clarifications no:1 to tender dossier · 2020. 11. 20. · installation, commissioning,...
TRANSCRIPT
![Page 1: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/1.jpg)
Page 1 of 61
CLARIFICATIONS No:1 to TENDER DOSSIER
Contract Title: Supply of Medical Equipment for the Secondary Healthcare Premises - Relaunch
Publication Reference: SIHHAT/2018/SUP/INT/10-BIS
CN: Contract Notice
TD: Tender Dossier
DOC: Document
ART: Article
ITT: c4b_itt_en [Instructions to Tenderers]*
DC: Draft Contract
TS: c4f_annexiitechspeciiitechoffer_en [Annex II + III: Technical Specifications + Technical Offer]*
GC: General Conditions
SC: c4d_specialconditions_en [Special Conditions]*
PG: c4h_perfguarantee_en [Performance Guarantee]
TG: c4n_tenderguarantee_en [Tender Guarantee]
App B: Appendix B to Annex II - Training Proposal [Appendix B to Annex II+III Training Proposal]*
Ann V: Annex V - Warranty Proposal [Annex V - Warranty Obligations Form]*
* In-parenthesis parts show the title inside the documents.
Further to the requests received from the tenderers, the following clarifications are provided.
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
1 CN 8
Participation is open to all natural persons who are
nationals of and legal persons (participating either
individually or in a grouping – consortium – of
tenderers) which are effectively established in a Member
State of the European Union or in an eligible country or
According to the requirements on the
left, we would like to clarify that; if we
offer a product, which is originating
from the United Kingdom for the
relevant tender on 29.03.2019 and if
Please be informed that the goods originating
from the United Kingdom will be threatened
as goods originating from third countries
provided that;
- The United Kingdom withdraws
![Page 2: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/2.jpg)
Page 2 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
territory as defined under the Regulation (EU) No
236/2014 establishing common rules and procedures for
the implementation of the Union's instruments for
external action (CIR) for the applicable instrument under
which the contract is financed (see also heading 22
below). Participation is also open to international
organisations.
All supplies under this contract must originate in one or
more of these countries.
Please be aware that after the United Kingdom's
withdrawal from the EU, the rules of access to EU
procurement procedures of economic operators
established in third countries and of goods originating
from third countries will apply to candidates or tenderers
from the United Kingdom and to all candidates or
tenderers proposing goods originating from the United
Kingdom depending on the outcome of negotiations. In
case such access is not provided by legal provisions in
force at the time of the contract award, candidates or
tenderers from the United Kingdom, and candidates or
tenderers proposing goods originating from the United
Kingdom could be rejected from the procurement
procedure.
the United Kingdom will not
withdrawal from the European Union
until the date of signing the contract by
us, will our contract be valid? After
the signing the contract; We will give
medical device order from the United
Kingdom. So that please clarify; If
United Kingdom will withdrawal from
the European Union after the signing
the contract, will our contract be valid?
from the European Union prior to the
award decision (contract signature
accordingly),
- No rules/regulations developed
covering the United Kingdom
originated goods within the scope of
the rules of access to EU
procurement procedures (depending
on the outcome of negotiations).
Please be aware that the statements cover the
procurement procedure (please see on the lef
: “will apply to candidates or tenderers”
and “candidates or tenderers from the United
Kingdom, and candidates or tenderers
proposing goods originating from the United
Kingdom could be rejected from the
procurement procedure” and “at the time of
the contract award”).
Please also be aware that following the
contract signature, the contract remains valid
regardless of the origin of the goods and the
outcomes of the withdrawal of the United
Kingdom from the European Union.
2 CN 16(1)
……
The average annual turnover of the tenderer in
the last three years must be equal or exceed the
50% tenderer’s financial offer.
……
We kindly request to revise the
requirement as “The average annual
turnover of the tenderer in the last
three years must be equal or exceed the
10% tenderer’s financial offer.”
Considering the volume of the tender and the
risk assessment of the Contracting Authority,
the existing requirement assessed to be
sufficient.
![Page 3: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/3.jpg)
Page 3 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
3 CN 16(3)
…..
The tenderer has delivered supplies under at
most three contracts with a budget of at least
one-fourth (¼) of the financial offer of the lot
being tendered in supply of medical
equipment(s).
…..
We kindly request to revise the
requirement as “The tenderer has
delivered supplies under at most three
contracts with a budget of at least one-
fifth (1/5) of the financial proposal of
the lot being tendered in supply of
medical equipment(s).”.
Considering the volume of the tender and the
risk assessment of the Contracting Authority,
the existing requirement assessed to be
sufficient.
4 ITT 1.1
The subject of the contract is the supply, delivery,
installation, commissioning, maintenance, after-sales
service by the contractor of the following goods:
Lot 1 High Level Ventilator - Neonatal 227
Lot 2 Transport Ventilator 106
Lot 3 C-Arm Digital X-Ray Device 70
Lot 4 Haemodialysis Device 145
in 4 lots to the points at the provinces of Turkey (please
refer to the Appendix-A, delivery points list), within 120
(one hundred and twenty) calendar days for Lot-1 and
Lot-2; 150 (one hundred and fifty) calendar days for Lot-
3; 90 (ninety) calendar days for Lot-4 as also stated in
Special Conditions, DDP , in accordance with point 15
of the Contract Notice. Please be aware that the
Contracting Authority reserves the right to update the
quantities per delivery point at any time based on the
possible fluctuations on the number of migrants.
Taking into consideration that the
production, delivery, installation and
commissioning of 70 systems for 49
different distribution sites, we kindly
request to extend the implementation
period for Lot-3 to 210 calendar days.
Considering the current official deadline of
the project under which this tender is being
funded (SIHHAT), there is no possibility to
extend the period as requested.
5 ITT 1.3
None of the supplies must be accompanied by an
additional ‘lot’ consisting of spare parts and/or
consumables.
We request a clarification on what is
exactly demanded in the sentence.
The item has been revised as:
The supplies described under all lots must be
accompanied by an additional ‘lot’ consisting
of spare parts and/or consumables. Neither
the unit price, nor the overall price of spare
![Page 4: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/4.jpg)
Page 4 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
parts will influence the evaluation of the
tenders, except where they vary substantially
between the tenders received. Lists of spare
parts must be drawn up by tenderers on the
basis of their professional experience and the
expected places of use; they must show the
unit prices of the parts, calculated as
specified in Article 11 (below). The
contracting authority reserves the right to
alter the list of spare parts; any changes will
appear in the contract.
Please also refer to Changes No:2 to TD.
6 ITT 3.1
Participation is open to all natural persons who are
nationals of and legal persons (participating either
individually or in a grouping – consortium – of
tenderers) which are effectively established in a Member
State of the European Union or in an eligible country or
territory as defined under the Regulation (EU) No
236/2014 establishing common rules and procedures for
the implementation of the Union's instruments for
external action (CIR) for the applicable instrument under
which the contract is financed (see also heading 22 of the
contract notice). Participation is also open to
international organisations. All supplies under this
contract must originate in one or more of these countries.
However, they may originate from any country when the
amount of the supplies to be purchased (as a whole or, if
divided into lots, per lot) is below EUR 100 000.
Should we understand the total Lot
price or the unit price from the
definition of “below EUR 100.000”
that is mentioned in the article in
question as follows: “However, they
may originate from any country when
the amount of the supplies to be
purchased (as a whole or, if divided
into lots, per lot) is below EUR
100.000?
As clearly stated in the existing item in
bracelet “(as a whole or, if divided into lots,
per lot)”; the item refers to the total value of
each Lot.
![Page 5: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/5.jpg)
Page 5 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
7 ITT 28
The tenderers and, if they are legal entities, persons who
have powers of representation, decision-making or
control over them, are informed that, should they be in
one of the situations of early detection or exclusion, their
personal details (name, given name if natural person,
address, legal form and name and given name of the
persons with powers of representation, decision-making
or control, if legal person) may be registered in the early
detection and exclusion system, and communicated to
the persons and entities listed in the above-mentioned
decision, in relation to the award or the execution of a
procurement contract.
We request a clarification on what is
exactly demanded or what is
emphasized in the sentence on the left.
The contracting authority shall take all
necessary measures to ensure full compliance
with the principles of PRAG and also by
facilitating the detection of unreliable
economic operators and the protection of
the European Union's financial interests. To
this aim, an Early Detection and Exclusion
System (EDES) is established by the
Financial Regulation to reinforce the
protection of the Union's financial interests
and to ensure sound financial management.
This item empowers the Contracting
Authority to communicate the data of
Tenderer to the EDES if necessary.
8 DC 1.1
The subject of the contract shall be the supply, delivery,
installation, commissioning, testing, inspection, training
and warranty services of the following supplies:
Lot 1: High Level Ventilator – Neonatal
Lot 2: Transport Ventilator
Lot 3: C-Arm Digital X-Ray Device
Lot 4: Haemodialysis Device
The place of acceptance of the supplies shall be indicated
in the Appendix A of the Annex II+III technical
specifications and technical offer, the time limits for
delivery shall be 120 (one hundred and twenty) calendar
days for Lot-1 and Lot-2; 150 (one hundred and fifty)
calendar days for Lot-3; 90 (ninety) calendar days for
Lot-4 and the Incoterm applicable shall be DDP. The
implementation period of tasks shall run from the date of
commencement of the contract to the date for provisional
acceptance.
Taking into consideration that the
production, delivery, installation and
commissioning of 70 systems for 49
different distribution sites, we kindly
request to extend the implementation
period for Lot-3 to 210 calendar days.
Please see the answer to clarification number
4 above.
![Page 6: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/6.jpg)
Page 6 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
9 DC 1.3
The supplies described under all lots must be
accompanied by an additional ‘lot’ consisting of spare
parts and/or consumables. Neither the unit price, nor the
overall price of spare parts will influence the evaluation
of the tenders, except where they vary substantially
between the tenders received. Lists of spare parts must
be drawn up by tenderers on the basis of their
professional experience and the expected places of use;
they must show the unit prices of the parts, calculated as
specified in Article 11 (below). The contracting authority
reserves the right to alter the list of spare parts; any
changes will appear in the contract.
Could you consider removing the
following part “The contracting
authority reserves the right to alter the
list of spare parts; any changes will
appear in the contract." For the sake of
the principle of efficient use of public
resources?
The item has been revised as:
The supplies which form the subject of the
contract for all lots must be accompanied by
the spare parts described by the contractor in
its tender and by the accessories and other
items necessary for using the goods over a
period of 10 years, as specified in the
instructions to tenderers.
Please also refer to Changes No:2 to TD.
10 SC 19.1
The supply, delivery, installation, commissioning,
testing, inspection and training of goods will be
completed in 120 calendar days for Lot-1 and Lot-2; 150
calendar days for Lot-3 and 90 calendar days for Lot-4
starting from the date of commencement of the contract.
Taking into consideration that the
production, delivery, installation and
commissioning of 70 systems for 49
different distribution sites, we kindly
request to extend the implementation
period for Lot-3 to 210 calendar days.
Please see the answer to clarification number
4 above.
11 SC 25.2
Inspection and testing shall take place upon delivery, and
in accordance with Article 25 of the General Conditions
and Technical Specifications Annex II + III at the
delivery points in Appendix A to Special Conditions.
The Contractor shall be liable for any damage that might
be given to public or private property, or the general
public, during the installation or throughout the operation
period of the project in line with its purposes within the
warranty period. The Contractor shall be liable for
compensation of losses arising from such damages.
We request the article to be modified
as covering the direct losses only, as it
is not possible to find out the direct
and indirect losses when it is specified
as all losses.
The item remains unchanged considering the
needs of the Contracting Authority and the
public interest.
12 SC 26.3
By derogation, the final payment to the contractor of the
amounts due shall be made within 90 days after receipt
by the contracting authority of an invoice and of the
application for the certificate of provisional acceptance.
