carrier gases in capillary gc analysis-e-seminar

51
Carrier Gases in Capillary GC Abby Folk Capillary GC Applications Engineer February 15, 2011

Upload: others

Post on 12-Sep-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Carrier Gases in Capillary GC Analysis-e-seminar

Carrier Gases in Capillary GC

Abby Folk

Capillary GC

yApplications EngineerFebruary 15, 2011

Page 2: Carrier Gases in Capillary GC Analysis-e-seminar

Configurations for Carrier Gas Purifiers

Indicating Oxygen TrapOxygen Trap

High CapacityOxygen Trap

GCIndicatingMoisture Trap

GC

Group/Presentation TitleAgilent Restricted

Page 2

Page 3: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASMobile PhaseMobile Phase

Carries the solutes down the column

Selection and velocity influences efficiency and retention time

Must be inert to solutes and stationary phase

Must be free of detectable contaminants

Must have a leak free and very precise pressure delivery system

Group/Presentation TitleAgilent Restricted

Page 3

delivery system

Page 4: Carrier Gases in Capillary GC Analysis-e-seminar

COMMON CARRIER GASES

NitrogenNitrogen

Helium

Hydrogen

Group/Presentation TitleAgilent Restricted

Page 4

Page 5: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASPropertiesProperties

Compressible

Expands with temperature

Viscosity increases with temperaturey p

Group/Presentation TitleAgilent Restricted

Page 5

Page 6: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASCARRIER GASConstant Pressure Mode

Head pressure is constantFlow decreases with temperature

Group/Presentation TitleAgilent Restricted

Page 6

Flow decreases with temperature

Page 7: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASC t t Fl M dConstant Flow Mode

Carrier gas flow is constantPressure increases with temperature

Group/Presentation TitleAgilent Restricted

Page 7

Pressure increases with temperature

Page 8: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASConsistent Temperature

Set the velocity at the same temperatureSet the velocity at the same temperature

Initial temperature is the most convenientInitial temperature is the most convenient

Group/Presentation TitleAgilent Restricted

Page 8

Page 9: Carrier Gases in Capillary GC Analysis-e-seminar

van Deemter Equation

h = Height equivalent to a theoretical plate

Packed Columns

h = A + B + CuPacked Columns

uh A + B + Cu

O T b l C l

h = B + CuOpen-Tubular Columns

A = Multi-path termB = Longitudinal diffusion term

uh B + Cu

Group/Presentation TitleAgilent Restricted

Page 9

Carrier Gas Selection

Page 9

C = Mass transfer term

Page 10: Carrier Gases in Capillary GC Analysis-e-seminar

Packed Columnsvan Deemter Curves

HET

P

HETP

Packed ColumnsHETP = A + B/u + Cu

HETPmin

uA + C

B/u A

Uopt u

B/u A

TP Open-Tubular ColumnH

E

HETP

pHETP = B/u + Cu

HETPmin

U t

HETP

uCB/u

Group/Presentation TitleAgilent Restricted

Page 10

Uopt u

Carrier Gas Selection

Page 11: Carrier Gases in Capillary GC Analysis-e-seminar

van Deemter Curve

Excessive Diffusion

1.00

0.50

0.75h

Poor Mass Transfer

0.25

0.50

ūoptOPGV

10 20 30 40 50 60

ūopt

Group/Presentation TitleAgilent Restricted

Page 11

Carrier Gas Selection

Page 11

u (cm/sec)

Page 12: Carrier Gases in Capillary GC Analysis-e-seminar

uopt and OPGV

uopt:Maximum efficiency

OPGV:Optimal practical gas velocityOptimal practical gas velocity

Maximum efficiency per unit time

1.5 - 2x uopt

Group/Presentation TitleAgilent Restricted

Page 12

Page 13: Carrier Gases in Capillary GC Analysis-e-seminar

VAN DEEMTER CURVENitrogen

N2

1.00

0.50

0.75H

0.25

0.50

12-20 cm/sec

10 20 30 40 50 60u (cm/sec)

Group/Presentation TitleAgilent Restricted

Page 13

u (cm/sec)

Page 14: Carrier Gases in Capillary GC Analysis-e-seminar

VAN DEEMTER CURVEHelium

1.00

He

0.50

0.75H

He

0.25

0.50

22-35 cm/sec

10 20 30 40 50 60u (cm/sec)

Group/Presentation TitleAgilent Restricted

Page 14

u (cm/sec)

