validation of a particle size determination method for ... · analysis of particle size is of...

1
Analysis of particle size is of paramount importance in pharmaceutical product development. Although there are several methods for size characterization of particles, light scattering and optical microscopy methods are typically employed for particle size analysis in pharmaceutical product development and quality control activities. 1,2 Because the light scattering technique for size measurement is rapid, reproducible, easy to use and suitable for both wet and dry particles, and because the software is compliant with regulatory requirements of quality control/assurance laboratories, it is a preferred technique in pharmaceutical product development and stability studies. For wet particles smaller than 1 µm, dynamic laser light scattering (DLS), also known as photon correlation spectroscopy (PCS), is commonly used. 3 The transfer of methods for size determination from a method development laboratory to a quality group necessitates validation in accordance with FDA regulations. 4 The current study focuses on method development and validation of a size determination method for drug products comprised of nano-particles by DLS and focuses on the following validation parameters: 1) Linearity, 2) Robustness, 3) Specificity, 4) Reproducibility, 5) Intermediate Precision and 6) Limit of Detection. 5 In order to evaluate specificity, samples of emulsions and dispersed solid particles were characterized using both DLS and laser diffraction techniques. Robustness was evaluated by monitoring the effects of dispersant electrolyte concentration, sample concentration, measurement duration and temperature. 6 Linearity and reproducibility were determined using certified sizing standards, and intermediate precision was evaluated using a second analyst. LOD was determined by sample concentration and correlating it with attenuation index, turbidity and transmittance. RESULTS INTRODUCTION Data showed that measured value of diameters of standard particles was linear in the range of 60 nm to 350 nm; however, for particles of ≥1 µm, deviation of >10% from the certified value is observed, suggesting that the method is not suitable for large particles. The concentration of the particles in suspension is critical for accurate measurement of size by dynamic light scattering. A sample of pure deionized water showed the presence of particles, suggesting that it is critical to review other parameters, including polydispersity index and attenuation index, when interpreting the data obtained from the instrument. The dynamic light scattering technique is not suitable for the particles that have a tendency to settle. For the product comprised of particles of >1 µm, it is important to study effect of “measurement duration” during method development. The study indicates that particle size determination method using a dynamic laser light scattering technique can be validated in accordance with the FDA guidelines. CONCLUSIONS 1. Jillavenkatesa, A., Dapkunas, S.J., and Lum, L-S.H. “Particle Size Characterization” NIST Special Publication 960-1; U.S. Department of Commerce, U.S. Government Printing Office: Washington, D.C., 2001; pp. 1-2. 2. Meijer, N., Abbes, H., and Hansen, W.G. Particle size distribution and dispersion of oil-in-water emulsions; an application of light microscopy. American Lab. 33(8), 28-31(2001). 3. Burgess, D.J., Duffy, E., Etzler, F., and Hickey, A.J. “Particle Size Analysis: AAPS Workshop Report, Cosponsored by the Food and Drug Administration and the United States Pharmacopeia” AAPS J. 6(3), 1-12 (2004). 