Could you consider revising the item
as “By derogation, the final payment to
the contractor of the amounts due shall
be made within 60 days after receipt
by the contracting authority of an
invoice and of the application for the
There is no contradiction between the
documents. Please be aware that the
particular items in the Special Conditions
serves for the purpose to derogate from
General Conditions. The term “by
derogation” at the very beginning of the item
![Page 7: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/7.jpg)
Page 7 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
certificate of provisional acceptance.”
in order to eliminate conflict between
documents as the document 'ANNEX
I:GENERAL CONDITIONS" Article
26.3 states that the payment would be
done within 60 days.
clearly states that the Contractşng Authority
here derogates from article 26.3 of the
General conditions with this particular item
in Special Conditions.
13 SC 32.6
- Repair time: Within 30 calendar days from the receipt
of the malfunctioning goods. If during 30 calendar days,
it is foreseen that the goods cannot be repaired and the
malfunction is not fault of the operator, corresponding
functional item should be provided until malfunctioning
goods is repaired.
Could the item considered to be
revised as “Repair time:Within 30
calendar days from the receipt of the
malfunctioning goods. If during 30
calendar days, It Is foreseen that the
goods cannot be repaired and the
malfunction is not fault of the operator,
In case, the warranty period wlll be
added free of charge for 2 days per
day.” considering various conditions
such as production, logistics,
installation, calibration of imaging
systems consisting of numerous units
having high technologic value, it is not
possible to leave a substitution device
within repair period?
The article has already been revised by
Changes No:1 to Tender Dossier, please refer
the public document.
14 GC 12.1(a)
Without prejudice to Article 32 (warranty obligations)
and Article 38 (force majeure), the contractor shall
assume (i) full responsibility for maintaining the
integrity of the supplies and (ii) the risk of loss and
damage, whatever their cause, until the final acceptance
as foreseen in Article 34.
Could you consider revising the item
as “Without prejudice to Article 32
(warranty obligations) and Article 38
(force majeure), the contractor shall
assume (i) full responsibility for
maintaining the integrity of the
supplies and (ii) the risk of loss and
damage, whatever their cause, until
The partial provisional acceptance as
foreseen in Article 34.” considering
the fact that in the current form of the
article, the risks and damages in the
Please be aware that “General Conditions” is
not a dynamic document subject to the
modification but a stable template document
itself.
![Page 8: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/8.jpg)
Page 8 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
systems delivered to the end users will
be delivered to the relevant hospitals
after 5 years following the provisional
acceptance. In its current form, the
amount of the Insurance to be made for
each system affects the unit price of
the system?
15 GC 17.2
Such license shall carry the right to grant sub-licenses
and shall be transferable by the contracting authority to
third parties without the consent of the contractor being
required.
Please clarify that grant of the
sublicense will be limited to the end
users which are the hospitals not any
other 3rd party.
Please see the answer to clarification number
14 above.
16 GC 26.7
For supplies not covered by a warranty period, the
payments listed above shall be aggregated. The
conditions to which the payments of pre-financing and
final payments are subject, shall be as stated in the
special conditions.
We request information about what is
exactly intended to express with the
sentence of "For supplies not covered
by a warranty period, the payments
listed above shall be aggregated.".
If the goods under the contract are not subject
to the warranty period, the payments shall be
aggregated; which is not a case for none of
the lots under the subject tender as all the
goods required to be covered by warranty
period.
17 GC 26.8
The payment obligations of the European Commission
under this contract shall cease at most 18 months after
the end of the period of implementation of the tasks,
unless the contract is terminated in accordance with these
general conditions.
Could you consider revising the item
same as with the warranty period as
“The payment obligations of the
European Commission under this
contract shall cease at most 120
months after the end of the period of
implementation of the tasks, unless the
contract is terminated in accordance
with these general conditions.” as the
judgment proceedings for any dispute
that might arise between the parties
within the scope of the tender and
similar proceedings are likely to
exceed 18 months?
Please see the answer to clarification number
14 above.
![Page 9: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/9.jpg)
Page 9 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
18 GC 32.2
The contractor shall be responsible for making good any
defect in, or damage to, any part of the supplies which
may appear or occur during the warranty period and
which:
a) results from the use of defective materials, faulty
workmanship or design of the contractor; and/or
b) results from any act or omission of the contractor
during the warranty period; and/or appears in the
course of an inspection made by, or on behalf of, the
contracting authority.
We kindly request to change the clause
as follows: “The contractor shall be
responsible for making good any
defect in, or damage to, any part of the
supplies which may appear or occur
during the warranty period and which:
a) results from the use of
defective materials by Contractor's
technical services, faulty workmanship
or design of the contractor; and/or
b) results from any act or
omission of the contractor during the
warranty period; and/or
c) appears in the course of an
inspection made by, or on behalf of,
the contracting authority.
Please see the answer to clarification number
14 above.
19 GC 32.3
The contractor shall at its own cost make good the defect
or damage as soon as practicable. The warranty period
for all items replaced or repaired shall recommence from
the date when the replacement or repair was made to the
satisfaction of the project manager. If the contract
provides for partial acceptance, the warranty period shall
be extended only for the part of the supplies affected by
the replacement or repair.
We kindly request the removal of the
item.
Please see the answer to clarification number
14 above.
20 GC 32.4
If any such defect appears or such damage occurs during
the warranty period, the contracting authority or the
project manager shall notify the contractor. If the
contractor fails to remedy a defect or damage within the
time limit stipulated in the notification, the contracting
authority may:
a) remedy the defect or the damage itself, or employ
someone else to carry out the tasks at the contractor's risk
and cost, in which case the costs incurred by the
contracting authority shall be deducted from monies due
to or from guarantees held against the contractor or from
both; or
We kindly request to change the clause
as follows: “If any such defect appears
or such damage occurs during the
warranty period, the contracting
authority or the project manager shall
notify the contractor.”.
Please see the answer to clarification number
14 above.
![Page 10: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/10.jpg)
Page 10 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
b) terminate the contract.
21 GC 32.5
In case of emergency, where the contractor cannot be
reached immediately or, having been reached, is unable
to take the measures required, the contracting authority
or the project manager may have the tasks carried out at
the expense of the contractor. The contracting authority
or the project manager shall as soon as practicable
inform the contractor of the action taken.
We kindly request the removal of the
item.
Please see the answer to clarification number
14 above.
22 GC 34.1
Upon expiry of the warranty period, or where there is
more than one such period, upon expiry of the latest
period, and when all defects or damage have been
rectified, the project manager shall issue the contractor a
final acceptance certificate and a copy thereof to the
contracting authority, stating the date on which the
contractor completed its obligations under the contract to
the project manager's satisfaction . The final acceptance
certificate shall be issued by the project manager within
30 days after the expiration of the warranty period or as
soon as any repairs ordered under Article 32 have been
completed to the satisfaction of the project manager .
Upon expiry of the warranty period, or
where there is more than one such
period, upon expiry of the latest
period, and when all defects or damage
have been rectified, the project
manager shall issue the contractor a
final acceptance certificate and a copy
thereof to the contracting authority,
stating the date on which the
contractor completed its obligations
under the contract to the project
manager's satisfaction. The final
acceptance certificate shall be issued
by the project manager by the end of
the warranty period or as soon as any
repairs ordered under Article 32 have
been completed to the satisfaction of
the project manager.
Please see the answer to clarification number
14 above.
23 PG NA
(....)We note that the guarantee will be released within 60
days of the issue of the final acceptance certificate
(except for such part as may be specified in the special
conditions in respect of after sales service). [and in any
case at the latest on (at the expiry of 18 months after the
We kindly request to change the clause
as follows: “(....) We note that the
guarantee will be released within 30
days of the issue of the provisional
acceptance certificate (except for such
Please be aware that the relevant part of the
document is not subject to the revision as it is
a standard template contained in the PRAG.
![Page 11: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/11.jpg)
Page 11 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
period of implementation of the tasks)] 1.(...) part as may be specified in the special
conditions in respect of after sales
service) . (...)”.
24 TG NA
(....)We note that the guarantee will be released at the
latest within 45 days of the expiry of the tender validity
period, including any extensions, in accordance with
Article 8 of the Instructions to tenderers
[and in any case at the latest on (1 year after the deadline
for submission of tenders)]2 .(....)
We kindly request to change the clause
as follows: “(... .) We note that the
guarantee will be released at the latest
within 30 days at the latest after the
provisional acceptance,, including any
extensions, in accordance with Article
8 of the Instructions to tenderers. (...)”.
Please be aware that the relevant part of the
document is not subject to the revision as it is
a standard template contained in the PRAG.
25
TS –
General
Require
ments
3.9
Unless otherwise stated (please see technical
specifications for each Lot for the specific
requirements of each Lot), the contractor at least 2
(two) days free training of at least 2 (two) staff to
determine the use, maintenance, calibration, care
and possible defects of the device with their trained
staff. These trainings will be repeated up to 3 times
for each device if requested during the warranty
period. This requirement will be certified by the
contractor in the tender file. The date and place
which will be determined by the center. Documents
and equipment’s required for training shall be met
by the Contractor.
We kindly request to change the clause
as follows: “Unless otherwise stated
(please see technical specifications for
each Lot for the specific requirements
of each Lot), the contractor at least 2
(two) days free training of at least 2
(two) staff to determine the use, daily
maintenance, calibration, first level
intervention in case of malfunctions
/care and possible defects of the device
with their trained staff. These trainings
will be repeated once more free of
charge for each device if requested
during the warranty period. This
requirement will be certified by the
contractor in the tender file.
Documents and equipment's required
for training shall be met by the
Contractor.”.
The requirement remains unchanged
considering the needs of the Ministry and
sustainability of the health services.
26 TS 1.3
The device should be in fully compact structure. All
parameter adjustments and values of measurements shall
be made and tracked on at least 12’’ internal monitor of
In order to increase the number of
qualified competitors we kindly
request to change the clause as
The specification remains unchanged
considering the needs of the Ministry.
![Page 12: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/12.jpg)
Page 12 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
the ventilator. The internal screen of the device should be
separated from the main unit and put on pendant if
wanted in the future.
follows: “The device should be in fully
compact structure. All parameter
adjustments and values of
measurements shall be made and
tracked on at least 12'' internal monitor
of the ventilator. The internal screen of
the device should be separated from
the main unit or ventilator should be
separated from the transport cart and
put on pendant if wanted in the
future.”
27
We kindly request to change the clause
as follows: “The device should be in
fully compact structure . All parameter
adjustments and values of
measurements shall be made and
tracked on at least 12" internal monitor
of the ventilator or on the ventilator
with 10" screen and having the
same brand integrated humidifier
device. The internal screen of the
device should be separated from the
main unit and put on pendant if wanted
in the future or should be in fully
compact structure.”.
28 TS 1.5
VG (Volume Guarantee) property should be present as
standard. Tidal volume limiting properties other than this
mode shall not be accepted.
(VG (Volume Guarantee) property and
VTV (Volume Targeted Ventilation)
properties are the same properties and
only the names of the properties differ.
Therefore, in order to increase the
number of qualified competitors we
kindly request to change the clause as
follows: VG (Volume Guarantee) or
VTV (Volume Targeted Ventilation)
The specification remains unchanged
considering the needs of the Ministry.
![Page 13: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/13.jpg)
Page 13 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
property should be present as standard.
Tidal volume limiting properties other
than these modes shall not be
accepted.”
29 TS 1.6
There should be freeze or save button on the device and
wave forms or loops should be frozen or saved as instant
images.
We kindly request to change the clause
as follows: “Freeze or save buttons
should be present on the device, wave
forms and loops should be frozen or
saved as instant images.”.
The specification remains unchanged. Please
also see the Changes No.1 to Tender Dossier
for the updated version of the item.
30 TS 1.7
HFOT (High Flow O2 Therapy) feature should be added
to the devices optionally. The price of this feature
shouldn’t exceed %15 of tender unit price.
