Without reference, identify principles relating to Diagnostic Ultrasound Imaging

Calibration with at least 70 percent accuracy.

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 2: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

General Ultrasound Knowledge Other names • Real-time scanners • 2-D scanners • Duplex scanners • Echocardiographs • Cardiac ultrasound imagers • Vascular ultrasound imagers

Common uses • Cardiology • Diagnostic imaging • Obstetrics and gynecology • Ophthalmology • Radiology • Surgery • Vascular lab • Emergency department

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 3: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

Scope • This procedure covers all diagnostic ultrasound scanners, including

general-purpose and dedicated systems • This procedure does not cover nonimaging diagnostic ultrasound

systems such as doppler blood-flow detectors and doppler fetal heart monitors

Test Equipment Required Multipurpose phantom - comprehensively evaluate the

performance of diagnostic ultrasound imaging systems with the following capabilities • Dead-zone or ring-down • Vertical and horizontal measurement calibration • Focal zone • Sensitivity • Axial and lateral resolution • Functional resolution • Grayscale and displayed dynamic range

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 4: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 5: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

• Material The phantom should be designed with a combination of monofilament

line targets and tissue-mimicking cylindrical targets of varying sizes and contrasts

The monofilament line targets should have a diameter of 0.1 mm to 0.5 mm ±5% to optimize the displayed image at frequencies typically used on general-purpose scanners

Monofilament line targets with diameters of approximately 0.5 mm should be present to optimize the displayed image at the high ultrasound frequencies typically used on small-parts scanners

Doppler flow phantom - comprehensive evaluation of ultrasound scanners that incorporate continuous-wave, pulsed-wave, or color-flow Doppler capabilities with the following capabilities: • Flow velocity • Location of flow • Directional discrimination

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 6: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

Material The phantom can be either the tissue-mimicking type with one or more fluid-flow

channels (containing a non-degradable blood-mimicking solution with calibrated reflecting targets) or the type that incorporates a moving string target within a fluid-filled container

The string phantom provides greater accuracy for flow velocity calibration

NOTE: To check only basic Doppler operation, it is not necessary to use the calibrated

Doppler phantoms described above A less-costly blood-flow simulator may be purchased, or one may be constructed

using a fluid pump and flexible tubing submerged within a fluid-filled container In lieu of the nondegradeable blood-mimicking solution with calibrated reflecting

targets, any echogenic fluid, such as a detergent-water solution, may be used

Blood flow simulator - for checking doppler ultrasound Ecg simulator Leakage current meter Ground resistance ohmeter

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 7: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 8: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

Overview Ultrasounds typically provide 2-D images of soft tissue Scanners also provide waveforms (M-mode) Newer scanners have 3-D capabilities Ultrasound refers to sound waves emitted at frequencies above the

level of human hearing For diagnostic imaging, frequencies ranging from 2 to 10 MHz are

typically used Ultrasound waves are mechanical vibrations that require a medium for

transmission Echoes are produced whenever the beam encounters an interface of

different acoustic impedances, such as the soft-tissue/bone interface Transducer • Placed on the skin after an acoustic coupling gel is applied • Converts an electrical signal into ultrasonic energy that can be transmitted

into tissues • Reconverts reflected signal into an electrical signal

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 9: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

Procedure Note: • Before beginning an inspection, carefully read this

procedure and the manufacturer's instruction and service manuals

• Know how to operate the equipment and the significance of each control and indicator. Also review special inspection and preventive maintenance procedures or frequencies recommended by the manufacturer

Transducer Identification and Scanner Settings for Tests with the Multipurpose Ultrasound Phantom (Record the Type, Serial #, Frequency, Settings, Power, Gain, Dynamic Range, etc.)

