애리조나 대학 아틀라스 xray 강의자료_phys586-lec05-xray

Post on 05-Apr-2018

243 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/2/2019 xray _phys586-lec05-xray

    1/60

    1

    X-rays

    Ouch!

  • 8/2/2019 xray _phys586-lec05-xray

    2/60

    2

    X-rays

    X-rays are produced when electrons areaccelerated and collide with a target

    Bremsstrahlung x-rays

    Characteristic x-rays

    X-rays are sometimes characterized by thegenerating voltage

    0.1-20 kV soft x-rays

    20-120 kV diagnostic x-rays

    120-300 kV orthovoltage x-rays

    300 kV 1 MV intermediate energy x-rays

    > 1MV megavoltage x-rays

  • 8/2/2019 xray _phys586-lec05-xray

    3/60

    3

    Bremmstrahlung

    Bremsstrahlung x-rays occur whenelectrons are (de)accelerated in theCoulomb field of a nucleus

  • 8/2/2019 xray _phys586-lec05-xray

    4/60

    4

    Bremsstrahlung

  • 8/2/2019 xray _phys586-lec05-xray

    5/60

    5

    Bremsstrahlung

    The power radiated from an acceleratingcharge is given by Larmors equation

    In the case of an electron in the Coulomb fieldof a nucleus

    3

    22

    3

    2

    c

    aeP

    Zmr

    Zek

    m

    Fa ~

    2

    2

  • 8/2/2019 xray _phys586-lec05-xray

    6/60

    6

    Bremsstrahlung

    The probability of bremsstrahlung goes as Z

    2

    ,hence high Z targets are more effective thanlow Z

    The energy of the x-rays varies from zero to

    the maximum kinetic energy of the electron(x-ray tube kVp)

    The energy spectrum from a thick target goesas 1/E but inherent (1mm Al eq) plusadditional (few mm Al) filtration removes thelower energy x-rays

    Here I am referring to diagnostic x-rays

  • 8/2/2019 xray _phys586-lec05-xray

    7/60

    7

    Bremsstrahlung

    The unfiltered energy spectrum isapproximately given by Kramers law whichwas an early application of quantummechanics

    ETKZEI e

  • 8/2/2019 xray _phys586-lec05-xray

    8/60

    8

    Bremsstrahlung

  • 8/2/2019 xray _phys586-lec05-xray

    9/60

    9

    Characteristic x-rays

    After excitation, ionswith a vacancy intheir inner shell cande-excite

    Radiatively throughx-ray fluorescence

    Non-radiativelythrough the emission

    of Auger electrons

  • 8/2/2019 xray _phys586-lec05-xray

    10/60

    10

    Characteristic X-raysThus an x-ray spectrum will also show

    characteristic x-rays arising from L to K and Mto K transitions after ionization of a K electron

    Usually transitions to higher shells

    absorbed by the filtration or are not x-rays

  • 8/2/2019 xray _phys586-lec05-xray

    11/60

    11

    Characteristic X-rays

    The probability of K shell fluorescenceincreases with Z

  • 8/2/2019 xray _phys586-lec05-xray

    12/60

    12

    Characteristic X-rays

  • 8/2/2019 xray _phys586-lec05-xray

    13/60

    13

    Characteristic X-rays

    Sometimes the characteristic x-rays areemphasized using the same material fortarget and filter

    Characteristic x-rays from molybdenum are

    effective in maximizing contrast in mammography

  • 8/2/2019 xray _phys586-lec05-xray

    14/60

    14

    Characteristic X-rays

    Mo target, filter, and result

  • 8/2/2019 xray _phys586-lec05-xray

    15/60

    15

    Directionality

    For MeV electrons, bremsstrahlung x-rays are preferentially emitted in theelectrons direction

    For keV electrons, bremsstrahlung x-

    rays are emitted at larger anglesCharacteristic x-rays are emitted

    isotropically since there is no angular

    correlation between the incidentelectron that causes the ionization andthe fluorescent photon

  • 8/2/2019 xray _phys586-lec05-xray

    16/60

    16

    X-ray Tube

    A simplified x-ray tube (Coolidge type)shows the idea behind most x-ray tubestoday