We kindly request to change the clause
as follows: “HFOT (High Flow 02
Therapy) feature should be added to
the devices optionally or in ncpap
mode flow can be ad justed
manually from 6-20 lpm . The price
of this feature shouldn't exceed %15
of tender unit price.”.
The specification remains unchanged
considering the needs of the Ministry.
31 TS 1.14.8 Tidal volume: 2-200 ml
We kindly request to change the clause
as follows: “Tidal Volume : 2-150
ml.”.
The specification has been revised as:
Tidal Volume : 2-150 ml.
Please also refer to Changes No.2 to Tender
Dossier.
32
TS 1.15
The device must have an integrated back lighted LCD
graphic display at least 12 inches.
- Airway pressures,
- Flow, Volume, Pressure Curves,
-Trend curves of measurements or numerical values
should be monitored on this screen.
We kindly request to change the clause
as follows: “The device must have an
integrated back lighted LCD graphic
display at least 10 inches.”.
The item has been revised as:
The device must have an integrated back
lighted LCD or LED graphic display at least
12 inches.
- Airway pressures,
- Flow, Volume, Pressure Curves,
- Trend curves of measurements or numerical
values should be monitored on this screen.
33
We kindly request to change the clause
as follows: “The device must have an
integrated back lighted LCD
or LED backlight monitor graphic
![Page 14: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/14.jpg)
Page 14 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
display at least 12 inches.
- Airway pressures,
-Flow, Volume, Pressure Curves,
Trend curves of measurements or
numerical values should be monitored
on this screen.”
Please also refer to Changes No.2 to Tender
Dossier.
34
TS 1.16
3 (three) waveforms and 2 breathing cycles or 4
waveforms should be monitored at the same time on the
internal screen of the device. In addition, respiratory
cycles (P / F, P / V, V / F) should be monitored.
We kindly request to change the clause
as follows: “3 (three) waveforms and 2
breathing cycles or 3 waveforms
should be monitored at the same time
in the internal screen of the device . In
addition, respiratory cycles (P / F, P /
V, V / F) should be monitored.”.
The specification remains unchanged.
35
We kindly request to change the clause
as follows: “3 (three) waveforms and 2
breathing cycles or 4 waveforms or
1(one) waveforms and 2 (two)
breathing cycles should be monitored
at the same time on the internal screen
of the device. In addition, respiratory
cycles (P I F,P I V, V I F) should be
monitored.”.
36 TS 1.17
The humidifier supplied with the device must have the
following features.
We kindly request to change the clause
as follows: “The ventilator should
have an integrated humidifier which
can be controlled from the ventilator
screen and functions according to
the parameters set by the user. The
companies which cannot fulfill this
feature must supply external
humidifier which is clearly defined
in items 1.17.1, 1.17.2, 1.17.3,
1.17.4.”.
The specification remains unchanged.
![Page 15: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/15.jpg)
Page 15 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
37 TS 1.19
The device must contain the following alarms:
- Air and oxygen shutdown alarm or gas pressures low alarm
-System failure alarm
-High airway pressure alarm
- Low airway pressure alarm or PEEP alarm
- High and low oxygen concentration alarm or oxygen
value deviation alarm
- Flow sensor failure alarm
- Oxygen sensor failure alarm
- Flow sensor clogged (dirty) alarm or flow sensor
failure
-Disconnection or cycle fail alarm (leakage or
connection error in the patient circuit)
- High minute volume alarm
- Low minute volume alarm
- Low tidal volume alarm or low minute volume alarm
- Apnea alarm
- Endotracheal tube clogged alarm or high pressure in the
inspiratory tube alarm or high airway pressure alarm or P
Op high alarm
- Clogging in the patient circuit or high airway pressure
or resistance height alarm or inspiration-expiratory tube
alarm (Against the increase in resistance due to clogging
or curling in the patient circuit)
As the requested version also fully
satisfies the requirements, we kindly
request to change the clause as
follows: “The device must contain the
following alarms:
- Air and oxygen shutdown alarm or
gas pressures low alarm or no gas
alarm -System failure alarm
-High airway pressure alarm or high
pressure threshold exceeded alarm
- Low airway pressure alarm or PEEP
alarm
- High and low oxygen concentration
alarm or oxygen value deviation
alarm
- Flow sensor failure alarm or floor
sensor defective alarm
- Oxygen sensor failure alarm or 02
calibration fail alarm
- Flow sensor clogged (dirty) alarm or
flow sensor failure or flow sensor
contaminated alarm
- Disconnection or cycle fail alarm
(leakage or connection error in the
patient circuit)
- High minute volume alarm
- Low minute volume alarm
- Low tidal volume alarm or low
minute volume alarm
- Apnea alarm
- Endotracheal tube clogged alarm or
high pressure in the inspiratory tube
alarm or high airway pressure alarm
The item has been revised as:
The device must contain the following
alarms:
- Air and oxygen shutdown alarm or gas
pressures low alarm
- System failure alarm
- High airway pressure alarm or high
pressure threshold exceeded alarm
- Low airway pressure alarm or PEEP alarm
- High and low oxygen concentration alarm
or oxygen value deviation alarm
- Flow sensor failure alarm or floor sensor
defective alarm
- Oxygen sensor failure alarm
- Flow sensor clogged (dirty) alarm or flow
sensor failure alarm
- Disconnection or cycle fail alarm (leakage
or connection error in the patient circuit)
- High minute volume alarm
- Low minute volume alarm
- Low tidal volume alarm or low minute
volume alarm
- Apnea alarm
- Endotracheal tube clogged alarm or high
pressure in the inspiratory tube alarm or high
airway pressure alarm or P Op high alarm
- Clogging in the patient circuit or high
airway pressure or resistance height alarm or
inspiration-expiratory tube alarm (Against
the increase in resistance due to clogging or
curling in the patient circuit).
Please also refer to Changes No.2 to Tender
![Page 16: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/16.jpg)
Page 16 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
or P Op high alarm or blocked fresh
gas alarm
- Clogging in the patient circuit or high
airway pressure or resistance height
alarm or blocked fresh gas alarm or
inspiration-expiratory tube alarm
(Against the increase in resistance
due to clogging or curling in the
patient circuit).”.
Dossier.
38 TS 1.23 The device must have RS 232 and USB output.
We kindly request to change the clause
as follows: “The device must have RS
232 or USB output.”.
The specification remains unchanged.
39 TS 1.26
There should be a led indicator on the device or on the
uninterruptible power supply indicating the electrical
connection. In the absence of the mains supply, the
battery indicator of the device or uninterruptible power
supply indicator must warn the user.
We kindly request to change the clause
as follows: “There should be a led
indicator or halo indicator on the
device or on the uninterruptible
power supply indicating the electrical
connection. In the absence of the
mains supply, the battery indicator of
the device or uninterruptible power
supply indicator must warn the user.”.
The item has been revised as:
There should be a led indicator or halo
indicator on the device or on the
uninterruptible power supply indicating the
electrical connection. In the absence of the
mains supply, the battery indicator of the
device or uninterruptible power supply
indicator must warn the user.
Please also refer to Changes No.2 to Tender
Dossier.
40 TS 1.29
HFO + VG feature and n-HFO feature or proportional
pressure support (SPN-PPS) can be added to devices to
which HFO feature can be added in the future.
We kindly request to change the clause
as follows: “HFO + VG feature or n-
HFO feature or proportional pressure
support (SPN- PPS) can be added to
device to which HFO feature can be
added in the future.”.
The specification remains unchanged.
![Page 17: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/17.jpg)
Page 17 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
41
We kindly request to change the clause
as follows: “HFO + VG feature and n-
HFO feature or HFO+VTV (Volume
Target Ventilation) feature and
nHFOV (Nasal High-Frequency
Oscillation Ventilation) feature or proportional pressure support (SPN-
PPS) can be added to devices to which
HFO feature can be added in the
future.”.
42 TS 1.30
Materials To Be Supplied With The Ventilator
The following accessories will be provided for the
operation of the device and their suitability for the
current system will be indicated.
1- Oxygen central system connection hose: 1 pc
2- Central system air connection hose: 1 pc
3- Dual line heated disposable air circuit (Inspiration and
expirium line heater assembly)should be provided. The
disposable Chamber with automatic filling must be
double floated. In addition, there must be a pressure flow
limiter for the mechanical ventilators with pressure
control in the set package. Devices of the companies that
offer mechanical ventilator with this feature should be
fully compatible. All parts (chamber, patient set, flow
limiter and other accessories) must be original and the
same brand with moisturizing device.. Disposable sets of
materials from different manufacturers shall not be
accepted. :20 pcs
4- Disposable nasal CPAP set: 10 sets
(Each nasal set will contain prong, mask, bone, and if
any, interconnection devices. Prong, mask and bone
should have different size options.)
5- Sling arm for patient circuit: 1 pc
6- For systems working with hot wire principle, reusable
flow sensor: 5 pcs
7- Test lung: 1 piece
8- Humidifier unit and accessories: 1 pc
We kindly request to change the clause
as follows: “3 Dual line heated
disposable air circuit (Inspiration and
expirium line heater assembly) should
be present. For the systems having
disposable Chamber, the disposable
chamber with automatic filling must
be double float. In addition, there must
be a pressure flow limiter for the
pressure controlled mechanical
ventilators in the set package. Devices
of the companies that offer mechanical
ventilator with this feature should be
fully compatible. All parts (chamber,
patient set, flow limiter and other
accessories) must be the same brand
with moisturizing device and original.
Disposable sets of materials from
different manufacturers shall not be
accepted : 20 pcs
4- Disposable nasal CPAP set: 10 sets
(Each nasal set will contain prong,
mask, bone, and if any,
interconnection devices. Prong, mask
The specification remains unchanged.
![Page 18: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/18.jpg)
Page 18 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
9- Expiration or exhalation set (full set): 1 pc
and bone should have different size
options.)
5- Sling arm for patient circuit: 1 pc
6- For systems working with hot wire
principle, reusable flow sensor: 5 pcs
7- Test lung: 1 piece
8- Humidifier unit and accessories: 1
pc
9- Expiration or exhalation set (full
set) : 1 pc (when available)
43 TS 2.5
In the ventilator, the parameters defined below should be
adjusted at least in defined intervals.
a) Tidal volume: 50-2000 ml
b) Respiration frequency: 5-60 breathes/minute
c) Inspiration time: 0.3-3.0 seconds or I:E ratio is
at least 1:4-3:1
d) PEEP/CPAP: 0/off-20 cmH2O
e) Pressure Control Ventilation Mode Pressure
(PCV pressure/Pinsp): 5-55 cmH2O
f) Spontaneous Respiration Pressure Support
(PSV Pressure): 2-35 cmH2O
g) FiO2: Should be adjusted at least between %40-
%100
h) Flow value should go up to 100 liters/minute at
least.
i) Pressure Triggering: Between (-9) and (-0.1)
cmH2O or flow trigger should be between 1 and
9 liter/minute at least.
We kindly request to change the clause
as follows: “In the ventilator, the
parameters defined below should be
adjusted at least in
defined intervals.
a) Tidal volume: 5-2500 ml
b) Respiration frequency: 5-60
breathes/minute
c) Inspiration time: 0.3-3.0 seconds or
I:E ratio is at least 1:4-3:1
d) PEEP/CPAP: 0/off-20 cmH2O
e) Pressure Control Ventilation Mode
Pressure(PCV pressure/Pinsp): 5-55
cmH2O
f) Spontaneous Respiration Pressure
Support (PSV Pressure): 2-35 cmH2O
g) FiO2: Should be adjusted at least
between %40-%100
h) Flow value should go up to 100
liters/minute at least.
i) Pressure Triggering: Between (-9)
and (-0.1) cmH2O or flow trigger
should be between 1 and
9 liter/minute at least.”.
The specification remains unchanged.
