vm-702b shinkawa
TRANSCRIPT
-
7/29/2019 VM-702B shinkawa
1/78
MANUAL No. 6G10-024
Published in Sep. 2010
Revised in Oct. 2010
VM-7 Series MonitorVM-702BAbsolute Vibration Monitor ModuleManual
zThis manual describes the VM-7B-type racks and modules.zThis manual is solely designed for technicians to set, maintain, check for the
VM-7 Series Monitor and deal with trouble on the monitoring system.Use of this manual is restricted to those who are qualified or authorized to
perform the service.z Read this manual thoroughly and understand the contents before setting,
maintaining, checking and trouble-shooting for VM-7 Series Monitor.z Keep this manual so that the operator can refer to it when needed.
SHINKAWA Sensor Technology, Inc.
-
7/29/2019 VM-702B shinkawa
2/78
-
7/29/2019 VM-702B shinkawa
3/78
VM-702B Preface
1
Thank you very much for purchasing the VM-7 Series Monitor.
The VM-7 Series Monitor (hereafter "the unit") comes with ten manuals which contain information for differentoperations as well as tasks.
This manual is one of the ten.
Name VM-7B System Manual
Intended users
of the manual
Personnel who maintain the unit, install it to the control panel and connect the wires
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Description x Description of the unit
x Specifications and outlinex Description of unpacking and accessories
x How to install the unit into the control panel
x How to connect the unit
x Regular inspection
x How to set the unit
x Troubleshooting
x Names and functions of the individual parts
Name VM-701B Vibration/Displacement Monitor Module Manual
Intended users
of the manual
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Description x How to set the unit
x How to adjust the unit
x Troubleshooting
Name VM-702B Absolute Vibration Monitor Module Manual
Intended users
of the manual
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Description x How to set the unitx How to adjust the unit
x Troubleshooting
Name VM-703B Tachometer & Eccentricity Monitor Module Manual
Intended users
of the manual
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Description x How to set the unit
x How to adjust the unit
x
Troubleshooting
-
7/29/2019 VM-702B shinkawa
4/78
VM-702B Preface
2
Name VM-704B Temperature Monitor Module Manual
Intended users
of the manual
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Description x How to set the unitx How to adjust the unit
x Troubleshooting
Name VM-706B Rod Drop Monitor Module Manual
Intended users
of the manual
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Description x How to set the unit
x How to adjust the unit
x
Troubleshooting
Name VM-721B 18-Channel Relay Module Manual
Intended users
of the manual
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Description x How to set the unit
x Troubleshooting
Name VM-741B Local Communication & Phase Marker Module Manual
Intended users
of the manual
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Description x How to set the unit
x Description of communication specifications
x Troubleshooting
Name VM-742B Network Communication Module Manual
Intended users
of the manual
Personnel who perform the detail settings and maintenance of the unit
The use of this manual is restricted to those who have completed the maintenance trainingcourse run by our company.
Description x How to set the unit
x Description of communication specifications
x Troubleshooting
Name VM-771B MCL View Operation Manual
Intended users
of the manual
Operators who are in charge of the maintenance of the unit and who confirm the screen in
their work
Description x Description of the screen
x How to operate the unit
-
7/29/2019 VM-702B shinkawa
5/78
VM-702B Preface
3
Be sure to read this manual thoroughly and understand the contents before using this unit.Keep this
manual so that the operator can refer to it when needed.
If you have any inquiry about this unit or manual, contact the sales office where you purchased the unit.
For the use of the unit, adequate knowledge about other instruments connected to the unit and transducers
is also required.In addition, not only the contents of this manual, but common sense also needs to be
applied for the safe use of the unit.
This unit has had its EMC (Electro-magnetic compatibility) evaluated by various noise tests.However,
since the performance of the unit varies depending on the environmental conditions such as below, follow
instructions in this manual to stabilize the system.
Intensity of noise
Temperature and humidity
Layout of the unit
When reselling the unit as an assembly or transferring the unit (to a third party), please provide the end
users with this manual.
When discarding the unit, follow local government regulations.
Company and product names contained herein are trademarks or registered trademarks of their respective
owners.
Copyright is retained by Shinkawa Sensor Technology, Inc.No part of this document may be disclosed or
reproduced in any form without prior consent of SHINKAWA Sensor Technology, Inc.
The appearance of the Device Config window differs by unit specifications.
-
7/29/2019 VM-702B shinkawa
6/78
VM-702B Preface
4
The company assumes no liability for any damage or injury caused by the user's fault including not
following the instructions of this manual or not installing or handling the unit in an appropriate manner.
The company shall not be liable for any direct or indirect damage, injury, or lost income resulting from
misuse, abuse, modification to or of any part of, or disassembling of the unit by the user.
Unless expressly stated herein, the company makes no other warranties, express or implied, including
without limitation, any warranty of merchantability, fitness for a particular purpose, or infringement of
intellectual property rights.
A Warranty Provision is included at the end of this manual.In the event that the unit failure occurs, the
company shall repair the unit in accordance with the Warranty Provision.
The company shall not be liable for any incidents resulting from the use of parts other than those of which
are recommended by the company.
-
7/29/2019 VM-702B shinkawa
7/78
VM-702B Table of Contents
5
Regarding the Manuals for the VM-7 Series Monitor.............................. 1
Important Notices ............................................................ 3
Disclaimer .................................................................. 4
Table of Contents ............................................................ 5
Chapter 1 For Safe Use of the Unit........................................................71.1 For Users of This Manual................................................................................................................71.2 Warning Symbols .............................................................................................................................71.3 Precautions for Handling.................................................................................................................8
1.3.1 Prohibited Uses..........................................................................................................................81.3.2 Compliances...............................................................................................................................9
Chapter 2 Overview of Absolute Vibration Monitor Module..............112.1 Functional Description ...................................................................................................................11
2.1.1 Measurement Function ...........................................................................................................112.1.2 Display Function.......................................................................................................................122.1.3 Alarm Function .........................................................................................................................122.1.4 Output Function........................................................................................................................132.1.5 Control Function.......................................................................................................................14
2.2 Names and Functions of the Individual Parts.............................................................................15
Chapter 3
Installation/Uninstallation of a Module..............................17
3.1 Module Installation Location .........................................................................................................183.2 Installing the Module ......................................................................................................................193.3 Uninstalling the Module .................................................................................................................20
Chapter 4 Module Setting.....................................................................214.1 Workflow ..........................................................................................................................................214.2 Setting the Internal Switch ............................................................................................................22
4.2.1 Recorder Output Selection Switch ........................................................................................224.3 Loading/Writing the Setting...........................................................................................................234.3.1 Loading the Setting..................................................................................................................23
4.3.2 Writing the Setting....................................................................................................................234.4 Editing the Channel Configuration...............................................................................................244.5 Common Module Settings.............................................................................................................264.6 Absolute Vibration Channel Setting.............................................................................................28
4.6.1 Basic Settings (Absolute Vibration Monitor) ........................................................................284.6.2 Analysis Settings......................................................................................................................334.6.3 Alarm Setting............................................................................................................................36
4.7 Relay Settings.................................................................................................................................384.7.1 Operation Settings ...................................................................................................................384.7.2
Configuring a Standard Relay Logic (Relay Logic1)...........................................................40
4.7.3 Configuring a Multistage Relay Logic (Relay Logic2) ........................................................42
-
7/29/2019 VM-702B shinkawa
8/78
VM-702B Table of Contents
6
Chapter 5 Adjustment of the Module ..................................................485.1 Connecting the Equipment (Connecting the Input Terminals) .................................................485.2 Connecting the Equipment (Connecting the Output Terminals)..............................................505.3 Adjusting the Gap (Bias) Voltage Measurement Value.............................................................515.4 Adjusting the Measurement Values .............................................................................................52
5.4.1 Absolute Vibration Monitor (REL: Relative Shaft Vibration) ..............................................535.4.2 Absolute Vibration Monitor (SEIS: Seismic Vibration)........................................................555.4.3 Absolute Vibration Monitor (ABS: Absolute Shaft Vibration) .............................................58
5.5 Adjusting the Recorder Output.....................................................................................................605.5.1 Absolute Vibration Monitor (REL: Relative Shaft Vibration) ..............................................605.5.2 Absolute Vibration Monitor (SEIS: Seismic Vibration)........................................................625.5.3 Absolute Vibration Monitor (ABS: Absolute Shaft Vibration) .............................................64
5.6 Adjusting the Voltage Measurement Values of the Analysis Board ........................................66
Chapter 6 Module Status Confirmation with MCL View.....................67
Chapter 7 Maintenance and Service ...................................................687.1 Exchange for Module.....................................................................................................................68
7.1.2 Setting the Module...................................................................................................................697.2 Service Life of the Unit (Replacement parts) .............................................................................697.3 Storage of the Module....................................................................................................................69
Chapter 8 Trouble Shoot ing .................................................................708.1 Troubleshooting..............................................................................................................................70
Warranty Provision
-
7/29/2019 VM-702B shinkawa
9/78
VM-702B Chapter 1For Safe Use of the Unit
7
Warning
Chapter 1 For Safe Use of the Unit
For the safe use of the unit, be sure to read the instructions below.
