validation presentation
TRANSCRIPT
-
8/7/2019 VALIDATION PRESENTATION
1/28
Module 1,, Part 4: QC-related validation Slide 1 of 28 WHO EDM 1/2002
Validation Part 4:
QC-related validation
Supplementary Training Modules onGood Manufacturing Practice
-
8/7/2019 VALIDATION PRESENTATION
2/28
Module 1,, Part 4: QC-related validation Slide 2 of 28 WHO EDM 1/2002
Validation
Introduction
Why is analytical monitoring necessary?
What is the purpose of analytical validation?
-
8/7/2019 VALIDATION PRESENTATION
3/28
Module 1,, Part 4: QC-related validation Slide 3 of 28 WHO EDM 1/2002
Validation
Objectives
To introduce the concepts of :
Protocol development Instrument qualification
Analytical procedure
Extent of validation
Method transfer
Chemical and physical, biological, andmicrobiological test validation
-
8/7/2019 VALIDATION PRESENTATION
4/28
Module 1,, Part 4: QC-related validation Slide 4 of 28 WHO EDM 1/2002
Validation of analytical procedures
requires:
Qualified and calibrated instruments
Documented methods
Reliable reference standards
Qualified analysts
Sample integrity
Validation
-
8/7/2019 VALIDATION PRESENTATION
5/28
Module 1,, Part 4: QC-related validation Slide 5 of 28 WHO EDM 1/2002
Validation protocol for analytical method
Statement of purpose and scope
Responsibilities
Documented test method
List of materials and equipment
Procedure for the experiments for each parameter
Statistical analysis Acceptance criteria for each performance parameter
Validation
-
8/7/2019 VALIDATION PRESENTATION
6/28
Module 1,, Part 4: QC-related validation Slide 6 of 28 WHO EDM 1/2002
Qualification of the instrument
Make, model and makers manual
Modifications
Installation and operational qualification
Calibration programs
Maintenance schedules
Validation
-
8/7/2019 VALIDATION PRESENTATION
7/28
Module 1,, Part 4: QC-related validation Slide 7 of 28 WHO EDM 1/2002
Characteristics of analytical
procedures (1)
Accuracy
Precision
Repeatability
Reproducibility
Validation
-
8/7/2019 VALIDATION PRESENTATION
8/28
Module 1,, Part 4: QC-related validation Slide 8 of 28 WHO EDM 1/2002
Inaccurate &
imprecise
Inaccurate but
precise
Accurate but
imprecise
Validation
Relationship between accuracy andprecision
Accurate AND Precise
-
8/7/2019 VALIDATION PRESENTATION
9/28
Module 1,, Part 4: QC-related validation Slide 9 of 28 WHO EDM 1/2002
Characteristics of analytical procedures (2)
Ruggedness
Robustness
Variability caused by:
Day-to-day variations
Analyst-to-analyst
Laboratory-to-laboratory
Instrument-to-instrument Chromatographic column-to-column
Reagent kit-to-kit
Instability of analytical reagents
Validation
-
8/7/2019 VALIDATION PRESENTATION
10/28
Module 1,, Part 4: QC-related validation Slide 10 of 28 WHO EDM 1/2002
Characteristics of analytical procedures (3)
Linearity and range
Specificity
Sensitivity
Limit of detection
Limit of quantitation
Validation
-
8/7/2019 VALIDATION PRESENTATION
11/28
Module 1,, Part 4: QC-related validation Slide 11 of 28 WHO EDM 1/2002
Lin ity
of an analyte in a material
0.010
0.015
0.020
0.025
0.030
0.035
0.040
0.01 0.015 0.02 0.025 0.03 0.035 0.04
Reference material mg/mlCalculatedanalytein
mg/mL
Table of values (x,y)
x y
#Reference
aterial g/ l
Calculated
g/ l
1 0.0100 0.0101
2 0.0150 0.0145
3 0.0200 0.0210
4 0.0250 0.0260
5 0.0300 0.0294
6 0.0400 0.0410
Validation
-
8/7/2019 VALIDATION PRESENTATION
12/28
Module 1,, Part 4: QC-related validation Slide 12 of 28 WHO EDM 1/2002
Linearity Statistics
Intercept -0.0002
Limit of Linearity and Range 0.005 0.040 mg/mL Slope 1.0237
Correlation coefficient
Pearson 0.9978
Olkin and Pratt 0.9985
Relative procedure standarddeviation 3.4%
Validation
-
8/7/2019 VALIDATION PRESENTATION
13/28
Module 1,, Part 4: QC-related validation Slide 13 of 28 WHO EDM 1/2002
LOQ, LOD and SNR
Limit of Quantitation
Limit of Detection
Signal to Noise Ratio
noise
Peak A
LOD
Peak BLOQ
Baseline
Validation
-
8/7/2019 VALIDATION PRESENTATION
14/28
Module 1,, Part 4: QC-related validation Slide 14 of 28 WHO EDM 1/2002
Different classes of analytical tests
Class A: To establish identity
Class B: To detect and quantitate impurities
Class C: To determine quantitatively the
concentration
Class D: To assess the characteristics
Validation
-
8/7/2019 VALIDATION PRESENTATION
15/28
Module 1,, Part 4: QC-related validation Slide 15 of 28 WHO EDM 1/2002
* A degree of bias may be allowed
Characteristic A Bquant.
