ambient air monitoring at letterkenny, co. donegal final rep… · ambient air was pumped through a...

25
Ambient Air Monitoring At Letterkenny, Co. Donegal 16 th May 2008 – 14 th July 2009

Upload: others

Post on 09-Aug-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

Ambient Air Monitoring

At

Letterkenny, Co. Donegal

16th May 2008 – 14th July 2009

Page 2: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

1

Contents Summary . . . . . . . . 3 Introduction . . . . . . . . 4 Time Period . . . . . . . . 5 Siting . . . . . . . . . 5 Monitoring Methods . . . . . . . 6 Results . . . . . . . . . 7 Carbon Monoxide . . . . . . . 7 Sulphur Dioxide . . . . . . . 10 Nitrogen Dioxide and Oxides of Nitrogen . . . . 14 Particulate Matter . . . . . . . 19 Lead . . . . . . . . . 22 Arsenic, Cadmium, Nickel . . . . . . 23

Page 3: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

2

List of Figures

Figure 1. Map of Site Location . . . . . . 5 Figure 2. Carbon Monoxide 8-hour Running Average . . . 8 Figure 3. Carbon Monoxide . . . . . . 9 Figure 4. Sulphur Dioxide 24-hour Averages . . . . 12 Figure 5. Sulphur Dioxide Hourly Averages . . . . 13 Figure 6. NO2 Hourly Values . . . . . . 17 Figure 7. NOX Hourly Values . . . . . . 18 Figure 8. PM10 Daily Values . . . . . . 21

Page 4: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

3

Summary An assessment of air quality was carried out in Letterkenny, Co. Donegal from 16th May 2008 to 14th July 2009. No limit values were exceeded during the measurement period. Concentrations of carbon monoxide, nitrogen dioxide, sulphur dioxide and lead were below their respective lower assessment thresholds. Concentrations of oxides of nitrogen exceeded the lower assessment threshold value concerning the protection of vegetation however this only applies in rural areas and thus does not apply at the Letterkenny site. Concentrations of PM10 exceeded the lower assessment threshold value concerning the protection of human health. Below Lower

Assessment

Threshold

Below Upper

Assessment

Threshold

Above Upper

Assessment

Threshold

Above Limit

PM10

NO2

NOx

CO

SO2

Pb

As

Ni

Cd

Letterkenny, Co. Donegal is located in Zone C. The implications of this assessment are that within Zone C (specified urban centres with population in excess of 15,000)

• Levels of PM10 can be assessed using a combination of measurement and modelling

• Levels of SO2, CO, NO2, benzene and lead can be assessed using modelling or

objective estimation techniques Directive 2008/50/EC, commonly refered to as the Cleaner Air for Europe, (CAFÉ), Directive states that modelling or objective estimation techniques may be used to assess ambient air quality if levels of the pollutant in question in that zone are below the lower assessment threshold. A combination of measurement and modelling is required if levels exceed the lower assessment threshold while continuous monitoring is required if levels exceed the upper assessment threshold. Indicative measurements of legislative heavy metals do not exceed the ambient air quality limit values described in the fourth daughter directive (2004/107/EC).

