애리조나 대학 아틀라스 xray 강의자료_phys586-lec05-xray
Post on 05-Apr-2018
243 views
TRANSCRIPT
-
8/2/2019 xray _phys586-lec05-xray
1/60
1
X-rays
Ouch!
-
8/2/2019 xray _phys586-lec05-xray
2/60
2
X-rays
X-rays are produced when electrons areaccelerated and collide with a target
Bremsstrahlung x-rays
Characteristic x-rays
X-rays are sometimes characterized by thegenerating voltage
0.1-20 kV soft x-rays
20-120 kV diagnostic x-rays
120-300 kV orthovoltage x-rays
300 kV 1 MV intermediate energy x-rays
> 1MV megavoltage x-rays
-
8/2/2019 xray _phys586-lec05-xray
3/60
3
Bremmstrahlung
Bremsstrahlung x-rays occur whenelectrons are (de)accelerated in theCoulomb field of a nucleus
-
8/2/2019 xray _phys586-lec05-xray
4/60
4
Bremsstrahlung
-
8/2/2019 xray _phys586-lec05-xray
5/60
5
Bremsstrahlung
The power radiated from an acceleratingcharge is given by Larmors equation
In the case of an electron in the Coulomb fieldof a nucleus
3
22
3
2
c
aeP
Zmr
Zek
m
Fa ~
2
2
-
8/2/2019 xray _phys586-lec05-xray
6/60
6
Bremsstrahlung
The probability of bremsstrahlung goes as Z
2
,hence high Z targets are more effective thanlow Z
The energy of the x-rays varies from zero to
the maximum kinetic energy of the electron(x-ray tube kVp)
The energy spectrum from a thick target goesas 1/E but inherent (1mm Al eq) plusadditional (few mm Al) filtration removes thelower energy x-rays
Here I am referring to diagnostic x-rays
-
8/2/2019 xray _phys586-lec05-xray
7/60
7
Bremsstrahlung
The unfiltered energy spectrum isapproximately given by Kramers law whichwas an early application of quantummechanics
ETKZEI e
-
8/2/2019 xray _phys586-lec05-xray
8/60
8
Bremsstrahlung
-
8/2/2019 xray _phys586-lec05-xray
9/60
9
Characteristic x-rays
After excitation, ionswith a vacancy intheir inner shell cande-excite
Radiatively throughx-ray fluorescence
Non-radiativelythrough the emission
of Auger electrons
-
8/2/2019 xray _phys586-lec05-xray
10/60
10
Characteristic X-raysThus an x-ray spectrum will also show
characteristic x-rays arising from L to K and Mto K transitions after ionization of a K electron
Usually transitions to higher shells
absorbed by the filtration or are not x-rays
-
8/2/2019 xray _phys586-lec05-xray
11/60
11
Characteristic X-rays
The probability of K shell fluorescenceincreases with Z
-
8/2/2019 xray _phys586-lec05-xray
12/60
12
Characteristic X-rays
-
8/2/2019 xray _phys586-lec05-xray
13/60
13
Characteristic X-rays
Sometimes the characteristic x-rays areemphasized using the same material fortarget and filter
Characteristic x-rays from molybdenum are
effective in maximizing contrast in mammography
-
8/2/2019 xray _phys586-lec05-xray
14/60
14
Characteristic X-rays
Mo target, filter, and result
-
8/2/2019 xray _phys586-lec05-xray
15/60
15
Directionality
For MeV electrons, bremsstrahlung x-rays are preferentially emitted in theelectrons direction
For keV electrons, bremsstrahlung x-
rays are emitted at larger anglesCharacteristic x-rays are emitted
isotropically since there is no angular
correlation between the incidentelectron that causes the ionization andthe fluorescent photon
-
8/2/2019 xray _phys586-lec05-xray
16/60
16
X-ray Tube
A simplified x-ray tube (Coolidge type)shows the idea behind most x-ray tubestoday
-
8/2/2019 xray _phys586-lec05-xray
17/60
17
X-ray Tube
In addition to bremsstrahlung and
characteristic x-ray production, electrons alsoloose energy through collisions