1.1 For Users of This ManualThis manual is solely designed for technicians to set, maintain, check for the VM-7 Series Monitor and deal
with trouble on the monitoring system.
Use of this manual is restricted to those who are qualified or authorized to perform the service.
1.2 Warning SymbolsThis manual uses the following symbols to increase awareness of possible dangers or harmful situations to the
user, and to guide in the safe operation of the unit.Be sure that you fully understand the contents of this manual
and follow the instructions herein.
Warning Symbols Description
This symbol is used to warn against the possibility of serious or fatal injury.
Follow the instructions to ensure safety.
This symbol is used to caution against the possibility of injury.
Follow the instructions to ensure safety.
This symbol is used to draw attention to the possibility of property damage.Follow the instructions to prevent damage to equipment.
Point In this manual, Point is used to highlight some important issues as well as to provide useful
information.
Caution
Caution
-
7/29/2019 VM-702B shinkawa
10/78
VM-702B Chapter 1For Safe Use of the Unit
8
Warning
1.3 Precautions for HandlingFor the safe use of the unit, be sure to follow the precautions below.
1.3.1 Prohibited Uses
Never disassemble or modify the unit.Do not use the unit under conditions that exceed itsspecifications.Failure to do so may result in malfunction of the unit, fire or injury.
Do not use radio equipment as a transceiver, or mobile phones around the unit.
If electromagnetic waves are radiated to the unit, signal cables or the transducer, the unit may not
provide satisfactory precision described in the specifications.
Avoid any contact with the component side, board or soldered parts of the instrument rack or the
module.
Failure to do so may result in damage to or malfunction of the unit.
The instrument rack and the module are precision instrument. Avoid any excessive shock or
vibration.
Failure to do so may result in damage to or malfunction of the unit.
Avoid a voltage resistance test or a withstand voltage test in any places other than specified.
Failure to do so may result in damage to the unit.
Caution
-
7/29/2019 VM-702B shinkawa
11/78
VM-702B Chapter 1For Safe Use of the Unit
9
Warning
1.3.2 Compliances
Use of this unit requires expert knowledge.
Be sure to follow the instructions in this manual when installing the unit.
Failure to do so may result in malfunction of the unit, fire or injury.
Be sure to connect the earth cable.
Failure to do so may result in an electric shock, electrocution or malfunction of the unit.
Before installation, be sure to read the manuals for this unit as well as other instruments that are to
be connected.The use of this unit requires expert knowledge of instruments, and the knowledgeof any associated equipment.
Before any contact with the unit, be sure to touch the metal objects, etc. so that the static electricity
is discharged.
Static electricity accumulated in the human body may damage the unit.
Use a time-delay circuit protector.
In-rush current caused by power activation may damage the unit.The value of in-rush current isfrom a few to ten times higher than the normal level.(The value of in-rush current variesdepending on the timing of power activation and the impedance of the power supply.)
When carrying out a megger test, ensure that all the cables are disconnected from the driver
(transducer) as well as the converter.
After carrying out a megger test, ensure that any accumulated charge is discharged by
short-circuiting prior to reconnection.
Reconnecting the driver or the converter while the cables still carry a charge may damage the unit.
Be sure to connect the power input terminal to the power supply with the voltage and frequency
advised in the specifications.
Failure to do so may result in damage to or malfunction of the unit.
Make sure to place the module in the instructed position when installing the module onto the
instrument rack.
Failure to do so may result in damage to the unit or malfunction of the turbine monitoring system.
Ensure that all the screws on the instrument rack are tightly fastened and the instrument rack is
securely fixed.
Also, be sure to use module fasteners to fix each module properly.
Caution
-
7/29/2019 VM-702B shinkawa
12/78
VM-702B Chapter 1For Safe Use of the Unit
10
Be sure to perform the following for the stability of the unit.
- Place surge killers around the unit, the related instrumentation equipment, and the relay around thesignal cables to which these are connected.
- Ensure that the screws are tightly fastened and the unit is securely fixed.- Ensure that an interlock function is added to the unit in order to prevent the unit from making
mechanical errors in the case it is damaged.
- Place the unit in a metal cabinet, etc. and ensure that the lid is closed while the unit is operated.- Use of ferrite cores (ZCAT3050-1330/TDK) can expect to reduce the noise induced by the cables to
a certain extent. It is recommended to use anti-noise devices in the case that the conditions
surrounding the unit are considered to be unfavourable.
Contact the sales office you purchased the unit from if any of the following occurs.Continued use of the unit under any of the conditions below may result in fire or malfunction.
- Abnormal odor is detected from the unit.- Overheats.- Is dropped.- Damaged packaging.
Ensure that an interlock function is added to the unit in order to prevent the unit from making
mechanical errors in the case it is damaged.
Install the unit in a place without the direct rays of the sun where the following conditions aresatisfied.Do not store it in a hot and humid location or any location where corrosive gases arepresent.
- Ambient temperature : -30 to 85C- Ambienthumidity : 20 to 90%RH (no condensation)
Caution
-
7/29/2019 VM-702B shinkawa
13/78
VM-702B Chapter 2Overview of Absolute Vibration Monitor Module
11
Chapter 2 Overview of Absolute VibrationMonitor Module
The absolute vibration monitor module is a dedicated vibration monitoring module that is capable of
simultaneously monitoring both the relative shaft vibration and absolute shaft vibration, or both the relative shaft
vibration and seismic vibration, of the rotating machinery.
The module is capable of receiving input from a maximum of two systems (relative vibration: two channels,
seismic vibration: two channels).
The recorder output or contact output corresponding to the input is externally output via the rear connector of the
instrument rack.
2.1 Functional Descript ionThis section provides the basic description of the measurement function, display function, alarm function, outputfunction and control function provided by the module.
2.1.1 Measurement Funct ionz Calculating the Gaps and Measurement Values
The offset voltage value and the amplitude value are detected from input signals to calculate the sensor set
gap voltage value and measurement value for each monitor.This module is capable of calculating the combination of following measurement values based on the input
signals for displacement vibration and velocity vibration.
Combination Output Channel 1/3 Output Channel 2/4
1 REL: Relative shaft vibration ABS: Absolute shaft vibration2 REL: Relative shaft vibration SEIS: Seismic vibration
3 SEIS: Seismic vibration ABS: Absolute shaft vibration
* REL (Relative): Relative shaft vibration, which is the displacement vibration output detected by the
displacement transducer* SEIS (Seismic): Seismic vibration, which is the output obtained by converting signals from the velocitytransducer to displacement vibration (amplitude)
* ABS (Absolute): Absolute shaft vibration, which is the output obtained by subtracting seismic vibration
from relative shaft vibration
z Calculating the Analysis Data (Optional)
Based on input signals, data required to analyze the vibration of the rotating machinery (1Xamplitude/phase, 2X amplitude/phase, etc.) is calculated.
This function is enabled only when the optional analysis function is added to the module.
z High-pass filter (low-cut filter)
A high-pole (10-pole) high-pass (low-cut) filter is available for selection, which is suitable under
circumstances in which the signal frequency is close to the noise frequency.
You can switch between the low-pole (4-pole or 2-pole) and high-pole high-pass (low-cut) filters byshorting between FILT1 and FILT2 terminals on the instrument rack.
z Suppression Function
This suppression is the function that cancels the floatation of the zero point by piping vibration or noise
when the machine is not running. The suppression can be set at 0.0 to 5.0% of the monitor range.
For details about the suppression setting, see "4.6.1 Basic Settings (Absolute Vibration Monitor) P.28".
-
7/29/2019 VM-702B shinkawa
14/78
VM-702B Chapter 2Overview of Absolute Vibration Monitor Module
12
Warning
2.1.2 Display Funct ionz Indicators on the Front Panel
The indicators on the front panel show the OK alarm status.
For details about each indicator, see "1.1 For Users of This Manual P.7".
zInformation Displayed on MCL ViewBy installing MCL View on the local PC, you can view various information regarding the module.
2.1.3 Alarm Funct ionFor details about the alarm setting, see "4.6.3 Alarm Setting P.36".
z OK (Abnormal Input) Alarm
If the input signal level of the transducer becomes abnormal, the OK alarm is output.
Even when the input signal level returns to normal, the OK alarm is retained for approximately another 15seconds by the off-delay timer.
z Gap AlarmThe ALERT Alarm is output when the input signal level falls outside the range set for the gap alarm.
(The gap alarm is ORed with ALERT Alarm and output to ALERT Alarm.)This function is available only for the displacement vibration input channel.
z DANGER Alarm
DANGER Alarm is output when the measurement value exceeds the DANGER setting value.
z ALERT Alarm
ALRET Alarm is output when the measurement value exceeds the ALERT setting value.
z First-out Indication
It is possible to specify the channel on which alarms are output the earliest from those channels that thefirst-out indication function is active.For details about the first-out indication function, see "4.6.1 Basic Settings (Absolute Vibration Monitor)
P.28".
z Sequence
To prevent alarm output due to excessive vibration generated temporarily while starting up the machine,
this function expands the DANGER Alarm and ALERT Alarm settings by shorting between SEQ1 andSEQ 2 terminals on the instrument rack.(Amplitude: magnification, by 1 to 10 times, phase: Enable/disable cutoff)For details about the sequence setting, see "4.6.3 Alarm Setting P.36".
