ANNEX II: TECHNICAL SPECIFICATIONS General Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: a) User manual and handbooks for maintenance in English and Bulgarian. b) The equipment must be capable of withstanding power failures and equipments should not be damaged due to 'trip outs'. The normal voltage and frequency condition available at the site are as follows: 230 V and 50 Hz unless otherwise stated. All electrical devices must be equipped with the necessary power cords and connection cables. The power cords for class I single-phase equipment must be equipped with “Schuko” plugs. Class II equipment power cords must have Euro-plugs. c) Conditions of use for the computer equipment: ambient temperature (10 – 40 ºC); relative humidity (30 – 85%). All instruments for outdoor use must have rugged and water- resistant design, with water proof cases, and operating temperature range: –10 ºC to +50 ºC. d) Service must be available for all items. The reliability and maintenance response times are considered very important. Within the warranty period repair service must be available within 48 hours of notification, or if repair not possible within one month replacement of the item. The tenderer must indicate in his tender how this will be achieved. e) Tenderers should include in their tender, provision for tools and initial stock of maintenance spares as are essential for proper operation and maintenance of the equipment. Full particulars of spare parts should be provided separately. In addition to above, estimated requirements of spares consumption per annum should be indicated. However, as this is for information only, should not be included in the Financial Offer, and respectively will not be subject of evaluation. The successful tenderer shall provide a statement giving full warranty that spare parts for the system would be available for a minimum reasonable period of 5 years after acceptance of the system/equipment.

Upload: others

Post on 09-Jul-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

ANNEX II: TECHNICAL SPECIFICATIONS

General Requirements

The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise:

a) User manual and handbooks for maintenance in English and Bulgarian.b) The equipment must be capable of withstanding power failures and equipments should not

be damaged due to 'trip outs'. The normal voltage and frequency condition available at the site are as follows: 230 V and 50 Hz unless otherwise stated. All electrical devices must be equipped with the necessary power cords and connection cables. The power cords for class I single-phase equipment must be equipped with “Schuko” plugs. Class II equipment power cords must have Euro-plugs.

c) Conditions of use for the computer equipment: ambient temperature (10 – 40 ºC); relative humidity (30 – 85%). All instruments for outdoor use must have rugged and water-resistant design, with water proof cases, and operating temperature range: –10 ºC to +50 ºC.

d) Service must be available for all items. The reliability and maintenance response times are considered very important. Within the warranty period repair service must be available within 48 hours of notification, or if repair not possible within one month replacement of the item. The tenderer must indicate in his tender how this will be achieved.

e) Tenderers should include in their tender, provision for tools and initial stock of maintenance spares as are essential for proper operation and maintenance of the equipment. Full particulars of spare parts should be provided separately. In addition to above, estimated requirements of spares consumption per annum should be indicated. However, as this is for information only, should not be included in the Financial Offer, and respectively will not be subject of evaluation. The successful tenderer shall provide a statement giving full warranty that spare parts for the system would be available for a minimum reasonable period of 5 years after acceptance of the system/equipment.

f) The successful tenderer shall be responsible for erection (where ever required) and installation of the equipment at destination site and for making it fully operational. Specific preparatory works for installation and protection from barbarism of monitoring equipment (wells – items 1, 2, 12, 13, 14) shall be implemented according to the specification for preparatory works given below. The supplier must provide complete details regarding space and all infrastructure needs of the equipment which should be arranged before the arrival of the equipment to ensure its early installation and smooth operation thereafter (the successful bidders may offer his advice an assistance to the beneficiary in the preparation of the site).

g) The successful tenderer shall supply equipment that must be clearly identified and must visibly carry the EU logo and the mention “Provided with the support of the EU” in the operational language of the EU programme and in Bulgarian language.

h) The tenderers shall make provision of imparting training to our Researchers/maintenance Engineers on operation and use of equipment, its accessories and software for the intended applications and trouble shooting repair and maintenance. These services must be included as an integral component of the proposal.

The following notes have to be taken into due account:o The Tenderer’s offer must be accompanied by technical brochures showing in

details type, dimensions, size and description of the offered material and/or

Page 2: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

technical documentation proving the required performance and parameters in order to verify the compliance with the technical specifications.

o The specifications given are minimum requirements based on the item’s projected function, required performance or intended purpose. Characteristics deviating from exact figures or values given in the specifications (unless exceeding the requirements), due to particular proprietary design of the actually offered product, are acceptable, as long as the capacity is at least equivalent to the given specification and permissible under the standards quoted in the Technical Specifications.

o Any specifications that might eventually refer to a particular product or manufacturer, in particular to types, models and brand names, are to be understood as “or equivalent”. Where equivalency shall be subject to technical evaluation, the respective documentation giving evidence of equivalency, and - if appropriate - an assessment by an independent recognized party shall be provided with the offer.

Page 3: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

ANNEX II: TECHNICAL SPECIFICATIONSContract title: Project BG-RO 2004/016-783.01.02 Integrated Management of Transboundary Groundwater between Bulgaria and Romania in Dobrudja/Dobrogea AreaPublication reference: EuropeAid/122864/D/SUP/BG

Item Number

Specifications Quantity Evaluation Committee’s notes

1 Wells: water level and temperature meters (set). Contact level and TºC gauge, average cable length 250 m Measuring tape - white polyethylene tape (two wires), meter graduation in red, centimetre marks and decimetre graduation in black on white tape.Probe body - brass galvanized,Tape drum - robust, plasticBattery power supply Working temperature- 30 °C up to + 75 °CTemperature measurement- LC-display 3 ½ digits- resolution 0,1 °C- autom. battery checking deviceMeasuring range: - 5 °C up to + 60 °CAccuracy: ± 0,1 °CPreparatory works for installation – see the specification for the preparatory works below.

17

Page 4: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

2 Wells: bottom depth meter up to 1000 m (set).White polyethylene tape (two wires), meter graduation in red, centimetre marks andDecimetre graduation in black on white tape.Probe body - brass galvanized,Tape drum - robust special plasticWorking temperature: -30 °C up to + 75 °CFrame - Alu cast, plastic coated. Probe (level and temperature probe) Signal lamp Measuring tape Battery power supply Crank Alu cast-frame, plastic coated, with robust tape drumMeasuring range - Up to 1000mPreparatory works for installation – see the specification for the preparatory works below.

2

3 Spring discharges: flow velocity measurements. Electro magnetic flow portable current meter for low velocities and shallow waters.Electromagnetic Flow Sensor Portable: Material: proof epoxy resin Range: 0 to 2.5 m/sAccuracy: 1% of measured value

2

4 Spring and river discharges: flow velocity measurements (set for rod measurements)Current meter including: brass propellers 125mm and 80mm, calibration certificates and tables, minimum 6 m rod, sinking weight and attachments, water speed and discharge calculation portable logger with LCD display, supporting at least 10 measuring verticals, possibility for labeling the verticals and cross-sections, memory for at least 10 cross-sections observation, PC connection cable and current meter cable.