Page 5: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

4

Introduction

An updated and integrated approach to monitoring, assessment and management of air quality within the European Union was introduced through the Clean Air for Europe directive, (CAFE, 2008/50/EC) on the 21st May 2008. The directive replaced the pre-existing Air Quality framework directive (96/62/EC, 2nd September 1996) and three of the four preceding Air Quality framework daughter directives. It came into effect as of June 2010. The basic principle of the CAFE directive is that each country should be divided into zones and that the monitoring, assessment, management and reporting of air quality will be undertaken in relation to these zones. For the purposes of the directive, Ireland has been divided into four zones; Dublin (Zone A), Cork Urban Area (Zone B), specified population centres > 15,000 inhabitants (Zone C) and non-urban areas (Zone D). Limit values are set for each individual pollutant, which need to be met by a specific attainment date. Upper and lower assessment thresholds are also set for each pollutant, assessment thresholds are levels below the limit value, used solely in the determination of the level of monitoring needed for that pollutant in a particular zone. The extent of monitoring in any zone is determined by population size and air quality status. Measurement is mandatory in agglomerations (population >250,000) and where concentrations are above the lower assessment threshold. The greatest monitoring effort applies if concentrations are above the upper assessment threshold. Less intensive monitoring is required when concentrations are between the two assessment thresholds and maybe subsequently supported or replaced with ambient air modelling. Limit values, assessment thresholds, measurement techniques and other specifics for each pollutant which were previously described in the series of Daughter Directives, are now detailed in CAFE, with the exception of the most recent Fourth Daughter Directive . The first Daughter Directive was adopted in April 1999 (1999/30/EC) and covered SO2, NOX, particulate matter and lead. The second Daughter Directive was adopted in November 2000 (2000/69/EC) and covers CO and Benzene. The third Daughter Directive relates to ozone (2002/3/EC) while the fourth Daughter Directive relates to heavy metals and polycyclic aromatic hydrocarbons (2004/107/EC). The first three Directives were transposed into Irish law as the Air Quality Standard Regulations 2002 (S.I. No 271 of 2002) and the Ozone in Ambient Air Regulations 2004 (S.I. No 53 of 2004). To comply with the directive the Environmental Protection Agency uses mobile laboratories to carry out assessments in areas with no history of air pollution measurements. These trailers contained the following instruments:

• Monitoring instruments which continuously measure and record concentrations of the pollutants sulphur dioxide, nitrogen oxides, carbon monoxide and PM10.

• Sampler for lead and other metals in air (collection on filter for determination in the laboratory).

The sample inlets are at a height of ~3m. For further information please contact

Page 6: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

5

Micheál O’Dwyer, Lin Delaney, Kevin Delaney or Barbara O’Leary. Time Period The mobile laboratory was in place from the from 16th May 2008 to 14th July 2009. Siting The assessment was positioned at the County Museum, High Street, Letterkenny (Figure 1).

Figure 1. Map of site location (courtesy of OSI)

Page 7: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

6

Monitoring Methods Carbon Monoxide

Carbon monoxide was monitored using a Gas Filter Correlation CO Analyser (Model 300, Advanced Pollution Instrumentation, 6565 Nancy Ridge Drive, San Diego, California). This is a continuous analyser whose measurement technique is based on the absorption of infrared radiation by CO molecules at wavelengths near 4.7µm. Sulphur Dioxide

Sulphur dioxide was monitored using an Advanced Pollution Instrumentation SO2 Fluorescent Analyser - Model 100A. This is a continuous analyser, which measures the fluorescence of SO2 molecules after excitation by ultraviolet radiation. Nitrogen Dioxide and Oxides of Nitrogen

NOx species were monitored using an Advanced Pollution Instrumentation Chemiluminescent NO/NO2/NOx Analyser - Model 200A. This is a continuous analyser which utilises the chemiluminescent reaction between nitric oxide in the sample and ozone to measure NO concentrations. Any NO2 present is then reduced to NO by a molybdenum converter giving a second value for total NOx concentration. The amount of NO2 present is found by subtraction. Particulate Matter

Concentrations of PM10 were measured using an instrument which employed tapered element oscillating microbalance technology (TEOM, Rupprecht & Patashnick Co. Inc., 25 Corporate Circle, Albany, New York). This is a continuous method in which the air from the sampling head is passed through a filter placed on a tapered element. A mass transducer relates changes in the frequency of the tapered element to changes in particulate matter on the filter, the difference between the filter’s current weight and its initial weight gives the total mass of collected particulate matter. An inertial impactor sampling head restricted the sampled particles to those with a diameter less than 10µm. A filter dynamic measurement system, (FDMS), was coupled with the TEOM to compensate for the loss of volatile components which thus remove the need for a correction factor. Lead , Arsenic, Nickel and Cadmium

Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8µm) situated in a calming chamber. The filters were changed every 3-4 weeks. They were digested in conc. HNO3 and analysed for lead , arsenic, nickel and cadmium using ICP-MS (Inductively Coupled Plasma-Mass Spectrometry). All results for CO, SO2, NOX and the continuous particulate monitor were integrated to give 1-hour average values as required for comparison with the Directive limit values.

