bassett healthcare point of care i-stat 1 procedurestore at 2 to 8 c. do not use after expiration...
TRANSCRIPT
![Page 1: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/1.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 1 of 38
Bassett Healthcare
Point of Care
i-STAT 1 PROCEDURE
A. PRINCIPLE
The i-STAT System incorporates comprehensive components needed to perform blood
analysis at the point of care. The System consists of the i-STAT analyzer, with the
docking station and printer, test cartridges, and middleware, RALS 3.
B. METHOD
1. i-STAT Analyzer
When a sample-filled i-STAT cartridge is inserted into an analyzer for analysis, the
analyzer automatically controls all functions of the testing cycle including fluid
movement within the cartridge, calibration and continuous quality monitoring.
Results are reported in approximately 120 seconds for cartridges with sensors for
blood gases, electrolytes, chemistries and hematocrit. ACT and PT/INR cartridge
results within the time it takes to complete the test. The Troponin cartridge takes 10
minutes to complete the test.
2. Cartridges
A single-use disposable cartridge contains microfabricated sensors, a calibrant
solution, a fluidics system, and a waste chamber. Sensors for analysis of pH, pCO2,
pO2, sodium, potassium, chloride, ionized calcium, glucose, creatinine, urea nitrogen
(BUN), lactate and hematocrit are available in a variety of panel configurations.
Reactions for a particular test are the same across cartridges. Cartridges are also
available for ACT, PT/INR and troponin.
3. Point-of-Care Central Workstation (Database or middleware)
A dedicated desktop computer with the i-STAT program provides the primary
information management capabilities for the i-STAT System. IR links for Portable
Clinical Analyzers, down loaders for the i-STAT 1 Analyzers and RALS allow for
transmission of patient records from a widely distributed network of analyzers. Data
is stored, organized, edited, and transferred to the laboratory information system or
other computer system. Cartridge usage and efficiency reports can be generated for
management of the system.
C. SPECIMEN REQUIREMENTS - A whole blood sample of approximately 1 to 2 drops,
sample size from 17 ul to 95 ul, is dispensed into the cartridge sample well and the sample
well is sealed.
I. MODERATELY COMPLEX TESTS
a. REQUIRED for G3, EG7, and CG4, pH, pCO2, PO2 (blood gas component
only) - Acceptable Samples: Arterial, Venous, Cord, Capillary and Mixed Venous
TEST IMMEDIATELY:
Fresh whole blood collected in a plastic syringe, no anticoagulant – run
immediately
![Page 2: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/2.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 2 of 38
TEST WITHIN 30 MINUTES:
Fresh whole blood collected in a Lithium Heparinized plastic syringe -
minimum draw 0.8 ml. If testing is delayed stored capped, NOT on ice. Mix
well and dispense 1-2 drops before sampling.
Fresh whole blood collected in a Lithium Heparinized Vacutainer submitted at
room temperature submitted to the Clinical Lab ONLY.
Fresh whole blood collected in a Balanced Heparin capillary collection tube at
room temperature.
b. SUITABLE for EG7, Chem8 and CG4 for Electrolytes only, ICAL,
Hematocrit. Troponin and Lactate
Fresh whole blood collected in a Lithium Heparin (light green) Vacutainer
tube. Minimum draw half full. Test within 30 minutes.
NOTE: Lactate is drawn without a tourniquet and an unclenched fist.
c. SUITABLE for Troponin and CG4
Fresh whole blood collected in a Lithium Heparin (light green) Vacutainer
tube. Minimum draw half full. Test within 30 minutes.
d. SUITABLE for PT/INR – Testing for patients on Sodium Warfarin therapy only.
NOTE: Do not excessively squeeze patient finger. Pre-warm hands to increase
capillary perfusion.
Fingerstick specimens (preferred sample): Using a lancet, follow fingerstick
protocol to obtain a capillary sample. Gently squeeze the finger to develop a
hanging drop of blood. DO NOT WIPE away the first DROP. Collect sample
in i-STAT sample well (20 uL). Test immediately.
Fresh whole blood without anticoagulant collected in a plastic syringe. If
from an indwelling line, flush the line with 5 ml saline and discard the first 5
ml of blood or three to six dead space volumes of the catheter. Test
immediately.
e. SUITABLE for ACT
Fresh whole blood without anticoagulant collected in a plastic syringe. If
from an indwelling line, flush the line with 5 ml saline and discard the first 5
ml of blood or three to six dead space volumes of the catheter. Test
immediately.
II. WAIVED TESTS
a. REQUIRED for E3
Fresh whole blood collected in a Lithium Heparin (light green) Vacutainer
tube. Minimum draw half full. Test within 30 minutes. Transfer the sample
with a needless device or a pipette.
![Page 3: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/3.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 3 of 38
b. SUITABLE for Creatinine
Fresh whole blood collected in a Lithium Heparin (light green) Vacutainer
tube. Minimum draw half full. Test within 30 minutes. Transfer the sample
with a needless device or a pipette.
NOTE: Waived cartridges – E3 and Creatinine must be tested on venous
samples collected in evacuated tubes with Lithium Heparin.
Specimen Labeling
Unless the specimen is analyzed immediately after collection and then discarded, the
specimen container must be labeled with the following information:
Patient first and last name
Patient ID number (visit #)
Time and date of collection
Phlebotomist ID
Sample Volumes
E3 – 65 uL G3 – 95 uL
Creatinine – 65 uL CG4 – 95 uL
Chem8 – 95 uL EG7 – 95 uL
ACT – 40 uL Tropo nin – 17 uL
PT/INR – 20 uL
TABLE #1 – SPECIMEN TYPES BY TEST
Heparin
Syringe
No Anticoagulant
Syringe
Capillary Fingerstick Lithium
Heparin
Vacutainer
Measured Analyte
pH X X X Lab Only
pCO2 X X X Lab Only
pO2 X X X Lab Only
ICAL X X X** X
Hct X X X X
Na X X X X
K X X X** X
Troponin X X X
Lactate X X X
Creatinine X* X X
BUN X X X
CL X X X
Glucose X X X
TCO2 X X X
PT X
ACT X
*Chem8 cartridge only
**Capillary blood is known to show statistically different values to venous blood in glucose,
potassium and calcium.