17

Page 5: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

5 Additional set of accessories for suspended cable measurements by current meter.Rope graduated at 1/10cm, winch, sinking weight, winch for bridge measurements

4

6 Digital River and spring water temperature probesMax T - at least 70 deg. CResolution – 0,1deg.C

21

7 Spring discharge measurements - volumetric method Set of plastic tanks Volume :5,10,50 l

7

8 River cross-section profiles measurementsGeodesic levelling and distance measuring instruments for river cross-section profiles, laser accessories for measuring distances. Accuracy better than 1 cm.

3

9 Longitude, latitude, elevation GPS unit with coordinates and elevation measuring possibilities (better than 0.5 m for standalone applications) GPS using a local stationary base station and a mobile unit to achieve vertical positional accuracy < 0.50 mThree-dimensional position measurementsAt least WGS84 coordinate system displayASCII, RTCM SC-104 and XMODEM serial protocols1.EGNOS.2 Serial clip 3 Wireless 4 Compatible with USB 2.0

4

10 Longitude, latitude GPS unit with coordinates and elevation measuring possibilities Accuracy: minimum 8 m. horizontal coordinatesUSB Data Port for PC Connection/Data InputCase: Armoured rubber

5

11 Time chronometerLCD display for minutes, seconds and 1/100 sec., time and day, date, start –stop, addition stop and split functions. Integrated, replaceable battery

7

Page 6: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

12 Wells: water level and temperature metersAutomatic level meter, up to 250m.Material: Stainless steelMeasuring cell: CeramicChangeable at sight Li batteriesAccuracy: 0.05% of full rangeInterface: IrDA/RS232GSM/GPRS real time data transmission functionalityPreparatory works for installation – see the specification for the preparatory works below.

40

13 Automatic stations – (For Wells) with real time data transmissionWith real time data transmission suitable for wells (17 stations total). All of them with water level/temperature and precipitation/evaporation gauges, 5 - with additional air temperature and humidity sensorsPreparatory works for installation – see the specification for the preparatory works be-low.

13.1 Data acquisition / transmission unitData logger with the functionality to send alarm SMS when certain monitoring parameter exceeds user specified threshold, 230V AC power supply, buffer accumulator, modem supporting GPRS protocols, RS232 communication, SDI 12 interface and Infrared (IrDA) interface, lightning protection.Min. 8 free configurable input channelsMin. 4 MB internal memory Water proof case

17

13.2 Water level / temperature pressure probeMeasuring range - 0… 10 m Accuracy: 0.05% of full range

17

Page 7: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

13.3 Air temperature and relative humidity Temperature sensor with thermal radiation shield:Temperature-compensated working range:-5 °C +45 °C (ice-free)Measuring range, temperature: -25 °C +70 °C (ice-free)Resolution, temperature: 0,1 °CAccuracy, temperature: ±0,5 °C Humidity sensor: (Relative humidity)Humidity sensor:Relative humidityMeasuring range: 0...100% rel. hum. Output: 0 ... 1 V Measuring element: capacitive sensorAccuracy: ± 2 %

5

13.4 Combined evaporation, precipitation/snow intensity.Pluviometer for rain/snow/evaporation (resolution minimum 0.05mm)Installation: 2“-dia steel pipeWorking temperature range: -5 °C ... +70 °C (without heating device)Intensity 0 ... 50 mm/minMeasuring principle: Weighing

17

14 Automatic stations – (For springs and open channel flow) with real time data transmission station All of them with water level/temperature and precipitation/evaporation gauges, 2 - with additional air temperature and humidity sensorsPreparatory works for installation – see the specification for the preparatory works below.

Page 8: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

14.1 Data acquisition / transmission unit.Data logger with the functionality to send alarm SMS when certain monitoring parameter exceeds user specified threshold, 230V AC power supply, buffer accumulator, modem supporting GPRS protocols, RS232 communication, SDI 12 interface and Infrared (IrDA) interface, lightning protectionMin. 8 free configurable input channelsMin. 4 MB internal memory Water proof case

5

14.2 River/spring: water level measurements.Drift free water level probe: range 0 – 10m, accuracy better than 1 cm, resistant in high turbidity and pollution; measuring principle – differential pressure, bubbling chamber, the sensor is at 50 – 100 m distance from the logger

5

14.3 Water temperature meterTemperature sensor

5

14.4 Air temperature and relative humidity. Temperature sensor with thermal radiation shieldHumidity sensor: Relative humidity. Thermal radiation shield of anodised aluminium and improved ventilation for exact measurementHumidity sensor: Relative humidity

2

14.5 Combined evaporation, precipitation/snow intensity.Pluviometer for rain/snow/evaporation (resolution minimum 0.05mm)Measuring principle: Weighing

5

Page 9: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

15 Servers for data handling and distribution at the regional offices2 X CPU - 2.86 GHz or greater, L2 cache 2MB, FSB 800 MHz ; RAM - 2GB ECC DDR2 400, min 4x PCI extension slots 2 x HDDs min 160 GB/7200 rpm/8MB, RAID 0,1 Controller, optical device - CD&DVD R/W Keyboard Lat/ CyrMonitor 17 inch, TFT typeLinUx based operating system

5

16 Laser (or LED) printer/colour Printer Type: Colour Laser (or equivalent LED)Print speed 10 ppm b/w, 5 ppm colourPaper size A4

5

17 Computer for field data collection and data pre-processing at the local officesPortable PC (Notebook)Processor no less than 1.86 GHz, 1MB L2 cache, System Bus Speed 400 MHz, memory capacity 1 GB DDR2, HDD 80 GB, WLAN 802.11 b/g, 14” LCD displayPreloaded OS Windows XP OEM

15

18 Software for reporting, basic calculations.MS Office 2003 Professional or equivalent

15

Page 10: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

19 Software for data storage and analysis for 5 serversPlatform independency (possibility to run on Linux platforms)Functionality related requirements:Transaction supportRecord level locking mechanism for resolving concurrency during records updateProcedural language and stored procedures creation and their storing in the data baseTriggers definition capabilityIntegrity constrainsSoftware APIXML capabilitiesPossibility to create object types Bitmap indexing Administration related requirements:Possibility for performance tuning Online backup capabilitiesRobustness and extendibility requirements: - Recovery from hardware crashes- Clustering possibility- ReplicationIntegration with other software tools requirements:- To be used by mostly spread GIS; - Possibility for integration (to be used) with modelling software; Training: 4 days of training of 4 DB administrators

5

Page 11: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

20 Software for statistical analysisThe software should be menu driven. It should possess procedures for Exploratory Data Analysis, standard unvaried parametric statistical inference, multivariate statistical analysis (linear and quadratic discriminant analysis, principal component, factor analysis, canonical correlation, cluster analysis), univariant and multivariant linear multiple and nonlinear regression, multivariate adaptive regression splines for data smoothing, nonparametric statistical analysis, generalized linear models, generalized additive models, neural networks with multilayer perceptrons and general classification/regression tree models for classification and/or regression problems for predicting categorical and/or continuous variables, spectral time series analysis, and time series modeling and forecasting techniques based on Box and Jenkins. The software should be able to handle large data sets and to allow clustering of continuous and/or categorical variables, and to provide the functionality for complete supervised and unsupervised learning for pattern recognition and probability-based clustering. Graphics interface(The functionality should cover at least that of Stat Soft Statistica)