Page 15: Carrier Gases in Capillary GC Analysis-e-seminar

VAN DEEMTER CURVEHydrogenHydrogen

1.00

0.50

0.75H

0.25

0.50H2

35-60 cm/sec

10 20 30 40 50 60u (cm/sec)

Group/Presentation TitleAgilent Restricted

Page 15

( )

Page 16: Carrier Gases in Capillary GC Analysis-e-seminar

VAN DEEMTER CURVES

N1.00

He

N2

0.50

0.75H

He

0.25

0.50

H2

10 20 30 40 50 60u (cm/sec)

Group/Presentation TitleAgilent Restricted

Page 16

u (cm/sec)

Page 17: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASFlow Rate (mL/min)( )

"Volume"Volume

Measurement:

At column exitAt column exit

Calculate

Electronic Pressure Control (EPC)

Group/Presentation TitleAgilent Restricted

Page 17

Page 18: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASFlow Rate (mL/min)( )

“Handy Gizmos” for Flow Measurement:

FID Flow Measuring Insert (p/n 19301-60660)

“Little Red Cap” (p/n 325-0506)Little Red Cap (p/n 325-0506)

Group/Presentation TitleAgilent Restricted

Page 18

Page 19: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASAverage Linear Velocity (cm/sec)g y ( )

"Speed"

Lu =

Ltmm

L = column length (cm)tm = retention time of non-retained peak (sec)

Group/Presentation TitleAgilent Restricted

Page 19

Page 20: Carrier Gases in Capillary GC Analysis-e-seminar

FLOW RATE CALCULATION

Provides average flow rate

Fπ r2 L

F = tm

r = column radius (cm)( )L = column length (cm)tm = retention time of a non-retained peak (min)

Group/Presentation TitleAgilent Restricted

Page 20

Page 21: Carrier Gases in Capillary GC Analysis-e-seminar

NON-RETAINED COMPOUNDS

Detector CompoundDetector CompoundFID Methane, Butane

TCD, MS Methane, Butane, Air

ECD Vinyl chloride, SF6

Methylene Chloride (vapors)*

NPD Acetonitrile (vapors)*NPD Acetonitrile (vapors)

PID, ELCD Vinyl chloride

Group/Presentation TitleAgilent Restricted

Page 21

* at elevated temperatures

Page 22: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASAverage Linear Velocity CalculationAverage Linear Velocity Calculation

Ltm =u m ____

L l l th ( )L = column length (cm)tm = retention time of non-retained peak (sec)u = desired average linear velocity (cm/sec)

Group/Presentation TitleAgilent Restricted

Page 22

g y ( )

Page 23: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASCARRIER GASAverage Linear Velocity Calculation

Ltm =

3000 cm

u m

32 cm/sec3000 cmtm = = 93.8 sec = 1.56 min

tm = retention time of non-retained peak (sec)L = 30 meters = 3000 cm

Group/Presentation TitleAgilent Restricted

Page 23

u = 32 cm/sec

Page 24: Carrier Gases in Capillary GC Analysis-e-seminar

Figuring Carrier Gas Flow Rate – the easy way

USER CONTRIBUTED SOFTWARE

GC Pressure/Flow CalculatorGC Pressure/Flow Calculator Software

http://www.chem.agilent.com/en-US/S /D l d /U ili i /US/Support/Downloads/Utilities/Pages/GcPressureFlow.aspx

Group/Presentation TitleAgilent Restricted

Page 24

Page 25: Carrier Gases in Capillary GC Analysis-e-seminar

RESOLUTION VS. LINEAR VELOCITYRESOLUTION VS. LINEAR VELOCITYHelium

4 50 3 84 3 364.50 3.84 3.36

R = 1.46 R = 1.31 R = 0.9730 cm/sec 35 cm/sec 40 cm/sec

DB-1, 15 m x 0.32 mm ID, 0.25 um60°C isothermal

4.4 psig 5.1 psig 5.8 psig

Group/Presentation TitleAgilent Restricted

Page 25

60 C isothermal1,3- and 1,4-Dichlorobenzene

Page 26: Carrier Gases in Capillary GC Analysis-e-seminar

Effect of dimensional tolerances in capillary GC columnscolumns

I understand why RT changes occur with “in use” columns, but why with new columns?, y

– Normal dimensional differences in capillary tubingLengthInner diameter

– Minor differences in film thickness (β)– Variability in phase selectivity (RI)Variability in phase selectivity (RI)

More of an issue with high polarity phases

Group/Presentation TitleAgilent Restricted

Page 26

Page 27: Carrier Gases in Capillary GC Analysis-e-seminar

Capillary Tubing Dimensional Tolerances

Capillary columns manufactured by Agilent Technologies have a general dimensional specification of:specification of:

– Length within about 0.5 meter (≈ 1 “loop”)– ID ± 6 μmμ

Variation in internal diameter is a normal distribution around nominal. Approximating the range (12 m) as 6 times the standardthe range (12 μm) as 6 times the standard deviation, there is a 95.5% probability that tubing will be within ± 4 μm.

