Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc. March 1999 Axopatch 200B Patch Clamp Theory and Operation Part Number 2500-121 Rev E, Printed in U.S.A. Copyright 1997-1999 Axon Instruments, Inc. No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical, photocopying, microfilming, recording, or otherwise, without written permission from Axon Instruments, Inc. QUESTIONS? Call +1(510) 675-6200, Fax +1(510) 675-6300

Upload: faggotcatch4395

Post on 26-Dec-2015

365 views

Category:

Documents


11 download

DESCRIPTION

Axopatch 200B manual

TRANSCRIPT

Page 1: Axopatch 200B Manual Final

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

March 1999

Axopatch 200B Patch Clamp

Theory and Operation Part Number 2500-121 Rev E, Printed in U.S.A.

Copyright 1997-1999 Axon Instruments, Inc.

No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical, photocopying, microfilming, recording, or otherwise, without written permission from Axon Instruments, Inc.

QUESTIONS? Call +1(510) 675-6200, Fax +1(510) 675-6300

Page 2: Axopatch 200B Manual Final
Page 3: Axopatch 200B Manual Final

IMPORTANT INFORMATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Disclaimer THIS EQUIPMENT IS NOT INTENDED TO BE USED AND SHOULD NOT BE USED IN HUMAN EXPERIMENTATION OR APPLIED TO HUMANS IN ANY WAY.

Warning IF THIS EQUIPMENT IS USED IN A MANNER NOT SPECIFIED BY THE MANUFACTURER, THE PROTECTION PROVIDED BY THE EQUIPMENT MAY BE IMPAIRED.

Power-Supply Voltage Selection and Fuse Changing

Supply Voltage The Axopatch 200B can be directly connected to all international supply voltages. The input range is from 100 to 240 V~. No range switching is required.

Changing the Fuse The Axopatch 200B uses a 250 V~, T2A, 5 x 20 mm fuse.

In the event of fuse failure, disconnect the power cord.

Before changing the fuse investigate the reason for its failure.

To change the fuse: 1. Disconnect the power cord. 2. Use a screwdriver or a similar device

to rotate the fuse holder counterclockwise.

3. Replace the fuse with another fuse of the same rating.

4. Reconnect the power cord.

Basic Equipment Setup and Safety

1. Supply and Earthing Connections: Use the included IEC

power cord to connect the instrument to a GROUNDED power receptacle.

2. Mounting: Table or rack. 3. Assembly: The headstage connects

to the instrument through the rear panel, 15 pin D-sub connector marked "headstage".

4. Use: Do not operate this equipment with covers or panels removed.

5. Cleaning: Wipe the headstage connector with a damp cloth to clean salt spills. Avoid spilling liquids on the headstage.

The Teflon input connector should be kept very clean. Effective cleaning can be done by spraying with alcohol or swabbing carefully with deionized water. If possible, avoid the use of Freon since it is thought to be detrimental to the environment.

Safe Environmental Conditions 1. Indoor use. 2. Mains supply fluctuations: not to

exceed ±10% of the nominal voltage.

3. Temperature: between 5 ºC and 40 ºC.

4. Altitude: up to 2000 m 5. This instrument is designed to be

used under laboratory conditions. Operate in a clean, dry environment only. Do not operate in a wet or damp environment.

Static Precautions The headstage can normally be safely handled. However, if you are in a laboratory where static is high (i.e., you hear and feel crackles when you touch things), you should touch a grounded metal object immediately before touching the headstage.

You should not switch off power to the Axopatch 200B when handling the headstage input since this will upset the thermal equilibrium.

WARNING Shipping the Axopatch 200B

The Axopatch 200B is a solidly built instrument designed to survive shipping around the world. However, in order to avoid damage during shipping, the Axopatch 200B must be properly packaged.

In general, the best way to package the Axopatch 200B is in the original factory carton. If this is no longer available, we recommend that you carefully wrap the Axopatch 200B in at least three inches (75 mm) of foam or "bubble-pack" sheeting. The wrapped Axopatch 200B should then be placed in a sturdy cardboard carton. Mark the outside of the box with the word FRAGILE and an arrow showing which way is up.

We do not recommend using loose foam pellets to protect the Axopatch 200B. If the carton is dropped by the shipper, there is a good chance that the Axopatch 200B will shift within the loose pellet packaging and be damaged.

If you need to ship your Axopatch 200B to another location, or back to the factory, and you do not have a means to adequately package it, Axon Instruments can ship the proper packaging material to you for a small fee. This may seem like an expense you would like to avoid, but it is inexpensive compared to the cost of repairing an instrument that has sustained shipping damage.

It is your responsibility to package the instrument properly before shipping. If it is not, and it is damaged, the shipper will not honor your claim for compensation.

Page 4: Axopatch 200B Manual Final

RENSEIGNMENTS IMPORTANTS

Axopatch 200B, Copyright March 1997-1999, Axon Instruments, Inc.

LIMITE DE RESPONSABILITE CE MATERIEL N'A PAS ETE CONCU POUR DES EXPERIENCES SUR LES ETRES HUMAINS; ET NE DOIT DONC PAS ETRE UTILISE A CETTE FIN.

ATTENTION L'EMPLOI DE CE MATERIEL D'UNE MANIERE DIFFERENTE A CELLE SPECIFIEE PAR LE FABRICANT AFFECTERA LE NIVEAU DE PROTECTION FOURNIT PAR L'APPAREIL.

Sélection du voltage et changement du fusible

Voltage d'alimentation L'Axopatch 200B peut être directement branché sur toutes alimentations comprises entre 100 et 240 V~. Aucun changement n'est nécessaire afin de sélectioner le voltage de l’appareil.

Changement du fusible L'Axopatch 200B emploie un fusible de 250 V~, T2A, 5 × 20 mm.

En cas de rupture du fusible, débrancher la prise de courant.

Avant de changer le fusible, chercher la raison de la panne.

Pour changer le fusible: 1. Débrancher la prise de courant. 2. A l'aide d'un tournevis ou autre outil

de ce genre, faire tourner le support du fusible dans de sens opposé des aiguilles d'une montre.

3. Remplacer le fusible par un fusible de même valeur.

4. Rebrancher la prise de courant.

Installation du matériel et sécurité 1. Branchement: Employer le fil

electrique IEC fourni pour brancher l'appareil a une prise de courant comprenant UNE TERRE.

2. Pose: Table ou rack. 3. Montage: La tête de l'amplificateur

(“headstage”) est connectée à l'appareil sur le panneau arrière, par l'intermediere d'une prise D-sub à 15

fiches portant l'indication “headstage”.

4. Emploi: Ne pas utiliser ce matériel sans son couvercle et ne pas le couvrir lors de son utilisation.

5. Nettoyage: Essuyer la prise du “headstage” avec un linge humide pour nettoyer les traces de sel. Eviter de renverser des liquides sur le “headstage”.

La prise d'entrée en Téflon doit être maintenue trés propre. Un nettoyage efficace consiste à vaporiser de l'alcool ou á essuyer soigneusement avec de l'eau désionisée. Si possible, éviter l'emploi de Fréon, ce produit étant considéré comme nuisible pour l'environnement.

Conditions à respecter pour un emploi sans danger 1. Emploi à l'intérieur. 2. Fluctuations du réseaux

d'alimentation: ne doivent pas dépasser ±10% de la tension nominale.

3. Température: entre 5 °C et 40 °C. 4. Altitude: jusqu'à 2000 m. 5. Cet appareil a été étudié pour l'emploi

en laboratoire et il doit être situé dans un environnement sec et propre. Ne pas l'utiliser dans un environnement mouillé ou humide.

Précautions statiques Le “headstage” peut être maniée sans danger. Cependant, dans un laboratoire avec un niveau élevé d'electricité statique (c'est-à-dire lorsque vous sentez et voyez des décharges électriques), touchez un objet métallique pour une mise à la terre avant de toucher le “headstage”.

Ne pas d'ébrancher l'Axopatch 200B lors de la manipulation de l'entrée du “headstage”, ceci risque de déranger son équilibre thermique.

ATTENTION Expédition de l'Axopatch 200B

L'Axopatch 200B est un appareil de construction robuste, étudié en vue d'expéditions dans le monde entier. Cependant, l'appareil doit être correctement emballé pour éviter tout domage pendant son transport.

En général, la meilleure façon d'emballer l'Axopatch 200B est de le mettre dans son carton d'origine. Si celui-ci n'est plus disponible, il est recommandé d'envelopper soigneusement l'Axopatch 200B dans au moins trois inches (75 mm) de mousse ou de feuilles d'emballage à bulles. L'Axopatch 200B ainsi protégé devra alors être placé dans un carton solide. Indiquer la mention FRAGILE sur l'extérieur de la boîte ainsi qu'une flèche vers le haut montrant la position verticale.

Il n'est pas recommandé d'employer des boulettes de mousse pour protéger l'Axopatch 200B. En cas de chute de la boîte durant son transport, l'Axopatch 200B pourrait se déplacer à l'intérieur et être endommagé.

Si vous devez expédier l'Axopatch 200B à un autre endroit, ou le renvoyer au fabricant, et si les matériaux d'emballage nécessaires corrects ne sont pas disponibles, ces derniers peuvent être obtenus chez Axon Instruments pour un prix minime. Bien que ceci puisse sembler être une dépense que vous pourriez éviter, elle est cepandant insignifiante en comparaison à celle que coûterait la réparation d'un appareil endommagé pendant le transport.

La responsabilité vous incombe de bien emballer l'appareil avant son expédition. Si ceci n'est pas fait, le transporteur ne pourra pas satisfaire vos réclamation de compensation en cas d'avaries.

Page 5: Axopatch 200B Manual Final

WICHTIGE INFORMATIONEN

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

UNZULÄSSIGE VERWENDUNG DIESER APPARAT IST NICHT VORGESEHEN, BEI MENSCHLICHEN VERSUCHEN VERWENDET ZU WERDEN UND AUCH NICHT AN MENSCHEN IN IRGENDEINERWEISE ANWENDBAR. WARNUNG WEN DIESER APPARAT IN EINER ART UND WEISE ANGEWENDET WIRD, DIE NICHT VOM HERSTELLER SPEZIFISCH ERWÄHNT WIRD, KANN DIE SCHUTZVORRICHTUNG DES APPARATES BEEINTRÄCHTIGT WERDEN. Spannungswahl für die Stromversorgung und Auswechseln der Sicherung

Netzspannung Der Axopatch 200B kann direkt an alle internationalen Netzspannungen angeschlossen werden. Die Eingangsspannung reicht von 100 bis 240 V~. Ein Umschalten des Spannungsbereichs ist nicht erforderlich. Auswechseln der Sicherung Der Axopatch 200B verwendet eine 250V~, T2A, 5 × 20 mm Sicherung.

Im Falle des Ausfalls der Sicherung das Netzkabel ausschalten.

Vor dem Auswechseln der Sicherung den Grund für ihren Ausfall untersuchen.

Schritte zum Auswechseln der Sicherung: 1. Das Netzkabel ausschalten. 2. Die Fassung der Sicherung mit einem

Schraubenzieher oder einem ähnlichen Werkzeug entgegen dem Uhrzeiger drehen.

3. Die Sicherung mit einer anderen Sicherung mit gleicher Nennleistung ersetzen.

4. Das Netzkabel wieder anschließen. Grundlegende Hinweise zu Installation und Sicherheit der Ausrüstung 1. Netz- und Erdungsanschlüsse: Das Instrument mit

dem beigefügten IEC Netzkabel an einen Erdungsschalter anschließen.

2. Anbringung: Tisch oder Rahmengestell.

3. Montage: Der Vorverstärker (“headstage”) wird über einen mit der Aufschrift “headstage” gekennzeichneten 15 Pin D-Unterstecker an der Rückwand des Instrumentes verbunden.

4. Gebrauch: Dieser Apparat darf nicht mit abgenommenen Abdeckungen oder Platten in Betrieb gesetzt werden.

5. Reinigung: Zur Reinigung von verschüttetem Salz den Vorverstärkeranschluß mit einem feuchten Tuch abwischen. Das Verschütten von Flüssigkeiten auf den “headstage” ist zu vermeiden.

Der Teflon-Eingangsstecker sollte in sehr sauberem Zustand gehalten werden. Durch Besprühen mit Alkohol oder vorsichtigem Abtupfen mit entionisiertem Wasser ist eine wirksame Reinigung möglich. Die Benutzung von Freon ist nach Möglichkeit zu vermeiden, da diese Substanz als umweltschädigend angesehen wird.

Umweltsichere Betriebsbedingungen 1. Verwendung in Innenräumen. 2. Netzschwankungen: darf nicht ±10%

der Nennspannung überschreiten. 3. Temperatur: zwischen 5 °C und 40 °C. 4. Höhe: bis zu 2000 m. 5. Dieses Instrument ist für den Gebrauch

unter Laborbedingungen vorgesehen. Nur in sauberer, trockener Umgebung in Betrieb setzen. Nicht in nasser oder feuchter Umgebung in Betrieb setzen.

Schutzmaßnahmen gegen statische Aufladung Der “headstage” kann normalerweise sicher gehandhabt werden. Falls Sie sich jedoch in einem Labor mit höher statischer Aufladung befinden (D.h. Sie hören und fühlen beim Berühren von Objekten ein Knacken), sollten Sie unmittelbar vor dem Berühren der “headstage” ein geerdetes Objekt aus Metall anfassen.

Bei Handhabung des Vorverstärkereingangs sollten Sie die Stromzufuhr zum Axopatch 200B nicht abschalten, um das Temperaturgleichgewicht nicht zu stören.

WARNUNG Versand des Axopatch 200B Bei dem Axopatch 200B handelt es sich um ein solide gebautes Instrument, das beim weltweiten Versand keinen Schaden nehmen sollte. Um jedoch Versandschäden zu verhindern, muß der Axopatch 200B ordnungsgemäß verpackt werden.

Im allgemeinen läßt sich der Axopatch 200B am besten im Originalkarton des Werks verpacken. Ist dieser nicht mehr vorhanden, empfehlen wir, den Axopatch 200B vorsichtig in mindestens 75 mm starkem Schaumstoff oder Bubblepackungen einzuwickeln. Der so eingewickelte Axopatch 200B sollte dann in einen festen Pappkarton gesetzt werden. Die Außenseite des Kartons ist mit dem Worten ZERBRECHLICH (FRAGILE) und einem Pfeil, der auf die Oberseite des Kartons weist, zu kennzeichnen.

Sollte der Karton vom Spediteur fallengelassen werden, besteht eine gute Möglichkeit, daß der Axopatch 200B innerhalt der losen Schaumstoffkugelverpackung bewegt wird und dadurch beschädigt werden kann.

Wenn Sie den Axopatch 200B an einen anderen Ort oder zurück ans Werk senden müssen und Ihnen kein angemessenes Verpackungsmaterial zur Verfügung stehen, kann Axon Instruments Ihnen das geeignete Verpackungsmaterial gegen eine kleine Gebühr zustellen. Sie mögen dies zwar als unnötige Zusatzkosten betrachten, doch ist dieser Aufwand im Vergleich zu den Reparaturkosten fur ein während des Transports beschädigtes Instrument gering.

Sie sind selbst für das richtige Verpacken des Instruments vor dem Versand verantwortlich. Bei einer nicht ordnungsgemäßen Verpackung, die eine Beschädigung zur Folge hat, wird der Spediteur ihren Schadensersatzanspruch nicht anerkennen.

Page 6: Axopatch 200B Manual Final

INFORMACION IMPORTANTE

Axopatch 200B, Copyright March 1997-1999, Axon Instruments, Inc.

LÍMITE DE RESPONSABILIDADES ESTE EQUIPO NO ESTÁ DISEÑADO PARA USO EN HUMANOS Y NO DEBE USARSE PARA EXPERIMENTACIÓN O APLICACIÓN EN SERES HUMANOS BAJO NINGUNA CIRCUNSTANCIA.

ADVERTENCIA SI ESTE EQUIPO SE USA DE MANERA NO ESPECIFICADA POR EL FABRICANTE SE PODRÍA PERDER LA PROTECCIÓN PROVISTA POR EL EQUIPO.

Selección del suministro de corriente y cambio de fusibles

Voltaje de entrada El Axopatch 200B puede conectarse directamente a todos los suministros de energía. El límite de voltaje va entre 100 y 240 V~. No es necesario efectuar cambios en el selector.

Cambio de fusible El Axopatch 200B utiliza un fusible de 250 V~, T2A, 5 × 20 mm. En el caso de que un fusible falle, desconecte el cordón eléctrico. Antes de cambiar el fusible investigue la causa de la falla. Para cambiar el fusible: 1. Desconecte el cordón eléctrico. 2. Use un destornillador o un

dispositivo similar para girar el portafusibles en sentido contrario al de las manecillas del reloj.

3. Reemplace el fusible existente con otro de la misma capacidad.

4. Conecte nuevamente el cordón eléctrico.

Instalación básica y seguridad del equipo 1. Suministro de corriente y conexión a tierra: Use el cordón

eléctrico IEC incluido para conectar el instrumento a una toma de corriente CON CONEXIÓN A TIERRA.

2. Montaje: Sobre una mesa o en un estante.

3. Ensamblaje: El cabezal (“headstage”) se conecta al instrumento en el tablero posterior con el conector de 15 clavijas D-sub, marcado “headstage” .

4. Uso: No utilice este equipo sin las cubiertas o paneles.

5. Limpieza: Limpie el conector del “headstage” con un paño húmedo a fin de quitar los derrames de sales. Evite derramar líquidos sobre el “headstage”.

El conector de entrada fabricado de Teflon debe mantenerse muy limpio. Puede hacerse una limpieza efectiva rociando con alcohol o con un algodón humedecido con agua desionizada. En la medida de lo posible evite el uso del gas freón, puesto que es dañino para el medio ambiente.

Condiciones de seguridad ambiental 1. Para uso interior. 2. Fluctuaciones eléctricas en la fuente

de suministro: no deben exceder ±10% del voltaje nominal.

3. Temperatura: entre 5 °C y 40 °C. 4. Altitud: hasta 2.000 m 5. Este instrumento está diseñado para

ser usado en condiciones de laboratorio. Debe operarse únicamente en un ambiente limpio y seco. No lo use en un ambiente húmedo ni mojado.

Precauciones contra la estática El “headstage” puede manejarse con seguridad, bajo condiciones normales. Sinembargo, si usted se encuentra en un laboratorio donde la estática es alta (por ejemplo, si escucha y percibe chispas cuando toca los objetos), usted debería tocar inmediatamente un objeto metálico que esté en contacto con tierra, antes de tocar el “headstage”.

No apague el interruptor principal del Axopatch 200B cuando manipule la entrada del “headstage” ya que esto afectará el equilibrio térmico.

ADVERTENCIA Envío del Axopatch 200B El Axopatch 200B es un instrumento de construcción sólida, diseñado para soportar el transporte a cualquier parte del mundo. Sinembargo, para evitar los daños que pudieran ocurrir durante su envío, el Axopatch 200B debe empacarse adecuadamente.

En general, la mejor manera de empacar el Axopatch 200B es en la caja original de fábrica. Si ésta ya no se encuentra disponible, le recomendamos que envuelva cuidadosamente el Axopatch 200B en una funda o sábana de espuma o de “empaque de burbujas” con un espesor mínimo de 3 pulgadas (75 mm). El Axopatch 200B, envuelto así, deberá colocarse en una caja de cartón resistente. Marque el exterior de la caja con la palabra FRÁGIL y una flecha que indique la posición hacia arriba.

No recomendamos el uso de bolitas de espuma sueltas para proteger el Axopatch 200B. Si la caja se cae accidentalmente durante el transporte, es muy probable que el Axopatch 200B se desplace dentro del contenedor con las bolitas de espuma sueltas y se dañe.

Si necesita enviar su Axopatch 200B a otra localidad, o de regreso a la fábrica, y no posee el empaque adecuado, Axon Instruments puede enviarle el material necesario por un cargo mínimo. Esto podría parecerle un gasto superfluo que preferiría evitar, pero es económico comparado con lo que costaría la reparación de un instrumento que ha sufrido daños durante el envío.

Es su responsabilidad empacar el instrumento adecuadamente antes de enviarlo. Si no lo hace así y resulta dañado, el transportista no será responsable ni aceptará su reclamo de indemnización.

Page 7: Axopatch 200B Manual Final

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Explanation of symbols Explication des symboles Erklärung der verwendeten symbole Explicación de símbolos Symbol Symbole Symbol Símbolo

Description Description Beschreibung Descripción

Direct current Courant continu Gleichstrom Corriente continua

~

Alternating current Courant alternatif Wechselstrom Corriente alterna

On (Supply) Allumé (alimentation) An (Netz) Encendido (suministro)

o

Off (Supply) Éteint (alimentation) Aus (Netz) Apagado (suministro)

On (Supply) Allumé (alimentation) An (Netz) Encendido (suministro)

Off (Supply) Éteint (alimentation) Aus (Netz) Apagado (suministro)

Protective conductor terminal Borne du conducteur de protection Schutzleiterpol Terminal de conductor protector

Page 8: Axopatch 200B Manual Final
Page 9: Axopatch 200B Manual Final

i

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

COPYRIGHT

THE CIRCUITS AND INFORMATION IN THIS MANUAL ARE COPYRIGHTED AND MUST NOT BE REPRODUCED IN ANY FORM WHATSOEVER WITHOUT WRITTEN PERMISSION FROM AXON INSTRUMENTS, INC.

VERIFICATION

THIS INSTRUMENT IS EXTENSIVELY TESTED AND THOROUGHLY CALIBRATED BEFORE LEAVING THE FACTORY. NEVERTHELESS, RESEARCHERS SHOULD INDEPENDENTLY VERIFY THE BASIC ACCURACY OF THE CONTROLS USING RESISTOR/CAPACITOR MODELS OF THEIR ELECTRODES AND CELL MEMBRANES.

DISCLAIMER

THIS EQUIPMENT IS NOT INTENDED TO BE USED AND SHOULD NOT BE USED IN HUMAN EXPERIMENTATION OR APPLIED TO HUMANS IN ANY WAY.

Page 10: Axopatch 200B Manual Final

ii

Axopatch 200B, Copyright March 1997-1999, Axon Instruments, Inc.

LATE INFORMATION

Please check the back of this manual. If there are pages in yellow, these should be read first. They contain errata and information that became available too late for inclusion in the body of the manual.