3

Page 12: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

21 OS and software for running WEB applications Minimal requirements for OS:- Multi users support- Multi processing and multi threading support- Networking features: Network drivers TCP/IP Sockets- Support of multiple processors- Security features: Identification and authentication Authorization Access control- Shell and user interface Support of shell scripts Graphical user interface

Web Application servers:- Protocols: HTTP/HTTPS XML SSL CGI- Security management- Clustering- Load balancing- Connection pooling- Session support- Internationalization (UTF 8)- Server site pages support- Possibility for component based development

Training: 4 days of training of 4 administrators Windows/Linux platforms and web Application servers

3

Page 13: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

22 Software for GIS data visualization, with hydro tools, geostatistical analysis tools, 3D analysis tools, maps creationCreating and editing GIS layers – point, polygon and polyline themes, tabular data.Create a map topology - topological relationships.Geostatistical tools supporting spatial modeling and features’ analysis, possibility of georeferencing for scanning maps or photographs and satellite images to a map coordinate system, digitizer support tools for work with 3D vector features, projections tools, DTM generating tools from point, line, and polygon features, grid manipulating tools etc.Geo database definition and administration / annotation, multi-user versioned data management, topology, vector data model, raster datasets and catalogs.Groundwater tools, hydrology tools for drainage basins delineation.To support different development environments: JavaScript, VBScript, Visual Basic, VBA, C++, Visual Studio.NET. Compatibility to the OGC standards, COM and ISO standard and Geographic Information—Metadata standards.Import/Export of variety of raster and vector GIS formats (coverage, shapefile, geodatabase, XML, TAB, MIF, DGN, DXF, PDF, AI, TIFF, JPEG, PNG, EMF, BMP).

4

23 Software for hydrogeological and geological data visualisation with the functionality for: borehole log plotting, 3D visualization and animation, cross-section editing, hydro-chemical data handling, graphical user interface (The functionality should cover at least that of Hydro Geo Analyst of Waterloo Hydrogeologic Inc.)

4

Page 14: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

24 Sofware for modelling platformThe modelling platform should comprise wide range of possibilities for constructing a specific model related to surface and ground-water interaction, groundwater infiltration recharge, surface and groundwater flow, as well as various tolls related to analysis, management and planning, well developed graphical interface.With respect to groundwater modelling the platform should include the following functionalities:

Tree – dimensional finite-difference groundwater model Simulation of steady and non-steady flow in an irregularly shaped flow system in which aquifer layers can be confined, unconfined, or a combination of confined and unconfined.

Prediction of Impacts from Salt Water Intrusion from Over Pumping in Coastal Regions Delineation of well capture zones for municipal drinking water supplies Design and optimisation of pumping well locations for dewatering projects Evaluation of groundwater remediation systems (pump & treat, funnel & gate, etc.) Simulation of natural attenuation of contaminated groundwater Estimation of reductive dechlorination of TCE, PCE and DCE in groundwater Determination of contaminant fate and exposure pathways for risk assessment

With respect to surface waters the platform should include functionalities related to estimation and modelling of:

Surface water impact from groundwater withdrawal Conjunctive use of groundwater and surface water Wetland surface Directions and rates of exchange of water between the groundwater body and associated surface system

Environmental impact assessments (changes in land use and climate Aquifer vulnerability mapping with dynamic recharge and surface water boundaries Floods/Floodplains Agricultural practices impact including irrigation, drainage and nutrient and pesticide use.

3

25 Bulgarian GIS layers. 425.1 High accuracy DTM (25-50m GRID) 425.2 Soils: vectors with accuracy and parameters corresponding to map scale 1:400 000

with attribute tables containing at least soil class, texture, profile thickness, maximal field water capacity, infiltration rates, others)

3

25.3 Detailed geological map vectors with accuracy and parameters corresponding to map scale 1:100 000 with attribute tables

4

Page 15: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

26 Mobile Laboratory Off road car – pickup for equipment, 4+1 people, samples, able to reach observation sights in difficult weather and road conditions including:- four doors, air-conditioned, luggage compartment at least 150 x 150 x 90 with hard top cover;- maximum load: at least 900 kg- clearance at least 23 cm, - maximum gradient - at least 40°.- diesel engine min 120 kW at 4000 rpm, torque at least 400 Nm- 4 x 4, manual gear box; 4 channel ABS; front and back independent suspensions;- Safety/ security: immobilizer, central door locking system, alarm system- additional 4kW 230V AC el. portable generator, diesel engine.

7

Page 16: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

27 Portable photometer and thermo reactorMeasured parameters: Ammonium, nitrite, nitrate, orthophosphate, chloride, sulfate, cyanide, calcium, magnesium, COD, heavy metalsPhotometer working in visible range (350-800 nm), 2 nm;At least 8 different photofilters covering visible range;Photometric reproducibility not less than 0.001 A;Photometric resolution not less than 0.001 A;Calibration curves - integrated in the memory of the photometer;Types of measurement: absorbance/transmission and concentration;Measuring range of absorbance up to 3.000A;Possibility for measurement with 10, 20, 50 mm and round cuvettes;Possibility for free method updating.Automatic correction of turbidity;Possibility for bar code selection of the methods.Implemented system for quality assurance.Memory: min. 800 data sets;Possibility for freely programmable methods;Rechargeable battery and AC power supply;Software and connection cable for data transfer to PC;Protective caseThermoreactor (heating digestion system):Power supply: ACSimultaneous digestion with at least two different temperatures Temperature selection: from 25 oC to at least 150 oC;Current temperature shown on the display;Temperature stability: at least ±1 oCStored heating programs and possibility for freely selectable programs for temperature and time.

1

28 Portable Water SamplerWater sampler portable, with microprocessor control, vacuum dosing system, insulating box with 12 x 1 litre bottles

1

29 Sampler swingingSampler swinging, with bottles 500 ml, 1000 ml

1

Page 17: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

30 Portable pH meterMeasuring range/ resolution: pH: 0 ... +14;mV: -1999 … +1999 mV.Temperature: at least 0.0 … +80.0 °C.Accuracy ± 1 digit:- pH 0.005 pH; - mV ±0.3 mV at;- Temp. ±0.1 °C. Automatic temperature compensation; Calibration: 2 point automatic calibration with technical buffers and standard buffers.Data storage: at least 500 data setsData output: Via real-time graphic display and interface for export of logged data to PC (connection cable and software).Operating time: min. 48 hours with batteries.Protective case

1

Page 18: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

31 Portable conductivity / TDS / salinity meterMeasuring range/ resolution for:conductivity: 0.0 µS/cm to 500 mS/cm (in 5 ranges)temperature:- 5.0 ºC to +105.0 ºCsalinity: 0 to 70.0TDS: 0 mg/l to 1999 mg/lAccuracy (±1 digit):

Conductivity ±0,5 % of valueTemperature: ±0.1 ºC;Ref. Temperature: T ref 20 or 25 °C, selectable;

Cell constant: fixed and adjustable;Temperature compensation: automaticLCD display;Data storage: min. 500 data sets;Data output: Via display or interface;RS 232 Interface;Protective case;Measuring conductivity cell

Electrode cell with built-in temperature sensor;Cable length: min. 3 m;