Page 8: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

7

Results The following sections details the results observed at the Letterkenny site during the assessment period. Summary statistics and graphical representations of the data are provided. Relevant threshold and limit values per parameter are stated.

Carbon Monoxide - CO No. of hours 10179 Missing values 76 (including routine maintenance) 19

No. of measured values 10103 Percentage covered 99.2%

Maximum hourly value 5.7 mg.m-3 98 percentile for hourly values 1.3 mg.m-3 Mean hourly value 0.3 mg.m-3 Maximum 8-hour mean 4.23 mg.m-3 98 percentile for 8-hour mean 1.25 mg.m-3 Directive 2008/50/EC – Carbon Monoxide - Limit values and Assessment

Thresholds

Averaging Period

Limit Value

Date by which

limit value is to be

met

Limit Value for

the protection of

human health

8-hour running

average

10 mg m-3

1 January 2005

Upper assessment

threshold

8-hour running

average

7 mg m-3

Lower assessment

threshold

8-hour running

average

5 mg m-3

Figure Two displays eight hour average carbon monoxide (CO) concentrations at the site. No exceedances of the Limit Value or Thresholds occurred. Figure 3 displays both hourly and eight hour average CO concentrations.

Page 9: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

8

0.0

0

2.0

0

4.0

0

6.0

0

8.0

0

10

.00

16/05/2008

30/05/2008

13/06/2008

27/06/2008

11/07/2008

25/07/2008

08/08/2008

22/08/2008

05/09/2008

19/09/2008

03/10/2008

17/10/2008

31/10/2008

14/11/2008

28/11/2008

12/12/2008

26/12/2008

09/01/2009

23/01/2009

06/02/2009

20/02/2009

06/03/2009

20/03/2009

03/04/2009

17/04/2009

01/05/2009

15/05/2009

29/05/2009

12/06/2009

26/06/2009

10/07/2009

mg/m3

Date

Fig

ure

2: C

arb

on

Mo

nix

ide

8-

Ho

ur R

un

nin

g A

ve

rag

e C

on

ce

ntra

tion

s, L

ette

rke

nn

y, 1

6/0

5/0

8 -

14

/07

/09

8-h

r CO

Lo

wer A

ssessm

ent T

hre

sho

ld

Up

per A

ssessm

ent T

hre

sho

ld

Lim

it Valu

e

Page 10: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

9

0.0

0

2.0

0

4.0

0

6.0

0

8.0

0

10

.00

16/05/2008

30/05/2008

13/06/2008

27/06/2008

11/07/2008

25/07/2008

08/08/2008

22/08/2008

05/09/2008

19/09/2008

03/10/2008

17/10/2008

31/10/2008

14/11/2008

28/11/2008

12/12/2008

26/12/2008

09/01/2009

23/01/2009

06/02/2009

20/02/2009

06/03/2009

20/03/2009

03/04/2009

17/04/2009

01/05/2009

15/05/2009

29/05/2009

12/06/2009

26/06/2009

10/07/2009

mg/m3

Date

Fig

ure

3:C

arb

on

Mo

no

xid

e 1

-Ho

ur &

8-H

ou

r Av

era

ge

Co

nc

en

tratio

ns

, Le

tterk

en

ny, 1

6/0

5/0

8 -

14

/07

/09

1-h

our

8-h

our

Page 11: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

10

Sulphur Dioxide – SO2

No. of hours 10179 Missing values 297 (including routine maintenance) 78 No. of measured values 9882 Percentage covered 97.1% Maximum hourly value 131.9 µg.m-3 98 percentile for hourly values 25.5 µg.m-3 Mean hourly value 6.3 µg.m-3 Maximum 24-hour value 33.9 µg.m-3 98 percentile for 24-hour values 17.9 µg.m-3 Directive 2008/50/EC – Sulphur Dioxide - Limit values and Assessment

