volatile organic compounds in energy drinks as …...cynthia elmore and j. garrett slaton oi...
TRANSCRIPT
Cynthia Elmore and J. Garrett SlatonOI Analytical, P.O. Box 9010, College Station, TX 77845-9010 • (800) 653-1711 • www.oico.com
Volatile Organic Compounds in Energy Drinks asDetermined by GC-MS with Purge and Trap Sample Concentration
Introduction
Experimental
The use of purge and trap for foaming liquid samples has been demonstrated for a range of somewhat difficult matrices. The samples were purged in the vials, rather than in a fritted sparge vessel and usable results for samples typically analyzed by headspace were obtained.
Trace levels of chloroform, attributed to drinking water disinfectant byproducts, were found in several samples. Benzene was found in a few of the drinks containing benzoate preservatives.
The presence of other compounds, such as p-cymene, limonene, and butanoate esters is also to be expected based on their use as flavorings. Ethanol is detected in nearly all of the energy drinks to an appreciable level, but its source cannot be determined from the data.
Summary and Conclusions
References1. “The American Energy Drink Craze in Two Caffeinated Charts”, Roberto A.
Ferdman, Quartz, March 26, 2014.2. “Energy Drinks: Busting Your Health for the Buzz”, Pennsylvania Medical
Society, September 19, 2008.3. USEPA Method 8260B Volatile Organic Compounds by Gas
Chromatography/Mass Spectrometry (GC/MS) Revision 2 December 1996.4. ICBA Guidance Document to Mitigate the Potential for Benzene Formation
in Beverages, International Council of Beverages Associations, 2006.5. Data on Benzene in Soft Drinks and Other Beverages, U.S. FDA, 2007.
Chloroform and benzene are compounds on the NPDWRs list. Chloroform was found in several of the samples at reportable levels (>2 ppb), still well below the United States Environmental Protection Agency (USEPA) maximum contaminant level (MCL). Chloroform is known to develop in drinking water as a disinfectant byproduct. Flavoring agent p-cymene, used as part of the calibration but not on the list, was also detected above the reporting limit.
Benzene is carcinogenic at high concentrations and is monitored in drinking water analysis. It may form when benzoate salts and ascorbic acid are present. Various factors foster the formation, such as exposure to heat and light, while sugar and EDTA can inhibit it4. In 2005, the United States Food and Drug Administration (FDA) received private laboratory results showing low levels of benzene in some soft drinks that contained benzoate salts and ascorbic acid. Because the FDA has no standard for benzene in beverages other than bottled water, the NPDWR limit of 5 ppb was adopted as a standard5. Benzene was detected in two evaluated drinks at trace levels (2.17 and 1.85 ppb). The drinks contained benzoate salts, but no ascorbic acid.
All samples contained varying amounts of other compounds (TICs) identified as natural and artificial flavorings frequently found in fruit-flavored beverages. Most drinks also contained appreciable (>200 ppb) ethanol, perhaps left over as an aid to dissolve essential oils or from decomposition of ethyl esters.
Results and DiscussionEnergy drinks and shots are approximately a $30 billion dollar industry with major beverage companies, such as PepsiCo, and focused energy drink manufacturers, like Red Bull GmbH, participating1. A 2014 World Health Organization report estimated 30% of adults, 68% of adolescents, and 18% of children under the age of 10 consume energy drinks2.
Prior to public outcry in the wake of the discovery of benzene in soft drinks in the early 1990s, prepared beverages were not regulated. They are still not broadly held to the National Primary Drinking Water Regulations, though the benzene content is now regulated to 5 ppb. Beverages can be foamy when analyzed by purge and trap methodology, and thus, are traditionally analyzed by headspace coupled to a GC-MS. We have analyzed a selection of these challenging samples using GC-MS with a new purge and trap sample concentrator and present both quantitative and some qualitative data on identified VOCs both on and off the National Primary Drinking Water Regulations (NPDWRs) list with good results.