![Page 19: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/19.jpg)
Page 19 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
44 TS 2.6
The device should have respiration modes with the
parameters written below.
a) Assisted/ Volume-controlled Mandatory
Respiration (A/CMV-V.AC)
b) Pressure Controlled Synchronized Interval
Mandatory Respiration (P-SIMV-PC-
BIPAP/PS)
c) Volume Controlled Synchronized Interval
Mandatory Respiration ( /V-SIMV-VC-SIMV)
d) Pressure Supported Ventilation (PSV) or
Spontaneous Respiration (SPONT-
SpnCPAP/PS)
e) Continuous Positive Airway Pressure (CPAP-
SpnCPAP)
f) Back-up ventilation or apnea back-up
ventilation
The device should have respiration
modes with the parameters written
below.
a) Assisted/Volume-controlled
Mandatory Respiration (A/CMV-
V.AC) or VCNAC
b) Pressure Controlled
Synchronized Interval Mandatory
Respiration (P-SIMV-PC-
BIPAP/PS) or P SIMV+PS -APCV
(BILEVEL ST) c) Volume Controlled Synchronized
Interval Mandatory Respiration ( N -
SIMV-VC-SIMV)
d) Pressure Supported
Ventilation (PSV) or Spontaneous
Respiration (SPONT-SpnCPAP/PS)
e) Continuous Positive Airway
Pressure (CPAP-SpnCPAP)
f) Back-up ventilation or apnea back-
up ventilation
The item has been revised as:
The device should have ventilation modes
with the parameters written below.
a) Assisted/Volume-controlled
Mandatory Ventilation (A/CMV-
V.AC) or equivalent mode
b) Pressure Controlled Synchronized
Interval Mandatory Ventilation (P-
SIMV-PC-BIPAP/PS) or equivalent
mode c) Volume Controlled Synchronized
Intermittent Mandatory Ventilation
( /V-SIMV-VC-SIMV)
d) Pressure Supported Ventilation
(PSV) or Spontaneous Ventilation
(SPONT-SpnCPAP/PS)
e) Continuous Positive Airway Pressure
(CPAP-SpnCPAP)
f) Back-up ventilation or apnea back-
up ventilation
Please also refer to Changes No.2 to Tender
Dossier.
45 TS 2.8
The device should have an internal gas mixer. Oxygen
concentration in the inspiration air should be adjusted
between %40-%100. Oxygen concentration should be
adjusted through the device and the change in the
amount of oxygen flow shouldn’t be required for
concentration change.
Could you consider revising the
specification as: “The device should
have an internal or external gas
mixer. Oxygen concentration in the
inspiration air should be adjusted
between %40 -%100. Oxygen
concentration should be adjusted
through the device or with original
external mixer for concentration
change.” ?
The item has been revised as:
The device should have an internal or
integrated gas mixer. Oxygen concentration
in the inspiration air should be adjusted
between %40 -%100. Oxygen concentration
should be adjusted through the device or
with original integrated mixer for
concentration change. The change in the
![Page 20: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/20.jpg)
Page 20 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
amount of oxygen flow shouldn’t be required
for concentration change.
Please also refer to Changes No.2 to Tender
Dossier.
46 TS 2.15
The structure of the ventilator should have an at least 4,5
inch internal TFT or LCD and Electro-Luminescence
(EL) screen from which the monitorized parameters and
alarm status can be monitored in writing.
We kindly request to change the clause
as follows: “There should be an
internal TFT or LCD or
ElectroLuminescence (EL) screen at
least 8 inch in dimension, through
which monitorized parameters
and alarm statuses can be monitored in
writing, in the structure of the
ventilator.”.
The specification remains unchanged.
47 TS 2.20
The followings shall be provided with each device:
d) 3 adult type, 3 pediatric type reusable respiration
circuits and current sensors and/or 3 expiratory valve
systems or cassettes.
e) One original carrier bag and stand, 3.5 lt oxygen tube
and regulator clock, assembly apparatus for patient’s bed
or stretcher.
f) At least 2 meters long oxygen hose and required lock
clip, sleeves, connectors etc.
We kindly request to change the clause
as follows: “…..
a) 3 adult type reusable, 3 pediatric
type reusable or disposable
respiration circuit and current sensors
and/or 3 expiratory valve systems or
cassettes.
b) One original carrier bag and stand,
oxygen tube of 2 or 3,5 It and
regulator clock, assembly apparatus for
patient 's bed or carrier.
The item has been revised as:
The followings shall be provided with each
device:
a) 3 adult type, 3 pediatric type reusable
respiration circuits and current sensors and/or
3 expiratory valve systems or cassettes.
b) One original carrier bag and stand, 3.5 lt
oxygen tube and regulator clock, assembly
apparatus for patient’s bed or stretcher.
c) At least 2 meters long oxygen hose and
required lock clip, sleeves, connectors etc.
Please also refer to Changes No.2 to Tender
Dossier.
48 TS 2.23
During the warranty period, the contractor shall realize
the periodical maintenances of the device (not being less
than 4 times a year) in place, all expendables being its
own.
Could you consider revising the
specification as: “During the warranty
period, the contractor shall realize the
periodical maintenances of the device
(not being less than 2 times a year) in
The specification remains unchanged.
![Page 21: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/21.jpg)
Page 21 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
place, all expendables being its own.”?
49 TS
2.38
(New) -
The device must be suitable for use in
ambulances, aircraft and on the
ground.
The insertion of such requirement has been
regarded unnecessary considering the
intended purpose of the device.
50
TS 3.1
The system will include an image amplifier, an x-ray
tube, 2 (two) monitors mounted on a C-armed stative,
and an integrated monitor on the stative or wheeled table.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “In the system, the image
intensifier and X-ray tube shall be
mounted on a C-arm stative. 2 (two)
monitors shall be on an C arm stative
or separate cart stand.”.
The item has been revised as:
The System will include an image amplifier
and an x-ray tube which mounted on a C-
armed stative. There should be 2 (two)
monitors or 1 (one) integrated monitor on the
stative of wheeled table.
Please also refer to Changes No.2 to Tender
Dossier. 51
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “In the system, the Flat
detector, X-ray tube shall be mounted
on a C-arm stative. 2 (two) monitors
shall be on separate cart stand.”.
52 TS 3.3
There will be continuous and pulsed fluoroscopy in the
system and there will be digital radiography or single
image mode. All of these features will be shown in the
current product catalog obtained from the Product
Tracking System.
We kindly request you to change the
clause as follows “There will be
continuous and pulsed fluoroscopy in
the system and there will be digital
radiography or single image mode. All
of these features will be shown in the
current product catalog.”.
The catalogues are welcomed to be registered
to the Product Tracking System. The item
will remain unchanged.
53 TS 3.4
At least 12.5 pulse / second pulsed fluoroscopy will be
done in the system.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
The specification remains unchanged
considering the needs of the Ministry.
![Page 22: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/22.jpg)
Page 22 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
follows “At least 10 pulse/second
pulsed fluoroscopy will be done in the
system.”.
54
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “At least 25 pulse/second
pulsed fluoroscopy will be done in the
system.”
55 TS 3.8
If required, DSA option must be added to the system
including pixel shift or remask or masking, landmark,
peak opacification, roadmark or roadmap or subtraction
modes.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “If required, DSA option must
be added to the system including pixel
shift or remask or masking, landmark
or peak opacification, roadmark or
roadmap or subtraction modes.”.
The specification remains unchanged
considering the needs of the Ministry.
56 TS 3.9
The system should have anatomic programs such as
general, orthopedic, spine, vascular, or the imaging
technique that is automatically activated in the system to
achieve optimum image quality. Companies will explain
these features.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The system should have
anatomic programs such as general,
orthopedic, spine or the imaging
technique that is automatically
activated In the system to achieve
optimum Image quality. Companies
will explain these features.”.
The item has been revised as:
The system should have anatomic programs
such as general, orthopedic, spine or the
imaging technique that is automatically
activated In the system to achieve optimum
Image quality. Companies will explain these
features.
Please also refer to Changes No.2 to Tender
Dossier.
57 TS 3.10.1 C-Arm opening should be at least 75 cm.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “C-Arm opening should be at
least 83 cm.”.
The specification remains unchanged
considering the needs of the Ministry.
![Page 23: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/23.jpg)
Page 23 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
58 TS 3.10.3 SID in C-Arm shall be at least 94 cm.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “SID in C-Arm shall be at
least 103 cm.”.
The specification remains unchanged
considering the needs of the Ministry.
59 TS 3.10.9
For ease of use in the system, different color codes
should be used for each brake and movement.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “For ease of use in the system,
different color codes should be used
for each brake and movement or the
system should be equipped with a
warning sign indicating whether the
device is in the brake position or not.
The item has been revised as:
For ease of use in the system, different color
codes should be used for each brake and
movement or the system should be
equipped with a warning sign indicating
whether the device is in the brake position
or not.
Please also refer to Changes No.2 to Tender
Dossier.
60
TS 3.11.1 The x-ray generator must have at least 15 kW and at
least 40 kHz high frequency.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The x-ray generator must
have at least 2,3 kW and at least 30
kHz high frequency.”. The item has been revised as:
The x-ray generator must have at least 12
kW and at least 40 kHz high frequency.
Please also refer to Changes No.2 to Tender
Dossier. 61
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The x-ray generator must
have at least 12 kW and at least 40
kHz high frequency.”.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The x-ray generator must
have at least 2,2 kW and at least 20
kHz high frequency.”.
![Page 24: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/24.jpg)
Page 24 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
62 TS 3.11.2
The fluoroscopic tension shall be at least 40 kV to 120
Kv.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The fluoroscopic tension
shall be at least 40 kV to 110 kV.”.
The item has been revised as:
The fluoroscopic tension shall be at least 40
kV to 110 kV.
Please also refer to Changes No.2 to Tender
Dossier.
63
TS 3.11.3 At continuous fluoroscopy, the current value shall be at
least 15 mA.
In the specifications item 3.12.4,
Anode cooling capacity required is at
least 73000HU/min. with 15mA
continuous fluoroscopy by definition,
15mA x 120KV=1800W at least
required value. 1800W x (4/3) x
60=144 000 HU/min which is double
the required cooling capacity.
Therefore, there is a contradiction in
the required specifications. To
overcome the contradiction, the item
requested to be revised as “At
continuous fluoroscopy, the current
value shall be at least 7.5 mA or
maximum current value shall be at
least 15mA”.
The item has been revised as:
At continuous fluoroscopy, the current value
shall be at least 13 mA.
Please also refer to Changes No.2 to Tender
Dossier.
64
We kindly request the item to be
revised as “At continuous fluoroscopy,
the current value shall be at least 13
mA.”
65 We kindly request the item to be
revised as “At fluoroscopy , the current
value shall be at least 75 mA.”.
66 We kindly request the item to be
revised as “At continuous fluoroscopy,
the current value shall be at least 5,4
![Page 25: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/25.jpg)
Page 25 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
mA.”.
67
TS 3.11.4 In pulsed fluoroscopy, the current value shall be at least
12 mA.
60 mA or more current will add more
powerful working conditions.
Therefore we kindly request the item
to be revised as “In pulsed
fluoroscopy, the current value shall be
at least 60 mA.”
The specification remains unchanged
considering the needs of the Ministry. 68
We kindly request the item to be
revised as “In pulsed fluoroscopy , the
current value shall be at least 60 mA or
fluoroscopy; the current value shall
be at least 110 mA.”.
69
We kindly request the item to be
revised as “In pulsed fluoroscopy, the
current value shall be at least 7,2
mA.”.
70
TS 3.11.5
The system will include digital boost or digital spot or
digital radiography or digital exposures or single image
mode. In this mode which operates separately from
fluoroscopy, the current value will be at least 75 mA.
We kindly request the item to be
revised as “The system will include
digital boost or digital spot or digital
radiography or digital exposures or
single image mode. In this mode which
operates separately from fluoroscopy,
the current value will be at least 13
mA.”. The specification remains unchanged
considering the needs of the Ministry.