Criteria for Specimen Rejection
![Page 4: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/4.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 4 of 38
Evidence of clotting
Specimens collected in vacuum tubes with anticoagulant other than lithium or
sodium heparin
Specimen for ACT collected in glass syringe or tube or with anticoagulant of
any kind
Syringe for pH, pCO2, and pO2 with air bubbles in sample
Incompletely filled vacuum tube for the measurement of ionized calcium or
pCO2
Other fluid types such as urine and CSF – POC use only. (Lab may analyze
fluids according to outlined procedure)
Syringes sent to laboratory with needle attached
Creatinine patient receiving Hydroxyurea
D. REAGENTS AND MATERIALS/INSTRUMENTATION
1. Cartridges
a. Storage:
Cartridges are sealed in individual pouches. Cartridges must be used within 10
minutes from removal of original packaging. Store the main supply of cartridges
refrigerated at 2C to 8C. Remove cartridges from refrigeration prior to use, allow
to equilibrate at room temperature. Cartridges cannot be returned to the refrigerator
once they have been at room temperature. Do not allow cartridges to freeze.
Freezing will cause higher than expected ionized calcium results. Do not store
cartridges at temperatures greater than 30C.
NOTE:
Cartridges for testing must acclimate to room temperature.
1 cartridge = 5 minutes
1 box cartridges – 1 hour
Date all boxes with RT (room temperature) expiration date
Date individual cartridges with RT (room temperature) expiration date
DO NOT USE cartridges after labeled expiration date OR manufacturer’s
expiration date
TABLE #2 – CARTRIDGE STABILITY AT ROOM TEMPERATURE
Cartridges Room Temp Storage
Moderately Complex
ACT - Activated clotting time 14 Days
*PT/INR – INR 14 Days
cTnI – Troponin 14 Days
G3+ - pH, pCO2, pO2 2 Months
CG4 - pH, pCO2, pO2, Lactate 2 Months
CHEM 8 - NA, K, CL, TCO2, iCal, Glucose, Bun,
Creat 14 Days
Waived
Crea - Creatinine 14 Days
E3+ - Na, K, HCT 2 Months
![Page 5: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/5.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 5 of 38
*PT/INR cartridges are only used in the CVRU, MRI and ASU for pre-procedure
clearance for patients taking Sodium Warfarin. Any INR’s >1.9 require a venous
confirmation.
i-STAT Testing Locations and Approved Test by Site: i-STAT testing is approved by patient
care area. Please refer to Table 3 to confirm approved testing.
TABLE #3 – APPROVED TESTS BY LOCATION MIBH POC
TEST SITE G3 or EG7 ACT CG4 PT/INR Troponin
Creatinine
E3
ASU X
X
Cardiac OR X X
Cath Lab X X X
CVRU X
X
ED
X X
GI Lab
X
ICU X X X
Little Falls
Dialysis
X
OB InPt X
OR X X
OSS Dialysis
X
PACU OPCU X X X
Respiratory X X
MRI/CT Scan
X X
b. Cartridge Storage Conditions
Refrigerated Cartridges – Store 2-8C
Verify that the cartridges stored in the refrigerator are within the
expiration date printed on the boxes. Discard expired cartridges.
Verify that the refrigerator did not exceed the limits of 2 to 8C. Do not
use cartridges stored below 2C, discard. Cartridges stored above 8C
must be dated for room temperature expiration.
Document refrigerator temperature on the temperature log daily.
Room Temperature Cartridges (18-30°C)
Verify that the cartridges stored at room temperature have not exceeded
the room temperature expiration date or manufacturer’s expiration date.
Discard all cartridges that have expired.
Verify that room temperature is between 18-30°C.
Document temperature on temperature log if applicable.
c. Ordering information:
Cartridges: Notify the Laboratory Point of Care Office at X6728 (M-F
between 0800 and 1630, or the Chemistry Lab at X3731 during other times
when cartridges are required. Cartridges are hand delivered to the testing
location as cartridges cannot be delivered through the pneumatic tube system.
2. i-STAT 1 – Downloadable version (gray) or wireless version (blue).
![Page 6: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/6.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 6 of 38
3. i-STAT 1 Downloader/Recharger Kit – Abbott reference #04P73-04, DRC 300
The Downloader comes in two formats:
Downloader: A low-profile table-top unit with “arms” between which the
analyzer is placed
Downloader/Recharger (DR): A cradle that the analyzer is placed within.
4. Quality Controls
a) TriControls for blood gases, electrolytes, chemistries and hematocrit – Abbott
Manufacturer reference #05P71-01, 05P72-01, 05P73-01
Store at 2 to 8C. Do not use after expiration date on the ampules and box.
Warm a minimum of 4 hours at room temperature before use.
Room temperature storage is 5 days.
b) Controls for ACT and PT/INR – Abbott Manufacturer reference # ACT 06P17-
15, 06P17-16; PT/INR 06P17-13, 06P17-14
Store at 2 to 8C. Do not use after expiration date on the vials and box.
Warm minimum 45 minutes at room temperature before use.
Room temperature storage is 4 hours.
c) Control for Troponin – Abbott Manufacturer reference #06P17-09, 06P17-10,
06P17-11
Store 2 to 8C.
