emrp eng06 - d3.4: an ultrasound based temperature ... · an ultrasound based temperature...
TRANSCRIPT
1
EMRP ENG06 - D3.4: An ultrasound based temperature measurement method
Theory, calibration method, equipment, uncertainty
Version Date Author Description
1 17 October 2011 GJK Initial document
2 2 February 2012 GJK First complete draft version
3 21 February 2012 GJK Inclusion of several comments of PK and MvdB
4 29 February 2012 GJK Reordering of some sections
Gertjan Kok VSL Thijsseweg 11 2629 JA Delft the Netherlands
2
1. Introduction ........................................................................................................................................ 4
2. Background ......................................................................................................................................... 4
3. Theory of ultrasound based temperature measurement ................................................................... 5
3.1. General idea ................................................................................................................................. 5
3.2. Ultrasound path average ............................................................................................................. 7
3.3. Representativeness of the path(s) ............................................................................................. 11
3.4. Enthalpic temperature/ energy flow ......................................................................................... 13
3.5. Effect of composition on the speed of sound ............................................................................ 18
3.6. Effect of flow speed ................................................................................................................... 19
3.7. Conclusions of theoretical background part .............................................................................. 20
4. Calibration method for ultrasound based temperature measurement ........................................... 21
4.1. Complete calibration by manufacturer ...................................................................................... 21
4.2. Calibration by user ..................................................................................................................... 22
4.2.1. Ultrasonic path lengths ....................................................................................................... 22
4.2.2. Temperature expansion coefficient .................................................................................... 23
4.2.3. Pressure expansion coefficient ........................................................................................... 23
4.2.4. Transmitter delays .............................................................................................................. 24
4.2.5. Speed of sound reference data ........................................................................................... 25
4.2.5.1. Off-line reference data................................................................................................. 25
4.2.5.2. Continuous on-line calibration reference data ............................................................ 28
4.2.6. Calibration of other auxiliary equipment ............................................................................ 29
4.3. Conclusion about calibration method........................................................................................ 30
5. Required equipment ......................................................................................................................... 31
5.1. Ultrasound sensor for temperature measurement ................................................................... 31
5.2. Auxiliary equipment ................................................................................................................... 32
5.2.1. Pressure sensor ................................................................................................................... 32
5.2.2. Composition determination ................................................................................................ 32
5.2.4. Equipment to be used for calibration ................................................................................. 32
5.2.5. Equipment to be used for validation of the method .......................................................... 34
5.2.6. Required equipment for speed of sound reference data ................................................... 38
5.2.6.1. Required equipment for off-line laboratory reference data ....................................... 38
5.2.6.2. Required equipment for continuous on-line calibration ............................................. 38
3
5.3. Conclusion concerning required equipment ............................................................................. 38
6. Overall theoretical uncertainty budget ............................................................................................ 39
7. Experimental implementation of the different calibration methods ............................................... 40
7.0. Initial wet calibration ................................................................................................................. 40
7.1. Use of a real-time calibrator ...................................................................................................... 40
7.2. Use of external reference data .................................................................................................. 41
7.3. Creation of a traceable reference database .............................................................................. 41
7.4. Conclusion concerning experiments with calibration methods ................................................ 42
8. Overall conclusion ............................................................................................................................. 42
References ............................................................................................................................................ 42
Appendix A: Measured temperature per path ..................................................................................... 43
Appendix B: Full uncertainty budget .................................................................................................... 44
4
1. Introduction This report has been made in the scope of the EU funded EMRP research project ENG06: ‘Metrology
for improved power plant efficiency’. This report is in fulfillment of the deliverable D3.4: ‘Report on
the development of a temperature calibration method and equipment for an ultrasound based
temperature method’.
This report gives theoretical background information concerning a method to measure temperature
with ultrasound in pipe flow. The equipment needed will be described together with a way how it
can be calibrated. In report D3.6 the quality of this measurement principle is described based on
experimental tests, whereas D3.5 deals with a multipoint reference sensor. In the scope of
deliverable D5.4 a webpage-handbook for performing ultrasound temperature measurements in
pipes will be made and in D5.5 a webpage linking to traceable speed of sound/ temperature data will
be created.
2. Background The background of this research topic is the question how the mean temperature of water flowing in
pipe lines where the water has a non-homogeneous temperature distribution can be best measured.
This problem is especially of interest in the hot leg of the primary circuit of nuclear power plants of
pressurized water reactors (PWR). In these pipe lines hot water with temperatures ranging from 310
to 330 °C flows at high speed with local temperature differences up to 20 °C (estimate based on
simulations performed by a nuclear power plant owner). The current temperature measurement
technique is estimated to have an uncertainty of at best 1 °C. In this research project it will be
investigated if the current measurement technique can be improved by means of using an ultrasonic
technique. Another advantage of the ultrasound based measurement would be the high speed with
which the measurement result is acquired and the fast reaction time to process fluid temperature
changes.
5
3. Theory of ultrasound based temperature measurement In this chapter the general ideas behind the method are described. Only wetted ultrasonic sensors
are considered. For a clamp-on version the travel distance of the beam through the wall has to be
considered as well, adding some additional travel time and uncertainty. The trip of the beam
through the fluid itself follows the same logic as for wetted sensors. The application under
consideration will be pipe flows with water. All data in the report will be for water. However the
method can in principal be used for any fluid. The thermodynamic reference data in this report have
been obtained with the help of the NIST Refprop program (see reference [1]) and calculations have
been performed using Microsoft Excel.
3.1. General idea The speed of sound of a fluid changes with temperature. Therefore one can use speed of sound
measurement to derive the fluid temperature. As the speed of sound also depends on fluid pressure
and composition, these parameters need to be known as well. In case of homogeneous temperature
distribution in the pipe the temperature can accurately be measured by means of the speed of
sound (0.02 °C uncertainty may be possible). The idea now is to use this method also for
inhomogeneous temperature distributions. The ultrasound beam will travel through different parts
of the fluid having different temperatures. The beam will thereby consecutively travel at the speed
corresponding to the local fluid temperature. The total travel time is the sum of the travel times in
the different parts of the fluid and the resulting measured speed of sound over the whole traversed
path will be a somehow averaged speed of sound. The idea is to use this average speed of sound to
calculate an average pipe fluid temperature. In Figure 1 the temperature dependency and the
pressure dependency of the speed of sound of pure water are shown. The phase transition from
liquid to vapor is clearly visible. The temperature dependency is much stronger than the pressure
dependency. In Table 1 a few numerical values of the derivatives of c to T and P are shown. The last
column is a measure for the error in calculated temperature in °C by the ultrasound method when
the measured pressure is incorrect by 1 bar.
Figure 1: Speed of sound as function of temperature and as function of pressure for pure water. The phase transition from liquid to vapor is clearly visible.
6
T [°C] P [barg] dc/ dT [m/s/°C] dc/ dP [m/s/bar] (dc/dP) · (dT/dc)
(=dT/dP) [°C /bar]
25 0 2.672 0.168 0.063
25 10 2.676 0.168 0.063
300 100 -6.373 0.919 -0.144
300 150 -5.702 0.807 -0.141
Table 1: Some numerical values of the derivatives of the speed of sound c of water to temperature T and pressure P.
The maximal allowable uncertainty for the measurement of speed of sound and pressure in order to
get a desired uncertainty in temperature (provided the fluid has homogeneous temperature) is
shown in Table 2. The maximal uncertainty is actually divided by 2 in order to allow for different
uncertainty contributions and not just one single maximal one.
Desired uncertainty in
T, u(T)
Nominal
temperature and
pressure
Maximal uncertainty in
c divided by 2, u(c)
Maximal uncertainty in
P divided by 2, u(P)
0.1 °C 20 °C/ 0 barg 0.13 m/s 0.8 bar
0.1 °C 300 °C/ 150 barg 0.3 m/s 0.3 bar
0.01 °C 20 °C/ 0 barg 0.013 m/s 0.08 bar
0.01 °C 300 °C/ 150 barg 0.03 m/s 0.03 bar
Table 2: Maximal uncertainty in speed of sound measurement and pressure to be able to attain a desired uncertainty in the temperature measurement.
The uncertainty aim of 0.1 °C should be possible at both low and high temperatures in view of the
required measurement uncertainty of speed of sound and pressure. At low temperatures 0.01 °C
should also be possible. At high temperatures it might be difficult to measure 150 barg with 0.03 bar
uncertainty (0.02%). Around the temperature of T0 = 70 °C the speed of sound attains a maximum
and the variation of the speed of sound with temperature is relatively small. Therefore it is more
difficult to get accurate values for the temperature in this region. Furthermore, in all cases it should
be known from other sources if the temperature is above or below T0.
7
3.2. Ultrasound path average In this section it will be calculated what the ultrasonic temperature meter theoretically measures for
several inhomogeneous temperature profiles. For this purpose, consider the cross section of a pipe
with coordinate origin in the center, horizontal x-axis and vertical y-axis. Let a vertical ultrasound
path be parameterized by y ranging from R to -R with R the pipe radius and L = 2·R the pipe
diameter. So the beam travels from a point at the top of a horizontal pipe (inside of pipe) to a point
diametrically at the opposite at the bottom of the pipe (inside). The temperature at spot (x,y) equals
T(x,y) and the speed of sound will equal c = c(T(x,y), P) with pressure P. The ultrasound beam transit
time for the vertical path through the pipe center is now given by τ = ∫-RR 1/c(T(0,y), P) dy and the
measured average speed of sound will be cmeas = L/τ, which can be converted to a temperature when
the pressure is known. In Table 4 the difference between the true (arithmetic) average temperature
on a line (Tline,vol = (1/L)·∫-RR T(0,y) dy) and the ‘ultrasound temperature’ Tus (theoretical temperature
that will be measured by means of the ultrasound method) is shown for a few temperature profiles
described in Table 3. The ultrasound path under consideration goes vertically from top to bottom
through the center of the pipe.
Nr
. Temperature profile name Temperature profile mathematical description
Visualizati
on of
profile
1. Homogeneous, T = 25 °C T(x, y) = 25 °C, P = 2 barg
2. Homogeneous, T = 300 °C T(x, y) = 300 °C, P = 150 barg
3. Homogeneous with boundary layer
of 1% of L, Tin = 30 °C, Tout = 20 °C
T(x, y) = 30 °C for 0 < √(x2+y
2) < 0.98 R and T(x, y) = 20 °C
for 0.98 R < √(x2+y
2) < R. P = 2 barg.
4.
Homogeneous with boundary layer
of 0.1% of L, Tin = 300 °C, Tout = 250
°C
T(x, y) = 300 °C for 0 < √(x2+y
2) < 0.998 R and T(x, y) =
250 °C for 0.998 R < √(x2+y
2) < R. P = 150 barg.