4. FDA Draft CMC Guidance for Industry. “Analytical Procedures and Methods Validation” Section XI, Part F, August (2000). 5. USP General Chapter <429> “Light Diffraction Measurement of Particle Size” Pharmacopeial Forum 28(4), 1293-1298 (2002). 6. “Method Validation for Laser Diffraction Measurements” Malvern Application Note MRK671-01. 7. “The Measurement of Latex Standards by Photon Correlation Spectroscopy” Malvern Application Note MRK111-01. 8. “Measuring NIST Traceable Standards Using Dynamic Light Scattering (DLS)” Malvern Application Note MRK481-01. REFERENCES Validation of a Particle Size Determination Method for Nano-particles by Dynamic Laser Light Scattering J. Brunotte, M. Darter, J. Vaughn and V. Kulkarni DPT Laboratories, San Antonio, Texas Poster# W4110 1.866.CALL.DPT WWW.DPTLABS.COM Particle Diameter, Z Average (nm) Concentration of KCI (mM) 72 71 70 69 68 67 66 65 64 63 62 0 5 10 15 20 25 Robustness Fig. 2: Influence of an electrolyte in the media (KCl) on the measured diameters for standard latex particles of 60 nm. Particle Diameter, Z Average (nm) DPT 10% emulsion Intralipid 20% Temperature (°C) 350.00 300.00 250.00 200.00 150.00 100.00 10 15 30 35 20 25 Robustness Fig. 4: Particle size (Z average) for Intralipid ® and the proprietary nano-emulsion were measured at different temperatures. Data show that, for both products, particle size is not influenced by the sample temperature in the range of 15-30 o C. Particle Diameter, Z Average (μm) Measurement Duration (s) 2.50 2.40 2.30 2.20 2.10 2.00 1.90 0 10 20 50 60 70 30 40 Robustness Fig. 3: Effect of measurement duration on 2 µm standard particles. The error bars indicate variation of ±1 standard deviation in the measured diameters. RESULTS Measurement of Intralipid ® Analyst I Analyst II Particle Size (nm) 295.8 312.4 296.7 316.2 300.9 315.7 295.8 314.6 296.7 314.7 300.9 313.8 Average 297.8 314.6 SD 2.4 1.4 % RSD 0.8 0.4 Pooled Average 306.2 Pooled SD 8.97 % RSD 2.93 Table 2: Particle size (Z average, in nm) data for Intralipid ® obtained by two analysts is shown. 1 10 100 1000 10000 Intensity (%) 16 14 12 10 8 6 4 2 0 Size (d.nm) Size Distribution by Intensity System Temperature (°C): 25.0 Duration Used (s): 20 Count Rate (kcps): 332.9 Measurement Position (mm): 4.65 Cell Description: Disposable sizing cuvette Attenuator: 8 Results Size (d.nm) % Intensity Width (d.nm) Z Average (d.nm): 315.8 Peak 1: 344.3 100.0 92.41 Pdl: 0.070 Peak 2: 0.000 0.0 0.000 Intercept: 0.924 Peak 3: 0.000 0.0 0.000 Result Quality: Good Record 853: 50 ppm Intralipid 2 Fig. 10: At an appropriate dilution, Intralipid ® showed a single sharp peak with low polydispersity index (0.07) and attenuator of 8. 1 10 100 1000 10000 Intensity (%) 10 8 6 4 2 0 Size (d.nm) Size Distribution by Intensity System Temperature (°C): 20.0 Duration Used (s): 70 Count Rate (kcps): 230.9 Measurement Position (mm): 0.85 Cell Description: Disposable sizing cuvette Attenuator: 2 Results Size (d.nm) % Intensity Width (d.nm) Z Average (d.nm): 277.1 Peak 1: 479.5 89.8 240.4 Pdl: 0.452 Peak 2: 71.68 8.5 21.55 Intercept: 0.743 Peak 3: 22.25 1.7 4.240 Result Quality: Good Record 344: 20% Intralipid IV Emulsion Fig. 11: A very concentrated sample of Intralipid ® showed multiple peaks with high polydispersity index (0.45) at attenuator of 2, suggesting that the sample concentration is too high for accurate size analysis. 