beam loss monitoring r&d

31
Beam Loss monitoring R&D Arden Warner Fermilab MPS2014 Workshop March 5-6, 2014

Upload: others

Post on 05-Dec-2021

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Beam Loss monitoring R&D

Beam Loss monitoring R&D Arden Warner

Fermilab

MPS2014 Workshop

March 5-6, 2014

Page 2: Beam Loss monitoring R&D

Outline

• PXIE

• Technical Concerns

PXIE Study plans

Preliminary sCVD R&D

Cold Ionization chambers

MPS2014; Arden Warner2

Page 3: Beam Loss monitoring R&D

Loss Mechanisms at PXIE

MPS2014; Arden Warner3

Low energy beam will transition from RT to SC at 2.1 MeV in the

PXIE front-end

Complicated chopping scenarios

LEBT Kicker and Ion

source /primary actuators

for beam

Require feedback and

monitoring of

intensity/losses

Chopper must be

closely monitored

BeamDump

IonSource

LEBT RFQ MEBT HWR

Diagnostics

30keV 2.1MeV 11MeV

Page 4: Beam Loss monitoring R&D

Technical Concerns

• CW Beam:

Losses develop continuously; no large, peak beam

intensities to trigger loss monitors.

MPS must allow pulsed, diagnostic beam operation when

CW losses inhibit normal operation.

• MEBT Chopper:

Kicks out individual bunches to reduce average current of

beam from 5mA to 1mA. Diagnostics needed to insure

SRF is not overloaded.

• Low Energy SRF Transition:

Most loss monitors are ineffective at low energies.

Reduced penetration depth of beam energy impacting

niobium.

Need a way to detect losses before cavities quench.

MPS2014; Arden Warner4

Page 5: Beam Loss monitoring R&D

PXIE Study Plans

• Warm section:

Develop understanding of acceptable loss rates in

LEBT

RFQ

MEBT

• Develop strategy to monitor chopped beam from the MEBT

• Low energy cold section:

Detection of beam loss in cryomodules

Develop understanding of low energy beam loss mechanisms and

their instruments

Develop/select sensor technology to directly measure low energy

beam loss

Understand impact of dark current effects on MPS instruments

MPS2014; Arden Warner5

Page 6: Beam Loss monitoring R&D

Preliminary sCVD Detector Test

MPS2014; Arden Warner6

CVD diamonds are being investigated because of their sensitivity to single particles,

nanosecond time response and the excellent radiation resistivity which makes them

perfectly suitable for applications in high-radiation environments. In addition operation in

vacuum at both room temperature and 1.8 Kelvin has been demonstrated.

Interest for PIXIE/PIP-II:

1. Develop an effective loss monitor to detect low energy proton losses (2.1 MeV) as a

diagnostic to protect the PIXIE cryogenic system at injection.

2. Provide a detector capable of single particle detection as an effective diagnostic for

beam extinction measurements in the accelerator.

Characteristics of crystal that were tested:

Parameters:

Substrate material sCVD diamond

Substrate size 4.6 mm x 4.6 mm x 0.5 mm

Electrode size 4 mm x 4 mm

Electrode material Gold

Detector capacitance 3 pF

Bias Voltage 500 Volts

Page 7: Beam Loss monitoring R&D

CVD General Principle and Characteristics

MPS2014; Arden Warner7

Page 8: Beam Loss monitoring R&D

Diamond Beam Loss Monitor Response

MPS2014; Arden Warner8

Beam Toroid signal

Single particle detection of scattered electrons and secondary protons from dump

Reflections due to cable capacitance is understood and correctableZoom

Detector located inside the vacuum

with the crystal angled toward the

dump. 160 µs beam pulses, 200 µA

– 3 mA, 2.5 MeV and 500 volt bias

on the detector. Window in case

exposes crystal to the beam.

The ideal crystal should be thin

(100 micron limit possible

dependent on area).

detector Fast Charge Amplifier: 100 MHz, 4 mV/fC

Page 9: Beam Loss monitoring R&D

Arden Warner, FNAL9

Installation Inside CM2

• sCVD Diamond test also underway at ASTA

Installation in cryo-module near BPM and quadrupole

Page 10: Beam Loss monitoring R&D

Proton Interaction

Single particles

Absorption

Particle bunches

Absorption

Single particles

Traversing

LHCTEVATRON

P.T.