5. Two homogeneous circular regions,
Tin = 25 °C, Tout = 24 °C
T(x, y) = 25 °C for 0 < x2 + y
2 < (R/2)
2, T(x, y) = 24 °C for
(R/2)2 < x
2 + y
2 < R
2. P = 2 barg.
6. Two homogeneous circular regions,
Tin = 301 °C, Tout = 300 °C
T(x, y) = 301 °C for 0 < x2 + y
2 < (R/2)
2, T(x, y) = 300 °C for
(R/2)2 < x
2 + y
2 < R
2. P = 150 barg.
7. Two homogeneous circular regions,
Tin = 28 °C, Tout = 22 °C
T(x, y) = 28 °C for 0 < x2 + y
2 < (R/2)
2, T(x, y) = 22 °C for
(R/2)2 < x
2 + y
2 < R
2. P = 2 barg.
8
Nr
. Temperature profile name Temperature profile mathematical description
Visualizati
on of
profile
8. Two homogeneous circular regions,
Tin = 303 °C, Tout = 297 °C
T(x, y) = 303 °C for 0 < x2 + y
2 < (R/2)
2, T(x, y) = 297 °C for
(R/2)2 < x
2 + y
2 < R
2. P = 150 barg.
9. Two equal homogeneous regions, Tup
= 25 °C, Tdown = 24 °C
T(x, y) = 24 °C for y < 0 and T(x, y) = 25 °C for y > 0. P = 2
barg.
10. Two equal homogeneous regions, Tup
= 301 °C, Tdown = 300 °C
T(x, y) = 300 °C for y < 0 and T(x, y) = 301 °C for y > 0. P =
150 barg.
11. Two equal homogeneous regions, Tup
= 28 °C, Tdown = 22 °C
T(x, y) = 22 °C for y < 0 and T(x, y) = 28 °C y > 0. P = 2
barg.
12. Two equal homogeneous regions, Tup
= 303 °C, Tdown = 297 °C
T(x) = 297 °C for y < 0 and T(x, y) = 303 °C for y > 0. P =
150 barg.
13. Two equal homogeneous regions, Tup
= 30 °C, Tdown = 5 °C
T(x, y) = 5 °C for y < 0 and T(x, y) = 30 °C for y > 0. P = 2
barg.
14. Two equal homogeneous regions, Tup
= 330 °C, Tdown = 300 °C
T(x, y) = 300 °C for y < 0 and T(x, y) = 330 °C for y > 0. P =
150 barg.
15. Linear profile from 22 °C to 28 °C T(x, y) = 25 + 3 y/R. P = 2 barg.
16. Linear profile from 297 °C to 303 °C T(x) = 300 + 3 y/R. P = 150 barg.
Table 3: Some theoretic temperature profiles. The temperature at the common boundary of two regions is assumed to be equal to the average of the temperatures of the two adjacent regions.
9
Temperature profile True (arithmetic)
mean temperature
of path, Tline,vol [°C]
Ultrasound
measured path
temperature, Tus
[°C]
Difference Tus
- Tline,vol [°C]
Homogeneous, T = 25 °C 25.000 25.000 0.000
Homogeneous, T = 300 °C 300.000 300.000 0.000
Homogeneous with boundary layer of 1% of L, Tin = 30 °C, Tout
= 20 °C
29.800 29.765 -0.035
Homogeneous with boundary layer of 0.1% of L, Tin = 300 °C,
Tout = 250 °C
299.900 299.933 0.033
Two homogeneous circular regions, Tin = 25 °C, Tout = 24 °C 24.500 24.496 -0.004
Two homogeneous circular regions, Tin = 301 °C, Tout = 300 °C 300.500 300.503 0.003
Two homogeneous circular regions, Tin = 28 °C, Tout = 22 °C 25.000 24.856 -0.144
Two homogeneous circular regions, Tin = 303 °C, Tout = 297 °C 300.000 300.093 0.093
Two equal homogeneous regions, Tup = 25 °C, Tdown = 24 °C 24.500 24.496 -0.004
Two equal homogeneous regions, Tup = 301 °C, Tdown = 300 °C 300.500 300.503 0.003
Two equal homogeneous regions, Tup = 28 °C, Tdown = 22 °C 25.000 24.856 -0.144
Two equal homogeneous regions, Tup = 303 °C, Tdown = 297 °C 300.000 300.093 0.093
Two equal homogeneous regions, Tup = 30 °C, Tdown = 5 °C 17.500 15.161 -2.339
Two equal homogeneous regions, Tup = 330 °C, Tdown = 300 °C 315.000 318.149 3.149
Linear profile from 22 °C to 28 °C 25.000 24.952 -0.048
Linear profile from 297 °C to 303 °C 300.000 300.031 0.031
Table 4: Theoretically measured ultrasound path average compared with true path average for various temperature profiles. The homogeneous circular temperature distributions of Table 3 give the same results as the case of two homogeneous sections with the same temperature difference as the distance travelled by the beam in each temperature area is equal.
The last column of Table 4 shows that the difference between arithmetic mean temperature and
temperature measured by means of ultrasound can be considerable when big temperature
differences exist in the pipe. For temperature differences up to 6 °C the difference may be up to 0.14
°C. In case of differences of 25 °C the ‘ultrasound temperature’ may deviate more than 2 °C from the
arithmetic mean temperature.
To clarify Table 4 the graph in Figure 2 shows the measurement deviation of the ultrasound based
temperature measurement in case that the ultrasound beam travels half of the path in water of T1 =
20 °C and the other half of the path in water of T2 = 20 to 35 °C (pressure 2 barg). As the speed of
sound at lower temperatures is lower in this temperature domain, the beam will spend more time in
10
the colder part and the deviation is negative. The same has been done for temperatures T1 = 300 °C
and T2 = 300 to 330 °C at a pressure of 150 barg, see Figure 3. As the speed of sound in this
temperature domain decreases with increasing temperature, the deviation is positive. The reference
temperature in both cases is the average of the temperatures of both sections, called Tvol
(‘volumetric temperature average’).
Figure 2: Deviation of ultrasound based temperature measurement in case of beam traveling half of the path length through water with T1 = 20 °C and the other half through water with temperature T2. The pressure is 2 barg.
11
Figure 3: Deviation of ultrasound based temperature measurement in case of beam traveling half of the path length through water with T1 = 300 °C and the other half through water with temperature T2. The pressure is 150 barg.
It is seen that in the extreme case of a sharp 10 °C temperature difference in a pipe the ultrasound
based temperature method may be off by -0.4 °C at low temperatures and by 0.3 °C at high
temperatures. With increasing temperature differences this deviations increase rapidly. An
ultrasound temperature meter with multiple paths and smart software could possibly decrease this
deviation. It has to be noted that the studied profile is a very extreme case.
3.3. Representativeness of the path(s) The temperature that an ultrasound beam measures is an estimate for the average temperature
along the beam path. In general this is a better estimate of the average cross section temperature
than a spot temperature measured by a single Pt100 temperature sensor. However, it might still not
be completely representative for the entire cross section. In Figure 4 the difference between the
true section averaged temperature and the ‘ultrasound temperature’ is shown for different path
configurations (see Table 5) and temperature profiles (see Table 3). The numerical data used for this
figure are shown in Table 20 in appendix A. No (boundary) effects like reflection, refraction and
absorption at the interface of warm and cold water have been modeled. The refraction effect might
change the travel path and its length of the ultrasound beam, whereas absorption and reflection
would only cause an attenuation of the measurement signal.
12
Path configuration name Path configuration description Visualization
One vertical path (yielding
temperature Tus,v1)
One path from the top of the
pipe to the bottom of the pipe,
perpendicular on pipe axis.
Three paths through center
(yielding temperature Tus,c3)
Three paths perpendicular on
pipe axis and through pipe
center. One from top to
bottom (0° to 180°), one from
50° to 230° and one from -50°
to 130°. All paths have weight 1
in the temperature calculation.
Three horizontal paths (yielding
temperature Tus,h3)
Three horizontal paths
perpendicular to pipe axis. One
at 28% of (vertical) pipe
diameter, one at 50% and one
at 72%. All paths have weight 1
in the temperature calculation.
Three vertical paths (yielding
temperature Tus,v3)
Three vertical paths
perpendicular to pipe axis. One
at 28% of (horizontal) pipe
diameter, one at 50% and one
at 72%. All paths have weight 1
in the temperature calculation.
Table 5: Different path configurations for ultrasonic temperature sensor. All paths are perpendicular on the pipe axis.
13
Figure 4: Deviation of ultrasound based mean temperature for different path configurations and several temperature profiles. The reference temperature Tref is the volumetric mean section temperature Tsec,vol.
From Figure 4 it can be seen from the boundary layer and circular section temperature profiles (nr. 3
to 8) that all ultrasound path configurations miss the relative large section area that is in the outer
ring and therefore the temperature estimate is somewhat too high. For energy flow measurements
in practice this effect may however be compensated by a flow speed profile in which the speed close
to the wall is lower than in the center. Thus the larger boundary area travels at lower speed than the
smaller center area.
Furthermore, the points in the figure for profiles 13 and 14 seem to suggest that a design with three
horizontal layers gives the best results. However this is only true in this particular case, because of
the perfectly symmetrical temperature patterns and the three horizontal beams having the same
perfect symmetry. Further research with irregular flow patterns is needed. For this purpose a
software tool is planned which may be available on the Webpage-handbook of D5.4.
3.4. Enthalpic temperature/ energy flow Until now the focus was on the arithmetic path averaged and (arithmetic) section averaged
temperatures. However what really matters in the application in power plants is the energy or
enthalpy flow which is given by the enthalpy flow integral over the cross section A of the pipe
(section coordinates x and y, dA = dx·dy): H = ∫A v(x,y) · ρ(T(x,y), P) · h(T(x,y), P) · dA. Remark that the
local enthalpy is weighted with local (mass) density ρ(T(x,y), P) and local flow speed v(x,y)
(orthogonal to cross section dA). The enthalpic temperature Tflow,enth is now defined as the
14
temperature that a fluid of homogeneous temperature of the same mass flow rate would have if it
transported the same amount of enthalpy. In formula form this yields:
Tflow,enth = h-1 P [ (∫A v(x,y) · ρ(T(x,y), P) · h(T(x,y), P) · dA) / (∫A v(x,y) · ρ(T(x,y), P) · dA) ],
where h-1 P[ . ] denotes the inverse enthalpy function as function of temperature at (constant)
pressure P. As hP(T) is almost linear at low temperatures, this function can for practical applications
at low temperatures be simplified to:
Tflow,enth ≈ Tflow,mass = (∫A v(x,y) · ρ(T(x,y), P) · T(x,y) · dA) / (∫A v(x,y) · ρ(T(x,y), P) · dA).