  • 8/2/2019 xray _phys586-lec05-xray

    17/60

    17

    X-ray Tube

    In addition to bremsstrahlung and

    characteristic x-ray production, electrons alsoloose energy through collisions

    Collision losses dominate in this energy region

    For 100 keV electrons in W

    Thus >99% of the electron energy goes intoheating the target rather than x-rays

    Removing heat from the anode in a vacuum is an

    issue

    MeVin820losscollisionlossradiation

    EEZ

    009.0820

    741.0

    losscollision

    lossradiation

  • 8/2/2019 xray _phys586-lec05-xray

    18/60

    18

    X-ray Tube

    Efficiency of x-ray production depends on thetube voltage and the target material

    W (Z=74) in this example

    ZVP

    P

    IZVP

    VIP

    deposited

    radiated

    radiated

    deposited

    9

    29

    109.0Efficiency

    109.0

    kVp

    (V)

    Heat(%)

    X-rays(%)

    50 99.7 0.3200 99 1

    6000 65 35

  • 8/2/2019 xray _phys586-lec05-xray

    19/60

    19

    X-ray Tube

    X-ray tubes

  • 8/2/2019 xray _phys586-lec05-xray

    20/60

    20

    X-ray Tube

    More detail

  • 8/2/2019 xray _phys586-lec05-xray

    21/60

    21

    X-ray Tube

    Housing for shielding (Pb) and cooling(oil)

  • 8/2/2019 xray _phys586-lec05-xray

    22/60

    22

    X-ray Tube

    More detail

  • 8/2/2019 xray _phys586-lec05-xray

    23/60

    23

    X-ray Tube

    The main parts of the x-ray tube are Cathode/filament

    Typical electron current is 0.1-1.0 A for shortexposures (< 100 ms)

    Anode/target

    Glass/metal envelope

    Accelerating voltage

    Typical voltage is 20-150 kVp

  • 8/2/2019 xray _phys586-lec05-xray

    24/60

    24

    Cathode

    Cathode consists of Low R tungsten wire for thermionic emission

    Tungsten has a high melting point (3370C) and minimumdeposit on the glass tube

    Tube current is controlled by varying the filament currentwhich is a few amps

    A focusing cup Uses electric field lines to focus the electrons

    Typically there are two filaments Long one: higher current, lower resolution

    Large focal spot

    Short one: lower current, higher resolution Small focal spot

  • 8/2/2019 xray _phys586-lec05-xray

    25/60

    25

    Cathode

    Dual focus filament is common

  • 8/2/2019 xray _phys586-lec05-xray

    26/60

    26

    Anode

    Usually made of tungsten in copper becauseof high Z and high melting point Molybdenum and rhodium used for soft tissue

    imaging

    Large rotating surface for heat distribution

    and radiative heat loss Rotation of 3k-10k revolutions/minute Resides in a vacuum (~10-6 torr) Thermally decoupled from motor to avoid

    overheating of the shaft

    Target is at an tilted angle with respect toaxis Bremsstrahlung is emitted at ~ right angles for

    low energy electrons

    Determines focal spot size

  • 8/2/2019 xray _phys586-lec05-xray

    27/60

    27

    Anode

  • 8/2/2019 xray _phys586-lec05-xray

    28/60

    28

    Anode

  • 8/2/2019 xray _phys586-lec05-xray

    29/60

    29

    AnodeThe heating of the anode limits the voltage,

    current, and exposure timeAn exposure rating chart gives these limits

  • 8/2/2019 xray _phys586-lec05-xray

    30/60

    30

    Anode

    Power = V x I (watts)

    Energy = Power x time = V x I x s(joules)

    HU (Heating Unit) ~ J Damaged anodes

  • 8/2/2019 xray _phys586-lec05-xray

    31/60

    31

    AnodeThe angle determines the projected focal spot

    The smaller the angle the better the resolution Typically 7-20 degrees

    Angle

    Incident electronbeam width

    Apparent focal spot size

    Actual focal

    spot size

    Film

    Angle

    Incident electronbeam widthIncreasedapparent

    focal spot size

    Actual focal

    spot size

    Film

  • 8/2/2019 xray _phys586-lec05-xray

    32/60

    32

    X-rays

    The energy of the photons depends onthe electron energy (kVp) and thetarget atomic number Z