ISOLDE

Compliments of Erich Griesmayer

PXIE

PIP-II

Page 11: Beam Loss monitoring R&D

Electron Interaction

400

Single particles

Absorption

Particle bunches

Absorption

Single particles

Traversing

Compliments of Erich Griesmayer

Page 12: Beam Loss monitoring R&D

LHC - DBLM

1 ms

DBLM

Ionization

Chambers

Compliments of Erich Griesmayer

Page 13: Beam Loss monitoring R&D

LHC - DBLM

1 us

Zoom x1000

Compliments of Erich Griesmayer

Page 14: Beam Loss monitoring R&D

LHC - DBLM

100 ns

Zoom x10’000

Compliments of Erich Griesmayer

Page 15: Beam Loss monitoring R&D

15

Cryogenic Beam Loss Monitors (Ionization chamber)

Arden Warner, FNAL ([email protected])

Highlights

operation in air and high vacuum

Operates from 5K to 350K

Stainless steel vessel, 120cm3, filled

with He-gas

He-gas filling at 1.0- 1.5 bar pressure

Sensitivity: 1.9 pA/(Rad/hr)

Readout via current-to-frequency

converter (1.9 Hz/(Rad/hr) and FPGA-

TDC

Pulses can be sent through long

cables

Features

Custom-built prototype detector for

operation as a beam loss monitor at

cryogenic temperatures

Helium filled ionization chamber with

signal current proportional to dose rate

All material radiation hard and suitable

for operation at 5K

Current is measured with a recycling

integrator I-F converter for low current

and a wide dynamic range. A Fermilab

designed FPGA based TDC measures

time intervals between pulses output

from the recycling integrator ensures a

fast response along with current

measurement resolution better than 10-

bits.

Monitors and recycling integrator electronics fabricated and calibrated by Bridgeport Instruments, LLC.

Page 16: Beam Loss monitoring R&D

Arden Warner, FNAL ([email protected])16

Cryogenic Beam Loss Monitors Specifications

Page 17: Beam Loss monitoring R&D

17

Cryogenic Ionization chamber 5k – 350K

The electronics is self-contained and requires no computer to operate.

Helium was chosen because of its properties (boiling point 5K) and the fact that it will be in a helium environment during operation anyway.

Arden Warner, FNAL ([email protected])

Fill port

Bias voltage and electronics

Page 18: Beam Loss monitoring R&D

Arden Warner, FNAL18

Source test and calibration

Page 19: Beam Loss monitoring R&D

The chamber housing is held at negative potential and negative charge is collected on the center electrode. The HV is -95 V and is kept well below the minimum breakdown voltage of 156V in Helium.

Cryogenic Loss Monitor operation

The electronics uses a recycling integrator as a current to frequency converter with a wide dynamic range. The charge per pulse is 1.63pC or 238µR at 1 atm (room temp) of He.

The recycling integrator consist of a charge integrating amplifier with a 0.50 pF capacitance followed by a discriminator which senses when the capacitor is fully charged.

The FPGA generates a fixed-width (1.2µs) discharge pulse with an amplitude of 3.3V. It connects to the amplifier input via a 13 MΩresistor, creating a 254 nA discharge current

19 Arden Warner, FNAL ([email protected])

Page 20: Beam Loss monitoring R&D

Bench top measurements

20

Pulses with 150nA input currentPulses with 300nA input current

The maximum periodic pulse rate at the output is close to 700 KHz. The corresponding maximum chamber current is 1.60 µA or 842 KR/hr. Pulses can be sent loss-free over great distances and the technique allows to measure radiation levels with dynamic range of 100,000: 1

Arden Warner, FNAL ([email protected])

Page 21: Beam Loss monitoring R&D

Arden Warner, FNAL ([email protected])21

Cryogenic Loss Monitor Connection and signal path

NIM to LVDS FPGA-TDC

Output signal

Recycling integrator

1.2 µs NIM pulses (1.63 pC or 238 µR

@ 1 atm

Page 22: Beam Loss monitoring R&D

Arden Warner, FNAL22

Typical digitization scheme with a counter.

digitization scheme using an FPGA based TDC.

In typical digitization/readout schemes as shown in Fig., a

counter is utilized to accumulate the number of pulses generated

by the recycling integrator to digitize the total charge. In order to

calculate current with reasonable resolution, a long period must

be waited for each sample. For example, to achieve 7-bit

resolution, the sampling period corresponds to 128 pulses when

input current is at upper limit. This scheme provides a total

dosage of the radiation over long period but is not fast enough for

accelerator beam protection.