71
We kindly request the item to be
revised as “The system will include
digital boost or digital spot or digital
radiography or digital exposures or
single Image mode. In this mode
which operates separately from
fluoroscopy, the current value will be
at least 8 mA.”
![Page 26: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/26.jpg)
Page 26 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
72
TS 3.11.7 X-Ray generator must be single tank or monoblock or
split block structure.
Monoblock system helps acquiring
images with sharp pulses and without
radiating patient unnecessarily. Non-
monoblock system means extra dose to
the patient. The requested statement is
defined same as the technical
specification of Boarders & Coast
Office Type-2 C-arm system bulk
tender in 2015 (Tender No: 2015-
56699). To define an advanced C-arm
system it is kindly requested to revise
the item as follows “X-Ray generator
must be monoblock structure.”.
The specification remains unchanged.
73
The C-Arms which have separate tube
and generator blocks cause artefacts in
sensitive medical devices (especially
ultrasound etc.) due to high voltage
between them. Therefore, we kindly
request the item to be revised as “X-
Ray generator must be single tank
or monoblock structure.”.
74
TS 3.12.2
X-ray tube should be double focused. The small focus
dimensions shall be 0.3 mm at most and the large focus
dimension shall be at least 0.6 mm.
We kindly request the item to be
revised as “The X-ray tube should be
single or double-focused. In the
systems with X-ray tube double
focus, small focus dimensions should
not be more than 0.3 mm, the largest
focus dimension should be no more
than 1.0 mm. For single-focus systems,
the size of the focus shall be 0.6 mm.”.
The specification remains unchanged.
75
We kindly request the item to be
revised as “The X-ray tube should be
double-focused. Small focus size
should not be more than 0.6 mm, the
![Page 27: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/27.jpg)
Page 27 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
largest focus size should be no more
than 1.5 mm.
76
TS 3.12.3 The anode heat capacity of the device shall be at least
311.000 HU.
Systems with rotating anode and high
anode heat capacity allows working on
much more patients quickly without
causing any hesitance during
procedures. In addition to that, X-ray
tubes with rotating anode have much
more lifetime because of erosion will
be less in rotating anodes comparing to
fixed anodes. The requested statement
is defined same as the technical
specification of Boarders & Coast
Office Type-2 C-arm system bulk
tender in 2015 (Tender No: 2015-
56699). To define an advanced C-arm
system in the technical specification, it
is kindly requested to revise the item
as “The anode heat capacity of the X-
Ray tube with rotating anode shall
be at least 311.000 HU.”.
The item has been revised as:
The anode heat capacity of the device shall
be at least 300.000 HU.
Please also refer to Changes No.2 to Tender
Dossier.
77
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The anode heat capacity of
the device shall be at least 300.000
HU”.
78
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The anode heat capacity of
the device shall be at least 61.000
HU.”.
![Page 28: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/28.jpg)
Page 28 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
79
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The anode heat capacity of
the device shall be at least 46.000
HU.”.
80
TS 3.12.4 Anode cooling capacity shall be at least 73.000 HU
/ min.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “Anode cooling heat capacity
shall be at least 60.000 HU/min”.
The item has been revised as:
Anode cooling capacity shall be at least
72.000 HU / min.
Please also refer to Changes No.2 to Tender
Dossier.
81
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “Anode cooling heat capacity
shall be at least 72.000 HU/min”.
82
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “Anode cooling capacity shall
be at least 18.000 HU/ min.”.
83
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “Anode cooling heat capacity
shall be at least 48.000 HU/min.”.
![Page 29: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/29.jpg)
Page 29 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
84
TS 3.12.5
The heat capacity of the x-ray haube should be at least
1900 kHu or the system's heat capacity should be 10000
kHU.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The heat capacity of the x-ray
haube should be at least 1890 kHu or
the system's heat capacity should be
10000 kHU.”.
The item has been revised as:
The heat capacity of the x-ray haube should
be at least 1800 kHu or the system's heat
capacity should be 10000 kHU.
Please also refer to Changes No.2 to Tender
Dossier.
85
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The heat capacity of the x-ray
haube should be at least 1800 kHu or
the system's heat capacity should be
10000 kHU.”.
86
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The heat capacity of the x-ray
haube should be at least 1100 kHu or
the system's heat capacity should be
10000 kHU.”.
87
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The heat capacity of the x-ray
haube should be at least 950 kHu or
the system's heat capacity should be
10000 kHU.”.
![Page 30: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/30.jpg)
Page 30 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
88 TS 3.12.6
There will be 0.1 mm copper (Cu) and 3 mm aluminum
(Al) filter at the tube outlet to filter the harmful emission
rays.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “There will be 0.1 mm copper
(Cu) or 3 mm aluminum (Al)
equivalent filter to filter out the
harmful emission rays at the tube
outlet.”.
The specification remains unchanged.
89
TS 3.12.7
The system must have at least one of the following
special software that improves the image quality.
Companies will show this feature in their original
catalogues:
- EASY (Enhanced Acquisition System) with automatic
dose adjustment
- ODDC (Object Detected Dose Control) that can make
automatic dose adjustment by detecting object and
motion
- Anatomically Programmed Fluoroscopy (APF) control
system
- A feature that automatically adjusts contrast and
brightness and ensure the production of sharp, low-dose
images (IDEAL: Intelligent Dose Efficiency Algorithm)
In order to increase the number of
qualified competitors we kindly
request you to remove this item from
the Technical Specifications.
The item has been revised as:
The system must have at least one of the
following special software that improves the
image quality. Companies will show this
feature in their original catalogues:
- EASY (Enhanced Acquisition System) with
automatic dose adjustment
- ODDC (Object Detected Dose Control) that
can make automatic dose adjustment by
detecting object and motion
- Anatomically Programmed Fluoroscopy
(APF) control system
- A feature that automatically adjusts contrast
and brightness and ensure the production of
sharp, low-dose images (IDEAL: Intelligent
Dose Efficiency Algorithm)
- the filter which reduces motion blur by
detecting dynamic motion by means of
motion adaptation and double leaf collimator
which can function asymmetrically
- DDC (Dynamic Density Compensation)
which provides the optimum images to be
obtained by automatic dose adjustment
according to the anatomical features of
each different patient.
90
In order to increase the number of
qualified competitors we kindly
request you to add the following to the
item “DDC (Dynamic Density
Compensation) which provides the
optimum images to be obtained by
automatic dose adjustment
according to the anatomical features
of each different patient.”.
![Page 31: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/31.jpg)
Page 31 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
Please also refer to Changes No.2 to Tender
Dossier.
91 TS 3.13.2
The image amplifier section will have a laser or lighted
centering device.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The Flat Detector section
will have a laser or lighted centering
device.”.
The specification remains unchanged.
92
TS 3.13.3 There will be both an iris and a parallel layer collimator
in the system.
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “There will be both an iris and
a parallel layer or slot diaphragm
collimator in the system.”.
The item has been revised as:
There will be both an iris and a parallel layer
collimator or equivalent technology in the
system.
Please also refer to Changes No.2 to Tender
Dossier. 93
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “There will be both an iris and
a parallel layer or symmetric and
asymmetric collimator in the system.
![Page 32: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/32.jpg)
Page 32 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
94
TS 3.13.5 Collimator DQE rate should be at least %65.
The terminology of DQE means
“Detective Quantum Efficiency” and,
it is an Image Intensifier related
terminology. Therefore, the
amendment request is for clarifying the
specification wording and the
meaning. Please also note that our
Image Intensifier has 65% DQE rate
so, we will comply with the technical
specifications, if the wording and the
meaning can be clarified. The item is
requested to be revised as follows
“Image Intensifier DQE rate should
be at least %65.”. The specification has been removed from the
Technical Specifications.
Please also refer to Changes No.2 to Tender
Dossier.
95
Detective quantum efficiency (DQE) is
one of the fundamental physical
variables related to image quality in
radiography and refers to the
efficiency of a detector in converting
incident x-ray energy into an image
signal.
The words "quantum efficiency" have
a precise meaning, because the DQE
measures the quantum efficiency of an
equivalent virtual detector, which
produces at its output the same signal-
to-noise ratio (SNR) of the real
detector even when the input signal is
the same.
The DQE is generally defined by the
ratio of the squared output signal-to-
![Page 33: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/33.jpg)
Page 33 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
noise ratio (SNRo)2 to the squared
input signal-to-noise ratio (SNRi)2 of
the imaging detector.
DQE is dependent on radiation
exposure, spatial frequency, MTF, and
detector material. The quality (voltage
and current) of the radiation applied is
also an important influence on DQE.
DQE is therefore directly proportional
to the MTF of a detector; as an
imaging system, however, adds noise
to the output information, the DQE is
also in relation to SNR, with inverse
proportionality. Thus, the DQE
measures the SNR and MTF at various
spatial frequencies.
Therefore, we request this
specification to cover the “image
intensifier” not the collimator or to be
removed from the specifications.
96 We kindly request you to change the
clause as follows “Image intensifier
DOE rate should be at least %60.”.
97 We kindly request you to change the
clause as follows “Flat Detector DOE
rate should be at least %65.”.
98 TS 3.14 Image Amplifier
We kindly request the item to be
revised as “Flat Detector”.
The specification remains unchanged
considering the needs of the Ministry.
![Page 34: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/34.jpg)
Page 34 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
99
TS 3.14.1 The diameter of image amplifier must be at least 12
inches.
We kindly request you to change the
clause as follows “The diameter of
image amplifier must be at least 9
inches.”. The specification remains unchanged
considering the needs of the Ministry.
100
We kindly request you to change the
clause as follows “The size of Flat
Detector must be at least 29 cm x 29
cm.”.
101
TS 3.14.2 Images should be processed in at least 32 bits.
32 bit image processing is available on
PC used for post processing. On the
other hand, this requirement is located
in Image Amplifier section. However,
Image Intensifiers have 16 bit image
processing in general. To comply with
the technical specifications and
participation in the tender, it is kindly
requested to revise the item as follows
“Images should be processed in at least
16 bits.”.
The item has been revised as:
Images should be processed in at least 16
bits.
Please also refer to Changes No.2 to Tender
Dossier.
102
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “Images should be processed
in at least 16 bits for each monitor.”.
![Page 35: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/35.jpg)
Page 35 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
103
With the state of the art technology in
medical field, grayscale images
represented by max 16bit which leads
to 65536 different grayscales, which is
beyond human perception. So
maximum available bit size is 16bit,
therefore maximum possible image
processing for grayscale images 16bit.
But because of the attainable DQE
(detective quantum efficiency ) with
the current technology, meaningful
processed bit should be at least 12bit-
which is 4096 grayscale which is still
not detectable difference by the most
human. Therefore, we kindly request
you to change the clause as follows
“Images should be processed in at
least 12 bits.”.
104
In order to increase the number of
qualified competitors we kindly
request you to revise the item as
“Images should be processed in at least
14 bits.”.
105 TS 3.14.3
There will be at least 1K X 1K CCD high resolution
camera in the device.
We kindly request you to revise the
item as “There will be at least 1,5K X
1,5K Flat Detector in the device.”.
The item has been revised as:
There will be at least 1K X 1K CCD or
CMOS high resolution camera in the device.
Please also refer to Changes No.2 to Tender
Dossier.
We kindly request you to revise the
item as “There will be at least 1K X
1K CCD or CMOS high resolution
camera in the device.”.
106 TS 3.14.4 The image amplifier must have at least 3 areas.
In order to increase the number of
qualified competitors we kindly
request you to revise the item as “The
The specification remains unchanged
considering the needs of the Ministry.
![Page 36: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/36.jpg)
Page 36 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
image amplifier must have at least 2
areas”.
107 We kindly request you to revise the
item as “The Flat Detector must have
at least 3 areas.”.
