basic capillary gc theory and practical troubleshooting

62
Basic Capillary GC Theory and Practical Troubleshooting Katherine K. Stenerson Supelco, 595 North Harrison Road, Bellefonte, PA 16823 T404077 HCE Part 4, Troubleshooting

Upload: others

Post on 21-Mar-2022

9 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Basic Capillary GC Theory and Practical Troubleshooting

Basic Capillary GC Theory and Practical Troubleshooting

Katherine K. StenersonSupelco, 595 North Harrison Road, Bellefonte, PA 16823

T404077HCE

Part 4, Troubleshooting

Page 2: Basic Capillary GC Theory and Practical Troubleshooting

GC Troubleshooting: Topics

• Basic Troubleshooting Strategy• Preventing Problems

- Gas Purification- Injection Technique- Liner Selection and Care- Column Installation - Guard Columns

• Identifying Common Problems• Recommended Reading• Discussion

Page 3: Basic Capillary GC Theory and Practical Troubleshooting

Troubleshooting Strategy

• Have appropriate equipment and supplies on hand.• Establish a systematic approach.• Know what to look for first.• Record what you did to correct the problem.

Page 4: Basic Capillary GC Theory and Practical Troubleshooting

Troubleshooting Strategy

• Suggested equipment to have on hand for troubleshooting:- Electronic Leak Detector- Flow Meter- “Test” Column- Replacement Accessories (Syringes, Ferrules, Septa, Liners)- Replacement Purifiers

Page 5: Basic Capillary GC Theory and Practical Troubleshooting

Troubleshooting Strategy

• Isolate the source of the problem:

Problem was in the Original

detector

Not OKOK

Check operatingparameters

RunReference Standard

Correct the parameter

OK Not OK OK

Problem was column

relatedNot OK

Problem was sample

related

Install Test Column

Switch Detector OK

Problem was in the inlet or with

the carrier gas Not OK

Page 6: Basic Capillary GC Theory and Practical Troubleshooting

Troubleshooting Strategy

• Approaching the problem… - Check first to see if a “fix” for the problem is already known.- Check the Supelco Capillary GC Troubleshooting Guide

(Bulletin 853.)- Check the instrument maintenance record.- Talk to others in your lab.

Page 7: Basic Capillary GC Theory and Practical Troubleshooting

Troubleshooting Strategy

• Five major sources of chromatographic problems: - Operator Error- The Sample- The Column- The Gas Flow System (both internal and external to the GC)- The GC Electrical System

Page 8: Basic Capillary GC Theory and Practical Troubleshooting

Troubleshooting Strategy

• When reviewing method parameters, consider these questions:- Should I be doing split or splitless injection?- Is my starting temperature low enough to allow sufficient

sample focusing?- For splitless injections, is my splitter opening at the appropriate

time?- Is my column flow set to give me maximum efficiency at the

most critical point?- Are heated zones (injectors, detectors, interfaces) set

appropriately?- Am I using the appropriate liner type?

Page 9: Basic Capillary GC Theory and Practical Troubleshooting

Preventing Problems

• The best way to solve problems is to prevent them!- Install and maintain proper purification for all gases in the GC

system.- Maintain the injector by periodically inspecting and changing

the liner, septa, and seal.- Use the proper injection technique - this includes using the

right liner for the job.- Install the column at the recommended insertion distances.- When necessary, use a guard column to protect the analytical

column.

Page 10: Basic Capillary GC Theory and Practical Troubleshooting

Gas Purification

• Carrier Gas - At minimum, remove hydrocarbons, water, and oxygen.

• Hydrogen (FID)- At minimum, remove hydrocarbons.

• Air (FID)- At minimum, remove water and hydrocarbons.

• Nitrogen make-up (FID, ECD)- At minimum, remove hydrocarbons.

• P-5 make-up (ECD)- At minimum remove hydrocarbons, halocarbons, and oxygen.