Use immediately, return to refrigerator.
Open expiration date is 30 days.
Do not use after expiration date on the vials and box.
5. Linearity – Tricontrols Abbott Manufacturer reference #05P70-01 – 5 levels.
Store at 2 to 8C. Do not use after expiration date on the ampules and box.
Warm a minimum of 4 hours at room temperature before use.
Room temperature storage is 5 days.
6. Electronic Simulator Store at room temperature and protect contact pads from contamination by replacing
the plastic cap and placing the Electronic Simulator in its protective case after use.
![Page 7: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/7.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 7 of 38
7. Specimen Collection Equipment a) Blood Gas, Electrolytes, Chemistries and Hematocrit. Lithium Heparin is the
preferred anticoagulant.
NOTE: Sodium Heparin can be used for all tests EXCEPT ELECTROLYTES
(sodium and potassium)
Arterial Puncture – ProVent Arterial Blood Sampling Kit - reference 4699P-1
Smiths Medical. 23.5 IU dry Lithium Heparin – neutralized for ionized
calcium. Minimum sample 0.8 mL or plain untreated plastic syringe.
Line Draw - Line Draw Arterial Blood Sampling Kit – reference 4043W-2
Smiths Medical. 23.5 IU dry Lithium Heparin – neutralized for ionized
calcium. Minimum sample 0.8 ml or plain untreated plastic syringe.
Venipuncture – Lithium Heparinized collection tube and disposable transfer
device (1 cc syringe; 16-20 gauge needle)
Capillary – Safe-T-Fill Balanced Heparin capillary collection – reference 06-
0186 (150 uL). FOR BLOOD GAS ONLY
b) ACT - Untreated plain plastic syringe, NO ANTICOAGULANT.
c) PT/INR – Lancet fingerstick supplies or a plain untreated plastic syringe.
8. Clorox Germicidal Wipes – Warehouse order #554713
9. Rechargeable Battery Pack – Abbott i-STAT reference #06F23-55
10. 9V Lithium Batteries – Warehouse Stock #U9VL
11. Printer Paper – Warehouse Stock #06F17-11
Calibration/Linearity
1. For cartridges, calibration is automatically performed as part of the test cycle on each
cartridge type. Calibration is not required for ACT, PT/INR cartridges
2. Each new meter and each new lot G3, EG7, CG4, Chem8 and Troponin requires
linearity before use. Run the Linearity TriControls five sample calibration
verification kit for G3, EG7, CG4 and Chem8. Run Level 1-3 Troponin QC or
Linearity for Troponin. This procedure is performed by the POC office and the
information is used for instrument correlation as well as linearity.
E. LINEARITY REQUIREMENTS
TABLE #5 – LINEARITY REQUIREMENTS
Analytes Linearity Frequency
G3, EG7, CG4.Chem8 Level 1, Level 2, Level 3. Level 4, Level 5 Each shipment and each new lot #
Troponin Level 1, Level 2, Level 3 Each shipment and each new lot #
![Page 8: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/8.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 8 of 38
PROCEDURE – Linearity
1. Procedure for testing cartridges with i-STAT Level 1-5 Calibration Verification
a) Prior to testing cartridges that measures pO2, ampules should stand at room
temperature a minimum of 4 hours before use. When testing other cartridges,
ampules may be used once the fluid has reached room temperature, approximately
30 minutes for individual ampules. For best results, ampules, cartridges and
analyzers should be at the same temperature.
b) Immediately before use, shake the ampule vigorously for 5-10 seconds to
equilibrate the liquid and gas phases. To shake, hold the ampule at the tip and
bottom with forefinger and thumb to minimize increasing the temperature of the
solution. If necessary, tap the tip of the ampule to send solution back into the
bottom section of the ampule. Protect fingers with gauze, tissue, or glove, or use
an ampullate breaker to snap off the tip of the ampule at the neck.
c) Immediately transfer the solution from the ampule into a syringe.
When using a syringe (1 cc or 3 cc syringes with 16 to 20 gauge needles are
recommended), slowly draw approximately 1 mL of solution from the bottom
of the ampule. If air is trapped between the leading edge of the solution and
the plunger, do not invert the syringe to expel it; this will not affect solution
near the tip of the syringe. If air bubbles are continually drawn into the
syringe, or if a bubble is trapped near the tip of the syringe, discard the
ampule and syringe and use a fresh ampule or syringe. Expel one or two
drops from the syringe before filling the cartridge.
Do not use solution left in the syringe, ampule, or capillary tube for additional
testing of the cartridges that contain sensors for ionized calcium, pH, pCO2 or
pO2. However, cartridges without these sensors may be tested with remaining
fluids if within 10 minutes of opening the ampule.
d) Immediately transfer the solution to the cartridge.
e) On the i-STAT select MENU, #3 Quality Tests, #3 Cal Verification.
Linearity Documentation
Print the linearity record from RALS. Compare each level of linearity to the value
assignment sheet (print at www.abbottpointofcare.com), taking note of appropriate
cartridge, cartridge lot alpha character and CLEW version number. Date and initial
each sheet for review. Complete the i-STAT linearity coversheet. Attach i-STAT
linearity coversheet and the value assignment sheet to the linearity record printouts.
File completed record in the i-STAT linearity file cabinet by cartridge type. Sticker
cartridge boxes “This lot is Ready for Use” (green).
Remedial Action:
If any results are outside the published expected ranges:
DO NOT USE cartridges from the suspect lot.
Quarantine the suspected lot.
POC will follow up with technical support.
