Page 1: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

1

Liquid Injection Techniques

• Flash Vaporization

–Split injection

–Splitless injection

–Direct - Uniliner w/Hole

Page 2: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

2

Flash Vaporization - Split Injection

Advantages:

• Easy to use

• Easy to automate

• Ideal for dirty samples

Disadvantages:

• Sample degradation

• Backflash

• Discrimination

• Not a trace technique

Page 3: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

3

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

Injection Port ComponentsInjection Port Injection Port ComponentsComponents

Flash VaporizationSplit Injection

Page 4: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

3

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

Injection Port ComponentsInjection Port Injection Port ComponentsComponents

Flash VaporizationSplit Injection

Liner or ‘Sleeve’

Flash VaporizationSplit Injection

Page 5: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

3

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

Injection Port ComponentsInjection Port Injection Port ComponentsComponents

Flash VaporizationSplit Injection

Septum

Page 6: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

3

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

Injection Port ComponentsInjection Port Injection Port ComponentsComponents

Flash VaporizationSplit Injection

Inlet Seal

Flash VaporizationSplit Injection

Page 7: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

3

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

Injection Port ComponentsInjection Port Injection Port ComponentsComponents

Flash VaporizationSplit Injection

Carrier Gas In

Page 8: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

3

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

Injection Port ComponentsInjection Port Injection Port ComponentsComponents

Flash VaporizationSplit Injection

Septum purge

Page 9: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

3

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

Injection Port ComponentsInjection Port Injection Port ComponentsComponents

Flash VaporizationSplit Injection

Solenoid Valve

Page 10: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

3

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

Injection Port ComponentsInjection Port Injection Port ComponentsComponents

Flash VaporizationSplit Injection

To Split Vent

Flash VaporizationSplit Injection

Page 11: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

4

0 10

25:1 Split25:1 Split

0 10

50:1 Split50:1 Split

Flash Vaporization - Split Injection

Increasing the split ratio decreases the peak area, if all other variables are equal.

Page 12: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

5

Laminar Cup Cyclosplitter®

Split Injection Liner Designs

Split with wool or CarboFrit™

Use only a small amount of wool

Use only a small amount of wool Excellent

vaporization, expensive

Excellent vaporization, expensive

Keeps contamination out of column

Keeps contamination out of column

Page 13: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

6

Flash Vaporization - Splitless Injection

• Used for trace analysis

• Mode of operation of a split injector

• Slow sample transfer to column– Long contact time with the liner

promotes vaporization

– Long initial sample bandwidth, sosolvent focusing is important

– Requires careful optimization

Page 14: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

7

Column Head Pressure

0 100

Split Vent Purge Vent

Total Flow Septum Purge

SplitlessInjection

Solenoid valve controls the flow

Page 15: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

8

Splitless Injection Characteristics

• Slow sample velocity through liner

• Long residence time can lead to compound breakdown– Injection port inertness is critical

• Requires careful optimization – Oven temperature, solvent type, liner volume, solenoid

hold time...

Page 16: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

9

Splitless InjectionFactors Affecting Performance

• Hold time

• Solvent focusing:– Solvent boiling point

– Solvent polarity

• Incomplete sample vaporization due to rapid auto injection

• Surface deactivation

Page 17: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

10

3 min1.5 min45 sec30 sec

Factors Affecting Splitless InjectionHold Times

Column ID(mm)

0.180.250.320.53

Note: based on a 2µL injection volume of CH2Cl2 = 0.8 mL sampleexpansion value @ 250°C & 10 psig.

Column Flow Rate (cc/min.)

0.30.71.22.6

He

Approx.Hold Time

Determineempirically

Page 18: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

11

Factors Affecting Splitless InjectionHold Times

0 10 20 0 10 20

Hold Too LongHold Too Long ——no purgeno purge

Proper Hold TimeProper Hold Timewith purgewith purge

Time (min)

Page 19: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

12

Factors Affecting Splitless InjectionHold Time Optimization

Sol

vent

Are

a (m

illio

ns)

Com

ponent Area (thousands)

0.32mm ID

Hold Time (sec.)