This is the mass flow averaged section temperature. The enthalpic temperature will be of interest in
the experimental testing of the method, which will be described in report D3.6.
In Table 7 the differences between the various definitions of average temperature are shown.
Hereby the notation and definitions of Table 6 are being used. The calculation of the enthalpic
temperature Tflow,enth also needs a definition of a flow pattern. When a homogeneous bulk flow
pattern is assumed (i.e. v(x,y) = constant) however, the flow speed v(x,y) drops out from the defining
equation for Tflow,enth. The symbol c-1P denotes the inverse function of speed of sound in order to
calculate temperature from a given speed of sound at a pressure P. To simplify notation the variable
P and subscript P for pressure will be dropped from ρ(T,P), h(T,P), c-1P and h-1
P.
Symbol Description Formula
Tus Temperature that a (wetted) ultrasound sensor
will measure on a path from top to bottom (or
bottom to top) of a pipe.
Tus = c-1 [L/(∫-RR 1/c(T(0,y)) dy)]
Tline,vol Arithmetic mean temperature on a line (e.g.
ultrasound path) from top to bottom of a pipe.
As there is no weighing with mass, it’s volume
based. The average of a big number of
temperature sensors on a row would tend
towards this value.
Tline,vol = (1/L)·∫-RR T(0, y) dy
Tline, mass Arithmetic mean temperature on a line (e.g.
ultrasound path) weighted with density on a
line from top to bottom of a pipe.
Tline, mass = (∫-RR ρ(T(0,y))·T(0,y) dy) /
∫-RR ρ(T(0, y)) dy
Tline, enth Arithmetic mean temperature on a line (e.g.
ultrasound path) using enthalpy on a line from
to bottom of a pipe. Physical true mean
temperature after mixing the fluid on the line.
Tline, enth =h-1[ (∫-RR ρ(T(0, y))· h(T(0, y))
dy) / ∫-RR ρ(T(0, y)) dy]
15
Symbol Description Formula
Tsec,vol Arithmetic mean temperature over entire
circular cross section of a pipe. One could also
call it the volume mean temperature. The
average of a big number of temperature
sensors spread over the cross section would
tend towards this value.
Tsec,vol = (1/A)·∫A T(x, y) dA
Tsec, mass Arithmetic mean temperature over entire
circular cross section of a pipe weighted with
density.
Tsec, mass = (∫A ρ(T(x, y))·T(x,y) dA) /
∫A ρ(T(x, y)) dA
Tsec, enth Arithmetic mean temperature over entire
circular cross section of a pipe using enthalpy.
Physical true mean temperature after mixing
the fluid in the section.
Tsec, enth =h-1[ (∫A ρ(T(x, y))·h(T(x, y)) dA)
/ ∫A ρ(T(x, y)) dA]
Tflow,vol Arithmetic mean temperature over entire
circular cross section of a pipe weighted with
flow speed.
Tflow,vol = (∫A v(x,y) · T(x,y) dA) / (∫A v(x,y)
dA)
Tflow,mass Arithmetic mean temperature over entire
circular cross section of a pipe weighted with
flow speed and density.
Tflow,vol = (∫A v(x,y) · ρ(T(x,y)) · T(x,y) dA)
/ (∫A v(x,y) · ρ(T(x,y)) dA)
Tflow,enth Arithmetic mean temperature over entire
circular cross section of a pipe weighted with
flow speed and using enthalpy. This is the true
energy or enthalpy flow mean temperature.
Tflow,enth = h-1 [ (∫A v(x,y) · ρ(T(x,y)) ·
h(T(x,y)) dA) / (∫A v(x,y) · ρ(T(x,y)) · dA)]
Table 6: Different ways of calculating an average line, section and bulk flow temperature.
The most relevant temperatures are Tus as it is the temperature the ultrasound sensor will give,
Tsec,vol as it is the temperature that a set of temperature sensors mounted in the pipe section (e.g.
Pt100s) would have as its mean temperature and Tsec, enth as it represents the true physical
temperature of the fluid when the cross section would be (adiabatically) mixed to a homogeneous
temperature. The other temperatures are mainly approximations of the true physical enthalpic
temperatures which are easier to calculate. It is instructive to see the size of the differences. The
flow speed weighted enthalpic temperatures Tflow,sec, Tflow,mass, Tflow,enth are not shown Table 7, as that
would need an assumption of the flow profile over the cross section of the pipe. This is not done
here. (If the flow profile is homogeneous bulk flow v(x,y) = constant, then Tflow,xxx = Tsec,xxx).
16
Temperature profile Tus
[°C]
Tline,vol
[°C]
Tline, mass
[°C]
Tline, enth
[°C]
Tsec,vol
[°C]
Tsec, mass
[°C]
Tsec, enth
[°C]
Homogeneous, T = 25 °C 25.000 25.000 25.000 25.000 25.000 25.000 25.000
Homogeneous, T = 300 °C 300.000 300.000 300.000 300.000 300.000 300.000 300.000
Homogeneous with boundary layer
of 1% of L, Tin = 30 °C, Tout = 20 °C
29.765 29.800 29.799 29.799 29.604 29.603 29.603
Homogeneous with boundary layer
of 0.1% of L, Tin = 300 °C, Tout = 250
°C
299.933 299.900 299.890 299.890 299.800 299.780 299.781
Two homogeneous circular regions,
Tin = 25 °C, Tout = 24 °C
24.496 24.500 24.500 24.500 24.250 24.250 24.250
Two homogeneous circular regions,
Tin = 301 °C, Tout = 300 °C
300.503 300.500 300.500 300.500 300.250 300.250 300.250
Two homogeneous circular regions,
Tin = 28 °C, Tout = 22 °C
24.856 25.000 24.998 24.997 23.500 23.498 23.498
Two homogeneous circular regions,
Tin = 303 °C, Tout = 297 °C
300.093 300.000 299.974 299.995 298.500 298.481 298.496
Two equal homogeneous regions,
Tup = 25 °C, Tdown = 24 °C
24.496 24.500 24.500 24.500 24.500 24.500 24.500
Two equal homogeneous regions,
Tup = 301 °C, Tdown = 300 °C
300.503 300.500 300.500 300.500 300.500 300.500 300.500
Two equal homogeneous regions,
Tup = 28 °C, Tdown = 22 °C
24.856 25.000 24.998 24.997 25.000 24.998 24.997
Two equal homogeneous regions,
Tup = 303 °C, Tdown = 297 °C
300.093 300.000 299.974 299.995 300.000 299.974 299.995
Two equal homogeneous regions,
Tup = 30 °C, Tdown = 5 °C
15.161 17.500 17.473 17.456 17.500 17.473 17.456
Two equal homogeneous regions,
Tup = 330 °C, Tdown = 300 °C
318.149 315.000 314.174 314.969 315.000 314.174 314.969
Table 7: Differences between various ways of calculating a mean temperature. Tflow,vol , Tflow,mass and Tflow,enth are not calculated, as that would need an assumption about the flow profile.
In order to clarify the numbers of Table 7 the true temperature after mixing using enthalpy has been
compared with the approximations based on volume weighted averaging and mass weighted
averaging. Half of the volume has a fixed temperature T1 = 20 °C (c.q. 300 °C) and the other half has
temperature T2 ranging from 20 to 35 °C (c.q. 300 to 330 °C). See Figure 5 and Figure 6.
17
Figure 5: Error involved in approximate formulae to calculate temperature after mixing in case of half of the volume having temperature T1 = 20 °C and half of the volume having temperature T2 =20 to 35 °C. Pressure is 2 barg.
Figure 6: Error involved in approximate formulae to calculate temperature after mixing in case of half of the volume having temperature T1 = 300 °C and half of the volume having temperature T2 =300 to 330 °C. Pressure is 150 barg.
18
It is concluded that in case of low temperature the volume weighted approximate temperature has
an error less than 0.01 °C for temperature differences less than 10 °C. The mass weighted
approximate temperature is very good and has an error of 0.002 °C for a temperature difference of
15 °C.
In case of high temperatures the volume weighted approximate temperature has an error less than
0.05 °C for temperature differences up to 30 °C. The mass weighted approximate temperature has
an error less than 0.05 °C for temperature differences less than 8 °C. The fact that the mass weighted
temperature deviates is related to the fact that the water density is very sensitive to temperature
variations. The calorific value cp also changes with temperature. The final result is that the volume
weighted temperature is a very good approximation of the true temperature after mixing at high
temperatures.
3.5. Effect of composition on the speed of sound The speed of sound is directly dependent on the fluid composition. In this section it is studied what
the effect is of impurities in water on the ultrasound based temperature method. It is assumed that
the used reference data is valid for pure water, whereas the real fluid is water with a tiny bit of
ammonia dissolved or salt water with different salinities. In Table 8 the error induced in the
ultrasound based temperature measurement is shown for an ammonia solution at different
temperatures, for drinking water and for sea water. The ammonia concentration of 3.8 mg/l is the
highest concentration that was found to be used in power plants according to the survey that was
made.
The speed of sound in drinking and sea water is calculated by means of the equation of Coppens
(1981), see reference [4].
Real fluid Used data
base fluid
Nominal T and P Resulting error in
measured T
Ammonia solution (3.8 mg/l) Pure water 20 °C/ 2 barg 0.006 °C
Ammonia solution (3.8 mg/l) Pure water 300 °C/ 150 barg < 0.001°C
Drinking water (salinity 100 ppm) Pure water 20 °C/ 2 barg 0.036 °C
Limit drinking water (salinity 1000 ppm) Pure water 20 °C/ 2 barg 0.36 °C
Sea water (salinity 35 000 ppm) Pure water 10 °C/ 0 barg 13 °C
Table 8: The influence of water composition on measured temperature by means of ultrasound method.
19
According to the survey, the following chemicals are or were used in power plants (process water
and cooling water):
- Ammonia (0.2 – 4 mg/L) - Boric acid (0 - 1500 ppm) - Chlorine - Copper (0.02 - 0.2 mg/L) - Ferrous sulphate - Hydrazine (0.025 - 0.3 mg/L)
- HCL - H2 - Iron (0.01 - 10 mg/L) - Lithin - NaOH - Silicon (0.02-0.1 mg/L)
As concentrations are low, no significant influence on the speed of sound is expected. This was
verified for ammonia both based on RefProp reference data and experimentally, and for boric acid
experimentally.
3.6. Effect of flow speed In this section possible effects of the flow speed on the ultrasound based temperature measurement
method are studied.