Thresholds

Averaging Period

Limit Value

Date by which

limit value is to be

met

Hourly limit value

for the protection

of human health

1 hour

350 µg m-3 not to be exceeded more

than 24 times a calendar year

1 January 2005

Daily limit value

for the protection

of human health

24 hours

125 µg m-3 not to be exceeded more

than 3 times a calendar year

1 January 2005

Limit value for

the protection of

vegetation

Calendar year and winter (1 October

to 31 March)

20 µg m-3

-

Alert threshold

-

500 µg m-3 over

three consecutive hours

-

Page 12: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

11

Directive 2008/50/EC – Sulphur Dioxide - Limit values and Assessment

Thresholds Continued

Averaging Period

Limit Value

Date by which

limit value is to be

met

Upper assessment

threshold for the

protection of

human health

24 hours

75 µg m-3 not to be

exceeded more than 3 times a calendar year

-

Lower assessment

threshold for the

protection of

human health

24 hours

50 µg m-3 not to be

exceeded more than 3 times a calendar year

-

Upper assessment

threshold for the

protection of

vegetation

Calendar year and winter (1 October

to 31 March)

12 µg m-3

-

Lower assessment

threshold for the

protection of

ecosystems

Calendar year and winter (1 October

to 31 March)

8 µg m-3

-

Figure 4 displays the average 24 hour SO2 concentrations. There were no exceedances of the 24 hour limit value of 125 µg.m-3 or either assessment threshold. The directive stipulates that the lower assessment threshold should not be exceeded more than three times in a calendar year. Figure 5 displays hourly SO2 concentrations over the monitoring period. No exceedances of the hourly limit value of 350µgm-3 were measured.

Page 13: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

12

0.0

20

.0

40

.0

60

.0

80

.0

10

0.0

12

0.0

17/05/2008

31/05/2008

14/06/2008

28/06/2008

12/07/2008

26/07/2008

09/08/2008

23/08/2008

06/09/2008

20/09/2008

04/10/2008

18/10/2008

01/11/2008

15/11/2008

29/11/2008

13/12/2008

27/12/2008

10/01/2009

24/01/2009

07/02/2009

21/02/2009

07/03/2009

21/03/2009

04/04/2009

18/04/2009

02/05/2009

16/05/2009

30/05/2009

13/06/2009

27/06/2009

11/07/2009

ug/m3

Date

Fig

ure

4:S

ulp

hu

r Dio

xid

e 2

4-H

ou

r Av

era

ge

Co

nc

en

tratio

n, L

ette

rke

nn

y, 1

7/0

5/0

8 -

13

/07

/09

SO

2 D

aily

Ave

rage

Lim

it Valu

e

Up

per A

ssessm

ent T

hre

sho

ld

Lo

wer A

ssessm

ent T

hre

sho

ld

Page 14: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

13

0.0

50

.0

10

0.0

15

0.0

20

0.0

25

0.0

30

0.0

35

0.0

16/05/2008

30/05/2008

13/06/2008

27/06/2008

11/07/2008

25/07/2008

08/08/2008

22/08/2008

05/09/2008

19/09/2008

03/10/2008

17/10/2008

31/10/2008

14/11/2008

28/11/2008

12/12/2008

26/12/2008

09/01/2009

23/01/2009

06/02/2009

20/02/2009

06/03/2009

20/03/2009

03/04/2009

17/04/2009

01/05/2009

15/05/2009

29/05/2009

12/06/2009

26/06/2009

10/07/2009

ug/m3

Date

Fig

ure

5: S

ulp

hu

r Dio

xid

e 1

-H

ou

r Av

era

ge

Co

nc

en

tratio

ns

, Le

tterk

en

ny, 1

6/0

5/0

8 -

14

/07

/09

Ho

urly

SO

2

Ho

urly

Lim

it Valu

e

Page 15: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

14

Nitrogen Dioxide and Oxides of Nitrogen – NO2 , NOX

No. of hours 10179

Missing values 315

(including routine maintenance) 74

No. of measured values 9864

Percentage covered 96.9%

Maximum hourly value (NO2) 111.9 µg.m-3 99.7 percentile for hourly values (NO2) 76.6 µg.m-3 Mean hourly value (NO2) 13.1 µg.m-3

