calibration/intermediate check of measuring and · pdf fileas per nabl 142 - when traceability...

32
Calibration/Intermediate Check of Measuring and Volumetric Equipment S.SUBRAMANIAN

Upload: vandung

Post on 06-Mar-2018

1.078 views

Category:

Documents


49 download

TRANSCRIPT

  • Calibration/Intermediate Check

    of Measuring and VolumetricEquipment

    S.SUBRAMANIAN

  • WHAT IS CALIBRATION?

    Calibration is the process of comparing a measuring instrument against a measurement standard to establish the relationship between

    the values indicated by the instrument and those of the

    standard (which has traceability to national/ international standard of

    measurement) 2

  • WHAT IS INTERMEDIATE CHECK?

    Intermediate check is the process of verifying the calibration status of the measuring instrument in

    between calibration.

    This is done to verify whether there is any drift in the measurement

    indicated by the instrument over a period of time 3

  • MANDATORY REQUIREMENTS FOR CALIBRATION AS PER NABL 142 -

    WHEN TRACEABILITY TO SI UNITS IS POSSIBLE

    CALIBRATION OF INSTRUMENTS BY

    National Physical Laboratory (NPL, India) or any NMI whose service is suitable for the intended need and is covered by the MRA.

    NABL accredited calibration laboratory or a calibrationlaboratory accredited by an Accreditation Body covered by theILAC Arrangement or by Regional Arrangements recognised byILAC e.g. by APLAC, IAAC & EA whose service is suitable forthe intended need (i.e, the scope of accreditation specificallycovers the appropriate calibration).

    BOTH ARE TAKEN AS EVIDENCE OF TRACEABILITY 4

  • MANDATORY REQUIREMENTS FOR CALIBRATION AS PER NABL 142 -

    WHEN TRACEABILITY TO SI UNITS IS NOT POSSIBLE

    CALIBRATION SHALL PROVIDE CONFIDENCE IN MEASUREMENTS BY ESTABLISHING TRACEABILITY TO APPROPRIATE MEASUREMENT STANDARDS SUCH AS

    Use of Certified Reference Materials provided by acompetent supplier to give a reliable physical orchemical characterization of a material; Use ofmatrix CRM and/or CRM spike

    Use of specified methods and/or consensusstandards that are clearly described and agreed byall parties concerned.

    Participation in a suitable programme of interlaboratory comparisons is required where possible.

    5

  • ESSENTIAL REQUIREMENTS OF CALIBRATION

    Trained & Authorized manpower

    Documented procedure

    Suitable Reference standard

    Multi-point calibration covering working range

    Estimation of Measurement Uncertainty

    Acceptable limit in calibration

    Calibration frequency is documented

    Controlling environmental conditions6

  • CLASSIFICATION OF EQUIPMENT

    General service equipment not used for making measurements or with minimal influence on measurements (eg hotplates, stirrers, non-volumetric glassware and glassware used for rough volume measurements such as measuring cylinders)

    Volumetric equipment (e.g. flasks, pipettes, pyknometers, burettes etc);

    Measuring instruments (e.g. hydrometers, U-tube viscometers, thermometers, timers, spectrometers, chromatographs, electrochemical meters, balances etc);

    Physical standards (weights, reference thermometers);

    7

  • CALIBRATION OF GENERAL SERVICE EQUIPMENT

    Maintained by appropriate cleaning and checks for safety as necessary

    Calibrations or performance checks will be necessary where the setting can significantly affect the test or analytical result (e.g temperature of a muffle furnace or constant temperature bath)

    8

  • For the highest accuracy, measurements can often be made bymass depending on properly calibrated weighing mechanismwith traceability to accredited calibration laboratories (inINDIA or abroad APLAC/EA Member Countries) rather than byvolume i.e GRAVIMETRIC METHOD

    GRAVIMETRIC METHOD consists on weighing the volumetricapparatus under calibration when empty and again when full.The procedure adopted the use of the reference line or marksto have the purpose to provide an exact measure of liquidvolume and the draining or drying procedures shall befollowed carefully because they all effect the measurements.The difference obtained in the weighing measurements givesthe mass of contained or delivered liquid.

    CALIBRATION OF VOLUMETRIC EQUIPMENT

    9

  • Balance of suitable capacity and resolution is required as per NABL 122-04

    For volume in range of 100 l to 10ml, 0.1 mg is required

    For volume in range of 10 ml to 1000ml, 1 mg is required

    For volume in range of 1000 ml to 2000 ml, 10 mg is required

    CALIBRATION OF VOLUMETRIC EQUIPMENT

    10

  • Balance requirement for Micropipette

    For volume in range of 100 l, 0.1 mg

    CALIBRATION OF VOLUMETRIC EQUIPMENT

    11

  • Instruments such as Balance, Hydrometers, Viscometers, Thermometers and Timers should be done by accredited calibration laboratories

    Other instruments such as pH meter, Conductivity meter, Chromatographs and Spectrophotometers are to be done using valid, reputed/ traceable RMs.

    Guideline for calibration frequency of equipment:

    SlNo.