Page 11: Basic Capillary GC Theory and Practical Troubleshooting

Gas Purification

• Acceptable purity levels for chromatography grade gases:

Impurity / Maximum Concentration

Gas O2 H2O CO2 COTotal

HydrocarbonsHelium <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppmNitrogen <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppmAir 20-22% <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppmHydrogen <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppmArgon/ Methane <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm

Page 12: Basic Capillary GC Theory and Practical Troubleshooting

Gas Purification

• Suggested gas purifiers:

Hydrocarbons Water Oxygen Carrier Supelcarb HC

Supelpure HC Mol Sieve 5A OMI-2

H2

Air Mol Sieve 5A

N2 makeup

P-5 OMI-2 OMI-2 P-5

Page 13: Basic Capillary GC Theory and Practical Troubleshooting

Gas Purification

• Hydrocarbon Traps- Noise in the baseline (FID)- Increase in background peaks

on tune (MSD)- Higher than normal baseline

reading on FID- Extra peaks visible in run

• Molecular Sieve 5A- Increase in column bleed- Water visible in MS

background- Poor peak shapes for gaseous

VOCs (purge and trap)- Extra peaks visible in run- OMI™-2 color change

• What are some signs that my purifiers need to be changed?

Page 14: Basic Capillary GC Theory and Practical Troubleshooting

Injector Maintenance

• Change (as needed):- Septa- Liner and O-ring- Seal and washer

• Inspect the Inlet Periodically- Look for contamination in the liner- Look for residue on the seal

Page 15: Basic Capillary GC Theory and Practical Troubleshooting

Injection Technique

• It is important to choose the injection technique that is appropriate for your analysis. In Capillary GC, the techniques used are:- Split - Splitless - Direct- On-column

Page 16: Basic Capillary GC Theory and Practical Troubleshooting

Injection Technique

• Split Injection- A vaporizing type injection designed to limit the amount of

sample reaching the capillary column.- Sample is split and a small portion flows to the column while

the bulk is typically vented through the split vent port.- Split injection can be used in an isothermal or temperature

programmed analysis.

Page 17: Basic Capillary GC Theory and Practical Troubleshooting

Injection Technique

• Splitless Injection - Sample is introduced into a heated injection port operating in a

nonsplitting mode.- Sample vaporizes and sample cloud is mixed with carrier gas

and transferred into the column.- Sample condenses on head of the capillary column due to the

oven temperature being 10-20ºC below matrix solvent boiling point.

- After 1.5 to 2 injector volumes have entered the column, split vent is opened and inlet purged.

Page 18: Basic Capillary GC Theory and Practical Troubleshooting

Injection Technique

• The volume of a splitless liner is important:

Typical Splitless Injection Liner Volumes

Liner Length Liner ID Liner Volume

78.5mm 4.0mm 986µL78.5mm 2.0mm 246µL78.5mm 1.0mm 62µL

Page 19: Basic Capillary GC Theory and Practical Troubleshooting

Solvent BP (°C) 10 20 30 10 20 30Ethyl acetate 77 236 168 131 286 204 58Hexane 68.7 177 126 98 214 153 119Isooctane 99.2 140 100 77 170 121 94Methanol 64.5 570 406 315 691 492 382ychloride 40.1 360 257 200 437 311 241MTBE 55 194 138 108 235 167 130Water 100 1279 910 706 1548 1102 855

Head pressure 300°C Inlet Temp.

Head pressure 200°C Inlet

Injection Technique

• Solvent expansion volumes of 1µL injection at specified temperatures and pressures:

Page 20: Basic Capillary GC Theory and Practical Troubleshooting

Injection Technique

• Direct Injection - A vaporizing type injection typically used with wide bore

capillary columns in a converted packed column GC.- Sample is injected into a heated injection sleeve, vaporized,

and transported directly to the column in the carrier gas flow. Similar to a packed column flash vaporization injector.

- Analyses can be isothermal or temperature programmed.

Page 21: Basic Capillary GC Theory and Practical Troubleshooting

Injection Technique

• Cold On-Column Injection - A non-vaporizing type injection in which the liquid sample is

directly deposited at the inlet of the capillary columns.- All analyses are temperature programmed analyses since a

liquid sample is deposited in the column.- No liners are typically required.- Special syringes are required.- 0.53mm ID fused silica created to allow insertion of 26 gauge

needle into column.

Page 22: Basic Capillary GC Theory and Practical Troubleshooting

Injection Technique

0 2 4 6 8 10Time (min)

0 2 4 6 8 10Time (min)

Rapid, smooth injection

Slow, “choppy” injection

• Injection speed can have an effect.

