A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 [email protected] Eleventh International Magnet Measurement Workshop Eleventh International Magnet Measurement Workshop September 21 - 24, 1999 September 21 - 24, 1999 Brookhaven National Laboratory, Upton, New York, USA Brookhaven National Laboratory, Upton, New York, USA A Mole for warm magnetic and A Mole for warm magnetic and optical measurements of LHC optical measurements of LHC dipoles dipoles 1 Thursday, 23 Sept.1999 Berkner B 10:50-11:20 L Bottura 1 , M Buzio 1 , G Deferne 1 , H Jansen 2 , C Glöckner 2 , A Köster 2 , P Legrand 1 , A Rijllart 1 , P Sievers 1 , F Villar 1 1 CERN, European Organization for Nuclear Research - 1211 Geneva 23, Switzerland 2 Frauenhofer Institut IPT, Aachen, Germany 1 Also published in the proceedings of the 16 th International Conference on Magnet Technology, Ponte Vedra Beach, IL, 26 Sept- 2 Oct 1999

Upload: clifton-lamb

Post on 11-Jan-2016

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Eleventh International Magnet Measurement WorkshopEleventh International Magnet Measurement WorkshopSeptember 21 - 24, 1999September 21 - 24, 1999

Brookhaven National Laboratory, Upton, New York, USABrookhaven National Laboratory, Upton, New York, USA

A Mole for warm magnetic and optical A Mole for warm magnetic and optical measurements of LHC dipolesmeasurements of LHC dipoles11

Thursday, 23 Sept.1999Berkner B

10:50-11:20

L Bottura1, M Buzio1, G Deferne1, H Jansen2, C Glöckner2, A Köster2,

P Legrand1, A Rijllart1, P Sievers1 , F Villar1

1 CERN, European Organization for Nuclear Research - 1211 Geneva 23, Switzerland2

Frauenhofer Institut IPT, Aachen, Germany

1Also published in the proceedings of the 16th International Conference on Magnet Technology, Ponte Vedra Beach, IL, 26 Sept- 2 Oct 1999

Page 2: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

IntroductionIntroduction Mechanical layoutMechanical layout VR DemoVR Demo11

Magnetic performance Magnetic performance Optical performanceOptical performance ConclusionsConclusions

A Mole for Warm Magnetic and Optical A Mole for Warm Magnetic and Optical Measurements of LHC DipolesMeasurements of LHC Dipoles

A Mole for Warm Magnetic and Optical A Mole for Warm Magnetic and Optical Measurements of LHC DipolesMeasurements of LHC Dipoles

1 http://home.cern.ch/m/mtauser/www/applications/mole/virtualmole.wrl

Page 3: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

IntroductionIntroductionIntroductionIntroduction

Why use a mole ?Why use a mole ? Purpose: measure magnetic axis Cold measurements done for max. efficiency with 15 m long coil shaft, obstructing line-of-sight to the coil center travelling probe is needed Additional duties: warm field quality of all dipoles, cold field of samples

Requirements for series measurementsRequirements for series measurements Coil center position: precision better than 0.1 mm Warm measurements: 50 mm, B 25mT Rugged, quick and simple to setup and operate As many off-the-shelf components as possible

Page 4: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Mole test bench overviewMole test bench overviewMole test bench overviewMole test bench overview

rear reference bench(transport motor, cable connections)

magnet

front reference bench(telescope)

Page 5: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Front reference benchFront reference benchFront reference benchFront reference bench

CCD camera

Telescope

mole support shell (closed)

Sliding support

focusing system

Page 6: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Rear reference benchRear reference benchRear reference benchRear reference bench

Transport motor

Cable receptacle

Axis transfer jigMole

Page 7: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Front reference benchFront reference benchFront reference benchFront reference bench

Laser for alignment measurement

optical alignment targets (4)

axis transfer carbon fiber jig

Telescope/CCD/DSP unit

transport belt

granite table

Page 8: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

StatusStatusStatusStatus Warm Mole Warm Mole

First unit accepted and under commissioning Two more units to come, will perform warm test of all LHC dipoles Can be readily adapted to warm quadrupole tests

Cold MoleCold Mole Order started, last design details now being finalized Will perform cold tests for a sample populations of LHC dipoles to get warm-cold axis correlation Further units may be slightly modified for cold quadrupole tests

Page 9: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Mechanical layoutMechanical layoutMechanical layoutMechanical layout

Main featuresMain features three mmlong radial coils coil drive: Shinsei travelling-wave ultrasonic piezo motor 4096 CPR Heidenhein angular encoder On-board feedback auto-leveling system Spring-loaded roller system Motorized traction belt drive Two granite benches to establish reference positions LED+lens system to create virtual light-spot in the coil centre Telescope + CCD camera + DSP to measure light-spot position All functions remotely controlled via RS232 Auxiliary optical system to transfer axis to magnet fiducials

