2006md 7 - troubleshooting

Upload: skvemula

Post on 08-Apr-2018

225 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/7/2019 2006MD 7 - Troubleshooting

    1/17

    1

    HPLC

    Troubleshooting

    A.NARENDER

    Asst.Service Manager

    Spinco Biotech Pvt Ltd.

    2

    Daily Maintenance

    Pre-analysis checks (1)

    Filter mobile phase

    Wait for mobile phase to reach room temperature

    Purge all the flow lines with mobile phase

    Replace mobile phases as necessary

    If buffer is used as mobile phase, wash the back of plungerseal

    If autosampler is used, verify that the autosampler rinse bottle is

    full

    Purge the autosampler rinse solution

    Check for leaks

    Check the pump pressure

    Check the oven temperature

    Perform a baseline check

  • 8/7/2019 2006MD 7 - Troubleshooting

    2/17

    2

    3

    Daily Maintenance

    Post-analysis checks

    Wash the column

    If the column is not going to be used for long time, take off it

    from the system after washing, and cap both ends of the

    column, store it in cool and dark place

    Clean the flow line or the rest of the instrument, as necessary

    Store the solvent reservoir filters in rinse solvent to prevent

    them from drying out

    4

    Outline

    - Troubleshooting Process

    - Categories of Problems

    Pressure Issues

    Baseline Issues

    Peak Shape Issues

    Retention Issues

    - Troubleshooting Examples

  • 8/7/2019 2006MD 7 - Troubleshooting

    3/17

    3

    5

    Troubleshooting Process

    Gather the facts not theories.

    Compare the performance obtained to the expected performance.

    List possible causes.

    Check the simplest things first its easier.

    Work through the possible causes in a step-by-step manner

    checking the outcome from any changes made.

    As a last resort get help from elsewhere, for example your

    instrument supplier help desk or your local technical support

    department.

    6

    Potential Sources of Problems

    Mobile phase

    Injector

    Pump

    Sample

    Column and guard column

    Detectors

    Tubing and connectors

    etc.

    Analysts should simplify the problem source

    chemistry or mechanical?

  • 8/7/2019 2006MD 7 - Troubleshooting

    4/174

    7

    Troubleshooting Guide

    Main Troubles with HPLC1. Pressure

    Low pressure

    Pressure fluctuation

    High pressure

    2. Unstable Baseline

    Drift/noisy

    Spikes

    Sawtooth

    Cyclic

    No change

    3. Distorted Peak Shape

    Broading

    Tailing

    Leading

    Splitting

    No peaks

    4. Retention Time

    Changing Retention

    8

    Pressure Low pressure

    Leaking in pumps, check

    valves, seals or connectors

    Pump cavitation

    No pumping

  • 8/7/2019 2006MD 7 - Troubleshooting

    5/17

    5

    9

    Possible causes :

    1. Air bubbles in pump head

    2. Leaks in system

    3. Check valvesare dirty

    4. Plunger seal is damaged

    5. Plunger is spoilt.

    Remedy:

    1. Purge the pump2. Check for leaks in all the connections, especially new connection.

    3. Clean check valves in ultrasonicator

    4. Replace plunger seal

    5. Replace plunger

    Pressure Fluctuation in System

    10

    High Pressure in System

    Possible Causes

    Clogging in

    Detector cell Column

    Pump

    Mixer

    Piping

    Too much work ?

    Too much nagging ?

    Medical problem ?

  • 8/7/2019 2006MD 7 - Troubleshooting

    6/17

    6

    11

    Locate the point of the pressure increases

    Column or system?

    Check the system methodically:

    Disconnect column from detector, if the pressure problem goes way,

    the problem is clogging of detector cell or piping to and from flow

    cell.

    If pressure persist, disconnect guard column and main column. If

    pressure becomes normal, clean/replace main column.

    If pressure persist, remove guard column. If pressure is normal,replace guard column.

    Continue in this manner from auto-injector (if present) to mixer and

    finally to pump until find out the high pressure spot.

    12

    Possible reason:

    Clogging in system. It may occur in any part of system.

    Remedy :

    If clogging in piping, it may be possible to flush out using

    IPA or nitric acid at higher flow rate. In this case, do not

    connect column or detector.

    Clean or replace in-line filter.

    High Pressure in System

    Troubleshooting Guide

  • 8/7/2019 2006MD 7 - Troubleshooting

    7/17

    7

    13

    Possible reason:

    Column clogging/failure.

    Accumulation of particulates and/or adsorbed substances on column

    Remedy :

    1.Clean guard column and column.

    2.Replace filter frit on column if possible.

    3.Flush the column reversely or repair the column (This is recommended

    only as a last resort).

    Precaution:

    In these cases, it is always recommended to use guard column

    whenever

    possible. The guard columns are cheaper and can be easily replaced

    when clogging or adsorption occurs.