Page 23: Basic Capillary GC Theory and Practical Troubleshooting

Liner Selection

• Some liners used for split injection:

Cup

Split/splitless, wool packed

Baffle (Varian)

Cup

Baffle (Varian)

Split/splitless, wool packed

Cup

Baffle (Varian)

Page 24: Basic Capillary GC Theory and Practical Troubleshooting

Dual-tapered

Liner Selection

• Some liners used for splitless injection:

2 mm ID, straight

Single-tapered

Page 25: Basic Capillary GC Theory and Practical Troubleshooting

Liner Selection

• PROs- Packing liners helps aid in the

vaporization process- Packing liners can help

improve reproducibility of area counts by minimizing droplets reaching the head of the column

- Packing can act as a particle trap

• CONs- The packing does act as a

short packed column and can influence results

- Can cause discrimination of higher molecular weight compounds

- Can cause adsorption & sample degradation

• Packed liners, PROs and CONs:

Page 26: Basic Capillary GC Theory and Practical Troubleshooting

Splitless injection, 2mm vs. 4mm ID liner

Liner Selection

• The ID of the liner can affect sensitivity:

Page 27: Basic Capillary GC Theory and Practical Troubleshooting

Liner Care

• If you must clean a liner…. - Handle liners with gloves or forceps.- Use clean compressed gas and/or a fine brush to remove

particles.- Rinse liner in an appropriate solvent and dry with clean

compressed gas.- Use mineral acid and/or detergent only if absolutely necessary.

Be sure to deactivate the liner after after this process.- If repacking with glass wool, make sure it has been

deactivated.

Page 28: Basic Capillary GC Theory and Practical Troubleshooting

18 20 22 24 26Time (min)

18 20 22 24 26Time (min)

Undeactivatedglass wool

1

2

3 4

5

1

2

3

45

6

6

1. 4,4’-DDE2. Endrin3. 4,4’-DDD4. Endrin aldehyde5. 4,4’-DDT6. Endrin ketone

Liner Care

• The results of using undeactivated glass wool in 4mm ID liner used for pesticide analysis:

Deactivated glass wool

Page 29: Basic Capillary GC Theory and Practical Troubleshooting

Column Installation

• Installing the column too low in the inlet can result in peak tailing.

0 2 4 6 8 10 12 14Time (min)

This column was installed too low in the inlet.

Page 30: Basic Capillary GC Theory and Practical Troubleshooting

Guard Columns

• Choose a guard column that has been deactivated. • Usually, the ID of the guard matches the analytical

column. • A 5-10 meter length is normally used.• Connect with either a GlasSeal™ or butt connector.

Page 31: Basic Capillary GC Theory and Practical Troubleshooting

Common Problems

1. Poor Peak Shapes (either tailing, fronting, or just generally ugly.)

2. Nonlinearity3. Baseline Noise and /or Drift4. Ghost Peaks5. Missing Peaks / Poor Response6. Insufficient Resolution

Page 32: Basic Capillary GC Theory and Practical Troubleshooting

Poor Peak Shape

• In Gas-liquid chromatography, fronting may indicate column overload.

||

||

• In Gas-liquid chromatography, fronting may indicate column overload.

• Tailing may indicate activity in the system or improper column installation.

Page 33: Basic Capillary GC Theory and Practical Troubleshooting

Poor Peak Shape

• Generally ugly peaks, such as α,α-dimethylphen-ethylamine, can be caused by a variety of problems.

Page 34: Basic Capillary GC Theory and Practical Troubleshooting

Nonlinearity

• The most common causes are:- Column overload- Detector overload- Standards preparation- Poor peak shape resulting in improper integration

Page 35: Basic Capillary GC Theory and Practical Troubleshooting

6.0 7.0

Nonlinearity and Column Overload

• An Example of Column Overload:

Fronting due to overload

Page 36: Basic Capillary GC Theory and Practical Troubleshooting

An Example of Column Overload:

• Preventing column overload: - Inject a smaller amount and/or increase split ratio.- Use a thicker film column.- Use a column with a wider ID.- Decrease upper limit of calibration range.- Use a column of slightly different polarity.

Page 37: Basic Capillary GC Theory and Practical Troubleshooting

Nonlinearity and Poor Peak Shape

• An example of poor peak shape affecting linearity: - The poor peak shape of benzoic acid here is caused by

solubility problems with the 5% phenyl methylpolysiloxanephase.