To activate the contact for external operation, you need to short between terminals for more than 1
second.
-
7/29/2019 VM-702B shinkawa
15/78
VM-702B Chapter 2Overview of Absolute Vibration Monitor Module
13
Warning
z Alarm Delay
You can set the delay time (0.0 to 99.0 seconds) for each channel to prevent DANGER Alarm or ALERT
Alarm from being output by mistake due to noise or other reasons.For details about the alarm delay time setting, see "4.6 Absolute Vibration Channel P.28".
z Alarm Logic
Alarms for several channels can be output via AND or OR.
For details about the logic setting, see "4.7 Relay Settings P.38".
z Alarm Relay Mode
Either "normally de-energized" or "normally energized" can be selected for the alarm contact.For details about the alarm relay mode setting, see "4.7 Relay Settings P.38".
The use of "normally energized" (non-energized alarm) is recommended for the alarm relay mode.
The use of "normally de-energized" (energized alarm) may prevent the relay from being energizedwhen internal power supply is in an abnormal state.
z Alarm Reset Options
From the following options, select how the alarm should be handled when the input signal returns to normal:
Auto-Reset: The alarm contact output is cancelled automatically.
Self-hold: The alarm status is retained until a contact is operated to cause a short-circuit between RES1 and
RES2 terminals on the instrument rack, or until the alarm reset operation is performed on the VM-771BMCL View, or until alarm reset operation is performed using Modbus communication (optional).
For details about the alarm reset options for the absolute vibration channel, see "4.6.3 Alarm Setting P.36".For details about the reset options for alarm relay, see "4.7 Relay Settings P.38".
The alarm reset operation does not work while the alarm is output continuously.
2.1.4 Output Funct ionz Timed OK Channel Defeat
By enabling this function, shuts off DANGER Alarm, ALERT Alarm, indication values and recorder
outputs when the OK alarm is output.For details about how to set timed OK channel defeat, see "4.6 Absolute Vibration Channel P.28".
z Monitor Output
The buffered output of the input signal is output from the monitor/pulse output connector on the front
panel and the monitor output connector on the VM-76*B instrument rack.
The monitor output signals are not isolated from the input.The signal to be assigned to monitor output connectors 2 and 4 can be selected from velocity vibration(input signal obtained via buffer amplifier), seismic vibration and absolute shaft vibration.
For details about the monitor output setting, see "4.6 Absolute Vibration Channel P.28".
Caution
-
7/29/2019 VM-702B shinkawa
16/78
VM-702B Chapter 2Overview of Absolute Vibration Monitor Module
14
Warning
z Recorder Output
The analog signals (voltage signals or current signals) proportional to the vibration value are output from
the recorder output connector on the VM-76*B instrument rack.For details about the recorder output setting, see "4.6 Absolute Vibration Channel P.28".
2.1.5 Control Functionz Danger Bypass
This function shuts off DANGER Alarm on channels. While the alarm is shut off, both the DANGER
bypass indicator on Local Communication & Phase Marker Module and the DANGER bypass indicatoron the MCL View window light up.This function can be operated from MCL View.
z Channel Bypass
This function shuts off DANGER Alarm, ALERT Alarm, OK Alarm, indication values and recorder outputs
for channels. While the alarm output is shut off, the OK Alarm indicator remains off and the ChannelBypass indicator remains lit on the MCL View window.
This function can be operated from MCL View.
While the bypass is enabled, the machine is not protected.
z Power-on Timer Mode
This mode is activated for 15 seconds immediately after power-on. While power-on timer mode isactivated, the OK indicator flashes and all alarm functions are shut off.
-
7/29/2019 VM-702B shinkawa
17/78
VM-702B Chapter 2Overview of Absolute Vibration Monitor Module
15
2.2 Names and Functions of the Individual Parts
No. Name Function
1 CH1 OK indicator
CH1 is normal: On (green)
OK alarm occurs in CH1: Flash
2 CH2 OK indicatorCH2 is normal: On (green)
OK alarm occurs in CH2: Flash
3 CH3 OK indicatorCH3 is normal: On (green)
OK alarm occurs in CH3: Flash
4 CH4 OK indicatorCH4 is normal: On (green)
OK alarm occurs in CH4: Flash
5Monitor output connector1
Connector that outputs the CH1 input signals via buffer amplifier.
6Monitor output connector2
Specify the signal to be output using Device Config, from one of thefollowing three types:
x CH2 input signals, which are output via buffer amplifier
x CH2 input signals (velocity vibration), which are integrated and
converted to displacement vibration
x CH1 and CH2 signals, which are computed and then converted to theabsolute vibration (displacement vibration) signals
7Monitor output connector
3Connector that outputs the CH3 input signals via the buffer amplifier.
-
7/29/2019 VM-702B shinkawa
18/78
VM-702B Chapter 2Overview of Absolute Vibration Monitor Module
16
No. Name Function
8
Monitor output connector
4
Specify the signal to be output using Device Config, from one of thefollowing three types:
x CH4 input signals, which are output via buffer amplifier
xCH4 input signals (velocity vibration), which are integrated and
converted to displacement vibration
x CH3 and CH4 signals, which are computed and then converted to theabsolute vibration (displacement vibration) signals
9 Module fastenersFasteners used to fix the absolute vibration monitor module to the
instrument rack.
10 Main board Board on which electronic parts are mounted
11 Connector A Connector used to connect the unit to the motherboard
12 Connector B Connector used to connect the unit to the motherboard
13 Connector C Connector used to connect the unit to the motherboard
-
7/29/2019 VM-702B shinkawa
19/78
VM-702B Chapter 3 Installation/Uninstallation of a Module
17
Warning
Chapter 3 Installation/Uninstallation of aModule
This chapter describes how to install and uninstall a module.
Even while VM-7 is running, a module can be installed or uninstalled without affecting other modules.
Once a module is uninstalled, the alarm relay contact output terminal is in open mode.
If the alarm output of the module is assigned to the relay of other module, uninstall the module
using Channel Bypass.
When the Channel Bypass function is operating, no alarm is output even if the measurement value
exceeds the alarm setting value.
Before any contact with the unit, be sure to touch the metal objects, etc. so that the static electricity
is discharged.
Avoid any contact with the component side, board or soldered parts of the instrument rack or the
module.
Failure to do so may result in damage to or malfunction of the unit.
The instrument rack and the module are precision instrument.Avoid any excessive shock orvibration.
Caution
-
7/29/2019 VM-702B shinkawa
20/78
VM-702B Chapter 3 Installation/Uninstallation of a Module
18
Warning
3.1 Module Installation LocationVM-702B Absolute Vibration Monitor Module can be installed in Slot 1 to 11.
Do not install the unit in slots other than the one specified above.Otherwise unit failure couldoccur.
Slot 1 to 11
-
7/29/2019 VM-702B shinkawa
21/78
VM-702B Chapter 3 Installation/Uninstallation of a Module
19
3.2 Installing the Module1 The circuit board of the module is set to the slit on the top and
bottom of the instrument rack. And insert it straight.
2 An upper and lower module fastener is evenly turned by the
hand. It fastens until the module shoulder hits the instrument
rack edge.
(Do not use a tool to tighten the fasteners.
The module is inserted into the connector located inside therack.)
3 Turn on the power of the module. If it is operating normally, the
LED on the front panel will be lit or flashing.
(Tighten the fasteners until the shoulder of the module come in
contact with the turned edge of the instrument rack.If the LED
does not light up within 10 seconds, the module may be faulty.
In such a case, immediately uninstall the module to prevent a
heating hazard etc.)
Over-tightening the module fasteners could deform the module.
Make sure to install the module in one of Slots 1 to 11 on the instrument rack.
Ensure that all the fasteners on the instrument rack are tightly fastened and the instrument rack is
securely fixed.
4 If MCL View and Device Config are connected, update rack configuration data.
For details about how to update the software, see Updating the information displayed on MCL View
P.20 and Updating the information displayed on Device Config P.20.
Turned edge ofthe instrumentrack
Shoulder of themodule
Module fastener
Caution
-
7/29/2019 VM-702B shinkawa
22/78
VM-702B Chapter 3 Installation/Uninstallation of a Module
20
3.3 Uninstalling the Module1 If the alarm output of the module is assigned to other module,
bypass that channel.
(See "4.2.2 Channel Reset Window" in MCL View OperationManual.)
2 Loosen the module fasteners of the uninstalled module evenly.
Turn it from the instrument rack to the omission of the screw.
(The module disengages from the connector located inside the
rack.)
3 Slowly slide out the module from the instrument rack.
4 If MCL View and Device Config are connected, update
software data.