1 10 100 1000 10000 Intensity (%) 20 15 10 5 0 Size (d.nm) Size Distribution by Intensity System Temperature (°C): 25.0 Duration Used (s): 20 Count Rate (kcps): 237.0 Measurement Position (mm): 4.65 Cell Description: Disposable sizing cuvette Attenuator: 11 Results Size (d.nm) % Intensity Width (d.nm) Z Average (d.nm): 390.9 Peak 1: 353.0 90.0 76.95 Pdl: 0.203 Peak 2: 2528 9.0 183.5 Intercept: 0.864 Peak 3: 0.000 0.0 0.000 Result Quality: Refer to quality report Record 850: 1.0 ppm IL Suspension 2 Fig. 12: A very dilute sample of Intralipid ® showed two peaks and high polydispersity index (0.2) at attenuator of 11, suggesting that the sample concentration is too dilute for accurate size analysis. 1 10 100 1000 10000 Intensity (%) 50 40 30 20 10 0 Size (d.nm) Size Distribution by Intensity System Temperature (°C): 25.0 Duration Used (s): 20 Count Rate (kcps): 78.3 Measurement Position (mm): 1.25 Cell Description: Disposable sizing cuvette Attenuator: 11 Results Size (d.nm) % Intensity Width (d.nm) Z Average (d.nm): 3933 Peak 1: 2478 100.0 210.9 Pdl: 0.984 Peak 2: 0.000 0.0 0.000 Intercept: 0.124 Peak 3: 0.000 0.0 0.000 Result Quality: Refer to quality report Record 882: di water 1 Fig. 13: A sample of deionized water showed polydispersity index of 0.98 and attenuator of 11, suggesting that when attenuator of 11 is reached, sample dilution is too high to accurately measure the size. 1 10 100 1000 10000 Intensity (%) 50 40 30 20 10 0 Size (d.nm) Size Distribution by Intensity System Temperature (°C): 20.0 Duration Used (s): 70 Count Rate (kcps): 197.7 Measurement Position (mm): 4.65 Cell Description: Disposable sizing cuvette Attenuator: 7 Results Size (d.nm) % Intensity Width (d.nm) Z Average (d.nm): 636.0 Peak 1: 305.7 66.0 47.37 Pdl: 0.560 Peak 2: 2617 33.0 113.6 Intercept: 0.887 Peak 3: 0.000 0.0 0.000 Result Quality: Refer to quality report Record 45: 100 ppm Kaolin Fig. 15: A sample of 100 ppm Kaolin showed multiple peaks and high polydispersity index of 0.560. Polydispersity Index (PDI) Particle Size (nm) Attenuator 0.500 0.450 0.400 0.350 0.300 0.250 0.200 0.150 0.100 0.050 0.000 390.0 370.0 350.0 330.0 310.0 290.0 270.0 250.0 0 1 2 3 4 5 6 7 8 9 10 11 12 Limit of Detection PDI Size (nm) Fig. 5: Trend of polydispersity index and particle size with the attenuator (inversely related to the sample concentration) for Intralipid ® . Polydispersity Index (PDI) Turbidity (NTU) 700.0 600.0 500.0 400.0 300.0 200.0 100.0 0.0 0.250 0.200 0.150 0.100 0.050 0.000 0 50 100 150 200 250 Concentration of Sample (μg/mL) Limit of Detection PDI Turbidity Fig. 7: The graph shows dependence of polydispersity index (PDI) and turbidity on the concentration of the sample of Intralipid ® . Transmittance at 700 nm (% T) Turbidity (NTU) Attenuator Limit of Detection 0 1 2 3 4 5 6 7 8 9 10 11 12 2500.0 2000.0 1500.0 1000.0 500.0 0.0 120 100 80 60 40 20 0 Transmittance Turbidity Fig. 9: The trend shows that for the nano-emulsion, with increasing attenuator, transmittance of light increases while turbidity decreases. Both transmittance and turbidity reach equilibrium for attenuators 6-11. 0 1 2 3 4 5 6 7 8 9 10 11 12 Polydispersity Index (PDI) Particle Size (nm) 160.0 155.0 150.0 145.0 140.0 0.140 0.120 0.100 0.080 0.060 0.040 0.020 0.000 Attenuator Limit of Detection PDI Particle Size Fig. 8: Trend of polydispersity index and particle size with the attenuator (inversely related to the sample concentration) for a nano-emulsion. 0 1 2 3 4 5 6 7 8 9 10 11 12 Transmittance at 700 nm (% T) Turbidity (NTU) 700 600 500 400 300 200 100 0 120 100 80 60 40 20 0 Attenuator Limit of Detection Turbidity % Transmittance Fig. 6: The trend shows that for Intralipid ® with increasing attenuator, transmittance of light increases while turbidity decreases. Both transmittance and turbidity reach equilibrium for attenuators 9-11. 