Page 38: Basic Capillary GC Theory and Practical Troubleshooting

Baseline Noise and Drift

• Common causes: - Column bleed- Septa bleed- Dirty detector- Contaminants in carrier gas / carrier gas purity

Page 39: Basic Capillary GC Theory and Practical Troubleshooting

Column Bleed

• Results from the normal degradation of the stationary phase.

• All columns bleed to some extent.• Bleed increases with temperature.• The amount of bleed will increase in the presence of

oxygen.

Page 40: Basic Capillary GC Theory and Practical Troubleshooting

4

Bleed measured as the difference between 1 and 2.

1

2

Column Bleed

• A Typical Bleed Profile:

Page 41: Basic Capillary GC Theory and Practical Troubleshooting

Column Bleed

• Column bleed and an MSD: - Visible as baseline rise in the TIC.- Check spectra for key bleed ions:

• Equity-1: 73, 207, 281• Equity-5: 207, 281• Equity-1701: 207, 269• SPB™-624: 207, 269

• Make sure interface temp. is < column max. temp.

Page 42: Basic Capillary GC Theory and Practical Troubleshooting

207: D3

Si OSi

SiO

Si

O

O

CH3

CH3CH3

CH3

CH3

CH3CH3

+

OSi

SiO

Si

O

CH3

CH3

CH3

CH3

CH3

+

284: D4

Si +

CH3

CH3

CH3

73: TMS

Column Bleed

• Common Bleed Ions

Page 43: Basic Capillary GC Theory and Practical Troubleshooting

40 60 80 1001201401601802002202402602803003203403603804004204400

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

m/z-->

Abundance

Scan 8569 (43.681 min): EQ12506A.D207

28173

17913396

45239

299 327 355115 161 373 401 447

D3

D4TMS

Column Bleed

• Equity-1: MS Spectra of Bleed

Page 44: Basic Capillary GC Theory and Practical Troubleshooting

40 60 80 1001201401601802002202402602803003203403603804004200

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

7500

8000

8500

9000

9500

10000

10500

11000

11500

12000

12500

13000

m/z-->

Abundance

Scan 4393 (22.148 min): 0326003.D207

28173

96 133 253

177331115 405156 355 42951 311

73

207

281

Column Bleed

• Equity-5: MS Spectra of Bleed

Page 45: Basic Capillary GC Theory and Practical Troubleshooting

40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 3400

2000

4000

6000

8000

10000

12000

14000

16000

18000

20000

22000

24000

m/z-->

Abundance

Scan 8581 (33.704 min): 1022006.D207

269

44253

19113373 96177147

283119 322 343239163 22359 298

207

269

Column Bleed

• SPB™-624: MS Spectra of Bleed

Page 46: Basic Capillary GC Theory and Practical Troubleshooting

40 60 80 1001201401601802002202402602803003203403603804004204400

1000

2000

3000

4000

5000

6000

7000

8000

9000

10000

11000

12000

13000

14000

m/z-->

Abundance

Scan 9126 (84.849 min): 0215001.D207

269

96

13373 177 38451 309 343159115 240 438

207

269

Column Bleed

• Equity-1701: MS Spectra of Bleed

Page 47: Basic Capillary GC Theory and Practical Troubleshooting

40 60 80 1001201401601802002202402602803003203403603804004200

2000

4000

6000

8000

10000

12000

14000

16000

18000

20000

22000

24000

26000

28000

30000

32000

34000

36000

38000

40000

42000

44000

m/z-->

Abundance

Scan 604 (7.460 min): 1201001.D73

147

281

32745

207415399131 191 343 383251 29795 223163115

73

147

281

Septa Bleed

• Septa Bleed: MS Spectra

Page 48: Basic Capillary GC Theory and Practical Troubleshooting

Column & Septa Bleed

• Minimize bleed! - Sufficiently purge column with carrier gas before ramping it up

in temperature.- Make sure carrier gas is scrubbed for water and oxygen.- Check integrity of all fittings leading to the column.- Do not heat the column above its maximum temp.- Precondition the column prior to use.- Use a high quality, high temperature septa and ferrules.