108 TS 3.15.2
The memory capacity of the device must be at least
30.000 images.
We kindly request you to revise the
item as “The memory capacity of the
device must be at least 15.000
images.”.
The specification remains unchanged
considering the needs of the Ministry.
109 TS 3.15.4
There should be special software in the device which
improves the image quality. There should be ODDC
(Object Detected Dose Control) or EASY (Enhanced
Acquisition System) which can make automatic dose
adjustment by detecting object and motion or the filter
which reduces motion blur by detecting dynamic motion
by means of motion adaptation and double leaf
collimator which can function asymmetrically or the
property (IDEAL= Intelligent Dose Efficiency
Algorithm) which enables the formation sharp images
with low dose by automatically setting the contrast and
brightness in the device. The companies shall indicate
this property in original catalogues.
In order to increase the number of
qualified competitors we kindly
request you to remove this item from
the Technical Specifications.
The item has been revised as:
The system must have at least one of the
following special software that improves the
image quality. Companies will show this
feature in their original catalogues:
- EASY (Enhanced Acquisition System) with
automatic dose adjustment
- ODDC (Object Detected Dose Control) that
can make automatic dose adjustment by
detecting object and motion
![Page 37: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/37.jpg)
Page 37 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
110
In order to increase the number of
qualified competitors we kindly
request you to revise the item as
“There should be special software in
the device which improves the image
quality. There should be ODDC
(Object Detected Dose Control) or
EASY (Enhanced Acquisition System)
which can make automatic dose
adjustment by detecting object and
motion or the filter which reduces
motion blur by detecting dynamic
motion by means of motion adaptation
and double leaf collimator which can
function asymmetrically or the
property (IDEAL= Intelligent Dose
Efficiency Algorithm) which enables
the formation sharp images with low
dose by automatically setting the
contrast and brightness in the device
or DDC (Dynamic Density
Compensation or companies specify
their own similar software and / or
hardware features. The companies
shall indicate this property in original
catalogues.”.
- A feature that automatically adjusts contrast
and brightness and ensure the production of
sharp, low-dose images (IDEAL: Intelligent
Dose Efficiency Algorithm)
- the filter which reduces motion blur by
detecting dynamic motion by means of
motion adaptation and double leaf collimator
which can function asymmetrically
- DDC (Dynamic Density Compensation)
which provides the optimum images to be
obtained by automatic dose adjustment
according to the anatomical features of each
different patient.
Please also refer to Changes No.2 to Tender
Dossier.
![Page 38: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/38.jpg)
Page 38 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
111
We kindly request you to revise the
item as “There must be special
software in the system that Improves
the Image quality. There should be
ODDC (Object Detected Dose
Control) which can adjust dose
automatically by perceiving object and
motion,EASY (Enhanced Acquisition
System) or the property which
generates sharp and low dose images
by automatically adjusting the contrast
and brightness (IDEAL: Intelligent
Dose Efficiency Algorithm) or the
filter which reduces the motion blur by
perceiving the dynamic motion due to
motion adaptation and asymmetrical
double-leaf collimator. Companies will
show these features in their original
catalogs.”.
112 TS 3.15.8 The image should be converted to negative and can be
enlarged.
In order to increase the number of
qualified competitors we kindly
request you to revise the item as “The
image should be able to be
enlarged.”.
The specification remains unchanged
considering the needs of the Ministry.
113 TS 3.15.16 Optical Magnification
We kindly request you to revise the
item as “Optical Magnification or
Detector Zoom”.
The specification remains unchanged
considering the needs of the Ministry.
![Page 39: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/39.jpg)
Page 39 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
114
TS 3.15.18 The UPS will be given to supply the system's
workstation for at least 10 minutes.
The C-Arm systems have to be used in
operating room where is totally
covered by a central UPS system
specifically. Using an additional
standalone UPS system will cause
contamination in the hospital which is
totally against the International
regulations. Besides, some devices
have “Linux” based operating systems
and because of that unique feature,
there is no loss of image in case of loss
of electricity. Therefore, we kindly
request the item to be revised as “The
UPS will be given to supply the
system's workstation for at least 10
minutes or captured images can be
saved automatically without the need
of UPS.”.
The item has been revised as:
There should be an external UPS or an
integrated UPS to supply the system's
workstation for at least 10 minutes.
Please also refer to Changes No.2 to Tender
Dossier.
115
In order to increase the number of
qualified competitors we kindly
request you to revise the item as “The
UPS will be given to supply the
system's workstation for at least 10
minutes or there should have an
integrated UPS that can feed the
workstation of the system.”.
116 TS 3.15.20
Multi-modality surgical workstation imaging or pulsed
fluoroscopy should be able to reach at least 175 mA to
reach high level modality in the system.
Multi-modality surgical workstation is
available with the system. However,
mA is measured only for X-ray tube
and not for the multi-modality
workstation. In addition to that, max.
mA values of continuous fluoroscopy,
pulsed fluoroscopy and digital
Based on the needs of the Ministry, the
specification remains unchanged.
![Page 40: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/40.jpg)
Page 40 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
exposure mode are defined and
requested at the specification items of
3.11.3., 3.11.4 and 3.11.5. To comply
with the technical specifications and
participation in the tender, it is kindly
requested to revise the item as “Multi-
modality surgical workstation should
be available with the system.”.
117
In order to increase the number of
qualified competitors we kindly
request you to remove this item from
the Technical Specifications.
118
To comply with the technical
specifications and participation in the
tender, it is kindly requested to revise
the item as “Multi-modality surgical
workstation imaging or pulsed
fluoroscopy should be able to reach at
least 24 mA to reach high level
modality in the system.
119
We kindly request the item to be
revised as “Multi-modality surgical
workstation imaging or pulsed
fluoroscopy should be able to reach at
least 8 mA to reach high level
modality In the system.”.
120 TS 3.15.21
The workstation unit in the device must have at least 1.5
processors, at least 4 GB of RAM, and at least 2 USB 3.0
compatible drives.
Our system has 4 USB 2.0 ports with
the system. To comply with the
technical specifications and
participation in the tender, it is kindly
requested to revise the item as “The
workstation unit in the device must
have at least 1.5 processors, at least 4
GB of RAM, and at least 2 USB 2.0
Based on the needs of the Ministry, the
specification remains unchanged.
![Page 41: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/41.jpg)
Page 41 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
compatible drives.”.
121
Comparing the “Windows” based
systems, “Linux” systems are
specifically optimized for medical
devices according to the needs of the
device and they work with higher
performance without needing more
RAM or processors. Therefore, we
kindly request the item to be revised as
“The workstation unit in the device
must have at least 1.5 processors, at
least 4 GB of RAM, and at least 2
USB 3.0 compatible drives Tech or
unit should have Linux operating
system.”
122
We kindly request the item to be
revised as “The workstation unit in the
device must have at least 1.5
processors, at least 4 GB of RAM, and
at least 2 USB drives.”.
123
We kindly request the item to be
revised as “The workstation unit in the
device must have at least 1.5
processors, at least 4 GB of RAM,and
at least 1 USB 3.0 compatible drives.
![Page 42: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/42.jpg)
Page 42 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
124 TS 3.16.2
There should be at least two 18 '' diagonal sized LCD or
TFT monitors in the system.
The brightness of the LCD monitor is
an important feature. Therefore, we
kindly request the item to be revised as
“There should be at least two 18 ''
diagonal sized LCD monitors in the
system and the brightness should be
at least 500 cd/m2”.
Based on the needs of the Ministry, the
specification remains unchanged.
125
TS 3.16.3
Monitors should be on a separate transport stand, if
needed only the monitors should be rotated at 180°
without moving the tripod or the live image should be
rotated up to 180° in total with the touch control panel
displayed and they should be able to tilt.
We kindly request the item to be
revised as “Monitors should be on a
separate transport stand or the
monitors of the device must be able
close on each other.”.
The item has been revised as:
Monitors should be on a separate transport
stand, if needed only the monitors should be
rotated at 180° without moving the tripod or
the live image should be rotated up to 180° in
total with the control panel displayed and
they should be able to tilt.
Please also refer to Changes No.2 to Tender
Dossier.
126
We kindly request the item to be
revised as “Monitors should be on a
separate transport stand if needed only
the monitors should be able to rotate
up to 180 degree in total and the
height of the monitors should be
adjustable .”.
127
We kindly request the item to be
revised as “Monitors should be on a
separate transport stand, only the
monitors should be rotated 180 degrees
without moving the tripod and should
be able to tilt or the live image should
be rotated up to 180 degrees in total
with the touch control paneldisplayed
and the monitors shall have a viewing
angle of 160 degrees (vertical and
horizontal).”.
![Page 43: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/43.jpg)
Page 43 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
128
TS 3.16.5
Patient information, cumulative dose, kV and mA values
should be monitored on the test monitor at the touch
mobile imaging station during the procedure.
We kindly request the item to be
revised as “Patient information,
cumulative dose, kV and mA values
should be monitored from the
monitor on the mobile imaging
station during the procedure.”.
The item has been revised as:
Patient information, cumulative dose, kV and
mA values should be monitored on the test
monitor at the mobile imaging station during
the procedure.
Please also refer to Changes No.2 to Tender
Dossier.
129
We kindly request the item to be
revised as “Patient
Information,cumulative dose,kV and
mA values should be monitored on the
display monitor mounted on a stand-
alone stand or on a stand-up bracket
separate from the C-arm system during
the procedure.”.
130 TS 3.17.1
2 foot pedal and 2 hand buttons will be delivered with
the device.
Foot pedal is integrated with the
system so, another second foot pedal
will not have any function for the
system. In addition to that, one foot
pedal and one hand button are already
requested at the specification item of
3.11.6. On the other hand, the system
configuration includes one foot pedal
only. Optionally, companies can also
provide remote control which also
helps user reviewing acquired images
and parking them on reference, also
magnification can be changed without
touching the system. Therefore, it
helps standard workflow very
efficiently. Please see what other
functions it has: Run loop, Overview
run/exam, Retrieve previous
image/run, Retrieve next image/run,
Park image on Reference monitor,
Based on the needs of the Ministry, the
specification remains unchanged.
![Page 44: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/44.jpg)
Page 44 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
Retrieve image from Reference
monitor, Protect image/release image,
SmartMask, Fluoroscopy mode
selection, II-format selection,
Subtraction on/off, Image grab. To use
public interest properly, it is kindly
requested to revise the item as “1 foot
pedal and 1 hand buttons and 1
wireless remote control will be
delivered with the device.”.
131 It is kindly requested to revise the item
as “1 foot pedal and 1 hand buttons
will be delivered with the device.”.
132
We kindly request the item to be
revised as “1 foot pedal and 2 hand
buttons will be delivered with the
device.” given such revision does not
make any clinical and functional
restriction.
133 TS 3.17.2
Radiation shielding with the following features and
amounts shall be delivered.
- 4 skirt-vest lead aprons which are easy to wear
(with touch fasteners on the shoulder) in sizes
of 1 large, 1 small and 2 medium sizes, which
don’t contain lead, the weight of which will be
at most 10 kg in 0.5 mm Pb equivalent or at
most 4 kg in 0,5 mm Pb equivalent.
- 1 adult male, 1 adult female and 1 pediatric
gonad protectors which do not contain lead in
0,5 Pb equivalent or equal to 0,5 mm Pb
- 2 adult, 1 pediatric thyroid protectors which do
not contain lead in in 0,5 Pb equivalent or equal
to 0,5 mm Pb equivalent.
To comply with the technical
specifications via our authorized
supplier, it is kindly requested to revise
the first bullet of the item as
“- 4 skirt-vest lead aprons which are
easy to wear (side adhesive instead
of shoulder), 1 large, 1 small and 2
medium sizes, which don’t contain
lead, the weight of these aprons
will be at most 8 kg in 0.5 mm Pb
equivalent.”.
The specification remains unchanged.