Let’s look at the effects of these actual tubing dimensions on the pressures required to

i t i t ti ti f th dGroup/Presentation Title

Agilent RestrictedPage 27

maintain retention times for a method…

Page 28: Carrier Gases in Capillary GC Analysis-e-seminar

Relationship of Tubing Dimensions and Pressure:30m x 0.25mm ID column

L (m) ID (μm) P (psi) % of nominal P29.5 255 8.2 82%29.5 255 8.2 82%30.0 250 10.0 100%30.5 245 11.9 119%Conditions: vacuum outlet (MSD), helium carrier,

100°C oven temperture,maintained Tm at 1.38 min

L (m) ID (μm) P (psi) % of nominal PL (m) ID (μm) P (psi) % of nominal P29.5 255 9.3 93%30.0 250 10.0 100%30.5 245 10.9 109%Conditions: atmospheric outlet (FID), helium carrier,

100°C oven temperture,maintained Tm at 2.60 min

Group/Presentation TitleAgilent Restricted

Page 28

Calculated using Column Pressure/Flow Calculator available from Agilent (www.agilent.com/chem)

Page 29: Carrier Gases in Capillary GC Analysis-e-seminar

Relationship of Tubing Dimensions and Pressure:12m x 0.20mm ID column

L (m) ID (μm) P (psi) % of nominal P11.5 205 6.9 69%12.0 200 10.0 100%12.5 195 13.5 135%Conditions: vacuum outlet (MSD), helium carrier,

100°C oven temperture100 C oven temperture,maintained Tm at 0.345 min

L (m) ID (μm) P (psi) % of nominal PL (m) ID (μm) P (psi) % of nominal P11.5 205 8.7 87%12.0 200 10.0 100%12.5 195 11.5 115%Conditions: atmospheric outlet (FID), helium carrier,

100°C oven temperture,maintained Tm at 0.651 min

Group/Presentation TitleAgilent Restricted

Page 29

Calculated using Column Pressure/Flow Calculator available from Agilent (www.agilent.com/chem)

Page 30: Carrier Gases in Capillary GC Analysis-e-seminar

Comparison of Influence of Length and I.D. onComparison of Influence of Length and I.D. on Required Head pressure (.25 mm, MSD)

14

16

re (p

si)

I.D.length

10

12

ad P

ress

u lengthExtremes

81 2 3

Hea

“Extremes” denotes combination of smallest ID/longest l d i ( 1 l th 6 ID)

Group/Presentation TitleAgilent Restricted

Page 30

column and vice versa. (±1 m length; ±6 µm ID)

Page 31: Carrier Gases in Capillary GC Analysis-e-seminar

Comparison of Influence of Length and I D onComparison of Influence of Length and I.D. on Required Head Pressure (.2 mm, MSD)

121416

re (p

si)

I.D.

68

10

d pr

essu

r

length

Extremes

24

1 2 3

head

“Extremes” denotes combination of smallest ID/longest l d i ( 1 l th 6 ID)

Group/Presentation TitleAgilent Restricted

Page 31

column and vice versa. (±1 m length; ±6 µm ID)

Page 32: Carrier Gases in Capillary GC Analysis-e-seminar

Impact of Dimensional Differences on Required PRequired P

I t i i l ith l th d IDImpact varies inversely with length and ID- The relative percentage of impact increases as

nominal length and ID decrease.nominal length and ID decrease.

Additionally, vacuum outlet (MSD) greatly exaggerates pressure drop across theexaggerates pressure drop across the tubing. This in turn amplifies the differences in head pressure required to maintain Tm

Group/Presentation TitleAgilent Restricted

Page 32

Page 33: Carrier Gases in Capillary GC Analysis-e-seminar

“I don’t use retention time windows. Why should I care if my retention times shift?”care if my retention times shift?

Retention time changes in temperature programmed analyses can also alter the

l ti f l telution sequence of solutes…

Solutes elute in an order mandated by their “net” vapor pressures - i.e., vapor

d th ipressures under their gas chromatographic conditions.