Collision losses dominate in this energy region
For 100 keV electrons in W
Thus >99% of the electron energy goes intoheating the target rather than x-rays
Removing heat from the anode in a vacuum is an
issue
MeVin820losscollisionlossradiation
EEZ
009.0820
741.0
losscollision
lossradiation
-
8/2/2019 xray _phys586-lec05-xray
18/60
18
X-ray Tube
Efficiency of x-ray production depends on thetube voltage and the target material
W (Z=74) in this example
ZVP
P
IZVP
VIP
deposited
radiated
radiated
deposited
9
29
109.0Efficiency
109.0
kVp
(V)
Heat(%)
X-rays(%)
50 99.7 0.3200 99 1
6000 65 35
-
8/2/2019 xray _phys586-lec05-xray
19/60
19
X-ray Tube
X-ray tubes
-
8/2/2019 xray _phys586-lec05-xray
20/60
20
X-ray Tube
More detail
-
8/2/2019 xray _phys586-lec05-xray
21/60
21
X-ray Tube
Housing for shielding (Pb) and cooling(oil)
-
8/2/2019 xray _phys586-lec05-xray
22/60
22
X-ray Tube
More detail
-
8/2/2019 xray _phys586-lec05-xray
23/60
23
X-ray Tube
The main parts of the x-ray tube are Cathode/filament
Typical electron current is 0.1-1.0 A for shortexposures (< 100 ms)
Anode/target
Glass/metal envelope
Accelerating voltage
Typical voltage is 20-150 kVp
-
8/2/2019 xray _phys586-lec05-xray
24/60
24
Cathode
Cathode consists of Low R tungsten wire for thermionic emission
Tungsten has a high melting point (3370C) and minimumdeposit on the glass tube
Tube current is controlled by varying the filament currentwhich is a few amps
A focusing cup Uses electric field lines to focus the electrons
Typically there are two filaments Long one: higher current, lower resolution
Large focal spot
Short one: lower current, higher resolution Small focal spot
-
8/2/2019 xray _phys586-lec05-xray
25/60
25
Cathode
Dual focus filament is common
-
8/2/2019 xray _phys586-lec05-xray
26/60
26
Anode
Usually made of tungsten in copper becauseof high Z and high melting point Molybdenum and rhodium used for soft tissue
imaging
Large rotating surface for heat distribution
and radiative heat loss Rotation of 3k-10k revolutions/minute Resides in a vacuum (~10-6 torr) Thermally decoupled from motor to avoid
overheating of the shaft
Target is at an tilted angle with respect toaxis Bremsstrahlung is emitted at ~ right angles for
low energy electrons
Determines focal spot size
-
8/2/2019 xray _phys586-lec05-xray
27/60
27
Anode
-
8/2/2019 xray _phys586-lec05-xray
28/60
28
Anode
-
8/2/2019 xray _phys586-lec05-xray
29/60
29
AnodeThe heating of the anode limits the voltage,
current, and exposure timeAn exposure rating chart gives these limits
-
8/2/2019 xray _phys586-lec05-xray
30/60
30
Anode
Power = V x I (watts)
Energy = Power x time = V x I x s(joules)
HU (Heating Unit) ~ J Damaged anodes
-
8/2/2019 xray _phys586-lec05-xray
31/60
31
AnodeThe angle determines the projected focal spot
The smaller the angle the better the resolution Typically 7-20 degrees
Angle
Incident electronbeam width
Apparent focal spot size
Actual focal
spot size
Film
Angle
Incident electronbeam widthIncreasedapparent
focal spot size
Actual focal
spot size
Film
-
8/2/2019 xray _phys586-lec05-xray
32/60
32
X-rays
The energy of the photons depends onthe electron energy (kVp) and thetarget atomic number Z
The number of photons depends on thethe electron energy (kVp), Z, and thebeam current (mA)
A typical number / area is ~ 1013 / m2
About 1% will hit the film ~ 1011 / m2
Absorption and detection efficiency willfurther reduce this number