![Page 9: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/9.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 9 of 38
F. QC REQUIREMENTS
TABLE #4 – EXTERNAL QC
From each lot of cartridges received, run the appropriate levels of control material. Controls
must be run before putting cartridges into use and monthly thereafter on each lot available.
See below for Point of Care stickers indicating use. The QC samples must be rotated amongst
all testing personnel from all testing sites. The Point of Care Team will coordinate the
scheduling of staff and maintain quality control documentation in the laboratory.
1. Internal Calibration The performance of each handheld analyzer in the i-STAT system is verified by the
internal Electronic Simulator with every cartridge use. If Simulator FAIL is
displayed on the analyzer screen, run the external simulator.
2. External Simulator – The external electronic simulator is a quality control device for
the analyzer’s cartridge signal-reading function. It simulates two levels of electrical
signals that stress the analyzer’s cartridge signal detection function both below and
above measurement ranges.
Remedial Action:
Turn the analyzer on
Select menu
#3 Quality Tests
#4 Simulator
Scan or enter operator ID
Scan or enter simulator ID
Insert simulator
Analytes Controls Frequency
WAIVED
E3 Level 1, Level 2, Level 3 3 levels, once per shipment
Creatinine Level 1, Level 2, Level 3 3 levels, once per shipment
MODERATELY COMPLEX
G3, EG7, CG4 Level 1, Level 2, Level 3 3 levels monthly, each lot in use
ACT Level 1 and Level 2 ACT Control 2 levels monthly, each lot in use
PT/INR Level 1 and Level 2 PT/INR
Control
2 levels monthly, each lot in use
Troponin Troponin Level 1, Level 2, Level 3 2 levels monthly – one normal
level, one abnormal level, each
lot in use
Chem8 Level 1, Level 2, Level 3 3 levels monthly, each lot in use
Refrigerate Immediately Store 2-8C, do not allow to come to room temperature
DO NOT USE this lot number QC has not been run, do not use for patient testing
This lot is ready for use QC has been run, satisfactory for patient testing
![Page 10: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/10.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 10 of 38
Allow simulator to complete 60 second test cycle. Verify that the simulator has
passed.
Remove by paging to #1 Test Options.
If PASS is displayed use the analyzer as required.
If FAIL is displayed repeat the procedure with a different external Electronic
Simulator.
If PASS is displayed with the second Electronic Simulator:
Use the analyzer as required.
If FAIL is displayed with the second Electronic Simulator:
DO NOT analyze patient samples with the analyzer. Return i-STAT and
Simulator to the POC office.
G. PROCEDURE – QC
1. Procedure for testing cartridges with i-STAT TriControls Level 1, 2 and 3
Controls
a) Prior to testing cartridges that measures pO2, ampules should stand at room
temperature a minimum of 4 hours before use. When testing other cartridges,
ampules may be used once the fluid has reached room temperature, approximately
30 minutes for individual ampules. For best results, ampules, cartridges and
analyzers should be at the same temperature.
b) Immediately before use, shake the ampule vigorously for 5-10 seconds to
equilibrate the liquid and gas phases. To shake, hold the ampule at the tip and
bottom with forefinger and thumb to minimize increasing the temperature of the
solution. If necessary, tap the tip of the ampule to send solution back into the
bottom section of the ampule. Protect fingers with gauze, tissue, or glove, or use
an ampullate breaker to snap off the tip of the ampule at the neck.
c) Immediately transfer the solution from the ampule into a syringe.
When using a syringe (1 cc or 3 cc syringes with 16 to 20 gauge needles are
recommended), slowly draw approximately 1 mL of solution from the bottom
of the ampule. If air is trapped between the leading edge of the solution and
the plunger, do not invert the syringe to expel it; this will not affect solution
near the tip of the syringe. If air bubbles are continually drawn into the
syringe, or if a bubble is trapped near the tip of the syringe, discard the
ampule and syringe and use a fresh ampule or syringe. Expel one or two
drops from the syringe before filling the cartridge.
Do not use solution left in the syringe, ampule, or capillary tube for additional
testing of the cartridges that contain sensors for ionized calcium, pH, pCO2 or
pO2. However, cartridges without these sensors may be tested with remaining
fluids if within 10 minutes of opening the ampule.
d) Immediately transfer the solution to the cartridge.
e) On the i-STAT select MENU, #3 Quality Tests, #1 Control.
f) Compare the results to the value assignment sheets taking note of appropriate
cartridge, cartridge lot alpha character and CLEW version. QC ranges that have
been uploaded to RALS for eVAS should be corrected and appear on the QC
report.
![Page 11: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/11.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 11 of 38
g) Remedial Action:
If any results are outside the published expected ranges:
DO NOT USE cartridges from the suspect lot.
Quarantine the suspected lot.
POC will follow up with technical support.
2. Procedure for testing cartridges with i-STAT Level 1 and Level 2 ACT and
PT/INR Control: a) On the i-STAT select MENU, #3 Quality Test, #1 Control.
b) Prior to use, allow one vial each of the lyophilized plasma and calcium chloride
reconstituting fluid to stand at room temperature for a minimum of 45 minutes.
c) Remove the cap and stopper from the vials and pour the entire contents of the
calcium chloride vial into the lyophilized plasma vial. Place the stopper back on
the reconstituted vial.
d) Allow the vial to sit for 1 minute and then mix the contents by swirling gently for
1 minute, then inverting slowly for 30 seconds.
e) Use a plastic pipette, syringe, or capillary tube without anticoagulant to transfer
the solution to an ACT cartridge.
f) Immediately seal the cartridge and insert it into an analyzer. This process must be
completed within 30 seconds of the complete reconstitution of the control sample.
g) Compare the results to the value assignment sheets taking note of appropriate
cartridge, cartridge lot alpha character and CLEW version. QC ranges that have
been uploaded to RALS for eVAS should be correct and appear on the QC report.
h) Remedial Action: If any results are outside the published expected ranges:
Quarantine and DO NOT USE cartridges from the suspect lot.