1

Page 19: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

32 Portable dissolved oxygen meter - Accuracy ±1 digit for:- O2 ±0.5% of measured value- Temperature ±0.1 K;- Temperature compensation;- Air pressure compensation;- Salinity correction;- LCD display;- RS232 interface;- Protective case;Measuring oxygen sensor:- With temperature compensation;- Working life – at least 3 months with one electrolyte filling;- Waterproof sensor; - With 1-3 m cable and waterproof plug;- With storage vessel;- With calibration vessel;

1

33 Turbidimeter portable with accessories- Measuring principle: Nephelometric (90° scattered light)- Light source: Nephelometric or IR - Measuring range: from 0 to 1000 NTU - Resolution: 0.01 NTU from 0.00 to 9.99 NTU and 0.1 NTU from 10.0 to 99.9

NTU- Accuracy: ±2% of value - Reproducibility: ±1% of value - Short response time for measurement;- Automatic calibration - LCD display - Battery operation for at least 1000 measurements with one battery set - Carrying case with separate compartment for turbidity meter and cuvettes- Calibration standards from 0.02 to 1000 NTU Empty cuvettes for turbidity meter, 5 pcs.

1

34 Ceramic top hot plate (Max. 1500 W)

1

Page 20: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

35 Hot plate stirrer (Laboratory type)

1

36 Balance portable Weighting capacity: >200g; Readability: 0.0001 g; Reproducibility: ±0.0001 g or better; Linearity: ±0.0002 g or better; Response time (average): ≤ 4 s; Sensitivity drift within +10°C.. .+30°C: ±2.106 1°C or better; Pan size: grater then 60 mm diameter;Analytical draft shield;Dust cover; AC adapter; Level indicatorData printer

1

37 Digital burettes capacity 25 ml

3

38 Digital burettes capacity 50 ml

3

39 Timers Mechanical, traditional brass or polished chrome finishRange - Main Dial is calibrated for 0-60 seconds in 1/5th seconds Upper sub dial 0-30minutes Lower sub dial 0-60 seconds

2

40 Set of Pipettes Variable volume, automatic

15

41 Membrane filtration unit - Funnel: glass with capacity 250 – 300 ml- Glass filter disk with glass rim- Filter head: made from glass- Filter holder- Suction flask with capacity: approx. 1 l

1

Page 21: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

42 Water bath Six nestsVolume: 10-15 l.Microprocessor controlled circulation thermostatTemperature readout: LED typeTemperature stability: ± 0.2 °C stepsOver temperature protectionTemperature range: 5 °C to 100 °CTemperature control: on/offCover with various diameter rings

1

43 Thermostat Low temperature incubator - 20 0C

1

44 Refrigerator With temperature indication, two point temperature controlVolume approx. 300 lRange: 0÷12 oCRefrigerating agent: Ozone friendlyEnergy supply: AC

1

45 Submersible pump - up to 100 m water column, - generator - accessories

1

46 Portable winch - mechanical or hydraulic, - single conductor armoured sea-cable, - cable block, - connection kit

1

47 Water sampler - sample volume – 2 l- integrated thermometer with range - 0 to +60o C- guiding rails- drop messenger or other type of distant closing

1

Page 22: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

48 Sampler with cartridge kits for spontaneously releasing gasesStainless Steel CylindersTemperature Rating 200°F (93°C)Pressure Rating 1800 psi (125 bar)

1

49 Radiation counter, portableDetector Connector: MHV.Serial Connection: 25 pin DB25S connector, RS-232.Meter: 76 mm (3") scale, 50 divisions, 0 - 500 CPM or 0 - 5 CPS linear scale.Fast and slow response time programming feature (1-300 seconds)Linearity: Within 5% of full scale.Dead Time Correction: 1 - 999 microseconds.Alarm Point: Adjustable from 0 10 100% full meter scale.Low Count Fail: User adjustable 0 to 255 minutes (10 minutes default setting).Over-range Alarm: Adjustable from 1 to 1010 cps (106 cps default setting).Audible AlarmVisual AlarmSpeakerSensitivity: Lower Threshold: 0 - 5.1 mV, upper Threshold: 0 - 60 mV.Anti-coincidence: Programmable switch permits discrimination between beta and alpha pulses.Power: 230 V AC/ 50HzBattery Backup: 8 hours operation.

1

Page 23: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

50 Compact ion chromatography system for determination of cations and anionsAnions: fluoride, chloride, nitrate, nitrite, bromide, phosphate, sulfate, sulfide and cyanide. Cations: lithium, sodium, potassium, magnesium, calcium, barium and ammonium.Detection limits for anions - according ISO 10304-1 :1995 and REGULATION No. 1 on the Exploration, Use and Protection of Groundwater Anions: Fluoride - 0,005 mg/l , chloride - 0,05 mg/l, nitrate - 0,05mg/l, nitrite - 0.01 mg/l, bromide - 0,01 mg/l, phosphate - 0,05 mg/l, sulfate - 0,05 mg/l, sulfide - 0,002 mg/l and cyanide - 0,004 mg/l; Cations: lithium, sodium, potassium, magnesium, calcium, barium - 0,05 mg/l, and ammonium - 0,01 mg/l

IC Pumps:High class IC pump Chemical inert constructionComputer control of all pump parametersMinimum flow rate : 0,2 ÷ 0.5 ml/min

Autosampler:Software control of the autosampler parametersAt least 25 vials with variable volume

Suppressor module:Suppressor module with regeneration for chemical suppression for anion determinationSoftware control of the suppressor parameters100 % solvent compatible

2

Page 24: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

Detector: Conductivity detector: - Measuring range of conductivity minimum: 0 ÷ 1000 μS - Electronic suppression and auto zero function - Adjustable detector temperature in range: 25 ÷ 45 °C - Temperature stability: ≤ 0.01 oC

Computer system: PC with at least 2.4 GHz Intel Pentium or AMD or similar processor, 512 MB RAM, 120 GB SATA HDD, 3.5”FDD, 40xCD R/W, MS-WIN XP Pro or equivalent OS, keyboard, mouse, 17” TFT color display, printer with at least 1200x1200 dpi, 18ppm/A4 B/W, 8 MB RAM

Application software features: - data acquisition from detectors and samplers, - full system control: pumps parameters autosamplers parameters suppressor parameters- Possibility for setting limits and thresholds - Reporting and alerting messages to the operator feature- Possibility for executing various data analysis - Possibility for module configuration and sequencing

Columns:Columns for determination of the a.m. anions and cationsReactives and standards for the a.m. anions and cations

Page 25: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

51 UV-VIS Spectro-photometric SystemUV-Visible instrument based on diode-array technologyWavelength range minimum from 190 to 1000 nm- Resolution test >1.6 toluene in hexane, ratio abs. at 269 nm/266/ 260/ nm- Stray light <0.05% at 340 nm (NaNO2, ASTM) or < 0.1% at 220 nm (NaI, ASTM)- Wavelength accuracy <+/- 0,5 nm- Photometric accuracy +/- 0.01 A- Wavelength reproducibility +/- 0,05 nm- Cell holder for 10 mm path-length standard cell - Adjustable cell holder for cylindrical and rectangular cells up to 10cm path-length.