For details about how to update the software, see Updating the
information displayed on MCL View P.20 and Updating the
information displayed on Device Config P.20.
Updating the information displayed on MCL View After installing the module:
The information regarding the installed module is not
displayed on the window.
After uninstalling the module:A cross is displayed over the position of the slot from which
the module has been uninstalled.
1 Click the [CONNECT] icon.
The latest information is reflected on the window.
Updating the information disp layed on Device Config After installing the module:
The information regarding the installed module is not
displayed on the window.
After uninstalling the module:A cross is displayed over the position of the slot from which
the module has been uninstalled.
1 In the [Communication] menu,
click [Get Configuration].
2 Click [OK].The latest information is reflected on the window.
Modulefasteners
1
1
-
7/29/2019 VM-702B shinkawa
23/78
VM-702B Chapter 4 Module Setting
21
Chapter 4 Module SettingThis chapter describes how to set the VM-702B absolute vibration monitor module.
4.1 WorkflowThis module can be set as follows:
Set the internal switch.
Recorder output selection switch See "4.2.1 Recorder Output Selection Switch".
Load the setting See "4.3 Loading/Writing the Setting".
Edit the channel configuration See "4.4 Editing the Channel Configuration".
Edit the setting
Module common setting See "4.5 Common Module Settings".
Absolute vibration channel setting See "4.6 Absolute Vibration Channel".
Relay setting See "4.7 Relay Settings".
Write the setting See "4.3 Loading/Writing the Setting".
-
7/29/2019 VM-702B shinkawa
24/78
VM-702B Chapter 4 Module Setting
22
4.2 Setting the Internal Switch4.2.1 Recorder Output Selection Switch
To select the recorder output range, you need to change the switch on the module board.
After using the switch, change the Recorder Output Range setting for the channel using Device Config.
The switches corresponding to each channel are as follows:
CH1 SW600
CH2 SW601
CH3 SW602
CH4 SW603
1 Uninstall the module.
(See "3.3 Uninstalling the ModuleUninstalling the Module P.20".)
2 Select the switch position appropriate to the desired recorder output mode.
V: Voltage output (Example: 1V to 5 V, 0V to 5 V, 0V to 10 V)I: Current output (Example: 4mA to 20 mA)
3 After changing the switch setting, insert the module back in the slot.
(See "3.2 Installing the Module P.19".)
4 Check that the fasteners on the instrument rack are securely fastened.
-
7/29/2019 VM-702B shinkawa
25/78
VM-702B Chapter 4 Module Setting
23
Warning
4.3 Loading/Writ ing the SettingTo change the module operation, you have to load the setting from the module, edit it, and then write it back to the
module again.
Load or write the setting according to the steps described below.
4.3.1 Loading the Setting
1 In the [Communication] menu,
click [Get Configuration].
2 Click [OK].
Loading is completed.
4.3.2 Writing the Setting
When the channel configuration and settings are changed, alarm contacts may be output.
As the setting is written, the measurement values in MCL View and host network may disappear.
1 In the [Communication] menu,
click [Set Configuration].
2 Click [OK].
Writing is completed.
Point Writing Mode
Two modes are available for writing the setting from Device Config to VM-7.
- With [Set Configuration], the setting is written to all modules on the instrument rack. It may take some time
to transfer the setting to all modules.
- With [Set Slot Configuration], the setting is written to the modules in the selected slots.
Caution
1
1
-
7/29/2019 VM-702B shinkawa
26/78
VM-702B Chapter 4 Module Setting
24
4.4 Editing the Channel ConfigurationCheck that the internal switch settings are correct before proceeding to edit the channel configuration.
Connect the PC with VM-7 and start Device Config.
1 From [Device Tree] in the main window, select the module whose channel configuration you want toedit.
2 Click [Change Channel Type] on the [Edit] menu.
The Channel Configuration window appears.
1
2
-
7/29/2019 VM-702B shinkawa
27/78
VM-702B Chapter 4 Module Setting
25
3 From [Channel List], select the channel whose function you want to change.
4 From the [Monitor Type] list, select the function to be assigned to the channel.
The following functions can be assigned to a channel:
(None) Select this function if you are not using the channel.
Absolute Vibration Select this function if you are using the absolute vibrationmonitor.
5 Click the [Change] button.
6 The change is reflected in [Channel List]. Click the [OK] button to confirm the change.
3 4
5
6
-
7/29/2019 VM-702B shinkawa
28/78
VM-702B Chapter 4 Module Setting
26
4.5 Common Module SettingsThis section describes how to set the common information and functions for the modules.
1 From [Device Tree] on the left of the window, select the [Abs. Vibration] node.
2 Select the [Common] button on the upper right of the window to display the related setting items.
3 Set or check the following items:
Information
Serial Number The serial number is displayed.
Firmware ID The firmware ID is displayed.
Function
Timed-PowerON NOT-OK Output Select the OK alarm output status for the channel uponpower-on.
x Off: Do not output the OK alarm upon power-on.
x On: Output the OK alarm upon power-on.
Alarm / Relay Control Select Select the output for which to determine the delay time andreset method.
x Alarm: Determine the delay time and reset method for the
output generated immediately after alarm identification.
x Relay: Determine the delay time and reset method for the
output generated immediately after relay logic calculation.
* Make sure to set the same value for all modules in the same
rack.
1
23
-
7/29/2019 VM-702B shinkawa
29/78
VM-702B Chapter 4 Module Setting
27
Information(Analysis)
Serial Number The serial number for the analysis board is displayed.
Firmware ID The firmware ID for the analysis board is displayed.
-
7/29/2019 VM-702B shinkawa
30/78
VM-702B Chapter 4 Module Setting
28
4.6 Absolute Vibration Channel SettingThis section describes how to set the absolute vibration channel.
4.6.1 Basic Settings (Absolute Vibration Monitor)Set the basic functions related to the absolute vibration channel.
1 From [Device Tree] on the left of the window, select the [Abs. Vibration] node.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 From [Group List] on the left side of the window, select [Basic1-3]. (Under development)
4 Set or check the following items:
The absolute vibration monitor computes the input signals from two channels to measure relative shaft vibration,
seismic vibration and absolute shaft vibration. Thus, two channels are required to set the absolute vibrationmonitor.
Signals to be measured with this monitor can be selected from the combinations shown below. (The setting is
performed using [Measure Mode].)
z If REL (relative shaft vibration) is assigned to the first output channel and SEIS (seismicvibration) is assigned to the second output channel:
1
2
3
4
Input Ch1
Input Ch 2 Integration
Output Ch1
Output Ch 2
Displacement vibration REL
Velocity vibration
Phasecorrection
Absolute
vibration B
channel
Absolute
vibration A
channel
SEIS
-
7/29/2019 VM-702B shinkawa
31/78
VM-702B Chapter 4 Module Setting
29
z If REL (relative shaft vibration) is assigned to the first output channel and ABS (absolute shaftvibration) is assigned to the second output channel:
z If SEIS (seismic vibration) is assigned to the first output channel and ABS (absolute shaftvibration) is assigned to the second output channel:
Information
Channel Name Enter the channel name as necessary. The entered name will
be displayed in the bar graph window of MCL View. Amaximum of 60 characters can be entered.
* If over 20 single-byte characters are entered, or if the number
of double-byte characters or a total number of single-byte anddouble-byte characters exceeds 15 characters, the entered
characters may not be displayed in MCL View.
Tag Name Enter the tag name as necessary. The entered name will be
displayed in the bar graph window of MCL View. A
maximum of 60 characters can be entered.
* If over 20 single-byte characters are entered, or if the numberof double-byte characters or a total number of single-byte and
double-byte characters exceeds 15 characters, the entered
characters may not be displayed in MCL View.
Monitor
Monitor Type The monitor type representing the monitoring target isdisplayed.
Channel Type The channel type is displayed.
Measure Mode
(Absolute vibration A channel only)
Select the combination of signals to be measured.
Absolute vibration A channel (channel 1):
x 1st:REL(Ch1), 2nd:ABS(Ch2)
x 1st:REL(Ch1), 2nd:SEIS(Ch2)
x 1st:SEIS(Ch2), 2nd:ABS(Ch2)
Absolute vibration A channel (channel 3):
x
1st:REL(Ch3), 2nd:ABS(Ch4)x 1st:REL(Ch3), 2nd:SEIS(Ch4)
x 1st:SEIS(Ch4), 2nd:ABS(Ch4)
Input Ch1
Input Ch 2 Integration
Output Ch1
Output Ch 2
Displacement vibration
Velocity vibration
Phase
correction
Differential
calculation
Absolute
vibration A
channel
Absolute
vibration B
channel
SEIS
ABS
Input Ch1
Input Ch 2 Integration
Output Ch1
Output Ch 2
Displacement vibration
Velocity vibration
Phase
correction
Differential
calculation
Absolute
vibration B
channel
Absolute
vibration A
channel REL
ABS
-
7/29/2019 VM-702B shinkawa
32/78
VM-702B Chapter 4 Module Setting
30
Input
Input Signal Select the input transducer.