Record the QC failure in the i-STAT QC Action Log along with the action
taken.
3) Procedure for testing cartridges with i-STAT cTnl (Troponin) Control Level 1, 2
and 3:
a) On the i-STAT select MENUS, #3 Quality Test, #1 Control.
b) Remove vial from refrigerator, run immediately.
c) Thoroughly mix by gently swirling the bottle. Avoid foaming of the sample.
d) Dispense a drop of sample directly from vial into the i-STAT cTnl cartridge and
seal the cartridge. Tightly recap the bottle immediately after all sampling is
complete and store at 2-8C. Open vials are stable for 30 days at 2-8C.
e) Compare the results to the value assignment sheets taking note of appropriate
cartridge, cartridge lot alpha character and CLEW version. QC ranges that have
been uploaded to RALS for eVAS should be corrected and appear on the QC
report.
f) Remedial Action: If any results are outside the published expected ranges:
Quarantine and DO NOT USE cartridges from the suspect lot.
Record the QC failure in the i-STAT QC Action Log along with the action
taken.
H. QC DOCUMENTATION: Print QC record from RALS for each lot of QC run. Date
and initial each sheet for review. Acceptable ranges should appear on QC record. If
ranges are not available, notify technical specialist and refer to value assignment sheet
![Page 12: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/12.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 12 of 38
(print at www.abbottpointofcare.com). Place in monthly i-STAT QC book. File by
cartridge type.
I. EPIC ORDER CODES
Arterial blood gas line draw (RT85)
Arterial blood gas puncture (RT43)
POC i-STAT blood gas other (POC215) – (cord, venous, mixed venous or
capillary)
J. PROCEDURE FOR ANALYSIS
1. Procedures for Specimen Collection a) In-Dwelling Line
Back fill the line with sufficient amount of blood to remove intravenous
solution, heparin, or medications that may contaminate the sample.
Recommendation: three to six times the volume of the catheter, connectors,
and needle. If collecting sample for ACT, clear the line first with 5 mL saline
and discard the first 5mL of blood.
b) Venous Sample
Collect the sample using phlebotomy procedure. Use appropriate collection
device for cartridge/test required. Refer to Table 1.
NOTE: Mix blood and anticoagulant by rolling the syringe between palms for at least
5 seconds and inverting the syringe 5 seconds. Mix vacutainer by inverting for 5
seconds. Remove blood from tube using needless transfer pipette.
Precautions for Sample Collection
Do not draw from an arm with an IV.
Do not draw from a site if tourniquet is left on greater than one minute.
Do not have patient pump fist.
Avoid traumatic puncture – hemolysis.
Do not ice the samples.
Do not delay before closing cartridge – ACT or PT/INR.
Do not expose the sample to AIR.
Do not draw a lactate sample using a tourniquet.
2. Procedures for Patient Testing
a) Turn on the instrument
b) Select #2 – iSTAT cartridge
c) Scan or enter correct operator ID number.
d) Scan or enter the patient ID number (billing/visit #). MIB patients have a blue
wristband
e) Scan the cartridge lot #
f) Remove the cartridge from its pouch. Avoid touching the contact pads or exerting
pressure over the calibrant pack in the center of the cartridge.
g) Direct the dispensing tip, the capillary tube containing the blood or a hanging
drop from a fingerstick into the sample well.
h) Dispense the sample until it reaches the fill mark on the cartridge and the well is
about half full.
![Page 13: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/13.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 13 of 38
i) Close the cover over the sample well until it snaps into place. (Do not put
pressure over the calibrant pack)
j) Insert the cartridge into the cartridge port until it clicks into place. Never attempt
to remove a cartridge while the message “Cartridge Locked” is displayed. Place
the instrument on a stationary surface. Do not move during testing. The analyzer
must remain horizontal during the testing cycle.
k) Enter additional parameters on the Chart Page
Required Field
Sample Type: Choose the number corresponding to the sample type used
when prompted at the Sample Type field. Select CPB for patient in
cardiopulmonary bypass surgery (only available on specific cardiac OR
instruments)
Optional Field
Patient temperature must be entered as degrees C (Celsius only). Only
values between 10 and 45 are interpreted as degrees Celsius. Use the
. (period) key on the i-STAT Portable Analyzer for a decimal point.
Temperature corrected results will be calculated.
FIO2 can be entered as the number of liters or as a percentage of the
oxygen a patient is receiving. Enter this only if analyzing a blood gas
sample.
Fields 1, 2, and 3 are user-defined fields, typically used for ventilator
settings such as PIP or PEEP.
l) If partial results are shown, any parameter(s) resulting with * repeat with a new
cartridge selecting the entire panel.
m) At the completion of the test cycle (audible alarm), view results shown on the
analyzer’s display screen.
n) If Action Range Flag, Scan or Enter Code ____ appears, enter “1” as the comment
code for critical values. Critical values are indicated with or following the
value(s). Print results option will not work until Comment Code is entered.
o) Remove the cartridge after testing. Scroll until Test Options appears on the
bottom of the screen, Enter “1”. Follow prompt for continued testing.
p) Clean the i-STAT analyzer after each patient use. Wipe exterior with Clorox
Germicidal wipes for 1 full minute or 3 minutes for isolation patients. Refer to
Point of Care Device Cleaning and Disinfecting Guide.
3. Alternative Procedure Should the i-STAT System become inoperable for any reason, specimens other than
ACT can be collected and submitted to the laboratory in accordance with the
Laboratory Procedure Manual.