Computer system: PC with at least: 2.6GHz processor, 512MB RAM, 80GB HDD, 3.5” FDD, DVD combo, Video card 64 MB, MS-Win XP Pro or equivalent OS, keyboard, Optical mouse; 17” color TFT display; InkJet printer with at least 1200x1200 dpi, 10ppm/A4 colour, USB port

Original UV-VIS software package: Win32 application UV-VIS software, which consists of modules for full instrument control, measurement, data acquisition and evaluation of spectrometer data; Various data analysis modes, spectrum presentation and processing modes; GLP functions.

Automated sipper system with peristaltic pump, 10-mm path length flow cell and all proper tubing’s, fittings and cables

Rectangular cells- 10mm path-length rectangular quartz cell- 20mm path-length rectangular quartz cell- 50mm path-length rectangular quartz cell

1

Page 26: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

52 Magnetic stirrer with heating - Heating power: 600-800 W- Temperature range: up to 300 oC- Speed range: up to 1100 rpm- Set of stirring bars- Power supply: AC

1

53. Glassware for on-site evaluation of components characterizing ground water status

53.1 Adjustable-volume pipettes with 100tips 100-1000μl 553.2 Adjustable-volume pipettes with 100tips 500- 5000μl 553.3 Adjustable-volume pipettes with 200tips 1 -10 ml 553.4 105Bends with two conical ground sockets of the same size, borosilicate glass, ST-

cones 29/322

53.5 Beakers, low form, with spout and graduation, borosilicate glass 100 ml 1053.6 Beakers, low form, with spout and graduation, borosilicate glass 150 ml 2053.7 Beakers, low form, with spout and graduation, borosilicate glass 400 ml 2053.8 Beakers, low form, with spout and graduation, borosilicate glass 600 ml 3053.9 Beakers, low form, with spout and graduation, borosilicate glass 1000 ml 1553.10 Beakers, low form, with spout and graduation, borosilicate glass 2000 ml 553.11 Beakers, tall form, with spout and graduation, borosilicate glass 50 ml 2053.12 Beakers, tall form, with spout and graduation, borosilicate glass 150 ml 4053.13 Evaporating dishes, hemispherical form, porcelain, with spout, glazed inside and out,

ф80mm, capacity 90 ml20

53.14 Automatic burettes with PTFE stopcock, PTFE valve, intermediate stopcock, class A, includes 2l clear glass reservoir 25 ml

1

53.15 Automatic burettes with PTFE stopcock, PTFE valve, intermediate stopcock, class A, includes 2l clear glass reservoir 50 ml

1

53.16 Separating funnel conical shape, with PTFE plug and polyethylene stopper, graduated, borosilicate glass, ST-socket 29/32 500ml

3

Page 27: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

53.17 Separating funnel conical shape, with PTFE plug and polyethylene stopper, graduated, borosilicate glass, ST-socket 29/32 1000 ml

3

53.18 Bottle top dispenser 1-10 ml 253.19 Bottle top dispenser (for dispending concentrated acids(HCl,HNO3))10 - 50 ml 253.20 Erlenmayer flasks with conical ground socket, narrow neck, with graduation,

borosilicate glass, socket 19/26 100ml50

53.21 Erlenmayer flasks with conical ground socket, narrow neck, with graduation, borosilicate glass, socket 29/32 300 ml

30

53.22 Erlenmayer flasks with conical ground socket, narrow neck, with graduation, borosilicate glass, socket 29/32 500ml

20

53.23 Erlenmayer flasks with conical ground socket, wide neck, with graduation, borosilicate glass 300 ml

70

53.24 Desiccators with NS-24 stopcock in lid, lid is interchangeable, with plates 300 ml 253.25 Test tubes graduated with stopper, neck NS-12/21 10 ml 3053.26 Erlenmayer flasks with conical ground socket, narrow neck, with graduation,

borosilicate glass, socket 29/32, with PP stoppers NS 29/32, 300 ml 40

53.27 Evaporating dishes with spout, fused rim, borosilicate glass, diameter- 95mm, height- 55mm,capacity-170ml

30

53.28 Macro cuvet, optical special glass, 100-OS 10mm two-unit sets 253.29 Volumetric flasks with hollow glass stoppers, with graduation, borosilicate glass, class

A, ST-10/19 25ml30

53.30 Volumetric flasks with hollow glass stoppers, with graduation, borosilicate glass, class A, ST-12/21 50 ml

60

53.31 Volumetric flasks with hollow glass stoppers, with graduation, borosilicate glass, class A, ST-12/21 100 ml

80

53.32 Volumetric flasks with hollow glass stoppers, with graduation, borosilicate glass, class A, ST-14/23 250 ml

5

53.33 Volumetric flasks with hollow glass stoppers, with graduation, borosilicate glass, class A, ST-19/26 500 ml

25

53.34 Volumetric flasks with hollow glass stoppers, with graduation, borosilicate glass, class A, ST-24/29 1000 ml

10

Page 28: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

53.35 Eppendorf pipette tip 1-10 ml, bags of 100 tips 253.36 Eppendorf pipette tip 50- 1000µl, bags of 500 tips. 253.37 Eppendorf pipette tip 100- 5000 µl, 120mm, bags of 100 tips 553.38 Petri dishes, diameter x height 100х15 1053.39 Graduated pipettes with piston syringe, class A, 1,5,10,25 ml 1053.40 Volumetric pipettes with one mark, class A, 1,5,10,25,50 ml 753.41 Wide neck jars cylindrical, PP (PE), with screwcaps, neck internal 65-70mm, height

153mm, 1000 ml150

53.42 Narrow neck bottle, with NS glass stopper, neck NS 14/15, amber glass 100 ml 1053.43 Narrow neck bottle, with NS glass stopper, neck NS 19/26,amber glass 250 ml 1553.44 Narrow neck bottle, with NS glass stopper, neck NS 24/29,amber glass 500 ml 1553.45 Narrow neck bottle, with NS glass stopper, neck NS 29/32,amber glass 1000ml 1553.46 Wide neck bottle, with screw cap, thread 55, amber glass, 500ml 2553.47 Test tubes(straight) use for Kjeldahl determination, 250 ml 1353.48 Analytical sieves -0,063mm,0,125mm,1mm and 2mm 153.49 Weighing botties d.50, h 30 4053.50 Weighing botties d.50, h 50 1053.51 Weighing botties d.40, h 80 1053.52 Laboratory thermometer, range -20ºС …+50ºС, division 0.1ºС 653.53 Laboratory thermometer, range -10ºС …+100ºС, division 0.5ºС 553.54 Glass fiber filters, pore size 0.45 μm, Ø 55 mm, 100 pcs. pack 253.55 Glass fiber filters GF8, Ø 47mm, 200pcs pack 153.56 Quartz fiber filter discs QF 20, Ø 47mm 100 pcs. pack 353.57 Filter paper -sheets-80g/m2 10 -12 kg /packs 2