In our new scheme, the same recycling integrator

output is sent to an FPGA in which a time-to-

digital converter (TDC) is implemented. The TDC

is based on a multi-sampling scheme in which the

input transition is sampled with four different

phases of the system clock. With system clock

rates of 250 MHz and four-phase sampling, a 1-

ns time measurement resolution can be achieved.

Page 23: Beam Loss monitoring R&D

TDC Implemented with FPGA

There are two popular schemes for FPGA TDC:

Multiple sampling based scheme: LSB: 0.6 to 1 ns.

Delay line based scheme: LSB: 40 to 100 ps.

We are currently working on a variation of the delay line based TDC called

Wave Union TDC.

pulse input to 8

Channel LVDS

Cyclone III FPGA

Arden Warner, FNAL ([email protected])

Page 24: Beam Loss monitoring R&D

Multi-Sampling TDC FPGA

24

Ultra low-cost: 48 channels in $18.27 EP2C5Q208C7.

Sampling rate: 250 MHz :

x4 phases = 1 GHz.

LSB = 1 ns.

Multi-Sampling TDC FPGA

This picture represent a placement in Cyclone FPGA

4Ch

c0

c90

c180

c270

c0

MultipleSampling

ClockDomainChanging

Trans. Detection& Encode

Q0

Q1

Q2

Q3QF

QE

QD

c90

Coarse TimeCounter

DV

T0T1

TS

Logic elements with non-critical timing are freely placed by the fitter of the compiler.

Arden Warner, FNAL ([email protected])

Page 25: Beam Loss monitoring R&D

Bench-top measurements with FPGA-

TDCI=Q/dt

Input Current

100 nA/div

Output Pulses

Input current and out-put pulses

FPGA-TDC data generated by measuring time

between pulses

A Scheme using FPGA-based time-to-digital converter (TDC) to measure time

intervals between pulses output from the recycling integrator is employed to

ensure a fast beam loss response along with a current measurement resolution

better than 10-bit.

Arden Warner, FNAL ([email protected])

Page 26: Beam Loss monitoring R&D

Arden Warner, FNAL26

VME based counter/Timer

boardHTS installation

Counter/timer show 630 counts = 150 mR

Dark current measurements at Photo-injector and HTS

Test cavity

Initial cold measurements were done at the

horizontal test stand (HTS) shown here. A VME

based counter timer board was used to count

pulse in ACNET: Counter/timer showed counts

corresponding to 150mR

Test are now being done using the FPGA-TDC

method which is faster with better resolution

Loss due to Dark current background at A0-photo-injector.

Measured to be ~ 400 nA downstream of bend magnet

40 µs rf gate (dark current only no

photo-electrons injected)

Page 27: Beam Loss monitoring R&D

Design improvements and modifications

Arden Warner, FNAL27

Final test are underway with FNAL designed FPGA-TDC at HTS

We have increased the pressure from 1bar to 1.5bars to establish the best operating point for

the device. The calibration “S” of the monitors is almost completely determined by the volume

“V” of the enclosed gas and by the type of gas:

S ≈ V ρ . e/Eion (“ρ” is gas density and Eion is mean energy deposition to create electron-ion pairs)

We had Bridgeport Instruments modify they FPGA code in the recycling integrator electronics

box so that the leading edge of the discriminator can be seen at the NIM port output. This

would improve the resolution of the TDC measurement between pulses.

A mechanical scheme to easily mount the loss monitor in a cryomodule near the quads and

BPMs is being done as shown in the following example.

Page 28: Beam Loss monitoring R&D

SRF=Superconducting Radio Frequency

1.3 GHz 9-Cell SRF Cavity

8 Cavity SRF Cryomodule

SCRF Research at Fermilab

Arden Ayube Warner 6/15/2010

Page 29: Beam Loss monitoring R&D

Arden Warner, FNAL29

Page 30: Beam Loss monitoring R&D

Arden Warner, FNAL30

Cryogenic Loss Monitor proposed installation in CM2

Page 31: Beam Loss monitoring R&D

Proposed installation in cryomodule II

Arden Warner, FNAL31

G10/11 Plate

with t-slot

BLM with

t-slot flange

(welded)

SS clamp plates

BPM

QUAD