    EQUIPMENT FREQUENCY REMARKS

    1 Balance Once a year By accredited lab only

    2 Weight Once in 2 years By accredited lab only

    3 Oven, Incubator, Furnace

    Once a year By accredited lab only

    4 Water Bath Once a year Use suitable reference standard

    5 Thermometer (Liquid in glass)

    Once in 3 years By accredited lab only

    CALIBRATION OF MEASURING INSTRUMENTS

    12

  • CALIBRATION OF MEASURING INSTRUMENTS

    SlNo.

    EQUIPMENT FREQUENCY REMARKS

    6 Stop watch / clock Once a year By accredited lab only

    7 Hygrometer Once a year By accredited lab only

    8 Pressure Gauge Once a year By accredited lab only

    9 Volumetric Labware(Vol. Flask, Burette, Pipette)

    ---- In-house; Use suitableBalance

    10 Micropiette Once a year In-house; Use suitableBalance

    11 UV-Visible / IR / FT-IR/ Spectrophotometer / Colorimeter

    QuarterlyPhotometric Absorbance & wavelength accuracy for the working range

    In-house; Use suitable CRMs

    13

  • CALIBRATION OF MEASURING INSTRUMENTS

    SlNo.

    EQUIPMENT FREQUENCY REMARKS

    12 Atomic Absorption Spectrophotometer

    Quarterly/ Half yearly

    Performance check using Cu CRM as per manufacturers instruction.

    13 pH Meter Once a year In-house; Use suitable CRMs

    14 Conductivity Meter Once a year In-house; Use suitable CRMs

    14

  • CALIBRATION OF MEASURING INSTRUMENTS

    SlNo.

    EQUIPMENT FREQUENCY REMARKS

    15 Gas Chromatograph, GC-MS/HPLC / LC-MS

    Quarterly System performance including: Resolution, sensitivity, repeatability, retention time and noise level

    Use suitable CRMs/RMs 15

  • CALIBRATION OF UV VIS SPECTROPHOTOMETER

    Use NIST traceable K2Cr2O7 for absorbance calibration

    Use Holmium Oxide Filter or solution with NISTtraceability for wavelength calibration

    If Holmium Oxide Filter is used it has to be calibrated only once in its life time, provided storage conditions are not extreme

    For stray light calibration use KCl 1.2% w/v in distilled water. Absorbance should be > 2.0

    16

  • CALIBRATION OF UV VIS SPECTROPHOTOMETER

    WAVELENGTH CALIBRATION-TOLERANCE ALLOWED :

    1 nm in range of 200 400 nm

    3 nm in range of > 400 nm

    ABSORBANCE CALIBRATION-TOLERANCE ALLOWED:

    122.9 to 126.2 @ 235 nm

    142.8 to 146.2 @ 257 nm

    47.0 to 50.3 @ 313 nm

    105.6 to 109.0 @ 350 nm

    15.7 to 16.1 @ 430 nm17

  • CALIBRATION OF VOLUMETRIC GLASSWARE AS PER NABL 122 -04 & IS/ISO 4787:2010 and ISO 8655-6

    MANDATORY REQUIREMENTS

    Balance of suitable capacity

    For volume in range of 100 l to 10ml, 0.1 mg is required

    For volume in range of 10 ml to 1000ml, 1 mg is required

    For volume in range of 1000 ml to 2000 ml, 10 mg is required

    Reagent Grade Water complying with ISO Grade 3 water

    Calibrated weights complying with F1 class of weights

    Air temperature, Relative Humidity and Atmospheric Pressure Monitor

    Controlled environment as given below:

    Temperature = (23 + 2) C

    Relative Humidity = (50 to 85) %

    Intensity of light above 250 lux18

  • CALIBRATION OF VOLUMETRIC GLASSWARE AS PER NABL 122 -04 & IS/ISO 4787:2010 and ISO 8655-6

    PRECAUTIONS TO BETAKEN

    CLEAN GLASSWARE

    TEMPERATURE STABILIZATION OF GLASSWARE, BALANCE, WATER & WEIGHTS

    WEIGHING IS DONE USING CONTAINER WITH LID

    ENSURE WATER IS FREE FROM AIR BUBBLES

    NO DROPLETS OF WATER ON SIDES OF GLASSWARE

    NO WETTING OF UPPER AREA OF GLASSWARE 19

  • CALIBRATION OF VOLUMETRIC GLASSWARE AS PER NABL 122 -04 & IS/ISO 4787:2010 and ISO 8655-6

    Mass of the empty vessel + mass of the standard weight (W1) = Balance reading R1

    Mass of the empty vessel+ distilled water + standard weight (W2) = Balance reading R2

    Mass of Water, Mw (in g) = (R2 R1) + (W1 W2)

    Volume of Water, Vm (in ml) = [Mw / (w - a)] x [1- (a /s)]

    where, w is the density of water during calibration (g/cc)

    a is the density of air during calibration (g/cc)

    s is the density of weights used for calibration (g/cc)

    Volume of Water at 270C, V27 (in ml) = Vm [1 + (27 - tw)]

    where, is the coefft. of cubical thermal expansion of glassware

    tw is the temperature of water used during calibration 20

  • CALCULATION OF MEASUREMENT UNCERTAINTY

    Uncertainty in Measurement repeatability (type A)

    Uncertainty of the balance including linearity*

    Uncertainty of the balance due to readability

    Uncertainty in water temperature measurement

    Uncertainty in Air temperature, pressure & relative humidity

    Uncertainty Coefficient of expansion of the material