Page 29: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

53.58 Filter papers slow Ø150mm, 100pcs. pack 2053.59 Funnels, glass d40мм 2053.60 Funnels, glass d70мм 4053.61 Measuring cylinders with spout, round base, class A, with graduation, 100ml 3553.62 Measuring cylinders with spout, round base, class A, with graduation, 500ml 753.63 Measuring cylinders with spout, hexagonal base, with graduation, class A, PE-

stopper, 50 ml15

53.64 Karlsruhe bottle including 60 mm ground stopper, suitable for BOD5 determination, 300ml, 10 pcs./pack

20

53.65 Stirring rods 154 Chemicals and reagents for on-site evaluation of components characterizing

ground water status54.1 Test kits –nitrate 23-225 mg/l- 25 tests 454.2 Test kits –nitrite 0.1-2.0 mg/l- 400 tests 154.3 Test kits –phenol 0,002-5,00mg/l-50- 250 tests 254.4 Test kits – fluoride 0,1-20 mg/l- 100 tests 154.5 Test kits –phosphate 0,01-5,0mg/l- 420 tests 154.6 Test kits –chlorine 0,05-7,50mg /l- 200 tests 154.7 Test kits –COD 4-40 mg /l- 25 tests 454.8 Test kits –COD10-150 mg /l- 25 tests 454.9 Test kits –COD15-300 mg /l- 25 tests 254.10 Test kits –COD 25-1500 mg /l- 25 tests 454.11 Test kits –COD 50-500 mg /l- 25 tests 254.12 Test kits –COD 300-3500 mg /l- 25 tests 254.13 Test kits –COD 500-10000 mg /l- 25 tests 2

Page 30: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

54.14 Test kits –cyanide 0,01-0,5mg /l- 25 tests 354.15 Test kits –surfactants(anion.) 0,05-2,0mg/l - 25 tests 254.16 Analytical quality assurance for photometer Combi Check 10, 1set 154.17 Analytical quality assurance for photometer Combi Check 20, 1set 154.18 Analytical quality assurance for photometer Combi Check 30, 1set 154.19 Analytical quality assurance for photometer Combi Check 50, 1set 154.20 Analytical quality assurance for photometer Combi Check 60, 1set 154.21 Analytical quality assurance for photometer Combi Check 70, 1set 154.22 Analytical quality assurance for photometer Combi Check 80, 1set 154.23 N-Allylthiourea 50g. for synthesis, ISO. 254.24 Aluminium oxide 90 active neutral 1kg. 154.25 Ammonium monovanadate 25g GR for Analysis, ISO 154.26 Ammonium chloride 500g. GR for Analysis, ISO 154.27 Ascorbic acid 250g. . GR for Analysis, ISO 454.28 Barium chloride 2Н2О 500 g. GR for Analysis, ISO 154.29 pH buffer 4,01 /25ºС/ 500 ml Traceable to SRM of NIST 254.30 pH buffer 6,863 /25ºС/ 500 ml Traceable to SRM of NIST 254.31 pH buffer 9,18 /25ºС/ 500 ml. Traceable to SRM of NIST 254.32 Vanadate- molybdate reagent 500ml 154.33 D-Glucose 250g. for biochemistry 254.34 Dimethylglyoxime 100g GR for Analysis, ISO 154.35 Ethylenedinitrilotetraacetic acid, disodium salt. 2Н2О 250 g. GR for Analysis, ISO 254.36 Ethylenedinitrilotetraacetic acid, disodium salt solution 0.1 mol/l , 1amp. Traceable to

SRM of NIST2

Page 31: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

54.37 Iron, particle size 10 µm, 100g GR for Analysis, ISO 154.38 Iron (II) sulfate 50g GR for Analysis, ISO 154.39 Iron(III) chloride.6H2O, 250g GR for Analysis, ISO 454.40 Mercury(II) iodide 50g. GR for Analysis, ISO 254.41 Mercury(II) sulfate 50g GR for Analysis, ISO 454.42 Potassium antimony tartrate 250g. GR for Analysis, ISO 154.43 Potassium dichromate 500g. GR for Analysis, ISO 254.44 Potassium permanganate 250g. GR for Analysis, ISO 154.45 Potassium hydrogen phthalate 250g.GR for Analysis, ISO 154.6 Potassium dihydrogen phosphate /КН2РО4//1kg. GR for Analysis, ISO 154.47 Calcium chloride.2H2O, 500g GR for Analysis, ISO 254.48 Quartz wool 100g 1054.49 Magnesium sulfate.7H2O, 500g GR for Analysis, ISO 154.50 Manganese (II) chloride 500g GR for Analysis, ISO 254.51 Sodium azide 100g. GR for Analysis, ISO 154.52 Sodium carbonate 500g. GR for Analysis, ISO 154.53 Sodium dodecyl sulfate 5g. GR for Analysis, ISO 154.54 Sodium oxalate 500g. GR for Analysis, ISO 154.55 Sodium hydroxide 1kg. GR for Analysis, ISO 1054.56 Sodium pyrophosphate Н2О, 500g. GR for Analysis, ISO 154.57 Sodium salicylate 250g. GR for Analysis, ISO 854.58 Sodium tetraborate 10Н2О, 1kg GR for Analysis, ISO 154.59 Sodium thiosulfate.5Н2О, 250g. GR for Analysis, ISO 1

Page 32: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

54.60 Sodium thiosulfate solution 0.01N , 1amp. Traceable to SRM of NIST 254.61 Sodium thiosulfate solution 0.1N, 1amp. Traceable to SRM of NIST 254.62 Sodium hydrogen carbonate 500g. GR for Analysis, ISO 154.63 tri- Sodium citrate 2Н2О, 500g. GR for Analysis, ISO 454.64 Starch 100g. 254.65 Silica gel 5kg. 254.66 Potassium chloride solution (nominal 1.41 mS/cm)Certified reference material for the

measurement of electrolytic conductivity 30 packs /box Traceable to SRM of NIST1

54.67 Iron standard solution 500ml Traceable to SRM of NIST 154.68 Calcium standard solution 500ml Traceable to SRM of NIST 154.69 Potassium hydrogen phthalate solution for COD determination /0,17g/l./ 100ml

Traceable to SRM of NIST1

54.70 Magnesium standard solution 500ml Traceable to SRM of NIST 154.71 Manganese standard solution 500ml Traceable to SRM of NIST 154.72 Sulfate standard solution 500ml Traceable to SRM of NIST 154.73 TOC standard solution 100ml Traceable to SRM of NIST 154.74 Fluoride standard solution 500ml Traceable to SRM of NIST 154.75 Phosphorus standard solution 100ml Traceable to SRM of NIST 154.76 Chloride standard solution 500ml Traceable to SRM of NIST 154.77 Sulfuric acid 97% 2.5l GR for Analysis, ISO 754.78 Carbon tetrachloride for spectroscopy Uvasol -2.5l 1054.79 Titanium (II) oxide 500g. GR for Analysis, ISO 154.80 Phenol 1 kg. GR for Analysis, ISO 154.81 Phenolphthalein 50g. 154.82 Ammonium iron(II) sulfate .6Н2О, 500g GR for Analysis, ISO 4

