Zen and the Art of On-Wafer Probing†
A Personal Perspective
Rob Sloan School E&EE, University of Manchester
† - after Robert Pirsig
Device or Circuit Measurement at Microwave/ Millimetre-wave/ THz frequencies
On-wafer Probes, typically GSG, connected to SMUs and VNAs
Device Modelling
From Physical layers Engineering layers
Solid-state Power vs. frequency
Several high frequency processes and research programmes, dot7, DARPA THz Transistor, Teledyne, HRL, NGST, Infineon…. Designing into these processes requires good models got from good device measurements.
20 dB of gain at 670 GHz
Fabricated by Northrop Grumman under DARPA THz Electronics Program*
On-Wafer Probing - Calibration
In measuring a device we need to calibrate the network analyser. This removes the systematic errors associated with cables and the receiver. Usual on-wafer model is based upon a 16-term error model which can be simplified if the two leakage terms are is zeroed and the source switch is behind the reflectometers usually down to an 8-term error model (or 7 normalised).
DUT
S11
S12
S22
S21
e20
e13
e10
e01
e00
e11
e30
e03
e23
e21
e12
e22
e33
e32
e31
e02
a0
a0
a1
a1
b2
b1
a2
b2
b3
b0
a3
b3
b0
b1
a3
a2
One of the 16 error terms can be normalized to yield 15 error terms
e00
, e33
Directivity
e11
, e22
Port Match
e10
, e01
, e32
, e23
Tracking
e30
, e03
Primary Leakage
All others are lower level
leakage paths
16-Term Error Model1 Errors Corrected: • Ports match • Coupler directivity • Tracking • Leakage
1. JV Butler, DK Rytting, MF Iskander, RD Pollard and MV Bosche, ‘16-Term Error Model and Calibration Procedure for On-Wafer Network Analysis Measurements’, IEEE Trans. on Microwave Theory & Tech. VOL. 39, NO. 12. 1991, pp. 2211 – 2217.
Summary of Calibration Methods for on-wafer
probing
• Based on the 8-term error model; seven or more known conditions must be measured for calibration. Z0 must be known.
SOLT
(Standard Cal
for VNAs)
Known Thru (T)
[4 Terms]
3 known Reflects (R)
(Open, Short & Load) on Port 1
[3 Terms]
3 known Reflects (R)
(Open, Short & Load) on Port
2
[3 Terms]
LRM Known Thru (T)
or Line (L)
[4 Terms]
Unknown but equal (symmetry)
Reflect (R) on Port-1 & Port-2
[1 Term]
Known Match (Load) (M) on
Port-1 & Port-2
[2 Terms]
LRRM &
eLRRM*
Known Thru (T)
or Line (L)
[4 Terms]
2 Unknown but equal (symmetry)
Reflects (RR – open & short) on
Port-1 & Port-2
[2 Terms]
Known Match (Load) (M) on
Port-1
[1 Term]
TRL & mTRL
Known Thru (T)
[4 Terms]
Unknown but equal (symmetry)
Reflect (R) on Port-1 & Port-2
[1 Term]
Known (S11 and S22) Line or
multiple Lines (L)
[2 Terms]
* - usually through Cascade Microtech’s WinCAL
On-Wafer Probing - Calibration
In measuring a transistor calibrating the VNA removes the systematic errors associated with cables and the test setup and supposedly the leakage.
Probes – GGB Picoprobe, Cascade Microtech plus loads of others, down to 50µm pin/contact spacing and to greater than 1THz Can be GSG but also almost any other combination, eg. balanced
On-Wafer Probes
GSG – ground signal ground, 3 pins
Coax connector eg, 1mm. RF plus DC feed via VNA and Bias-T
Choice of reference plane for RFOW Probing ? Probe tip vs. On-wafer at the DUT
Probe tip calibration
On Wafer calibration
100µm Probe pitch
Probe Tip Cal Using ISS – Impedance Standard Substrate
• Example here is often used for LRRM - the industry standard. Owned by Cascade Microtech and is often used in conjunction with their software WinCAL. Menu driven.
Uses a standard SOLT set of standards for probe tip calibration.
Calibration Checklist for Probe Tip Cal Using ISS –
Impedance Standard Substrate
• Switch phone off. It’s really important not to do anything else when doing RFOW cal and measurement.
• Probes are expensive and not easily repaired. And an wafer may be a lot more! • Visually check and clean probes • Mount probes on prober if necessary. • Planarise, 3 even scratches GSG (contact substrate for planarity) or on wafer area.
• Then set the probe skate.