![Page 45: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/45.jpg)
Page 45 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
134 TS 3.18
Vertical radiation scattering for patient and user safety
when the system is in the upright position (image
amplifier at the top, bottom of tube) with a standard 20-
25 cm water conjugated acrylic phantom at 70-80 kV
voltage, at 2-4 mA current or at maximum power
according to IEC standard 30 cm from image amplifier
should not exceed 2.0 mGy/hour within the range 0 cm
to 50 cm or 1 mGy / h within 50 cm-100 cm range.
When the image amplifier is at the top and the tube is at
the bottom, 2.3 mGy/hour shouldn’t be exceeded when
the image amplifier 10 cm above the ground in 110 kV
voltage and 3 mA current. This measurement shall be
reported by the A-type inspection body of the Turkish
Accreditation Agency to the ISO 17020 standard
independent of the contractor following the installation
of the instrument.
The manufacturer companies have ISO
17025 standard for C-arm systems. To
comply with the technical
specifications and participation in the
tender, it is kindly requested to revise
the item as “Vertical radiation
scattering for patient and user safety
when the system is in the upright
position (image amplifier at the top,
bottom of tube) with a standard 20-25
cm water conjugated acrylic phantom
at 70-80 kV voltage, at 2-4 mA current
or at maximum power according to
IEC standard 30 cm from image
amplifier should not exceed 2.0
mGy/hour within the range 0 cm to 50
cm or 1 mGy / h within 50 cm-100 cm
range or at 120 kV voltage, at 2-4
mA current or at maximum power
according to IEC standard 25 cm
from image amplifier should not
exceed 10.0 mGy/hour within the
range 0 cm to 50 cm or 2 mGy / h
within 50 cm-100 cm range or when
the image amplifier is at the top and
the tube is at the bottom, 2.3 mGy/hour
shouldn’t be exceeded when the image
amplifier 10 cm above the ground in
110 kV voltage and 3 mA current. This
measurement shall be reported by the
A-type inspection body of the Turkish
Accreditation Agency to the ISO
17020 or ISO 17025 standard
independent of the contractor
The specification remains unchanged.
![Page 46: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/46.jpg)
Page 46 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
following the installation of the
instrument.”.
![Page 47: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/47.jpg)
Page 47 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
135
We kindly request the item to be
revised as “When the system is in the
upright position (the image intensifier
is at the top and the tube is at the
bottom) the vertical radiation
scattering shall not exceed 2.0 mGy/h
at 0 cm - 50 cm or 1.0 mGy/h at 50 cm
- 100 cm at shooting at a distance of
30 cm away from the image intensifier
and at maximum power according to
the IEC standard or with standard 20-
25 cm water-conjugated phantom, 70-
80 kV voltage and 2-4 mA current or
when the image intensifier at the top,
tube is at the bottom, the image
intensifier shall not exceed 2.3 mGy/
hour at the 110 kV voltage , 3 mA
current, and at 10 cm height from the
flor or when the tube at the top, the
image intensifier is at the bottom or
when the tube at the bottom, the image
intensifier is at the top , the image
intensifier shall not exceed 2 mGy/
hour at the 110 kV voltage, 5,3 mA
current, and at 10 cm height from the
floor.”.
![Page 48: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/48.jpg)
Page 48 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
136
We kindly request the item to be
revised as “Vertical radiation
scattering for patient and user safety
when the system is in the upright
position (Flat Detector at the top,
bottom of tube) with a standard 20-25
cm water conjugated acrylic phantom
at 70-80 kV voltage, at 2-4 mA current
or at maximum power according to
IEC standard 30 cm from Flat
Detector should not exceed 2.0
mGy/hour within the range 0 cm to 50
cm or 1 mGy/ h within 50 cm-100 cm
range or when the Flat Detector is at
the top and the tube is at the bottom,
2.3 mGy/hour shouldn't be exceeded
when the Flat Detector 10 cm above
the ground in 110 kV voltage and 3
mA current or when the system is in
the upright position (Flat detector at
the top, bottom of tube) fluoroscopy
shooting at a current of 20 mA at a
voltage of 125 kV, at a rate of 15 p/ s
(at 18x18cm) the tube have a height
of 10 cm should not exceed 2.5 Mg/
hour.
137 TS 3.19
The device and all external accessories must be
registered in the Product Tracking System.
The personal protective equipment is
regulated by 89/686 EEC Personal
Protective Equipment Regulations.
Therefore, there is no X-ray protective
apron, gonad protector, thyroid
protector etc. in the market which are
registered to Product Tracking System.
Therefore, we kindly request the item
Please be aware that the UPS’ are not
considered as accessories within the scope of
this article. Therefore, the item will remain
same.
Please be informed that all the medical
devices and accessories required to be
registered to Product Tracking System
![Page 49: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/49.jpg)
Page 49 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
to be revised as “The device must be
registered in the Product Tracking
System.”.
according to the 93/42 EEC Medical Device
Regulations.
138 TS 3.20
The devices to be delivered by the contractor shall have
at least 5 (five) years of warrantee period after they are
accepted. The contractor is responsible for arranging the
warranty documents belonging to these devices in the
name of the administration and presenting their original
copies to the administration. In case it is not possible to
arrange warranty documents of the received devices in
the name of the administration, the Contractor is obliged
to present a document containing warranty undertakings
to the administration. The contractor shall undertake the
removal of the faults, defects and deficiencies
determined in the device under the scope of warranty
within the duration of the contract by the warranty
provider person or institution. In case this obligation is
not fulfilled by the contractor, the administration shall
deduct all cost for the provision of the warranty from the
receivables of the contractor or collect the costs by
forfeituring.
5 (five) years of warranty is demanded.
5 (five) year warranty can be given to
the devices. But after 2 (two) year
warranty period which is standard,
manufacturers is charging extra fee for
each year. In this case, we as
representatives have to add the price
difference arising from the extra
warranty period to our tender offer.
When the warranty period is 5 (five)
years, offered prices to the tender will
be higher than expected and market
price. After the standard 2 (two) year
warranty period of the devices, it will
be more advantageous for the user
institutions to make an annual
maintenance-repair agreement with the
authorized service of device.
In the item no 3.34 of the technical
specifications, the annual maintenance
and repair conditions (3% excluding
spare parts, 5% including spare parts)
is specified; the signing of a
maintenance-repair contract with these
conditions will be more effective and
in the interest of the institutions as the
service, authority brings the company
directly to the service provider.
The specification remains unchanged.
![Page 50: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/50.jpg)
Page 50 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
139 TS 3.21
During the warranty period, the seller shall remove the
defects of the manufacturing, assembly, material and
workmanship and the uncorrected defective parts shall
be replaced by a new one. During the warranty period,
there will be no charge for maintenance, repair and spare
parts.
We kindly request the item to be
revised as “During the warranty
period, the seller shall remove the
defects of the manufacturing,
assembly, material and workmanship
and the uncorrected defective parts
shall be replaced by a new one. During
the warranty period, there will be no
charge for maintenance, repair and
spare parts (excluding usage errors,
environmental conditions related
errors).
The specification remains unchanged.
140 TS 3.22
During the warranty period, all software updates,
upgrades and re-installations will be made free of charge
by the contractor. The Contractor shall digitally deliver a
copy of all image processing, operating and service
software to be used in the system. The installation and
access limitation of these systems shall not be applied.
For the sake of terminological usage of
the terms, we kindly request the item
to be revised as “During the warranty
period, all software updates will be
made free of charge by the contractor.
The Contractor shall deliver together
with the system all image processing,
operating and service software to be
used in the system. The installation
and access limitation of these systems
shall not be applied.”.
The specification remains unchanged.
141 TS 3.23
The Contractor shall carry out protective / preventive
maintenance and calibrations of the device and all
components (including accessories) at least four (4)
times a year for three (3) month periods in place.
Maintenance with remote connection will not be
accepted.
To ensure that the requirement is in
line with the period recommended by
the manufacturer, we kindly request
the item to be revised as “The
Contractor shall carry out
protective/preventive maintenance and
calibrations of the device and all
components (including accessories) at
least once (1) a year for in place.
Maintenance with remote connection
will not be accepted.
The specification remains unchanged.
![Page 51: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/51.jpg)
Page 51 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
142 TS 3.24
The Contractor shall perform periodic metrology
(testing, control, calibration, inspection, validation)
activities of the device and all components (including
accessories) during the warranty period, at least two (2)
times in six (6) month periods per year in accordance
with the Circular on Property and Service Purchase
Regarding Medical Devices and the Regulation on
Testing, Control and Calibration of Medical Devices and
Biomedical Metrology Activities Manual. The
Contractor shall not charge any additional fee for these
services during the warranty period.
The removal of this item is kindly
requested. The specification remains unchanged.
143 TS 3.25
Intervention time is at most 36 hours starting from the
date of fault notification. This duration starts when the
device malfunction is notified to the contractor or
authorized service by the related health facility or
affiliated institution. When a fault notification about the
device is made to the technical service, the device shall
be delivered in working condition within 3 working days
in case spare parts are not needed or within 5 working
days in case spare parts are needed. In case of spare part
change requires import permission, this duration shall
not exceed 20 working days after the intervention.
We kindly request you to change the
clause as follows “Intervention time is
at most 24 hours within official
working hours starting from the date of
fault notification . This duration starts
when the device malfunction is
notified to the contractor or authorized
service by the related health facility or
affiliated institution. When a fault
notification about the device is made
to the technical service, the device
shall be delivered in working condition
within 3 working days in case spare
parts are not needed or within 5
working days in case spare parts are
needed. In case of spare part change
requires import permission, this
duration shall not exceed 20 working
days after the intervention.”.
The specification remains unchanged.
144 TS 3.26
If the device fault cannot be removed within 5 working
days after the intervention, the contractor provides
another device with similar properties to the
administration until the end of the repair process.
We kindly request you to change the
clause as follows “If the device fault
cannot be removed within 5 working
days after the intervention, warranty
period of the corresponding system
The specification remains unchanged.
![Page 52: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/52.jpg)
Page 52 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
should be extended by 1(one) calendar
day. The maximum extension is
limited to 45 days.”.
145 The article in question is requested to
be removed from the Technical
Specification.
146
TS 3.27
The device will be guaranteed a 95% uptime per year
during the warranty period. The device will be counted
down if the patient is not taken completely after the
initial intervention. If the 5% down time is exceeded, an
administrative fine shall be imposed in the amount of the
multiplication of the average of the total number of
transactions for the last 10 days of the device and the
maximum SUT transaction price for each working day.
We kindly request you to change the
clause as follows “The device will be
guaranteed a 95% uptime per year
during the warranty period. The device
will be counted down if the patient is
not taken completely after the initial
intervention . If the 5% down time is
exceeded, for each day of additional
down-warranty period of the
corresponding system should be
extended by 1(one) calendar day. The
maximum extension is limited to 45
days.”. The specification remains unchanged.
147
We kindly request you to change the
clause as follows “The device will be
guaranteed a 95% uptime per year
during the warranty period.The device
will be counted down if the patientIs
not taken completely after the initial
intervention. If the defective time
exceeds 5%,the warranty period will
be added free of charge for 2 days per
day.”.
![Page 53: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/53.jpg)
Page 53 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
148 TS 3.31
Before the contract is signed, the contractor is obliged to
present the price list of all spare parts, accessories and
expendables including lifed parts with no exceptions in
the tender offer file, which will not exceed %150 of the
unit device price. All parts, the types or amounts of
which are not included in the price list of all spare parts,
accessories and expendables including lifed parts with no
exceptions by the contractor shall be provided if the
administration demands free of any charges.