Page 33: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Item Number

Specifications Quantity Evaluation Committee’s notes

54.83 Ferroin solution 100ml 154.84 Formaldehyde 37% 1l GR for Analysis, ISO 154.85 Hexamethylenetetramine 100g GR for Analysis, ISO 154.86 Hydrazinium sulfate 100g GR for Analysis, ISO 154.87 Hydroxylammonium chloride 50g. GR for Analysis, ISO 254.88 Chloroform 1l GR for Analysis, ISO 1054.89 Mercury(II) chloride 50g GR for Analysis, ISO 154.90 Redox Buffer Solution, 220mv / рН 7/ , 250ml Traceable to SRM of NIST 1

Page 34: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Specification of the preparatory works needed to install and protect from barbarism the monitoring equipment (wells)

1.Type A. Wells (boreholes having a diameter of the well head from 10 to 50 cm) – 39 locations1.1. Making a reinforced concrete small platform at the ground level around the well head with an area of 2 – 3 square meters and thickness

10 cm.1.2. Cutting/shortening the well head to the standard height of 20 cm. above the concrete platform.1.3. Making a cover of the well head, Reinforced concrete cube (B:1m, H:1m, L:1m) with wall thickness 10 cm and lockable iron lid at the

upper part having diameter 30 cm assuring access to the well.1.4. Pulling out cables and/or a pump if there are any into the borehole.1.5. Placing the Reinforced concrete cover cube over the platform above the well head.

2.Type B. Wells (having a diameter from 10 to 50 cm – 25 locations) 2.1. Cutting/shortening the well head to the standard height of 20 cm. above the concrete platform;2.2. Making steel iron cover with a small lockable hole on the top and fixing it over the well head.

3.Type C. Shaft Wells (having a diameter from 100 to 200 cm – 26 locations).3.1. Making a lockable steel iron cover and fixing it over the well head.

4.Type D. Spring and river level measuring point – 5 locations4.1.Concrete fundament for the equipment and foot gauge installation, mesh fence 4 x 4m.

5. For type A, B and D – 22 locations (including 9 deep wells; 8 shallow wells; 5 spring/rivers)5.1.230V/50Hz/10A mains power supply at 300m \average distance, including consumption meter.

Page 35: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

Sm a ll Fe rro -c o nc re tep la tfo rm (2-3 sq . m .)

Fe rro -c o nc re te C ub e(100/100/100 c m )

Lo c ka b le Iro n Lid (D ~ 70 c m )

70 7030

20

10

1

00

40

60

35 35

Se c tio na l Vie w

10

10 1020 20

30 60 70 100

170

Bird -Eye Vie w

Ground water quantity and quality monitoring equipment will play a key role in future monitoring program in accordance with requirements of European Legislation (WFD 2000/60/EC). All the beneficiaries have some monitoring equipment but it is quite insufficient at all. There are no mobile laboratories for water quantity and quality monitoring equipment among the beneficiaries. These mobile labs with additional portable monitoring equipment are important tools for monitoring and control of surface and ground waters and wastewaters and also mostly in case of accidental/incidental pollution. Taking in to account the real condition in the country as well as the fact that the most of the observation points are situated in an open field without protection it is highly recommended the above shown structure against barbarism to be built (fulfilled).

Page 36: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

PLACE OF DELIVERY

Item Nr

Denomination Q’ty Destination

1 Wells: water level and temperature meters (set) 17 NIMH1 in: Dobrich 3, Varna 4, Shumen 1, Sindel 1; Vaklino 1; Shabla 1, Pleven 2, Silistra 2,

ExEA RL in: Shumen 1, Varna 1.

2 Wells: bottom depth meter up to 1000 m (set) 2 NIMH in: Dobrich 1; Varna 1.

3 Spring discharges: flow velocity measurements: Electro magnetic flow portable current meter for low velocities and shallow waters.

2 NIMH in: Dobrich 1; Varna 1.

4 Spring and river discharges: flow velocity measurements (set for rod measurements)

17 NIMH in: Dobrich 4, Varna 4, Shumen 2, Sindel 1; Abrit 1; Draganovo 1; Silistra 1, Pleven 3

5 Additional set of accessories for suspended cable measurements by current meter.

4 NIMH in: Dobrich 1; Varna 1, Pleven 2

6 Digital River and spring water temperature probes 21 NIMH in: Dobrich 2, Varna 2, Shumen 2, Pleven 2, Sindel 2; Abrit 1; Draganovo 1; Silistra 2; Shabla 1.

ExEA RL2 in: Varna 3, Shumen 3

7 Spring discharge measurements - volumetric method Set of plastic tanks

7 NIMH in: Dobrich 2, Varna 2, Shumen 1, Abrit 1; Draganovo 1

8 River cross-section profiles measurements:

Geodesic instruments, accessories for measuring distances.

3 NIMH in: Varna 2, Pleven 1

9 Longitude, latitude, elevation GPS unit with coordinates and elevation measuring possibilities

4 NIMH in: Varna 1; Dobrich 1;

ExEA RL in: Varna 1, Shumen 1

10 Longitude, latitude. GPS unit with coordinates and elevation measuring possibilities

5 NIMH in: Dobrich 1; Varna 1; Pleven 1.ExEA RL in: Varna 1, Shumen 1

1 National Institute for Meteorology and Hydrology2 Executive Environmental Agency Regional Laboratories

Page 37: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

11 Time chronometer 7 NIMH in: Dobrich 1, Varna 2, Shumen 1, Sindel 1; Abrit 1; Draganovo 1.

12 Wells: water level and temperature meters 40 NIMH Stations: see Table 1 below

13 Automatic stations – (For Wells)

13.1 Data acquisition / transmission unit 17 NIMH Wells: see table 2 below

13.2 Water level / temperature pressure probe 17 NIMH Wells: see table 2 below

13.3 Air temperature and relative humidity. 5 NIMH Wells: see table 2 below

13.4 Combined evaporation, precipitation intensity. 17 NIMH Wells: see table 2 below

14 Automatic stations – (For springs and open channel flow) with real time data transmission station

14.1 Data acquisition / transmission unit 5 NIMH Stations: see table 3 below

14.2 River/spring: water level measurements 5 NIMH Stations: see table 3 below

14.3 Water temperature meter 5 NIMH Stations: see table 3 below

14.4 Air temperature and relative humidity sensor 2 NIMH Stations: see table 3 below

14.5 Combined evaporation, precipitation intensity. 5 NIMH Stations: see table 3 below

15 Servers for data handling and distribution 5 RBD3 in: Varna: 1NIMH in: Varna 1, Dobrich 1

ExEARL in: Shumen 1, Varna 1

16 Laser printer/colour 5 RBD in: Varna: 1NIMH in: Varna 1, Dobrich 1RBD in: Pleven 1

ExEA RL in: Shumen 1

17 Computer for field data collection / Portable PC 15 RBD in: Varna 2, Pleven 1ExEARL in: Shumen 3, Varna 2NIMH in: Varna 2, Dobrich 1;

Sindel 1; Shumen 1; Pleven 2.

3 River Basin Directorate

Page 38: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

18 Software for reporting, basic calculations. 20 RBD in: Varna 3, Pleven 1,NIMH in: Varna 3, Dobrich 2;Sindel 1; Shumen 1; Pleven 2.