0 1 2 3 4 5 6 Linearity Measured Diameters, Z Average (μm) Measured Diameters, Z Average (μm) Certified Diameters of Standard Particles (μm) R 2 = 0.9994 R 2 = 0.9982 0.400 0.350 0.300 0.250 0.200 0.150 0.100 0.050 0.000 5.0 4.0 3.0 2.0 1.0 0.0 Fig. 1: Measured diameters of standard particles are plotted against the certified values. A good correlation (R 2 >0.999) is seen for the particles in size range of 60 nm to 350 nm and R 2 of 0.998 for range of 60 nm to 5 µm. However, large deviation (≥10%) from the standard value is observed for the large particles. 0 200 400 600 800 1000 1200 Particle Size (μm) Distribution (Mastersizer) Z Average (μm) from Nano-S 0.800 0.750 0.700 0.650 0.600 0.550 0.500 0.450 0.400 16 14 12 10 8 6 4 2 0 Specificity Concentration (μg/mL) Dv10 Dv50 Dv90 Z Average from Nano-S Fig. 14: Comparison of size distribution data obtained from Mastersizer and that from the Nano-S for Kaolin. Standard Particles of 100 nm Z Average (nm) Polydispersity Index Mean Intensity Peak (nm) 104 0.006 104.7 103.3 0.003 104.1 103.8 0.015 104.1 102.7 0.002 103.5 103.9 0.006 104.8 103.5 0.007 104.7 104.4 0.006 105.1 102.6 0.011 103.8 104.3 0.032 104.2 104 0.015 104.3 103.9 0.011 104.4 104.2 0.017 104.4 103.4 0.005 104.7 103.6 0.013 103.9 102.6 0.011 103.9 103.2 0.004 103.9 103.5 0.012 104.9 103.8 0.016 105.6 Average 103.6 0.011 104.4 Minimum 102.6 0.002 103.5 Median 103.7 0.011 104.4 Maximum 104.4 0.032 105.6 SD 0.55 0.01 0.53 % RSD 0.53 66.60 0.51 Table 1: Data given in the Table shows particle size (Z average, nm), polydispersity index and mean intensity peak (nm) for 100 nm standard particles. Measurements were made on triplicate sample preparations with six replicate measurements per sample preparation. Average, median, standard deviation (SD), percent relative standard deviation (% RSD), minimum and maximum values determined from the 18 measurements are shown in the Table. Reproducibility Intermediate Precision Specificity of the method was evaluated by using different types of particles, including nano-emulsions and suspended particles. Attenuation index data from the Zetasizer was correlated to the turbidity of the sample and transmittance at the visual wavelength range of 600-700 nm. The studies indicated that the size of standard nano-particles determined by dynamic light scattering technique showed out-of-specification- high results in the dispersant without a strong electrolyte, whereas in the presence of an electrolyte the size data passed the specifications, suggesting that the presence of small amounts of strong electrolyte is essential for accurate size measurements. This observation is consistent with previously reported data. 7 Results obtained for standard particles >1 µm showed >10% deviation from certified value, indicating that additional sample manipulation may be required, including density matching. 8 DISCUSSION 1. Particle Size Analyzer: Zetasizer Nano S (ZEN 1600) and Mastersizer 2000 (both from Malvern Instruments, Westborough, MA). 2. Turbidimeter, 2100N (Hach Company, Loveland, CO) and UV Spectrophotometer (UV-1700, Pharma Spec, Shimadzu Scientific Instruments, Columbia, MD). 3. Suspensions of Standard Particles of 60 nm to 5 µm diameter (Thermo Scientific Inc., Fremont, CA). 4. Intralipid ® 20% (Baxter Healthcare Corp., Deerfield, IL) and Proprietary Nano-emulsion. 5. Kaolin Colloidal Powder (hydrated aluminum silicate), USP (Mineral and Pigment Solutions Inc., South Plainfield, NJ). MATERIALS AND METHODS AAPS Annual Meeting 2010