An ultrasound transit time (flow) meter measures transit times t1 and t2 of two ultrasonic pulses
travelling in opposing directions between the transmitters. From the known path length L between
the two transducers and the angle α between the path and the pipe axis, flow speed in pipe axis
direction v and (approximate) speed of sound capp are calculated in the following way:
v = L/(2·cos(α)) · (t2 – t1)/(t1·t2)
capp = L/2 · (t1 + t2)/(t1·t2)
The speed of sound calculated in this way, is an approximation if the ultrasound transducers are
modeled as point sized and the water is homogeneous and flowing with speed v in the direction of
the pipe axis. The approximation lies in the fact that the effect of the ultrasound beam being
transported in axial direction by the fluid flow is neglected. At least theoretically this would imply
that the beam will not land on the point-sized transducer at the other side of the pipe, but
somewhat downstream. The correct formula for calculating the speed of sound in this setting is:
c = capp √[1 + (v/capp)2 · sin2(α)]
capp = c √[1 - (v/c)2 · sin2(α)]
and thus (slightly) depends on flow speed v. With transducers orthogonal to the pipe axis (α = 90°)
the inaccuracy is maximal and shown in Table 9.
This effect is not present if the point approximation of the transducers is invalid. For small pipes this
might indeed be the case: in a 0.1 m diameter pipe, with v = 10 m/s, the beam is displaced 0.7 mm
parallel to the pipe axis. As the transducer diameter will be much bigger than 0.7 mm, there won’t
20
probably any effect. For a pipe diameter of 1 m and a flow speed of v = 20 m/s, the beam will be 21
cm displaced parallel to the pipe axis. In this case, the effect may be present, as the transducers will
have a much smaller diameter than 21 cm.
Another effect is that by Bernoulli’s law the hydrostatic pressure is dependent on v, as the
hydrostatic pressure is diminished by the dynamic pressure ½ · ρ · v2. The fluid feels the static
pressure which is the total pressure diminished by the dynamic pressure. If the total pressure is
measured instead of the static pressure, then this will result in a final temperature measurement
error, as shown in Table 9.
This effect is not present if the pressure transducers measure static pressure and not total pressure.
This is probably the case. Care should be taken that they measure at a representative spot.
If the just described effects are present, in case of very high flow speeds and when additionally very
high accuracies are desired a flow meter for estimating the flow speed might be needed. In case of
using an ultrasonic flow meter for the ultrasound based temperature measurement this flow speed
information is readily available.
Nominal T
and P
Flow speed
v
Error in T due
to use of capp
Error in T due to
unmeasured reduced
hydrostatic pressure
Total resulting error in
measured T
20 °C/
2 barg
5 m/s -0.003 °C -0.007 °C -0.009 °C
20 °C/
2 barg
10 m/s -0.011 °C -0.027 °C -0.038 °C
300 °C/
150 barg
10 m/s 0.009 °C 0.051 °C 0.060 °C
300 °C/
150 barg
20 m/s 0.036 °C 0.206 °C 0.242 °C
Table 9: Induced temperature measurement error due to flow speed for orthogonal transducers (α = 90°). Both effects
probably don’t occur in practice for small size meters. For big size meters the first effect may occur, but the influence remains very small.
3.7. Conclusions of theoretical background part The method of measuring temperature with ultrasound is a priori a powerful idea. For homogeneous
temperatures measurement uncertainties down to 0.01 °C are theoretically possible for
temperatures around 20 °C. For high temperatures 0.1 °C seems a more realistic goal. In case of
temperature differences below 1 °C, the ultrasound method can still perform with the same
uncertainty. In case of higher temperature differences, the ultrasound temperature method path
mean will start to deviate from the path mean or true enthalpic mean temperature more and more.
For heterogeneous irregular temperature distributions with local temperature difference up to 10 °C
a measurement uncertainty of a few 0.1 °C is theoretically possible. At higher temperatures the
21
ultrasound method as presented here may start to deviate more. This can possibly be amended by
using multiple paths and some smart software.
Around the temperature of T0 ≈ 70 °C temperature measurement based on speed of sound is
difficult, as the speed of sound attains a maximum at this temperature. Furthermore, in all cases it
should be known from other sources if the temperature is above or below T0.
The flow speed has no significant effect on the method. Also the small amount of additives added to
the process water in power plants are not expected to influence the measurement in a noticeable
way.
In the next chapter the temperature calibration method will be described before moving on with a
more detailed description of the equipment.
4. Calibration method for ultrasound based temperature
measurement As is the case for most measurement devices, also an ultrasonic temperature measurement device
needs calibration. It depends however on the device, its manufacturer and the application of the
user how much work the user needs to perform for the calibration. Different cases will be
considered.
4.1. Complete calibration by manufacturer In this case the user can just install the ultrasound sensor in his pipe work and start working with it
without any calibration to be performed by the end user. The sensor should be equipped with its
own pressure measurement and the used fluid should be (sufficiently) equal to the fluid the
manufacturer had in mind (and possibly used in a factory calibration). Also the device should be very
robust such that ultrasonic path lengths don’t change when installing the device in the piping.
Therefore clamp on sensors delivered without any piping seems to be excluded for the ideal case, as
the determination of ultrasonic path length will need at least some calibration or configuration
effort.
In case the device doesn’t have its own pressure measurement, it should be possible to feed the
device with a representative pressure measurement by means of an analog or digital signal to the
temperature computing unit.
Finally the manufacturer should have implemented device software in which the temperature is
calculated from the measured ultrasonic transit times and pressure for the fluid in the piping.
One important issue to address is that the device should know if the temperature is above or below
T0 , the temperature at which the speed of sound attains a maximum (T0 ≈ 70 °C). This can either be
established by a conventional temperature measurement performed by the device, by an external
22
temperature measurement fed to the device or, most commonly, by prior knowledge about the
approximate fluid temperature. Furthermore, temperature measurement based on speed of sound
is problematic around T0.
4.2. Calibration by user In the basic case the end user only gets the ultrasonic transit times from the device. In this case he
has to determine some numerical factors himself. The user can either plug some theoretically
determined factors in the device or perform a real calibration (experimentally). For the calibration a
flow loop is needed in which homogeneous flow at different temperatures and pressures can be
created. Experimental calibration data are strongly preferred as they better reflect reality and can
also take into account or at least show not modeled effects, which would otherwise be overlooked.
These calibration factors will now be discussed as well as their approximate size and importance.
4.2.1. Ultrasonic path lengths
To calculate the speed of sound from transit times, the ultrasound path lengths should be divided by
the corresponding transit times. The path lengths could be measured geometrically before
installation of the device, taking into account the wall thickness for clamp on sensors. Another
possibility is to let flow a homogeneous fluid with known temperature and pressure and known
pressure-temperature-speed of sound relationship through the sensor. The homogeneous
temperature could be read from a reference temperature sensor close to the device and the same
for the pressure. The speed of sound can now be taken from the database and from this the path
lengths can be recalculated.
The contribution of the ultrasonic path length uncertainty to the total measurement uncertainty in
the temperature is shown in Table 10.
Parameter Nominal
value of
parameter
Uncertainty in
parameter
Nominal T
and P
Resulting
uncertainty in
measured T
Ultrasonic path
length
100 mm 0.001 mm 25 °C/ 0
barg
0.006 °C
Ultrasonic path
length
1000 mm 0.010 mm 300 °C/ 150
barg
0.002 °C
Table 10: Contribution of the ultrasonic path length uncertainty to the total measurement uncertainty of the temperature.
An inaccuracy in path length of 1 μm on a total path length of 100 mm gives an inaccuracy of 1/105
in the speed of sound which amounts to 1/105·1500 = 0.015 m/s or 0.006 °C. For comparison: a
human hair has a thickness of about 60 μm.
23
4.2.2. Temperature expansion coefficient
Depending on the pipe wall material the pipe diameter and thus also the ultrasonic path lengths will
change with temperature. Therefore the path lengths need correction with the best known wall
temperature estimate before calculating the speed of sound from path length and transit time. In
good approximation the temperature expansion correction can be taken in form of the linear
expansion coefficient α times the temperature increase from calibration reference temperature.
One can take a database value if the pipe material is known (e.g. 16 ppm/K for SS316). Another
possibility is to let flow a homogeneous fluid with known temperature and pressure and known
pressure-temperature-speed of sound relationship through the sensor and repeating this at different
temperatures. From the resulting data (transit times), the measured pressure and reference
temperature and the database data, the temperature expansion coefficient can be deduced.
The uncertainty contribution of the pipe temperature expansion to the total measurement
uncertainty of the measured temperature can be found in Table 11.
Parameter Nominal
value of
parameter
Uncertainty
in
parameter
Nominal T
and P
Resulting
uncertainty in
measured T
Temperature expansion
coefficient α (used for
extrapolation from 20 °C)
16 ppm/K 1 ppm/K 25 °C/ 0
barg
0.003 °C
Temperature expansion
coefficient α (used for
extrapolation from 20 °C)
16 ppm/K 1 ppm/K 300 °C/ 150
barg
0.05 °C
Wall temperature 25 °C 1 °C 25 °C/ 0
barg
0.009 °C
Wall temperature 300 °C 5 °C 300 °C/ 150
barg
0.01 °C
Table 11: Uncertainty contribution of the pipe temperature expansion to the total uncertainty in temperature measurement.
When the ultrasonic temperature measurement is corrected for pipe wall expansion, the remaining
uncertainty component is at most 0.05 °C, which is small compared to other effects. Nevertheless
the temperature expansion correction needs to be implemented somehow.
4.2.3. Pressure expansion coefficient
Depending on the pipe wall material and thickness the pipe diameter and thus also the ultrasonic
path lengths will change with pressure. However for most materials and low pressures, this effect
24
will be very minor. In the general case the path lengths need correction for the pressure. In good
approximation the pressure expansion correction can be taken of linear form. One can take a
theoretical value for the Young modulus E if the pipe material is known (e.g. E = 200 GPa for SS316),
measure or look up the pipe wall thickness b and use an expansion formula for thin or thick walled
pipes (e.g. dL/L = P L/ (2 b E) = β P) for thin walled pipes with diameter L and wall thickness b) to
calculate the pressure expansion coefficient. Another possibility is to let flow a homogeneous fluid
with known temperature and pressure and known pressure-temperature-speed of sound
relationship through the sensor and repeating this at different pressures. From the resulting data
(transit times), the measured pressure and reference temperature and the database data, the
pressure expansion coefficient can be deduced (which might also take into account some other
pressure effects).
The contribution of neglecting the pressure expansion for thin walled pipes to the total
measurement uncertainty (deviation/ error) in temperature is shown in Table 12.
Parameter Nominal
value of
parameter
Uncertainty in
parameter
Nominal T
and P
Resulting
uncertainty/
error in
measured T
Uncompensated
pressure expansion
(thin wall
coefficient: β= L/ (2
b E), L = 100 mm, b =
3 mm, E = 200 GPa)
β = 8.3 10-11
bar-1
8.3 10-11 bar-1 25 °C/ 2
barg
0.009 °C
Table 12: Error in ultrasound temperature measurement due to uncertainty in or negligence of pressure expansion of pipe for thin walled pipes.