ExEARL in: Shumen 4,Varna 3

19 Software for data storage and analysis 5 RBD in: Varna 1, Pleven 1NIMH in: Varna 1; Dobrich 1.ExEA RL in: Varna 1

20 Software for statistical analysis 3 RBD in: Varna 1NIMH in: Varna 1

ExEARL in: Shumen 1

21 Software for WEB design and service 3 RBD in: Varna 1NIMH in: Varna 1

ExEA RL in: Shumen 1

22 Software for GIS data visualization, with hydro tools, geostatistical analysis tools, 3D analysis tools, maps creation

4 RBD in Varna 1, Pleven 1NIMH in: Varna 1

ExEA RL in: Shumen 1

23 Hydrogeological and geological data visualisation 4 RBD in Varna 1, Pleven 1NIMH in: Dobrich 1

ExEA RL in: Shumen 1

24 Sofware for modelling platform 3 RBD in Varna 1NIMH in Varna 1

ExEA RL in: Shumen 1

25 Bulgarian GIS layers. 4 RBD in Varna: 1, Pleven 1NIMH in: Varna 1

ExEA RL in: Shumen 1

25.1 High accuracy DTM 4 RBD in: Varna: 1, Pleven 1NIMH in: Varna 1

ExEA RL in: Shumen 1

25.2 Soils 3 RBD in: Varna 1, Pleven 1

NIMH in: Varna 1

25.3 Detailed geological map vectors with attribute tables

4 RBD in: Varna 1, Pleven 1NIMH in: Varna 1

ExEA RL in: Shumen 1

26 Mobile laboratory 7 RBD in: Varna 1, Pleven 1NIMH in: Varna 1; Dobrich 1, Pleven 1;

ExEA RL in: Shumen 1, Varna 1

27 Portable photometer and thermoreactor 1 ExEA RL in Shumen

Page 39: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

28 Portable Water Sampler 1 ExEA RL in Shumen

29 Sampler swinging 1 ExEA RL in Shumen

30 Portable pH meter 1 ExEA RL in Shumen

31 Portable conductivity / TDS / salinity meter 1 ExEA RL in Shumen

32 Combined set of Portable pH meter, conductivity /TDS / salinity meter and dissolved oxygen meter

1 ExEA RL in Shumen

33 Turbidimeter portable with accessories 1 ExEA RL in Shumen

34 Ceramic top hot plate 1 ExEA RL in Shumen

35 Hot plate stirrer 1 ExEA RL in Shumen

36 Balance portable 1 ExEA RL in Shumen

37 Digital burettes capacity 25 ml 3 ExEA RL in Shumen

38 Digital burettes capacity 50 ml 3 ExEA RL in Shumen

39 Timers 2 ExEA RL in Shumen

40 Set of Pipettes, variable volume, automatic 15 ExEA RL in Shumen

41 Membrane filtration unit 1 ExEA RL in Shumen

42 Water bath 1 ExEA RL in Shumen

43 Thermostat – Low temperature incubator 1 ExEA RL in Shumen

44 Refrigerator 1 ExEA RL in Shumen

45 Submersible pump 1 ExEA RL in Shumen

46 Portable winch (mechanical or hydraulic), single conductor armored sea-cable, cable block, connection kit

1 ExEA RL in Shumen

47 Water sampler 2 litre 1 ExEA RL in Shumen

48 Sampler with cartridge kits for spontaneously releasing gases

1 ExEA RL in Shumen

49 Radiation counter, portable 1 ExEA RL in Shumen

Stationary laboratory equipment

50 Compact ion chromatography system 2 ExEA RL in: Shumen, Varna

51 UV-VIS Spectro-photometric System 1 ExEA RL in Shumen

52 Magnetic stirrer with heating 1 ExEA RL in Shumen

53 Glassware for on-site evaluation of components characterizing ground water status

1 ExEA RL in Shumen

Page 40: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

54 Chemicals and reagents for on-site evaluation of components characterizing ground water status

1 ExEA RL in Shumen

Table 1. List of NIMH Wells:Nr Location Nr Location1 Senokos 21 Konare2 Dropla. 22 Bozhanovo3 Rakovski 23 Krapets4 Ograjden 24 Gurkovo5 Karapelit 25 Momchil/Tsarichino6 Dolina 26 Dabrava7 Chernevo 27 Dropla8 Vetrino 28 Gorichane9 Stan 29 Durankulak10 Gurkovo 30 Neykovo11 Mogilishte 31 Bezvoditsa12 Chernookovo 32 Zmeevo13 Prisad 33 Spasovo14 General Toshevo 34 Vaklino15 Dabovik 35 Malina16 Tsarevets 36 Snop17 Kochmar 37 Abrit18 Abrit 38 Pchelarovo19 Opanets 39 Ograzhden20 Seltse 40 Tulenovo

Table 2: List of NIMH Wells with Automatic Stations:Nr Location Note

1 “Makedonka” Pump Station + Air temperature and relative humidity

2 Vaklino 3 Chelopechene 4 General Toshevo, Pastir Quarter 5 Metodievo6 Lovchantsi7 Telerig

8 Smolnitsa +Air temperature and relative humidity

9 Dabrava10 Trigortsi11 Shabla12 Balchik13 Belgun

14 Zahari Stoyanovo +Air temperature and relative humidity

15 Uzovo +Air temperature and relative

Page 41: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

humidity16 Metodievo

17 Dobrich +Air temperature and relative humidity

Table 3: List of NIMH Meteorological Stations - Springs/open channelsNr Location Note1 Devnia2 Devnia3 Devnia4 Sindel +Air temperature and relative

humidity5 Odrintsi +Air temperature and relative

humidity

Page 42: ANNEX II: · Web viewGeneral Requirements The following requirements are valid for all items in the Technical Specifications except where explicitly stated otherwise: User manual

ADDRESS LIST

INSTITUTION Address Director Contact Person Code

Main telephone

with extention

Business phone

numberE-mail

River Basin Directorate Varna

33, Alexander Diakovich str., 9000 Varna

Dipl. Eng. Ventsislav Nikolov

Dipl. Eng. Ventsislav Nikolov 052 687 431 [email protected]

River Basin Directorate Pleven

1, Vassil Levski str., fl.6,

Dipl. Eng. Tsvetanka Dimitrova

Dipl. Eng. Tsvetanka Dimitrova 064 803-279

ExEA Regional Laboratory Varna

4, Yan Palah str.9000 Varna

Mrs. Bangieva Mrs. Bangieva 052 634 576 0889 270

337 [email protected]

ExEA Regional Laboratory Shumen

71 Saedinenie str., floor 4 9700 Shumen Mrs.Nina

Mihaylova Mrs.Nina Mihaylova 054 54 996 0889 270 340 [email protected]

NIMH - Branch Varna

10, Sveti Nikola site, 9005, Varna Radoslava

Manyarova Vesela Stoeva 052302 257

0887979166 [email protected]

NIMH - Branch Pleven

60 Hadzhi Dimitar str., 5800 Pleven, Ivan Matev Dipl. Eng. Marin

Marinov 05264 82 70 57

0889422399 [email protected]

NIMH - Hydromet observatory in Dobrich

80, 25 September blvd.9300 Dobrich

Radoslava Manyarova

Dipl. Eng. Veselina Pavlova 052

604 2890886262162 [email protected]