Planarity 3 even
scratches
Probe tip Cal on ISS: LRRM – Line, 2x Reflect, Match
Short Standard 1 Reflect [R]
Thru or Line Standard, 1ps [L]
Match Standard [M]
Open standard is taken as the two probes lifted off the ISS tile > 250um and spaced sufficiently far apart >1mm 1 Reflect [R]
• Set the skate using the alignment marks from X to Y. This distance depends upon the probes used.
Good calibration
Good device/ circuit measurements
Good Modelling & design
Good working circuits On-wafer Probing
Virtuous Circle for MMIC Design
So what’s it got to do with Zen & the Art…?
• It’s about QUALITY in this case • Whether it’s maintaining a motorbike or making a measurement right attitude yields
quality. In our case a good calibration. That gets good measurements and then good device models.
However for Millimetre-wave DUTs there’s more work to do….
So if you want to measure a transistor/diode right up to the DUT there are 2 ways: 1. ISS Cal and then de-embed back to the deviceǂ. The problem with this are the
parasitics. The error associated with de-embedding far outweighs the device parasitics that you’re trying to measure.
Probe tip cal plane
De-embed De-embed ǂ H Cho and , DE Burk, ‘A three-step method for the de-embedding of high-frequency S-parameter measurements’ , IEEE Trans. on Electron Devices. Vol. 38, No. 6. 1991, pp. 1371 – 1375.
On-wafer Calibration Standards
2. Use on-wafer cal standards. LRRM or multiline TRL. Brings the cal plane up to the device and eliminates the need to de-embed the structures connecting to the device.
On-wafer cal plane
On-wafer Calibration Standards
• LRRM is a popular choice since the standards are compact and take up little valuable wafer space.
• Multiline TRL is a better cal standard for millimeter-wave frequencies and there is free software available from NIST to do mTRL. However mTRL requires longer lengths of line, which is more wafer space! For more information see ‘Prescription for THz Transistor Characterisation’, Dylan Williams, NIST, Boulder, Colorado.
• Guidelines on Cal standard spacing between circuits should be >5x the substrate thickness. CPW circuits screen better than microstrip but bring with them a whole bunch of modal issues. Also spacing between standards depends upon the probe used – check with the manufacturer.
Angle of probe and use of absorber
On-wafer Cal Standards - Thru
On a 50µm GaAs pHEMT process using ADS & simulated using Momentum. Pp10 WIN Semiconductor. Then check using LRRM cal on ISS.
On-wafer Cal Standards - Open
50µm GaAs pHEMT process pp10 WIN Semiconductor. Checked using LRRM cal on ISS tile.
On-wafer Cal Standards - Short
50µm GaAs pHEMT process pp10 WIN Semiconductor. Checked using LRRM cal on ISS tile.
Pull in of higher frequency resonance
On-wafer Cal Standards - Load
Notice this is not a particularly good load above ~ 60GHz, via inductance is becoming a feature. There are tricks that can be employed to improve the appearance of a good load.
50µm GaAs pHEMT process pp10 WIN Semiconductor. Checked using LRRM cal on ISS tile.
Finally the On-wafer Cal can be done and DUT can be measured
• Calibrate then check by re-measuring the standards. LRRM is self consistent so that the shorts do not have to sit at zero Ohms and can on occasion appear to give gain especially at frequencies above 80GHz!
Note these small devices have huge gain at low frequencies so in measuring small-signal we mustn’t drive the device too hard. RF signal level for modelling transistors is about -40dBm (VNA -30dBm plus cable/probe losses) or around 2mV RMS or pk-pk ~6mV. Usually small signal1 is VT/10.
1. ICCAP Manual, Franz Sischka, October 2010
Future of On-Wafer Cal
• LRRM and MTRL are the way forward. I can see a frequency partition calibration DC -80GHz LRRM and >80GHz mTRL. This requires an extra set of loads to be designed on wafer but saves the space of the lines needed for low frequency cal.
Example of Multiline TRL Kit: M Ferndahl, K Andersson & C Fager, ‘Mulit-line TRL Calibration
Compared to a General De-embedding Method’, 73rd ARFTG, 2009
Good calibration good models complex circuit design
For TWA accurate models allow for designs to be optimised and kept stable. 10 stages of cascode.
Conclusions
• There’s a lot of work to do… mm-wave devices are here and need exploiting! Devices exist capable of working >400GHz. The applications will follow and devices and pull outs need measuring and characterising.
Co-workers: Dr Chris Duff, Miss Priya Shinghal, Mr Geoffrey Barigye (UoM) And Finally,
Conclusions
• The zen master and the university professor vs. • Zen and the student
“In the beginner’s mind there are many possibilities, but in the expert’s there are few.” ― Shunryu Suzuki, Zen Mind, Beginner’s Mind