1© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | 003-1904-00 Rev J (11/11/15)
VetPro® 5000 Wall / Cabinet Mount Installation Applies to Models:8001-001 8001-002 8001-005 8001-006
warning Equipmentnotsuitableforuseinthepresenceofaflammableanestheticmixture.
noteUse clamps and fittings for tubing connections shown in VetPro Solo and or VetPro 5000 Junction box installation instructions. Clamp sleeves are only needed if you need to shorten the water line and lose your existing clamp sleeves.
water Bottles (2)
grease (1) P80 (PN SMSC-0031)
Dowel Pin (1)
Clamp Sleeves (5)
Barb Fittings (2)
Uni-Clamps (20)
instruments and accessories May Vary Depending on Model
MaxTensionSprings
Loose ComponentsBearing Plate and Cover
2© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
SuggestedMountingLocationFromProcedureTableDeliveryUnitcanbeinstalledonleftorrightsideoftableorcabinet.
180° arm Length Standard - 28” Short - 13 1/2”
250° arm Length
Standard-45.3”Short - 37 5/16”
Mounting RequirementsMountingStructureNote:Themountingstructuremustbecapableofsupportingtheweightandtorquelistedbelowwithoutdeflection.Unit / Arm Assembly MaximumWeight 50lbs (22.6 kgs)
Torque 170.8 ft. lbs. (231.5 N•M)
Note: To prevent unit from drifting, the mounting plate must deflect no more than 1/16” (1.6 mm) over a 12” (30.5) distance.
Cabinet or wall
Cab
inet
or w
all
noteWHIP head folded against wall measures 30.6” from center of main axis to outermost point.
Space requirements
3© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
range of Motion
4© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
VA1327
Cabinet Mount - WallMountStartsOnNextPage
noteDelivery Unit can be installed on left or right side of cabinet.
The bearing plate can be mounted with the needle bearing at the top or bottom. It is recommended that the needle bearing is positioned at the Top.
If user desires, hang plastic sleeve for tri-fold quick reference with training CD on side of cabinet next to bearing plate for easy access.
Continue Cabinet Mount with Step 3
Step2:ForsteelcabinetsONLY.A) Drill two # 21 (.221) holes in the side of the cabinet for the two bottom backer plate screws. This is not required when steel cabinets are supplied with a wood end panels.
Bearing Plate
Shim
Level
needle Bearing
Equipment alert The cabinet must be fastened to a wall stud similar to the wall mount installation. Failure to do that will cause cabinet to tip over. Step1:Mounttocabinet.A) Screw cabinet to wall studs on delivery mount side as close to corner as possible.B) Mark and drill top two 5/16” mounting holes through bearing plate and cabinet. C) Align mounting holes and install appropriate bolts (1” or 2 1/4”) in top two holes. Repeat(B)and(C)forlowertwoholesonbearingplate. D) Loosen all four bolts and level bearing mounting plate. Use shims if necessary, then tighten bolts.
Mount Cabinetto wall
Backer Plate
Drill Bottom Holes on Steel Cabinets Only
Chase
note Refer to chase installation manual for mounting hole locations in chase cabinet. To install VetPro 5000, use hardware included with chase cabinet. (1/4” screws, washers and nuts)
Chase Cabinets
5© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
wallStuds
MountingCover
MountingPlate
wall Mount
noteThe bearing plate can be mounted with the needle bearing at the top or bottom. It is recommended that the needle bearing is positioned at the Top.If user desires, hang plastic sleeve for tri-fold quick reference with training CD on side of wall next to bearing plate for easy access.
Continue wall Mount with Step 3
Step2:Installbearingplate.A) Align mounting holes in bearing plate and frame cover. B) Install four bolts. C) Loosen bolts and level bearing mounting plate. Use shims, if necessary, then tighten bolts.
BearingPlate Shim
needleBearing
Equipment alert Mounting holes MUST be aligned with wall studs. Failure to comply could cause delivery unit to fall damaging unit and wall structure. Step1:Mounttowall.A) Locate wall studs and align mounting plate with studs. B) Level then mark and drill 1/4” pilot holes (36” top to floor). C) Secure frame to wall with (8) (5/16” x 2 1/2”) lag screws. D) Install mounting cover using (4) (#10-32 x 3/8”) screws.
Level
6© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step3:Installunitarm.A) Loosen set screw on bearing. B) Apply grease (supplied) to fixed arm. C) Route umbilical tubing thru bearing and slide fixed arm into bearing. D) Size and trim umbilical sleeve length to Jbox or Solo unit. Slide sleeve over tubing & wiring. E) Install and secure umbilical retainer.