Upload: dangtruc

Post on 11-Jul-2018

229 views

Category:

Documents


0 download

TRANSCRIPT

Analysis of particle size is of paramount importance in pharmaceutical product development. Although there are several methods for size characterization of particles, light scattering and optical microscopy methods are typically employed for particle size analysis in pharmaceutical product development and quality control activities.1,2 Because the light scattering technique for size measurement is rapid, reproducible, easy to use and suitable for both wet and dry particles, and because the software is compliant with regulatory requirements of quality control/assurance laboratories, it is a preferred technique in pharmaceutical product development and stability studies. For wet particles smaller than 1 µm, dynamic laser light scattering (DLS), also known as photon correlation spectroscopy (PCS), is commonly used.3

The transfer of methods for size determination from a method development laboratory to a quality group necessitates validation in accordance with FDA regulations.4 The current study focuses on method development and validation of a size determination method for drug products comprised of nano-particles by DLS and focuses on the following validation parameters: 1) Linearity, 2) Robustness, 3) Specificity, 4) Reproducibility, 5) Intermediate Precision and 6) Limit of Detection.5

In order to evaluate specificity, samples of emulsions and dispersed solid particles were characterized using both DLS and laser diffraction techniques. Robustness was evaluated by monitoring the effects of dispersant electrolyte concentration, sample concentration, measurement duration and temperature.6 Linearity and reproducibility were determined using certified sizing standards, and intermediate precision was evaluated using a second analyst. LOD was determined by sample concentration and correlating it with attenuation index, turbidity and transmittance.

RESULTS

INTRODUCTION

• Data showed that measured value of diameters of standard particles was linear in the range of 60 nm to 350 nm; however, for particles of ≥1 µm, deviation of >10% from the certified value is observed, suggesting that the method is not suitable for large particles.• The concentration of the particles in suspension is critical for accurate measurement of size by dynamic light scattering.• A sample of pure deionized water showed the presence of particles, suggesting that it is critical to review other parameters, including polydispersity index and attenuation index, when interpreting the data obtained from the instrument. • The dynamic light scattering technique is not suitable for the particles that have a tendency to settle. For the product comprised of particles of >1 µm, it is important to study effect of “measurement duration” during method development. • The study indicates that particle size determination method using a dynamic laser light scattering technique can be validated in accordance with the FDA guidelines.

CONCLUSIONS

1. Jillavenkatesa, A., Dapkunas, S.J., and Lum, L-S.H. “Particle Size Characterization” NIST Special Publication 960-1; U.S. Department of Commerce, U.S. Government Printing Office: Washington, D.C., 2001; pp. 1-2.2. Meijer, N., Abbes, H., and Hansen, W.G. Particle size distribution and dispersion of oil-in-water emulsions; an application of light microscopy. American Lab. 33(8), 28-31(2001).3. Burgess, D.J., Duffy, E., Etzler, F., and Hickey, A.J. “Particle Size Analysis: AAPS Workshop Report, Cosponsored by the Food and Drug Administration and the United States Pharmacopeia” AAPS J. 6(3), 1-12 (2004).4. FDA Draft CMC Guidance for Industry. “Analytical Procedures and Methods Validation” Section XI, Part F, August (2000).5. USP General Chapter <429> “Light Diffraction Measurement of Particle Size” Pharmacopeial Forum 28(4), 1293-1298 (2002).6. “Method Validation for Laser Diffraction Measurements” Malvern Application Note MRK671-01.7. “The Measurement of Latex Standards by Photon Correlation Spectroscopy” Malvern Application Note MRK111-01.8. “Measuring NIST Traceable Standards Using Dynamic Light Scattering (DLS)” Malvern Application Note MRK481-01.

REFERENCES

Validation of a Particle Size Determination Method for Nano-particles by Dynamic Laser Light Scattering

J. Brunotte, M. Darter, J. Vaughn and V. KulkarniDPT Laboratories, San Antonio, Texas

Poster# W4110

1 . 8 6 6 . C A L L . D P TW W W. D P T L A B S. C O M

Part

icle

Dia

met

er, Z

Ave

rag

e (n

m)

Concentration of KCI (mM)

72

71

70

69

68

67

66

65

64

63

620 5 10 15 20 25

Robustness

Fig. 2: Influence of an electrolyte in the media (KCl) on the measured diameters for standard latex particles of 60 nm.