Group/Presentation TitleAgilent Restricted

Page 33

Page 34: Carrier Gases in Capillary GC Analysis-e-seminar

Impact of Dimensional Differences on RNot only may absolute retention times change, but resolution may change as well due to changes in carrier gas linear velocity (efficiency; HETP)

MSD

MSD

Group/Presentation TitleAgilent Restricted

Page 34

Page 35: Carrier Gases in Capillary GC Analysis-e-seminar

Net Vapor PressureThe net vapor pressure is the intrinsic vapor pressure reduced by the sum of all solute-stationary phase interactions(e g dispersion proton sharing and dipole interactions all of(e.g., dispersion, proton sharing and dipole interactions, all of which are influenced by temperature).

If considerations for the inter-effects of length, diameter and g ,gas velocity are not factored into the method’s Standard Operating Procedure (SOP) the end result can be:

• Loss of resolution• Complete reversal of elution

More common in mixtures of disparate functionalities; it does not occur– More common in mixtures of disparate functionalities; it does not occur with homologues.

– Also more common with solutes and stationary phases employing multiple modes of solute-stationary phase interactions

Group/Presentation TitleAgilent Restricted

Page 35

multiple modes of solute stationary phase interactions.

Page 36: Carrier Gases in Capillary GC Analysis-e-seminar

Example -- Compounds of InterestExample -- Compounds of Interest

d Camphor Linalool Linalyl acetated-Camphor, b.p 176-180°Cketo functionWeak H-Bonding, Van der Waals

Linaloolb.p. 199°Chydroxy and alkene functionsStrong H-Bonding, weak dipole,Van der Waals

Linalyl acetateb.p. 220°Cacid ester and alkene functions2 x Weak H-BondingVan der Waals Van der Waals 2 x Weak H-Bonding,weak dipole, Van der Waals

Group/Presentation TitleAgilent Restricted

Page 36

Page 37: Carrier Gases in Capillary GC Analysis-e-seminar

Effects of Head Pressure on Elution Order: 26 psi

N FID1 A, (072701A\005B0801.D)

6 6

linaloolLinalyl acetateColumn: 20m x 0.10mm ID x 0.2um, DB-WAX

Oven: 50C (0.33 min), 10C/min to 200C and hold

Norm.

100

120

12.

456

12.

706

Lavender Oil, Spiked, 26 psi Constant Pressure

80

4

camphor

40

60 12.

534

Camphre, 12.534 min

20

10.

029

10.

642

11.

038

11.

152

11.

566

1302

3

Group/Presentation TitleAgilent Restricted

Page 37

min9 9.5 10 10.5 11 11.5 12 12.5

Page 38: Carrier Gases in Capillary GC Analysis-e-seminar

Effects of Head Pressure on Elution Order: 30 psi

Norm FID1 A, (072701A\005B0601.D)

4 7

Column: 20m x 0.10mm ID x 0.2um, DB-WAX

Oven: 50C (0.33 min), 10C/min to 200C and hold

Norm.

100

120

11.

954

12.

187

Lavender Oil, Spiked, 30 psi Constant Pressure

80

40

60

12.

782 1

2.91

6

Camphre, co-elutes with Linalool

20

9.5

43

10.

143

10.

550

10.

637

11.

045

12.

481

12.

568

12.

694

Group/Presentation TitleAgilent Restricted

Page 38

min9 9.5 10 10.5 11 11.5 12 12.5

Page 39: Carrier Gases in Capillary GC Analysis-e-seminar

Effects of Head Pressure on Elution Order: 38 psi

FID1 A, (072701A\005B0501.D)

7 0

Column: 20m x 0.10mm ID x 0.2um, DB-WAX

Oven: 50C (0.33 min), 10C/min to 200C and hold

Norm.

100

120

11.

187

11.

400

.182Lavender Oil, Spiked, 38 psi, Constant Pressure

80 11.

096

086

12

Camhpre, 11.096 min

40

60

11.

986

12.

0

12.

750

20

9.3

77 9.8

03 9

.845

10.

248

11.

655

11.

733

11.

861

12.

332

12.

465

12.