Page 11: Axopatch 200B Manual Final

TABLE OF CONTENTS • iii

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

TABLE OF CONTENTS

Page LIST OF FIGURES AND TABLES ....................................................................................vii

CHAPTER 1: INTRODUCTION..........................................................................................1

CHAPTER 2: FUNCTIONAL CHECKOUT........................................................................3

CHAPTER 3: USE OF THE PATCH CLAMP - A TUTORIAL .......................................11 Single Channel Recording (Model Cell) ......................................................................12 Pipette Offset Adjustment ............................................................................................12

Using the TRACK Mode ......................................................................................12 Using the V-CLAMP Mode..................................................................................13

Adjustment of Pipette Capacitance Compensation.......................................................14 Adjustment of Leak Subtraction...................................................................................14 Whole-Cell Recording (Model Cell) ............................................................................15 Pipette Offset Adjustment ............................................................................................16 Whole-Cell Capacitance Compensation .......................................................................16 Series Resistance Compensation ..................................................................................19

Brief Method For Setting Series Resistance Compensation .................................19 Detailed Method For Setting Series Resistance Compensation............................22

Current Clamp (Model Cell).........................................................................................28 Single-Channel Recording (Real Cell) .........................................................................30 Whole-Cell Recording (Real Cell) ...............................................................................35

CHAPTER 4: INTERFACING A COMPUTER TO THE AXOPATCH 200B .................41

Page 12: Axopatch 200B Manual Final

iv • TABLE OF CONTENTS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

CHAPTER 5: LOW NOISE RECORDING TECHNIQUES..............................................43 Glass Type And Coating.......................................................................................44 Seal .......................................................................................................................47 Signal Generator ...................................................................................................48

CHAPTER 6: REFERENCE SECTION: INSTRUMENT OPERATION..........................49 Bridge Mode .................................................................................................................49 Capacitance Measurements...........................................................................................49 Command Potentials.....................................................................................................50

Holding Command................................................................................................50 Seal Test................................................................................................................51

Current Clamp ..............................................................................................................51 Membrane Potential ..............................................................................................51 Whole-Cell Current Clamp...................................................................................51 Whole-Cell Parameters .........................................................................................52 Electrode Capacitance Compensation...................................................................52 Limitations ............................................................................................................53

Electrochemistry ...........................................................................................................54 Headstage......................................................................................................................54

Offset Adjustment For Headstage.........................................................................54 Frequency Boosting ..............................................................................................55 Case Ground Connector........................................................................................55 Mounting the Headstage .......................................................................................55 Cleaning................................................................................................................55 Static Precautions..................................................................................................56 Optical Pick-up .....................................................................................................56 Acoustic Pick-up...................................................................................................56 Tuning the Headstage ...........................................................................................56

Holders..........................................................................................................................56 Features.................................................................................................................56 Parts ......................................................................................................................58 Use ........................................................................................................................59 Glass Dimensions .................................................................................................61

Page 13: Axopatch 200B Manual Final

TABLE OF CONTENTS • v

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Adapters................................................................................................................61 Input BNCs...................................................................................................................61 Output Section ..............................................................................................................63

Filter......................................................................................................................63 Output Gain (α) ....................................................................................................63 Output BNCs ........................................................................................................64

Panel Meter...................................................................................................................67 Power-Supply Voltage Selection and Fuse Changing..................................................68

Supply Voltage .....................................................................................................68 Changing the Fuse ................................................................................................68

Probe Temperature .......................................................................................................68 Zap................................................................................................................................69

CHAPTER 7: REFERENCE SECTION: GENERAL INFORMATION ...........................71 Grounding and Hum.....................................................................................................71 Model Cell ....................................................................................................................72 Model Bilayer...............................................................................................................73 Power-Supply Glitches.................................................................................................74 Ten-Turn Potentiometers..............................................................................................74

CHAPTER 8: REFERENCE SECTION: PRINCIPLES OF OPERATION.......................75 Headstages....................................................................................................................75

Principals of Operation .........................................................................................75 Resistor Feedback .........................................................................................75 Capacitor Feedback.......................................................................................76 If the Probe Output is Slow, How Can Voltage Clamping Be Fast? ............78 What is Clamped During Voltage Clamping? ..............................................79

Capacitance Compensation...........................................................................................79 Pipette Capacitance Compensation...............................................................79 Whole-Cell Capacitance Compensation .......................................................81 Absolute Value .............................................................................................83 Limitations ....................................................................................................83

Series Resistance ..........................................................................................................83

Page 14: Axopatch 200B Manual Final

vi • TABLE OF CONTENTS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Absolute Value..............................................................................................83 Correction, Prediction, and Lag............................................................................83

Saturation Effects..........................................................................................87 Oscillations ...................................................................................................90 Using Lag to Prevent Oscillations ................................................................90 Limitations ....................................................................................................90

Pipette Offset ................................................................................................................91 Leak Subtraction...........................................................................................................92 Zap................................................................................................................................93 Lipid Bilayers ...............................................................................................................93

Experimental Techniques .....................................................................................93 Noise vs. Access Resistance .........................................................................95

Current and Voltage Conventions ................................................................................96 Trouble Shooting ........................................................................................................101

Chapter 9: SPECIFICATIONS..........................................................................................103

Chapter 10: REFERENCES ..............................................................................................119

Chapter 11: HEADSTAGE TUNING PROCEDURE ......................................................121

INDEX................................................................................................................................127

WARRANTY.....................................................................................................................129

WARNING.........................................................................................................................131

CIRCUIT DIAGRAMS REQUEST...................................................................................133

Page 15: Axopatch 200B Manual Final

FIGURES AND TABLES • vii

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

LIST OF FIGURES AND TABLES Page

Figure 1 Connections for testing Axopatch 200B and PATCH-1U model cell................. 6

Figure 2 Pipette capacity compensation .......................................................................... 15

Figure 3 Whole-cell capacity compensation.................................................................... 19

Figure 4 Maximum % PREDICTION as a function of voltage step................................ 21

Figure 5 Series resistance compensation, brief method................................................... 22

Figure 6 Series resistance compensation, detailed method......................................... 26,27

Figure 7 Pipette resistance measurement using SEAL TEST.......................................... 32

Figure 8 Change in resistance while forming a seal ........................................................ 33

Figure 9 Going whole cell: Capacity transients observed when rupturing the patch ..... 37

Figure 10 Going whole cell: Capacity transients observed during Amphotericin partitioning .................................................... 38

Figure 11 Exploded view of the holder ............................................................................. 59

Figure 12 Ag/AgCl pellet assembly................................................................................... 60

Figure 13 PATCH-1U model cell ...................................................................................... 73

Figure 14 MCB-1U bilayer model..................................................................................... 73

Figure 15 Resistive headstage............................................................................................ 75

Figure 16 Capacitive headstage ......................................................................................... 76

Figure 17 Signal handling during resets ............................................................................ 78

Figure 18 Pipette capacitance compensation circuit .......................................................... 80

Page 16: Axopatch 200B Manual Final

viii • FIGURES AND TABLES

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Figure 19 Whole-cell capacitance compensation circuit............................................... 81

Figure 20 Using the injection capacitor to charge the membrane capacitance ............. 82

Figure 21 Integrator driving bilayer model................................................................... 94

Figure 22 Typical current noise in bilayers as a function of series resistance .............. 96

Figure 23 Total current noise as a function of bandwidth ................................... 105,106

Figure 24 Frequency tuning of whole cell configuration............................................ 123

Figure 25 Reset transient compensation ..................................................................... 126

Table I Glass electrical and thermal properties ......................................................... 46

Page 17: Axopatch 200B Manual Final

INTRODUCTION • 1

Chapter 1

Chapter 1

INTRODUCTION

The Axopatch 200B is an improved and enhanced version of the Axopatch 200A amplifier that ushered in a new generation of patch-clamp amplifiers. It offers unsurpassed performance intight-seal patch clamp for single-channel and whole-cell voltage and current clamping. Its noise in single-channel recording is the lowest of any available patch clamp; its series-resistance compensation in whole-cell mode is the most effective of any existing patch clamp, and it includes series resistance compensation in current clamp. Three headstage settings offer an extremely wide range of recording capability, from single-channel to loose patch to whole-cell. In addition, it includes built-in support for capacitance and voltammetric measurements, and a novel slim headstage design that allows easy pipette access to the experimental preparation.

The Axopatch 200B achieves its ultra-low noise performance by cooling components of the innovative capacitive feedback circuitry used for single-channel recording. This technology was developed by Axon Instruments engineers over a period of several years, with extensive design and test contributions from scientific consultants.

The capacitive feedback technique is complicated. The headstage output has to be periodically reset every time the voltage across the capacitor increases towards the limits of the power supply of the instrument. Much of the circuitry of the Axopatch 200B is devoted to eliminating the transients introduced by these resets. In most circumstances, the transient elimination is so good that the resets will not affect the recording and the benefits of the capacitor feedback will be available without penalty.

For whole-cell recording, the noise benefits of the capacitive feedback technique are swamped by the noise sources of the electrode and the membrane capacitance. Since the noise benefits cannot be realized, and at the same time the frequency of resets is high

Page 18: Axopatch 200B Manual Final

2 • INTRODUCTION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

because of the larger average currents, the Axopatch 200B uses the traditional resistive-feedback technique for whole-cell recording.

Patch clamping is a powerful technique that permits the direct observations of the behavior of the small ionic currents that flow through a single ion channel protein. To extract the most from the technique, meticulous attention to detail is required. We have designed the Axopatch 200B with great care and we are confident that it is an excellent tool. But the benefits of this instrument will not be realized if the user is not well versed in its operation and at the same time knowledgeable about the experimental techniques. Therefore, we have written this manual with two goals in mind. The first is to explain the operation of the instrument and many of the principles that underlie its design. The second is to provide the user with a guide to experimental techniques that our customers and scientific consultants have found useful during many years of experience in patch clamping.

We hope that you will find this manual to be a useful laboratory companion. We aim to revise it periodically and we look forward to receiving any suggestions that you might have for its improvement.

NOTE The CV 203BU Headstage will feel warm to the touch during normal use (with headstage cooling ON). This is due to the operation of the Peltier device that cools the critical headstage circuitry to achieve the extremely low noise levels possible with the Axopatch 200B.

Page 19: Axopatch 200B Manual Final

FUNCTIONAL CHECKOUT • 3

Chapter 2

Chapter 2

FUNCTIONAL CHECKOUT

When you receive your Axopatch 200B, you should first run a functional checkout to ensure the proper functioning of the instrument. All units are burned-in and thoroughly tested at the factory before shipping. If you observe any damage caused by shipping or if you encounter problems with the functional checkout, please call the factory.

Startup Procedure For the initial checkout, the Axopatch 200B should be situated on a benchtop away from other equipment. Do not install in a rack until the checkout is complete. Make sure that the power is OFF. An oscilloscope is the only other piece of equipment required for these tests. A large sheet of aluminum foil is needed.

1) The only connections to the Axopatch 200B should be: a) the power cable, b) the headstage, c) a cable from the SCALED OUTPUT BNC to one channel of the oscilloscope. Take care to prevent static discharge near the headstage input connector. Turn the power on.

Page 20: Axopatch 200B Manual Final

4 • FUNCTIONAL CHECKOUT

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

2) Set the front panel controls of the Axopatch 200B as follows:

PIPETTE OFFSET: About 5.0 ZAP 0.5 ms PIPETTE CAPACITANCE COMP.: Minimum

(fully counterclockwise) SERIES RESISTANCE COMP. % PREDICTION: 0 %, OFF SERIES RESISTANCE COMP. % CORRECTION: 0 %, OFF SERIES RESISTANCE COMP. LAG: 1 µs WHOLE CELL CAP.: 0 pF, OFF SERIES RESISTANCE: 0 MΩ HOLDING COMMAND: 0 mV, x1, OFF SEAL TEST: OFF METER: Set switch to I MODE: V-CLAMP CONFIG.: WHOLE CELL (β = 1) OUTPUT GAIN: α = 10 LOWPASS BESSEL FILTER: 5 kHz LEAK SUBTRACTION: ∞ MΩ, OFF

Test the noise:

3) Shield the headstage with a large sheet of aluminum foil. Wrap the foil loosely and completely around the headstage. Leave room at the headstage input for the model cell in the next test. Connect the foil shield to the ground input of the headstage (gold-plated 1 mm socket at the rear of the probe) using a clip-lead. The easiest way to do this is to connect the clip-lead directly to the foil and to the 1 mm pin inserted into the headstage ground socket. (Ground is available from the gold-plated 1 mm socket at the rear of the probe, from the yellow 4 mm socket on the rear-panel of the main unit, or from the BNC shields.)

Page 21: Axopatch 200B Manual Final

FUNCTIONAL CHECKOUT • 5

Chapter 2

4) Turn the meter switch to IRMS. Note the reading on the rms noise display for the three headstage configurations (toggle the configuration between the PATCH and WHOLE CELL positions). The expected values under optimal conditions are:

CV 203BU PATCH β = 1 ≈ .045 pA rms WHOLE CELL β = 1 ≈ 0.55 pA rms

WHOLE CELL β = 0.1 ≈ 1.60 pA rms

IRMS is shown on the panel meter always in a 5 kHz bandwidth using a Butterworth filter that is independent of the front-panel lowpass Bessel filter. The front-panel gain does not affect this reading.

If the observed values are more than twice the expected values or if the meter is blanked due to exceeding its range, then check to see that the foil shield is correctly grounded and that all controls are in the positions noted in 2 above. Also check on the screen of the oscilloscope for 60 Hz interference or other noise pickup.

Test the PIPETTE CAPACITANCE COMPENSATION and FILTERS:

5) Set CONFIG. to PATCH. Set the oscilloscope gain to 0.5 V/div, trigger to line, and the sweep speed to 2 ms/div. Connect the oscilloscope to the Scaled Output with a BNC cable. Turn on the SEAL TEST switch. You should see oppositely going capacitance transients at about 8 ms intervals (about 10 ms for 50 Hz line frequency). Turn the FAST MAG and FAST τ controls and verify that the capacitance transients change their size and amplitude. Do the same with the SLOW MAG and SLOW τ controls. The change is subtle. You may need to change settings on the oscilloscope to see an effect. Switch from PATCH to WHOLE CELL (β = 1) and repeat the above procedures. Switch between the 1, 2, 5, 10, and 100 kHz positions on the LOWPASS BESSEL FILTER control and verify that the capacitance transient signals seen on the oscilloscope screen are affected.

Turn off SEAL TEST.

Page 22: Axopatch 200B Manual Final

6 • FUNCTIONAL CHECKOUT

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Test WHOLE CELL and PATCH configurations:

Figure 1. Connections for testing Axopatch 200B and PATCH-1U model cell.

6) Remove the foil shield temporarily. Connect the black lead (2 mm pin at one end) to the PATCH-1U model cell ground (2 mm socket at central position) then connect this lead to the ground input of the headstage (1 mm pin).

7) Connect the model cell BATH connection to the headstage input. Insert the Teflon collet of the model cell into the collar of the headstage (the fit will be snug). Make sure that the collet is inserted all the way so that the connector pin is fully inserted into the headstage input jack.

8) Shield the headstage and model cell with the foil. Reconnect the foil shield to ground.

9) Place the CONFIG. switch in the WHOLE CELL (β = 1) position. Turn the panel meter switch to the I position. Zero the panel meter using the PIPETTE OFFSET knob. If the meter cannot be easily zeroed within ±10 pA or if it constantly drifts, then the foil shield is probably not correctly connected to ground.

10) Turn the panel meter switch to VHOLD/IHOLD and set the HOLDING COMMAND switch to "+" and the toggle to "x1". Turn the HOLDING COMMAND potentiometer until the meter reads 10 mV. Turn the panel meter switch back to I.

Page 23: Axopatch 200B Manual Final

FUNCTIONAL CHECKOUT • 7

Chapter 2

This clamps 10 mV across the 10 MΩ resistor in the model cell BATH position. The panel meter should read the correct current (1 nA) within error limits (2% for the meter, 1% for the resistor and 1% for the HOLDING COMMAND generator plus errors for zeroing, etc.).

11) Turn the HOLDING COMMAND switch OFF. Place the CONFIG. switch in the PATCH position. Zero the panel meter using the PIPETTE OFFSET knob.

Switch the HOLDING COMMAND to "+" with the toggle to "x1". This again clamps 10 mV across the 10 MΩ resistor in the model cell. The panel meter should read 1 nA.

At this point the INTEGRATOR RESET LED should be illuminated continuously because of the large number of resets needed to pass a continuous current of 1 nA.

Test I-CLAMP and TRACK (I=0):

12) Set mode switch to TRACK and turn the panel meter to VTRACK. The meter should read 0 mV. Turn the PIPETTE OFFSET control one full turn clockwise. Meter should read +50 mV. This is the voltage applied by the TRACK circuit to keep the current at zero.

13) Turn the PIPETTE OFFSET control counterclockwise until the meter again reads 0 mV.

14) Switch mode to I-CLAMP NORMAL. Switch meter to VHOLD/IHOLD. Rotate HOLDING COMMAND control until meter reads 1.0 nA. Turn meter switch to Vm. Meter should read +10 mV. Confirm that you get the same reading when I-CLAMP FAST mode is chosen. You may need to adjust the FAST PIPETTE CAPACITANCE COMPENSATION to remove oscillations.

Test SERIES RESISTANCE and WHOLE-CELL CAPACITANCE COMPENSATION:

15) Turn HOLDING COMMAND OFF. Then set CONFIG. to WHOLE CELL (β = 1) and mode to V-CLAMP.

16) Again remove the foil from the headstage and unplug the model cell. Now connect the model cell in the CELL position to the headstage input connector and reattach the foil making sure it is grounded as before. Turn on the SEAL TEST switch. Again you should see capacitance transients on the oscilloscope screen but now they should be

Page 24: Axopatch 200B Manual Final

8 • FUNCTIONAL CHECKOUT

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

larger and longer lasting. Turn on the WHOLE CELL CAP. switch. Using the SERIES RESISTANCE and WHOLE CELL CAP. controls, observe a change in shape and size of the capacitance transients. The capacity transient must be nulled to read correct values of the membrane capacitance (Cm) and the series resistance (Rs) from these front panel controls.

17) Without completely nullifying the transient, turn on the % PREDICTION control and rotate it clockwise. Observe the capacitance transient getting larger and faster.

18) Now switch off the WHOLE CELL CAP. and turn the SERIES RESISTANCE control to about 10 MΩ. Set the LAG control to its maximum value and then turn on the % CORRECTION control. Rotate this to the right and again observe that the capacitance transients get larger and faster. Set the control to 100%. Rotate the LAG control counterclockwise and verify that the circuit will oscillate at lower LAG values.

Test CAPACITANCE DITHER:

19) Set up as in 15 and 16 above. Connect square-wave voltage source (switching from approximately 0 V to +5 V at a frequency of between 0.5 to 5 Hz) to CAP DITHER CONTROL BNC connection on back panel. Set lowpass filter setting to 100 kHz. View transients on oscilloscope screen. Note small changes in compensation as dither unit is activated by square pulses from square wave voltage source. These changes will be very small and may require sensitive adjustment of gain and triggering on the oscilloscope.

Test ZAP:

20) Turn off SEAL TEST, PREDICTION & CORRECTION. Set the ZAP control to MANUAL and set the METER to I. While the ZAP button is depressed you should see a reading of approximately +2.6 nA.

21) Set the ZAP control to 0.5 ms. Depress the ZAP button several times. The OVLD light in the SCALED OUTPUT section should flash each time the button is pushed.

Page 25: Axopatch 200B Manual Final

FUNCTIONAL CHECKOUT • 9

Chapter 2

Test TEMPERATURE CONTROL:

22) Set HEADSTAGE COOLING switch to OFF (switch is located on back panel). Set Meter Switch to TEMP. Verify that Probe Temp light is off and that temperature reading on Panel Meter is near 20 °C (note that reading is not strictly accurate at room temperature values). Next, set HEADSTAGE COOLING switch to ON. Verify that Probe Temp light comes on when temperature reading falls below 0 °C.

You have now verified that all the circuits are in working order.

Page 26: Axopatch 200B Manual Final
Page 27: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 11

Chapter 3

Chapter 3

Use of the Patch Clamp — A Tutorial

The purpose of this chapter is to ease you into the use of your Axopatch 200B. The controls have been carefully grouped for clarity. Many of them can be switched off and ignored until you become more familiar with the instrument and patch clamping.

Operation of the Axopatch 200B will be described initially in the context of a real experiment while using the PATCH-1U model cell supplied with the unit and illustrated in the Model Cell portion of the REFERENCE SECTION: GENERAL INFORMATION. An oscilloscope, a step generator, and some aluminum foil will be required.

Note that your headstage contains three feedback elements, a capacitor in the PATCH position and resistors in the WHOLE CELL β = 0.1 and β = 1 positions. On each setting, the headstage gain is β mV/pA and this gain is multiplied by the setting on the rotary gain switch, α, to achieve the total output gain

The PATCH-1U model cell emulates three experimental configurations:

BATH: 10 MΩ "electrode" resistor to ground. PATCH: 10 GΩ "patch" resistor to ground. Approximately 5 pF stray capacitance to ground. CELL: 10 MΩ "electrode" resistor. 500 MΩ "cell membrane" resistor in parallel with 33 pF "cell membrane" capacitor. Approximately 5 pF stray capacitance to ground.

Page 28: Axopatch 200B Manual Final

12 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Single Channel Recording (Model Cell) Set the front panel controls of the Axopatch 200B as follows:

PIPETTE OFFSET: About 5.0 PIPETTE CAPACITANCE COMP. Minimum (fully counterclockwise) ZAP: 0.5 ms SERIES RESISTANCE COMP. % PREDICTION: 0 %, OFF SERIES RESISTANCE COMP. % CORRECTION: 0 %, OFF SERIES RESISTANCE COMP. LAG: 1 µs WHOLE CELL CAPACITANCE: 0 pF, OFF SERIES RESISTANCE: 0 MΩ HOLDING COMMAND: 0 mV, x1, OFF SEAL TEST: ON METER: Set switch to VTRACK MODE: TRACK CONFIG.: PATCH OUTPUT GAIN: α = 10 LOWPASS BESSEL FILTER: 5 kHz LEAK SUBTRACTION: ∞ MΩ, OFF

Pipette Offset Adjustment Install the PATCH-1U model cell in BATH position into your headstage connector as described in the FUNCTIONAL CHECKOUT section. Again, surround it with aluminum foil grounded to the headstage ground connector. Connect the SCALED OUTPUT BNC to one channel of your oscilloscope. Set the oscilloscope gain to 1 V/div and set its sweep speed to 2 ms/div. Select line synchronization.

Using the TRACK Mode You will see on the oscilloscope a rectangular, somewhat drooping pulse lasting about 8 ms (10 ms for 50 Hz line frequency). This is the current driven through the 10 MΩ "electrode"

Page 29: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 13

Chapter 3

in the model cell by the 5 mV SEAL TEST command. The total height of the pulse is about 5 volts. (Note: With a 10 MΩ resistance a 5 mV step will generate 0.5 V; the x10 gain provides a 10-fold amplification resulting in a 5 V output). While the SEAL TEST circuit puts out only a +5 mV signal, the current signal you see will be 2.5 volts above zero and 2.5 volts below. This is because the TRACK circuit keeps the total current at zero. It achieves this by supplying the appropriate command that, when summed with the 5 mV command, keeps the integral of the current at zero. What you see on the oscilloscope is comparable to what you would see immediately following the immersion of the pipette tip in the bathing solution surrounding your cells.

Now turn off the SEAL TEST and look at the value on the meter. Adjust the PIPETTE OFFSET control until the TRACK voltage is zeroed. At this point, the tracking circuit does not have to put out a voltage in addition to the pipette offset voltage to achieve zero current. This specifies the proper setting of the PIPETTE OFFSET control to zero junction potentials and electrode asymmetries.

Using the V-CLAMP Mode Some investigators prefer not to use a tracking circuit but do their offset adjustments and sealing in V-CLAMP. The choice is a matter of personal preference. Both methods enable you to follow drifts in the electrode. In TRACK mode, drifts are observed as changing voltages on the panel meter when VTRACK is selected. In V-CLAMP mode, drifts are observed as changing currents. To test this approach, change the OUTPUT GAIN to 5, switch to V-CLAMP mode, turn the meter switch to I, and again turn on the SEAL TEST. Now you will see a 2.5 V, about 8 ms (10 ms for 50 Hz line frequency) pulse that goes in the positive direction. Change the position of the PIPETTE OFFSET control and notice that the DC position of the rectangular pulse is altered as you do so. You could achieve rough zeroing of pipette offsets by simply adjusting the control until the rectangular pulse starts from zero on the oscilloscope screen. For more accurate adjustment, turn off the SEAL TEST and adjust the PIPETTE OFFSET control until the meter reads zero.

Page 30: Axopatch 200B Manual Final

14 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Adjustment of Pipette Capacitance Compensation Now connect the model cell in the PATCH position to the headstage connector and surround with grounded aluminum foil. With the panel meter set to IRMS, the reading should be less than 0.200 pA rms. Return the meter setting to I.

Turn the gain switch (α) to 10 and turn on SEAL TEST. On the oscilloscope, you will see two capacitance transients, one near the beginning of the sweep and the other near the end. They will be of opposite polarity. This is comparable to what you would see immediately following a gigohm seal between your pipette and cell. Using the FAST MAG control, reduce the size of the capacitance transient as far as possible. Switch the gain to 500. Now use both the FAST MAG and FAST τ controls to minimize the transient. This is done iteratively. Turn the FAST MAG slightly and then readjust the FAST τ for minimum transient. Do this over and over again until you find the setting where the transient is finally minimized (Figure 2). With some practice, you will be able to use the two controls simultaneously to compensate stray capacitance rapidly. With real electrodes, but probably not with your model cell, the capacitance transient will have more than one component. SLOW MAG and SLOW τ controls are provided for minimizing a second electrode component in real experiments.

Adjustment of Leak Subtraction With the model cell still in the PATCH position, turn off the SEAL TEST. The current should still read zero on the meter. Turn the HOLDING COMMAND switch to "+" (HOLDING COMMAND toggle set to "x1") and use the HOLDING COMMAND control to move the current substantially off zero. While looking at the oscilloscope screen, turn the LEAK SUBTRACTION control until the current is returned to zero. This circuit sums a scaled version of the command signal with the current. The scaling is determined by the LEAK SUBTRACTION control. Now change the HOLDING COMMAND switch from "+" to "-". If the LEAK SUBTRACTION circuit is properly adjusted, the current trace will stay on zero as the HOLDING COMMAND is switched back and forth between "+" and "-".

Page 31: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 15

Chapter 3

CONDITIONS: PATCH CONFIG. SEAL TEST ON 5kHz FILTER

Figure 2. Electrode (pipette) capacitance compensation.

Whole-Cell Recording (Model Cell) Set the front panel controls of the Axopatch 200B as follows:

PIPETTE OFFSET: About 5.0 ZAP: 0.5 ms SERIES RESISTANCE COMP. % PREDICTION: 0 %, OFF SERIES RESISTANCE COMP. % CORRECTION: 0 %, OFF SERIES RESISTANCE COMP. LAG: 1 µs WHOLE CELL CAP.: 0 pF, OFF SERIES RESISTANCE: 0 MΩ HOLDING COMMAND: 0 mV, toggle x1, OFF SEAL TEST: OFF METER: Set switch to VTRACK MODE: TRACK CONFIG.: WHOLE CELL (β = 1) OUTPUT GAIN: α = 1 LOWPASS BESSEL FILTER: 5 kHz LEAK SUBTRACTION: ∞ MΩ, OFF

↑ ↑ PULSE PULSE ON OFF

1 pA

1 ms

Page 32: Axopatch 200B Manual Final

16 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Pipette Offset Adjustment These are done exactly as described in the Single Channel Recording section in the TUTORIAL. By the end of the pipette offset adjustment procedure a number of front panel controls will be changed from the above list. These include: MODE will be switched from TRACK to V-CLAMP, SEAL TEST will be ON, and OUTPUT GAIN will change from 1 to 5.

Whole-Cell Capacitance Compensation There are a variety of approaches for adjusting the parameters to cancel whole-cell capacity transients. The approach we suggest here works well and is convenient for rapid and complete cancellation of whole-cell transients.

1) Select the whole-cell (resistive feedback) mode. (Set CONFIG. to WHOLE CELL (β = 1); set MODE to V-CLAMP; set METER to I.)

2) Attach the model cell to the headstage input in the CELL position. Insure that 60 Hz interference is sufficiently small. If necessary, this can be achieved by shielding with aluminum foil. If necessary, adjust PIPETTE OFFSET knob to zero the METER reading.

3) For a Config. setting of W. Cell β = 1 (500 MΩ feedback resistor), a gain setting of 0.5 is convenient; for a Config. setting of W. Cell β = 0.1 (50 MΩ feedback resistor) set the gain at 5.

4) Begin with the PIPETTE CAPACITANCE COMPENSATION controls fully off (counter-clockwise). Both WHOLE-CELL PARAMETERS should be off, i.e., WHOLE CELL CAP. and SERIES RESISTANCE potentiometers should be fully turned counterclockwise, and the WHOLE CELL CAP. switch should be OFF. PREDICTION and CORRECTION should also be OFF.

5) Be sure that the HOLDING COMMAND switch is off and turn on the EXT. COMMAND switch. Apply a square wave with a frequency of about 50 Hz and an amplitude of about 2 V to the EXT. COMMAND input. (Alternatively, if a pulse generator is not conveniently available you can use the SEAL TEST.) Trigger the scope from the external source. This will produce large transients followed by steps in

Page 33: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 17

Chapter 3

the command potential of 40 mV amplitude and about 10 ms duration. If you are using the SEAL TEST it will generate a 5 mV command instead of a 40 mV command. You should scale the amplitude values in the following paragraphs accordingly.