Absolute vibration A (displacement vibration input) channel
x VK-202A
x FK-202F for Vibration
Absolute vibration B (velocity vibration input) channel
x CV-86
x CV-88
Intrinsically Safe Select the explosion-proof specification for the connectedinput transducer.
x No: Non-intrinsically safe
x Yes: Intrinsically safe
Standard Sensitivity Txhe standard sensitivity of the input transducer is displayed.
If you have selected [Other Sensor] for input transducer, enter
the standard sensitivity of your choice.
Setting value: From -100.0 to +100.0 (except 0)
Actual Sensitivity Enter the actual sensitivity of the input transducer.
Setting value: From -100.0 to +100.0 (except 0)
* If the actual sensitivity of the input transducer is known,loop adjustment can be enabled by changing the [Actual
Sensitivity] value. However, be sure to calibrate with thestandard sensitivity before changing the [Actual Sensitivity]
value.
Measure
Measure Unit Select the unit for the measurement target.
um, mils, 1/100mm, cmm
Expression Style of Measure Select the number of digits following the decimal point to
display.
x 0.000 : Display three digits following the decimal point.
x 0.00 : Display two digits following the decimal point.
x 0.0 : Display one digit following the decimal point.
x 0 : Display an integer.
Monitor Range Monitor Range
Enter the full scale for the monitor range.
Suppression Enter the threshold value for the suppression function.
If this function is used, there may be no output if the input
signal level is small.
(Point Suppression Function)
Setting value: From 0.0% to 5.0%
Function
Timed-OK Channel Defeat Select whether to enable or disable the Timed-OK Channeldefeat function.
x Off: Disabled
x On: Enabled
-
7/29/2019 VM-702B shinkawa
33/78
VM-702B Chapter 4 Module Setting
31
Burn-out Select the output method for the measurement value andrecorder output, which will be applied to restrict output while
Timed-OK Channel defeat is enabled or while starting up the
unit, or while bypass is enabled.
x Down Scale (0%): Output the value equivalent to 0% of the
monitor range.
x Down Scale (0V/0 mA): Output 0 V or 0 mA.
First Out Select whether to enable or disable the first-out function.
If you enable this function, it is possible to specify the channelon which alarms are output the earliest from those channels
that the first out functionality is active.
x Off: Disabled
x On: Enabled
High-cut Filter Frequency Select the cutoff frequency of the low-pass filter (high-cut
filter).
There are restrictions on the combination of low-pass filter andhigh-pass filter (described below) that can be used. Make sure
that the setting conforms to the conditions specified in the table
below.
(Point Restrictions on Filter Combinations)
Low-cut Filter Select Select how to apply the high-pass filter (low-cut filter).
x Selectable: Switch the filter by input from an external
contact.
x LCF: Fixed to the 4-pole HPF (LCF)
x LCF: Fixed to the 10-pole HPF (LCF)
If you select [Selectable], you need to short between FILT1 and
FILT2 terminals on the instrument rack to enable the 10-pole
high-pass filter.
If this is done, the 4-pole high-pass filter will be disabled.
Low-cut Filter * In case of P-P, "2pole" is used
instead
Select the cutoff frequency for the 4-pole or 2-pole high-pass(low-cut) filter.
This filter features a -24 dB/oct. cutoff. In general, ample
damping effect can be anticipated with this filter.
Low-cut Filter Select the cutoff frequency of the 10-pole high-pass (low-cut)filter.
This filter features an approximately -60 dB/oct cutoff, and is
suitable if the cutoff frequency and the noise frequency areclose.
Phase Marker
Phase Marker Source
(Absolute vibration A channel only)
Select the phase marker to be used for analytic arithmetic.
x None: No phase marker is used (simulated rotation speed
enabled)
x Ch1 of Phase Marker Module
x Ch2 of Phase Marker Module
x Ch3 of Phase Marker Module
x Ch4 of Phase Marker Module
-
7/29/2019 VM-702B shinkawa
34/78
VM-702B Chapter 4 Module Setting
32
Output
Recorder Output Range Select the recorder output range. If you change the outputrange, you may need to change the internal switch setting. For
details about how to set the internal switch, see 4.2.1Recorder Output Selection Switch P.22.
x 4 to 20 mA
x 1 to 5 V
x 0 to 5 V
x 0 to 10 V
2nd Monitor Output
(Absolute vibration B channel only)
Select the signal to be assigned to the monitor output connector
for the absolute vibration B channel.
x BUFF : Output velocity vibration.
x SEIS : Output seismic vibration.
x ABS : Output absolute shaft vibration.
Point Suppression Function
This function prevents the zero point of the monitor from floating, which is caused by the piping vibration or
noise when the machine is not running.Any value within the monitor range, namely, 0.0% to 5.0% (in 0.1%
steps), can be set.
With the suppression function for the absolute vibration monitor, any input signals below the specified value are
converted to the 0% output signals. Input signals that are above the specified value but below 5.0% are converted
to a value lower than the true value and then output.
Example) Suppression Setting of 2.0% and 5.0%
0% 100(%)
INPUT
100(%)OUTPUT&
DISPLAY
0%5%
*1 Suppression settingvalue
0% 100(%)
INPUT
100(%)OUTPUT&
DISPLAY
0%5%
2.0% (*1) 5.0% (*1)
-
7/29/2019 VM-702B shinkawa
35/78
VM-702B Chapter 4 Module Setting
33
4.6.2 Analysis SettingsSet the basic functions related to analysis.
1 From [Device Tree] on the left of the window, select the [Abs. Vibration] node.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 From [Group List] on the left of the screen, select [Analysis1]. (Under development)
4 Set or check the following items:
Analysis
Target Machine The measurement mode appropriate to the machine to be
monitored is displayed.
Critical: For a large-sized rotating machine (sleeve bearing)
Sampling Interval : Trend The sampling interval for the trend data is displayed.
Sampling Interval : Waveform The sampling interval for the waveform data is displayed.
Gear Ratio Enter the gear ratio if the phase marker and the measurement
channel are installed on different shafts.
Setting value: From 0.001 to 10.000
(Point Setting the Gear Ratio)
Simurate RPM Enter the simulated rotation speed.
Setting value: 0 (not used), from 1 to 60000 rpm
* This function is enabled only when [None] is selected for
[Phase Marker Source] (basic setting).
Sampling Count Select the sampling count per cycle for synchronous sampling.
Setting value: 32, 64, 128
1
2
3
4
-
7/29/2019 VM-702B shinkawa
36/78
VM-702B Chapter 4 Module Setting
34
Sampling Frequency Select the sampling frequency for asynchronous sampling.
Setting value: 2, 5, 10, 20, 50 kHz
Waveform Length The length (number of lines) of the waveform data is displayed.
Setting value: Fixed to 800 lines
Low-cut Filter Frequency Select the frequency for the high-pass filter (low-cut filter).
Setting value: None (not used), 5Hz, 10Hz
PM Setting Angle Enter the mounting angle for the phase marker. Measure themounting angle from the top side, proceeding in the direction
opposite to the rotation direction of the shaft.
Setting value: From 0 to 359 degrees
Setting Angle Enter the mounting angle for the measurement channel.Measure the mounting angle from the top side, proceeding in
the direction opposite to the rotation direction of the shaft.
Setting value: From 0 to 359 degrees
Smax Calculation Method Select how to calculate Sp-p max.x None: No calculation
x Method A (Calculate the square-root of the sum of squares
of the orthogonal pk-pk data)
x Method B (Pick the maximum value of the orthogonal pk-pk
data)
Smax Pair Channel Select the other channel for the calculation of Sp-p max.
Setting value: From Channel 1 to Channel 4
NX1 Set Value Enter the value for the first NX.
Setting value: From 0.01 to 0.99, 0 (not used), from 3.00 to
10.00* NX is either the order of a synchronous component (example:2X, 3X, etc.) that is an integral multiple of therotation-synchronized frequency component (1X), or the order
of a synchronous component less than 1X (example: 0.1X,
0.2X, etc.).
NX2 Set Value Enter the value for the second NX.
Setting value: From 0.01 to 0.99, 0 (not used), from 3.00 to10.00
NX3 Set Value Enter the value for the third NX.
Setting value: From 0.01 to 0.99, 0 (not used), from 3.00 to
10.00
NX4 Set Value Enter the value for the fourth NX.
Setting value: From 0.01 to 0.99, 0 (not used), from 3.00 to10.00
FX1 Set Value Enter the value for the first FX.
Setting value: 0 (not used), from 0.01 to 20000.00 Hz
* FX represents the frequency of a specified frequencycomponent.
FX2 Set Value Enter the value for the second FX.
Setting value: 0 (not used), from 0.01 to 20000.00 Hz
-
7/29/2019 VM-702B shinkawa
37/78
VM-702B Chapter 4 Module Setting
35
Point Setting the Gear Ratio
In case where the phase marker cannot be mounted on the rotation shaft on which the sensor is installed, you can
use the gear ratio to obtain the rotation speed of the phase marker via the gear.Enter the value obtained from the following formula, where the gear on the rotation shaft of the phase marker is
the primary gear and the gear on the rotation shaft of the sensor is the secondary gear.
gearsecondarytheofteethofNumber
gearprimarytheofteethofNumberratioGear =
There is a following restriction on using the gear ratio
The trend data is calculated using asynchronous sampling waveform data.