Mean hourly value (NOx) 22.1 µg.m-3 NO2

Directive 2008/50/EC – Nitrogen Dioxide and Oxides of Nitrogen - Limit Values

and Assessment Thresholds

Averaging Period

Limit Value

Date by which

limit value is to be

met

Hourly limit value

for the protection

of human health

1 hour

200 µg m-3 NO2 not

to be exceeded more than 18 times

a calendar year

1 January 2010

Annual limit value

for the protection

of human health

Calendar year

40 µg m-3 NO2

1 January 2010

Annual limit value

for the protection

of vegetation

Calendar year

30 µg m-3 NOx

19 July 2001

Alert threshold

400 µg m-3 NO2

over three consecutive hours

Page 16: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

15

Directive 2008/50/EC – Nitrogen Dioxide and Oxides of Nitrogen - Limit Values

and Assessment Thresholds continued

Averaging Period

Limit Value

Date by which

limit value is to be

met

Upper assessment

threshold for the

protection of

human health

1 hour

140 µg m-3 NO2 not

to be exceeded more than 18 times

a calendar year

-

Upper assessment

threshold for the

protection of

human health

Calendar year

32 µg m-3 NO2

-

Lower assessment

threshold for the

protection of

human health

1 hour

100 µg m-3 NO2 not

to be exceeded more than 18 times

a calendar year

-

Lower assessment

threshold for the

protection of

human health

Calendar year

26 µg m-3 NO2

-

Upper assessment

threshold for the

protection of

vegetation

Calendar year

24 µg m-3 NOx

-

Lower assessment

threshold for the

protection of

vegetation

Calendar year

19.5 µg m-3 NOx

-

Figure 6 displays the hourly NO2 concentrations at the site for the entire monitoring period. There was one exceedance of the lower threshold value concerning the protection of human health. No more than 18 exceedances each of the lower

Page 17: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

16

assessment threshold, upper assessment threshold and limit value are allowed per year. The mean hourly NO2 concentration was 13.1 µg.m-3 . This was below the lower assessment threshold value of 26 µg.m-3 for the protection of human health. The mean annual NOx concentration was 22.1 µg.m-3. This exceeds the lower assessment threshold concentration concerning protection of vegetation and natural ecosystems. However, this threshold applies only in rural areas and thus is not applicable to the Letterkenny site. Figure 7 displays hourly NO, NO2 and NOx concentrations in µg.m-3 . NO2 and NOX are measured as ppb (parts per billion) by volume. To convert to µg m-3, a factor (1.25 for NO, 1.91 for NO2) is used. No formula is specified for NOX, the directive requires it to be expressed as NO2 (i.e. ppb*1.91). This applies even when most of the NOX is present as NO.

Page 18: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

17

(1) Limit Value 200ug/m3 EU Directive 2008/50/EC - > 18 exceedances per year denotes limit value breach (2) Upper Assessment Threshold EU Directive 2008/50/EC - > 18 exceedances per year denotes threshold breach (3) Lower Assessment Threshold EU Directive 2008/50/EC - > 18 exceedances per year denotes threshold breach

Figure 6: Nitrogen Dioxide Hourly Concentrations, Letterkenny, 17/05/08 - 14/08/09