At high pressures and temperatures, the pipe wall will be very thick, so that the thin wall
approximation is not valid anymore. If the pressure effect on the wall is bigger than desired, the pipe
wall thickness can be increased in the design (usual approach), or a correction factor can be applied
in the meter software. In any case the remaining uncertainty component is estimated to be relatively
small. Usually the pressure expansion correction doesn’t need to be determined.
4.2.4. Transmitter delays
The transit times measured by of the ultrasonic device τtransmitter might not be completely corrected
for delays τdelay in electronics and pipe walls. One can model this by the equation τtransit = τtransmitter –
τdelay for the true transit time τtransit. One can determine these delays by making a best fit through the
calibration data taken at a flow of homogeneous fluid with known temperature and pressure and
known pressure-temperature-speed of sound relationship through the sensor and repeating this at
different temperatures and pressures. From the resulting data (transit times), the measured
pressure and reference temperature and the database data, the temperature expansion coefficient,
25
the pressure expansion coefficient and the transmitter delays can be deduced. Normally this is not
needed however.
The uncertainty contribution of the transit time measurement is shown in Table 13.
Parameter Nominal
value of
parameter
Uncertainty in
parameter
Nominal T
and P
Resulting
uncertainty in
measured T
Transmitter delay
(D = 100 mm pipe,
τtransit = 67.446 μs )
0 μs 0.001 μs 25 °C/ 2
barg
0.008 °C
Transmitter delay
(D = 1000 mm pipe,
τtransit = 1034.531 μs )
0 μs 0.001 μs 300 °C/ 150
barg
0.0002 °C
Table 13: Uncertainty of ultrasound temperature measurement method due to uncertainty in transit time.
An inaccuracy in measured transit time of 1 ns on a total path length of 100 mm gives an inaccuracy
of 1.5/105 in the speed of sound which amounts to 1.5/105·1500 = 0.022 m/s or 0.008 °C at low
temperature and pressure. The uncertainty of this component is estimated to be relatively small and
normally this correction term doesn’t need to be determined.
4.2.5. Speed of sound reference data
The speed of sound reference data are not a calibration factor, but they are needed to convert
measured speed of sound (together with knowledge about pressure and composition) to a
temperature. This reference data can come under different forms: e.g. tables with values, an
equation of state, some polynomial functions, etc. All these forms will simply be called reference
data. In the following subsections several cases are considered for the type (not the form) of
reference data that can be used.
4.2.5.1. Off-line reference data
In this case an independent database with traceable speed of sound is being used. The database
gives the relationship between temperature, pressure and speed of sound for a given fluid. Once it is
established that the ultrasonic device is able to measure speed of sound correctly at different
pressures and temperatures, nothing else needs to be calibrated. Only the reference data have to be
looked up.
In deliverable D5.5 of this research project a webpage with references to existing traceable speed of
sound data will be created. If the end user wants to create high quality traceable reference data by
himself, considerable investments and work is needed. Description of the needed equipment
26
surpasses the scope of this report. Reference values can for example be taken from the equation of
state made by IAWPS for water (see Ref. [3]), or from the NIST RefProp program (see [1]).
The speed of sound reference data possess an uncertainty. The NIST RefProp program states an
uncertainty of 0.1 % (see reference [2]). The uncertainty for the speed of sound of water given by
IAWPS is shown in Figure 7. It has to be noted that here the speed of sound is calculated from a
general equation of state, which is in turn fitted based on a certain number of measurements.
Particular measurements of the speed of sound might have lower uncertainty. These data are
however harder to find and may only cover a small part of all interesting temperatures and
pressures. There also exist some particular equations which give speed of sound as function of
temperature and pressure with lower uncertainty in some small regions, especially at ambient
pressure and relatively close to ambient temperature.
In Table 14 the contribution of the uncertainty in speed of sound reference data to the total
measurement uncertainty is shown. This uncertainty can be reduced by a direct calibration of the
device with the correct fluid, temperature and pressure.
27
Figure 7: Uncertainty of speed of sound reference values for water given by IAPWS (see reference [3]).
Parameter Nominal
value of
parameter
Uncertainty in
parameter
Nominal T
and P
Resulting
uncertainty in
measured T
Speed of sound
reference value
1496.70
m/s
0.03 % / 0.45
m/s
25 °C/ 2
barg
0.17 °C
Speed of sound
reference value
966.6 m/s 0.2 % / 1.9 m/s 300 °C/ 150
barg
0.3 °C
Table 14: Uncertainty in ultrasound temperature measurement method due to uncertainty in database reference values.
From Table 14 it is seen that speed of sound reference data actually possesses a relatively high
uncertainty when converted to the induced uncertainty in ultrasonic temperature measurement.
28
This means that it is not possible to get low uncertainties in the ultrasound temperature
measurement without a wet temperature calibration of the device at approximately the working
temperature. The error in the speed of sound reference data will be systematic for small
temperature ranges, but might be random for larger temperature ranges. This means that
calibration data and speed of sound reference data cannot be extrapolated from e.g. 20 °C to 300 °C.
4.2.5.2. Continuous on-line calibration reference data
In this case the idea is that the needed speed of sound reference data is created by the end user
himself, using exactly the same fluid as in the process. Some liquid is taken out of the process, is
heated to the desired temperature and then pressure, temperature and speed of sound are
measured simultaneously in order to find their relationship. In Figure 8 this is schematically shown.
This method avoids the need for determining the fluid composition and the search for database
reference data for the pressure/ temperature/ speed of sound relationship for the particular fluid
composition. Another advantage is the quick response to changes in fluid composition. What is still
needed is an initial calibration of the main ultrasound device used in the process together with the
ultrasound device used for establishing the curve at different temperatures and preferably also
different pressures and compositions. This is needed to verify that both devices measure the same
(non-traceable) approximate speed of sound at different conditions. After this initial calibration the
on-line calibrator can be used for establishing the current reference curve for the current
composition.
29
Figure 8: Set-up for on-line calibrator of a speed of sound based temperature meter.
The disadvantages are the required additional investments for the measurement set-up and,
sometimes more important, the need of having a sampling point in the process. In the application in
the primary circuit of a nuclear power plant, this would mean taking out some radio-active water out
of the primary circuit. In practice this will never be done if it can be prevented. Furthermore, from
discussions with plant owners it followed that the fluid composition (concentration of chemicals
dissolved in the water) changes only very slowly. The interesting time scale is rather months than
hours. Therefore the off-line laboratory reference data approach seems to be more relevant for
practical applications together with monitoring of the water composition on a regular basis. Besides
that, the low concentration of chemicals used in power plants are expected to change the speed of
sound only in a very minor way, irrelevant in the current measurement uncertainty. This has been
verified for two chemicals (ammonia and boric acid) in the experimental tests and for ammonia also
using NIST RefProp reference data.
4.2.6. Calibration of other auxiliary equipment
The calibration of other auxiliary equipment (like pressure and temperature sensors, see next
chapter) is standard and will therefore not be described in this report.
30
4.3. Conclusion about calibration method In this chapter several issues related to the calibration needed for the ultrasonic temperature
measurement method were considered. Because of the limited accuracy of speed of sound
reference data and because it is difficult to measure speed of sound well without prior wet
calibration (e.g. only geometrical and theoretic calibration), a wet calibration of the device is needed
at the temperature at which the device will be used to get a good estimate of the path length.
Dependent on the construction of the device, the envisaged accuracy and on the range of working
temperatures, pressures and compositions, the temperature and pressure expansion coefficients
and transmitter delays may be calibrated as well. Often some theoretical values are good enough to
get a fair accuracy for the ultrasound based temperature measurement.
In case the process fluid only slowly changes composition and when reference data in literature
exist, it will be convenient to use the existing reference data instead of using a separate device to
determine the reference data concerning the temperature-pressure-speed-of-sound relationship
oneself.
In Table 15 a summary of the different calibration methods is given and when each of them is the
method of choice.
Calibration method Advantages Disadvantages Preferred when
Method 1:
Initial wet calibration,
use of real-time
calibrator
- Fast response to
changing composition
- No traceability
needed for speed of
sound
- Additional equipment
needed
- Additional calibration
and simultaneous
testing of devices
needed
- Fast changing
composition of process
fluid
- Unknown fluid
composition or use of
a fluid for which no
reference data exists
Method 2:
Initial wet calibration,
use of existing
(traceable) database
with reference data for
speed of sound
- No additional
equipment needed
- Database for all
relevant compositions,
pressures and
temperatures has to
be found
- Fluid composition
measurement might
be needed
- Fluid composition is
known and reference
data exists
31
Calibration method Advantages Disadvantages Preferred when
Method 3:
Initial wet calibration,
creation of traceable
database for speed of
sound
- Arbitrary
composition, pressures
and temperatures can
be used
- Database creation
and process can be
physically separated,
no simultaneous
calibration needed.
- Costly equipment is
needed for the
creation of a high
quality database and
additional work is
involved
- Composition
measurement of
process fluid might be
needed
- No reference data
exists for the fluid
- Establishment of
reference data and
main process should
be physically
separated
- Desire of universal
and traceable
reference data for
speed of sound
Table 15: Summary of different calibration concepts for an ultrasound based temperature method.
5. Required equipment This section will describe the needed equipment for an ultrasound based method for measuring
temperature in more detail. The main component is the ultrasound sensor, but also some other
equipment is needed.
5.1. Ultrasound sensor for temperature measurement The ultrasound sensor is the main part of the equipment. In order to measure the speed of sound, a
transit time based ultrasonic flow meter can be used. An alternative is to have a spool piece with
ultrasound transducers mounted orthogonally to the pipe axis on it. In both cases transit times of
short sound pulses at some ultrasonic frequency (e.g. 1 MHz) are being measured. These transit
times can either be a directly transferred to the user or converted to a speed of sound or
temperature by the (flow) meter transmitter itself.
The orthogonal design is more compact and the cross section measured by the ultrasound beams is
an orthogonal cross section and not a section on an angle, which may be advantageous when
dealing with irregular flow patterns. It may also allow for more accurate speed of sound
measurements. The advantage of an ultrasonic flow meter is obviously that it also measures the flow
rate.
Transit time ultrasonic flow meters are available from a number of commercial manufacturers.
However it depends on the manufacturer if measured transit times, speed of sound and/or
temperature are available as output from the device. Calibration of the device was discussed in the
last chapter.
32
5.2. Auxiliary equipment In order for the ultrasonic temperature method to work, some auxiliary measurement equipment is
needed. This consists of equipment directly used for the determination of the temperature (namely
pressure and possibly fluid composition measurement) and equipment used for calibration and
validation of the device.