Part

icle

Dia

met

er, Z

Ave

rag

e (n

m)

DPT 10% emulsion Intralipid 20%

Temperature (°C)

350.00

300.00

250.00

200.00

150.00

100.0010 15 30 3520 25

Robustness

Fig. 4: Particle size (Z average) for Intralipid® and the proprietary nano-emulsion were measured at different temperatures. Data show that, for both products, particle size is not influenced by the sample temperature in the range of 15-30oC.

Part

icle

Dia

met

er, Z

Ave

rag

e (µ

m)

Measurement Duration (s)

2.50

2.40

2.30

2.20

2.10

2.00

1.900 10 20 50 60 7030 40

Robustness

Fig. 3: Effect of measurement duration on 2 µm standard particles. The error bars indicate variation of ±1 standard deviation in the measured diameters.

RESULTS

Measurement of Intralipid®

Analyst I Analyst IIParticle Size (nm)

295.8 312.4296.7 316.2300.9 315.7295.8 314.6296.7 314.7300.9 313.8

Average 297.8 314.6SD 2.4 1.4

% RSD 0.8 0.4

Pooled Average

306.2

Pooled SD 8.97% RSD 2.93

Table 2: Particle size (Z average, in nm) data for Intralipid® obtained by two analysts is shown.

1 10 100 1000 10000

Inte

nsi

ty (%

)

16

14

12

10

8

6

4

2

0

Size (d.nm)

Size Distribution by Intensity

System

Temperature (°C): 25.0 Duration Used (s): 20

Count Rate (kcps): 332.9 Measurement Position (mm): 4.65

Cell Description: Disposable sizing cuvette Attenuator: 8

Results

Size (d.nm) % Intensity Width (d.nm)

Z Average (d.nm): 315.8 Peak 1: 344.3 100.0 92.41

Pdl: 0.070 Peak 2: 0.000 0.0 0.000

Intercept: 0.924 Peak 3: 0.000 0.0 0.000

Result Quality: Good

Record 853: 50 ppm Intralipid 2

Fig. 10: At an appropriate dilution, Intralipid® showed a single sharp peak with low polydispersity index (0.07) and attenuator of 8.

1 10 100 1000 10000

Inte

nsi

ty (%

)

10

8

6

4

2

0

Size (d.nm)

Size Distribution by Intensity

System

Temperature (°C): 20.0 Duration Used (s): 70

Count Rate (kcps): 230.9 Measurement Position (mm): 0.85

Cell Description: Disposable sizing cuvette Attenuator: 2

Results

Size (d.nm) % Intensity Width (d.nm)

Z Average (d.nm): 277.1 Peak 1: 479.5 89.8 240.4

Pdl: 0.452 Peak 2: 71.68 8.5 21.55

Intercept: 0.743 Peak 3: 22.25 1.7 4.240

Result Quality: Good

Record 344: 20% Intralipid IV Emulsion

Fig. 11: A very concentrated sample of Intralipid® showed multiple peaks with high polydispersity index (0.45) at attenuator of 2, suggesting that the sample concentration is too high for accurate size analysis.

1 10 100 1000 10000

Inte

nsi

ty (%

)

20

15

10

5

0

Size (d.nm)

Size Distribution by Intensity

System

Temperature (°C): 25.0 Duration Used (s): 20

Count Rate (kcps): 237.0 Measurement Position (mm): 4.65

Cell Description: Disposable sizing cuvette Attenuator: 11

Results

Size (d.nm) % Intensity Width (d.nm)

Z Average (d.nm): 390.9 Peak 1: 353.0 90.0 76.95

Pdl: 0.203 Peak 2: 2528 9.0 183.5

Intercept: 0.864 Peak 3: 0.000 0.0 0.000

Result Quality: Refer to quality report

Record 850: 1.0 ppm IL Suspension 2

Fig. 12: A very dilute sample of Intralipid® showed two peaks and high polydispersity index (0.2) at attenuator of 11, suggesting that the sample concentration is too dilute for accurate size analysis.