    The number of photons depends on thethe electron energy (kVp), Z, and thebeam current (mA)

    A typical number / area is ~ 1013 / m2

    About 1% will hit the film ~ 1011 / m2

    Absorption and detection efficiency willfurther reduce this number

  • 8/2/2019 xray _phys586-lec05-xray

    33/60

    33

    Automatic Exposure Control

    AEC detectors can ionization chambers or solid-state detectors

    X Ray tube

    Collimator

    Beam

    Softtissue

    BoneAir

    Patient

    Table

    Grid

    Cassette

    AEC detectors

  • 8/2/2019 xray _phys586-lec05-xray

    34/60

    34

    Automatic Exposure Control

    Most modern x-rays machines are equippedwith automatic exposure control also called aphototime

    The AEC sets the technical parameters of the

    machine (kV, mA, time, ) in order to avoidrepeated exposures

    AEC is used to keep the radiographic quality(film density) equal on all patients

    AEC detectors can be ionization chambers orsolid state detectors

  • 8/2/2019 xray _phys586-lec05-xray

    35/60

    35

    Grid

    To reduce the number of secondaryscattered photons making it to the film,a grid between the patient and film isused

  • 8/2/2019 xray _phys586-lec05-xray

    36/60

    36

    GridDetails

    Grid bars are usually lead whereas the gridopenings are usually made of aluminum orcarbon

    Grid thickness is typically 3 mm

    Grid ratio is H/W and 10/1 is typical Grid frequency of 60 lines / cm is typical

    B/W/H on the figure might be 0.045, 0.120,1.20 in mm

    The Bucky factor is the entrance exposurew/wo the grid while achieving the samefilm density 4 is average

  • 8/2/2019 xray _phys586-lec05-xray

    37/60

    37

    Accelerating Voltage

    The potential difference betweencathode and anode must be generatedby 60 Hz 220V AC power

    High voltages are produced using atransformer

  • 8/2/2019 xray _phys586-lec05-xray

    38/60

    38

    Accelerating Voltage

    Electrons are accelerated when thefilament is at a negative potential withrespect to the target

    Diode circuits can be used to providerectification (AC to DC voltage)

    Three phase power (6 pulse or 12 pulse)can be used to reduce ripple

    Constant potential operation can beachieved by using constant potential(voltage regulations) or high frequency x-ray generators

  • 8/2/2019 xray _phys586-lec05-xray

    39/60

    39

    Half-wave Rectifier

    Not very efficient

  • 8/2/2019 xray _phys586-lec05-xray

    40/60

    40

    Full-wave Bridge Rectifier

    This circuit allows the entire inputwaveform to be used

  • 8/2/2019 xray _phys586-lec05-xray

    41/60

    41

    Accelerating Voltage

    100%

    13%

    4%

    Line voltage

    Single phase single pulse

    Single phase 2-pulse

    Three phase 6-pulse

    Three phase 12-pulse

    0.02 s

    0.01 s

    kV ripple (%)

  • 8/2/2019 xray _phys586-lec05-xray

    42/60

    42

    ImagesAnalog radiography

    Film based still widely used Fluorescent screens are used to convert x-rays into

    visible light that is then recorded on film

    Screens are more efficient at stopping x-rays than

    the film (CaWO4 or Gd2O2S:Tb or other rare earth)

  • 8/2/2019 xray _phys586-lec05-xray

    43/60

    43

    Analog Radiography

    The film itself has excellent spatialresolution but

    Film detects 0.65% of incident x-rayenergy

    Gd2O2S detects 29.5% of incident x-rayenergy

    Thus using phosphor screens greatly

    reduces the radiation dose to thepatient

    And also reduces load on the x-ray tube

  • 8/2/2019 xray _phys586-lec05-xray

    44/60

    44

    Analog Radiography

    There are two efficiency considerations

    Absorption efficiency or QDE

    Fraction of incident x-rays that interactwith the screen

    Depends on kVp and screen thickness

    Gd2O2S has a QDE of ~ 60% for 80 kVp beam,20 cm patient, 120 mg/cm2 screen thickness