+ Solvent

5

4

3

2

1

0

20

16

12

8

4

00 15 30 45 60 75 90 105 120

• High MW Component

Page 20: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

13

pentane 36methylene chloride 40acetone 56chloroform 62methanol 64.5hexane 691,1,1 trichloroethane 74.1ethyl acetate 77ethanol 78.3

(°C)

Subambient cooling will improve chromatography

Subambient cooling will improve chromatography

Note: When the first component Note: When the first component has a much higher bp than the has a much higher bp than the solvent, you may inject above the solvent, you may inject above the bp of the solvent (Analyte bp of the solvent (Analyte Focusing, e.g., analysis of Focusing, e.g., analysis of chlorinated pesticides)chlorinated pesticides)

Factors Affecting Splitless InjectionSolvent Focusing, cont’d

Page 21: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

14

Mismatched Polarity: The solvent will bead or puddle, causing non-

uniform analyte focusing and split peaks

Polarity Match:The solvent wets the stationary phase

Solvent FocusingSolvent Polarity, cont’d

Page 22: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

15

Beading causes Beading causes nonnon--uniform analyte uniform analyte focusingfocusing

Split Peaks Split Peaks

0 1 2 min.

Solvent FocusingSolvent Polarity

Page 23: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

16

Splitless Liner Designs

Straight GooseneckDouble

GooseneckCyclo Double-

Gooseneck

Page 24: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

Split/splitless Injection using theDrilled Uniliner

• For trace analysis

• Inlet sleeve has a press-fit connection with column at bottom of sleeve

• More inert sample pathway

• Helps eliminate injection port discrimination

Page 25: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

18

Drilled Uniliner w/ Hole

• Allows DI and Splitless injection methods

• Minimizes injection port discrimination

• Reduces loss of active compounds for more accurate results

• Hole can be placed on bottom or top of liner (each has benefits)Column seals via

a press tight connection

Page 26: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

19

Drilled Uniliners

4mm IP deactivated

4mm Siltek deactivated

2mm Siltek deactivated

Page 27: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

20

Injection Port LinerDrilled Uniliner w/ hole

The Uniliner with a hole is a unique liner that can be used for both direct and splitless injection. The column is connected at the bottom of the liner, eliminating any sample contact with metal below the liner. In order for the carrier gas to be routed through the split vent, a hole has been placed on the side of the liner. This hole allows the carrier gas to be vented through the split vent line during the split operation of the injection port.

Page 28: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

21

Injection Port LinerDrilled Uniliner w/ hole

There are two styles of Drilled Uniliner® liners, one with a hole near the top, and a second with a hole near the bottom of the liner. The choice of liner is dependent on

how closely the first compound elutes to the solvent peak. If the first compound elutes near the solvent peak, as in Method 8270, the drilled Uniliner® with the hole at the bottom of the liner should be used. This allows all the solvent to be flushed out of the entire inlet liner, reducing any solvent tailing, which could interfere with the first

eluting compounds. The drilled Uniliner® with the hole near the top of the liner works well when the first compounds elute away from the solvent peak as in ethylene glycol

in water, or chlorinated pesticide analysis.

The Drilled Uniliner ® also helps eliminate injection port discrimination for the late eluting PAHs. This is shown in figures 5A and 5B. Figure 5A shows the difference between a pressure pulsed injection versus constant flow for the single

gooseneck liner. Figure 5B shows the difference between the Drilled Uniliner ® and single gooseneck liner under constant flow condition. The Drilled Uniliner ® exhibits

the least amount of injection port discrimination.

Page 29: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

22

Figure 5a - Single Gooseneck Liner(Constant Flow vs Pressure Pulsed Injection)

White = Single Gooseneck LinerWhite = Single Gooseneck Liner

Pressure Pulsed InjectionPressure Pulsed Injection

Red = Single Gooseneck LinerRed = Single Gooseneck Liner

Constant FlowConstant Flow

Page 30: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

23

Figure 5b - Single Gooseneck vs Drilled Uniliner® Sleeve(Constant Flow)

White = Drilled UnilinerWhite = Drilled Uniliner

Red = Single Gooseneck LinerRed = Single Gooseneck Liner

Page 31: • Flash Vaporization –Split injection –Splitless …Flash Vaporization - Splitless Injection • Used for trace analysis • Mode of operation of a split injector • Slow sample

24

Leak Detectors