Phase marker Primary
Secondary
-
7/29/2019 VM-702B shinkawa
38/78
-
7/29/2019 VM-702B shinkawa
39/78
VM-702B Chapter 4 Module Setting
37
Alarm Set Value
H-DANGER Enter the DANGER Alarm setting for the high side.
* Deselecting this checkbox disables the DANGER Alarmdetection function for the high side.
H-ALERT Enter the ALERT Alarm setting for the high side.
* Deselecting this checkbox disables the ALERT Alarmdetection function for the high side.
Alarm Delay
DANGER Enter the output delay time for DANGER Alarm.
Setting value: From 0.0 to 99.0 seconds
ALERT Enter the output delay time for ALERT Alarm.
Setting value: From 0.0 to 99.0 seconds
Alarm Reset
OK Select the OK alarm reset method.
x Auto Reset: Automatic reset method
x Self Hold: Self-hold method
DANGER Select the DANGER Alarm reset method.
x Auto Reset: Automatic reset method
x Self Hold: Self-hold method
ALERT Select the ALERT Alarm reset method.
x Auto Reset: Automatic reset method
x Self Hold: Self-hold method
Sequence
DANGER Enter the sequence setup magnification number for DANGERAlarm.
Setting value: 0 (not used), by 1.0 to 10.0 times
To enable the sequence function, you need to short between
SEQ1 and SEQ2 terminals on the instrument rack.
ALERT Enter the sequence setup magnification number for ALERTAlarm.
Setting value: 0 (not used), by 1.0 to 10.0 times
The value set for DANGER Alarm and ALERT Alarm should not exceed 110% of the monitor range.
Setting a value in excess of 110% can prevent the alarm from being identified.
If the output delay time set for DANGER Alarm and ALERT Alarm is short, chattering may occur in
the alarm output. If the input signal is unstable, set an appropriate delay time that is long enough.
To expand the alarm setting using the sequence function, set it to 110%or less of the monitor
range.
Setting a value in excess of 110% can prevent the alarm from being identified.
Caution
-
7/29/2019 VM-702B shinkawa
40/78
VM-702B Chapter 4 Module Setting
38
4.7 Relay SettingsThis section describes how to set the relay.
4.7.1 Operation SettingsSpecify the settings related to the relay operation.
1 From [Device Tree] on the left of the window, select the [Abs. Vibration] node.
2 On the right of the window, select the [Relay] button to display the relevant settings.
3 Set or check the following items:
Relay
Relay Mode Select the energization method for the relay.
x Normally De-Energized
x Normally Energized
(Point ALARM RELAY MODE)
Relay Delay Set the output delay time for the relay.
(This setting is applicable only when [Relay] is selected for
[Alarm / Relay Control Select])
Setting value: From 0.0 to 99.0 (in 0.1 steps)
Relay Reset Select the reset method for the relay.
(This setting is applicable only when [Relay] is selected for
[Alarm / Relay Control Select])x Auto Reset
x Self Hold
Relay Logic The relay logic is displayed.
1
23
-
7/29/2019 VM-702B shinkawa
41/78
VM-702B Chapter 4 Module Setting
39
Point Alarm Relay Mode
Each alarm relay contact can be set to either normally energized or normally de-energized.
Monitor Power ONMonitor Power
OFF Normal State Alarm state
NO contact, normally de-energized OPEN OPEN CLOSE
NO contact, normally energized OPEN CLOSE OPEN
NC contact, normally de-energized CLOSE CLOSE OPEN
NC contact, normally energized CLOSE OPEN CLOSE
-
7/29/2019 VM-702B shinkawa
42/78
VM-702B Chapter 4 Module Setting
40
4.7.2 Configur ing a Standard Relay Logic (Relay Logic1)Specify the settings related to the relay operation.
1 From [Device Tree] on the left of the window, select the [Abs. Vibration] node.
2 On the right of the window, select the [Relay] button to display the relevant settings.
3 Double-click the [Relay Logic] from the [Relay], to bring up the [Relay Logic Edit Dialog] window.
2
3
1
-
7/29/2019 VM-702B shinkawa
43/78
VM-702B Chapter 4 Module Setting
41
4 Set or check the following items.
Relay Logic1
LogicCode Set the relay output conditions.
Operand1 Set the channel and signal to be assigned to operand 1.
Operand2 Set the channel and signal to be assigned to operand 2.
Operand3 Set the channel and signal to be assigned to operand 3.
5 Click [Close] to close the window.
Point Alarm Logic
Each alarm logic is described in the table below.
Alarm Logic
LogicCode Description
None Disables the contact output.
Direct Outputs the alarm signal of the channel specified for [Operand1].
This logic can only be set for [Operand1].
AND Outputs the AND logic signals set for [Operand1], [Operand2] and [Operand3].
OR Outputs the OR logic signal specified for [Operand1], [Operand2] or [Operand3].
RACK COM-OK Outputs a signal indicating abnormal communication within the rack (including the
removal of modules).
5
4
-
7/29/2019 VM-702B shinkawa
44/78
VM-702B Chapter 4 Module Setting
42
Point Alarm Signals
Details of each alarm signal are shown in the table below.
Operand Description
None Disabled
Not-OK OK alarm signal
Alert ALERT Alarm signal (including the gap alarm)
Danger DANGER Alarm signal
4.7.3 Configur ing a Multistage Relay Logic (Relay Logic2)You can combine several alarms within the same rack and output the contacts.
This section describes how to output contacts using the logic described below.
Alarm A Alarm B Alarm C Description of the Logic
S3C1-Alert S3C3-Alert S3C2-Danger ! ( ( S3C1-Alert + S3C3-Alert ) * S3C2-Danger )
Alarm C
Alarm B
Alarm A
-
7/29/2019 VM-702B shinkawa
45/78
VM-702B Chapter 4 Module Setting
43
1 From [Device Tree] on the left of the window, select the [Abs.Vibration] node.
2 On the upper right of the window, select the [Relay] button to display the relevant settings.
3 Double-click the [Relay Logic] from the [Relay], to bring up the [Relay Logic Edit Dialog] window.
1
2
3
-
7/29/2019 VM-702B shinkawa
46/78
VM-702B Chapter 4 Module Setting
44
4 Select the alarm contact to be set.Click [RelayLogic2].
5 Select the [Enable Relay Logic2] checkbox.
If you select this checkbox, the [RelayLogic1] tab setting is disabled.
6 Click [(].
Example: (
7 From the pull-down menu, select alarm A (S3C1-Alert).
5
7
6
4
-
7/29/2019 VM-702B shinkawa
47/78
VM-702B Chapter 4 Module Setting
45
8 Click [Set].
Example: ( S3C1-Alert
If the field is blank or if calculation cannot be performed, the output of the contacts is disabled.
9 Click [OR(+)].
Example: ( S3C1-Alert +
10 From the pull-down menu, select alarm B (S3C3-Alert).
11 Click [Set].
Example: ( S3C1-Alert + S3C3-Alert
12 Click [)].
Example: ( S3C1-Alert + S3C3-Alert )
13 Click [] to move the cursor to the beginning of the line.
14 Click [(].
Example: ( ( S3C1-Alert + S3C3-Alert )
15 Click [] to move the cursor to the end of the line.
16 Click [AND(*)].
Example: ( ( S3C1-Alert + S3C3-Alert ) *
14
8, 11
12
13
15
16
9
10
-
7/29/2019 VM-702B shinkawa
48/78
VM-702B Chapter 4 Module Setting
46
17 From the pull-down menu, select alarm C (S3C2-Danger).
18 Click [Set].
Example: ( ( S3C1-Alert + S3C3-Alert ) *S3C2-Danger
19 Click [)].
Example: ( ( S3C1-Alert + S3C3-Alert ) *S3C2-Danger )
20 Click [] to move the cursor to the beginning of the line.
21 Click [!].
Example: ! ( ( S3C1-Alert + S3C3-Alert ) * S3C2-Danger )22 Click [OK] to close the window.
18
19
20
21
2217
-
7/29/2019 VM-702B shinkawa
49/78
VM-702B Chapter 4 Module Setting
47
Setting a Redundant Relay
1 Select the alarm contact to be set.
Click the [RelayLogic2] tab.
2 Select the [Enable Relay Logic2] checkbox.
If you select this checkbox, the [RelayLogic1] tab setting is disabled.
3 From the pull-down menu, select a relay.
The relay logic corresponding to the selected relay is output.4 Click [OK] to close the window.
12
3
4
-
7/29/2019 VM-702B shinkawa
50/78
VM-702B Chapter 5 Adjustment of the Module
48
Chapter 5 Adjustment of the Module
This chapter describes how to adjust the VM-702B absolute vibration monitor module.
5.1 Connecting the Equipment (Connecting the Input Terminals)Refer to the following figure and connect each piece of equipment.