5.2.1. Pressure sensor
As speed of sound depends on both pressure and temperature, it is not enough to measure only the
speed of sound for deducing the temperature. The hydrostatic pressure of the fluid at the position of
the ultrasound beam is needed as well. Thus a pressure sensor is needed. The required uncertainty
is determined by the sensitivity of the speed of sound to pressure variations (see Table 1 and Table
2). The influence of pipe wall expansion due to pressure is in general less important (see Table 12).
5.2.2. Composition determination
The speed of sound varies with the composition of the fluid. If one decides to use off-line speed of
sound reference data, the composition should be known with sufficient accuracy. The composition
should either be known beforehand and not change, or, when larger changes in composition are
expected to occur, should be measured on a regular basis with a suitable device. There is a number
of companies offering equipment for chemical analysis and monitoring the composition of fluids.
Depending on the estimated composition and the desired accuracy a device has to be selected.
If an on-line calibration reference data method is being used, the composition doesn’t need to be
determined. The on-line calibration device will automatically produce adapted calibration values in
case fluid composition changes.
In the case of power plants the concentration of solved additives into the process water is so low,
that no noticeable effect on the speed of sound is expected (deviating from the speed of sound in
pure water) and thus composition determination is not needed for the ultrasound based
temperature measurement method.
5.2.4. Equipment to be used for calibration
In case that ultrasonic path lengths are being measured geometrically a caliper or laser based
method is needed to determine these lengths. This is however unconventional and won’t include
electronic delays and other effects, which can integrally be determined in a direct calibration.
Therefore, this method is not advised.
A better idea is using a flow loop for determining the calibration settings. Besides the flow loop and
knowledge concerning pressure and composition, a reference temperature sensor is needed. Even
better are several reference sensors. With a so-called Pt100-Array, consisting of 9 Pt100
33
temperature sensors mounted at different positions in the cross section of a pipe, the homogeneity
of the fluid temperature can easily be verified, see Figure 9.
Figure 9: A Pt100-array: a spool-piece with 9 Pt100 temperature sensors mounted at different positions in the cross section.
The required uncertainty of a caliper or laser for a geometric calibration can be found in Table 10.
However a wet calibration is preferred. An overview of uncertainty sources in a wet calibration in a
flow loop is given in Table 16.
Uncertainty source Explanation Formula or
symbol
Numerical
estimate of error
or uncertainty
Reference
temperature
uncertainty
Depends a.o. on calibration of
reference, installation effects,
ambient temperature effects,
drift, etc.
δTcal + δTinstallation
+ δTamb + δTdrift
0.02 °C
Imperfect
homogeneous
temperature
Water temperature is non-
uniform over the pipe section.
Can be measured with a Pt100-
array.
(max(Ti) –
min(Ti))/2
0.01 °C
34
Uncertainty source Explanation Formula or
symbol
Numerical
estimate of error
or uncertainty
Heat leak between
ultrasound device
and reference
temperature sensor
Depends on insulation quality
(k), distance between
ultrasound device and reference
(L), mass flow rate (Qm), specific
heat of fluid (cp) and the
temperature difference
between ambient air and the
fluid (Tamb – T)
(Tamb – T) · k · L /
(Qm · cp)
0.001 °C
Time delay between
ultrasound device
and reference
temperature sensor
Effect depending on the size of
the temperature variation of
the water and the flow speed v
and the distance L
(L/ v) * max(
|dT/dt|)
0.01 °C
Fluid heating due to
viscous effects
Due to pipe wall friction and
viscous forces, fluid pressure
will be converted to heat
(pressure drop dP)
dP/(ρ·cp) 0.000 °C
Table 16: Uncertainties involved in calibrating an ultrasonic temperature sensor in a flow loop. The numerical values have to be estimated for each particular case. Example values are shown.
5.2.5. Equipment to be used for validation of the method
The validation of the ultrasound based temperature measurement method for homogeneous water
temperatures can be done with the same equipment as used in the experimental calibration, thus
comparing the ultrasound temperature with a reference temperature.
In order to validate the method for inhomogeneous water temperatures a method for creating
inhomogeneous temperature profiles in the fluid is needed. This can for example be done by
combining two flows of known flow rate and of known homogeneous temperature by using a T-
piece in the flow loop (see Figure 10 left) and two flow meters and two temperature sensors. One
can also design a special flow injector simulating special conditions, e.g. a stratifying flow injector
simulating stratified flow (see Figure 10 middle and right). The Pt100-Array described before can be
used to monitor the actual created temperature profile. The water temperature after perfect
(adiabatic) mixing is the enthalpy flow mean temperature. This temperature will be taken as
reference temperature. For further information see the chapter about the ‘Theory of ultrasound
based temperature measurement’.
35
Figure 10: Classical T-piece (left) and spool-piece for creating stratified flow pattern. There is a flow injection nozzle below the plate.
In Table 17 and Table 18 uncertainty budgets for the reference temperature in the case of mixing of
two flows with different temperature is given. Due to instable conditions, test set-up and delays,
some additional components may have to be taken into account in practice. See also Table 16. In the
last rows of Table 17 and Table 18 the approximate mixing temperature is calculated using mass
flows in order to see if volume flow and mass flow weighted temperatures form a good
approximation of the true temperature determined by enthalpy flows. The measurement equation is
given by
Tmix = h-1[(Q1·ρ1·h1 + Q2·ρ2·h2)/ (Q1·ρ1 + Q2·ρ2)]
and by
Tmix,vol = (Q1·T1 + Q2·T2)/ (Q1 + Q2)
and
Tmix,mass = (Q1·ρ1·T1 + Q2·ρ2·T2)/ (Q1·ρ1 + Q2·ρ2)
for the approximate temperatures using volume respectively mass flows.
Parameter Nominal
value of
parameter
Uncertainty in
parameter
Sensitivity
coefficient
Resulting
uncertainty in
measured T
Q1, flow 1 25 m3/h 0.05% (= 0.013
m3/h)
0.06 °C/ ( m3/h) 0.0007 °C
Q2, flow 2 5 m3/h 0.05 % (= 0.003
m3/h)
-0.28 °C/ (
m3/h)
0.00007 °C
T1, temperature of flow 1
30 °C 0.02 °C 0.8 °C/°C 0.017 °C
36
Parameter Nominal
value of
parameter
Uncertainty in
parameter
Sensitivity
coefficient
Resulting
uncertainty in
measured T
T2, temperature of
flow 2
20 °C 0.02 °C 0.17 °C/°C 0.003 °C
ρ1, water density in
flow 1
995.74 kg/m3 0.001 % (= 0.01
kg/m3)
0.0014 °C/
(kg/m3)
0.00001 °C
ρ2, water density in
flow 2
998.30 kg/m3 0.001 % (= 0.01
kg/m3)
-0.0014 °C/
(kg/m3)
0.00001 °C
Assuming constant
specific heat cp (*)
4.18 kJ/kg K 0.001 °C
Tmix, temperature after
mixing (using enthalpy
flow)
28.329 °C 0.017 °C
Tmix,mass, approximate
temperature after
mixing using mass
flow
28.330 °C 0.018 °C
Tmix,vol, approximate
temperature after
mixing using volume
flow
28.333 °C 0.021 °C
Table 17: Example uncertainty budget for reference (enthalpic) temperature at 2 barg with input flow temperatures 20 °C and 30 °C. The volume and mass flow weighted temperatures are shown for the purpose of comparison. In experiments some additional components may have to be taken into account. See also Table 16. It’s assumed that all uncertainty sources are independent and possess a normal distribution.
(*) The specific heat is not constant with changing temperature, but varies slightly with
temperature. The absolute value of the specific heat constant (or enthalpies) is not important for the
calculations, only its variation. Here, the assumption of constant cp is compared with using the best
known reference data values of h(T) in the temperature calculation. The difference is taken as very
small additional uncertainty which is linearly added to the final uncertainty. If one uses the
uncertainties stated for absolute enthalpy (1%) or of the (absolute) specific heat coefficient cp, one
will find higher values for the uncertainty (0.025 °C), but this neglects the fact that the uncertainty of
cp for temperatures close to each other is very correlated (‘the error is constant’).
37
Parameter Nominal value
of parameter
Uncertainty in
parameter
Sensitivity
coefficient
Resulting uncertainty
in measured T
Q1, flow 1 25 000 m3/h 0.5% (= 125 m
3/h) 0.0001 °C/ (
m3/h)
0.014 °C
Q2, flow 2 5 000 m3/h 0.5 % (= 25 m
3/h) -0.0006 °C/ (
m3/h)
0.014 °C
T1, temperature of flow 1 330 °C 0.02 °C 0.02 °C/°C 0.017 °C
T2, temperature of flow 2 310 °C 0.02 °C 0.003 °C/°C 0.003 °C
ρ1, water density in flow
1
650.0 kg/m3 0.1 % (= 0.7
kg/m3)
0.003 °C/ (kg/m3) 0.003 °C
ρ2, water density in flow
2
703.8 kg/m3 0.1 % (= 0.7
kg/m3)
-0.003 °C/
(kg/m3)
0.003 °C
h1, specific enthalpy in
flow 1 (**)
1518.5 kJ/kg 0.1 % (= 0.125
kJ/kg)
0.13 °C/ (kJ/kg) 0.19 °C
h2, specific enthalpy in
flow 2 (**)
1394.3 kJ/kg 0 % (uncertainty included in h1)
0.03 °C/ (kJ/kg) 0.0 °C
Induced error by
assuming constant
specific heat cp (**)
6.56 kJ/kg K 0.25 °C
Tmix, temperature after
mixing (using enthalpy
flow)
326.69 °C 0.19 °C
Tmix,mass, approximate
temperature after
mixing using mass flow
326.44 °C 0.44 °C
Tmix,vol, approximate
temperature after
mixing using volume
flow
326.67 °C 0.21 °C
Table 18: Example uncertainty budget for reference enthalpic temperature at 150 barg with input flow temperatures 310 °C and 330 °C. The volume and mass flow weighted temperatures are shown for the purpose of comparison. In experiments some additional components may have to be taken into account. See also Table 16.
(**) At high temperatures the assumption of constant specific heat capacity cp is not valid. It
would result in an error of 0.25 °C. Therefore the correct enthalpies should be used. In Table 18 the
full uncertainty of 0.1 % for cp has been used for the enthalpy 1, not taking into account correlations
between the error in cp at different temperatures, which would considerably reduce the total final
uncertainty.
38
5.2.6. Required equipment for speed of sound reference data
Two methods for getting the pressure-temperature-speed-of-sound relationship were described.
The first was the use of off-line laboratory reference data and the second was on line measurement
with the fluid being used. The equipment needed will now be described in brief.