  • 8/2/2019 xray _phys586-lec05-xray

    45/60

    45

    Analog Radiography

    Conversion efficiency Fraction of absorbed x-ray energy that is

    emitted as light

    5% for CaWO4

    15% for Gd2O2S

    50,000 eV x 0.15 = 7500 eV

    7500 eV / 2.7 eV = 2800 photons produced per

    absorbed x-ray 50-90% reduction in photon diffusion to film

  • 8/2/2019 xray _phys586-lec05-xray

    46/60

    46

    Analog Radiography

    Film is an emulsion containing silver-halide grains (AgBr and AgI) coated onmylar

    Body

    Film

    X-Raysource

    DarkLight

  • 8/2/2019 xray _phys586-lec05-xray

    47/60

    47

    Analog Radiography

  • 8/2/2019 xray _phys586-lec05-xray

    48/60

    48

    Film Badge

    A film badge consists of a photographic film

    with various filtersThe film is a gelatin emulsion containing silver-

    halide grains (95% AgBr and 5% AgI) on a

    supporting material Grain diameter is ~ 1mm

  • 8/2/2019 xray _phys586-lec05-xray

    49/60

    49

    Film Badge

    The film is exposed by light by An electron is released from Br- and moves about

    the 1m diameter crystal

    The electron may be captured by a trap such as acrystal imperfection or AgS speck

    The trapped electron attracts mobile Ag+ ionswhere it is subsequently neutralized

    Additional Ag atoms are formed by repeatedtrapping and neutralization

    These Ag atoms are called a latent image center The developing process effectively amplifies this

    process turning the grains with latent imagecenters into a visible silver deposit

  • 8/2/2019 xray _phys586-lec05-xray

    50/60

    50

    Film Badge

  • 8/2/2019 xray _phys586-lec05-xray

    51/60

    51

    Film Badge

    Silver atoms at latent image centers

  • 8/2/2019 xray _phys586-lec05-xray

    52/60

    52

    Images

    Digital radiography

    Detector based

  • 8/2/2019 xray _phys586-lec05-xray

    53/60

    53

    Digitial Radiography

    CCD systems CCD systems use a scintillator like

    gadolinium disulphide to convert x-rays tovisible light

    Light is collected by optics to demagnifythe 35x45cm2 film to 2-4 cm2 CCD

    Well talk about CCDs much later in the

    course but essentially visible light isconverted into charge that is amplified andreadout

    A negative is the thickness of the detector

    system because of the optical system

  • 8/2/2019 xray _phys586-lec05-xray

    54/60

    54

    Digital Radiography

    Indirect or direct conversion thin-filmtransistor (TFT) arraysAlso called FPD (flat panel detectors)

    Well cover these later in the course as well

    probably through a student talk The idea is that charge proportional to the

    x-rays received is stored on a capacitor

    The charges are conducted out by

    transistors one row at a time andsubsequently amplified, multiplexed, anddigitized

    The readout is very fast

  • 8/2/2019 xray _phys586-lec05-xray

    55/60

    55

    Digital Radiography

    Indirect or direct conversion thin-filmtransistor (TFT) arrays Indirect conversion uses a scintillator layer

    (like CsI:Tl) to convert x-rays to visible

    light and amorphous silicon photodiodes toconvert visible light into charge

    Direct conversion uses an x-rayphotoconductor layer (usually amorphous

    selenium) to convert x-rays to chargeAn applied electric field directs the charges

    to the charge collection electrodes

  • 8/2/2019 xray _phys586-lec05-xray

    56/60

    56

    Digital Radiography

  • 8/2/2019 xray _phys586-lec05-xray

    57/60

    57

    Digital Radiography

  • 8/2/2019 xray _phys586-lec05-xray

    58/60

    58

    Digital Radiography

    Readout

  • 8/2/2019 xray _phys586-lec05-xray

    59/60

    59

    Images

    Digital radiography

    The battle over image quality, however,may be incomprehensible to anyonewithout a background in high-energy

    physics.

  • 8/2/2019 xray _phys586-lec05-xray

    60/60

    X-rays

    For bone tissue, the linear attenuationcoefficient is much greater than that forsoft body tissue