Recommended Equipment
DC voltage generator: 7651 (YOKOGAWA)
Function generator: FG120 (YOKOGAWA)
If you are adjusting the gap (bias) voltage, connect only the DC voltage generator.
-
7/29/2019 VM-702B shinkawa
51/78
VM-702B Chapter 5 Adjustment of the Module
49
If you are adjusting the measurement values, connect both the DC voltage generator and thefunction generator.
Function Generator or DC Voltage Generator Connected Terminals
Channel A Channel B Channel C Channel D
Terminal
No.
DescriptionTerminal
No.
DescriptionTerminal
No.
DescriptionTerminal
No.
Description
A1 IN1 A6 IN2 A11 IN3 B1 IN4
A2 COM1 A7 COM2 A12 COM3 B2 COM4
-
7/29/2019 VM-702B shinkawa
52/78
VM-702B Chapter 5 Adjustment of the Module
50
5.2 Connecting the Equipment (Connecting the Output Terminals)Refer to the following figure and connect each piece of equipment.
A common method is used to connect the equipment, regardless of the monitor type.
Recommended Equipment
Digital multimeter: 34401A (Agilent technology)
Digital Multimeter Connected Terminals
Channel A Channel B Channel C Channel D
Terminal
No.Description
Terminal
No.Description
Terminal
No.Description
Terminal
No.Description
A1 REC1 A3 REC2 A5 REC3 B1 REC4
A2 COM1 A4 COM2 A6 COM3 B2 COM4
-
7/29/2019 VM-702B shinkawa
53/78
VM-702B Chapter 5 Adjustment of the Module
51
Warning
5.3 Adjusting the Gap (Bias) Voltage Measurement ValueThis section describes how to adjust the gap (bias) voltage measurement value of channels.
This adjustment should only be carried out if a significant error in gap (bias) voltage arises.
If this adjustment is carried out, alarm contacts may be output.
Perform adjustment on all channels to which transducers are connected, according to the steps described below.
1 From [Device Tree] on the left of the main window, select the [Abs. Vibration] node.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 Click the [Open] button for [Adjustment (Gap)] to bring up the Adjustment (Gap) window.
1
2
3
-
7/29/2019 VM-702B shinkawa
54/78
VM-702B Chapter 5 Adjustment of the Module
52
4 Output -20 V from the DC voltage generator.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 0% side of the
gap voltage.
5 Output +20 V from the DC voltage generator.Wait for approximately five seconds (settling time) and then click the [Set] button for the 100% side of
the gap voltage.
6 Click [Close] to close the window.
5.4 Adjusting the Measurement ValuesThis section describes how to adjust the measurement values for the absolute vibration monitor.
The adjustment procedure differs depending on the combination of measurement values used in the absolute
vibration monitor.
Select the adjustment procedure from the combination of signals selected in [Measure Mode].
Measure Mode Absolute Vibration A Channel Absolute Vibration B Channel
1st: REL(Ch1/3),
2nd: ABS(Ch2/4)
5.4.1 Absolute Vibration Monitor (REL:
Relative Shaft Vibration)
5.4.3 Absolute Vibration Monitor (ABS:
Absolute Shaft Vibration)
1st: REL(Ch1/3),2nd: SEIS(Ch2/4)
5.4.1 Absolute Vibration Monitor (REL:Relative Shaft Vibration)
5.4.2 Absolute Vibration Monitor (SEIS:Seismic Vibration)
1st: SEIS(Ch2/4),2nd: ABS(Ch2/4)
5.4.2 Absolute Vibration Monitor (SEIS:Seismic Vibration)
5.4.3 Absolute Vibration Monitor (ABS:Absolute Shaft Vibration)
4
56
-
7/29/2019 VM-702B shinkawa
55/78
VM-702B Chapter 5 Adjustment of the Module
53
Warning
5.4.1 Absolute Vibration Monitor (REL: Relative Shaft Vibration)This section describes how to adjust the measurement values when REL (relative shaft vibration) is assigned to the
output channel of the absolute vibration monitor.
If this adjustment is carried out, alarm contacts may be output.
1 From [Device Tree] on the left of the main window, select the appropriate module.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 Click the [Open] button for [Adjustment (Measure / Recorder)] of the absolute vibration A
(displacement vibration input) channel to bring up the Adjustment (Measure / Recorder) window.
Output Ch1
Output Ch 2
Output Ch1
Output Ch 2
Displacement vibration REL
Velocity vibration
Absolute
vibration A
channel
Absolute
vibration B
channel
1
2
3
-
7/29/2019 VM-702B shinkawa
56/78
VM-702B Chapter 5 Adjustment of the Module
54
4 The transducer to be connected to the absolute vibration A (displacement vibration input) channel is
either the VK or FK series. Output the offset voltage of -6 V to -12 V from the DC voltage generator.
5 From the function generator for the absolute vibration A (displacement vibration input) channel,
output AC voltage equivalent to 20% of the full scale.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 20% side of
[Measure].
List of Inputs for the Displacement Vibration Monitor
Input (Input Value) (m Vrms)
Non-intrinsically Safe Intrinsically SafeMonitorRange
Input Signal
20% 100% 20% 100%
100 mp-p VK-202A, FK-202F 55.6 278.2 55.1 275.5
125 mp-p VK-202A, FK-202F 69.6 347.8 68.9 344.3
200 mp-p VK-202A, FK-202F 111.3 556.5 110.2 550.9
400 mp-p VK-202A, FK-202F 223 1113 220 1102
500 mp-p VK-202A, FK-202F 278 1391 275 1377
3 mils p-p VK-202A, FK-202F 42.4 212.0 42.0 209.9
5 mils p-p VK-202A, FK-202F 70.7 353.4 70.0 349.8
10 mils p-p VK-202A, FK-202F 141.3 706.7 139.9 699.7
15 mils p-p VK-202A, FK-202F 212 1060 210 1050
5
6
7
The following calculation is applied for other ranges and input signals:
Non-intrinsically safe: Range x Input signal sensitivity/22 = mVrms
Example: 100 m x 7.87 mV/m/22 = 278.2 mVrms Intrinsically safe: Range x Input signal sensitivity/22 x 0.99 = mVrms
Example: 100 m x 7.87 mV/m/22 x 0.99 = 275.4 mVrmsFor explosion-proof specification, multiply the input by 0.99.
-
7/29/2019 VM-702B shinkawa
57/78
VM-702B Chapter 5 Adjustment of the Module
55
Warning
6 From the function generator for the absolute vibration A (displacement vibration input) channel,
output AC voltage equivalent to 100% of the full scale.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 100% side of
[Measure].
7 Click [Close] to close the window.
5.4.2 Absolute Vibration Monitor (SEIS: Seismic Vibration)This section describes how to adjust the measurement values when SEIS (seismic vibration) is assigned to the
output channel of the absolute vibration monitor.
If this adjustment is carried out, alarm contacts may be output.
z If either output channel 2 or output channel 4 is seismic vibration:
1 From [Device Tree] on the left of the main window, select the appropriate module.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 Click the [Open] button for [Adjustment (Measure / Recorder)] of the absolute vibration B (velocity
vibration input) channel to bring up the Adjustment (Measure / Recorder) window.
Input Ch1
Input Ch 2 Integration
Output Ch1
Output Ch 2
Displacement vibration
SEISVelocity vibration
Phase
correction
Absolute
vibration A
channel
Absolute
vibration B
channel
-
7/29/2019 VM-702B shinkawa
58/78
VM-702B Chapter 5 Adjustment of the Module
56
4 The transducer to be connected to the absolute vibration B (velocity vibration input) channel is either
the CV-86 or CV-88. Output the offset voltage of +6 V to +12 V from the DC voltage generator.
5 From the function generator for the absolute vibration B (velocity vibration input) channel, output
AC voltage equivalent to 20% of the full scale.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 20% side of
[Measure].
List of Inputs for the Velocity Vibration Monitor
Input (Input Value) (m Vrms)
Non-Intrinsically SafeIntrinsically Safe
(only CV-86)
MonitorRange
Input Signal
20% 100% 20% 100%
10mm/s pk CV-86, CV-88 5.57 27.86 5.52 27.58
20mm/s pk CV-86, CV-88 11.14 55.72 11.03 55.16
50mm/s pk CV-86, CV-88 27.9 139.3 27.6 137.9100 m p-p CV-86, CV-88 17.5 87.52 17.33 86.65
200 m p-p CV-86, CV-88 35.0 175.0 34.7 173.3
500 m p-p CV-86, CV-88 87.5 437.6 86.6 433.2
10mm/s rms CV-86, CV-88 7.88 39.40 78.0 39.01
20mm/s rms CV-86, CV-88 15.76 78.80 15.60 78.01
50mm/s rms CV-86, CV-88 39.4 197.0 39.0 195.0
100 m rms CV-86, CV-88 24.8 123.8 24.51 122.54
200 m rms CV-86, CV-88 49.5 247.6 49.0 245.1
500 m rms CV-86, CV-88 123.8 618.9 122.5 612.7
5
6
7
-
7/29/2019 VM-702B shinkawa
59/78
VM-702B Chapter 5 Adjustment of the Module
57
6 From the function generator for the absolute vibration B (velocity vibration input) channel, output
AC voltage equivalent to 100% of the full scale.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 100% side of
[Measure].