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 10: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Measurements • Dead Zone - place the transducer over the dead zone line target

group. Determine the minimum distance (in mm) at which the scanner can resolve individual structures

• Axial resolution - place the transducer over one of the axial resolution line target groups. Determine the minimum reflector separation (in mm) along the axis of the transducer beam required to produce separate reflections

• Lateral resolution Place the transducer over each lateral resolution line target group Determine the minimum reflector separation (in mm) perpendicular to the

sound path needed to produce discrete reflections Because lateral resolution can vary with depth, multiple transducer

locations should be used Record the lateral resolution for each depth checked If the transducer has adjustable transmit focus, verify its operation and

perform each lateral resolution measurement in the respective focal zone, if possible

Page 11: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

• Vertical Distance Place the transducer over a vertical distance calibration line target group Use the scanner's digital calipers and/or marker grids to determine the accuracy of

linear measurements along the axis of the transducer beam at 20 mm and 100 mm measured distances

• Horizontal Distance Place the transducer over a horizontal distance calibration line target group Use the scanner's digital calipers and/or marker grids to determine the accuracy of

linear measurements perpendicular to the sound path On images produced by flat sequenced linear-array transducers, check at 20 mm and

60 mm measured distances, using any left to right location On images produced by sector transducers, check at 20 mm and 60 mm measured

distances, but at multiple left to right positions. If the transducer has adjustable transmit focus, perform measurements within the focal zone, if possible

• Doppler Calibration If a calibrated Doppler ultrasound phantom is available, verify the accuracy of flow

velocity, directional discrimination, and, if applicable, pulsed-Doppler gate positioning Measurement parameters and display quality should not vary between inspections

with the same transducer configuration, scanner settings, and technique Operator Calibration - ensure operator calibrations are performed according

to the manufacturers manuals

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 12: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

Other Checks Accessories

Confirm the presence and condition of accessories (e.g., transducers and coupling gel)

Verify that expiration dates have not been exceeded A copy of the instruction manual should be readily available

to the operators Chassis/housing

Examine the exterior of the scanner for cleanliness and general physical condition

Be sure that plastic housings are intact, that all hardware is present and tight, and that there are no signs of spilled liquids or other serious abuse

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 13: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

AC plug/receptacles • Examine the AC power plug for damage • Wiggle the blades to check for security • Shake the plug and listen for rattles • If damage is suspected, open the plug and inspect it • For receptacles mounted on the scanner - verify the presence of line

power, and insert an AC plug into each and check that it is held firmly. Line cord • Inspect the cord for signs of damage • If damaged, replace the entire cord or, if the damage is near one

end, cut out the defective portion • Be sure to wire a new power cord or plug with the correct polarity

Strain reliefs • Examine the strain reliefs at both ends of the line cord • Be sure that they hold the cord securely

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 14: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

Circuit breaker/fuse • If the scanner has a switch-type circuit breaker, check that it moves

freely • If the scanner is protected by an external fuse, check its value and type

against that marked on the chassis Cables • Inspect any cables (e.g., transducer) and their strain reliefs for general

condition • Carefully examine cables to detect breaks in the insulation and to ensure

that they are gripped securely in the connectors at each end to prevent rotation or other strain

• Verify that there are no intermittent faults by flexing electrical cables near each end and looking for erratic operation or by using an ohmmeter

Transducers • Check the surface areas of ultrasound transducers for deterioration,

cracks, or dents in the membrane • Check the acoustic fluid of mechanically steered transducers; refill with

the recommended fluid if air bubbles are present

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 15: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

Controls/switches • Before changing any controls, check their positions • If any settings appear unusual (e.g., a gain control at

maximum), consider the possibility of inappropriate clinical use or of incipient device failure

• Record the setting of those controls that should be returned to their original positions following the inspection Physical condition Secure mounting, and correct motion Check that control knobs have not slipped on their shafts Control fixed-limit stops, alignment, as well as positive

stopping Each control and switch performs its proper function

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION

Page 16: Without reference, identify principles relating to Diagnostic Ultrasound Imaging Calibration with at least 70 percent accuracy

Indicators/displays • Record the reading of an hour meter, if present • During the course of the inspection, confirm the operation of all lamps,

indicators, meters, gauges, and visual displays on the scanner • Be sure that all segments of a digital display function • Observe an image on the CRT display, and check its quality Also verify

image quality of peripheral accessories Audible signals (Confirm appropriate volume, as well as the operation

of a volume control) ECG - Using an ECG simulator, verify normal operation of ultrasound

scanners configured for echocardiography, according to the following criteria: • Baseline - constant thickness; horizontal and no vertical drift • ECG waveform should be clearly visible, including the P wave and QRS

complex • 50/60 Hz noise should not be present • Intervals between the timing marks displayed by the scanner should be

consistent with the timing of the ECG simulator

DIAGNOSTIC ULTRASOUND IMAGING CALIBRATION