![Page 14: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/14.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 14 of 38
K. REFERENCE RANGE/REPORTABLE RANGE/CRITICAL VALUE
TABLE #6 – REFERENCE AND REPORTABLE RANGES
Test Units Bassett
Reportable Range
Measured Reference Range
Critical Values
Arterial Venous Cord Arterial Cord
Venous Capillary Below Above
Na mmol/L 100-180 135-145 135-145 120 155
K mmol/L 2.0-9.0 3.4-4.5 3.4-4.5 2.8 6.0
Chloride
BUN
Glucose Fasting
Glucose Random
Creatinine***
iCal
Hct
mmol/L
mg/dL
mg/dL
mg/dL
mg/dl
%PCV
65-140
3-140
25-600
0.1-20.0
1.0-10.0
15-68
98-110
70-100
70-139
F 7-17
M 9-20
4.2-5.4
F 34-46
M 35-50
98-110
70-100
70/139
4.2-5.4
F 34-46
M 35-50
40
3.0
21
450
6.5
60
pH 6.5-8.0 7.35-7.48 7.33-7.43 7.20-7.45
CV <7.10
7.20-7.45
CV <7.10
7.35-7.45 7.25 7.60
PCO2 mmHg 20-110 32-48 40-50 38-54 38-54 32-48 23 (art) 60 (art)
PO2 mmHg 10-500 83-108 29-49 10-24 10-24 80-90 45
(art/cap)
Lactate
Troponin
mmol/L
ng/mL
0.3-18.0
0.1-35
0.4-1.3
0.9-1.7
<0.1
>3.9
>0.6
HCO3*
TCO2*
SO2*
BE*/**
mmol/L
mmol/L
%
mmol/L
1-85
5-50
0-100
(-30)-(+30)
21-28
19-32
95-98
(-2)-(3)
22-29
22-34
60-85
<3
15-25
20-34
60-98
(-6.2)-(-0.5)
15-25
20-34
60-98
(-6.2)-(-0.5)
22-27
33-43
90-100
Below**
* = Calculated values
**= Newborn-1 week=(-10)-(-2), 1 week-1 yr=(-7)-(1), 1-16 yr=(-4)-(+2), >16 yr=(-3)-(+3)
***=GFR calculation uses the IDMS-traceable MDRD equation.
GFR (ml/min/1.73m2) = 186 x (PCR)
-1.154 x (Age)
-0.203 x (0.742 if female) x (1.210 if African American)
Test Units Reportable
Range
Reference Range Critical Values
Arterial Venous Capillary Below Above
ACT Sec. 50-1000 Baseline 100-140 Cath Lab Interventional Cases Therapeutic Range=200-300 sec. Cardiac OR Baseline 100-140 sec.
100-140
INR**** <0.9-1.9 Routine Oral-Anticoagulants Mechanical Valves Normal Range
2.0-3.0 2.5-3.5 0.9-1.1
4.4
****Send Confirmatory sample to lab when INR >1.9
Reference Ranges , Reportable Ranges: Reference range means the range of test values
expected from 95% of fasting individuals presumed to be healthy. Reportable range means the
range of test values throughout which the measurement system’s results have been shown to be
valid. The table contains the Reference Ranges (for adults) and Reportable Ranges applicable to
the i-STAT System.
![Page 15: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/15.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 15 of 38
L. RESULTS REPORTING
1. Calculations
The i-STAT analyzer contains a microprocessor that performs all calculations
required for reporting results. BE, HCO3, CO2, O2 SAT and Hgb are calculated.
2. Displayed Results
Results are displayed numerically with their units. Electrolyte, chemistry and
hematocrit results are also depicted as bar graphs with reference ranges marked under
the graphs.
3. Point of Care Testing Critical Value Reporting:
The i-STAT instrument will prompt the operator when there is a critical value. The
critical value comment code must be entered before proceeding to print or download
results. Enter comment code “1”, “Critical Value Policy Followed.” The appropriate
comment will appear in Epic after downloading. i-STAT values are considered
definitive and no confirmations are required, EXCEPT for any INR’s >1.9 require a
venous confirmation.
The Acute Care Nursing areas and Respiratory Therapy have protocols for repeating
and reporting critical values based on patient condition and history. For non-acute
care nursing units, critical values are reported to the provider within 30 minutes.
4. Suppressed Results
There are three conditions under which the i-STAT System will not display results:
a) Results outside the System’s reportable ranges are flagged with a < or >,
indicating that the result is below the lower limit or above the upper limit of the
reportable range, respectively. (See Table 6 Reportable Ranges.) The <> flag
indicates that the results for this test were dependent on the result of a test flagged
as either > or <.
Action:
Send specimen(s) to the laboratory for analysis.
b) Cartridge results which are not reportable based on internal QC rejection criteria
are flagged with ***.
Action:
Analyze the specimen again using a fresh sample and another cartridge. Retest
for the full panel. If the result is suppressed again, send specimen(s), other than
ACT, to the laboratory for analysis in accordance with the Laboratory Procedure
Manual.
c) A Quality Check message will be reported instead of results if the analyzer
detects a problem with the sample, calibrant solution, sensors, or mechanical or
electrical functions of the analyzer during the test cycle.
Action:
![Page 16: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/16.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 16 of 38
Squeeze cup lid to gain access to paper roll
Take the action displayed with the message that identifies the problem. Contact
the Point of Care office at 6728 who will then refer to i-STAT System Manual’s
Troubleshooting section if necessary.
5. Printing and Transmitting Results
a) Printing Results from the i-STAT Analyzer to the Portable Printer:
Align the IR windows of the analyzer and the portable printer within 1 to 6
inches of each other or place analyzer in a Downloader or
Downloader/Recharger if printer attached.