1 10 100 1000 10000

Inte

nsi

ty (%

)

50

40

30

20

10

0

Size (d.nm)

Size Distribution by Intensity

System

Temperature (°C): 25.0 Duration Used (s): 20

Count Rate (kcps): 78.3 Measurement Position (mm): 1.25

Cell Description: Disposable sizing cuvette Attenuator: 11

Results

Size (d.nm) % Intensity Width (d.nm)

Z Average (d.nm): 3933 Peak 1: 2478 100.0 210.9

Pdl: 0.984 Peak 2: 0.000 0.0 0.000

Intercept: 0.124 Peak 3: 0.000 0.0 0.000

Result Quality: Refer to quality report

Record 882: di water 1

Fig. 13: A sample of deionized water showed polydispersity index of 0.98 and attenuator of 11, suggesting that when attenuator of 11 is reached, sample dilution is too high to accurately measure the size.

1 10 100 1000 10000

Inte

nsi

ty (%

)

50

40

30

20

10

0

Size (d.nm)

Size Distribution by Intensity

System

Temperature (°C): 20.0 Duration Used (s): 70

Count Rate (kcps): 197.7 Measurement Position (mm): 4.65

Cell Description: Disposable sizing cuvette Attenuator: 7

Results

Size (d.nm) % Intensity Width (d.nm)

Z Average (d.nm): 636.0 Peak 1: 305.7 66.0 47.37

Pdl: 0.560 Peak 2: 2617 33.0 113.6

Intercept: 0.887 Peak 3: 0.000 0.0 0.000

Result Quality: Refer to quality report

Record 45: 100 ppm Kaolin

Fig. 15: A sample of 100 ppm Kaolin showed multiple peaks and high polydispersity index of 0.560.

Poly

dis

per

sity

Ind

ex (P

DI)

Part

icle

Siz

e (n

m)

Attenuator

0.500

0.450

0.400

0.350

0.300

0.250

0.200

0.150

0.100

0.050

0.000

390.0

370.0

350.0

330.0

310.0

290.0

270.0

250.00 1 2 3 4 5 6 7 8 9 10 11 12

Limit of Detection

PDI

Size (nm)

Fig. 5: Trend of polydispersity index and particle size with the attenuator (inversely related to the sample concentration) for Intralipid®.

Poly

dis

per

sity

Ind

ex (P

DI)

Turb

idit

y (N

TU)

700.0

600.0

500.0

400.0

300.0

200.0

100.0

0.0

0.250

0.200

0.150

0.100

0.050

0.0000 50 100 150 200 250

Concentration of Sample (µg/mL)

Limit of Detection

PDI

Turbidity

Fig. 7: The graph shows dependence of polydispersity index (PDI) and turbidity on the concentration of the sample of Intralipid®.

Tran

smit

tan

ce a

t 70

0 n

m (%

T)

Turb

idit

y (N

TU)

Attenuator

Limit of Detection

0 1 2 3 4 5 6 7 8 9 10 11 12

2500.0

2000.0

1500.0

1000.0

500.0

0.0

120

100

80

60

40

20

0

Transmittance

Turbidity

Fig. 9: The trend shows that for the nano-emulsion, with increasing attenuator, transmittance of light increases while turbidity decreases. Both transmittance and turbidity reach equilibrium for attenuators 6-11.

0 1 2 3 4 5 6 7 8 9 10 11 12

Poly

dis

per

sity

Ind

ex (P

DI)

Part

icle

Siz

e (n

m)

160.0

155.0

150.0

145.0

140.0

0.140

0.120

0.100

0.080

0.060

0.040

0.020

0.000

Attenuator

Limit of Detection

PDI

Particle Size

Fig. 8: Trend of polydispersity index and particle size with the attenuator (inversely related to the sample concentration) for a nano-emulsion.

0 1 2 3 4 5 6 7 8 9 10 11 12

Tran

smit

tan

ce a

t 70

0 n

m (%

T)

Turb

idit

y (N

TU)

700

600

500

400

300

200

100

0

120

100

80

60

40

20

0

Attenuator

Limit of Detection

Turbidity

% Transmittance

Fig. 6: The trend shows that for Intralipid® with increasing attenuator, transmittance of light increases while turbidity decreases. Both transmittance and turbidity reach equilibrium for attenuators 9-11.