0

50

100

150

200

17/0

5/2

00

8

31/0

5/2

00

8

14/0

6/2

00

8

28/0

6/2

00

8

12/0

7/2

00

8

26/0

7/2

00

8

09/0

8/2

00

8

23/0

8/2

00

8

06/0

9/2

00

8

20/0

9/2

00

8

04/1

0/2

00

8

18/1

0/2

00

8

01/1

1/2

00

8

15/1

1/2

00

8

29/1

1/2

00

8

13/1

2/2

00

8

27/1

2/2

00

8

10/0

1/2

00

9

24/0

1/2

00

9

07/0

2/2

00

9

21/0

2/2

00

9

07/0

3/2

00

9

21/0

3/2

00

9

04/0

4/2

00

9

18/0

4/2

00

9

02/0

5/2

00

9

16/0

5/2

00

9

30/0

5/2

00

9

13/0

6/2

00

9

27/0

6/2

00

9

11/0

7/2

00

9

Date

ug

/m3

LV (1)

UAT (2)

LAT (3)

Page 19: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

18

0

10

0

20

0

30

0

40

0

50

0

60

0

70

0

17/05/2008

31/05/2008

14/06/2008

28/06/2008

12/07/2008

26/07/2008

09/08/2008

23/08/2008

06/09/2008

20/09/2008

04/10/2008

18/10/2008

01/11/2008

15/11/2008

29/11/2008

13/12/2008

27/12/2008

10/01/2009

24/01/2009

07/02/2009

21/02/2009

07/03/2009

21/03/2009

04/04/2009

18/04/2009

02/05/2009

16/05/2009

30/05/2009

13/06/2009

27/06/2009

11/07/2009

ug/m3

Date

Fig

ure

7: N

Ox

Ho

urly

Co

nc

en

tratio

ns

, Le

tterk

en

ny, 1

7/0

5/0

8 -

14

/08

/09

NO

x

NO

NO

2

Page 20: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

19

Particulate Matter – PM10

PM10 : TEOM method

No. of days 353 Missing values 41 (including routine maintenance) 6 No. of measured values 312 Percentage covered 88.4% Maximum daily value 160.2 µg.m-3 Mean daily value 18.0 µg.m-3

Directive 2008/50/EC – Particulate Matter - Limit Values and Assessment

Thresholds

Averaging Period

Limit Value

24-hour limit

value for the

protection of

human health

24 hour

50 µg m-3 PM10 not to be

exceeded more than 35 times a calendar year

Annual limit value

for the protection

of human health

Calendar year

40 µg m-3 PM10

Upper assessment

threshold for the

protection of

human health

24 hour

35 µg m-3 PM10 not to be

exceeded more than 35 times a calendar year

Upper assessment

threshold for the

protection of

human health

Calendar year

28 µg m-3 PM10

Page 21: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

20

Directive 2008/50/EC – Particulate Matter - Limit Values and Assessment

Thresholds, Continued

Averaging Period

Limit Value

Lower assessment

threshold for the

protection of

human health

24 hour

25 µg m-3 PM10 not to be

exceeded more than 35 times a calendar year

Lower assessment

threshold for the

protection of

human health

Calendar year

20 µg m-3 PM10

Figure 8 displays 24 hour average concentration of PM10 at the site. The 24 hour limit value for the protection of human health (50 µg.m-3) was exceeded 15 times during the measurement period. The directive stipulates that the limit value should not be exceeded more than 35 times in a calendar year. The upper assessment threshold (35 µg.m-3) was exceeded on 33 days, the lower assessment threshold (25 µg.m-3) was exceeded on 53 days. The directive stipulates that each of the assessment thresholds should not be exceeded more than 35 times in a calendar year. The mean of the daily values during the measurement period (18.0 µg.m-3) is below the annual limit value for the protection of human health (40 µg.m-3).