6) Initially select a 10 kHz or 100 kHz bandwidth since this will allow the fast capacity transient to be more readily distinguished from the slower whole-cell transient (Figure 3a). At this band-width use the FAST MAG and FAST τ of the PIPETTE CAPACITANCE COMPENSATION to eliminate the fast capacity transient as previously described. The fast transient can be distinguished from the slower whole-cell transient by its more rapid time course. At a sweep speed of 50 or 100 µs per division and a vertical sensitivity of 0.5 or 1.0 V per division attempt to make the leading edge of the whole-cell transient look similar to Figure 3b.

7) Reduce the bandwidth to 5 kHz. At a slow sweep speed (e.g., 5 ms/division) and high vertical resolution (e.g., 20 mV/division) adjust the LEAK SUBTRACTION potentiometer (see Adjustment of Leak Subtraction above) to eliminate leak current from the 500 MΩ resistor simulating the membrane resistance in the model cell as in Figure 3c.

8) Turn the WHOLE CELL CAP. switch on. Reduce the vertical resolution to 1 V/div and increase the sweep speed to 1 ms/div. Using the WHOLE CELL CAP. and SERIES RESISTANCE controls simultaneously, minimize the capacity transients as in Figure 3d. There may be a small wiggle at the leading edge of the capacity transient due to minor misadjustment of the fast electrode compensation controls which can be removed by readjustment of those controls. Final adjustment should be done with a vertical resolution of about 50 mV/div. (Note: Be sure that the trace is flat at times beyond about 1 ms after its start. This is best accomplished with a sweep speed of 1 or 2 ms/div.) With a little practice it should be simple to reduce the capacity transient into the noise. The values of the series resistance and whole-cell capacitance determined in this way will be quite close to the correct values.

Page 34: Axopatch 200B Manual Final
Page 35: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 19

Chapter 3

Series Resistance Compensation The Axopatch 200B is capable of series-resistance compensation equal to that of the Axopatch 200A and substantially better than has been possible with any other commercial patch clamp to date. In order to achieve the outstanding performance of the Axopatch 200B it is critical to set its controls properly. Because of the importance of this issue we go to great lengths here to describe in detail the methods for properly setting the instrument. We present both a brief method for experienced users and a more detailed method for those not yet skilled in patch clamping. (See REFERENCE SECTION: PRINCIPLES OF OPERATION for a discussion of the theory of whole-cell and series-resistance compensation).

Brief Method For Setting Series Resistance Compensation At this point you have eliminated all capacitive current from the measured output. However, the cell membrane response to a step voltage command will still proceed with the time constant RsCm of about 330 µs for the model cell (where Rs is the series resistance and Cm is the membrane capacitance). The Axopatch 200B uses dual controls to speed this response and to compensate for IR drops resulting from membrane current and series resistance, and for the filtering effect of the membrane capacitance and series resistance. These controls are labeled PREDICTION and CORRECTION. Non-ideal circuit characteristics require minor readjustments of WHOLE CELL CAP., SERIES RESISTANCE, and PIPETTE CAPACITANCE COMPENSATION potentiometers to achieve the best performance. The procedure here is intended to familiarize the user with these controls and with the necessary readjustments.

Users who are already skilled at adjusting both the fast electrode capacitance and whole-cell capacitance compensation circuits may prefer to set both PREDICTION and CORRECTION at the same time prior to readjusting the various controls to eliminate the final transient. Be aware that if you exceed the maximum achievable % PREDICTION you will produce unacceptable non-linearities (see Figure 4 and detailed explanation in the Series Resistance section in PRINCIPLES OF OPERATION).

Page 36: Axopatch 200B Manual Final

20 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

MAXIMUM ACHIEVABLE% PREDICTION (%)

VSTEP(mV)

Figure 4. Maximum % PREDICTION as a functioning voltage step.

After adjusting the whole-cell capacitance as described above (Figure 5a; same as Figure 3d), advance the PREDICTION potentiometer setting to 95 or 96%. Without any other adjustments, set the LAG at 10 µs and advance the CORRECTION potentiometer setting to about 95%. A rather large transient should appear in the current at the beginning and end of the command step. Its peak-to-peak amplitude should be 2-4 nA and it should undergo several distinct "rings" requiring 2-4 ms to disappear into the noise (Figure 5b). To completely eliminate this transient, begin by reducing the setting of the SERIES RESISTANCE control by about 2-3%. As you reduce this setting, the amplitude of the transient first decreases and then begins to increase. A distinct minimum exists and the desired setting of the SERIES RESISTANCE control is at this minimum (Figure 5c).

Next, slightly decrease the FAST MAG setting. The fast leading edge transient should decrease along with the peak-to-peak magnitude of the overall transient. Stop when the leading edge transient disappears as in Figure 5d.

If all controls were set correctly from Figure 3d (or 5a), you should be able to completely obliterate the remaining transient with a small increase in the WHOLE-CELL CAP. setting. If the transient cannot be completely eliminated in this way, minor readjustments of the SERIES RESISTANCE control may be required, followed by further readjustments of FAST MAG and FAST τ. An iterative procedure works best. It should be possible to reach this high degree of compensation without any residual transient as shown in Figure 5e.

Page 37: Axopatch 200B Manual Final
Page 38: Axopatch 200B Manual Final

22 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Detailed Method For Setting Series Resistance Compensation New patch clamp users who are not yet skilled in setting all of the capacity compensation controls may find it impossible to achieve the exceptional series resistance compensation shown above. We, therefore, present here a detailed description of a method for setting the controls that ensures outstanding series resistance compensation. In this approach, PREDICTION and CORRECTION are set sequentially rather than concurrently as in the above section. Minor readjustments are then done to achieve the final result. We also provide the user with a detailed explanation of the series-resistance circuit operation.

If you are satisfied with your ability to set all of the capacity compensation controls, you can skip the following indented sections and resume at the Current Clamp (Model Cell) section.

Note that PREDICTION is an open loop process, i.e., it does not involve feedback, and instability is only possible if the internal circuitry that develops the prediction signals is pushed too far. Generally, the circuit is stable up to values of about 98%, but it can become non-linear, depending on the magnitude of VSTEP (Figure 4). If the PREDICTION potentiometer has been advanced too far it may not be noticeable in 5 kHz bandwidth until the current begins to oscillate. However, in 100 kHz bandwidth you will observe ringing developing in the transient as the PREDICTION percentage becomes too large. For best results you should start with whole-cell transients canceled as in Figure 6a (same as Figure 3d). To begin, advance the PREDICTION potentiometer gradually up to a final value of 95 or 96%. A brief transient will emerge at the leading and trailing edge of the command potential step. The amplitude of this transient is typically about 500 pAp-p

1 (in 5 kHz bandwidth) for a 40 mV step using the cell model (Figure 6b). The residual transient at this stage is typically biphasic, with a total duration of about 200 µs in a bandwidth of 5 kHz, but it can be eliminated.

To eliminate the residual transient that has resulted from the use of PREDICTION, small readjustments in the settings of the PIPETTE CAPACITANCE COMPENSATION FAST MAG control, the WHOLE CELL CAP., and SERIES RESISTANCE settings are needed. To begin, slightly reduce the setting of the FAST MAG control in PIPETTE CAPACITANCE COMPENSATION. The amplitude of the initial negative-going component of the residual transient will decrease. Continue to change the control setting until the waveform of the residual transient changes to a monophasic positive going response. At the point where the

1 pAp-p - pA peak-to-peak

Page 39: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 23

Chapter 3

negative component has just been eliminated, you may observe a small wiggle at the leading edge of the residual transient. Continue to reduce the FAST MAG control until the leading edge of the transient is smooth and "S-shaped". This may lead to a small (5-10%) increase in the peak amplitude of the residual (positive-going) transient. Note that further reductions in the FAST MAG control will cause the residual transient to grow further in amplitude; stop when the leading edge is smooth. This is best accomplished at a sweep speed of 100-200 µs/div and a vertical sensitivity of about 100 mV/div, assuming a 40 mV command step (Figure 6c).

Reduce the value of the SERIES RESISTANCE control setting by about 2-3%. A notch should develop in the residual waveform and a slower positive-going component should develop at the trailing edge of the transient. Stop when the overall transient is centered around the baseline. A typical waveform at this point is shown in Figure 6d. If the initial direction of the waveform is negative, the FAST MAG may need to be decreased a little more.

Slightly increase the setting of the WHOLE CELL CAP. potentiometer (1-2%) until the slow component of the residual transient disappears. The resulting waveform should be biphasic with positive-going initial component and a negative-going final component. In 5 kHz bandwidth its peak-to-peak amplitude should be about 100-200 pA and its total duration about 200 µs (Figure 6e).

Adjust the FAST τ of the PIPETTE CAPACITANCE COMPENSATION until the residual transient seems to be minimized. Note that it will still be biphasic; keep the waveform as smooth as possible. The readjustment of FAST τ is usually not very large. Now readjust the FAST MAG control of PIPETTE CAPACITANCE COMPENSATION and minimize the final transient. A few iterative adjustments of the FAST MAG and FAST τ may be required to achieve best performance. If a reasonably fast (< 200 µs) component remains that can not be compensated with the FAST MAG and FAST τ controls, it probably means that a very small readjustment of the SERIES RESISTANCE potentiometer is needed. At a slower sweep speed (e.g., 1-2 ms/div) make sure that there is no slow component visible; if you observe a slow component you need to slightly readjust the WHOLE CELL CAP. potentiometer. With a little practice it should be possible to reduce the residual transient to an amplitude of about 10 pA so that it essentially disappears into the noise (Figure 6f ). You are now clamping the model cell membrane in about 30 µs and have completely eliminated all the capacity current from the output of the headstage. Similar performance can normally be achieved with real cells but the use of the slow component of the PIPETTE CAPACITANCE COMPENSATION is likely to also be required. Set the vertical sensitivity to 0.5 V/div and the sweep speed to 1-2 ms/div, and turn OFF and ON the WHOLE CELL CAP. switch (turning this switch OFF also disables PREDICTION but not CORRECTION). This is a striking demonstration of the improvement in

Page 40: Axopatch 200B Manual Final

24 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

performance that has been achieved by the adjustment procedure. Also note that in 5 kHz bandwidth turning ON and OFF this switch does not noticeably change the noise riding on the waveform.

Despite the fact that the membrane potential is now being changed extremely rapidly (about 30 µs) in response to a step command of potential, two important effects of series resistance (Rs) have NOT yet been eliminated. These are IR drops resulting from the flow of ionic currents (I) in the cell membrane and the filtering effect of the series resistance and cell membrane capacitance on measured membrane currents. An ionic current of 2 nA flowing across a series resistance of 10 MΩ will produce a 20 mV error in the true membrane potential relative to the command potential; this effect of series resistance is well known. It is far less well understood that series resistance in conjunction with membrane capacitance will have the effect of filtering the measured current with a one pole RC filter with a corner frequency given by 1/2πRsCm. For Rs= 20 MΩ and Cm= 50 pF, this is only about 160 Hz.

The CORRECTION potentiometer is used to greatly reduce both of these errors. To set CORRECTION properly, first set the LAG potentiometer at about 10 µs. Without changing any of the settings previously established, advance the CORRECTION potentiometer gradually up to about 95%. Observe the current at a vertical gain of about 100 mV/div and a sweep speed of about 500 µs/div. As the CORRECTION percentage increases, a small transient will emerge at the leading and trailing edges of the command step. At the same time the noise in the measured current will increase. By the time the CORRECTION potentiometer setting has reached 95%, a ringing transient with a peak-to-peak amplitude of roughly 400 pA will have emerged (Figure 6g). The transient should "ring" (i.e., oscillate) and will reverse polarity 2 or 3 times before disappearing into the noise somewhat less than 1 ms (in 5 kHz bandwidth) after the beginning of the command step.

This residual transient is easily eliminated. Usually all that is required is to slightly (about 1%) increase the setting of the WHOLE CELL CAP. potentiometer setting to make the residual transient disappear into the noise (Figure 6h). If this is not sufficient, very small readjustments of the SERIES RESISTANCE control and FAST MAG and FAST τ may also be required.

Page 41: Axopatch 200B Manual Final
Page 42: Axopatch 200B Manual Final
Page 43: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 27

Chapter 3

At this point, 95% of the approximately 10 MΩ series resistance has been compensated; the residual series resistance is 500 kΩ. An ionic current of 2 nA amplitude would now cause only a 1 mV error in the membrane potential relative to the command potential, i.e., a 20-fold reduction from the situation prior to the use of CORRECTION. Moreover, the true membrane potential is established within about 30 µs after the start of the step command without overshoot or ringing. In addition, the bandwidth of current measurement has been increased from 480 Hz to about 9.6 kHz (of course the measurement bandwidth is still restricted to 5 kHz by the output filter). It is this increase in the bandwidth of current measurement that is responsible for the increased noise as the CORRECTION percentage is increased.

Turn on and off the WHOLE CELL CAP. switch and observe the improvement in performance. Set the vertical gain of the oscilloscope to 2 V/div and set the sweep speed to 1 or 2 ms/div. With the switch ON, the trace should be essentially flat. Recall that turning off this switch not only eliminates the correction signal applied to the 5.1 pF capacitor in the headstage used to compensate for whole-cell capacity transients, but also disables PREDICTION; however, CORRECTION is not disabled. With the WHOLE CELL CAP. switch turned off, series resistance is still compensated via positive feedback of the measured current. Turning off this switch will result in a large ringing capacity transient, with a peak-to-peak amplitude of more than 10 nA. At a higher vertical gain on the oscilloscope (e.g., 200 mV/div), 7 or 8 discernible peaks can be observed in this transient before they disappear into the noise (about 2 ms following the beginning of the step command). The membrane potential will also ring severely and have a 1% settling time of nearly 2 ms. Turning the switch back ON completely eliminates the transient and results in a large improvement in stability: the true membrane potential changes smoothly without ringing to its new value in about 30 µs following the step command. Similar results can usually be achieved with real cells.

The percentage CORRECTION can be increased beyond 95%; 100% can often be achieved with the 10 µs LAG. However, with the parameters of the model cell, as CORRECTION is increased beyond about 95% the current record shows periodic noise (about 200 µs period; which is essentially the same as the ring frequency observed above) that may interfere with current measurement. This can be eliminated by increasing the LAG setting. However, increasing the LAG setting filters the signal used in CORRECTION (e.g., 10 µs corresponds to a 16 kHz 1 pole RC filter, 20 µs corresponds to an 8 kHz filter, etc.). This oscillatory "noise" will also virtually disappear with the output filter set at 1 or 2 kHz. Similarly, also note that at the 95% level of CORRECTION the LAG control setting can be reduced to about 2-3 µs before the system becomes unstable. However, once again, noise will increase.

Page 44: Axopatch 200B Manual Final

28 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Current Clamp (Model Cell) The usual mode of operation of a patch clamp is voltage clamp. In V-CLAMP mode the membrane potential is controlled and the current needed to maintain that potential is recorded. It is, however, often useful to allow the membrane potential to change while keeping the current constant. This can be done in I-CLAMP mode in which the membrane current is controlled (often at zero) while the membrane potential is recorded.

Current-clamp mode can be used in either the WHOLE CELL or PATCH configurations. However, it is most often used in the WHOLE CELL configuration. All time varying current command signals must be generated externally and brought in through the EXT. COMMAND BNCs. A DC holding current can be generated in I-CLAMP using the HOLDING COMMAND control; in current clamp, SEAL TEST generates a 50 pA (β = 1) or 500 pA (β = 0.1) current steps.

To obtain the most accurate cell parameter measurements while in I-CLAMP, it is important to have the PIPETTE CAPACITANCE COMPENSATION set correctly before entering I-CLAMP mode. This is because in I-CLAMP (for a first approximation) the pipette capacitance (Cp) appears to be in parallel with the membrane capacitance (Cm). If the PIPETTE CAPACITANCE COMPENSATION is set too low, the measurement of Cm will be too high and vise versa.

Typically, you will change to I-CLAMP mode after establishing whole-cell recording in V-CLAMP mode. Assuming you have just completed going through the whole-cell voltage clamp tutorial, you have already set the PIPETTE CAPACITANCE COMPENSATION controls to minimize fast transients. You may turn off CORRECTION, PREDICTION and WHOLE CELL CAP. if you wish; they will all be disabled automatically when you switch from V-CLAMP to TRACK (I=0) or I-CLAMP. Do not change the PIPETTE OFFSET control as that will lead to erroneous Vm measurements.

Switch the MODE to TRACK (I=0). You are now in a slow current clamp, but all commands are being ignored and the current is being clamped at zero. Now is the time to set the external command signal and holding current command desired when you go into I-CLAMP mode.

Page 45: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 29

Chapter 3

Set the external command for a 5 Hz, 100 mVp-p1 rectangular waveform. This will cause a

200 pAp-p2 current to be forced through the model cell. Switch the MODE to I-CLAMP

NORMAL. The signal at the SCALED OUTPUT is now Vm. Set the output gain to x10. There should be approximately 1 Vp-p

3 square waveform (Vm ≈ 100 mVp-p) with a time constant (τ) of 16.5 ms (± 15%, 10 - 90% rise time of 36.3 ms) that is identical to the τ of the model cell.

To observe the two different speeds of the I-CLAMP loop and at the same time confirm that the current clamp loop settles much faster than Vm of the model cell, connect an oscilloscope channel to the I OUTPUT BNC. Switch the rear panel slide switch near the BNC to 100 βmV/pA ("up" position).

The observed current signal will have an amplitude of 20 Vp-p. With the MODE switch set to I-CLAMP NORMAL the 10 - 90% rise time will be about 250 µs with about 10% overshoot. Switch the MODE to I-CLAMP FAST and the response will change to a rise time of about 35 µs with less than 10% overshoot (the internal rise time under these conditions is about 10 µs, the signal at the I OUTPUT BNC is filtered at 10 kHz and thus is limited to a rise time of 35 µs).

SERIES RESISTANCE compensation is functional in I-CLAMP mode when CORRECTION is on, and may be used to compensate for the pipette resistance. This is most important during intracellular or whole-cell measurements when one wishes to measure the membrane potential or pass current to control the membrane potential. Since the fraction of series resistance compensation applied is determined by the amount of CORRECTION, set the CORRECTION knob to 100% when using series resistance compensation in I-CLAMP mode. However, remember to reduce the amount of CORRECTION before returning to V-CLAMPmode in order to avoid oscillations. Viewing SCALED OUTPUT, use the SERIES RESISTANCE knob to bring the voltage level during the pulse to the baseline level. Membrane changes may now be more accurately recorded and injected current (using the HOLDING COMMAND controls) more reliably used to control membrane potential.

It is good experimental practice to monitor both the current and the voltage. With good current clamp, the current step should be square. In current clamp, monitoring the current 1 mVp-p - mV peak-to-peak 2 pAp-p - pA peak-to-peak 3 Vp-p - V peak-to-peak

Page 46: Axopatch 200B Manual Final

30 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

via the back-panel I OUTPUT connector as well as the voltage output via the SCALED OUTPUT will allow you to assess the fidelity of the clamp, and to determine whether I-CLAMP FAST or I-CLAMP NORMAL is more suited to a particular cell and electrode combination.

Single-Channel Recording (Real Cell) Set the front panel controls of the Axopatch 200B as follows:

PIPETTE OFFSET: About 5.0 ZAP: 0.5 ms SERIES RESISTANCE COMP. % PREDICTION: 0 %, OFF SERIES RESISTANCE COMP. % CORRECTION: 0 %, OFF SERIES RESISTANCE COMP. LAG: 1 µs WHOLE CELL CAP.: 0 pF, OFF SERIES RESISTANCE: 0 MΩ HOLDING COMMAND: 0 mV, toggle x1, OFF SEAL TEST: ON METER: Set switch to VTRACK MODE: TRACK CONFIG.: PATCH OUTPUT GAIN: α = 10 or as desired LOWPASS BESSEL FILTER: 1,2, or 5 kHz LEAK SUBTRACTION: ∞ MΩ, OFF

Set the oscilloscope to line triggering. Switch off any external step generators.

Insert the pipette and its holder into the input connector of the patch clamp. Be sure to touch grounded metal before doing this to discharge any static charge that may have inadvertently built up on you or on the holder. Be sure to support the headstage with your other hand so that the micromanipulator will not have to absorb the force as you firmly insert the holder. Lower the pipette into the bath. Any voltage offset between the bath electrode and the patch electrode will show up as a non-zero tracking voltage on the meter. Adjust the PIPETTE OFFSET potentiometer until VTRACK is zero. (Some investigators

Page 47: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 31

Chapter 3

prefer not to use the tracking circuit but rather do their early adjustments and seal formation in voltage clamp mode. If you prefer this approach, place the MODE switch in the V-CLAMP position and the meter switch to I before placing the pipette tip into the bath. With the tip in the bath, adjust the PIPETTE OFFSET control until the current on the meter reads zero.)

Note: This is a good point to check the stability of your bath (ground) and patch (recording) electrodes. Drifting electrodes will cause a constantly changing VTRACK voltage in TRACK mode whereas in V-CLAMP the current will continually drift off zero.

Switch the external command switch to the SEAL TEST position. A 5 mV positive-going rectangular pulse is applied to the patch clamp input. The pulse frequency is the same as the AC line frequency and the duty cycle is 50%. This will result in a square pulse of current whose amplitude depends on the pipette resistance as shown in Figure 7.

5 nA1 nA100 pA1 M

10 M

100 M

PIPETTERESISTANCE

(ohms)

CURRENT PULSE

Figure 7. Pipette resistance measurement using seal test.

Lower the pipette until it just touches the cell. You will be able to tell when this happens because the resistance will increase as the pipette makes contact with the cell membrane, causing a decrease in the size of the current pulse (Figure 8, three upper traces). Continue to lower the pipette until the fractional resistance increase is optimal to promote sealing to the cell. For many cells, a two-fold resistance increase is optimal (the size of the square

Page 48: Axopatch 200B Manual Final

32 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

pulse on the oscilloscope decreases in half) but you will have to determine the optimum for your particular cell type.

2 ms

250 pA

IN BATH

PUSHED AGAINST CELL

MORE PRESSUREAGAINST CELL

GIGOHM SEAL

Figure 8. Change in resistance while forming a seal.

At this point, apply suction to the tubing connected to the pipette holder. Some investigators prefer to use mouth suction while others find that a 10 ml syringe works well. With the latter approach, it is best to start with the syringe plunger pulled back to a volume of 5-7 ml so that approximately this volume exists in the system before suction is applied. Pulling the syringe back by about 1 ml from this starting volume will reduce the pressure by 0.2 to 0.14 atmospheres. Since many other schemes of applying suction are used by investigators, we do not provide here an exhaustive description. As suction is applied, there should be a sudden increase in the resistance. Once a gigohm seal is established, the rectangular current pulse will disappear entirely and be replaced by capacitance transients in synchrony with the rising and falling edges of the command pulse (Figure 8, lowest trace).

Page 49: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 33

Chapter 3

If the seal was made using the tracking circuit, this circuit can now be turned off by switching to the V-CLAMP mode. Turn off the SEAL TEST switch and turn the meter switch to the IRMS position. Check the noise level on the meter and decide if the quality of the seal and pipette coating are satisfactory. With the Axopatch 200B integrating headstage, proper pipette coating, low loss pipette glass, and proper grounding and shielding, it is possible to obtain noise levels in the 0.095-0.12 pA rms range if a high resistance seal is obtained. Forgetting to turn off the SEAL TEST signal will result in very high noise levels because of the noise contribution of the capacitance transients.

Turn the SEAL TEST signal back on and, if you plan to do pulsed experiments, proceed to cancel the capacitance transients. This is done using the fast and slow PIPETTE CAPACITANCE COMPENSATION circuits. First, turn the SLOW MAG control fully counterclockwise so that there will initially be no compensation for a slow component. Adjust the FAST MAG and FAST τ controls until the capacitance transient is minimized. (Note: with some practice you will be able to adjust MAG and τ simultaneously and decrease the time required to do the adjustments). If a slow component remains, minimize its size using the SLOW MAG and SLOW τ controls. Turn off the SEAL TEST switch and you are ready to record channel currents.

In some experiments, you may not require complicated pulse protocols and the application of a simple DC voltage to the Axopatch 200B command line is all that is required. For this, you can use the HOLDING COMMAND circuit included in the Axopatch 200B. Turn the meter switch to VHOLD/IHOLD and the meter will display the absolute value of the voltage set by the HOLDING COMMAND controls. This is true even with the HOLDING COMMAND switch in the OFF position. This feature allows you to preset the voltage without delivering it to the preparation until you are ready. Note that the HOLDING COMMAND toggle gives you the choice of two ranges, x1 and x5. The voltage actually delivered to the preparation will be positive or negative depending on the position you choose for the HOLDING COMMAND switch (+ or -).

At this point, you should be seeing single-channel currents if channels exist in the patch. Adjust the OUTPUT GAIN rotary switch to a position appropriate for the size of the currents.

If you plan to use an external function generator or the output of a computer driven D/A converter to provide command potentials, the signal should be connected to the EXT. COMMAND INPUT BNC (front switched or rear-switched) on the back of the

Page 50: Axopatch 200B Manual Final

34 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Axopatch 200B. This is switched with an attenuation of 20 mV/V when the EXT. COMMAND switch is set to the ON position.

