

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    14/22

    Calculated IMRT factor, C1 = 4.36

    Calculated workload = 1800 Gy/Week

    The workload for IMRT is applicable to calculatethe shielding thickness against the head leakage

    radiation only.Leakage radiation is generally (0.1-0.2)% of

    useful beam

    Cont.

    Workload for IMRT center

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    15/22

    Treatment time basis workload

    Average treatment time/patient was 9.64 Min

    Average MU delivered /treatment was 297

    2D/3DCRT treatment capacity per day (8h) = 50

    Workload = 148Gy/Day i.e 740Gy/week

    (3DCRT/2D treatments)

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    16/22

    Cont.

    Average treatment time/patient was 16.92 Min

    Average MU delivered /treatment was 1285

    IMRT treatment capacity per day (8h) = 28

    Workload = 360Gy/day i.e 1800Gy/week

    IMRT treatments

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    17/22

    Considering Maximum dose rate = 600MU/Min1 MU 1cGy at isocenter of the machine

    Max. Beam ON time for 3DCRT/2D = 102 hour/year

    (when machine fully occupied with 3DCRT/2D treatment only)

    Max. Beam ON time for IMRT = 253 hour/year(when machine fully occupied with IMRT treatment only)

    The proposed design limit for IDR is following an IAEArecommendation with 1 mSv/year for public and 20

    mSv/year for occupational workers

    IDR calculation for design limit

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    18/22

    Calculated IDR for design limit

    TreatmentModalities

    Calculated IDR (in Sv/h) fordesign limit

    Public Area Occupational Area

    3DCRT center1 IDR

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    19/22

    Permissible design limit

    TreatmentModalities

    OccupancyArea Permisible Design Limit (in Sv/h)

    IAEA UK USA Bangladesh

    N/A

    Public

    ---

    IDR

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    20/22

    Conclusion

    The IDR Survey (by NSRC) results in front of thedoor & linac control room was < 4 Sv/h

    (FS= 40 40 cm2, E = 10MV, DR = 600 MU/Min,G.A = 0, 90, 180, 270)

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    21/22

    Cont.

    The calculated radiation dose at the neardoor & control room of SQUARE hospital wasfound to be 0.2 mSv/year

    (Data analyzed within the year of 2010 and 2011 inthe basis of the total MU delivered within this time

    and NSRC survey data)

    Finally, the study reveals that the radiotherapydepartment in Square hospital is sufficiently

    radiation protected for radiation workers and public.

  • 8/3/2019 Comparison of input values of shielding design parameters for medical linac with IAEA Safety Report Series 47 and calculated values of shielding parameters with treatment deli

    22/22