beam loss monitoring for mps - frib · multi-layer detection requirement for fps, slide 7 threshold...

17
This material is based upon work supported by the U.S. Department of Energy Office of Science under Cooperative Agreement DE-SC0000661. Michigan State University designs and establishes FRIB as a DOE Office of Science National User Facility in support of the mission of the Office of Nuclear Physics. Zhengzheng Liu Beam Diagnostics Physicist Beam Loss Detection for MPS at FRIB

Upload: others

Post on 09-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

This material is based upon work supported by the U.S. Department of Energy Office of Science under Cooperative Agreement DE-SC0000661.

Michigan State University designs and establishes FRIB as a DOE Office of Science National User Facility in support of the mission of the Office of Nuclear Physics.

Zhengzheng Liu

Beam Diagnostics Physicist

Beam Loss Detection for MPS at FRIB

Page 2: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Beam loss induced damage

Layered requirement for beam loss detection

Diagnostic devices for beam loss detection• Radiation detectors

• Differential beam current monitor (DBCM)

• Differential beam position monitor (DBPM)

• Halo monitor ring (HMR)

Differential current monitoring for MPS

Early-operation / commissioning planning of loss detection

Beam loss monitoring system for commissioning

Outline

Z. Liu, March 2014 MPS workshop - 7B, Slide 2

Page 3: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Fast Beam Loss Induces Thermal DamageDeposited Energy Matters !

, Slide 3

The worst case (uranium beam ~20MeV/u): may damage a SS bellow in less than 40 µs

LS1

LS2

LS3

20 MeV/u

8.4 pmA

40 kW

149 MeV/u

8.4 pmA

297 kW35 ms

1.4 J

60 ms

17.8 J

Z. Liu, March 2014 MPS workshop - 7B

Page 4: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Errant Beam Loss Induces SRF DegradationExample: Errant beam mitigation at SNS*

, Slide 4

• Errant beam: off-energy beam generated and

transported to the downstream

• Errant beam hits cavity surface, desorbs

gas/particulates, increases possibilities of

arcing/discharge and leads to degradation

• e.g. two coupler windows leaks in 6c and 20d

presumably from 2009

• Historically most degradation has been

recovered by thermal cycling of cryomodules

• MPS detects errant condition from SNS RF system

or BLM and trigger MPS in 15µs

• <10% of BLM trips were due to ion source/LEBT

• Most ion sources induces BLM trips during the

first week of new source installation

• >90% of BLM trips were due to warm LINAC RF

faults

• Adequate BLM and shorter beam stop time is

wanted to reduce the degradationQuote from S. Kim’s talk on 3/3/2014 cryomodule workshop

Odd beam from front end

Z. Liu, March 2014 MPS workshop - 7B

Page 5: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

The ratios of dose rates from heavy-ion beams to the dose rate from the proton beam are very nearly the same as the ratios of the produced neutrons*

Scale the residual dose rate for heavy ions relative to the 1 W/m, 1GeV proton beam loss. In the case of slow losses, FRIB ions produce low activations

Activation from 1 W/m Slow Losses Is Much Lower Than the Hand-on Maintenance Limit

Ion

(Specific

Energy in

MeV/u)

Prompt

Neutron Fluxes

from 1 W/m

loss relative to

that from

proton beam

[%]

Residual

Dose Rate

[mrem/h]

H (1000) 100.0 98.9

238U (200) 1.5 1.5

18O (300) 12.2 12.1

, Slide 5

* R. Ronningen “Studies of limits on uncontrolled heavy-ion

beam losses for allowing hands-on maintenance”

Z. Liu, March 2014 MPS workshop - 7B

Page 6: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

The present slow loss detection criterion has been set to 1 W/m, corresponding to 10-5 ~ 10-6 fractional power loss per meter• An average 1 W/m beam loss adds ~250 W heat load to the cryoplant, which

is 10% of the total 2 K design load

• The design margin for cryoplant heat load is limited, ~100%

Will the SRF cavity experience long-term degradation from the low-level chronic beam loss?

The current slow loss threshold will not trigger FPS trips. It is an optional detection criteria for commissioning/tuning

Slow Beam Loss Adds Cryogenic Load

, Slide 6Z. Liu, March 2014 MPS workshop - 7B

Page 7: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

The criterion for fast beam loss MPS is to mitigate the beam energy deposition for each event

Adopt a lost beam energy limit, based on damage thresholds –

20 kW x 35 µs = 1.4 J

Example design detection/mitigation schema for Fast MPS (FPS)

Beam current from ion source fluctuates on order ~10% over tens of msMonitoring at fluctuation thresholds below ~10% will require feed-forward from the injector current monitoring system

Multi-Layer Detection Requirement for FPS

, Slide 7

Threshold of

full current

Beam

detection

time (ms)

Beam

mitigation

time (ms)

Maximum beam energy loss

1 100% 15 35 20 kW x 100% x 35 ms = 1.4 J

2 10% 330 350 20 kW x 10% x 350 ms = 1.4 J

2 1% 3480 3500 20 kW x 1% x 3500 ms = 1.4 J

Z. Liu, March 2014 MPS workshop - 7B

Page 8: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Traditional Radiation Detectors include Ion Chamber and Neutron Detector. However, they are not sufficient for FRIB superconducting linac segments because• Cavity X-ray background can be several rad/hr

• Radiation cross talk from LS3 overshadows LS1 and LS2

Radiation detectors are still useful at high power deposition areas and for tuning purpose

Radiation Detectors

, Slide 8

Gamma intensity from LS3 decreases to

7.5% for a line loss and 0.6% for a point

loss, it is still larger than LS1 loss signal.

Neutron radiation is more penetrating than

gamma and therefore its cross talk effect is

even worse.