When the HOLDING COMMAND is changed, the current baseline will move off zero current. This is due to the current flowing through the seal resistance. How much the trace moves will depend upon the magnitude of the seal resistance. For a large seal resistance, only small currents flow; but for low seal resistances, the currents can be large. For example, for a 1 GΩ seal and a 100 mV HOLDING COMMAND, the seal current will be 100 pA. If, for example, you have set the overall Axopatch 200B gain so that you can look at a 5 pA channel current taking up one division at 1 V/div on the oscilloscope screen, the application of this HOLDING COMMAND will cause the current trace to go right off the screen. If you were digitizing the current with an A/D converter, the baseline current would also exceed the range of some converters. The Axopatch 200B includes a leak subtraction circuit to alleviate this problem. With the HOLDING COMMAND applied, turn the LEAK SUBTRACTION control until the baseline current returns to zero. If the baseline current is really seal current, it should exhibit linear behavior. Once the LEAK SUBTRACTION is set properly it should be possible to toggle the HOLDING COMMAND switch between + and - and the baseline current should stay in the center of the oscilloscope screen. (Note: it is possible for patches to contain small channels or electrogenic transporters that do not produce discernible single-channel events. These will appear to be part of the seal current and may impart apparent non-linear behavior to the seal.)

The Axopatch 200B contains a 4-pole internal Bessel filter with five frequencies for filtering the current output. For many experiments, the 1, 2, 5, 10 and 100 kHz settings available will be sufficient. If other frequency cutoffs are required, they must be provided from an external filter. For this purpose, set the internal filter to 100 kHz and choose the desired bandwidth from your external filter. Note that the bandwidth achieved will be that of the cascaded filters.

Page 51: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 35

Chapter 3

Whole-Cell Recording (Real Cell) Set the front panel controls of the Axopatch 200B as follows:

PIPETTE OFFSET: About 5.0 ZAP: 0.5 ms SERIES RESISTANCE COMP. % PREDICTION: 0 %, OFF SERIES RESISTANCE COMP. % CORRECTION: 0 %, OFF SERIES RESISTANCE COMP. LAG: 1 µs WHOLE CELL CAP.: 0 pF, OFF SERIES RESISTANCE: 0 MΩ HOLDING COMMAND: 0 mV, toggle x1, OFF SEAL TEST: ON METER: Set switch to I MODE: TRACK CONFIG.: WHOLE CELL (β = 1) OUTPUT GAIN: α = 10 or as desired LOWPASS BESSEL FILTER: 1,2, or 5 kHz LEAK SUBTRACTION: ∞ MΩ, OFF

Insert the pipette and its holder into the input connector of the patch clamp. In this case, the pipette should usually contain a filling solution that is low in Ca2+ and the tip should be as large as possible to minimize the pipette resistance (the importance of low pipette resistance is obvious from the previous series resistance compensation discussion). At this point, proceed to insert the holder, lower the pipette into the bath and make a seal in the same manner as described in the preceding single-channel recording tutorial.

A pulse of strong suction is required to rupture the cell membrane. This can again be done by mouth suction or by a syringe. If you are using a 10 ml syringe, disconnect the suction line from the syringe end briefly and push the plunger most or all of the way in. Reconnect the suction line. Draw back slightly on the plunger until a large capacitance transient suddenly appears. An example of this is shown in Figure 9. Again, many schemes of rupturing the cell have been used by investigators.

Page 52: Axopatch 200B Manual Final

36 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

The Axopatch 200B contains a ZAP circuit to aid in breaking into the cell. This circuit delivers a pulse of 1.3 VDC volts for variable durations ranging from 0.5 to 50 ms. The user sets the duration which is optimal for the particular cells in use by adjusting the potentiometer that makes up the outer part of the control. When the center button on the control is depressed, the pulse is delivered. A successful result will again look like that in Figure 9. If the patch is not disrupted, the pulse duration can be increased and the pulse applied a second time, and so on. Some investigators have found that the application of moderate suction while the voltage pulse is given results in a higher incidence of successful patch disruption. The reappearance of the original rectangular pulse either means that you have lost the seal or that the cell does not have a large input resistance. It is not unusual for small cells to have an input resistance of several gigohms but with active conductances it might be as low as a few tens of megohms.

300 pA

2 ms

SEAL

ADJUST FAST MAG

AND FAST τ

GO WHOLE CELL

Figure 9. Going whole-cell: capacity transients observed when rupturing the patch.

Note: with some cells it has proven nearly impossible to go whole cell without loss of seal. If you have one of those cells, you might consider the PERFORATED PATCH technique. In this approach, the very tip of the pipette is filled with a normal filling solution and the rest of the pipette is backfilled with the same filling

Page 53: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 37

Chapter 3

solution to which 120-150 µg/ml of Nystatin or Amphotericin B [from a stock solution of 30 mg/ml in DMSO] has been added. Over a 5-30 min. time period these polyene antibiotics form myriad tiny cation-selective, voltage-independent channels in the membrane patch. These channels allow small ions to equilibrate between the cell and the pipette allowing the cell to be voltage clamped through the open channels. Since substances as large as, or larger than, glucose will not permeate these channels, cell contents are not washed out as in standard whole-cell techniques. This is an advantage or a disadvantage, depending on the experiment. With this technique, a rise in whole-cell capacity transients will be observed as the antibiotic partitions into the cell as shown in Figure 10.

400 pA

3 ms

1 MIN

2 MIN

3 MIN

4 MIN

5 MIN

Figure 10. Going whole-cell: capacity transients observed during amphortericin partitioning.

Now, turn off the SEAL TEST switch. The panel meter reads the output voltage of the tracking circuit that, at this time, represents the command voltage necessary to keep the whole-cell current zeroed. This is, by definition, the resting voltage of the cell. If you have made the seal and gone whole cell in V-CLAMP mode, you will see a DC shift in the

Page 54: Axopatch 200B Manual Final

38 • USE OF THE PATCH CLAMP – A TUTORIAL

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

current since the ionic gradients available to drive channel current will not be balanced by a resting voltage. Using the HOLDING COMMAND control with the HOLDING COMMAND switch set to (-) and the toggle set to (x1), dial the current to zero either using the trace on the oscilloscope as a null indicator or by switching the meter switch to the I position and making the meter read zero. At this point, the reading on the HOLDING COMMAND control or on the meter set to VHOLD/IHOLD will be the membrane resting potential of the cell. Note that the membrane resting potential recorded under these conditions may not be the same as the resting potential in intact cells because the membrane resting potential depends on the ionic compositions of the pipette and the bath solutions.

Many investigators prefer to prevent the cell from experiencing this loss of HOLDING COMMAND as they go whole-cell since some excitable cells die quickly once depolarized. To avoid this, some researchers set the HOLDING COMMAND value at the anticipated resting potential of the cell. This will be applied to the cell interior at the instant they achieve the whole-cell configuration. If there is still a DC current shift (i.e., their anticipated voltage was not quite correct), minor adjustments of the HOLDING COMMAND control can be made to zero the current.

Another possible scheme to go whole-cell without loss of resting potential is to disrupt the patch while in I-CLAMP mode. The current-clamp circuit is fast enough to keep the membrane current at zero and thus keep the cell at its resting potential during the patch disruption. It is important that no external command be applied while in current clamp; otherwise, cell voltage changes caused by the command will occur.

If you went whole-cell in either I-TRACK or I-CLAMP, set the desired HOLDING COMMAND and turn the MODE switch to V-CLAMP. Turn the SEAL TEST back on. Turn the WHOLE CELL CAP. switch to ON. You are now ready to adjust the whole-cell capacitance compensation. Using both the WHOLE CELL CAP. and SERIES RESISTANCE controls simultaneously, adjust them until the capacitance transient is minimized. It is also quite possible that the PIPETTE CAPACITANCE COMPENSATION will require slight readjustment at this time. (See Whole-Cell Capacitance Compensation in the Whole-Cell Recording (Model Cell) section of the TUTORIAL).

This procedure leads to unique settings of the SERIES RESISTANCE and WHOLE CELL CAP. controls corresponding to the electrode and cell being clamped. When the transient is minimized, the access resistance (pipette resistance and any resistive contribution from cell contents) and the cell capacitance can be read on the SERIES RESISTANCE and WHOLE

Page 55: Axopatch 200B Manual Final

USE OF THE PATCH CLAMP – A TUTORIAL • 39

Chapter 3

CELL CAP. controls, respectively. Usually, the access resistance will be about three times that of the pipette alone. The whole-cell capacitance value can be used to estimate the total surface area of the cell assuming that 1 cm2 of membrane has 1 µF capacitance.

If your currents are large, you will want to use series resistance compensation to increase the frequency response of the whole-cell clamp and to minimize the command voltage error due to the IR drop across the access resistance. The proper adjustment of these controls is discussed in the Whole-Cell Recording (Model Cell) section above.

The Axopatch 200B contains a 4-pole internal Bessel filter with five frequencies for filtering the current output. For many experiments, the 1, 2, 5, 10 and 100 kHz settings available will be sufficient. If other frequency cutoffs are required, they must be provided from an external filter. Again, be sure that you take into account that the final bandwidth is that of the cascaded internal and external filters.

We do not recommend the use of LEAK SUBTRACTION in the whole-cell configuration since with whole-cell recordings it is exceedingly difficult to determine the fraction of the leak current due to the seal vs. the fraction due to background currents which might have some dependence on voltage. Software packages like pCLAMP allow a user-specified after-the-fact leakage correction, which is a much safer procedure.

Page 56: Axopatch 200B Manual Final
Page 57: Axopatch 200B Manual Final

COMPUTER INTERFACING • 41

Chapter 4

Chapter 4

Interfacing a Computer to the Axopatch 200B

The Axopatch 200B has many features that allow extensive interactions with a laboratory computer. Some are inputs and some are outputs. This section describes the use of these features with a computerized patch clamp setup.

Most whole-cell experiments and many single-channel experiments require complex voltage-step protocols. These are best provided by D/A converters interfaced to a computer. For this purpose, Axon Instruments has developed pCLAMP and AxoScope software that works in conjunction with the Digidata series interfaces. The output voltage from these D/A converters can be delivered either through the front-switched or the rear-switched EXT. COMMAND inputs. We recommend the front-switched EXT. COMMAND because it is often useful to be able to disconnect the computer input via the front panel.

The scaled and filtered current is also available either from a front panel or rear panel BNC (SCALED OUTPUT). One channel of the computer's A/D converter can be connected to this output for simultaneously sampling the current as the computer is delivering command voltages through its D/A converters. The I-OUTPUT is provided for users who choose to record at a high bandwidth and filter digitally at a later time. This BNC provides the current at a 10 kHz bandwidth and a rear switched gain of either βmV/pA or 100 βmV/pA, where β is the headstage gain.

While you are busy conducting patch-clamp experiments it is very easy to forget to record the settings of switches that specify the size of the current, bandwidth, etc., making it impossible to analyze the currents. Therefore, the Axopatch 200B provides the majority of this information via telegraph outputs on the rear panel. By connecting these outputs to several channels of the computer's A/D converters, it is possible to have your software

Page 58: Axopatch 200B Manual Final

42 • COMPUTER INTERFACING

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

interrogate the outputs before recording data so as to automatically determine the important switch settings at the time the data is recorded. This information can be included with the data record so that analysis programs can read it and, thus, always have the correct information for analyzing the recording (e.g., scaling).

These outputs include telegraphs for settings of the following switches: gain, filter (frequency), and mode. In addition, an output is available which is proportional to the whole-cell capacitance setting on the front panel WHOLE CELL CAP. potentiometer. (See Output BNCs in REFERENCE SECTION: GENERAL INFORMATION section for details of actual voltages and settings.)

For both single-channel and whole-cell experiments, there is a rear panel BLANK ACTIVATE INPUT. It holds the current at the level that exists at the time a positive TRANSISTOR TRANSISTOR LOGIC (TTL) pulse is applied to this input. The current is held at this level as long as the pulse is HI. This feature is particularly useful for blanking out the capacitance transient associated with the leading edge of a voltage command. One way to utilize it is by using a second channel of D/A converter or a digital output to activate this line either just before or simultaneously with the application of a voltage step to the EXT. COMMAND input. The duration of the blanking input would be that expected for the duration of the capacitance transient.

Two rear panel BNCs are provided specifically for use in single-channel recording with the integrating headstage. One is the FORCED RESET INPUT. A disadvantage of integrating headstages is that they must be reset periodically. The larger the current that flows, the more often resets occur. Although the reset glitches are small they may be bothersome to some single-channel detection schemes. In pulsed experiments it is possible to minimize the occurrence of these glitches by resetting the integrator immediately before applying a voltage step. For many kinds of channels this ensures that no reset glitches will occur during the brief recording period following the onset of the command. This is again done by using a second channel of D/A converter or a digital output to apply a TTL pulse to the FORCED RESET line just before applying the command step. To ensure that one does not unknowingly utilize data taken during a reset or an external blanking, a DATA NOT VALID OUTPUT is supplied to the user. This line produces a TTL HI (positive) during the time that the output is held via the sample and hold circuit either during reset or blanking. The user is free to implement some scheme to interrogate this line during data collection to determine precisely when the data is not valid.

Page 59: Axopatch 200B Manual Final

LOW NOISE RECORDING • 43

Chapter 5

Chapter 5

Low Noise Recording Techniques

The PATCH configuration of the Axopatch 200B is capable of producing single-channel recordings with significantly lower noise than a standard resistive patch clamp because of the inherently low noise of the integrating headstage, particularly at low to moderate frequencies (below 10 kHz). Low-noise performance is further improved by active cooling of the critical circuitry. To realize this performance the user must pay close attention to other sources of noise. This is because the total rms noise of a patch clamp recording is the square root of the sum of the individual squared rms noise sources. This means that any particular noise source that is large will dominate the total noise and make other noise sources insignificant. Therefore, all potentially contributing noise sources must be minimized. Specifically, the headstage, the pipette glass, the holder, and the seal contribute significantly even under circumstances where extraneous noise pickup from the environ-ment is negligible. It is absolutely crucial that the entire preparation be properly shielded and hum from power supply, mains, and other sources be negligible, i.e., < 0.1 pAp-p. (Actually, < 0.01 pAp-p is possible with some effort). In this section, we suggest some approaches to low-noise recording of single channels. While these same approaches are a good idea for whole-cell recording, they are less important there since in whole-cell recording the dominant noise source comes from the access resistance in series with the whole-cell capacitance.

Page 60: Axopatch 200B Manual Final

44 • LOW NOISE RECORDING

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Glass Type and Coating The noise from pipette glass itself arises from the lossy characteristics of its walls1. Therefore, it is expected that glasses with the lowest inherent dielectric loss will have the lowest noise. Generally, the thicker the wall is, the lower the noise will be. These expectations have been largely born out by actual experiments. Table I presents the specifications from a large number of commercially available glasses that have been used for patch voltage clamping. Each of these glasses has been shown to be sealable to cell membranes in several different cells. Aluminosilicate glasses like Corning #1723 and high lead glasses like Corning #8161 are particularly noteworthy for their low inherent noise but have not found much acceptance for use in patch clamp studies. Aluminosilicate glasses are hard to pull because of their high softening temperature and some high lead glasses have been reported to modify channel currents (e.g., see Cota and Armstrong, Biophysical J. 53:107-109, 1988; Furman and Tanaka, Biophysical J. 53:287-292, 1988). Since any glass may potentially modify channel currents, one must be aware of this fact and control for it regardless of the glass one uses. We recommend two glasses: Corning #7052 and quartz. Both have been successfully sealed to many different cell types. Quartz, with its significantly lower-loss factor, has produced the lowest noise recordings known to us. However, because of its extremely high-softening temperature, quartz requires a special puller like the P-2000 from Sutter Instrument Company.

Pipette glass can be obtained from specialty glass houses such as:

Clark Electromedical Instruments P.O. Box 8, Pangbourne, Reading, RG8 7HU, England, (073) 573-888

Garner Glass 177 S. Indian Hill Road, Claremont, CA 91711, USA, (909) 624-5071

Jencons Scientific Cherycourt Way Industrial Estate, Stanbridge Road, Leighton Buzzard Bedfordshire LU7 8UA, UK, (0525) 372-010

Sutter Instrument Company 40 Leveroni Court, Novato, CA 94949, USA, (415) 883-0128

1 When a sine voltage is applied across a perfect dielectric, the alternating current should be 90° out of phase with the

voltage. The deviation from 90° is the "loss factor". The loss factor is related to the power dissipated in the dielectric. Since energy is lost in the dielectric, dielectrics (e.g., glasses) are commonly referred to as "lossy".

Page 61: Axopatch 200B Manual Final

LOW NOISE RECORDING • 45

Chapter 5

Even if one uses electrically superior glasses, low noise will not be obtained unless the outer surface of the glass is coated with a hydrophobic substance, such as Dow Corning Sylgard #184. This substance prevents the bathing solution from creeping up the outer wall of the pipette glass. This is important since a thin film of solution on the outer surface of the glass produces a distributed resistance that interacts with the glass capacitance to produce a noise source that rises with frequency. Since it becomes the dominant noise source, it must be eliminated. While many other hydrophobic substances have been used, none, to the best of our knowledge, produces as low noise as does Sylgard #184. Sylgard also decreases the capacitance of the pipette wall and so reduces the lossiness of the wall as well. It has been shown experimentally that Sylgard will improve the noise of any glass but it will not turn a poor electrical glass into a good one. Low-loss glasses coated with Sylgard give significantly less noise than poor glasses coated with Sylgard. Obviously, the closer to the tip that the Sylgard can be painted the lower the noise.

Sylgard can be obtained from:

Dow Corning 2200 Salzburg, Midland, Michigan 48611, USA, (517) 496-6000

K.R. Anderson 2800 Bowers Avenue, Santa Clara, CA 95051, USA, (800)538-8712

UTSU SHOJI Tokyo, Japan, 03-3663-5581

Page 62: Axopatch 200B Manual Final

46 • LOW NOISE RECORDING

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

TABLE I Glass Electrical And Thermal Properties

Log10 Loss Volume Dielectric Softening Glass Factor Resistivity Constant Temp. °C Description 7940 .0038 11.8 3.8 1580 Quartz (fused silica) 1724 .0066 13.8 6.6 926 Aluminosilicate 7070 .25 11.2 4.1 ---- Low loss borosilicate 8161 .50 12.0 8.3 604 High lead Sylgard .58 13.0 2.9 ---- #184 Coating cmpd. 7059 .584 13.1 5.8 844 Barium-borosilicate 7760 .79 9.4 4.5 780 Borosilicate EG-6 .80 9.6 7.0 625 High lead 0120 .80 10.1 6.7 630 High lead EG-16 .90 11.3 9.6 580 High lead 7040 1.00 9.6 4.8 700 Kovar seal borosilicate KG-12 1.00 9.9 6.7 632 High lead 1723 1.00 13.5 6.3 910 Aluminosilicate 0010 1.07 8.9 6.7 625 High lead 7052 1.30 9.2 4.9 710 Kovar seal borosilicate EN-1 1.30 9.0 5.1 716 Kovar seal borosilicate 7720 1.30 8.8 4.7 755 Tungsten seal borosilicate 7056 1.50 10.2 5.7 720 Kovar seal borosilicate 3320 1.50 8.6 4.9 780 Tungsten seal borosilicate 7050 1.60 8.8 4.9 705 Series seal borosilicate KG-33 2.20 7.9 4.6 827 Kimax borosilicate 7740 2.60 8.1 5.1 820 Pyrex borosilicate 1720 2.70 11.4 7.2 915 Aluminosilicate N-51A 3.70 7.2 5.9 785 Borosilicate R-6 5.10 6.6 7.3 700 Soda lime 0080 6.50 6.4 7.2 695 Soda lime

Page 63: Axopatch 200B Manual Final

LOW NOISE RECORDING • 47

Chapter 5

The holders supplied with the Axopatch 200B are made of polycarbonate. Polycarbonate was experimentally found to produce the lowest noise among ten substances tested. It was only slightly better than polyethylene, polypropylene, and Teflon, but was much better than nylon, Plexiglass, and Delrin. Axon Instruments holders avoid metal and shielding, which are noise sources. Holders, however, do become a significant noise source if fluid gets into them. Therefore, great care must be taken in filling pipettes with solution. They should be filled only far enough from the tip so that the end of the internal chlorided silver wire or silver/silver chloride pellet is immersed. Any solution that gets near the back of the pipette should be dried with dry air or nitrogen to keep it from getting into the holder. Holders that become contaminated with solution should be disassembled and sonicated in ethanol or pure deionized water, and allowed to dry thoroughly before being used again. It is also a good idea to periodically clean the holders by sonication even if no fluid has been observed in them.

The cleanliness of the holder can be checked before each attempt to make a seal. When the holder with a filled pipette has been inserted into the headstage connector, and the pipette tip is positioned just above the bathing solution, the rms current noise seen on the meter of the Axopatch 200B (meter switch in the IRMS position) should not significantly exceed 0.08 pA.

Seal The seal will usually be the dominant noise source if it is only a few gigaohms. Seal resistances in excess of 20 GΩ must be obtained if exceptionally low noise single-channel recordings are to be routinely achieved. Seal quality can be monitored by periodically observing the Axopatch 200B rms noise meter (IRMS). The noise depends also on the depth of the pipette tip below the surface of the bathing solution since the effective pipette capacitance increases as the depth of immersion increases. The voltage noise of the headstage interacts with the pipette capacitance to produce a noise source that rises with frequency. With excised membrane patches lifted to just under the surface of the bathing solution, the integrating headstage can produce background noise as low as 0.083 pA rms in a 5 kHz bandwidth in membrane patches from several preparations, when used with quartz glass pipettes; 0.095-0.12 pA rms is routinely achievable. These numbers can serve as guidelines for what is potentially possible in your experiments.

Page 64: Axopatch 200B Manual Final

48 • LOW NOISE RECORDING

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Signal Generator One last potential noise source to consider is the noise in the signal generator that provides the command. In the Axopatch 200B we have succeeded in minimizing this noise by heavily attenuating the external command. However, it is possible for this noise source to be significant, particularly if the command signal comes from a D/A converter.

Page 65: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 49

Chapter 6

Chapter 6

Reference Section: Instrument Operation

The major topics in this section are organized alphabetically.

Bridge Mode The Axopatch 200B may be used to follow cell membrane potential, in a way similar to the bridge mode of conventional microelectrode amplifiers, when in current clamp (I=0, I-CLAMP NORMAL and I-CLAMP FAST; please see Current Clamp section below). This is possible because the series resistance compensation circuitry is operable in current clamp, and can be used to correct for electrode resistance. Thus, with series resistance correctly compensated in current clamp mode, the voltage at the scaled output will accurately reflect the cell membrane potential. Changes in resting potential, spike activity, and synaptic activity may be monitored in this way (however, please note risetime limitations listed in the SPECIFICATIONS chapter for current clamp).

Compensating for electrode resistance in current clamp is straightforward. Monitor the voltage with SEAL TEST activated or when applying a step current command; you will note a response that reaches a steady-state level, then returns to baseline following the end of the step. Adjust the series resistance knob so as to bring the steady-state voltage level coincident with the baseline level. The electrode resistance is now compensated.

Capacitance Measurements Capacitance measurements may be made in voltage-clamp mode by providing the rear-panel capacitance dither BNC input with a time-varying TTL input to regularly activate the

Page 66: Axopatch 200B Manual Final

50 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

dither. This will effectively increase the measured membrane capacitance by 0.1 % of the full-scale range (thus, by 0.1 pF for β = 1 or 1 pF for β = 0.1) by transiently reducing the setting of the Whole-Cell Cap. knob by that amount. With appropriate software, these controlled changes in capacitance values may be used in the measurement of cell membrane capacitance changes during an experiment (for example, see Joshi and Fernandez, Biophysical Journal (1988) 53: 885-892).

In order to use the Phase-Tracking method of capacitance measurements (for example, see Fidler and Fernandez, Biophysical Journal (1989) 56: 1153-1162), one needs in addition to use the DR-1 series resistance dither unit. The series resistance dither provides a short-circuit link between the preparation and ground that increases to 500 kΩ resistor while a TTL signal is high.

Command Potentials Command potentials can be obtained from two internal sources, HOLDING COMMAND and SEAL TEST, and from external sources via two rear panel BNCs.

Holding Command In voltage clamp this control allows the user to apply a membrane holding potential of ±200 mV (with the toggle set to x1; the full range of ± 1V is available when the toggle is set to x5). In current clamp the control allows application of a holding current of ±2 nA with β = 1 (toggle set to x1) or ±20 nA with β = 0.1 (toggle set to x1). In current clamp, with the toggle set to x5, the control allows application of a holding current of ±10 nA with β = 1 or ±100 nA with β = 0.1. Turn the control fully counterclockwise for a zero command. Use the three position toggle switch to set the polarity or to switch off. This control is automatically switched OFF when either TRACK or I = 0 mode is chosen.

HOLDING COMMAND RANGES MODE Toggle x1 Toggle x5 β = 0.1 β = 1 β = 0.1 β = 1 VCLAMP ± 200 mV ± 200 mV ± 1 V ± 1 V ICLAMP ± 20 nA ± 2 nA ± 100 nA ± 10 nA

Page 67: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 51

Chapter 6

Seal Test

The SEAL TEST command generator provides a convenient source of test pulses to be used during seal formation. The step is gated by the internal line frequency oscillator; its duty cycle is between 40% and 60%. In voltage clamp mode, the step size is 5 mV; in current clamp, it is 50 pA (β = 1) or 500 pA (β = 0.1).

Current Clamp The Axopatch 200B can be used in current-clamp mode similarly to a conventional microelectrode amplifier. Series Resistance compensation is functional in this mode. There are three current-clamp modes. I=0 for voltage recording without an applied current command, I-CLAMP NORMAL for use with pipette resistances greater than 1 MΩ and I-CLAMP FAST for use with pipette resistances greater than 10 MΩ. The accuracy of the response can be improved by using the FAST and SLOW ELECTRODE CAPACITANCE COMPENSATION as well as the SERIES RESISTANCE COMPENSATION controls.

Membrane Potential In voltage-clamp mode, Vm is simply equal to the command potential. In current-clamp mode, Vm is not clamped and automatically becomes the voltage needed to allow the commanded current to flow.

Note: To avoid introducing errors, you should not change the PIPETTE OFFSET control after you switch from V-CLAMP to I-CLAMP.

Whole-Cell Current Clamp Current-clamp can be used from the very beginning of the experiment, or you can switch into current-clamp mode at any time (see Current Clamp section of the TUTORIAL). However, if whole-cell recording has been established, special care must be taken when switching between voltage-clamp and current-clamp modes because in voltage-clamp mode the EXT. COMMAND sensitivity is 20 mV/V and the HOLDING COMMAND sensitivity is 20 mV/turn (for x1; 100 mV/turn with the toggle set to x5). In current-clamp the EXT.