Page 10: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Main components Main components Main components Main components

Coil

Elastic coupling(with helical groove for the cable)

Ball-bearing support flange

Piezo motorWyler encoder

Potentiometer

3:1 reductiongearbox

Page 11: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Longitudinal transportLongitudinal transportLongitudinal transportLongitudinal transport

closed-loop pre-tensioned Inconel transport belt

feedback-controlled DC motor drive

referencing given by optical barriers at both ends of the magnet (self-calibration at the end of each trip)

max. velocity about 1 m/s

precision of positioning better than 0.5 mm

mole must be switched manually between apertures

Page 12: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Longitudinal reference positionLongitudinal reference positionLongitudinal reference positionLongitudinal reference position

lower half-shellsupport

Optical barrier(receptor)

spring-loadedroller

LED viewingaperture

front transport belt attachment

Page 13: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

SafetySafetySafetySafety

interlock system: mole will stop if covers are removed, transport cable gets tangled, transport motor drains overcurrent

reduced velocity close to the edges

longitudinal position cross-checked with two independent measurements (angular encoder + linear potentiometer)

Page 14: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

AutolevelingAutolevelingAutolevelingAutoleveling

Why autoleveling ?Why autoleveling ? imperfections of beam pipe/rollers rotation of container during travel observed rotation up to 2~3°/m transport belt & cable may be twisted coil must be level prior to measurement

How does it work ?

DC motor engaged in container to rotate coil+encoder+piezo motor feedback system using wide-range level meter precision achieved better than 4 mrad time needed less than 5 s autoleveling done at the end of each longitudinal movement

Page 15: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Auto-leveling mechanism Auto-leveling mechanism Auto-leveling mechanism Auto-leveling mechanism

Motor shaft gear

Wide-rangelevel meter Roller (fixed tangentially

by friction)

Inner gear (linked with container)

Page 16: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Auto-leveling mechanism Auto-leveling mechanism Auto-leveling mechanism Auto-leveling mechanism

Final configurationStart End of travelling

travel autolevel

Page 17: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Optical measurementOptical measurementOptical measurementOptical measurement

TelescopeTelescope accurately leveled and aligned wrt magnet mounted on rails to serve both apertures

CCD cameraCCD camera 752 pixel, 6.00 mm size self-aligned to telescope via reference marks

DSP image processorDSP image processor threshold filtering + center of gravity calculation measurements internally averaged over multiple takes software calibration map

Page 18: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Virtual lightspotVirtual lightspotVirtual lightspotVirtual lightspot

Virtual lightspot

LED

the virtual image remains in the coil center even when the mole rotates

Magnifying lens

Page 19: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Optical measurementOptical measurementOptical measurementOptical measurement

PerformancePerformance mm mm 20 m range total error 60 m measurement time 4 s at closest range

Page 20: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Optical system calibrationOptical system calibrationOptical system calibrationOptical system calibration

Mole XY calibration check (z=20m)

-3.5

-2.5

-1.5

-0.5

0.5

1.5

2.5

3.5

-12 -10 -8 -6 -4 -2 0 2 4 6 8 10 12

x [mm]

y [

mm

]

XYTable

Optical

Page 21: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Optical system calibrationOptical system calibrationOptical system calibrationOptical system calibration

Measured X Error vs. y position (z=20 m) (zoomed)

-80

-60

-40

-20

0

20

40

60

80

100

120

140

-4 -2 0 2 4

y [mm]

dx

[

m]

x=-10

x=-8

x=-6

x=-4

x=-2

x=2

x=4

x=6

x=8

x=10

Page 22: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Example: optical measurement Example: optical measurement Example: optical measurement Example: optical measurement

MBL1N1 beam pipe shapeLeft aperture, X (horizontal) coordinate

(avg. diff=74 m)

0100020003000400050006000700080009000

0 5 10 15 20

z [m]

x [

mic

ron

]

LHC/MTA Mole

EST/SU Mole

Page 23: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Focusing systemFocusing systemFocusing systemFocusing system

How does it workHow does it work no autofocus needed longitudinal position sent to the camera at the end of each movement focusing motor follows a software calibration table total error if out-of-focus 1 m/mm

Page 24: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Magnetic performanceMagnetic performanceMagnetic performanceMagnetic performance

Harmonic CoilsHarmonic Coils mm fiberglass/epoxy coil support 3 radial coils with 400 turns, 3.4 m2 area 1 coil for main dipole, 1 for dipole compensation + 1 spare

System precisionSystem precision Main dipole: better than 10-4

Dipole angle: better than 0.2 mrad Multipoles: generally better than 2·10-6 relative to B1 bucking factor 2000

Page 25: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Example: multipoles Example: multipoles Example: multipoles Example: multipoles

F ie ld h a rm o n ic s in A p . 1 o f M B L 1 N 1 @ I = 3 0 Al ine a rly c o rre c te d fo r fe e d do wn - a ll units @ 1 7 m m