We kindly request you to change the
clause as follows “Before the contract
is signed, the contractor is obliged to
present the price list of all spare parts,
accessories and expendables including
lifed parts with no exceptions in the
tender offer file, which will not exceed
%200 of the unit device price. All
parts, the types or amounts of which
are not included in the price list of all
spare parts, accessories and
expendables including lifed parts with
no exceptions by the contractor shall
be provided if the administration
demands free of any charges. In the
period covered by the commitment
(during the warranty period and for 5
years thereafter) in case of
discontinued I changed I new spare
parts and the condition is documented
by the manufacturer; the price of the
parts offered with the new codes shall
not be included in the first list of spare
parts provided, but shall be considered
by the Administration.”.
The specification remains unchanged.
149 TS 3.32
Being under the %150 amount determined, all parts need
for the function of the device which are not included in
the list of all spare parts, accessories and expendables
including lifed parts with no exceptions shall be met by
the companies free of charge. The contractor company
shall deliver a list of all spare parts, accessories and
expendables including lifed parts with no exceptions to
all health facilities to which the device is delivered
before the contract is signed. Within the 5 years after the
warranty, if the health facilities demand spare parts,
We kindly request you to change the
clause as follows “Being under the
%200 amount determined , all parts
need for the function of the device
which are not included in the list of all
spare parts, accessories and
expendables including lifed parts with
no exceptions shall be met by the
companies free of charge. The
The specification remains unchanged.
![Page 54: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/54.jpg)
Page 54 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
accessories and expendables including lifed parts with no
exceptions, the demands of the health facilities shall be
met on condition that the financial amount defined in this
list being the upper limit. Sales cannot be offered by the
contractor by demanding any price or additional price
under the name of exception or out of coverage.
contractor company shall deliver a list
of all spare parts including lifed parts
with no exceptions to all health
facilities to which the device is
delivered before the contract is signed.
Within the 5 years after the warranty ,
if the health facilities demand spare
parts, accessories and expendables
including lifed parts with no
exceptions, the demands of the health
facilities shall be met on condition that
the financial amount defined in this list
being the upper limit. Sales cannot be
offered by the contractor by
demanding any price or additional
price under the name of exception or
out of coverage. In the period covered
by the commitment (during the
warranty period and for 5 years
thereafter) in case of
discontinued/changed/new spare parts
and the condition is documented by the
manufacturer; the price of the parts
offered with the new codes shall not be
included in the first list of spare parts
provided, but shall be considered by
the Administration.”.
![Page 55: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/55.jpg)
Page 55 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
150
TS 3.34
If the hospital administration wants to make a
maintenance and repair contract after the warranty, it can
demand a contract;
- At the rate of %3 of the annual unit device cost
excluding the spare parts at most,
- At the rate of %5 of the unit device cost
including all spare parts at most. The contractor
company meets this demand immediately.
We kindly request you to change the
clause as follows “If the hospital
administration wants to make a
maintenance and repair contract after
the warranty, it can demand a contract;
- At the rate of %4 of the annual
unit device cost excluding the
spare parts at most,
- At the rate of %8 (excluding
tubes & detector) of the unit
device cost including all spare
parts at most. The contractor
company meets this demand
immediately . The repair and
spare part requirements arising
from user errors shall be excluded
from the scope specified. The specification remains unchanged.
151
We request exclusion of the spare parts
in imaging systems that do not fail
frequently and the sales price of which
accounts to a major portion of the
system's sales price, as exemplified in
the specifications of public tenders
published before. We kindly request
you to change the clause as follows “If
the hospital administration wants to
make a maintenance and repair
contract after the warranty, It can
demand a contract;
- At the rate of %3 of the annual
unit device cost excluding the
spare parts at most,
- At the rate of %6 of the unit
![Page 56: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/56.jpg)
Page 56 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
device cost including all spare
parts (excluding tube and
detector) at most. The contractor
company meets this demand
Immediately.
152
TS 3.38
The device must have a heat exchanger system to save
heat.
Could you consider moving the item
under the main title “A.
GENERAL DESCRIPTION OF
THE SYSTEM” as the requirement is
technical.
The specification has been moved to TS
3.15.22.
Please be informed that it is finely confirmed
by the Ministry that “heat exchanger” feature
does not necessarily addresses only one
brand.
153
The desired “heat exchanger” feature
is only available on one brand.
Therefore, the other specifications
requested in the technical
specifications become meaningless and
only one brand device can participate
in the tender with this exiting
specification. Therefore we kindly
request the removal of this
specification.
154
In order to increase the number of
qualified competitors we kindly
request you to change the clause as
follows “The device must have a heat
exchanger system to save heat or
“ILX” (Integrated Liquid
CoolingSystem) or similar systems.”.
155 TS 3.39
The data on the display of the device should be in
Turkish or the buttons should be expressed as symbols.
Could you consider moving the item
under the main title “A.
GENERAL DESCRIPTION OF
THE SYSTEM” as the requirement is
technical.
The current place of the specification deemed
sufficient.
![Page 57: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/57.jpg)
Page 57 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
156 TS 3.41
The device must be capable of self-test. The test program
should detect the current fault and calibration errors in
the system, alert the users, and cover all electronic,
electromechanical and hydromechanical sections.
Could you consider moving the item
under the main title “A.
GENERAL DESCRIPTION OF
THE SYSTEM” as the requirement is
technical.
The current place of the specification deemed
sufficient.
157 TS 3.45
The hospital administration shall provide the electrical
system up to the instrument panel, the panel of the
device and the next system shall be installed by the
contractor company.
We kindly request the removal of this
item. The specification remains unchanged.
158 TS 3.46
During the installation, necessary ceiling, cabling on the
floor walls, channel operations, lighting of the device
room and control room and other decoration works will
be done by the contractor.
C-arm x-ray devices are often used in
operating rooms. Since they are mobile
type devices and they need to be
moved to different operating rooms for
different operations, they do not need
cabling, channel operations, device
room, control room lighting and other
decoration works in ceiling and floor
walls. Therefore, we kindly request the
removal of this item from the technical
specifications.
The specification has been revised as:
During the installation, necessary ceiling,
lead coating of the room (in line with the
requirements of TAEK), cabling on the floor
walls, channel operations, lighting of the
device room and control room and other
decoration works will be done by the
contractor.
Please also refer to the Changes No.2 to TD.
159 TS 3.47
The appropriate table and 2 rotating type adjustable seats
will be provided for the X-ray control room where the
system will be installed.
These furnitures are used in mostly
Control Rooms of Imaging Systems.
However, C-arm systems are used in
Operating Rooms so, there is not any
Control Room. On the other hand, C-
arm systems are mobile systems and
not installed in any specific room. To
use public interest properly, it is kindly
requested the item to be removed from
the technical specifications.
The specification remains unchanged.
![Page 58: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/58.jpg)
Page 58 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
C-arm x-ray devices are generally used
in operating rooms and they are mobile
type devices so they do not need a
suitable table for the control room and
a swivel type adjustable seat.
Therefore, the removal of the item
from the technical specification is
kindly requested.
160
TS 3.48
24,000 BTU power split air conditioner will be supplied
by the contractor for cooling the place where the devices
will be installed and will be delivered in working
condition.
Split air conditioners are used in
mostly Control Rooms and Technical
Rooms of Imaging Systems. However,
C-arm systems are used in Operating
Rooms so, there is not any Control
Room and Technical Room. In
addition to that, Operating Rooms
have their own laminar flow system
for ventilation in a sterile way. So,
standard split air conditioners might
affect Operating Rooms sterility
negatively. To use public interest
properly, it is kindly requested the item
to be removed from the technical
specifications.
The specification remains unchanged.
161
Today, in every hospital, there is a
modern air conditioning infrastructure
to keep the ambient temperature at the
desired level. So the second one is
unnecessary. At the same time, an
extra air conditioner will raise the offer
price. Therefore, we kindly request the
item to be removed from the technical
specification.
![Page 59: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/59.jpg)
Page 59 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
TS 3.52
During the acceptance and inspection, when the
companies are asked for the tests regarding the technical
characteristics and performance of the device, the
necessary personnel and equipment will be provided free
of charge by the companies. The seller is responsible for
accidents and damages that may occur during the
acceptance and inspection.
We kindly request you to change the
clause as follows “During the
acceptance and inspection, when the
companies are asked for the tests
regarding the technical characteristics
and performance of the device, the
necessary personnel and equipment
will be provided free of charge by the
companies. The seller is responsible
for accidents and damages caused by
its own personnel that may occur
during the acceptance and inspection.”.
162 TS 3.53
The necessary trainings are provided by the Application
Specialists for at least three (3) days to use all the
functions of the system. These trainings will be given a
total of three times during the warranty period.
Application Experts should also be certified by TCESIS
Clinical Support Staff. These documents must be
notarized and presented in the tender dossier.
We kindly request you to change the
clause as follows “The necessary
trainings are provided by the
Application Specialists for at least
three (3) days to use all the functions
of the system. These trainings will be
repeated once more free of charge if
requested during the warranty period.
Application Experts should also be
certified by TCESIS Clinical Support
Staff. These documents must be
notarized and presented in the tender
dossier.”.
Considering the needs of the Ministry the
request has been found unacceptable.
163 TS 3.54
Training for at least three (3) days for basic maintenance,
first level malfunction detection and repair (fault codes
and interpretation, foot pedal and hand button repair,
software settings) shall be provided to carry out the tasks
and operations specified in the service manual for all
functions of the system. These trainings will be given a
total of three times during the warranty period. Technical
Service Specialists must have the training certificate
issued by the manufacturer. These documents must be
We kindly request you to change the
clause as follows “Training for at least
three (3) days for determining the use,
basic maintenance, first level
malfunction detection shall be
provided to carry out the tasks and
operations specified in the service
manual for all functions of the system.
These trainings will repeated once
Considering the needs of the Ministry the
requests have been found unacceptable.
![Page 60: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/60.jpg)
Page 60 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
notarized and presented in the tender dossier. more free of charge if requested during
the warranty period. Application
Specialists must have the training
certificate issued by the manufacturer.
These documents must be notarized
and presented in the tender dossier.”.
164
It is very important to carry out
maintenance, repair and failure
interventions to the Imaging systems
consisting of numerous units having
high technological value, by
authorized engineers trained by the
manufacturer because of patient
health,safety and system performance.
Therefore, the article is requested to be
removed from the Technical
Specification.
165 TS New -
Touch sensitive panel or monitors are
easy use features. Therefore, we kindly
request the insertion of the additional
item to the technical specifications as
“At least one of the monitors on the
stand or the control panel of live image
on the C-arm will be touch sensitive.”.
The current state of the TS has been found
sufficient.
166 TS New -
We kindly request the insertion of the
additional item to the technical
specifications as “Monitors on the
stand must have up / down vertical
movement, or monitors should be able
to move at least +/- 10 degrees tilt.”.
The current state of the TS has been found
sufficient.
167 TS New -
We kindly request the insertion of the
additional item to the technical
specifications as “With the wireless
remote control in the system, it shall
The current state of the TS has been found
sufficient.
![Page 61: CLARIFICATIONS No:1 to TENDER DOSSIER · 2020. 11. 20. · installation, commissioning, maintenance, after-sales service by the contractor of the following goods: Lot 1 High Level](https://reader035.vdocuments.mx/reader035/viewer/2022071609/61482386cee6357ef92528d4/html5/thumbnails/61.jpg)
Page 61 of 61
# DOC
ART
/
ITEM
CLAUSE QUESTION ANSWER
be possible to easily switch between
images, select image amplifier, select
fluoroscopy mode or call up images on
the separate stand and watch the live
image displayed on the C-arm
.Additionally live image rotation can
be done on the touch control panel or
on the touch sensitive monitor by
tapping fingertip.”.
168 TS New -
We kindly request the insertion of the
additional item to the technical
specifications as “The device must
contain an energy storage unit (ESU)
or a Boost Caps energy storage unit to
maintain the continuity of power
supply. Devices that make power
increase with battery system will not
be accepted. The companies will
display the features mentioned above
by marking them on the catalog with
their full name.”.
The current state of the TS has been found
sufficient.