To print the displayed results, press the Print key.
Do not move analyzer or printer until the printout is complete.
Optional: Write the patient’s name on the “Pt Name” line and the physician’s
name on the “Physician” line.
Note: Results printed on thermal paper will fade with time and are therefore not
acceptable as a permanent chartable record.
6. Uploading to RALS: Upload i-STAT non-wireless (gray) results after each test by placing the analyzer in
the table top downloader or downloader/recharger. View the analyzer connecting,
“Communication in Progress” is displayed and return to idle. DO NOT remove i-
STAT downloader during transmission (arrows form a circle). Remove i-STAT from
table top downloader. Wireless units (blue) will automatically upload following
completion of testing. Allow i-STAT to remain in downloader/recharger if the
instrument has a rechargeable battery. Dock analyzer if wireless network is down.
7. Data Review:
a. Last result
Turn meter on
Choose number 1 – Last result
The most recent result performed will display
b. All results
Turn meter on
Press “menu” button
Choose #2 – Data Review
Choose #1 – If Patient ID known
Choose #6 – To list all tests beginning with most recen
M. MAINTENANCE
1. Cleaning:
a. Clean the display screen and the exterior of the i-STAT using the Clorox
Germicidal wipes. Refer to the Point of Care Devices – Cleaning/Disinfecting
Guide for specific instructions and dwell.
2A. Martel Printer – paper replacement a. To replace the paper, open the paper cup
lid by squeezing the lid as shown in the
![Page 17: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/17.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 17 of 38
1.
illustration and remove any remaining
paper by pressing the paper feed button.
Do not pull paper through the printer
mechanism. b. Reel off a few centimeters from a new roll
of paper and check that the end has clean
straight edge.
c. Slide the leading edge of the paper
through the paper entry slot, with the
leading edge of the paper feeding
forwards from the bottom of the roll, until
you feel resistance.
d. Press the paper feed button, and feed the
paper through the printer mechanism.
e. Keep the paper feed button depressed
until enough paper is fed through the
printer mechanism to pass through the
paper exit slot.
f. Sit the new paper roll in the paper cup and
close the lid.
POSITION OF PAPER ROLL IN PRINTER
2B. i-STAT Printer – Paper Replacement a. The STATUS indicator will illuminate to indicate the print status:
Ready: Green
Out of Paper: Orange
Error: Red
b. Paper for the i-STAT Printer can be installed or replaced as follows:
1) Open the paper compartment lid by pulling the release lever as shown in the
printer illustration and remove any remaining paper.
2) Reel off a few centimeters of paper from the new paper roll, with the leading
edge of the paper feeing forward from the bottom of the roll.
3) Sit the new paper roll in the compartment such that the leading edge is resting
outside the compartment on the printer casing.
![Page 18: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/18.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 18 of 38
4) Close the lid until it snaps into place.
Note 1: Should the printer become creased or misaligned, simply reload the paper as
described above ensuring that the paper has a clean, straight edge.
Note 2: When removing a printout from the printer, pull the printout toward the front
of the printer and tear from one side to the other across the serrated edge.
3. Replacing Batteries
a. Battery Compartment is located at the display end of the analyzer.
b. Wait until any test in progress is completed, and turn off the analyzer before
replacing the batteries or the most recent set of results may be lost. Stored
results will not be lost when replacing the batteries.
c. Lithium batteries are available from the warehouse.
d. Slide the battery compartment door off.
e. Tilt the analyzer slightly to slide out the
battery carrier, which contains two 9-volt
lithium batteries.
f. Remove old batteries from the carrier. Pull
each battery out to the side, then lift back and
out.
g. Note the battery orientation symbol molded into the
carrier on each side of the center wall. Starting with
one side, orient the new battery so it matches the
symbol. Slide the battery into the carrier, pushing the
terminal end in first, under the plastic bar, and slide it
up as far as it will go. Then push the bottom of the
battery inward. The terminals of the battery should be
underneath the protective bar on the carrier. Repeat for
the second battery on the other side of the carrier.
![Page 19: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/19.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 19 of 38
h. Note orientation of the battery carrier illustrated on
label on the carrier. The label faces up and the electrical
contact end of the carrier goes into the instrument first.
Insert the carrier into instrument as shown on label. If
carrier is inserted incorrectly, battery door will not
close.
i. Slide the battery compartment door back into place.
N. TROUBLESHOOTING
Overview
When the analyzer is turned on using the On/Off key, the analyzer performs self-checks.
If a condition that should be corrected in the near future, but that will not affect results is
detected, a warning is displayed. The operator presses the 1 key to continue with testing.
Technical Support Analyzer coded messages can be found in the Point of Care i-STAT procedure manual.
Call the Point of Care Office at X6728 for technical assistance.
O. REFERENCES
1. Statland, B.E., Clinical Decision Levels for Lab Tests. Medical Economics Books,
1987.
2. Tietz, N.W., Tietz Textbook of Clinical Chemistry, second edition, Ed. C.A. Burtis,
E.R. Ashwood, W.B. Saunders Company, Philadelphia, 1994. Table 41-20,
Appendix.
3. Kost, Gerald J. Using critical limits to improve patient outcome. Medical Laboratory
Observer. March 1993; 25(3): 22-27.
4. i-STAT System Manual
5. Fundamentals of Clinical Chemistry, N. Tietz, Third Edition, Pages: 426-435, 614-
616, and 676-678.
6. Clinical Chemistry Theory, Analysis & Correlation, Kaplan/Pesce, 2nd
Edition, Pages
850-856, 872-875, 884-888, and 1021-1024.