Page 68: Axopatch 200B Manual Final

52 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

COMMAND sensitivity changes to 2 ÷ β nA/V while the HOLDING COMMAND sensitivity changes to 200 ÷ β pA/ turn (for toggle set to x1; the range is five-fold greater with the toggle set to x5).

There are at least two possible ways to achieve a well-behaved transition between voltage-clamp and current-clamp modes:

1) Dedicate one external source to voltage clamp and the other to current clamp. Connect each of them to one of the EXT. COMMAND BNCs on the rear panel. While in the I = 0 mode, turn off the EXT. COMMAND corresponding to voltage clamp and turn on the current clamp EXT. COMMAND. Also, while in the I = 0 mode, readjust (or turn OFF) to HOLDING COMMAND control for use in I-CLAMP.

2) Design your setup to be sensitive to the MODE TELEGRAPH BNC and have your program to automatically scale a single external command accordingly so that a smooth transition between modes is obtained (see Telegraph Output section for values).

Whole-Cell Parameters Although WHOLE-CELL CAP. is disabled in current-clamp mode, SERIES RESISTANCE compensation is active, and may be used to correct for pipette resistance. Thus, one may use the Axopatch 200B to accurately record synaptic, action and field potentials from cells and tissues as well as to record currents from patches and cells. Please note, however, the risetime limitations listed in the SPECIFICATIONS chapter for current clamp. Due to the nature of patch-clamp amplifiers, current clamp mode in a patch-clamp amplifier will not clamp cell current with the fidelity that is possible with a conventional microelectrode amplifier (see below).

Electrode Capacitance Compensation The speed of the pipette time constant depends on the pipette resistance (Rp) and the pipette capacitance (Cp). As a first-order approximation, the pipette time constant depends on the product RpCp. The FAST and SLOW ELECTRODE CAPACITANCE COMPENSATION controls can be used to electronically reduce the effective value of Cp, thus reducing the pipette time constant. This is achieved by injecting a transient current into the headstage input to charge and discharge Cp during signal changes.

Page 69: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 53

Chapter 6

Limitations Unlike most patch-clamp amplifiers, the Axopatch 200B current clamp mode includes Series Resistance Compensation. However, similar to many other patch clamps, current clamp in the Axopatch 200B is not as fast or as stable as current clamp in a conventional microelectrode amplifier such as the Axoclamp or the Axoprobe. This difference in current-clamp performance results from significant differences in the design of the headstages.

In a conventional current-clamp amplifier, the headstage is designed as a voltage follower. Current is injected through a resistor and the pipette voltage is continuously recorded. Alternatively, the headstage of a patch-clamp amplifier is designed as a current follower; the pipette voltage is controlled while the pipette current is measured. To simulate current clamp, a feedback circuit in the main unit of a patch-clamp amplifier automatically adjusts the pipette voltage to keep the pipette current at the desired value.

Like any feedback circuit, the stability is compromised if the open-loop gain is too high. When the headstage is grounded through a pipette, the voltage gain of the headstage is nearly equal the value of the feedback resistor divided by the value of the pipette resistance. In the WHOLE CELL mode and with a pipette resistance of 1 MΩ, this voltage gain is 500 (50 for β = 0.1). In order to guarantee stability with pipette resistances as low as 1 MΩ, the current-clamp circuitry must be deliberately slowed down, compromising the response time for high-resistance pipettes. For low-resistance pipettes the main problem is stability. In the extreme case of a zero-resistance pipette (i.e., a directly grounded input), the enormous voltage gain of the headstage guarantees instability. To compromise between the two conflicting requirements of speed and stability, the Axopatch 200B has been designed with a dual speed current clamp. For low-resistance pipettes (between 1 and 10 MΩ) the I-CLAMP NORMAL setting will guarantee that the loop will be stable. For higher resistance pipettes (above 10 MΩ), I-CLAMP FAST setting will be stable and have a response many times faster than the I-CLAMP NORMAL setting. (It has been observed that in some cases the I-CLAMP FAST setting will still be stable with pipette resistances down to 3 MΩ). Risetime values for different combinations of pipette resistance and cell membrane capacitance are listed in the SPECIFICATIONS chapter for current clamp.

Page 70: Axopatch 200B Manual Final

54 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Electrochemistry The Axopatch 200B may be used to perform electrochemical measurements. Typically, voltammetric experiments require a large command potential range in voltage-clamp mode. This may be accomplished by setting the Holding Command toggle to "x5" for internal commands and by using the 100 mV/V Rear-Switched external command input for external commands. Similarly, amperommetric measurements may be made in one of the I-CLAMP modes with the same command settings (in this case, to control the current instead of the voltage).

For an introduction to electrochemical methods, you may refer to Kawagoe, Zimmerman and Wightman, (1993) Journal of Neuroscience Methods 48: 225-240. Electrodes for electrochemistry are available from Axon Instruments.

Headstage The CV 203BU headstage incorporates several improvements over previous headstages. The overall size is smaller and narrower than other headstages, allowing for greater ease of placement during experiments. Three configuration choices (one patch and two whole-cell) provide a wide useful range of operation. Most importantly, due to improved circuitry and temperature control provided by an internal Peltier cooling unit, the CV 203BU offers the lowest single-channel noise recording available. The critical headstage circuit components are housed in a patented hybrid enclosure (U.S. Patent number 5,285,012), ensuring consistent and reliable low-noise recording for years.

Offset Adjustment For Headstage Headstage offsets drift slowly over time and, therefore, may require renulling every few weeks. To do this, allow the headstage to warm up for a minimum of one hour. In V-CLAMP mode, with PATCH mode selected and with open circuit input, adjust the rear panel PATCH HEADSTAGE OFFSET potentiometer for a zero current reading on the panel meter. Change to WHOLE CELL mode and adjust the rear panel WHOLE CELL HEADSTAGE OFFSET for zero current.

Page 71: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 55

Chapter 6

Frequency Boosting The WHOLE CELL mode of the Axopatch 200B uses a resistor in the headstage feedback that requires frequency boosting (tuning) for optimal performance. The frequency boosting circuit is tuned at the factory and will usually not require readjustment in the field. If tuning is required, see the HEADSTAGE TUNING PROCEDURE (page 121).

Case Ground Connector The metal headstage case and the gold-plated 1 mm socket at the rear of the headstage are connected to ground. Use this ground for grounding the preparation.

No provision is made for driving a shield since using a driven shield around the pipette increases the high-frequency noise.

Headstage Dimensions The headstage box measures 0.7" wide x 0.75" high x 4.2" long.

Mounting the Headstage The improved, slim design of the CV 203BU headstage makes it easier to guide your pipette towards your preparation without physical interference from other instruments in your set-up. For maximum mechanical rigidity, the headstage can be mounted directly to some manipulators using the mounting plate located on the bottom of the headstage. In addition, the bottom of the headstage has been engineered for stable attachment to user-designed mounts.

Cleaning Wipe the headstage connector with a damp cloth to clean salt spills. Avoid spilling liquids on the headstage.

The Teflon input connector should be kept very clean. Effective cleaning can be done by spraying with alcohol or swabbing carefully with deionized water. If possible, avoid the use of Freon since it is thought to be detrimental to the environment.

Page 72: Axopatch 200B Manual Final

56 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Static Precautions The headstage can normally be safely handled. However, if you are in a laboratory where static is high (i.e., you hear and feel crackles when you touch things), you should touch a grounded metal object immediately before touching the headstage.

You should not switch off power to the Axopatch 200B when handling the headstage input since this will upset thermal equilibrium.

Optical Pick-up The Teflon input connector and the glass walls of the hybrid package inside the headstage are translucent. High intensity light can get through in sufficient strength to activate the input transistors inside the hybrid. Therefore, you should prevent bright light from falling on the input connector with a shade or by dimming lights. If you notice line-frequency hum on the current record, it could be due to fluctuating light levels from a bright fluorescent light or equivalent. In general, low light levels are not a problem.

Acoustic Pick-up Rare instances have been reported where the headstage was susceptible to low amplitude acoustic pick-up. The most troublesome being a situation where the audible hum of a nearby isolation transformer was being acoustically coupled to the input of the headstage. This was traced to the silver wire of the electrode and was solved by trimming off a fraction of the wire, thus changing its resonant frequency.

Tuning the Headstage See the HEADSTAGE TUNING PROCEDURE (page 121).

Holders

Features The HL-U series holder provides for enhanced low-noise mechanically stable microelectrode recordings with or without suction. Because the new holder provides a

Page 73: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 57

Chapter 6

universal fit for a very wide range of pipette diameters and will fit any of our redesigned headstages, it is named the HL-U.

The barrel of the holder is made out of polycarbonate for lowest noise. There are two different barrel lengths. The shorter barrel length contributes less to the operating noise and, therefore, is ideally suited for single channel patch clamp recordings. Although the longer barrel will contribute more to the operating noise, the increased length may provide the needed clearance between the headstage and other components in the experimental setup. Maintenance is simple because the holder can be fully disassembled for cleaning and parts replacement.

Mechanical stability of the pipette is assured in several ways. For example, as the pipette cap is closed, the cone washer is compressed on the pipette from the force applied to the front and back of the cone washer. The holder mates with the special threaded Teflon connector on U-type Axon Instruments headstages and is secured in place with a threaded collar.

The holder is designed to emerge along the long axis of the headstage. A right-angle adapter can be purchased if it is necessary for the holder to emerge at 90° from the headstage.

The HL-U holder is designed to be used with Axon Instruments amplifiers, and fit all U-type CV and HS series of headstages. These headstages have a threaded white Teflon collet. To minimize the added noise contributed by the holder in single-channel recording, the holder uses a small (1 mm) pin for the electrical connection and a large amount of insulating Teflon. This noise problem is peculiar to single-channel recording.

Figure 11. Exploded view of the holder.

Page 74: Axopatch 200B Manual Final

58 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Parts The bore size of the HL-U accepts pipettes with an outer diameter (OD) of 1.0-1.7 mm. Pipettes are secured by a cone washer with an inner diameter (ID) that accommodates the pipette OD. Color-coding aids identification of the four sizes of cone washers: 1.0 mm (orange), 1.3 mm (clear), 1.5 mm (orange) and 1.7 mm (clear). Each HL-U is supplied with two barrel lengths, 16 mm and 28 mm.

It has been shown that a Ag/AgCl pellet offers no greater stability than properly chlorided silver wire. Moreover, the diameter of the Ag/AgCl (1 mm) restricts its use to pipettes with a large ID i.e., > 1.1 mm. Therefore, the HL-U is supplied with 0.25 mm silver wire.

Spare components included with each holder are as follows: one 50 mm length of silver wire, 40 cone washers (10 of each size) and one 70 mm length of silicone tubing. Cut into 2 mm lengths, the silicone tubing will yield approximately 30 replacement silicone seals. Additional pipette caps, cone washers, silicone tubing, pins and silver wire can be purchased from Axon Instruments, as well as optional Ag/AgCl pellet assemblies.

Optional Ag/AgCl Pellets The HL-U holder will accommodate a 1 mm diameter Ag/AgCl pellet that should provide many months of DC-stable recordings. The inner diameter (ID) of the pipette must be > 1 mm. The silver wire is surrounded by a wax-sealed Teflon tube. This ensures that the electrode solution only contacts the Ag/AgCl pellet. Three pellet assemblies are sold as HLA-003.

Figure 12. Ag/AgCl pellet assembly.

Page 75: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 59

Chapter 6

Use

Insertion of Pipette Make sure the electrode cap is loosened so that pressure on the cone washer is relieved, but do not remove the pipette cap. Push the back end of the pipette through the pipette cap and cone washer until it presses against the pipette seat. Gently tighten the pipette cap so that the pipette is gripped firmly.

To minimize cutting of the cone washer by the sharp back end of the pipette, you can smooth the pipette edges by rotating the back end of the pipette in a bunsen burner flame before filling the pipette with salt solution.

Cleaning For lowest noise, keep the holder clean. Frequently rinse the holder with distilled water. If more thorough cleaning is required, briefly wash in ethanol or mild soapy water. Never use methanol or strong solvents.

Filling Pipette Only the taper and a few millimeters of the shaft of the pipette should be filled with solution. The chlorided tip of the wire should be inserted into this solution. Avoid wetting the holder since this will increase the noise.

Page 76: Axopatch 200B Manual Final

60 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Silver Chloriding It is up to you to chloride the end of this wire as required. Chloriding procedures are contained in many electrophysiology texts1. Typically the chlorided wire will need to be replaced or rechlorided every few weeks. A simple, yet effective, chloriding procedure is to clean the silver wire down to the bare metal using fine sand paper and immerse the cleaned wire in CHLOROX bleach for about 20 minutes, until the wire is uniformly blackened. This provides a sufficient coat of AgCl to work reliably for several weeks as an internal reference pipette. Drifting or otherwise unstable offsets during experiments is suggestive of the need for rechloriding. The chlorided region should be long enough so that the pipetted solution does not come in contact with the bare silver wire.

Heat smoothing the back end of the pipette extends the life of the chloride coating by minimizing the amount of scratch damage. Another way to protect the AgCl coating is to slip a perforated Teflon tube over the chlorided region.

The chlorided region should be long enough so that the pipette solution does not come in contact with the bare silver wire.

Replacing the Silver Wire To replace the silver wire, insert the nonchlorided end through the hole of the silicone seal and bend the last 1 mm of wire over to an angle of 90°. Press the wire into the back of the barrel making sure that the silicone seal is flush with the back of the barrel. Slip the threaded collar over the back of the barrel. With the large end of the pin directed toward the bent-over wire screw the pin cap down firmly, but without excessive force. This assures good electrical contact. Screw the pin cap down firmly but without excessive force.

1For easy-to-use recipes see Microelectrode Methods for Intracellular Recording and Ionophoresis, by R.D. Purves, London: Academic Press, 1981, p. 51.

The Axon Guide. Foster City, CA: Axon Instruments, Inc., 1993, p. 83.

Page 77: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 61

Chapter 6

Glass Dimensions Use the HL-U for pipettes with outside diameter (OD) of 1.0-1.7 mm. The optimal dimensions should match the inner diameter (ID) of the four sizes of cone washers, 1.1, 1.3, 1.5 and 1.7 mm. When the pipette OD falls between two sizes of cone washers, the larger size cone washer should be used. For instance, if the pipette OD is 1.6 mm, then use a cone washer with an ID of 1.7 mm.

Adapters HLR-U right-angle adapters allow the HL-U series holder to emerge at 90° from the headstage. Use the HLR-U with the HL-U holder.

HLB-U BNC-to-Axon adapter allows conventional BNC-type holders to be used with Axon Instruments U-type headstages. Use the HLB-U with all U-type CV and HS headstages (e.g., CV-4-1/100U and HS-2A-x1MGU). These headstages have a threaded white Teflon collet.

Input BNCs

BLANK ACTIVATE INPUT

A positive TTL pulse delivered to this input causes the output current to be held at the value it had at the instant the pulse was activated and to hold it at this level for the duration of the pulse. This is useful, for example, to blank the capacitance transient associated with the leading edge of a voltage step.

CAPACITANCE DITHER CONTROL

Provides control for the internal capacitance dithering used in capacitance measurement experiments. A TTL HIGH signal enables the dither, which effectively causes a change of 0.1 pF (for β = 1; 1 pF for β = 0.1) in Cm.

Page 78: Axopatch 200B Manual Final

62 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

EXTERNAL COMMAND INPUT (front panel switched)

Applies external command signal divided by 50 to the command input of the Axopatch 200B when the front toggle switch is set to EXT. COMMAND.

EXTERNAL COMMAND INPUT (rear panel switched)

Applies external command signal divided by 10 to the command input when the rear panel switch is engaged. This input should be used for VOLTAMMETRY commands, since the 10-fold scaling factor allows for the greater voltage range required for such experiments (up to ± 1 V).

The two external command inputs are summed together inside the patch clamp.

FORCED RESET INPUT

Causes the integrator and differentiator to reset on the positive edge of a TTL pulse. Used, for example, to force a reset just before applying a command to a cell or membrane patch. This increases the probability that no reset will occur during the brief recording period following the command.

SPEED TEST INPUT

When switched on by the rear panel switch, the SPEED TEST injects a current into the headstage input through a 1 pF capacitor in the headstage. This is used for verifying the dynamic response of the headstage. Injected current waveform is the derivative of the voltage waveform applied at SPEED TEST input. For example, a 100 Hz 10 Vp-p triangle wave will inject a 1 nAp-p square wave into the headstage input.

Page 79: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 63

Chapter 6

Output Section All the controls in the Output Section affect only the signal on the Scaled Output.

Filter The filter is a 4-pole lowpass Bessel filter. The attenuation of signals and noise above the -3 dB frequency is 80 dB/decade (24 dB/octave). The Bessel characteristic is suitable for patch and voltage clamping because it introduces < 1% overshoot. The lowpass bessel filter control allows you to select a bandpass of 1 kHz, 2 kHz, 5 kHz, 10 kHz or 100 kHz.

All lowpass filters slow the rise time of the signal. For filters with < 10% overshoot the 10-90% rise time is: tr ≈ 0.35 fc

where fc is the -3 dB frequency in Hertz. For example, the 10-90 % rise time of a 1 kHz filter is approximately 350 µs.

Note: If you use an external filter: Some manufacturers specify the -3 dB frequency based on the phase response of the filter instead of its amplitude response, or based on a straight line approximation to the filter characteristics instead of the actual characteristics. You should check your external filter by checking tr for a step signal applied to its input.

When a signal with 10-90% rise time t1 is passed through a filter with 10-90% rise time t2, the rise time of the output signal is approximately:

Output Gain (α) There are ten output gain settings in a 1, 2, 5 ratio. These are: 0.5, 1, 2, 5, 10, 20, 50, 100, 200, and 500.

In patch clamp amplifier design, there are conflicting concerns when deciding where to place the Bessel filter and the gain amplifier in the signal pathway. On the one hand, to minimize the noise contribution of the circuitry of the Bessel filter itself, the gain amplifier should come first. On the other hand, if the gain amplifier comes first, the peaks of the amplified signal and noise might get clipped if the gain setting is high. After filtering, it is difficult to distinguish the clipped signals from biological channel currents.

t t tr ≈ +12

22

Page 80: Axopatch 200B Manual Final

64 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

The very effective compromise solution adopted in the Axopatch 200B is to put an initial gain of ten in front of the filter if the gain switch is set to 5 or more. The rest of the gain amplification is after the filter.

The overload (OVLD) light illuminates if either the input to the filter or the output of the final gain amplifier exceeds ±10 V for longer than the 100 µs. This signifies that the signal output is being clipped and has become non-linear.

The electronic gain internal to the Axopatch 200B output section is twice the α value shown on the gain switch. The reason is clear if you consider the output in the WHOLE CELL configuration. To achieve the best compromise between noise and dynamic range, the Axopatch 200B uses a 500 MΩ feedback resistor. The output of the current to voltage converter is, therefore, 0.5 mV/pA. To simplify the daily mental scaling task for the user, the headstage output is presented to the user as 1 mV/pA by including an additional two times (x2) gain. In PATCH mode, the output of the differentiator is also 0.5 mV/pA and, therefore, similar considerations apply.

Output BNCs

CELL CAPACITANCE TELEGRAPH Puts out a voltage between 0 and 10 V that linearly specifies the whole-cell capacitance between 0 and 100 pF (for β = 1; capacitance between 0 and 1000 pF for β = 0.1). The voltage is 0 to -10 V if the WHOLE CELL capacitance switch is turned off. If the WHOLE CELL CAP. switch is ON but the control is overridden (e.g., by current-clamp mode) the telegraph output will be between 0 and +10 V.

DATA NOT VALID OUTPUT Puts out a positive TTL voltage while the current output is blanked either during reset or during external blanking via the blank activate input.

Page 81: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 65

Chapter 6

FREQUENCY TELEGRAPH Provides a series of voltages that can be read by a computer to determine the setting on the filter switch. The voltages and their associated frequency settings are as follows:

Frequency (kHz): 1 2 5 10 100

Frequency Telegraph Voltage Output (V): 2 4 6 8 10

GAIN TELEGRAPH Provides a series of voltages that can be read by a computer to determine the setting of the gain switch. The gain takes α and β factors into account.

I (mV/pA): 0.051 0.11 0.21 0.5 1 2 5 10 20 50 1002 2002 5002

Vm (mV/mV): N/A N/A N/A 0.5 1 2 5 10 20 50 100 200 500

Telegraph Output (V): 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5

These values have been chosen to fit in the range 0-10 V so that they can be read on A/D converter channels of most computers.

I OUTPUT Current output filtered at 10 kHz (3-pole Bessel). Gain is chosen by rear panel switch to be βmV/pA or 100β mV/pA.

1 Applicable for β = 0.1 only 2 Applicable for β = 1 only

Page 82: Axopatch 200B Manual Final

66 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

MODE TELEGRAPH OUTPUT Provides a series of voltages that can be read by the computer to determine the setting on the MODE switch. The voltages and their associated switch settings are as follows:

I-CLAMP I-CLAMP Mode: TRACK V-CLAMP I = 0 NORMAL FAST Scaled Output: I I Vm Vm Vm Telegraph (volts): 4 6 3 2 1

SCALED OUTPUT Provides the current (filtered and scaled) in V-CLAMP and TRACK modes or the membrane voltage (filtered and scaled) in I-CLAMP modes. This BNC is duplicated on the front and back panels. On the front panel, two LEDs specify whether the output is current (I) or membrane voltage (Vm). The current gain is α βmV/pA. The voltage gain is α mV/mV.

10 Vm OUTPUT Provides the membrane voltage multiplied by ten (x10) for use in current-clamp experiments.

In V-CLAMP mode, 10 Vm is ten times the command potential. If the pipette current is zero, or if the series resistance correction is 100%, the command potential is identical to the actual membrane potential. The actual membrane potential is given by the following equation:

10 Vm (100 - CORR) Actual Membrane Potential (mV) =_______ _ ________ I Rs

10 100

where I = pipette current in nA CORR = % correction set on SERIES RESISTANCE CORRECTION control Rs = resistance in MΩ set on SERIES RESISTANCE control

In TRACK mode, 10 Vm is ten times the actual potential.

Page 83: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 67

Chapter 6

In I-CLAMP mode, if the current is zero, 10 Vm is ten times the actual membrane potential. If current is flowing in the pipette, 10 Vm includes the voltage drop across the pipette. The actual membrane potential is given by the following equation:

10 Vm Actual Membrane Potential (mV) = _______ _ I Rs 10

where I = pipette current in nA Rs = resistance in MΩ set on SERIES RESISTANCE control

Panel Meter The main panel meter displays one of five signals. Selection is made by a rotary knob. The five signals are:

1) VHOLD/IHOLD: The value set on the HOLDING COMMAND potentiometer. When VHOLD/IHOLD is selected, it will show the absolute value of the HOLDING COMMAND even if the HOLDING COMMAND control is OFF.

2) I: The pipette current. The reading is automatically scaled to suit the headstage gain. The operation is autoranging; that is, the decimal point and the units indicators automatically shift so that large DC currents can be displayed.

3) IRMS: The rms current noise. The noise is measured in a 5 kHz bandwidth using a Butterworth filter that is independent of the front-panel lowpass Bessel filter. The front-panel gain does not affect this reading.

4) Vm: The membrane potential.

5) VTRACK: The output of the automatic offset nulling circuit.

6) TEMP: Headstage circuitry temperature in °C (range ± 20 °C).

Page 84: Axopatch 200B Manual Final

68 • INSTRUMENT OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Power-Supply Voltage Selection and Fuse Changing

Supply Voltage The Axopatch 200B can be directly connected to all international supply voltages. The input range is from 100 to 240 V~. No range switching is required.

Changing the Fuse The Axopatch 200B uses a 2.0 A, 250 V T 2A 5 x 20 mm fuse.

In the event of fuse failure, disconnect the power cord.

Before changing the fuse investigate the reason for its failure.

To change the fuse:

1) Disconnect the power cord. 2) Use a screwdriver or a similar device to rotate the fuse holder

counterclockwise. 3) Replace the fuse with another fuse of the same rating.

4) Reconnect the power cord.

Probe Temperature The green HEADSTAGE COOLED light comes on when the temperature of the Peltier device that cools the critical components of the headstage drops below 0 °C. This light should go on very soon after power is switched on, and should stay lit during the entire time of operation. It signals that the unit is ready for stable, low-noise recording.

The temperature of the circuitry may be displayed on the panel meter by setting the panel meter switch to TEMP.

The Peltier cooling unit may be turned on or off by the rear-panel HEADSTAGE COOLING ON/OFF switch.

Page 85: Axopatch 200B Manual Final

INSTRUMENT OPERATION • 69

Chapter 6

Zap In order to go from cell-attached patch clamping to whole-cell patch clamping it is necessary to rupture the patch. This is normally done by carefully controlled suction.

Another, easier to apply, technique for rupturing the patch is ZAP. ZAP works by applying a large hyperpolarizing voltage (1.3 VDC) to the patch for a controlled duration. This often causes dielectric breakdown of the membrane.

Suggested Use Apply a repetitive test command (e.g., Seal Test). Start with Duration = 0.5 ms. Press Trigger to ZAP the membrane. Successful zapping is accompanied by an increase in the current noise and by large capacitance-charging current transients in response to the test command. Use the briefest Zap that will rupture the membrane. Too long a Zap could cause the seal resistance to deteriorate.

Page 86: Axopatch 200B Manual Final
Page 87: Axopatch 200B Manual Final

GENERAL INFORMATION • 71

Chapter 7

Chapter 7

Reference Section: General Information

Grounding and Hum A perennial bane of electrophysiology is line-frequency pickup (noise), often referred to as hum. Hum can occur not only at the mains frequency but also at multiples of it.

The Axopatch 200B has inherently low hum levels (less than 0.005 pAp-p). To take advantage of these low levels great care must be taken when incorporating the Axopatch 200B into a complete recording system. The following procedures should be followed.