Umbilical Sleeve
retainer
apply grease
Set Screw
7© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
VA1719
adjustment Screw
Step4:Adjustaretension,ifnecessary.Adjust arms at pivot points using a 3/32 allen wrench (5 places). • If arms have too much play, tighten them (Clockwise). • If they do not move freely, loosen them (Counter-Clockwise).
adjustment Screws
adjustment Screw (ThisLocationOnlyUse1/8”AllenWrench) adjustment
Screw
8© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
6
Step5:Installbearingcoverandwaterbottle.A) Install bearing cover. B) Install water bottle.
water Bottle
Bearing Cover
CABINETINSTALLSONLY!Equipment alert
Installing hard stop bracket on cabinet installations will prevent flex arm from swinging into cabinet and causing damage to unit. Step6:Installhardstopbracket. A) Remove cap screw. B) Install bracket.C) Install and tighten screw.
Hard Stop Bracket
9© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
wall or Cabinet
Equipment alert Stop pins prevent arm from turning 360 degrees which twists and kink tubing. You may install one or two stop pins. Verify desired arm turning positions with customer. Position may be changed by moving stop pins. Step7:Installstoppins.A) Position arm at 90 degree angle based on user preferred direction swing, right-hand swing or left-hand swing. B) Lift trim ring located at bend on arm, closest to wall / cabinet. C) Rotate arm to desired position and install stop pins in rear holes.
Trimring
Stop Pin Locationsrear Only
10© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step8:InstallJunctionBoxAssemblyORVetPro®Solo.Refer to: Junction Box Installation Instructions (Document 003-1914-00 included with Junction Box) Or VetPro® Solo Installation Instructions (Document # 003-1996-00 included with VetPro® Solo)
ContinuewithStep9afterinstallationofJunctionBoxorVetPro® Solo
Step9:Activateflexarmlock.A) Remove end cap. B) Slide cover back to access locking pawls. C) Cut and remove wire tie. D) Adjust jam nut if too much “drift” in the deliver head (if applicable). E) Position arm going downward toward deliver head to slide cover in position and install end cap.
Locking Pawls
WireTie
End Cap
Jam nut
Toadjustflexarmdrift...A) Tighten jam nut clockwise if flex arm drifts DOWN. B) Loosen jam nut counter clockwise if flex arm drifts UP.
11© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step10:Installinstrumenttubing.A) Remove packaging from dental delivery head. B) Move pivot assembly forward and remove set screw from pivot. C) Slide arm into pivot on back of delivery head. D) Tighten set-screw with 3/32” allen wrench. E) Repeat with each arm.
Set-Screw
arm
Arm
Inst
alla
tion
12© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step11:Installprophyangle.
A) Rotate chuck release ring to the right to unlock. B) Install Prophy angle, turn chuck release ring to the left to tighten.
Step12:Installburinhighspeedhandpiece.A) Push bur into handpiece. (The shank will only go in half-way) B) Then push button on back of handpiece and push the bur in until shank is all the way in.
Note: Friction grip burs (FG Style) are designed so that the entire straight portion of the shaft should be inside the chuck of the handpiece. You should only see the tapered portion of the bur if installed properly.
Push Bur into Handpiece
Push Button
Finish Sliding Bur into Handpiece
ThisPortionof the Bur Should be inside the Handpiece
Equipment alert Handpieces will not work properly if not installed as shown.
Shaft inside the Handpiece
Prophy angle
13© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step13:Installinstruments,adjusttubing.A) Install instruments. B) Adjust tubing length by sliding tube. (on scaler slide grommet). C) Change arm spring if arm moves back and forth too freely. Maximum tension springs are included with loose parts.
Handpiece Connection
Tochangearmtensionspring (ifmoretensiononarmiswanted)...A) Remove tool pad cover and upper shroud.B) Disconnect from pivot pin with pliers. C) Remove allen screw with 7/64 allen wrench. D) Replace spring and tighten allen screw. E) Install upper shroud and tool pad.
TensionSpring
allen Screw
Pivot Pin
ToolPad
Upper Shroud
Arm
Adj
ustm
ent
Swivel Coupler Connection
Lightly spray the exposed o-rings with one step conditioner.
Tighten coupler firmly with wrench.
Line up components properly to prevent damage resulting in hand piece failure.
14© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
note The air adjustment valve that corresponds to the ultrasonic scaler must be OPEN in order to operate the scaler with the foot pedal.
Step15:Checkhandpiecewaterflow.A) Lift and pull handpiece forward.B) Toggle foot control to Wet position. Depress foot control. C) Regulate water pressure by adjusting the corresponding knobs. D) Repeat for each handpiece that uses water.
Pressure gauge[Drive air] Coolant water
Scaler adjustment
Syringe water
Drive air
Coolant air
Front View
Bottom View
Syringe air
Pressureindicator
Coolant waterFlush Button
air
water
Step14:Checkhandpieceairflow.A) Lift and pull handpiece forward. B) Toggle foot control to Dry. Depress foot control. C) Regulate air pressure by adjusting the corresponding knobs. D) Repeat for each handpiece.
15© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
InstallationTroubleshooting
Problem Probable Cause Check CorrectionNowatertothesystem. City water and delivery head supply
tube installed incorrectly. Verify City water is connected to the ribbed portion of the tube.
Correct tubing.
Watertosyringebutnothingelse. Water toggle is toggled to dry. Verify Toggle is towards the blue dot. Redirect toggle.
Water control valves are closed. Verify water control knob is turned on.
Turn on water supply.
Waterisleakingfromthelowspeedinstrument. Water supply is turned on to low speed.
Verify water is turned off to low speed.
Turn water off.
Fiberopticlightonhighspeedandscalerarenotworking. GFI was tripped. Verify green light on GFI is on. Reset GFI.
Scalerhaswaterbutdoesnotvibrate. Transformer or universal power supply is not plugged in.
Verify power supply is plugged in. Connect scaler to power supply.
Inadequate PSI. Verify PSI is reaching mid 50’s to mid 60’s.
Adjust PSI
Tip is not tight. Verify tip is tight. Tighten tip with scaling tip wrench.
Frequency setting is not set correctly. Verify dial corresponds with tip color. Adjust dial.
Fiberopticlightisnotworkingonhighspeed. High speed is installed on the wrong tubing.
Verify high speed is installed on the fiber optic tubing.
Reinstall high speed.
Multiplehandpiecesrunatthesametime. On / Off switch to the delivery head is turned off.
Verify switch position. Turn switch on.
Actuator arm was fallen down on one or more handpiece positions.
Verify actuator arm position. Correct positioning
Airleakingfromfootcontroler. Foot controller tubes were routed incorrectly.
Verify dashed line connects to Solo or Junction Box. Verify brown tube is not being used.
Correct tubing.
Deliveryheadwillnotlock. Twist tie is still in place on the locking claws inside flex arm.
Verify twist tie is removed. Remove tie.
JunctionBoxonly:Nocitywater. City water regulator is turned down too low.
Verify regulator is set to 30. Adjust regulator.
16© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
PostInstallation&TrainingChecklist
Check Verify the following: Delivery Head
1 Delivery unit comes ON when master switch is turned ON. Verify user knows switch location.
2 Delivery head locks, unlocks and doesn’t drift. (Must have air supply to delivery head to function.)
3 Flex arm and delivery unit rotate easily and stay where placed.
4 Flex arm and delivery unit are level. Stop pins are in place.
5 User knows the coolant water flush button, potentiometer, and water control valves. Note:Keepwatershutofftolowspeedatalltimes.
6 Pressing foot control delivers air/water to handpieces as expected. Verify user knows the function of the wet/dry toggle on foot control. (Effects water flow to low speed, high speed, and scaler only.)
7 Review quick reference guide with user, making sure they understand procedures for deliver head.
8User knows prophy angle, burr, scaling tip and syringe installation and removal. Review using Midmark conditioner for instruments, note: Burrs shouldbechangedandinstrumentsconditionedaftereachpatient.Scalingtipsshouldbeoperatedwithinthe correspondingcoloronthepotentiometerdial.
9 Wipe all components with a soft dry cloth to remove dirt and smudges.
Junction Box (if applicable)
10 Water and air supply regulator gauges in J-box read appropriate pressures. 32-40 for the high speed, 35-50 for the low speed, full flow for scaler.
Solo (if applicable)11 Solo unit On/Off switch works and user is aware of switch location.
12 Drain valve is working properly. User can locate city water shut off valve.
13 User is aware of GFI (used with fiber optics and scaler) plug on outside of Solo unit and plug on the inside of Solo unit.
14 Installers Signature_____________________________________________________ Date _________________________________
15 Users Signature________________________________________________________Date___________________________________