Figure 1. 4760 Purge & Trap with 4100 Sample Processor
Figure 2. Energy drinks evaluated in the study
Figure 3. 50 ppb Calibration Standard
Figure 4. Energy Drink 1
Eth
ano
l 50
0 p
pb
SS1
& 1
S1SS
2 IS2 SS
3B
uta
no
ate
este
r 1
60
pp
bB
uta
no
ate
este
r 1
80
pp
bB
uta
no
ate
este
r 9
00
pp
b
IS4
IS3 O
rgan
ic a
cid
65
pp
bIS
4
Bu
tan
oat
e es
ter
23
0 p
pb
Bu
tan
oat
e es
ter
62
0 p
pb
Figure 5. Energy Drink 2
Eth
ano
l 12
60
pp
b
SS1
& 1
S1SS
2IS
2
SS3
Bu
tan
oat
e es
ter
12
50
0 p
pb
3-m
eth
yl-1
1-B
uta
no
l 36
90
pp
b
SS4
Hex
ano
ate
eth
yl e
ster
39
0 p
pb
Bu
tan
oat
e es
ter
14
20
0 p
pb
Bu
tan
oat
e es
ter
26
40
pp
b
Hex
ano
ate
este
r 2
96
0 p
pb
Pen
tan
oat
e es
ter
10
30
pp
b
Figure 6. Energy Drink 3
Eth
ano
l 24
0 p
pb
SS1
& 1
S1SS
2IS
2
SS3 .b
eta.
– P
hel
lan
dre
ne
40
0 p
pb
SS4
IS3
Ben
zald
ehyd
e 1
40
pp
b
Unknown alcohol 170 ppb
Lin
no
nen
e 2
67
00
pp
bg
amm
a Te
rpin
ene
21
40
pp
bTe
rpin
ole
ne
80
0 p
pb
Table 1. Method Parameters
Purge-and-Trap Eclipse 4760 P&T Sample Concentrator
Trap #10 trap; Tenax® / Silica gel / CMS
Purge Gas Zero grade Helium at 40 mL/min
Purge Time 11 min
Sparge Mount Temp. 45 °C soil
Sample Temp. 45 °C soil
Desorb Time 0.5 min
Bake Time 5 min
OI #10 Trap Temp. Ambient during purge180 °C during desorb pre-heat190 °C during desorb210 °C during bake
Water Management 120 °C during purgeAmbient during desorb240 °C during bake
Transfer Line Temp. 140 °C
Six-port Valve Temp. 140 °C
Gas Chromatograph Agilent 7890A
Column Restek Rxi-624 Sil MS 30 meter, 0.25 mm, 1.4 µm film
Carrier Gas Zero grade helium
Inlet Temp. 250 °C
Inlet Liner Agilent Ultra Inert 1 mm straight, taper
Column Flow Rate 0.8 mL/min
Split Ratio 150:1
Oven Program Hold at 40 °C for 1.5 min16 °C/minute to 180 °C 40 °C/minute to 220 °CHold at 220 °C for 2.0 minTotal GC Run is 13.25 min
Mass Spectrometer Agilent 5975C
Mode Scan 35 - 300 amu
Scans/Second 5.19
Solvent Delay 1.40 min
Transfer Line Temp. 250 °C
Source Temp. 300 °C
Quadrupole Temp. 200 °C
Draw-out Plate 6 mm
Table 2. Calibration DataAnalyte Compound AVG % RSD1 pentafluorobenzene (IS)
2 dichlorodifluoromethane 0.154 7.19
3 chloromethane 0.197 6.43
4 vinyl chloride* 0.251 5.40
5 bromomethane 0.248 9.68
6 chloroethane 0.145 8.16
7 trichlorofluoromethane 0.515 5.21
8 ethyl ether 0.158 4.22
9 1,1-dichloroethene* 0.398 6.06
10 carbon disulfide 1.121 5.24
11 1,1,2-trichloro-1,2,2-trifluoroethane 0.409 5.61
12 methyl iodide 0.826 5.22
13 allyl chloride 0.187 2.10
14 methylene chloride* 0.420 7.19
15 acetone 0.026 9.93
16 trans-1,2-dichloroethene* 0.460 9.31
17 methyl tert-butyl ether 0.784 3.34
18 chloroprene 0.666 1.93
19 1,1-dichloroethane 0.743 3.41
20 acrylonitrile 0.115 7.14
21 cis-1,2-dichloroethene* 0.510 3.00
22 2,2-dichloropropane 0.329 6.24
23 bromochloromethane 0.250 2.78
24 chloroform* 0.847 2.78
25 methyl acrylate 0.300 4.67
26 carbon tetrachloride* 0.576 2.38
27 dibromofluoromethane (SS) 0.528 1.82
28 1,1,1-trichloroethane* 0.623 3.13
29 2-butanone 0.029 5.01