0 .0 0 0 1

0 .0 0 1

0 .0 1

0 .1

1

1 0

1 0 0

Page 26: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Example: calibrated dipole angle Example: calibrated dipole angle Example: calibrated dipole angle Example: calibrated dipole angle

MBL1N1 Field Angle - Aperture 1

0.00

1.00

2.00

3.00

4.00

5.00

6.00

7.00

8.00

9.00

0 2000 4000 6000 8000 10000 12000 14000

z (mm)

an

gle

(m

rad

)

Mole (calibrated) 10/5/99, ±30A

Chaconsa scanning machine18/5/97, ±15A

Page 27: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Axis TransferAxis TransferAxis TransferAxis Transfer

Why transfer the magnetic axis ? Why transfer the magnetic axis ?

harmonics, and hence magnetic axis, are measured in the coil reference frame the center of the coil is measured in the telescope reference frame the magnetic axis is needed in the reference frame of the magnet fiducials

How the transfer will be doneHow the transfer will be donethe granite tables establish two reference positions per aperture,

vertically and laterally reproducible within 20 m reference positions are measured by the telescope relative position of magnet fiducials and granite tables is measured by means of an auxiliary laser system relative position of coil center wrt magnet fiducials can thus be computed

Page 28: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Axis TransferAxis TransferAxis TransferAxis Transfer

Telescope frame

Mole coil positionalong magnet

Granite table jig

Granite table reference line

Magnet

fiducial line

Granite jig reference line

Auxiliary alignment laser beam

Page 29: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Axis TransferAxis TransferAxis TransferAxis Transfer

Telescope reference frame

Mole coil position

Magnetic center (z)

Marble table jigs line

Magnet fiducial line

Marble table reference line

Y

Z

Page 30: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Problems encounteredProblems encounteredProblems encounteredProblems encountered

Unsteady velocity of piezo-motor, getting worse with motor heatingUnsteady velocity of piezo-motor, getting worse with motor heating Micro cables and connectors delicate and difficult to handleMicro cables and connectors delicate and difficult to handle Fine alignment of the lensFine alignment of the lens Image distortion due to thermal gradients in the magnetImage distortion due to thermal gradients in the magnet

Page 31: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Piezo motor rotation speedPiezo motor rotation speedPiezo motor rotation speedPiezo motor rotation speed

Rotation speed quality Rotation speed quality

encoder detects low-frequency rotation speed oscillations up to 8% oscillation amplitude gets larger as motor heats up oscillations interfere with measured harmonics (errors in b2 up to 0.5 units

observed)

Improvements under wayImprovements under wayoptimization of stiffness and damping of the coil/motor link

to minimize response fine tuning of motor speed map (t) to minimize excitationenhanced motor cooling automatic off-line filtering of spurious harmonics

Page 32: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Piezo motor rotation qualityPiezo motor rotation qualityPiezo motor rotation qualityPiezo motor rotation quality

Backward run

Forward run

average

Page 33: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Piezo motor rotation qualityPiezo motor rotation qualityPiezo motor rotation qualityPiezo motor rotation quality

Mole Rotation & Magnetic Measurement Test

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

1 11 21 31 41 51 61 71 81 91 101 111 121 131 141

M e asure me nt n.

b2

Speed variation p.p. (FW D) (rad/s )

-b3

Speed variation p.p. (BW D) (rad/s )

Page 34: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

Piezo motor rotation qualityPiezo motor rotation qualityPiezo motor rotation qualityPiezo motor rotation quality

Mole Rotation & Magnetic Measurement Test

-0.3

-0.2

-0.1

0

0.1

0.2

0.3

0.4

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7

Amplitude of lowest frequency component in(t) (rad/s)

Dev

iatio

n fr

om

avg

. of h

arm

on

ic (u

nit)

b3-b3avg

b2-b2avg (f=4 Hz)

b2-b2avg (f=2 Hz)

Page 35: A mole for warm magnetic and optical measurements of LHC dipoles IMMW XI, Brookhaven National Laboratory, 21-24 Sept. 1999 marco.buzio@cern.ch Eleventh

A mole for warm magnetic and optical measurements of LHC dipoles

IMMW XI, Brookhaven National Laboratory, 21-24 Sept. [email protected]

ConclusionsConclusionsConclusionsConclusions

Magnetic and optical performance generally exceeding specificationsMagnetic and optical performance generally exceeding specifications Well suited for industrial-type test environment Well suited for industrial-type test environment Test bed for further technological improvements (piezo-motor, remote Test bed for further technological improvements (piezo-motor, remote

encoders, etc.)encoders, etc.) Software integration under waySoftware integration under way Minor improvements under way: smoother velocity control, better optics Minor improvements under way: smoother velocity control, better optics

alignmentalignment