0 1 2 3 4 5 6

Linearity

Mea

sure

d D

iam

eter

s, Z

Ave

rag

e (µ

m)

Mea

sure

d D

iam

eter

s, Z

Ave

rag

e (µ

m)

Certified Diameters of Standard Particles (µm)

R2 = 0.9994

R2 = 0.9982

0.400

0.350

0.300

0.250

0.200

0.150

0.100

0.050

0.000

5.0

4.0

3.0

2.0

1.0

0.0

Fig. 1: Measured diameters of standard particles are plotted against the certified values. A good correlation (R2 >0.999) is seen for the particles in size range of 60 nm to 350 nm and R2 of 0.998 for range of 60 nm to 5 µm. However, large deviation (≥10%) from the standard value is observed for the large particles.

0 200 400 600 800 1000 1200

Part

icle

Siz

e (µ

m) D

istr

ibu

tio

n (M

aste

rsiz

er)

Z A

vera

ge

(µm

) fro

m N

ano

-S

0.800

0.750

0.700

0.650

0.600

0.550

0.500

0.450

0.400

16

14

12

10

8

6

4

2

0

Specificity

Concentration (µg/mL)

Dv10

Dv50

Dv90

Z Average from Nano-S

Fig. 14: Comparison of size distribution data obtained from Mastersizer and that from the Nano-S for Kaolin.

Standard Particles of

100 nm

ZAverage

(nm)

Polydispersity Index

Mean Intensity

Peak (nm)

104 0.006 104.7103.3 0.003 104.1103.8 0.015 104.1102.7 0.002 103.5103.9 0.006 104.8103.5 0.007 104.7104.4 0.006 105.1102.6 0.011 103.8104.3 0.032 104.2104 0.015 104.3

103.9 0.011 104.4104.2 0.017 104.4103.4 0.005 104.7103.6 0.013 103.9102.6 0.011 103.9103.2 0.004 103.9103.5 0.012 104.9103.8 0.016 105.6

Average 103.6 0.011 104.4Minimum 102.6 0.002 103.5Median 103.7 0.011 104.4

Maximum 104.4 0.032 105.6SD 0.55 0.01 0.53

% RSD 0.53 66.60 0.51

Table 1: Data given in the Table shows particle size (Z average, nm), polydispersity index and mean intensity peak (nm) for 100 nm standard particles. Measurements were made on triplicate sample preparations with six replicate measurements per sample preparation. Average, median, standard deviation (SD), percent relative standard deviation (% RSD), minimum and maximum values determined from the 18 measurements are shown in the Table.

Reproducibility

Intermediate Precision

Specificity of the method was evaluated by using different types of particles, including nano-emulsions and suspended particles. Attenuation index data from the Zetasizer was correlated to the turbidity of the sample and transmittance at the visual wavelength range of 600-700 nm. The studies indicated that the size of standard nano-particles determined by dynamic light scattering technique showed out-of-specification-high results in the dispersant without a strong electrolyte, whereas in the presence of an electrolyte the size data passed the specifications, suggesting that the presence of small amounts of strong electrolyte is essential for accurate size measurements. This observation is consistent with previously reported data.7 Results obtained for standard particles >1 µm showed >10% deviation from certified value, indicating that additional sample manipulation may be required, including density matching.8

DISCUSSION

1. Particle Size Analyzer: Zetasizer Nano S (ZEN 1600) and Mastersizer 2000 (both from Malvern Instruments, Westborough, MA).

2. Turbidimeter, 2100N (Hach Company, Loveland, CO) and UV Spectrophotometer (UV-1700, Pharma Spec, Shimadzu Scientific Instruments, Columbia, MD).

3. Suspensions of Standard Particles of 60 nm to 5 µm diameter (Thermo Scientific Inc., Fremont, CA).

4. Intralipid® 20% (Baxter Healthcare Corp., Deerfield, IL) and Proprietary Nano-emulsion.

5. Kaolin Colloidal Powder (hydrated aluminum silicate), USP (Mineral and Pigment Solutions Inc., South Plainfield, NJ).

MATERIALS AND METHODS

AAPS Annual Meeting 2010