BLimittest
C D
Accuracy X X X*
Precision X X X
Robustness X X X X X
Linearity and range X X X
Specificity X X X X XLimit of detection X
Limit of quantitation X
Validation
-
8/7/2019 VALIDATION PRESENTATION
16/28
Module 1,, Part 4: QC-related validation Slide 16 of 28 WHO EDM 1/2002
Extent of validation
New methods require complete validation
Pharmacopoeial methods require partialvalidation (or verification)
Significant changes mean partial revalidation
equipment changes
formula changed changed suppliers of critical reagents
Validation
-
8/7/2019 VALIDATION PRESENTATION
17/28
Module 1,, Part 4: QC-related validation Slide 17 of 28 WHO EDM 1/2002
Analytical method transfer
Method transfer protocol and procedure
precision accuracy
ruggedness
Written and approved specific test method
Proficiency check
Formal acceptance by new laboratory
Validation
-
8/7/2019 VALIDATION PRESENTATION
18/28
Module 1,, Part 4: QC-related validation Slide 18 of 28 WHO EDM 1/2002
Chemical laboratory validation requirements (1)
Balances
Chromatography
HPLC, HPTLC, GC, TLC
Dissolution or disintegration apparatus
Karl Fischer moisture determination
Melting, softening or freezing point apparatus
Ovens, refrigerators, incubators
Validation
-
8/7/2019 VALIDATION PRESENTATION
19/28
Module 1,, Part 4: QC-related validation Slide 19 of 28 WHO EDM 1/2002
Chemical laboratory validation requirements (2)
pH meter
Polarimeter - optical rotation Refractometer
Spectrophotometer UV/Vis, IR, FTIR, Raman, AA
Timers
Viscometer
Volumetric equipment
Validation
-
8/7/2019 VALIDATION PRESENTATION
20/28
Module 1,, Part 4: QC-related validation Slide 20 of 28 WHO EDM 1/2002
Validation
Typical validation of HPCL assay (1)
System suitability (performance check)
system precision column efficiency
symmetry factor
capacity factor
-
8/7/2019 VALIDATION PRESENTATION
21/28
Module 1,, Part 4: QC-related validation Slide 21 of 28 WHO EDM 1/2002
Validation
Typical validation of HPLC assay (2)
Method validation
specificity accuracy
precision
linearity
robustness
-
8/7/2019 VALIDATION PRESENTATION
22/28
Module 1,, Part 4: QC-related validation Slide 22 of 28 WHO EDM 1/2002
Biological assays
Can be difficult to "validate"
"Validity" on a case by case basis
Strictly adhere to the Biological Testing
monographs in pharmacopoeias
Validation
-
8/7/2019 VALIDATION PRESENTATION
23/28
Module 1,, Part 4: QC-related validation Slide 23 of 28 WHO EDM 1/2002
Microbiological testing requiring validation
Microbial limit testing
Microbial count
Sterility testing
Preservative effectiveness testing
Environmental monitoring program
Biological testing
Validation
-
8/7/2019 VALIDATION PRESENTATION
24/28
Module 1,, Part 4: QC-related validation Slide 24 of 28 WHO EDM 1/2002
Validation of microbial test procedures (1)
Virtually impossible to completely validate testprocedures for every microorganism
Neutralize /inactivate inhibitory substances, ordilute
Periodic media challenge
Media QC
Reliable methods
Validation
-
8/7/2019 VALIDATION PRESENTATION
25/28
Module 1,, Part 4: QC-related validation Slide 25 of 28 WHO EDM 1/2002
Validation of microbial test procedures (2)
Incubation temperature and time
Media may not grow all microorganisms
Variations in media may affect recovery
Inhibitory disinfectants or preservatives
Sample
procedures
handling, storage, transport
Validation
-
8/7/2019 VALIDATION PRESENTATION
26/28
Module 1,, Part 4: QC-related validation Slide 26 of 28 WHO EDM 1/2002
Microbiological viable count methodvalidation (1)
Methods
pour plate / spread plate
membrane filtration
Most Probable Number
Sample size
Test dilution
Inoculation size
Validation
-
8/7/2019 VALIDATION PRESENTATION
27/28
Module 1,, Part 4: QC-related validation Slide 27 of 28 WHO EDM 1/2002
Microbiological viable count methodvalidation (2)
Membrane filtration conditions Incubation conditions
Acceptance criteria
Validation
-
8/7/2019 VALIDATION PRESENTATION
28/28
Module 1,, Part 4: QC-related validation Slide 28 of 28 WHO EDM 1/2002
Sterility testing validation requirements
Media growth promotion, sterility, pH
Product validation
Stasis testing
Environmental monitoring
Negative controls
Challenge organisms
Validation