-
8/2/2019 xray _phys586-lec05-xray
33/60
33
Automatic Exposure Control
AEC detectors can ionization chambers or solid-state detectors
X Ray tube
Collimator
Beam
Softtissue
BoneAir
Patient
Table
Grid
Cassette
AEC detectors
-
8/2/2019 xray _phys586-lec05-xray
34/60
34
Automatic Exposure Control
Most modern x-rays machines are equippedwith automatic exposure control also called aphototime
The AEC sets the technical parameters of the
machine (kV, mA, time, ) in order to avoidrepeated exposures
AEC is used to keep the radiographic quality(film density) equal on all patients
AEC detectors can be ionization chambers orsolid state detectors
-
8/2/2019 xray _phys586-lec05-xray
35/60
35
Grid
To reduce the number of secondaryscattered photons making it to the film,a grid between the patient and film isused
-
8/2/2019 xray _phys586-lec05-xray
36/60
36
GridDetails
Grid bars are usually lead whereas the gridopenings are usually made of aluminum orcarbon
Grid thickness is typically 3 mm
Grid ratio is H/W and 10/1 is typical Grid frequency of 60 lines / cm is typical
B/W/H on the figure might be 0.045, 0.120,1.20 in mm
The Bucky factor is the entrance exposurew/wo the grid while achieving the samefilm density 4 is average
-
8/2/2019 xray _phys586-lec05-xray
37/60
37
Accelerating Voltage
The potential difference betweencathode and anode must be generatedby 60 Hz 220V AC power
High voltages are produced using atransformer
-
8/2/2019 xray _phys586-lec05-xray
38/60
38
Accelerating Voltage
Electrons are accelerated when thefilament is at a negative potential withrespect to the target
Diode circuits can be used to providerectification (AC to DC voltage)
Three phase power (6 pulse or 12 pulse)can be used to reduce ripple
Constant potential operation can beachieved by using constant potential(voltage regulations) or high frequency x-ray generators
-
8/2/2019 xray _phys586-lec05-xray
39/60
39
Half-wave Rectifier
Not very efficient
-
8/2/2019 xray _phys586-lec05-xray
40/60
40
Full-wave Bridge Rectifier
This circuit allows the entire inputwaveform to be used
-
8/2/2019 xray _phys586-lec05-xray
41/60
41
Accelerating Voltage
100%
13%
4%
Line voltage
Single phase single pulse
Single phase 2-pulse
Three phase 6-pulse
Three phase 12-pulse
0.02 s
0.01 s
kV ripple (%)
-
8/2/2019 xray _phys586-lec05-xray
42/60
42
ImagesAnalog radiography
Film based still widely used Fluorescent screens are used to convert x-rays into
visible light that is then recorded on film
Screens are more efficient at stopping x-rays than
the film (CaWO4 or Gd2O2S:Tb or other rare earth)
-
8/2/2019 xray _phys586-lec05-xray
43/60
43
Analog Radiography
The film itself has excellent spatialresolution but
Film detects 0.65% of incident x-rayenergy
Gd2O2S detects 29.5% of incident x-rayenergy
Thus using phosphor screens greatly
reduces the radiation dose to thepatient
And also reduces load on the x-ray tube
-
8/2/2019 xray _phys586-lec05-xray
44/60
44
Analog Radiography
There are two efficiency considerations
Absorption efficiency or QDE
Fraction of incident x-rays that interactwith the screen
Depends on kVp and screen thickness
Gd2O2S has a QDE of ~ 60% for 80 kVp beam,20 cm patient, 120 mg/cm2 screen thickness
-
8/2/2019 xray _phys586-lec05-xray
45/60
45
Analog Radiography
Conversion efficiency Fraction of absorbed x-ray energy that is
emitted as light
5% for CaWO4
15% for Gd2O2S