Page 49: Basic Capillary GC Theory and Practical Troubleshooting

10

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

7500

8000

8500

9000

9500

10000

10500

11000

11500

12000

H2 carrier from tank

H2 carrier from a generator

GCMS baseline comparisons

Baseline Noise and Drift

• Effect of carrier gas purity on baseline noise:

Page 50: Basic Capillary GC Theory and Practical Troubleshooting

Ghost Peaks

• Possible causes:- Residue in the inlet liner and at the head of the column- Contaminated syringe / and or wash solutions on an

autosampler- Sample carryover- Contaminated carrier gas

Page 51: Basic Capillary GC Theory and Practical Troubleshooting

0

200000

400000

600000

800000

1000000

1200000

1400000

1600000

1800000

2000000

2200000

Response test mix, before

Ghost Peaks

• If pieces of septa get into an inlet liner...

Page 52: Basic Capillary GC Theory and Practical Troubleshooting

0

500000

1000000

1500000

2000000

2500000

3000000

3500000

Time-->

Response test mix, after

Ghost Peaks

…even a simple analysis can be ruined.

Page 53: Basic Capillary GC Theory and Practical Troubleshooting

Missing Peaks / Poor Response

• Sample decomposition- Activity in the inlet or column- Injection port temperature too high- Sample not stable enough for GC- Standards not stable

• Coelution• Insufficient run time / final temperature• Sample not volatile enough for GC• Improper column installation

Page 54: Basic Capillary GC Theory and Practical Troubleshooting

0

50000

100000

150000

200000

250000

300000

350000

400000

450000

2,4-

dini

troph

enol

4-ni

tr oph

eno l

2-m

ethy

l-3,5

-din

itrop

heno

l

pent

achl

orop

heno

l

Before Sample Injection

Missing Peaks / Poor Response

• Nasty samples can damage a column by creating active sites.

Page 55: Basic Capillary GC Theory and Practical Troubleshooting

0

100000

200000

300000

400000

500000

600000

700000

800000

900000

Time-->

Abundance

2,4-

DN

P &

4 -N

P sh

ould

be

here

2-m

ethy

l-3,5

-din

itrop

heno

l

Pent

achl

orop

heno

l sho

uld

be h

ere

Missing Peaks / Poor Response

• Responses of some acidic compounds were affected.

After Sample Injection

Page 56: Basic Capillary GC Theory and Practical Troubleshooting

200000

400000

600,000

800000

1000000

1200000

1400000

1600000

1800000

2000000

2200000

2400000

2600000

2,800,000

0

8 mm above top of ferrule

Missing Peaks / Poor Response

• Response can also be affected by the position of the column in the inlet.

Page 57: Basic Capillary GC Theory and Practical Troubleshooting

0

200,000

400000

600000

800000

1000000

1200000

1400000

1600000

1800000

2000000

2,200,000

Missing Peaks / Poor Response

• Here, the column was not inserted far enough.

5 mm above top of ferrule

Page 58: Basic Capillary GC Theory and Practical Troubleshooting

0

200,000

400000

600000

800000

1000000

1200000

1400000

1600000

1,800,000

Missing Peaks / Poor Response

• Here, the column was inserted too far:

20 mm above top of ferrule

Page 59: Basic Capillary GC Theory and Practical Troubleshooting

Insufficient Resolution

• Column- Longer columns increase resolution- Smaller ID columns increase resolution- A different phase altogether may be needed

• Conditions- Carrier gas flow too fast or slow- Oven ramp rate too fast- Wrong starting or ending temperature

Page 60: Basic Capillary GC Theory and Practical Troubleshooting

Supelco Bulletins

741: The Supelco Guide to Leak-Free Connections783: Cleaning Flame Ionization Detectors853: Capillary Troubleshooting Guide875: Supelco Capillary GC Selection Guide895: Installation and Maintenance Instructions for 0.25 mm

and 0.32 mm ID Fused Silica Capillary Columns897: Installation and Maintenance Instructions for 0.53 mm

ID Fused Silica Capillary Columns898: Gas Management Systems for GC899: Capillary GC Inlet Sleeve Selection Guide916: Purge and Trap System Guide918: Selecting Purifiers for Gas Chromatography

Page 61: Basic Capillary GC Theory and Practical Troubleshooting

Supelco Service

• Supelco Technical Service- phone: 1-800-359-3041 (US only), 814-359-3041- email: [email protected]

• Supelco Customer Service - phone: 1-800-247-6628 (US only), 814-359-3441- email: [email protected]

• Sigma-Aldrich Website - www.sigma-aldrich.com

Page 62: Basic Capillary GC Theory and Practical Troubleshooting

Discussion