7 Click [Close] to close the window.
z If either output channel 1 or output channel 3 is seismic vibration:
1 From [Device Tree] on the left of the main window, select the appropriate module.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 Click the [Open] button for [Adjustment (Measure / Recorder)] of the absolute vibration A
(displacement vibration input) channel to bring up the Adjustment (Measure / Recorder) window.
Input Ch1
Input Ch 2 Integration
Output Ch1
Output Ch 2
Displacement vibration
Velocity vibration
Phase correction
SEIS
Absolute
vibration A
channel
Absolute
vibration B
channel
Note: The displacement range shown above is the input value when the calibration frequency is 100
Hz. The input value must be recalculated if a different calibration frequency is used.The following calculation is applied for other ranges and input signals (the range is converted to
mm):
Displacement range: 2 x x Calibration frequency x (Range/2) x Input signal sensitivity/2 =
mVrmsExample: 2 x x 100 Hz x (200 m/2) x 3.94 mV/2 = 175.0 mVrms Velocity range: Range x Input signal sensitivity/2 = mVrms
Example: 20 mm/s x 3.94 mV/2 = 55.72 mVrmsMultiply the input by 0.99 when the explosion proof specification is used.
In addition, when the rms rectification is used, multiply the range by 2 and then calculate the inputvalue and gain.
-
7/29/2019 VM-702B shinkawa
60/78
VM-702B Chapter 5 Adjustment of the Module
58
Warning
4 The transducer to be connected to the absolute vibration B (velocity vibration input) channel is either
the CV-86 or CV-88. Output the offset voltage of +6 V to +12 V from the DC voltage generator.
5 From the function generator for the absolute vibration B (velocity vibration input) channel, output
AC voltage equivalent to 20% of the full scale.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 20% side of
[Measure].
(See "List of Inputs for the Displacement Vibration Monitor P.54".)
6 From the function generator for the absolute vibration B (velocity vibration input) channel, outputAC voltage equivalent to 100% of the full scale.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 100% side of
[Measure].
7 Click [Close] to close the window.
5.4.3 Absolute Vibration Monitor (ABS: Absolute Shaft Vibration)This section describes how to adjust the measurement values when ABS (absolute shaft vibration) is assigned to
the output channel of the absolute vibration monitor.
If this adjustment is carried out, alarm contacts may be output.
5
6
7
Input Ch1
Input Ch 2 Integration Output Ch 2
Displacement vibration
Velocity vibration
Phasecorrection
Differential
calculation
ABS
Output Ch1Absolute
vibration A
channel
Absolute
vibration Bchannel
-
7/29/2019 VM-702B shinkawa
61/78
VM-702B Chapter 5 Adjustment of the Module
59
1 From [Device Tree] on the left of the main window, select the appropriate module.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 Click the [Open] button for [Adjustment (Measure / Recorder)] of the absolute vibration B (velocity
vibration input) channel to bring up the Adjustment (Measure / Recorder) window.
4 The transducer to be connected to the absolute vibration A (displacement vibration input) channel is
either the VK or FK series. Output the offset voltage of -6 V to -12 V from the DC voltage generator.
5 Set the output of the function generator for the absolute vibration A (displacement vibration input)
channel to 0 Vrms.
6 The transducer to be connected to the absolute vibration B (velocity vibration input) channel is either
the CV-86 or CV-88. Output the offset voltage of +6 V to +12 V from the DC voltage generator.
7 From the function generator for the absolute vibration B (velocity vibration input) channel, output
AC voltage equivalent to 20% of the full scale.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 20% side of
[Measure].
(See "List of Inputs for the Displacement Vibration Monitor P.54".)
8 From the function generator for the absolute vibration B (velocity vibration input) channel, output
AC voltage equivalent to 100% of the full scale.
Wait for approximately five seconds (settling time) and then click the [Set] button for the 100% side of
[Measure].
9 Click [Close] to close the window.
7
8
9
-
7/29/2019 VM-702B shinkawa
62/78
VM-702B Chapter 5 Adjustment of the Module
60
Warning
5.5 Adjusting the Recorder Output
This section describes how to adjust the recorder output of the absolute vibration monitor.
The adjustment procedure differs depending on the combination of measurement values used in the absolute
vibration monitor.
Select the adjustment procedure from the combination of signals selected in [Measure Mode].
Measure Mode Absolute Vibration A Channel Absolute Vibration B Channel
1st: REL(Ch1/3),
2nd: ABS(Ch2/4)
5.5.1 Absolute Vibration Monitor (REL:
Relative Shaft Vibration)
5.5.3 Absolute Vibration Monitor (ABS:
Absolute Shaft Vibration)
1st: REL(Ch1/3),2nd: SEIS(Ch2/4)
5.5.1 Absolute Vibration Monitor (REL:Relative Shaft Vibration)
5.5.2 Absolute Vibration Monitor (SEIS:Seismic Vibration)
1st: SEIS(Ch2/4),
2nd: ABS(Ch2/4)
5.5.2 Absolute Vibration Monitor (SEIS:
Seismic Vibration)
5.5.1 Absolute Vibration Monitor (ABS:
Absolute Shaft Vibration)
5.5.1 Absolute Vibration Monitor (REL: Relative Shaft Vibration)This section describes how to adjust the recorder output when REL (relative shaft vibration) is assigned to the
output channel of the absolute vibration monitor.
If this adjustment is carried out, alarm contacts may be output.
1
2
3
-
7/29/2019 VM-702B shinkawa
63/78
VM-702B Chapter 5 Adjustment of the Module
61
1 From [Device Tree] on the left of the main window, select the appropriate module.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 Click the [Open] button for [Adjustment (Measure / Recorder)] of the absolute vibration A
(displacement vibration input) channel to bring up the Adjustment (Measure / Recorder) window.
4 The transducer to be connected to the absolute vibration A (displacement vibration input) channel is
either the VK or FK series. Output the offset voltage of -6 V to -12 V from the DC voltage generator.
5 Set the output of the function generator for the absolute vibration A (displacement vibration input)
channel to 0 Vrms.
Adjust the [Offset] value for [Recorder Output] until the recorder output reads 0%.
(Click the [Set] button to reflect the changes to the recorder output.)
6 From the function generator for the absolute vibration A (displacement vibration input) channel,
output AC voltage equivalent to 100% of the full scale. Adjust the [Gain] value for [Recorder Output]
until the recorder output reads 100%.
(Click the [Set] button to reflect the changes to the recorder output.) (See "List of Inputs for the
Displacement Vibration Monitor P.54".)
7 Click [Close] to close the window.
5
6
7
Input Ch1
Input Ch 2
Output Ch1
Output Ch 2
Displacement vibration REL
Velocity vibration
Absolute
vibration A
channel
Absolute
vibration B
channel
-
7/29/2019 VM-702B shinkawa
64/78
VM-702B Chapter 5 Adjustment of the Module
62
Warning
5.5.2 Absolute Vibration Monitor (SEIS: Seismic Vibration)
This section describes how to adjust the recorder output when SEIS (seismic vibration) is assigned to the output
channel of the absolute vibration monitor.
If this adjustment is carried out, alarm contacts may be output.
z If either output channel 2 or output channel 4 is seismic vibration:
1 From [Device Tree] on the left of the main window, select the appropriate module.
2 Select the [Channel] button on the upper right of the window to display the related setting items.
3 Click the [Open] button for [Adjustment (Measure / Recorder)] of the absolute vibration B (velocity
vibration input) channel to bring up the Adjustment (Measure / Recorder) window.
Input Ch1
Input Ch 2 Integration
Output Ch1
Output Ch 2
Displacement vibration
SEISVelocity vibration
Phase
correction
Absolute
vibration A
channel
Absolute
vibration B
channel
-
7/29/2019 VM-702B shinkawa
65/78
VM-702B Chapter 5 Adjustment of the Module
63
4 The transducer to be connected to the absolute vibration B (velocity vibration input) channel is either
the CV-86 or CV-88. Output the offset voltage of +6 V to +12 V from the DC voltage generator.
5 Set the output of the function generator for the absolute vibration B (velocity vibration input) channel
to 0 Vrms.
Adjust the [Offset] value for [Recorder Output] until the recorder output reads 0%.
(Click the [Set] button to reflect the changes to the recorder output.)
6 From the function generator for the absolute vibration B (velocity vibration input) channel, output an
AC voltage equivalent to 100% of the full scale. Adjust the [Gain] value for [Recorder Output] until
the recorder output reads 100%.
(Click the [Set] button to reflect the changes to the recorder output.) (See "List of Inputs for the
Displacement Vibration MonitorP.54".)
7 Click [Close] to close the window.
z If either output channel 1 or output channel 3 is seismic vibration:
5, 12
6, 13
7, 14
Input Ch1
Input Ch