5.2.6.1. Required equipment for off-line laboratory reference data
Reference data for the pressure/ temperature/ speed of sound relationship is available for a large
number of pure fluids and mixtures in various pressure and temperature ranges (e.g. the NIST
RefProp program, many alternatives exist). Thus in this case a computer suffices. If there is no data
available for the application under consideration, a facility can be build to measure speed of sound
accurately. However this requires mayor investments, considerable technical knowledge and is
beyond the scope of this document.
5.2.6.2. Required equipment for continuous on-line calibration
In this case the reference data for the pressure/ temperature/ speed of sound relationship is
established on-line. In this case a representative sampling point (valve) for the process liquid is
needed. This liquid has to be heated or cooled down to the desired reference temperature and may
be pressurized or depressurized if desired. When the liquid is of homogeneous temperature it can
flow through an ultrasonic meter for determination of the speed of sound. The liquid temperature
can be homogenized by using some pipe work or hoses going up and down in order to stimulate
mixing or by active mixing. Furthermore, the diameter of this sampling line will in general be smaller
than the process line, which enhances homogenization of the temperature. The relationship
between the values found by the reference device and the device measuring the temperature in the
process has to be established in an initial calibration effort. See Figure 8 for a schematic drawing of
an on-line calibrator set-up and the required parts.
The uncertainty components for the on-line calibrator are basically the same as for the calibration of
the ultrasound device (section 5.2.4.). The size of the uncertainty depends on the quality of the
components and the whole set-up.
5.3. Conclusion concerning required equipment The equipment needed for an ultrasound based temperature measurement method has been
described. In the ideal case a working calibrated device can be obtained from the manufacturer. It
has also been described which individual components are needed otherwise. Additional components
were presented for testing and validation of the device in case of heterogeneous fluid temperature
distribution. Finally a set-up for on-line calibration has been presented.
39
6. Overall theoretical uncertainty budget In Table 21 in appendix B all the relevant uncertainty components that have been discussed for an
ultrasound based temperature measurement method are presented. In Table 19 here below a
summary of the uncertainty budget for two cases is given only presenting the main components. In
principal each case will need its own uncertainty estimate (esp. different flow profiles).
Uncertainty
source
Explication
(Uncertainty due to:)
See table nr: Effect on final
temperature
uncertainty (20
°C/ 2 barg)
Effect on final
temperature
uncertainty
(300 °C/ 150
barg)
Calibration
uncertainty
The temperature reference
has some uncertainty
Table 16 0.02 °C 0.02 °C
Pressure
measurement
with 0.1 %
uncertainty
Conversion of speed of sound
to temperature depends on
pressure
Table 1 0.00 °C 0.02 °C
Transit time
measurement
Speed of sound is calculated
from transit time
measurement
Table 13 0.01 °C 0.00 °C
Pipe temperature
expansion
Path length increase caused
by the temperature expansion
of the pipe
Table 11 0.01 °C 0.05 °C
Composition 10 ppm-mol of NH3 added to
water changes speed of sound
Table 8 0.01 °C 0.00 °C
Temperature
profile vs. speed
of sound
averaging effect
The fact that averaging speed
of sound is not the same as
averaging temperature. Case
ΔT = 6 °C has been taken.
Table 4 0.14 °C 0.09 °C
Non-
representativity
effect
The fact that the ultrasound
beams don’t cover the entire
cross section.
Figure 4 0.15 °C 0.1 °C
Final uncertainty
in temperature
0.21 °C 0.15 °C
Table 19: Summary of uncertainty budget for measuring temperature with ultrasound.
For very homogeneous fluids, an uncertainty of a few 0.01 °C is theoretically possible at low
temperatures. At high temperatures the uncertainty is somewhat bigger. For heterogeneous fluids it
depends on the temperature gradient and profile. In general the uncertainty will be substantially
40
higher than for homogeneous fluids, namely in the order of 0.2 °C for a temperature difference of 6
°C (see also Figure 2 and Figure 3).
7. Experimental implementation of the different calibration methods Three different calibration methods were described in chapter 4. The needed equipment for the
ultrasound based temperature method and calibration was described in chapter 5. In this chapter
the experimental work considering the calibration is described. As far as possible with the limited
means and room for investments that were available, the three calibration methods were
implemented in an experimental setting. The experiments evaluating the measurement quality of
the ultrasound based temperature measurement method are described in a separate report, namely
D3.6.
7.0. Initial wet calibration All three calibration methods include a wet calibration of the device at homogeneous temperature.
This was done by mounting the ultrasound meter in a water flow loop with possibility of varying
temperature and pressure. The software of the transmitter was already quite advanced and gave a
temperature output. In the calibration run the path length setting of the three paths of the
ultrasound meter were adjusted such that the calculated temperature best matched the reference
temperature of the facility. The initial temperature and pressure coefficients of the manufacturer
were taken over. At high temperature also the conversion factors to convert speed of sound to
temperature were set.
7.1. Use of a real-time calibrator It was evaluated if the speed of sound of water significantly changed when a small amount of
different chemicals was added. This was not the case. The concept of a real-time calibrator was
partially tested by installing some hoses, an ultrasound sensor and a temperature array after each
other, see Figure 11. The hoses upstream of the ultrasound sensor create some distance to mix the
liquid to a homogenous temperature. The liquid could also be heated up to some desired
temperature. The ultrasound meter measures the speed of sound of the liquid. After the ultrasound
meter a Pt100-array with 9 Pt100 temperature sensors is installed to establish the reference
temperature and to check for temperature homogeneity. Pressure is also measured. In operation
the whole set-up is thermally insulated in order to stabilize the temperature as much as possible.
The measured composition-pressure-temperature-speed-of-sound relationship can now be used for
an ultrasound meter measuring temperature inside the main process as long as is guaranteed that
both ultrasound meters react in the same way on changing fluid composition, pressure and
temperature. The latter can be verified by an initial simultaneous calibration of both devices with
the same fluid, pressure and temperature. (It would also suffice that both devices measure speed of
sound in a traceable way, but this is a stronger requirement.)
41
Figure 11 Concept of a real-time calibrator: hoses to create some distance to mix (and possibly heat) the liquid, an ultrasound temperature meter and a reference temperature measurement, also checking for temperature homogeneity.
7.2. Use of external reference data In most of the experiments external reference data was used to convert measured speed of sound to
temperature. The initial wet calibration is to make sure that the device correctly measures the speed
of sound at different temperatures, pressures and possibly compositions. The reference data are
then used to convert speed of sound (in combination with pressure and composition) to a
temperature for cases not covered by the initial calibration. Preferably this is a temperature-
pressure-composition interpolation, although a limited extrapolation can be done as well. The
reference data were implemented in the transmitter of the ultrasound meter, and checked by an
off-line conversion of measured speed of sound to temperature as well. See deliverable D3.6 for
more details.
7.3. Creation of a traceable reference database A third way of being able to use the ultrasound based temperature meter is by creation of a
traceable reference database for the speed of sound oneself. Low quality reference data can be
made with the real-time calibrator set-up (see Figure 11) once the fluid composition is known. For
high quality traceable reference data a completely different set-up is needed. Inquiries were made
at some National Metrology Institutes which resulted in estimated costs of 100 to 200 k€. This was
far above the budget for investments for the current research project.
42
The analysis of the low quality reference data yielded that for small portions of added chemicals
(boric acid, ammonia) to the water, no measurable change in speed of sound was noticed. The used
small portions of additives are representative for the water composition in power plants.
For the reasons mentioned above it was decided not to create new traceable reference data, but to
gather the existing traceable speed of sound reference data and to present them on a website
(deliverable D5.5).
7.4. Conclusion concerning experiments with calibration methods The main method used in the experiments was the initial wet calibration with use of external
reference data. The concept of a real-time calibrator was partially implemented and tested. The
creation of new traceable reference data was not pursued as existing data already fitted the needs.
See deliverable D3.6 for more details regarding the experimental tests.
8. Overall conclusion In this report a general and practical frame work for temperature measurement using ultrasound for
flow in pipe lines has been presented. The necessary equipment has been described and different
calibration procedures have been proposed. Estimated uncertainties have been given. In the reports
of deliverables D3.5 and D3.6 the practical realization of these ideas will be presented together with
a performance evaluation.
For very homogeneous fluids, an uncertainty of a few 0.01 °C is theoretically possible at low
temperatures. At high temperatures the uncertainty is somewhat bigger. For heterogeneous fluids it
depends on the temperature gradient and profile. In general the uncertainty will be substantially
higher than for homogeneous fluids, namely in the order of 0.1 °C or more. Figure 2 and Figure 3
give an indication of the reachable uncertainty for extreme temperature profiles.
For (nuclear) power plants the ultrasound based temperature measurement method is very
promising from a theoretical point of view, as it may reduce the current uncertainty substantially.
References [1] NIST Reference Fluid Thermodynamic and Transport Properties Database (REFPROP), see
http://www.nist.gov/srd/nist23.cfmhttp://www.nist.gov/srd/nist23.cfm.
[2] NIST RefProp 8.0 help file on water.
[3] Revised Release on the IAPWS Formulation 1995 for the Thermodynamic Properties of Ordinary Water Substance for General and Scientific Use, 2009, International Association for the Properties of Water and Steam.
[4] A.B. Coppens, Simple equations for the speed of sound in Neptunian waters (1981) J. Acoust. Soc. Am. 69(3), pp 862-863, see also http://resource.npl.co.uk/acoustics/techguides/soundseawater/content.html#COP
43
Appendix A: Measured temperature per path In Table 20 the numerical values of the ultrasound based temperature measurement are shown per
path. The temperature at the line y = 0 for profiles 9 to 14 has been assumed equal to (Tup + Tdown)/2.
This has the consequence that the configuration with three horizontal paths gives the bests results.
This seems to be rather due to the studied symmetrical temperature profile than that it is true in
general. In deliverable D5.4 this will be studied in more detail using a software tool.