1) Ground the preparation bath only by directly connecting it to the gold ground connector on the back of the headstage.

2) Place the Axopatch 200B in the rack in a position where it will not absorb radiation from adjacent equipment. A grounded, thick sheet of steel placed between the Axopatch 200B and the radiating equipment can effectively reduce induced hum.

3) Initially make only one connection to the Axopatch 200B — from the SCALED OUTPUT to the oscilloscope output. After verifying that the hum levels are low, start increasing the complexity of the connections one lead at a time. Leads should not be draped near transformers located inside other equipment. In desperate circumstances, the continuity of the shield on an offending coaxial cable can be broken.

Page 88: Axopatch 200B Manual Final

72 • GENERAL INFORMATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

4) Try grounding auxiliary equipment from a ground distribution bus. This bus should be connected to the Axopatch 200B via the yellow banana (4 mm) socket on the rear panel. This socket is connected to the signal ground of the Axopatch 200B (i.e., the outer conductors of all the BNC connectors). The signal ground in the Axopatch 200B is isolated from the chassis and power ground.

5) Experiment. While hum can be explained in theory (e.g., direct pickup, earth loops), in practice the ultimate theory is the end result. Following the rules above is the best start. The final hum level can often be kept to less than 0.1 pAp-p. One technique that should not be used to reduce hum is the delicate placement of cables so that a number of competing hum sources cancel out. Such a procedure is too prone to accidental alteration.

Model Cell The PATCH-1U model cell (Figure 13) can be used to assist with testing and setting up. The pipette is modeled by a 10 MΩ resistor, the cell is modeled by 500 MΩ in parallel with 33 pF (the membrane time constant is 16.5 ms), and the patch is modeled by a 10 GΩ resistor. The pipette capacitance is about 4-6 pF. The charging time constant is approximately 330 µs (10 MΩ x 33 pF).

The PATCH-1U model cell has been made without a switch to change the model between BATH, PATCH and CELL positions. This is because even the best switches have an enormous amount of leakage resistance and capacitance which increases the noise three to five times beyond what you can achieve with a good seal. Instead of switches, three separate plug positions have been provided and you can rotate the model cell into the position required. With this technique the noise contribution of the model cell is still somewhat more than can be achieved with a good seal, but the increase is less than 50%.

The PATCH-1U model cell can conveniently be used in conjunction with the tutorial at the front of this manual.

Page 89: Axopatch 200B Manual Final

GENERAL INFORMATION • 73

Chapter 7

5 pF

PATCH

CELL

33 pF

10M 500M

10G

BATH

EQUIVALENT CIRCUIT

4pF

4 pF

10M

3.3 pF +STRAY

PATCH

CELL

33 pF

500M STRAY

10G

BATH

ACTUAL CIRCUIT

10M

Ω

Ω

Ω

Ω

ΩΩ

Ω

Figure 13. PATCH-1U model cell.

Model Bilayer The MCB-1U model bilayer contains a 10 kΩ resistor that models the pipette in series with a 100 pF capacitor that models the bilayer membrane (Figure 14).

EQUIVALENT CIRCUIT

10 kΩ

ACTUAL CIRCUIT

100 pF

BILAYER10 kΩ 100 pF

BILAYER

4 pF

Figure 14. MCB-1U bilayer model.

Page 90: Axopatch 200B Manual Final

74 • GENERAL INFORMATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Power-Supply Glitches The Axopatch 200B has been designed to minimize the effects of power-supply transients (glitches). Some power-supply glitches do, however, get through. These can cause transients to appear on the voltage and current outputs which may corrupt high-sensitivity recordings.

The only completely effective way to gain immunity from mains glitches is to eliminate them at the source. Most glitches are due to the switching on and off of other equipment and lights on the same power-supply circuit. Precautions to be taken include:

1) Avoid switching equipment and lights on or off while recordings are being made.

2) Water baths, heaters, coolers, etc. should operate from zero-crossing relays.

3) RFI1 filters should be installed in glitch-producing equipment.

In most circumstances, occasional transients on the outputs are inconsequential and, therefore, no precautions have to be taken.

Ten-Turn Potentiometers The ten-turn potentiometers used in the Axopatch 200B are high-quality wirewound types.

An inherent problem of wirewound potentiometers is that the wire elements tend to oxidize. When this happens, significant instrument noise becomes noticeable when the potentiometer is turned. This condition is easily cured. If noise is observed when a potentiometer is turned, the potentiometer manufacturer recommends to rapidly spin the knob 20-30 times between full clockwise and full counterclockwise. This clears the oxide off the element and restores noise-free operation.

1 RFI - Radio Frequency Interference

Page 91: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 75

Chapter 8

Chapter 8

Reference Section: Principles of Operation

Headstages

Principals of Operation Patch-clamp headstages are current-to-voltage (I-V) converters. That is, the voltage output is proportional to the current input. In contrast, conventional microelectrode amplifier headstages are voltage followers in which the voltage output corresponds to the voltage input.

Resistor Feedback The essential parts of a resistive headstage are shown in Figure 15:

+

Vp

BOOST

-+

-

RI f

OFFSETAND

SCALING

HIGH-FREQUENCY

BOOST

I

Vo

CIRCUITf

PROBE

Figure 15. Resistive headstage.

Page 92: Axopatch 200B Manual Final

76 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

For an ideal op amp1 the pipette current is the same as the current through the feedback element (Rf). Since the op amp in the probe acts to keep the voltage at its two inputs equal to each other, we know that the potential at its negative input equals Vp (pipette potential). Thus, the voltage across Rf is Vo - Vp, which is calculated by the differential amplifier in the control box. Subsequent amplifiers are used to scale the gain and remove voltage offsets.

A fundamental problem of this circuit when used for patch clamping is that the output bandwidth of the probe is inherently low. To a first approximation, the bandwidth is set by the product of Rf and the stray capacitance across it. For example, if Rf is 500 MΩ and the stray capacitance is 0.5 pF, the bandwidth is about 600 Hz.

To overcome this limitation, the probe output is passed through a high-frequency boost circuit. The gain of this circuit is proportional to the frequency.

Capacitor Feedback An alternative to measuring current across feedback resistors is to measure current as the rate of change of the voltage across a capacitor. Figure 16 shows the essential parts of an integrating headstage.

Vp

DIFFERENTIATOR

+

-

= 1 pFI f

OFFSETAND

SCALINGI

Vo

INTEGRATOR

fc

+

-

+

-

Figure 16. Capacitive headstage.

1 op amp - operational amplifier

Page 93: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 77

Chapter 8

Nearly ideal capacitors exist whereas high-gigohm resistors found in patch clamp headstages possess intrinsic noise (in excess of thermal noise) and have limited bandwidth due to stray capacitance. The benefits of capacitors are taken advantage of in the PATCH configuration (capacitive feedback) of the Axopatch 200B headstage, which is designed for ultra-low noise single-channel recordings. The headstage measures the integral of the current which is subsequently differentiated to allow measurement of the current itself. Unlike the resistive headstage, the output of the capacitive headstage is fast. No boost circuit is required to increase its high-frequency response. The capacitor mode achieves a substantial reduction of noise and has much better linearity compared to resistive feedback headstages. This noise reduction is particularly significant in the frequency band of interest for single channel recordings (10 Hz - 10 kHz). In integrating headstages, the low-frequency asymptote of the noise depends on the gate current of the headstage input transistor rather than on the thermal noise of the feedback resistor. With the U430 transistors used in the Axopatch 200B, this low frequency noise can be substantially less than that of the 50 GΩ feedback resistor customarily used in resistive headstages. At the same time, the high frequency noise is less because the capacitor lacks the excess high frequency noise associated with gigohm value resistors.

While the integrating headstage is quieter and more linear than resistive feedback headstages there is one disadvantage. The voltage across the feedback capacitor cannot ramp in one direction forever (the rate of change describes the current at the input). At some point the capacitor voltage will approach the supply limits and the integrator must be reset to start again near zero volts. Thus, the current record must be interrupted for 50 µs while the integrator and differentiator reset. The frequency of resets depends on the current that passes through the headstage, with the larger current requiring more frequent resets.

When this reset occurs, a sample and hold circuit maintains the value of the current at the level it had just prior to the reset. It does this for the duration of the reset while the DATA NOT VALID line specifies that the reset is in progress. Following reset, the sample and hold is inactivated, the DATA NOT VALID line goes low and integration of the current again proceeds.

Figure 17 shows the signal pathway for the capacitor-feedback configuration. The output current of the capacitor-feedback headstage is normally connected through a

Page 94: Axopatch 200B Manual Final

78 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

switch to the output pre-filter amplifier, then to the lowpass filter and finally to the post-filter amplifier. The current signal also goes through a lowpass filter to a sample-and-hold amplifier. During reset, the switch shifts to the RESET position. Simultaneously, the sample-and-hold amplifier is switched to the hold mode so that the signal immediately before the reset transient occurs is presented to the output amplifiers.

SCALEDOUTPUT

DATANOT VALID

50 s

SAMPLEANDHOLD

LOW-PASSFILTER

RESET

NORMAL

RESET

PRE-FILTERAMPLIFIER

LOW-PASSFILTER

POST-FILTERAMPLIFIER

I

x1 or x10 x1, 2, 5, 10, 20, 50, 100

HOLD

µ

Figure 17. Signal handling during resets.

The Axopatch 200B contains both a resistive and capacitive feedback elements. The capacitive element is selected using the front panel CONFIG. switch in the PATCH position, while the resistive headstage is engaged when the switch is in the WHOLE CELL position.

The capacitive feedback (PATCH) is recommended for single-channel recording because it offers the lowest noise. In such recordings, the average current is not more than a few pA and so the resets are infrequent.

In whole cell experiments, the average current is generally tens or hundreds of picoamps. Resets would occur very frequently if capacitive feedback was selected. Therefore, we recommend resistive feedback (WHOLE CELL) for these experiments.

If the Probe Output is Slow, How Can Voltage Clamping Be Fast? In resistive headstages (but not capacitive headstages) the current output of the current-to-voltage (I-V) converter in the probe is slow. The high-frequency boost occurs afterwards and cannot influence the events at the electrode. Thus, one might conclude that the voltage clamp of the pipette must also be slow.

Page 95: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 79

Chapter 8

In fact, despite the slow current output of the I-V converter, the voltage clamp of the pipette is rapid. The pipette is connected to the negative input (summing junction) of the op amp. The command potential is connected to the positive input of the op amp. The operation of the op amp in this configuration is to force the potential at the summing junction to rapidly follow the potential at the positive input. If the command potential is a step, the potential at the summing junction, and hence the pipette, is also a step. The current required to achieve this step is passed through the feedback resistor (Rf ) and the associated stray feedback capacitance (CRf ) of the I-V converter. The output of the I-V converter settles to the final value with time constant Rf CRf. This relatively slow settling occurs despite the fact that the step at the summing junction is fast.

In this discussion, we have carefully referred to the fact that it is the "pipette" that is rapidly voltage clamped. The membrane potential is voltage clamped to its final value much more slowly. To a reasonable approximation, the time constant for voltage clamping the membrane is RpCm, where Rp is the pipette resistance and Cm is the membrane capacitance.

What is Clamped During Voltage Clamping? Voltage clamping is the intrinsic mode of operation of a patch clamp headstage. The series combination of the pipette and the patch/cell membrane is voltage clamped — its voltage remains constant at a user-specified value (Vc), assuming that the membrane potential equals the command potential. This is true only if the current causes a negligible voltage drop across the pipette resistance.

Capacitance Compensation

Pipette Capacitance Compensation The FAST and SLOW PIPETTE CAPACITANCE COMPENSATION controls are used to charge the pipette capacitance (Cp) during a voltage step. A simplified circuit of the fast and slow compensation circuitry is shown in the Figure 18.

Page 96: Axopatch 200B Manual Final

80 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

C1 = 1 pF

RfRpOPEN

CIRCUIT

10 Vp

FAST MAG

SLOW MAG

SLOW

FAST

+

Ip

Cp

Vp

I C1

I

+

+

+

τ

+

-

τ

Figure 18. Pipette capacitance compensation circuit.

When the pipette command potential (Vp) changes, current Ip flows into Cp to charge it to the new potential. If no compensation is used, Ip is supplied by the feedback element (Rf) resulting in a large transient signal on the output (I).

By properly setting the fast and slow magnitude and τ controls, a current (IC1) can be induced in capacitor C1 (connected to the headstage input) to exactly equal Ip. In this case no current needs be supplied by Rf, and there is no transient on the output.

The FAST controls compensate that part of Cp that can be represented by a lumped capacitance at the headstage input. This is the major part of Cp. A small amount of Cp can only be represented as a capacitor with a series resistance component. This takes longer to charge to its final value and is compensated by the SLOW controls.

Page 97: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 81

Chapter 8

Whole-Cell Capacitance Compensation The SERIES RESISTANCE and WHOLE-CELL CAP. controls are used to charge the membrane capacitance (Cm). Figure 19 is a simplified circuit of these controls.

Figure 19. Whole-cell capacitance compensation circuit.

Assume that the fast and slow electrode compensation controls have already been set to compensate for Cp. By appropriately adjusting the SERIES RESISTANCE and WHOLE CELL CAP. controls, the current injected through C2 will supply the transient membrane current (Im). These adjustments do not alter the time constant for charging the membrane. Their function is to offload the burden of this task from the feedback resistor, Rf. In many cells, even a small command voltage (Vc) of a few tens of millivolts can require such a large current to charge the membrane that it cannot be supplied by Rf. The headstage output saturates for a few hundred microseconds or a few milliseconds, thus extending the total time necessary to charge the membrane. This saturation problem is eliminated by the appropriate adjustment of the SERIES RESISTANCE and WHOLE CELL CAP. controls. This

Page 98: Axopatch 200B Manual Final

82 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

adjustment is particularly important during series resistance correction (see Series Resistance section in PRINCIPLES OF OPERATION) since it increases the current-passing demands on Rf . By moving the pathway for charging the membrane capacitance from Rf to C2, the SERIES RESISTANCE circuitry can operate without causing the headstage input to saturate. The value of C2 is 1 pF.

The effect of transferring the current-passing burden from Rf to C2 is illustrated in Figure 20.

BeforeWhole-cell Compensation

AfterWhole-cell Compensation

VC

mV

I Rf

C2I

Figure 20. Using the injection capacitor to charge the membrane capacitance.

After perfect whole-cell compensation is applied, the current to charge the membrane capacitor is removed from the IRf trace and only the steady state current remains. All of the transient current appears in the IC2 trace. (The IC2 trace in the figure was recorded using an oscilloscope probe connected to the internal circuitry). The I and Vm outputs on the Axopatch 200B show the IRf and Vc trace illustrated in Figure 20. It is easy to mistakenly think that the time course for charging the membrane is very fast but this is clearly not the case. Use of an independent electrode in the cell would show that the cell charging rate is not affected by these adjustments.

Page 99: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 83

Chapter 8

Absolute Value The absolute value of the membrane capacitance shows on the WHOLE CELL CAP. dial after the whole-cell current transient has been eliminated. This value may be used for estimating the surface area of the cell assuming that the capacitance of 1 cm2 membrane is 1 µF. The setting of this control is available for computer acquisition on the CELL CAPACITANCE output telegraph on the rear panel of the Axopatch 200B.

Limitations The measurement of series resistance (Rs) and cell capacitance (Cm) is accurate only if the membrane resistance (Rm) is significantly greater than Rs.

Series Resistance In patch clamping, the term refers to the pipette resistance plus any other access resistances to the patch or cell.

The SERIES RESISTANCE and WHOLE CELL CAP. controls, along with % PREDICTION and % CORRECTION, are active in WHOLE CELL mode. In PATCH mode, only SERIES RESISTANCE and % CORRECTION are functional.

Absolute Value The absolute range of the series resistance shows on the SERIES RESISTANCE dial after the whole-cell current transient has been eliminated. The range of the SERIES RESISTANCE values is 0 - 100 MΩ.

Correction, Prediction, and Lag As discussed in the previous section (Capacitance Compensation), eliminating the current transient by setting the SERIES RESISTANCE and WHOLE CELL CAP. controls does not improve the speed of clamping the cell.

Page 100: Axopatch 200B Manual Final

84 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Uncompensated series resistance (Rs) has several effects on the fidelity of whole-cell voltage clamp measurements. In the absence of series resistance compensation, these are:

1) Following a step change in command potential, Vc, the actual cell membrane potential, Vm, will respond with an exponential time course (Vm= Vc(1-e-t/τ

s)) with a time constant given by τs = RsCm, where Cm is the cell membrane capacitance. This time constant is 330 µs for the model cell provided with the Axopatch 200B (Rs = 10 MΩ, Cm = 33 pF). This means that the actual membrane potential response to a step voltage command will have a 10-90% risetime of more than 0.7 ms and will not settle to within 1% of its final value until about 1.5 ms after the start of a step command. It is typical that after achieving a whole-cell recording, the access (series) resistance is approximately twice as large as the original pipette resistance; a three-fold or higher resistance increase is not uncommon. Thus, series resistances of 20 MΩ or more may be encountered. With a 100 pF cell a series resistance of 20 MΩ will result in a membrane-charging time constant of 2 ms. Settling of the true membrane potential to within 1% of its final value will require nearly 10 ms after the start of a step command. Note that the uncompensated whole-cell capacity transient has the shape of the derivative of the true membrane potential and both will have the same time constant.

2) Uncompensated series resistance will also cause the membrane potential to deviate from the command potential when ionic membrane current, Im , flows. The magnitude of this error is given by RsIm; e.g., for Rs = 10 MΩ and Im = 2 nA, a 20 mV error will result. In extreme situations in the presence of voltage gated channels, complete loss of control of membrane potential can occur.

3) The first two types of errors associated with series resistance described above are well known to most investigators. The third type of error is less commonly recognized. Series resistance in conjunction with membrane capacitance forms a one-pole RC filter with a corner (-3 dB) frequency given by f=1/2πRsCm for the measurement of membrane currents. This filter will distort currents regardless of their amplitude. For the parameters of the whole-cell model provided with the Axopatch 200B (Rs = 10 MΩ, Cm = 33 pF) this filter restricts true measurement bandwidth to 480 Hz without series resistance compensation. For a situation with Rs = 20 MΩ and Cm = 100 pF (as, for example, may be encountered with isolated cardiac myocytes), the actual bandwidth of current measurement is only about 80 Hz with no compensation for series resistance.

Page 101: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 85

Chapter 8

The Axopatch 200B uses a dual approach for the correction of the above errors associated with series resistance. In this regard, the performance of the Axopatch 200B is unparalleled by any other commercial patch clamp.

The approach taken to whole-cell capacity transient cancellation and series-resistance compensation in the Axopatch 200B involves the following front panel controls:

1) WHOLE-CELL PARAMETERS: WHOLE CELL CAP. potentiometer and its ON/OFF switch, and SERIES RESISTANCE potentiometer. These controls are used to cancel the whole-cell capacity transient. Their action is coordinated with series resistance compensation controls described below. Note that the WHOLE CELL CAP. switch must be ON to cancel whole-cell capacity transients. Turning this switch OFF disables the signal injected through the C2 capacitor in the headstage used to cancel the capacity transient (Figure 19); it also disables the PREDICTION potentiometer (see below). With the switch ON (for now assume that PREDICTION is OFF), the signal injected through C2 capacitor has an amplitude that is determined by the setting of the WHOLE CELL CAP. control and a time constant that is determined by the setting of both the WHOLE CELL CAP. and SERIES RESISTANCE controls. Precise canceling the whole-cell capacity transient with these controls requires a unique setting in each case. These settings are accurate representations of Rs and Cm to within 2-3%. As will be described below, the use of PREDICTION will modify the time course of the signal applied to the capacitor C2.

2) SERIES RESISTANCE COMPENSATION: PREDICTION, CORRECTION, and LAG potentiometers. These controls are used to correct for the errors associated with series resistance.

3) PIPETTE CAPACITANCE COMPENSATION: FAST MAG and FAST τ, and SLOW MAG and SLOW τ. Note that when using series resistance compensation it is important that the fast capacity transient arising from stray and pipette capacitance be adequately canceled.

PREDICTION adds a transient signal to the command potential, speeding the rate at which the true membrane potential will change in response to a step voltage command. It is similar to the idea of SUPERCHARGING introduced by Armstrong and Chow (Armstrong, C.M. and Chow, R.H. (1987) Biophys. J. 52, 1333.). The signal added to the command is derived from the command input and from the setting of the WHOLE-CELL PARAMETERS (WHOLE CELL CAP. and SERIES RESISTANCE control settings). It

Page 102: Axopatch 200B Manual Final

86 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

enables the actual membrane potential to be a faithful replica of the command potential; i.e., the effects of series resistance in distorting the command potential at the cell membrane are removed up to the percentage setting of the control (e.g., a 98% setting means that, in effect, only 2% of the original series resistance remains in terms of command potential). PREDICTION works with the setting of the SERIES RESISTANCE potentiometer and the WHOLE CELL CAP. potentiometer that, in conjunction with the percentage set on the PREDICTION control, determine the magnitude and time constant of the compensating signal added to the command input. Simultaneously, the signal applied to the C2 capacitor used to cancel whole-cell capacity transients is appropriately modified as the percent PREDICTION is increased. (The value of C2 is 5 pF with β = 1 and 50 pF with β = 0.1).

For example, consider a whole-cell voltage clamp situation where Rs = 10 MΩ and Cm= 50 pF and the resting membrane resistance Rm is very large with respect to Rs. Assume that the SERIES RESISTANCE control is set at 10 MΩ and the WHOLE CELL CAP. control is set at 50 pF so that the whole-cell capacity transient is perfectly canceled. If the PREDICTION control is OFF (0%), the signal applied to the headstage 5 pF capacitor (50 pF for β = 0.1) in response to a step voltage command will have a time constant of 500 µs and an amplitude that is appropriate to cancel a whole-cell capacitance transient arising from these parameters (about 10 Vc). With 0% PREDICTION nothing is added to the command potential waveform. In response to a step voltage command the cell membrane potential will change to its new value with a time constant of 500 µs (RsCm). If the % PREDICTION control is advanced to 50%, a transient will be added to the command potential step, Vc, with a time constant of 250 µs and an amplitude equal to that of the command step itself. This will have the effect of changing the cell membrane potential in response to a step command with a time constant given by RsCm (1 - %PREDICTION/100); here this is 250 µs. More formally, the command potential with the PREDICTION signal included, Vcp, can be expressed in terms of the command input, Vc, by:

Vcp= Vc (1+sτs)/(1+sτsrp)

where τs = RsCm, τsrp = RsrpCm, where Rsrp is the residual (uncompensated) series resistance in terms of PREDICTION given by Rsrp = Rs (1 - %PREDICTION/100), and, in the frequency domain s = jw (w is the natural frequency, w = 2πf), or in the time domain s is the operator d/dt. Thus, Vcp = Vc (1 + (Rs /Rsrp -1)e-t/τ

srp).

Moreover, the membrane potential, Vm, is given by Vm = Vcp /(1+sτs) = Vc /(1+sτsrp), or Vm = Vc (1 - e-t/τ

srp). Therefore, advancing the PREDICTION potentiometer setting to 80%

Page 103: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 87

Chapter 8

gives Rsrp of 2 MΩ and τsrp of 100 µs. That is, the speed with which the membrane potential responds to a voltage command is improved 5 fold over that which is achieved with 0% PREDICTION. PREDICTION of 98% gives Rsrp of 200 kΩ and τsrp of 10 µs. The membrane potential will now respond to a step voltage command with a 10-90% risetime of about 22 µs and will settle to within 1% of its final value in less than 50 µs.

Saturation Effects Note that the equation presented above for Vcp (i.e., the command potential plus PREDICTION signal) can be used to define the maximum allowable %PREDICTION for a given size step voltage command (this limit should not be confused with limitations imposed by the stability of the PREDICTION circuit itself). The command plus PREDICTION signal is attenuated at the headstage by a 10:1 voltage divider. Since the circuitry in the Axopatch 200B mainframe will saturate at about ±11-12 V, Vcp is limited in absolute value to about 1.1 to 1.2 V. To be conservative, we will use 1.1 V in the following calculations:

The peak amplitude of Vcp for a step voltage command, Vc, is given by Vc (Rs /Rsrp ) which can be rewritten as Vc /(1 - %PREDICTION/100). So we may state the limitation on Vc as a function of % PREDICTION as:

Vc ≤ 1.1(1 - %PREDICTION/100)

or the limitation on % PREDICTION as a function of Vc as:

% PREDICTION ≤ 100(1 - Vc /1.1)

Thus, for example, if it is known that the maximum command step to be used in a particular experiment is 100 mV, PREDICTION may be set at 91% without fear of saturation of Vcp; this is true regardless of the value of Rs or Cm. In fact, this is a rather conservative estimate since it is derived on the assumption that the signal Vcp will instantly jump to its maximum value following a step voltage command. In fact, due to limitations in the speed of the PREDICTION circuitry, this over-estimates the maximum value of Vcp, particularly when % PREDICTION is large. In actual practice, PREDICTION can typically be set to about 94% for a 100 mV command step. Figure 4 shows maximum % PREDICTION as a function of voltage step.

Page 104: Axopatch 200B Manual Final

88 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

As the PREDICTION potentiometer is advanced the signal applied to the 5 pF capacitor (C2) in the headstage is modified appropriately so that it will continue to cancel the whole-cell capacity transient despite the fact that the speed of this transient has increased. This is simply accomplished by reducing the time constant of this signal as % PREDICTION is increased. If the circuitry worked perfectly, and if the whole-cell capacity transient had been perfectly canceled with 0% PREDICTION, no transient would appear as % PREDICTION is increased up to the maximum allowable values. However, due to the complexity of this circuitry and a variety of non-ideal characteristics, cancellation of whole-cell capacity transients does not remain perfect as % PREDICTION is increased. The small residual transient that emerges can, however, be completely removed by small readjustments of the setting of the WHOLE CELL CAP., SERIES RESISTANCE, FAST MAG and FAST τ controls. A detailed description of the required procedure is provided in the Series Resistance section of the TUTORIAL.