Page 22: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

21

(1) Limit Value 50ug/m3 EU Directive 2008/50/EC , > 35 exceedances of daily limit value in a year denotes an exceedance (2) Upper Assessment Threshold EU Directive 2008/50/EC, > 35 exceedances denotes threshold breach (3) Lower Assessment Threshold EU Directive 2008/50/EC, > 35 exceedances denotes threshold breach

0

20

40

60

80

100

120

140

160

18026

/07/

2008

09/0

8/20

08

23/0

8/20

08

06/0

9/20

08

20/0

9/20

08

04/1

0/20

08

18/1

0/20

08

01/1

1/20

08

15/1

1/20

08

29/1

1/20

08

13/1

2/20

08

27/1

2/20

08

10/0

1/20

09

24/0

1/20

09

07/0

2/20

09

21/0

2/20

09

07/0

3/20

09

21/0

3/20

09

04/0

4/20

09

18/0

4/20

09

02/0

5/20

09

16/0

5/20

09

30/0

5/20

09

13/0

6/20

09

27/0

6/20

09

11/0

7/20

09

ug

/m3

Date

Figure 8: Daily PM10 Concentrations, Letterkenny 26/07/08 - 13/08/09

LV (1)

UAT (2)

LAT (3)

Page 23: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

22

Lead - Pb

No. of days 455

Missing days 92

(including routine maintenance) 92

No. of measured days 363

Percentage covered 79.8%

Concentration of Pb 0.003 µg.m-3

Directive 2008/50/EC – Lead - Limit Values and Assessment Thresholds

Averaging Period

Limit Value

Date by which

limit value is to be

met

Annual limit value

for the protection

of human health

Calendar year

0.5 µg m-3

1 January 2005

Upper assessment

threshold

Calendar year

0.35 µg m-3

Lower assessment

threshold

Calendar year

0.25 µg m-3

The concentration of lead during the measurement period was well below the lower assessment threshold.

Page 24: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

23

Arsenic(As), Cadmium(Cd) and Nickel(Ni):

No. of days 455

Missing days 92

(including routine maintenance) 92

No. of measured days 363

Percentage covered 79.8%

Concentration of As

Concentration of Cd Concentration of Ni

0.2 0.05 3.6

ng.m-3 ng.m-3 ng.m-3

Directive 2004/107/EC – Arsenic, cadmium and nickel - Limit Values and

Assessment Thresholds

Averaging Period

Target Value

Upper Assessment

Threshold , Lower

Assessment Threshold

Arsenic

Calendar year

6 ng m-3

3.6 ng m-3 , 2.4 ng m-3

Cadmium

Calendar year

5 ng m-3

3.0 ng m-3 , 2.0 ng m-3

Nickel

Calendar year

20 µg m-3

14.0 ng m-3 , 10.0 ng m-3

Four metals in addition to lead must be considered when carrying out an air quality assessment These are cadmium, arsenic, nickel and mercury . Limit values and measurement methods for these metals as well as certain polycyclic aromatic hydrocarbons are set out in the fourth Daughter Directive (2004/107/EC). An indicative method was used during this assessment to measure prevailing concentrations of cadmium, nickel and arsenic in air. This method is detailed above and essentially involves pumping air through a filter for several weeks before digesting the filter and analysing the digest for lead and other metals using ICP-MS. With this method, the detection limit is influenced by any traces of metal in the filter paper as well as by the volume of air passed through the filter. The results, although indicative, do provide some indication of the concentrations of these metals in air. The average concentration of arsenic measured in air during the assessment period was 0.2 ng.m-3 . The target value is 6 ng m-3

Page 25: Ambient Air Monitoring At Letterkenny, Co. Donegal final rep… · Ambient air was pumped through a Metricel membrane filter (Gelman, 37mm, 0.8 µm) situated in a calming chamber

24

The average concentration of cadmium measured in air during the assessment period was 0.05 ng.m-3 . The target value is 5 ng m-3 The average concentration of nickel measured in air during the assessment period was 3.6 ng.m-3 . The target value is 20 ng m-3 The target values of 6 ng m-3 for arsenic, 5 ng m-3 for cadmium and 20 ng m-3 for nickel have cross Europe attainment target dates of 31st of December 2012