781 1

2.90

4

Group/Presentation TitleAgilent Restricted

Page 39

min9 9.5 10 10.5 11 11.5 12 12.5

Page 40: Carrier Gases in Capillary GC Analysis-e-seminar

Temperature Programming

COLUMN A

Effect Of Locking Software On Temperature Of Elution

Inner diameter = 0.240 mm, Head Pressure 9.70 psi

COLUMN B

Tm 1.52 min

COLUMN BInner diameter = 0.260 mm, Head Pressure 9.70 psi

f

Tm 1.29 min

Group/Presentation TitleAgilent Restricted

Page 40

Temperature Profile

Page 41: Carrier Gases in Capillary GC Analysis-e-seminar

Temperature Programming Eff t Of L ki S ft O T t Of El ti

COLUMN A

Effect Of Locking Software On Temperature Of Elution

Inner diameter = 0.240 mm, Head Pressure 9.70 psi

Tm 1.52 min

COLUMN BCOLUMN BInner diameter = 0.260 mm, Head Pressure 5.97 psi

T t P filTm 1.52 min

Group/Presentation TitleAgilent Restricted

Page 41

Temperature Profile

Page 42: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASCARRIER GASHydrogen Comments

Hydrogen is extremely diffusive in airy g y

Difficult to reach explosive level of ~4 %p

Many GC's are flow regulatedy g

Group/Presentation TitleAgilent Restricted

Page 42

Page 43: Carrier Gases in Capillary GC Analysis-e-seminar

VAN DEEMTER CURVES

N1.00

He

N2

0.50

0.75H

He

0.25

0.50

H2

10 20 30 40 50 60u (cm/sec)

Group/Presentation TitleAgilent Restricted

Page 43

u (cm/sec)

Page 44: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASCARRIER GASSelection Example

6 2 min

Nitrogen11.7 cm/sec

Nitrogen20 cm/sec

Helium23.2 cm/sec

Hydrogen48 cm/sec

24.5 min 17.4 min13.0 min 6.2 min

R = 1.76 R = 1.33 R = 1.67 R = 1.65

Group/Presentation TitleAgilent Restricted

Page 44

Page 45: Carrier Gases in Capillary GC Analysis-e-seminar

Carrier GasHelium vs HydrogenHelium vs. Hydrogen

Helium (35 cm/sec) Hydrogen (73 cm/sec)1 3

21

4

5

6

78

9

2

3

4 6 7 89

59

DB-1, 15 m x 0.25 mm I.D., 0.25 µm

0 2 4 6 8 10 12Time (min.)

0 2 4 6 8Time (min.)

10.5 min 7.8 min

Group/Presentation TitleAgilent Restricted

Page 45

50°C for 2 min, 50-110°C at 20°/min

Page 46: Carrier Gases in Capillary GC Analysis-e-seminar

Hydrogen as Carrier with MSD – Contamination?

Contamination of GC flow modules and lines.

Hydrogen acts as a scrubber.

On the MS this typically looks like Hydrocarbon yp y ycontamination

Most people report that it takes 2-4 weeks to clean out, depending on flowsdepending on flows.

The FID only sees a high background for that time.

Hydrogen seems to scrub the lines of that which Helium leaves behind.

Group/Presentation TitleAgilent Restricted

Page 46

Carrier Gas Selection

Page 46

Page 47: Carrier Gases in Capillary GC Analysis-e-seminar

Carrier Gas - Summary

Gas Advantages DisadvantagesNitrogen Cheap Readily available Long run timesNitrogen Cheap, Readily available Long run timesHelium Good compromise, Safe ExpensiveHydrogen Shorter run times Cheap ExplosiveHydrogen Shorter run times, Cheap Explosive

Hydrogen is difficult to explode under GC conditions

Group/Presentation TitleAgilent Restricted

Page 47

Hydrogen is difficult to explode under GC conditions

Page 48: Carrier Gases in Capillary GC Analysis-e-seminar

CARRIER GASSelection SummarySelection Summary

Hydrogen is best especially for wide k range analyses

Helium is acceptable

Nitrogen is not recommended

Group/Presentation TitleAgilent Restricted

Page 48

Page 49: Carrier Gases in Capillary GC Analysis-e-seminar

Agilent J&W Scientific Technical Support

800-227-9770 (phone: US & Canada)*

* Select option 3 option 3 then option 1* Select option 3, option 3, then option 1.

866-422-5571 (fax)

www.agilent.com/chem

Group/Presentation TitleAgilent Restricted

Page 49

Page 50: Carrier Gases in Capillary GC Analysis-e-seminar

Wrap-up e-Seminar Questions

Thank you for attending today’s Agilent e-Seminar. Our Seminar schedule is expanding regularly.

Please check our web site frequently at:

www.agilent.com/chem/eseminars

Or register for Stay current

to receive regular updates.

with e - notes

Page 51: Carrier Gases in Capillary GC Analysis-e-seminar