![Page 20: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/20.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 20 of 38
APPENDIX A
INTEFERING SUBSTANCES
The following substances are known not to significantly interfere with the i-STAT Lactate
assay at the stated test concentrations:
Substance Test Concentration
(mmol/L)
Acetaldehyde 0.04517
Acetaminophen 1.32
Acetylcysteine 10.2
Ascorbate 0.34
Bromide (therapeutic) 2.518-20
Dopamine 0.006
Formaldehyde 0.13317
Β-Hydroxybuterate 6.021
Pyruvate 0.31
Salicylate 4.34
Uric Acid 1.4
The following substances are known to interfere with the i-STAT Creatinine assay:
Substance Test Concentration
(mmol/L) Interference
Acetaminophen 1.32 Increased i-STAT creatinine results.
Acetylcysteine 10.2 Increased i-STAT creatinine results.
Ascorbate 0.34 Increased creatinine results by
approximately 0.3 mg/dL.
Bromide 37.5 Increased i-STAT creatinine results.
Creatine 0.382 Increased i-STAT creatinine results
by approximately 0.2 mg/dL.
Hydroxyurea 0.92 Increased i-STAT creatinine results.
Use another method.
The following substances are known to interfere with the i-STAT Ionized Calcium assay:
Substance Test Concentration
(mmol/L) Interference
Acetaminophen 1.32 Decreased i-STAT ionized calcium
results.
Acetylcysteine 10.2 Decreased i-STAT ionized calcium
results.
Bromide 37.5 Increased i-STAT ionized calcium
results.
Magnesium 1.0
Increased i-STAT ionized calcium
results by approximately 0.04
mmol/L.
Lactate 6.6
Decreased i-STAT ionized Calcium
results by approximately 0.07
mmol/L.
Salicylate 4.34 Decreased i-STAT ionized Calcium
results.
Salicylate (therapeutic) 0.510
Decreased i-STAT ionized Calcium
results by approximately 0.03
mmol/L. .
![Page 21: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/21.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 21 of 38
APPENDIX A – CONTINUED
INTERFERING SUBSTANCES
The following substances are known to interfere with the i-STAT glucose assay:
Substance Test Concentration
(mmol/L) Interference
Acetaminophen 1.32 Increased i-STAT glucose
results.
Acetylcysteine 10.2 Decreased i-STAT glucose
results.
Bromide 37.5 Decreased i-STAT glucose
results.
Bromide (therapeutic) 2.58,9,10
Decreased i-STAT glucose
results by approximately 5
mg/dl.
Hydroxyurea 0.92 Increased i-STAT glucose
results. Use another method.
Thiocyanate 6.9
Decreased i-STAT glucose
results by approximately 7
mg/dl.
The following substances are known to interfere with the i-STAT chloride assay:
Substance Test Concentration
(mmol/L) Interference
Acetylcysteine 10.2 Decreased i-STAT chloride
results.
Bromide 37.5 Increased i-STAT chloride
results. Use another method.
Bromide (therapeutic) 2.58,9.10
Increased i-STAT chloride
results. Use another method.
Salicylate 4.34 Increased i-STAT chloride
results. Use another method.
Thiocyanate 6.9 Increased i-STAT chloride
results. Use another method.
The following substances are known to interfere with i-STAT sodium assay:
Substance Test Concentration
(mmol/L) Interference
Bromide 37.5 Increased i-STAT sodium
results.
![Page 22: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/22.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 22 of 38
![Page 23: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/23.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 23 of 38
![Page 24: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/24.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 24 of 38
![Page 25: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/25.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 25 of 38
![Page 26: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/26.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 26 of 38
![Page 27: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/27.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 27 of 38
![Page 28: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/28.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 28 of 38
![Page 29: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/29.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 29 of 38
![Page 30: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/30.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 30 of 38
![Page 31: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/31.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 31 of 38
![Page 32: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/32.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 32 of 38
![Page 33: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/33.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 33 of 38
![Page 34: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/34.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 34 of 38
![Page 35: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/35.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 35 of 38
![Page 36: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/36.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 36 of 38
![Page 37: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/37.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 37 of 38
![Page 38: Bassett Healthcare Point of Care i-STAT 1 PROCEDUREStore at 2 to 8 C. Do not use after expiration date on the ampules and box. Warm a minimum of 4 hours at room temperature before](https://reader033.vdocuments.mx/reader033/viewer/2022042010/5e71dccf60511221f34df1bc/html5/thumbnails/38.jpg)
Shares3\lab\POCT\PROCEDURES-POC CURRENT\i-STAT Procedure – Rev: 12/12/2016 Page 38 of 38
ER TROPONIN TESTING TIP SHEET
Initial i-STAT Troponin and lab Troponin will be collected on
appropriate chest pain patients in green top tubes and labeled
appropriately – one tube for each.
Scan the patient wristband when prompted for the patient ID.
Mix the sample gently, uncap, fill a syringe or pipette with about 1
ml. of blood, and insert immediately into the cartridge. If not
testing immediately, make sure the sample is well mixed again prior
to filling the cartridge.
After completion of the test, download the instrument, and print
results.
Only ONE initial i-STAT Troponin can be run on each patient.
Results >35 ng/ml should not be reported as the actual value. They
are reported as >35 ng/ml.
If the results error out, the test should not be repeated on the i-
STAT. Wait for the laboratory result.
Once the test is done, send the labeled i-STAT Troponin sample to
the laboratory so it can be saved for the appropriate length of time.
No paperwork is required.
Cartridges are available from the Point of Care office by calling
6728 between 8 – 4:30 Monday through Friday and 3731 off hours.
Store cartridges refrigerated until manufacturer’s outdate or at
room temperature for up to 2 weeks. Label cartridges stored at
room temperature with the room temperature outdate.