30 1,1-dichloropropene 0.570 3.99
31 1,4-difluorobenzene (IS)
32 benzene* 1.045 8.69
33 methacrylonitrile 0.113 4.17
34 1,2-dichloroethane-d4 (SS) 0.043 1.57
35 1,2-dichloroethane* 0.374 6.83
36 trichloroethene* 0.310 3.74
37 dibromomethane 0.195 2.68
38 bromodichloromethane* 0.358 3.82
39 1,2-dichloropropane* 0.263 2.33
40 methyl methacrylate 0.118 6.88
41 2-chloroethyl vinyl ether 0.073 13.29
42 cis-1,3-dichloropropene 0.350 8.09
43 chlorobenzene-d5 (IS)
Analyte Compound AVG % RSD44 toluene-d8(ss) 1.301 1.72
45 toluene* 0.790 3.81
46 4-methyl-2-pentanone 0.021 4.12
47 tetrachloroethene* 0.329 5.37
48 trans-1,3-dichloropropene 0.346 9.58
49 ethyl methacrylate 0.262 6.78
50 1,1,2-trichloroethane* 0.237 3.12
51 chlorodibromomethane* 0.313 9.24
52 1,3-dichloropropane 0.335 2.28
53 1,2-dibromoethane* 0.305 3.12
54 2-hexanone 0.154 5.73
55 chlorobenzene* 0.945 3.95
56 ethylbenzene* 1.496 4.11
57 1,1,1,2-tetrachloroethane 0.276 1.99
58 m,p-xylenes* 0.588 4.86
59 o-xylene* 0.575 4.39
60 styrene* 0.956 3.97
61 bromoform* 0.181 9.29
62 isopropylbenzene 1.446 4.59
63 1,4-dichlorobenzene-d4 (IS)
64 4-bromofluorobenzene (SS) 1.088 1.90
65 bromobenzene 0.880 4.57
66 n-propylbenzene 3.822 2.51
67 1,1,2,2-tertrachloroethane 0.687 4.20
68 2-chlorotoluene 2.173 3.12
69 1,3,5-trimethylbenzene 2.669 1.84
70 1,2,3-trichloropropane 0.656 6.60
71 4-chlorotoluene 2.545 3.50
72 tert-butylbenzene 2.520 1.84
73 pentachloroethane 0.382 6.77
74 1,2,4-trimethylbenzene 2.682 2.11
75 sec-butylbenzene 3.346 1.81
76 p-isopropytoluene 2.635 2.95
77 1,3-dichlorobenzene 1.478 4.77
78 1,4-dichlorobenzene* 1.449 4.65
79 n-butylbenzene 2.536 2.84
80 1,2-dichlorobenzene* 1.339 4.00
81 1,2-dibromo-3-chloropropane* 0.090 12.86
82 hexachlorobutadiene 0.574 2.68
83 1,2,4-trichlorobenzene* 0.943 4.35
84 naphthalene 1.868 9.34
85 1,2,3-trichlorobenzene 0.858 5.72
* National Primary Drinking Water Regulated Compounds (NPDWR)
Sample concentration and introduction was done using an OI Analytical 4760 Purge and Trap and a 4100 Sample Processor, while an Agilent 5975C Mass Spectrometer and 7890A Gas Chromatograph were used for separation and detection. See Figure 1 for a photograph of the equipment and Table 1 for the method parameters. Twenty samples were analyzed from top selling brands in the U.S. (Figure 2) Due to the potential for excessive foaming, samples were run in the “soil mode” of the 4100 where the sample is purged in a 40-mL vial rather than in a frit sparger.
A multi-point calibration was run, which included the regulated compounds from the NPDWRs list with concentrations ranging from 2 ppb to 200 ppb. Calibrations were generated using USEPA Method 82603. (Table 2) All method criteria were met. Samples were quantitated using this calibration and then a library search was performed on peaks not identified by the calibration method, i.e., tentatively identified compounds (TICs).
VOCs-in-Energy-Drinks-by-GC-MS-purge-and-trap_r3.indd 1 2/25/16 5:22 PM