It should be noted that PREDICTION will work for any command waveform, not just steps. This may be useful for capacitance measurements using phase sensitive techniques or lock-in amplifiers.

Although PREDICTION can greatly speed the response time of the true membrane potential with respect to the command potential and, thus, overcome one important effect of series resistance, it does not correct for the effects of series resistance associated with the flow of membrane ionic current (i.e., IR drops and filtering effects described above). This is the role of the CORRECTION potentiometer. CORRECTION feeds back a portion of the measured membrane current; this signal is added to the command potential. The percentage set by the CORRECTION potentiometer refers to the setting of the SERIES RESISTANCE control of WHOLE-CELL PARAMETERS. For example, if the SERIES RESISTANCE control is set at 10 MΩ, a 90% setting of the CORRECTION control means that 9 MΩ of series resistance is compensated; the residual (uncompensated) series resistance in terms of CORRECTION, Rsrc, is 1 MΩ.

The LAG potentiometer is used to determine the time constant of a one-pole RC filter through which the CORRECTION signal is passed prior to being summed with Vc. The -3 dB bandwidth of this filter is given by 1/2πτLAG, where τLAG is the setting (in seconds) of the LAG control. For example, a LAG of 5 µs corresponds to filtering the CORRECTION signal at 32 kHz; 10 µs corresponds to 16 kHz, 20 µs corresponds to 8 kHz, etc. The LAG control is used to ensure stability when large amounts of CORRECTION are used. It is generally good practice to begin using CORRECTION with the LAG control set at

Page 105: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 89

Chapter 8

10-20 µs or more. However, once the desired level of CORRECTION has been achieved, it is usually possible (if desired) to significantly reduce the LAG setting; 5 µs is usually quite adequate for 90% CORRECTION.

Continuing with the example considered above, i.e., a cell with Rs= 10 MΩ and Cm = 50 pF, a 90% CORRECTION setting will reduce voltage errors in the true membrane potential resulting from the flow of ionic current to 10% of the error present with 0% CORRECTION. For example, a 2 nA ionic current would produce a 20 mV error in Vm

with 0% CORRECTION, whereas 90% CORRECTION will reduce this error to only 2 mV. At the same time, the use of CORRECTION will reduce the filtering effect of Rs and Cm on the measured current. With 0% CORRECTION the actual bandwidth of current measurement prior to any output filtering is limited to 1/2πRsCm, which will be about 320 Hz in this example. As % CORRECTION is increased this "filter" moves to 1/2πRsrcCm, so that for 90% CORRECTION the possible bandwidth for current measurement is increased to 3.2 kHz in this example. With 95% CORRECTION the possible bandwidth is increased to 6.4 kHz and with 98% it is further increased to 16 kHz (although the effects of LAG should not be forgotten).

If the capacity transient has been canceled prior to the use of CORRECTION (and for now assume that PREDICTION has already been set at 95%), then, in principle, there is no capacity current to feed back when CORRECTION is utilized. Note that the discussion here of capacity current should be distinguished from the discussions of the ionic current. Therefore, no transient should develop as CORRECTION is advanced. In practice, however, a small transient will emerge as % CORRECTION is increased. Again, this is due to non-ideal characteristics of the circuitry. Procedures for eliminating this transient by minor readjustments of SERIES RESISTANCE, WHOLE CELL CAP., FAST MAG and FAST τ controls are described in detail in the TUTORIAL.

There are many situations in which it will be desirable to have the % PREDICTION and the % CORRECTION controls set at different values. For example, for a 200 mV step command PREDICTION should be limited to about 80% (see Figure 4; however, somewhat higher values can often be used) to avoid saturation. However, it is usually possible to compensate series resistance up to 90 to 95% or more by use of the CORRECTION control. In other patch clamps the issue of saturation would limit the amount of compensation used for ionic currents; this is not true in the Axopatch 200B. On the other hand, in some cases it might be impossible to advance the CORRECTION percentage beyond about 70% without causing instability. Nevertheless, PREDICTION, which is inherently stable up to

Page 106: Axopatch 200B Manual Final

90 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

98% or more, can be set to a value substantially higher than 70% (about 95%), thereby ensuring that the true transmembrane potential changes rapidly in response to the command potential even though a substantial series resistance remains uncompensated in terms of ionic currents.

Oscillations One of the practical problems when using the % CORRECTION function of SERIES RESISTANCE compensation is that there is a great risk of oscillations because the CORRECTION circuitry is a form of positive feedback. The main cause of oscillations is the in-ability of the circuitry to distinguish between current that flows down the pipette and into the cell from current that flows through the stray capacitance of the pipette into the bath. The current that flows through the pipette resistance into the cell is the current that is intended to be compensated. The CORRECTION circuitry also tries to compensate for the current into the pipette capacitance. However, in this case there is no significant series resistance component to compensate, and the CORRECTION circuit will oscillate as soon as the CORRECTION control is advanced.

The tendency to oscillate depends, therefore, on the relative magnitude of the pipette resistance to the pipette capacitance and the degree of compensation of the pipette capacitance. Thus, one should take care that Cm is well compensated as one advances correction.

Using Lag to Prevent Oscillations The tendency to oscillate can be reduced by limiting the bandwidth of the positive-feedback circuit. This is the function of the LAG control.

Limitations Series-resistance compensation is an attempt to electronically reduce the effect of the pipette resistance. Because of practical limitations, it is never perfect. Even if 100% compensation could be used with stability, this would only apply to DC and medium-speed currents. Very fast currents cannot be fully corrected.

For best results, the cell membrane resistance should be many fold higher than the pipette resistance. This is normally the case for cells at rest carrying small drug-activated or synaptic currents. However, during voltage activation the cell

Page 107: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 91

Chapter 8

membrane resistance could fall a hundredfold or more to values similar to or less than the series resistance. In these cases it is probable that:

1) There will be a significant error due to the voltage drop across the pipette. This error is not obvious to the user because the patch clamp controls the combined voltage drop across the pipette and the cell.

2) The setting of the SERIES RESISTANCE and WHOLE CELL CAP. compensation controls will become erroneous because it is based on the time constant to charge the membrane capacitance before the change in membrane resistance occurred. Since this time constant depends on the parallel value of membrane resistance and the pipette series resistance, this error could become substantial. The effect will be a larger transient at voltage levels that activate the fall of membrane resistance.

If the cell input resistance becomes comparable to, or less than, the pipette resistance, the whole-cell patch technique will probably not work. Note that typical patch pipette resistances are in the range of 2 to 10 MΩ. In this case, it would be preferable to use a discontinuous (chopped) single-electrode voltage clamp, such as the Axoclamp, that will give more accurate results.

Pipette Offset The PIPETTE OFFSET control is a ten-turn potentiometer used to add up to ±250 mV to the pipette command potential (Vp ). in order to compensate for the total offset of the liquid-liquid and liquid-metal junction potentials in the electrode and bath, and the offset of the probe input amplifier. It is used at the beginning of each experiment to zero the pipette current (I) when the electrode first touches the solution, and may be used occasionally thereafter to manually adjust for any offset drift. The PIPETTE OFFSET control should not be adjusted after formation of a tight seal between pipette and cell.

A description of how to use the PIPETTE OFFSET is given in the Pipette Offset Adjustment section of the TUTORIAL.

Page 108: Axopatch 200B Manual Final

92 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

In TRACK mode, Vp is continuously adjusted to keep I = 0 (or near zero) even though the pipette offset may be changing at a fairly rapid rate. TRACK mode is most often used during seal formation to stop the I trace from jumping into saturation.

I is severely distorted during TRACK mode; the effect is similar to AC coupling.

Note: The Axopatch 200B should never be left in TRACK mode once data is being recorded.

The rate at which TRACK returns I to zero depends on the pipette resistance. If you are applying a test pulse, you will find that for pipette resistance of 10 MΩ or lower you will see an obvious droop.

After a seal is formed the TRACK circuit becomes very slow in its efforts to keep I = 0.

Leak Subtraction The passive membrane response to a voltage step consists typically of a transient and a steady-state component. It is often helpful to subtract these from the output so that only active responses are observed.

The transient component is eliminated by using the CAPACITANCE COMPENSATION controls as discussed in the Capacitance Compensation section above. The steady-state component is eliminated by the LEAK SUBTRACTION control. This circuit simply subtracts a scaled version of the command voltage from the current. The Adjustment of Leak Subtraction section in the TUTORIAL describes the use of this control.

Since both the CAPACITANCE COMPENSATION and LEAK SUBTRACTION controls are driven by the command voltage, the passive responses remain eliminated for all polarities and magnitudes of command.

Please note that the setting of the LEAK SUBTRACTION control is to be multiplied by β = 1 or 0.1.

LEAK SUBTRACTION is disabled in Current Clamp and Track modes.

Page 109: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 93

Chapter 8

Zap The conventional technique for rupturing a membrane patch to go to whole-cell recording is to apply a pulse of suction. Sometimes this technique damages the cell. Zap provides an alternative method. It applies a pulse of voltage across the patch that ruptures the patch, presumably by causing dielectric breakdown. A timing circuit lets you find a Zap duration that is most likely to achieve the desired result without damaging the seal.

Lipid Bilayers

Experimental Techniques The integrating (PATCH) configuration is uniquely suited for performing experiments on lipid bilayer membranes. This is because the probe can be triggered at the onset of a voltage step so that it can quickly charge the membrane. Also, because the Axopatch 200B is stable while driving purely capacitive loads (up to 1000 pF), one has the ability of minimizing noise by minimizing access resistance.

To realize the full potential of the design, the user should externally initiate a reset at each voltage step applied to the membrane. This is accomplished by applying a positive going TTL pulse to the FORCED RESET INPUT BNC on the rear panel. Because of internal timing delays, the positive going edge of the reset pulse can occur simultaneously with the command step. The reset time is independent of the length of the reset pulse (it is factory set at 50 µs).

If it becomes necessary to blank the output for longer time periods, the BLANK ACTIVATE INPUT BNC should be used. When a HI logic level pulse is applied to this input, the output signal will be blanked for the duration of the pulse. In this circumstance, both FORCED RESET and BLANK ACTIVATE inputs could be tied together. For example, if a 1 ms pulse is applied to both BNCs, a normal reset will be initiated while blanking the output for 1ms.

The following is an explanation of the reset charging process:

Page 110: Axopatch 200B Manual Final

94 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

When the Axopatch 200B probe is in the PATCH configuration, it has a 1 pF capacitor as its feedback element (Figure 21).

INTEGRATOR

1 pF

10 kΩS1

+

-Cm

VSTEP

Vo

Rs

Figure 21. Integrator driving bilayer model.

When a voltage step, Vstep, is applied to the command input, the output, Vo, of the integrator, will attempt to become:

Cm Vo = -Vstep ___ Cf

where Cm is the membrane capacitance in pF and Cf is the feedback capacitor, the value of which is 1pF.

If Vo ≥ 10 V (e.g., 100 mV on 100 pF), a reset of the integrator is initiated internally; an external reset pulse is not needed. In the reset state, switch S1 is closed and the 1 pF capacitor is shunted by the 10 kΩ reset resistor. This all happens within 2 µs from the time the integrator reaches + 10 V or - 10 V.

During the time the integrator is in reset (10 µs), it can pass up to 1 mA (106 nA) of current to charge the membrane. For example, if Cm = 100 pF and Rs = 10 kΩ, the charging time constant (τ) is 1 µs. Therefore, the integrator will be in its reset mode for 10τ. This is long enough for the membrane to essentially reach its final value.

If the voltage step is small enough so that it does not cause a reset (Vo<10 V), it is recommend that a reset signal be applied externally. This is not required in order to

Page 111: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 95

Chapter 8

help the membrane to charge faster, but to keep the succeeding circuitry (differentiator, gain blocks, etc.) from saturating and erroneously indicating a slowly charging membrane.

If you are not sure whether you need to apply an external signal, it is best to just include it. There is no conflict in having a reset initiated by both an external and internal signal.

Noise vs. Access Resistance The Axopatch 200B is quite comfortable with loads of up to 1000 pF of pure capacitance (the maximum bandwidth decreases to about 20 kHz, no overshoot). This can be used to great advantage when doing bilayer experiments; the lower the access resistance, the lower the noise. While there will be some lower limit on the value of the access resistance, it will not be set by stability criteria of the instrument.

In bilayer applications, one is typically working with bandwidths below 1 kHz. In this region, the enCin noise has not yet become the major contributor to the overall noise (where en is the voltage noise of the probe input FETs1 and Cin is the capacitance of the input of the headstage, which is primarily the bilayer membrane capacitance). However, a resistance in series with the bilayer membrane capacitance produces voltage noise just as though the headstage had high intrinsic noise. If this resistance is large enough, then it becomes the major noise contributor.

Figure 22 shows peak-to-peak noise versus series resistance in a 1 kHz bandwidth, for a given bilayer membrane capacitance.

1 FET - Field Effect Transistor

Page 112: Axopatch 200B Manual Final

96 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

BW =1 kHz

SERIES RESISTANCE (ohms)10 k1 k100

0.1

10

300 pF200 pF

100 pF

MEMBRANECAPACITANC

TOTAL NOISE 1

Figure 22. Typical current noise in bilayer experiments.

Current and Voltage Conventions The terminology used in this discussion applies to all amplifiers manufactured by Axon Instruments.

Positive Current The flow of positive ions out of the headstage into the microelectrode and out of the microelectrode tip into the preparation is termed positive current.

Inward Current Current that flows across the membrane, from the outside surface to the inside surface, is termed inward current.

Outward Current Current that flows across the membrane, from the inside surface to the outside surface, is termed outward current.

Page 113: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 97

Chapter 8

Positive Potential The term positive potential means a positive voltage at the headstage input with respect to ground.

Transmembrane Potential The transmembrane potential (Vm) is the potential at the inside of the cell minus the potential at the outside. This term is applied equally to the whole-cell membrane and to membrane patches.

Depolarizing / Hyperpolarizing The resting Vm value of most cells is negative. If a positive current flows into the cell, Vm initially becomes less negative. For example, Vm might shift from an initial resting value of -70 mV to a new value of -20 mV. Since the absolute magnitude of Vm is smaller, the current is said to depolarize the cell (i.e., it reduces the "polarizing" voltage across the membrane). This convention is adhered to even if the current is so large that the absolute magnitude of Vm becomes larger. For example, a current that causes Vm to shift from -70 mV to +90 mV is still said to depolarize the cell. Stated simply, depolarization is a positive shift in Vm. Conversely, hyperpolarization is a negative shift in Vm.

Whole-Cell Voltage and Current Clamp

Depolarizing / Hyperpolarizing Commands In whole-cell voltage clamping, a positive shift in the command voltage causes a positive shift in Vm and is said to be depolarizing. A negative shift in the command voltage causes a negative shift in Vm and is said to be hyperpolarizing.

Transmembrane Potential vs. Command Potential In whole-cell voltage clamp, the command potential controls the voltage at the tip of the intracellular voltage-recording microelectrode. The transmembrane potential is thus equal to the command potential.

Page 114: Axopatch 200B Manual Final

98 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Inward / Outward Current In a cell generating an action potential, depolarization is caused by a flow of positive sodium or calcium ions into the cell. That is, depolarization in this case is caused by an inward current.

During intracellular current clamping, a depolarizing current is a positive current out of the microelectrode tip into the interior of the cell. This current then passes through the membrane out of the cell into the bathing solution. Thus, in intracellular current clamping, a depolarizing (positive) current is an outward current.

An inward sodium current flows in some cells after a depolarizing voltage step. When the cell is voltage clamped, the sodium current is canceled by an equal and opposite current flowing into the headstage via the microelectrode. Thus it is a negative current. When two-electrode voltage clamping was first used in the early 1950's, the investigators chose to call the negative current that they measured a depolarizing current because it corresponded to the depolarizing sodium current. This choice, while based on sound logic, was unfortunate because it means that from the recording instrument's point of view, a negative current is hyperpolarizing in intracellular current-clamp experiments but depolarizing in voltage-clamp experiments.

To prevent confusion, Axon Instruments has decided to always use current and voltage conventions based on the instrument's perspective. That is, the current is always unambiguously defined with respect to the direction of flow into or out of the headstage. Some instrument designers have put switches into the instruments to reverse the current and even the command voltage polarities so that the researcher can switch the polarities depending on the type of experiment. This approach has been rejected by Axon Instruments because of the real danger that if the researcher forgets to move the switch to the preferred position, the data recorded on the computer could be wrongly interpreted. Axon Instruments believes that the data should be recorded unambiguously.

Page 115: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 99

Chapter 8

Patch Clamp By design, the patch-clamp command voltage is positive if it increases the potential inside the micropipette. Whether it is hyperpolarizing or depolarizing depends upon whether the patch is "cell attached", "inside out" or "outside out". The patch-clamp pipette current is positive if it flows from the headstage through the tip of the micropipette into the patch membrane.

Cell-Attached Patch In this mode, the membrane patch is attached to the cell, and the pipette is connected to the outside surface of the membrane. A positive command voltage causes the transmembrane potential to become more negative, therefore it is hyperpolarizing. For example, if the intracellular potential is -70 mV with respect to 0 mV outside, the potential across the patch is also -70 mV. If the potential inside the pipette is then increased from 0 mV to +20 mV, the transmembrane potential of the patch hyperpolarizes from -70 mV to -90 mV.

From the examples it can be seen that the transmembrane patch potential varies inversely with changes in the command potential, shifted by the resting membrane potential (RMP) of the cell. A positive pipette current flows through the pipette, across the patch membrane into the cell. Therefore a positive current is inward.

Inside-Out Patch In this mode of recording, the membrane patch is detached from the cell, with the surface that was originally the inside surface exposed to the bath solution. Now the potential on the inside surface is 0 mV (bath potential). The pipette is still connected to the outside surface of the membrane. A positive command voltage causes the transmembrane potential to become more negative, therefore it is hyperpolarizing. For example, to approximate resting membrane conditions of Vm = -70 mV, the potential inside the pipette must be adjusted to +70 mV. If the potential inside the pipette is increased from +70 mV to +90 mV, the transmembrane potential of the patch hyperpolarizes from -70 mV to -90 mV.

From the example it can be seen that the transmembrane patch potential varies inversely with changes in the command potential. A positive pipette current flows through the pipette, across the patch membrane from the outside surface to the inside surface. Therefore a positive current is inward.

Page 116: Axopatch 200B Manual Final

100 • PRINCIPLES OF OPERATION

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Outside-Out Patch In outside-out patch mode, the membrane patch is detached from the cell in such a way that the surface that was originally the outside surface remains exposed to the bath solution. The potential on the outside surface is 0 mV (bath potential). The pipette interior is connected to what was originally the inside surface of the membrane. A positive command voltage causes the transmembrane potential to become less negative, therefore it is depolarizing. For example, to approximate resting membrane conditions, assuming that Vm = -70 mV, the potential inside the pipette must be adjusted to -70 mV. If the potential inside the pipette is then increased from -70 mV to -50 mV, the transmembrane potential of the patch depolarizes from -70 mV to -50 mV.

The membrane potential changes directly with the command potential. A positive pipette current flows through the pipette, across the patch membrane from the inside surface to the outside surface. Therefore a positive current is outward.

Summary 1) Positive current corresponds to:

Cell-attached patch patch inward current Inside-out patch patch inward current Outside-out patch patch outward current Whole-cell voltage clamp outward membrane current Whole-cell current clamp outward membrane current

2) A positive shift in the command potential is: Cell-attached patch hyperpolarizing Inside-out patch hyperpolarizing Outside-out patch depolarizing Whole-cell voltage clamp depolarizing

Page 117: Axopatch 200B Manual Final

PRINCIPLES OF OPERATION • 101

Chapter 8

3) The correspondence between the command potential (VCMD) and the transmembrane potential (Vm) is:

Cell-attached patch Vm = RMP - Vc Inside-out patch Vm = - Vc Outside-out patch Vm = Vc Whole-cell voltage clamp Vm = Vc

Troubleshooting It has been our experience at Axon Instruments that the majority of troubles reported to us have been caused by faulty equipment connected to our instruments.

If you have a problem, please disconnect all instruments connected to the Axopatch 200B except for the headstage. Ideally, remove the Axopatch 200B from the rack. Work completely through the Functional Checkout. This can often uncover a problem that is in your set up. If the problem persists, please call us for assistance.

Another common problem is caused when dirt or corrosion build up in the headstage connector socket. This can cause unstable current and voltage offsets. It is important to keep the holders and the headstage input clean.

Page 118: Axopatch 200B Manual Final
Page 119: Axopatch 200B Manual Final

SPECIFICATIONS • 103

Chapter 9

Chapter 9

Specifications

Unless otherwise noted: TA = 20 °C, 1 hr warm-up time.

CV 203BU Headstage

Construction All critical components are in a sealed hybrid.

Configuration High-speed low-noise current-to-voltage converter.

Headstage Gain (β) 1 mV/pA in either PATCH or WHOLE CELL β = 1 mode. 0.1 mV/pA in WHOLE CELL β = 0.1 mode.

Feedback Element PATCH 1pF WHOLE CELL β = 1 500 MΩ in parallel with 1pF WHOLE CELL β = 0.1 50 MΩ in parallel with 1pF

Page 120: Axopatch 200B Manual Final

104 • SPECIFICATIONS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Feedback Element Selection FET1 switches in hybrid enable remote selection of either a capacitor (PATCH mode) or a parallel combination of a capacitor and resistor (WHOLE CELL mode).

Tuning (WHOLE CELL mode only) Tuning circuit to idealize response of the feedback resistor is contained in the main instrument. Tuning is automatically bypassed when the capacitive feedback is selected.

Pipette-Capacitance-Compensation Injection Capacitor Value 1 pF

Whole-Cell-Capacitance-Compensation Injection Capacitor Values PATCH mode: none WHOLE CELL mode: β = 1: 5 pF

β = 0.1: 50 pF

Case Case connected to ground. Case jack mates to 2 mm plugs.

Bandwidth Test signal applied via SPEED TEST input; PATCH mode: Internal: 140 kHz Max. External: 100 kHz (limited by output filter)

Capacitive Load Stability 1000 pF, 0 Ω in series

1 FET - Field Effect Transistor

Page 121: Axopatch 200B Manual Final

SPECIFICATIONS • 105

Chapter 9

Maximum Instrument Noise Measured with minimal external noise sources (i.e., radiated line frequency noise, mechanical vibration), 8-pole Bessel filter.

PATCH WHOLE CELL β = 1 β = 0.1 Line freq. Without holder: & harmonics 0.005 pAp-p 0.005 pAp-p 0.005 pAp-p 0.1-100 Hz 0.030 pAp-p 0.50 pAp-p 1.60 pAp-p 0.1-1 kHz 0.015 pA rms 0.25 pA rms 0.75 pA rms 0.1-5 kHz 0.060 pA rms 0.65 pA rms 1.65 pA rms 0.1-10 kHz 0.130 pA rms 1.10 pA rms 3.00 pA rms

With holder: 0.1-10 kHz 0.145 pA rms 1.10 pA rms 3.00 pA rms

22

10-29

-3010

10 Hz 100 HzFREQUENCY

1 kHz

700 G Ω

CURRENT NOISE DENSITY(A / Hz)

10-31

-3210

Figure 23A. Typical low frequency current noise spectrum, patch mode.

Page 122: Axopatch 200B Manual Final

106 • SPECIFICATIONS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

CURRENT NOISE DENSITY(A /Hz)

10-29

10

10

10

-30

-31

-3210 Hz 100 Hz 1 kHz 10 kHz

2

FREQUENCY

Figure 23B. Typical broadband current noise spectrum, patch mode.

0.1

1

0.01

0.001100 Hz 1 kHz 10 kHz

BANDWIDTH

TOTAL NOISE(pA RMS)

Figure 23C. Typical total current noise as a function of bandwidth, patch mode.

Page 123: Axopatch 200B Manual Final

SPECIFICATIONS • 107

Chapter 9

Reset Characteristics (Patch Mode only)

Total reset time 50 µs ± 10%. This includes: integrator reset 10 µs differentiator reset 30 µs other overhead 10 µs

Time between resets (TBR ) For DC currents: TBR = 10/(IDC - IBIAS) where IDC and IBIAS are in pA and TBR is in seconds.

IBIAS is typically 0.3 - 1.0 pA.

For transient currents: A reset will occur if the headstage must deliver more than 10 pC of charge to the membrane. For example, a 60 mV step imposed on a 200 pF bilayer membrane will cause a reset (12 pC of charge needed) whereas a 40 mV step will not (8 pC of charge needed).

Reset transients in current waveform at Scaled Output (typical) 100 Hz bandwidth ± 0.25 pA 1 kHz ± 0.5 pA 10 kHz ± 2 pA

Current Clamp The current clamp mode has two speed settings: I-CLAMP NORMAL and I-CLAMP FAST. I-CLAMP NORMAL is for use with electrode resistances greater than 1 MΩ. I-CLAMP FAST is for use with pipette resistances greater than 10 MΩ. (The speed in I=0 mode is the same as in I-CLAMP NORMAL. In addition, TRACK mode is a slow clamp to zero current.) Note that series resistance compensation remains active in current clamp mode, allowing measurement of pipette resistance and (when Rs is compensated) accurate

Page 124: Axopatch 200B Manual Final

108 • SPECIFICATIONS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

monitoring of cell membrane potential, but the speed setting is still determined by the actual electrode resistance and not only the remaining uncompensated resistance.

The speed of the current clamp depends on the MODE setting (NORMAL or FAST), the time constant of the cell and the pipette resistance.

10 - 90% rise time 10 - 90% rise time (overshoot) (overshoot) Rp Rm Cm I-CLAMP NORMAL I-CLAMP FAST 1 MΩ 0 MΩ 0 pF 15 µs (10%) N/A 1 MΩ 500 MΩ 33 pF 350 µs (0%) N/A 10 MΩ 0 MΩ 0 pF 200 µs (20%) 20 µs (<1%) 10 MΩ 500 MΩ 33 pF 250 µs (10%) 10 µs (<1%) 50 MΩ 500 MΩ 33 pF 500 µs (30%) 150 µs (<1%)

HL-U Pipette Holder HL-U holder mates to threaded Teflon input connector of the CV headstage. Post for suction tubing is 1 mm OD. HL-U holder accepts glass 1.0-1.7 mm OD. Supplied with silver wire. Optional HLR-U right-angle adapter and HLB-U BNC adapter are available.