Temperature profile True volumetric
mean
temperature of
section, Tsec,vol
[°C]
Ultrasound
temperature,
one vertical
path, Tus,v1
[°C]
Ultrasound
temperature,
three center
paths, Tus,c3 [°C]
Ultrasound
temperature, 3
horizontal
paths, Tus,h3 [°C]
Ultrasound
temperature, 3
vertical paths,
Tus,v3 [°C]
Homogeneous, T = 25 °C 25.000 25.000 25.000
(m: 25.000,
b: 25.000,
t: 25.000)
25.000 - (m:
25.000, b:
25.000, t:
25.000)
25.000 - (m:
25.000, b:
25.000, t:
25.000)
Homogeneous, T = 300
°C
300.000 300.000 300.000 - (m:
300.000, b:
300.000, t:
300.000)
300.000 - (m:
300.000, b:
300.000, t:
300.000)
300.000 - (m:
300.000, b:
300.000, t:
300.000)
Boundary layer 1%, Tin =
30 °C, Tout = 20 °C
29.604 29.765 29.765 - (m:
29.765, b:
29.765, t:
29.765)
29.727 - (m:
29.765, b:
29.708, t:
29.708)
29.727 - (m:
29.765, b:
29.708, t:
29.708)
Boundary layer 0.1%, Tin
= 300 °C, Tout = 250 °C
299.800 299.933 299.933 - (m:
299.933, b:
299.933, t:
299.933)
299.922 - (m:
299.933, b:
299.917, t:
299.917)
299.922 - (m:
299.933, b:
299.917, t:
299.917)
Two homogeneous
circular regions, Tin = 25
°C, Tout = 24 °C
24.250 24.496 24.496 - (m:
24.496, b:
24.496, t:
24.496)
24.340 - (m:
24.496, b:
24.261, t:
24.261)
24.340 - (m:
24.496, b:
24.261, t:
24.261)
Two homogeneous
circular regions, Tin = 301
°C, Tout = 300 °C
300.250 300.503 300.503 - (m:
300.503, b:
300.503, t:
300.503)
300.345 - (m:
300.503, b:
300.266, t:
300.266)
300.345 - (m:
300.503, b:
300.266, t:
300.266)
Two homogeneous
circular regions, Tin = 28
°C, Tout = 22 °C
23.500 24.856 24.856 - (m:
24.856, b:
24.856, t:
24.856)
23.938 - (m:
24.856, b:
23.479, t:
23.479)
23.938 - (m:
24.856, b:
23.479, t:
23.479)
Two homogeneous
circular regions, Tin = 303
°C, Tout = 297 °C
298.500 300.093 300.093 - (m:
300.093, b:
300.093, t:
300.093)
299.138 - (m:
300.093, b:
298.66, t:
298.66)
299.138 - (m:
300.093, b:
298.66, t:
298.66)
Two equal homogeneous
sections, Tup = 25 °C,
Tdown = 24 °C
24.500 24.496 24.496 - (m:
24.496, b:
24.496, t:
24.496)
24.500 - (m:
24.500, b:
24.000, t:
25.000)
24.496 - (m:
24.496, b:
24.496, t:
24.496)
44
Temperature profile True volumetric
mean
temperature of
section, Tsec,vol
[°C]
Ultrasound
temperature,
one vertical
path, Tus,v1
[°C]
Ultrasound
temperature,
three center
paths, Tus,c3 [°C]
Ultrasound
temperature, 3
horizontal
paths, Tus,h3 [°C]
Ultrasound
temperature, 3
vertical paths,
Tus,v3 [°C]
Two equal homogeneous
sections, Tup = 301 °C,
Tdown = 300 °C
300.500 300.503 300.503 - (m:
300.503, b:
300.503, t:
300.503)
300.5 - (m:
300.500, b:
300.000, t:
301.000)
300.503 - (m:
300.503, b:
300.503, t:
300.503)
Two equal homogeneous
sections, Tup = 28 °C,
Tdown = 22 °C
25.000 24.856 24.856 - (m:
24.856, b:
24.856, t:
24.856)
25.000 - (m:
25.000, b:
22.000, t:
28.000)
24.856 - (m:
24.856, b:
24.856, t:
24.856)
Two equal homogeneous
sections, Tup = 303 °C,
Tdown = 297 °C
300.000 300.093 300.093 - (m:
300.093, b:
300.093, t:
300.093)
300.000 - (m:
300.000, b:
297.000, t:
303.000)
300.093 - (m:
300.093, b:
300.093, t:
300.093)
Two equal homogeneous
sections, Tup = 30 °C,
Tdown = 5 °C
17.500 15.161 15.161 - (m:
15.161, b:
15.161, t:
15.161)
17.5 - (m: 17.5,
b: 5, t: 30)
15.161 - (m:
15.161, b:
15.161, t:
15.161)
Two equal homogeneous
sections, Tup = 330 °C,
Tdown = 300 °C
315.000 318.149 318.149 - (m:
318.149, b:
318.149, t:
318.149)
315.000 - (m:
315.000, b:
300.000, t:
330.000)
318.149 - (m:
318.149, b:
318.149, t:
318.149)
Table 20: Error in ultrasound temperature due to non-representativity of paths. In the cell the mean temperature of the paths is shown, together with the individual paths (m = middle path, b = bottom, diagonal or off-center vertical path, t = top, diagonal or off center vertical path). In case of two homogeneous regions, the horizontal path at 50% is assumed to contain fluid of the average temperature. (Note: part of the temperature error is due to the ultrasonic averaging effect described in Table 4.)
Appendix B: Full uncertainty budget In this appendix an extensive uncertainty budget of the ultrasound temperature method is shown,
see Table 21. The numerical values are estimates and may need adjustment for a particular case.
Furthermore in experimental work the estimates may prove to be too conservative or too optimistic.
For a homogeneous temperature measurement at ambient temperature the uncertainty is
estimated to be 0.03 °C in this particular case, whereas in the case of an irregular inhomogeneous
temperature distribution with temperature differences up to 6 °C, the uncertainty is estimated to be
0.24 °C (both estimates at coverage factor k = 2). The measurement equations are as follows
(symbols are explained in Table 21):
45
t = tm – tdel
L = L0 + δLT + δLP
δLT = L·α·(Tw – T0)
δLP = β·(P – P0)
β = D/(2·b·E)
c = L/t
Tus = Tc-P + δTcal + δTcomp + δTc + δTP + δTcorr(c,P) + δTnon-lin + δTnon-repr
Transit time uncertainty
Variable xi Description Unit Value Expand
ed
uncerta
inty
U(xi)
Expand
ed
relative
uncerta
inty
U*(xi)
Probabi
lity
distribu
tion
Divis
or
Sensiti
vity
coeffici
ent ci
Standar
d
uncerta
inty
u(xi)
Standar
d
relative
uncerta
inty
u*(xi)
Expand
ed
uncerta
inty
U(xi)
Relative
expand
ed
uncerta
inty
U*(xi)
%
tm Average measured transit time s 6.9E-
05
1.0E-09 0.00% normal 2.00 1.00 5.0E-10
tdel Delay of electronics in US sensor s 2.0E-
06
5.0E-10 normal 2.00 -1.00 2.5E-10
t Average corrected transit time s 6.7E-
05
normal 2.00 5.6E-10 0.001% 1.1E-09 0.0017
%
Path length uncertainty
Variable xi Description Unit Value Expand
ed
uncerta
inty
U(xi)
Expand
ed
relative
uncerta
inty
U*(xi)
Probabi
lity
distribu
tion
Divis
or
Sensiti
vity
coeffici
ent ci
Standar
d
uncerta
inty
u(xi)
Standar
d
relative
uncerta
inty
u*(xi)
Expand
ed
uncerta
inty
U(xi)
Relative
expand
ed
uncerta
inty
U*(xi)
%
L0 Path length at T0 and P0 m 0.100 1.0E-06 0.00% rectang
ular
1.73 1.00 5.8E-07
46
δLT Temperature correction of L m 4.8E-
06
1.6E-06 34.08% rectang
ular
1.73 1.00 9.5E-07
- T0 Reference temperature °C 20 0.1 0.50%
- Tw Average wall temperature °C 23 1 4.35%
- α Linear thermal expansion coefficient 1/°C 1.6E-
05
1.0E-06 6.25%
δLP Pressure correction of L m 1.5E-
06
7.3E-08 5.00% rectang
ular
1.73 1.00 4.2E-08
- P0 Reference pressure barg 0 0.001
- P Actual process pressure barg 5 0.01 0.20%
- β Linear pressure expansion coefficient
for thin walled pipes
1/bar
g
2.9E-
06
1.5E-07 5.00%
- D Pipe diameter m 0.1 n.a.
- b Pipe wall thickness (thin wall) m 0.009 n.a.
- E Pipe Young modulus Pa 2.0E+1
1
n.a.
L Path length at T and P m 0.100 normal 2.00 1.1E-06 0.001% 2.2E-06 0.0022
%
Speed of sound uncertainty
Variable xi Description Unit Value Expand
ed
uncerta
inty
U(xi)
Expand
ed
relative
uncerta
inty
U*(xi)
Probabi
lity
distribu
tion
Divis
or
Sensiti
vity
coeffici
ent ci
Standar
d
uncerta
inty
u(xi)
Standar
d
relative
uncerta
inty
u*(xi)
Expand
ed
uncerta
inty
U(xi)
Relative
expand
ed
uncerta
inty
U*(xi)
%
L Path length (from above) m 0.100 2.2E-06 0.0022
%
normal 2.00 1.5E+0
4
1.1E-06
t Corrected transit time (from above) s 6.7E-
05
1.1E-09 0.0017
%
normal 2.00 -
2.2E+0
7
5.6E-10
c Speed of sound m/s 1492.
631
normal 2.00 0.021 0.001% 0.041 0.003%
47
Speed of sound based average cross section
temperature uncertainty
Variable xi Description Unit Value Expand
ed
uncerta
inty
U(xi)
Expand
ed
relative
uncerta
inty
U*(xi)
Probabi
lity
distribu
tion
Divis
or
Sensiti
vity
coeffici
ent ci
Standar
d
uncerta
inty
u(xi)
Standar
d
relative
uncerta
inty
u*(xi)
Expand
ed
uncerta
inty
U(xi)
Relative
expand
ed
uncerta
inty
U*(xi)
%
Tc-P =
f(c,P)
Temperature derived from c and P °C 23.21 0.027 0.12% rectang
ular
1.73 1 0.015
- c Speed of sound (from above) m/s 1492.
631
0.041
- P Process pressure (from above) barg 5 0.01
δTcal Corr./Unc. of calibration reference
temperature
°C 0 0.02
δTcomp Corr./Unc. due to impure water
composition
°C 0 0.01
δTc Corr./Unc. due to uncertainty in c °C 0 0.015
- df/dc Sensitivity coefficient df/dc °C/(m
/s)
0.355 n.a.
δTP Corr./Unc. due to uncertainty in P °C 0 -0.0006
- df/dP Sensitivity coefficient df/dP °C/ba
r
-0.06 n.a.
δTcorr(c,P) Correlation between c and P due to
pipe pressure expansion
°C negligi
ble for
final
result
n.a.
δTnon-lin Corr./Unc. of T due to averaging of 1/c
instead of averaging T inclusive non-
linearity of T=f(c,P) for dT = 6 °C in
pipe
°C 0 0.14 rectang
ular
1.73 1.00 0.08
δTnon-repr Corr./Unc. due to non-
representativeness of the
measurement paths due to
temperature profile
°C 0 0.15 rectang
ular
1.73 1.00 0.09
Tus Corrected cross section average
temperature
°C 23.21 normal 2.00 0.12 0.5% 0.24 1.0%
Table 21: Theoretical uncertainty budget for an ultrasound based temperature measurement method.