50,000 eV x 0.15 = 7500 eV
7500 eV / 2.7 eV = 2800 photons produced per
absorbed x-ray 50-90% reduction in photon diffusion to film
-
8/2/2019 xray _phys586-lec05-xray
46/60
46
Analog Radiography
Film is an emulsion containing silver-halide grains (AgBr and AgI) coated onmylar
Body
Film
X-Raysource
DarkLight
-
8/2/2019 xray _phys586-lec05-xray
47/60
47
Analog Radiography
-
8/2/2019 xray _phys586-lec05-xray
48/60
48
Film Badge
A film badge consists of a photographic film
with various filtersThe film is a gelatin emulsion containing silver-
halide grains (95% AgBr and 5% AgI) on a
supporting material Grain diameter is ~ 1mm
-
8/2/2019 xray _phys586-lec05-xray
49/60
49
Film Badge
The film is exposed by light by An electron is released from Br- and moves about
the 1m diameter crystal
The electron may be captured by a trap such as acrystal imperfection or AgS speck
The trapped electron attracts mobile Ag+ ionswhere it is subsequently neutralized
Additional Ag atoms are formed by repeatedtrapping and neutralization
These Ag atoms are called a latent image center The developing process effectively amplifies this
process turning the grains with latent imagecenters into a visible silver deposit
-
8/2/2019 xray _phys586-lec05-xray
50/60
50
Film Badge
-
8/2/2019 xray _phys586-lec05-xray
51/60
51
Film Badge
Silver atoms at latent image centers
-
8/2/2019 xray _phys586-lec05-xray
52/60
52
Images
Digital radiography
Detector based
-
8/2/2019 xray _phys586-lec05-xray
53/60
53
Digitial Radiography
CCD systems CCD systems use a scintillator like
gadolinium disulphide to convert x-rays tovisible light
Light is collected by optics to demagnifythe 35x45cm2 film to 2-4 cm2 CCD
Well talk about CCDs much later in the
course but essentially visible light isconverted into charge that is amplified andreadout
A negative is the thickness of the detector
system because of the optical system
-
8/2/2019 xray _phys586-lec05-xray
54/60
54
Digital Radiography
Indirect or direct conversion thin-filmtransistor (TFT) arraysAlso called FPD (flat panel detectors)
Well cover these later in the course as well
probably through a student talk The idea is that charge proportional to the
x-rays received is stored on a capacitor
The charges are conducted out by
transistors one row at a time andsubsequently amplified, multiplexed, anddigitized
The readout is very fast
-
8/2/2019 xray _phys586-lec05-xray
55/60
55
Digital Radiography
Indirect or direct conversion thin-filmtransistor (TFT) arrays Indirect conversion uses a scintillator layer
(like CsI:Tl) to convert x-rays to visible
light and amorphous silicon photodiodes toconvert visible light into charge
Direct conversion uses an x-rayphotoconductor layer (usually amorphous
selenium) to convert x-rays to chargeAn applied electric field directs the charges
to the charge collection electrodes
-
8/2/2019 xray _phys586-lec05-xray
56/60
56
Digital Radiography
-
8/2/2019 xray _phys586-lec05-xray
57/60
57
Digital Radiography
-
8/2/2019 xray _phys586-lec05-xray
58/60
58
Digital Radiography
Readout
-
8/2/2019 xray _phys586-lec05-xray
59/60
59
Images
Digital radiography
The battle over image quality, however,may be incomprehensible to anyonewithout a background in high-energy
physics.
-
8/2/2019 xray _phys586-lec05-xray
60/60
X-rays
For bone tissue, the linear attenuationcoefficient is much greater than that forsoft body tissue