Capacitance Compensation

Pipette Capacitance Fast τ: 0.2- 2 µs Fast Magnitude: 0 - 10 pF Slow τ: 0.1 - 10 ms Slow Magnitude: 0 - 1 pF

These controls are used to charge pipette capacitance. In I-CLAMP modes they act as a negative capacitance.

Page 125: Axopatch 200B Manual Final

SPECIFICATIONS • 109

Chapter 9

Whole-Cell Capacitance β = 1: 0.3 - 100 pF β = 0.1: 3 - 1000 pF

Series Resistance 0 - 100 MΩ

These controls are used to charge membrane capacitance in whole-cell voltage clamp. For PATCH mode, whole-cell capacitance is not operative. In I-CLAMP modes only the SERIES RESISTANCE control is operative. The whole-cell capacitance control places an analog voltage proportional to setting on CELL CAPACITANCE TELEGRAPH OUTPUT. The range of values for the SERIES RESISTANCE compensation control is 0 - 100 MΩ .

Series Resistance Compensation

% Prediction OFF, 0-100 %. Acts with WHOLE-CELL PARAMETERS to speed up charging of the membrane. Maximum achievable % PREDICTION is limited by magnitude of voltage step (see Figure 4).

% Correction OFF, 0-100%. Acts with SERIES RESISTANCE setting to reduce series resistance errors and speed up response to ionic currents.

Lag 1-100 µs. Cuts high-frequency response of series-resistance correction circuit to enable a higher CORRECTION setting.

Page 126: Axopatch 200B Manual Final

110 • SPECIFICATIONS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Capacitance Dithering A TTL input at the rear-panel capacitance dither BNC input activates the dither. This effectively increases the measured membrane capacitance by 0.1 % of the full-scale range (thus, by 0.1 pF for β = 1 or 1 pF for β = 0.1) by transiently reducing the Whole-Cell Capacitance value by that amount.

Mode

V-Clamp Pipette voltage is clamped.

I-Clamp Normal or Fast Pipette current is clamped to command current from Holding Command knob or external input.

NORMAL mode stable for electrode resistances greater than 1 MΩ.

FAST mode stable for electrode resistances greater than 10 MΩ.

SERIES RESISTANCE control is active.

I=0 Slow I-CLAMP to zero current.

Track Slow I-CLAMP to zero current used to correct pipette offset.

Selected mode sets analog voltage on MODE TELEGRAPH OUTPUT.

Page 127: Axopatch 200B Manual Final

SPECIFICATIONS • 111

Chapter 9

Zap

Amplitude +1.3 VDC at pipette for chosen duration.

Duration 0.5-50 ms or Manual. Triggered by front-panel pushbutton. In Manual position ZAP amplitude is maintained as long as pushbutton is depressed.

Command Potentials

Seal Test 5 mV command at line frequency (voltage clamp mode)

50 pA (current clamp mode, β = 1) or 500 pA (current clamp, β = 0.1).

External Commands Two separate BNC inputs, one front switched, one rear switched.

Sensitivity FRONT-SWITCHED: 20 mV/V in V-CLAMP, 2 ÷ β nA/V in I-CLAMP,

disabled in TRACK and I=0.

REAR-SWITCHED: 100 mV/V in V-CLAMP, 2 ÷ β nA/V in I-CLAMP, disabled in TRACK and I=0.

Input impedance 10 kΩ. Inputs may be connected in parallel to increase sensitivity.

Page 128: Axopatch 200B Manual Final

112 • SPECIFICATIONS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Holding Command Ten-turn potentiometer with dial. Polarity switch. Toggle for x1 or x5 to determine range. Value can be previewed on meter.

V-CLAMP mode ±200 mV (± 1 V with toggle x5).

I-CLAMP modes ±2 nA for β = 1, ±20 nA for β = 0.1 (toggle x1)

±10 nA for β = 1, ±100 nA for β = 0.1 (toggle x5)

Disabled in TRACK and I=0 mode.

Pipette Offset

Manual ±250 mV. Ten-turn control with uncalibrated dial.

Track, I=0 ±200 mV. Nulling potential automatically adjusts to main-tain zero pipette current.

RMS Noise 3.5 digit meter displays rms current noise in pA. Measurement bandwidth is 30 Hz to 5 kHz. Upper -3 dB frequency is set by 4-pole Butterworth filter.

Inputs

Forced Reset Positive edge triggered. Initiates a reset of the integrator; has no control over the duration of reset.

Page 129: Axopatch 200B Manual Final

SPECIFICATIONS • 113

Chapter 9

Blank Activate Causes SCALED OUTPUT and I OUTPUT to hold their initial value for the duration of the blanking pulse. Does not affect 10Vm output.

Speed Test Injects current into headstage input through a 1 pF capacitor. Injected current waveform is the derivative of the voltage waveform applied at SPEED TEST input. For example, a 100 Hz 10 Vp-p triangle wave will inject a 1 nAp-p square wave into the headstage input.

Signal Outputs

Scaled Output Scaled and filtered by output control settings. Sample and hold pedestal compensation. Output is I (αβ mV/pA) when in V-CLAMP or TRACK (I=0) mode. Output is Vm (α mV/mV) when in I-CLAMP mode. BNCs on front and rear panels are identical.

I Pipette current. Rear switched gain of either βmV/pA or 100 βmV/pA; fixed filter: 10 kHz 3-pole Bessel. Output does not benefit from sample and hold pedestal compensation.

10Vm Membrane potential at x10 gain. Junction potentials removed.

Output Controls

Output Gain(α) 10 values from 0.5 - 500. Affects SCALED OUTPUT only. Selected value sets analog voltage on GAIN TELEGRAPH OUTPUT for reading by computer.

Page 130: Axopatch 200B Manual Final

114 • SPECIFICATIONS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Low Pass Bessel Filter 4-pole lowpass Bessel filter with five settings: 1, 2, 5, 10 and 100 kHz. Selected value sets an analog voltage on FREQUENCY TELEGRAPH OUTPUT.

Leak Subtraction Causes a signal proportional to the command to be subtracted from current record. Range: 100 βMΩ to ∞.

Telegraph Outputs

Gain Takes α and β gain factors into account

I (mV/pA): 0.051 0.11 0.21 0.5 1 2 5 10 20 50 1002 2002 5002

Vm (mV/mV): N/A N/A N/A 0.5 1 2 5 10 20 50 100 200 500

Telegraph Output (V): 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5

Frequency Filter Setting (kHz): 1 2 5 10 100 Telegraph Output (V): 2 4 6 8 10

Mode I-CLAMP I-CLAMP Mode: TRACK V-CLAMP I = 0 NORMAL FAST Scaled Output: I I Vm Vm Vm Telegraph (volts): 4 6 3 2 1

Cell Capacitance Telegraph Output: 0 to +10 V, proportional to setting 0-100 pF (for β = 1; 0-1000 pF for

β = 0.1 ) when WHOLE CELL CAP. switch is in the ON position. 1 Applicable for β = 0.1 only 2 Applicable for β = 1 only

Page 131: Axopatch 200B Manual Final

SPECIFICATIONS • 115

Chapter 9

0 to -10 V, when WHOLE CELL CAP. switch is in the OFF position.

Data Not Valid Output goes High during a reset in PATCH mode or for the duration of a BLANK ACTIVATE pulse in either PATCH or WHOLE CELL mode.

Panel Meter 3.5 digit meter displays TRACK potential (VTRACK) in mV, membrane potential (Vm) in mV, current noise (IRMS) in pA rms, membrane current (I) in pA or nA, HOLDING COMMAND (VHOLD/IHOLD) in mV or nA, or capacitor-feedback circuitry temperature in degrees Celsius (TEMP). Meter has autoranging feature when membrane current (I) or holding command (VHOLD/IHOLD) is chosen.

Grounding Signal ground is isolated from chassis and power ground. Signal ground is available on rear panel.

Control Inputs Above 3 V accepted as logic High. Below 2 V accepted as logic Low. Inputs protected to ±15 V.

Models Unit is supplied with two model assemblies, the PATCH-1U and the MCB-1U.

PATCH-1U Model Cell Emulates three experimental conditions:

Page 132: Axopatch 200B Manual Final

116 • SPECIFICATIONS

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

BATH: 10 MΩ electrode resistor to ground. 4 pF pipette capacitance. CELL: 10 MΩ electrode resistor connected to a 500 MΩ// 33 pF cell. 4 pF pipette

capacitance. PATCH: 10 MΩ electrode connected to a 10 GΩ patch. 5 pF pipette capacitance.

MCB-1U Model Bilayer Emulates a bilayer membrane. 10 kΩ resistor in series with a 100 pF capacitor.

Headstage Dimensions Case is 0.7" x 0.75" x 4.2" (17.8 x 19 x 106.7 mm). A removable mounting plate 2.5 x 2.0 x 0.25" is supplied. Cable length is 10 feet (3 m).

Cabinet Dimensions 3.5" (89 mm) high, 19" (483 mm) wide, 12.5" (317 mm) deep. Mounts in standard 19" rack. Handles included. Net weight 11.5 lbs (5.1 kg).

Supply Requirements

Line Voltage: 100 to 240 V~. Universal voltage input.

Line Frequency: 50-60 Hz.

Power 30 W

Fuse 0.5 A slow. 5 x 20 mm.

Line Filter: RFI filter is included.

Line cord: Shielded line cord is provided.

Page 133: Axopatch 200B Manual Final

SPECIFICATIONS • 117

Chapter 9

Accessories Provided Theory and Operation Manual HL-U Pipette Holder Spare fuse PATCH-1U Model Cell MCB-1U Model Bilayer DR-1 Series Resistance Dither Unit

Page 134: Axopatch 200B Manual Final
Page 135: Axopatch 200B Manual Final

REFERENCES • 119

Chapter 10

Chapter 10

References

Fidler, N., and Fernandez, J.M. Phase tracking: an improved phase detection technique for cell membrane capacitance measurements. Biophysical Journal 56:1153-1162, 1989.

Hamill, O.P., Marty, A., Neher, E., Sakmann, B., Sigworth, F.J. Improved patch-clamp techniques for high-resolution current recording from cells and cell-free membrane patches. Pflügers Archiv, 391, 85-100, 1981.

Hille, B. Ionic Channels of Excitable Membranes. Sunderland, MA; Sinauer Associates Inc., 1992.

Kawagoe, K.T., Zimmerman, J.B. and Wightman, R.M. Principles of voltammetry and microelectrode surface states. Journal of Neuroscience Methods 48:225-240, 1993.

Joshi, C. and Fernandez, J.M. Capacitance Measurements: An analysis of the Phase Detector Technique Used to Study Exocytosis and Endocytosis. Biophysical Journal 53:885-892, 1988.

Purves, R.D. Microelectrode Methods for Intracellular Recording and Ionophoresis. London: Academic press, 1981.

Rae, J.L., Levis, R.A. Patch voltage clamp of lens epithelial cells: Theory and practice. Molec. Physiol, 6, 115-162, 1984.

Sakmann, B., Neher, E. Single-Channel Recording. New York, N.Y.: Plenum Press, 1983.

Page 136: Axopatch 200B Manual Final
Page 137: Axopatch 200B Manual Final

TUNING THE HEADSTAGE • 121

Chapter 11

Chapter 11

Headstage Tuning Procedure

This procedure should be carried out if the user has done any of the following:

1) Purchased an additional headstage and needs to match it to the main unit.

2) Has found the step response of the WHOLE CELL configuration is unacceptable. This should be verified by applying a high quality triangle voltage waveform to the SPEED TEST BNC. (A 10 Vp-p 100 Hz triangle wave will inject a 1 nAp-p square wave into the headstage input).

3) Has determined that the reset transients in PATCH mode are more than twice the typical values (see SPECIFICATIONS). This should be verified by warming up the unit for at least one hour and connecting the PATCH-1U model cell (PATCH position) as a load. Set the HOLDING COMMAND to + or - 200 mV to induce resets (toggle set to x1). Monitor the SCALED OUTPUT BNC and trigger your scope from the DATA NOT VALID BNC.

Equipment Required Oscilloscope High quality triangle waveform generator 100 Hz, 4-pole (or greater) Bessel filter (optional) PATCH-1U model cell Various BNC cables Large piece of aluminum foil for shielding the headstage

Page 138: Axopatch 200B Manual Final

122 • HEADSTAGE TUNING PROCEDURE

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Gain and Offset Trim These trims are optional if you are just trimming your present headstage and not changing headstages.

Set CONFIG. on PATCH.

Connect the headstage to PATCH-1U model cell in the BATH position.

Carefully shield the headstage and model cell with foil. Ground foil to brass pin on headstage using suitable jumper.

Select I on FRONT PANEL METER.

Use JUNCTION NULL to zero the current.

Set HOLDING COMMAND to 100 mV (toggle set to x1).

Switch HOLDING COMMAND between + and -.

Trim RT4 so that the difference on the meter is 20 nA (nominally ±10 nA if the current has been exactly zeroed).

Repeat procedure with CONFIG. set to WHOLE CELL β = 1 and trim RT5.

Repeat procedure with CONFIG. set to WHOLE CELL β = 0.1 and trim RT33.

Remove PATCH-1U model cell.

Use appropriate trim pots on back panel to zero the meter for both PATCH and WHOLE CELL CONFIG.

Set CONFIG. to WHOLE CELL β = 1.

Set HOLDING COMMAND to 200 mV.

Switch HOLDING COMMAND between + and - .

Trim RT14 until meter reads zero for both polarities.

Page 139: Axopatch 200B Manual Final

TUNING THE HEADSTAGE • 123

Chapter 11

Frequency Tuning (Whole Cell Config.) Set CONFIG. to WHOLE CELL β = 1.

Set OUTPUT GAIN to x1.

Set FILTER to 100 kHz.

Connect scope to SCALED OUTPUT BNC.

Trigger scope from waveform generator.

Set scope to 0.2 V/div, 1 ms/div.

Apply a 10 Vp-p 100 Hz triangle wave to the SPEED TEST BNC.

On the scope you should see about a 1 Vp-p square wave.

If the square wave response is unacceptable in terms of rise time or linearity continue with procedure, otherwise move on to RESET TRANSIENT COMPENSATION.

Set FILTER to 10 kHz.

Adjust RT31 until you get a response which looks like either "trace a" of Figure 24.

Figure 24. Frequency tuning of whole cell configuration.

Page 140: Axopatch 200B Manual Final

124 • HEADSTAGE TUNING PROCEDURE

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Set scope to 10 µs/div.

Set FILTER to 100 kHz.

Trim RT29 for fastest rise time without overshoot. The 10 - 90% rise time should be about 6 - 8 µs.

Set CONFIG. to WHOLE CELL β = 0.1.

Set OUTPUT GAIN to x10

Repeat above steps for β = 0.1, but use RT32 and RT35 for tuning.

Reset Transient Compensation This is a fairly complicated procedure. The results rely on the patience and skill of the person who is doing the adjustments. Please confirm that the reset transients are indeed unacceptable before proceeding. Because of the consistency between units, you may not have to do these adjustments even if you are changing headstages.

Make sure the unit has been on continually for at least one hour.

Forced Reset Remove PATCH-1U model cell.

CONFIG. to PATCH.

OUTPUT GAIN to x100.

FILTER to 5 kHz.

Scope to 0.2 V/div, 0.5 ms/div.

Trigger scope from DATA NOT VALID BNC.

Connect pulse generator to FORCED RESET BNC.

Set pulse generator +5 V ( > 100 µs duration), 100 Hz.

Turn switch 1 of SW8 (near center of PC board) to OFF.

Page 141: Axopatch 200B Manual Final

TUNING THE HEADSTAGE • 125

Chapter 11

Find reset transient on scope (see Figure 25a).

Turn switch 1 of SW8 to ON.

Trim RT22 (TAIL TAU) and RT23 (TAIL MAG) to minimize transient

Internal Reset Remove pulse generator from Axopatch 200B.

Connect PATCH-1U model cell in PATCH position to headstage.

Set FILTER to 10 kHz.

Adjust PIPETTE OFFSET for zero current.

Set HOLDING COMMAND to 200 mV (toggle set to x1).

Switch HOLDING COMMAND between + and -.

The INTEGRATOR RESET LED should be flashing about twice a second at either + or -200 mV.

If there is an obvious time difference between + and -, trim RT24 (CHG INJ) to balance it out.

Set switches 3,4 and 5 of SW8 to OFF (internal reset compensation).

Find reset transient on scope. It should have about a 2-7 pA amplitude with an apparent decay time constant of roughly 0.5 ms (see Figure 25b). The transient should be riding on a DC level of + or -20 pA and should change polarity with DC level.

Set scope to AC coupling 0.5 V/div, 1 ms/div.

Set switch 3 of SW8 to ON.

Trim RT18 (DIAB-1 MAG) and RT19 (DIAB-1 TAU) to eliminate fastest portion of transient as in Figure 25c.

Switch HOLDING COMMAND between + and - and find a the best possible compromise. There will be some amount of unavoidable asymmetry.

Set FILTER to 1 kHz.

Page 142: Axopatch 200B Manual Final

126 • HEADSTAGE TUNING PROCEDURE

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Set scope to 0.1 V/div, 5 ms/div.

Set switch 4 of SW8 to ON.

Trim RT20 (DIAB-2 MAG) and RT21 (DIAB-2 TAU) as above to remove the most significant portion of the reset transient. See Figure 25d. It may be necessary to re-trim RT18 and RT19 to optimize response.

If a slow tail (τ > 5 ms) still remains, it can be compensated with RT25 (DIAB-3 MAG) and RT26 (DIAB-3 TAU). For the most accurate adjustment of these trims, a 100 Hz 4-pole (or greater) Bessel filter should be connected between the SCALED OUTPUT and the scope. The next steps assume the 100 Hz filter is installed.

Set scope to 20 mV/div, 50 ms/div.

Set switch 5 of SW8 to ON.

Trim RT25 and RT26 to minimize slowest transient. As above, some re-trimming of RT20 and RT21 may be necessary to achieve optimum compensation. See Figure 25e.

Figure 25. Reset transient compensation.

For clarity, only one reset polarity is shown (positive DC current).

Page 143: Axopatch 200B Manual Final

INDEX • 127

Chapter 11

INDEX

Access Resistance, 83, 95 Acoustic Pick-up, 56 Bilayers, 93

Noise, 94 Blank Activate, 62 Bridge Mode, 49 Capacitance Compensation

Current Clamp, 51 Electrode, 14 Electrode, Fast, 79 Electrode, Slow, 79 Whole-Cell, 16

Capacitance Measurements, 49 Cell Capacitance Telegraph, 64 Chloriding, Silver, 60 Cleaning

Headstage, 55 Holders, 59

Command Potentials, 50 Computer Interfacing, 41 Correction, 83 Current Clamp

Capacitance Compensation, 51 Initiation, 51 Model Cell, 28 Speed, 53 Whole-cell, 51

Current Convention, 96 Current Output, 68 Current, Display, 67 Data Not Valid Output, 64 Electrochemistry, 54 External Command, 62, 72

Filter

See Output, 63 Forced Reset, 62 Frequency Telegraph, 64 Fuse Changing, 68 Gain

See Output, 63 Gain Telegraph, 64 Glass

Dimensions, 61 Types, 46

Ground Connections, 4, 6, 55 Grounding, 71 Headstage

Adapters, 61 Capacitor Feedback, 76 Cleaning, 55 Frequency Boosting, 55 Gain, 11 Integrating, 76 Mounting, 55 Offset, 55 Optical Pick-up, 56 Reset, 77 Resistor Feedback, 75 Resistor Values, 103 Static, 55

Holder, Pipette Cleaning, 59 Glass Dimensions, 61 Use, 59

Holding Potential, 50 Display, 67

Hum, 71

Page 144: Axopatch 200B Manual Final

128 • INDEX

Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.

Input BNCs, 61 Lag, 83 Leak Subtraction, 92 Membrane Potential, 51

Display, 67 Output, 66

Meter see Panel Meter, 67

Mode, 66 I-CLAMP, 28 Track, 13 V-CLAMP, 13

Model Bilayer, 73 Model Cell, 11, 72 Models

Model Bilayer, 73 Model Cell, 72

Noise, 43

rms, 5 vs. Access Resistance, 94 vs. Bandwidth, 105

Offset

Nulling, 67 Pipette, 12, 67, 91

Optical Pick-up, 56 Oscillations, 39, 90 Output

Filter, 63 Gain, 63

Output BNCs, 64 Overload Light, 64 Panel Meter, 67 Percentage Compensation

See Series Resistance, 83 Perforated Patch, 36 Pipette, 55

Chloriding, 60 Cleaning, 59 Filling, 59

Glass, 46 Insertion, 59 Nulling, 67 Offset, 12, 91

potentiometers, 74 Power Supply

Glitches, 74 Voltage Selection, 68

Prediction, 83 Reset, 77

Forced Reset, 62 RMS Noise

Display, 5, 67 Scaled Output, 66 Seal Test, 5, 51 Series Resistance

Compensation, 19, 83 Correction, 19, 83 Fast, 19 Lag, 19, 83 Oscillations, 39

Using Lag, 90 Prediction, 19, 83 Slow, 19

Single-Channel Recording Model Cell, 11 Real Cell, 30

Specifications, 103 Speed

Voltage Clamping, 78 Speed Test, 62, 72 Static, 56 Track

Display, 67 Trouble Shooting, 101 Voltage Convention, 96 Whole-Cell Recording

Model Cell, 15 Real Cell, 35

Zap, 69, 93

Page 145: Axopatch 200B Manual Final

WARRANTY • 129

Warranty

WARRANTY

We warrant every Axopatch 200B and every headstage to be free from defects in material and workmanship under normal use and service. For 12 months from the date of receipt, we will repair or replace without cost to the customer, any of these products that are defective and that are returned to our factory properly packaged with transportation charges prepaid. We will pay for the return shipping of the product to the customer.

Before returning products to our factory the customer must contact us to obtain a Return Merchandise Authorization number (RMA) and shipping instructions. Failure to do so will cause long delays and additional expense to customer.

This warranty shall not apply to damage resulting from improper use, improper care, improper modification, connection to incompatible equipment, or to products which have been modified or integrated with other equipment in such a way as to increase the time or difficulty of servicing the product.

This warranty is in lieu of all other warranties, expressed or implied.

Page 146: Axopatch 200B Manual Final
Page 147: Axopatch 200B Manual Final

WARNING • 131

Warning

WARNING

Shipping the Axopatch 200B The Axopatch 200B is a solidly built instrument designed to survive shipping around the world. However, in order to avoid damage during shipping, the Axopatch 200B must be properly packaged.

In general, the best way to package the Axopatch 200B is in the original factory carton. If this is no longer available, we recommend that you carefully wrap the Axopatch 200B in at least three inches (75 mm) of foam or "bubble-pack" sheeting. The wrapped Axopatch 200B should then be placed in a sturdy cardboard carton. Mark the outside of the box with the word FRAGILE and an arrow showing which way is up.

We do not recommend using loose foam pellets to protect the Axopatch 200B. If the carton is dropped by the shipper, there is a good chance that the Axopatch 200B will shift within the loose pellet packaging and be damaged.

If you need to ship your Axopatch 200B to another location, or back to the factory, and you do not have a means to adequately package it, Axon Instruments can ship the proper packaging material to you for a small fee. This may seem like an expense you would like to avoid, but it is inexpensive compared to the cost of repairing an instrument that has sustained shipping damage.

It is your responsibility to package the instrument properly before shipping. If it is not, and it is damaged, the shipper will not honor your claim for compensation.

Page 148: Axopatch 200B Manual Final
Page 149: Axopatch 200B Manual Final

CIRCUIT DIAGRAMS REQUEST FORM • 133

Circuit Diagrams Request Form

Circuit Diagrams Request Form All the information that you require for operation of the Axopatch 200B is included in the operator's manual. In the normal course of events, the Axopatch 200B does not require any routine maintenance. If, for some reason, the headstage is changed, the Axopatch 200B must be recalibrated. In anticipation of this, the recalibration procedures are described in the operator's manual, and circuit diagrams are not required.

Should you need the circuit diagrams for the Axopatch 200B, Axon Instruments will be pleased to supply them to you. However, we caution you that the Axopatch 200B is a sophisticated instrument and that service should only be undertaken by talented electronics experts.

To request a copy of the circuit diagrams and the parts lists, please complete the form at the bottom of this page and mail it to:

Axon Instruments, Inc. Sales Department 3280 Whipple Road Union City, CA 94587 USA

This form must be completed in full and signed. Telephone orders will not be accepted.

Name of registered owner: ___________________________________________________________________

Department: ______________________________________________________________________________

University/Institute: ________________________________________________________________________

Street address: _____________________________________________________________________________

City: ____________________________ State: ______ Zip Code: __________ Country: _________________

Telephone: ________________________________________ Fax: __________________________________

Model: Axopatch 200B Serial number: ________________

Declaration Please send me the circuit diagrams and parts lists for the Axopatch 200B. I agree that I will only use the circuit diagrams and parts lists for service of the Axopatch 200B. I will not use them to create equivalent or competing products. If I transfer the circuit diagrams or copies thereof to someone who is assisting in the service of the Axopatch 200B, I will ask them to make the same undertaking that I am declaring herein.

Signature:_______________________________________________ Date:________________________

Name:__________________________________________________ Title:________________________

Page 150: Axopatch 200B Manual Final
Page 151: Axopatch 200B Manual Final

CIRCUIT DIAGRAMS REQUEST FORM • 135

Circuit Diagrams Request Form

Please fold out so that you may refer to this page while reading the manual.

Page 152: Axopatch 200B Manual Final
Page 153: Axopatch 200B Manual Final

Circuit Diagrams Request Form