Z. Liu, March 2014 MPS workshop - 7B

Page 9: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Example: LS1 differential currents• Current difference measurements are performed in <10 µs with a current

resolution better than 4 µA (~1% of full beam current)

Design Example of DBCM (Bergoz BCM)

Beamline

Segment

Time of flight

(µs)

LEBT 1.3

RFQ 0.7

MEBT 0.7

LS1 2.4

FS1 1.4

LS2 1.2

FS2 0.3

LS3 and BDS 0.9

, Slide 9Z. Liu, March 2014 MPS workshop - 7B

Page 10: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Experiment was set up at ReA3 with FermilabBPM receivers• Measures second harmonic 161 MHz

• 81 dB gain pre-amp; 1.32 dB cable loss

• Effective BW ~37kHz (ԏ~4.3µs)

For 37 kHz, the calculated intensity RMS resolution (std) is 126 nA

From experiment, the RMS intensity resolution (std) for single BPM is measured as 67-106 nA for 204800 samples, assuming beam is at the center. The differential intensity resolution is ~140 nA

FRIB BPM intensity resolution is comparable with ReA3 BPM, and it features fast evaluation (~15µs)

However, the BPM resolution is very sensitive to beam position and beam velocity. Calibration is necessary

Differential Current Monitoring by BPM Pair

, Slide 10

rms bunch lgth

beta MeV/u rms deg mm

rfq out 0.03275 0.5 2.7 0.9

ls1 in 0.03275 0.5 1.8 0.6

ls1 out 0.18647 16.63 0.9 1.7

ls2 out 0.50624 148.63 0.6 3.1

ls3 out 0.56985 202.06 4.7 27.7

target 0.56985 202.06 7.5 44.2

6

nA-rms

323

322

763

2023

2306

2355

FRIB BPM intensity resolution

Linear and Polynomial fit for BPM intensity

Z. Liu, March 2014 MPS workshop - 7B

Page 11: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

The halo ring monitor is a niobium ring designed to intercept ions in the halo of the beam that are likely to be lost farther downstream

It has high sensitivity (~0.1nA) for integrated small signal and fast response time (~10 µs) for large signal

Drawback: Its aperture is hard to determine by simulation. Have to use large aperture and may skip 1st phase commissioning to avoid possible limitation

Halo Monitor Ring (HMR)

, Slide 11

HMR measurement at NSCL with 18O3+ at 11 MeV/u

Z. Liu, March 2014 MPS workshop - 7B

Page 12: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

There are two hot spots of potential beam loss inside the cryomodule, the drift space before and after solenoid

We consider a localized slow loss on beam pipe, 0.05 W on a 1cm2

spot, the resulted rising time at the hot spot is ~2 seconds / 0.1 K. After 10 second, the temperature is 5 K

Beam Pipe Temperature Sensor Inside Cryomodule

, Slide 12Z. Liu, March 2014 MPS workshop - 7B

Page 13: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Differential Current Monitoring for FPS

Beam Current Monitoring

based on ACCTsCharge

Stripper

Differential current monitoring is mainly for large beam losses

Current from ion source fluctuates at 10% in a time scale of tens of ms,

hence the differential current monitoring is less effective for small losses

BPM and BCM differential current measurement should be cross-calibrated

, Slide 13

Beam Current Monitoring

based on BPMs

Z. Liu, March 2014 MPS workshop - 7B

Page 14: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

MPS trips trigger beam abort within 35 µs • Ion sources

» Removal of extraction HV ceases beam production

• LEBT E-bends» Removal of HV prevents low energy beam to further

acceleration downstream.

After abort up to 10 µs of beam remains in pipe• CW operation continuously generates beam

• Downstream BID may be required to further limit unplanned beam energy deposition

Beam Inhibit Devices for Machine Protection

, Slide 14

Front End

Ion sources

E-bends

RFQ

Z. Liu, March 2014 MPS workshop - 7B

Page 15: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

For initial pulsed beam (50 eµA, 50 µs, <1 Hz)• Total lost beam energy deposition is below damage threshold

• May not require FPS in this mode

For initial, low power CW beam (<340 enA K17+, 344 W full power)• Requires beam mitigation in ~50 ms

• Conductive thermal flux may alleviate the energy deposition limit

Fast beam loss monitoring could be performed by differential beam current measurements• Demonstrate FPS beam mitigation at fastest time scales

Slow loss beam monitoring will be calibrated and commissioned• Demonstrate detection limits, response time, crosstalk

Commissioning Approach

, Slide 15Z. Liu, March 2014 MPS workshop - 7B

Page 16: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Fast protection systems will be commissioned at necessary response times

Slow loss MPS can serve any mode once it has been established

Beam loss thresholds and faults will require study for various operating modes and ion species

Low power CW operation (mode O2) may require dedicated Faraday cups in warm sections to calibrate beam current monitors

Dynamic Ramping is a mode of special interest• Quickly evolving pulse length and duty cycle over time scale of seconds

• Data acquisition may require reconfiguration from gated to continuous mode

Operations Approach

, Slide 16Z. Liu, March 2014 MPS workshop - 7B

Page 17: Beam Loss Monitoring for MPS - FRIB · Multi-Layer Detection Requirement for FPS, Slide 7 Threshold of full current Beam detection time (ms) Beam mitigation time (ms) Maximum beam

Neutron detectors

Beam Loss Monitoring Network for Commissioning

Beam Pipe

Temperature Sensor

• Slow response time

• Commissioning segment by

segment

• 2 K/4 K header heater PS current

• Novel techniques?

• Si/CVD diamond detectors

• LHe ionization detectors

Ion chambers

, Slide 17Z. Liu, March 2014 MPS workshop - 7B