    High Pressure in Column

    Troubleshooting Guide

    14

    Precautions for Column

    Buffer solution must be filtered

    by 0.2 or 0.45 m membrane filter.

    Sample must be filtered

    by 0.2 or 0.45 m membrane filter.

    Be sure do not jump from buffer to pure organic or from organic tobuffer. This can lead to buffer precipitation, plugging and pressureproblems.

    Always use a washout, intermediate solvent.

  • 8/7/2019 2006MD 7 - Troubleshooting

    8/17

    8

    15

    l ti (1)

    Wash the column with suitable solvent

    Open the column end, carefully remove the

    filter/frit with the column in an upright

    position, and sonicate it.

    Change the filter/frit.

    Wash the column reversely at low flow rate.

    16

    C l m Clea i g

    Washing the column will remove non-polars.

    Buffered mobile phase:

    must wash out the buffer with the same mobile phase minusthe buffer first

    Aqueous organic solvent

    Wash with Acetonitrile or methanol for at ~20 column

    volumes.

  • 8/7/2019 2006MD 7 - Troubleshooting

    9/17

    9

    17

    Column Cleaning

    For reversed phase column

    1. Wash with mobile phase without buffer salts

    2. Wash with methanol or acetonitrile

    18

    Solution (2)

    only ~1 cm might be polluted

    column

    Last resort:

    Remove the contaminated part and

    repack with clean packing material.

    Always use the same size and type of

    material used originally in the column.

  • 8/7/2019 2006MD 7 - Troubleshooting

    10/17

    10

    19

    Baseline Troubleshooting

    Noisy baseline

    Sawtooth

    Drift

    Cyclic

    Spikes

    Positive & negative peaks

    Troubleshooting Guide

    20

    Unstable Baseline

    Contaminated column

    Detector cell is dirty

    System is contaminated

    Weak detector lamp

    Leaks

    Unstable pumping

    Mobile phase is not degassed properly

    Air bubbles in pumps/detector

    Electronic noise

    Temperature is fluctuating

    Wavelength is too low

    More

    en? Why is itnot flat?

  • 8/7/2019 2006MD 7 - Troubleshooting

    11/17

    11

    21

    Spike Baseline

    Bubbles in flow cell

    Apply back pressure to the flow cell

    Degas mobile phase

    Clean flow cell with IPA

    Pump pulsation

    Check pump function

    Install damper

    22

    Baseline - Sawtooth

    Bubbles

    Mixing problem

    Plugged lines

    Electrical problems

  • 8/7/2019 2006MD 7 - Troubleshooting

    12/17

    12

    23

    Temperature fluctuations

    Mixing problem

    Mobile phase was not degassed

    Pump problem

    Baseline - Cyclic

    24

    Baseline no change

    Faulty circuits

    replace any faulty parts

    Lamp is off

    set lamp parameter to turn on lamp power

    Range setting is not correct

    enter an appropriate range value

  • 8/7/2019 2006MD 7 - Troubleshooting

    13/17

    13

    25

    Peak Shape Issues

    Peak broading

    Peak tailing

    Peak fronting

    Peak splitting

    Other asymmetry peaks

    No Peaks

    Many peak shape issues are combined

    Troubleshooting Guide

    26

    Peak Broading

    Sample overload

    Large extra system volume

    Retention time is too long

    Poor column efficiency

  • 8/7/2019 2006MD 7 - Troubleshooting

    14/17

    14

    27

    Peak Tailing

    T e fl -line ntain e tra ea l e

    a le erl a ing

    ili a- a e l n egra e at ig r ig

    te erat re

    28

    Peak Fr nting

    a le erl a ing

    C anneling in l n

  • 8/7/2019 2006MD 7 - Troubleshooting

    15/17

  • 8/7/2019 2006MD 7 - Troubleshooting

    16/17

    16

    31

    Sample is too diluted

    Injection volume is too small

    Detector lamp is off

    Leaks

    Pump is not pumping

    Injector is damaged

    Zoom setting of chromatogram is not correct

    Column retains all compounds

    Incorrect mobile phase/sample

    o Peaks

    32

    Changing Retention Time

    Contamination was built up

    Equilibration time is insufficient

    Inconsistent mobile phase preparation

    Inconsistent on-line mobile phase mixing

    Selective evaporation of mobile phase component

    Unstable column temperature

    Leakage

    Unstable pumping

    Troubleshooting Guide

  • 8/7/2019 2006MD 7 - Troubleshooting

    17/17

    17

    33

    Tips

    Good habits

    Log book record pressure and mobile phase

    Wash the column after sample analysis

    Read manual/instruction

    If any problem happens

    Observe the instrument carefully look, smell, and feel

    Check the simple things first

    The hardware is not easy to be broken

    34

    Thank You!