Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 1 of 28
Models NPCU & TMCU Model TMLU
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 2 of 28
ContentsFeatures & Benefits 3
Introduction 4
General Instructions 4
InstallationControl Cabinet 5Headers 5Plumbing 6Electrical 8Headers, Pigtails & Cylinders 10
Start Up & Testing Procedures 12
Model TML Emergency Reserve 13
Cylinder Replacement & Handling 14
Line Delivery Pressure Adjustment 14
Replacement Parts 15
Trouble Shooting Guide 17
General Maintenance 18
Appendix A Technical Specifications 19
Appendix B Models NPC & TMC Piping Schematic 20
Appendix C Model TML Piping Schematic 21
Appendix D Models NPC & TMC Wiring Schematic 22
Appendix E Model TML Wiring Schematic 23
Appendix F Model TMC Heater Wiring Schematic 24
Warranty 25
Technical Assistance 1-888-769-7979
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 3 of 28
Features & Benefits• Fully automatic changeover • Novalvesorleverstoresetaftereachchangeover.
• Field upgradeable design • Kitsallowunittobechangedfrom-i.e.cylindersto portablebulkorfromstandardflowtohighfloworfrom lowerdeliverypressuretohigherdeliverypressure
• 400 psig pressure differential rated solenoid valve
• Exclusive single solenoid pressure differential rated changeover on cylinder x cylinder models
• Circuit board triggers secondary use alarm
• Easy to service layout/design
• Circuit board incorporates LEDs • Easilyvisibleeveninpoorlightingconditions
• Analog gauges provided
• Built in DISS gas specific emergency feed port
• Built in emergency reserve bank ports
• Input power 120 VAC, 50 to 60 Hz • Singlepointconnection,evenforunitswithheaters
• Dual line pressure regulators available
• Make unit NFPA 99 compliant
• Gas specific header bar with integral check valves and cylinder pigtail assemblies
• Variety of header configurations available to meet the available space requirements of your installation
• Optional part# PX-88-1075 single point vent kit available for NPCU models
• Optional part# PX-17-0169 and PX-17-0271 union assembly vent kits available for all models
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 4 of 28
IntroductionPowerex manifolds are cleaned for use with oxygen. Eachsystem is tested forchangeover, triggeringof secondary inusealarm, leakage and flow. Each unit is designed and preparedfor the indicated gas service. Powerex manifolds are built inaccordance with the National Fire Protection Association andCompressedGasAssociationguidelines.
WarrantyAll Powerex manifolds are warranted against defects in material and workmanship for the period of one year from date ofpurchase. All circuit boards are warranted against defects inmaterial andworkmanship for the period of three years fromdateofpurchase.
General Instructions/Location & ShelterManifolds should be installed in accordance with guidelinesstatedbytheNationalFireProtectionAssociation,theCompressedGasAssociation,OSHA, andall applicable local codes. Centralsupplysystemsandcylindersshouldnotbeplacedinalocationwherethetemperaturewillexceed125°F(51.6°C)orfallbelow-20°F(-29°C).Amanifoldplacedinanopenlocationshouldbeprotectedagainstweatherconditions.Duringwinter,protectthemanifoldfromiceandsnow.Insummer,shadethemanifoldandcylindersfromcontinuousexposuretodirectsunlight.
Leaveallprotectivecoversinplaceuntiltheirremovalisrequiredfor installation. This precautionwill keepmoisture anddebrisfromthepipinginterior.
Failure to follow the following instructions can result in personal
injury or property damage:
• Never permit oil, grease, or other combustible materials to come in contact with cylinders, manifold, and connections. Oil and grease may react with explosive force when ignited while in contact with some gases – particularly oxygen and nitrous oxide.
• Cylinder and master valves should always be opened very slowly. Heat of recompression may ignite combustible materials creating an explosive force.
• Pigtails should never be kinked, twisted, or bent into a radius smaller than 3 inches. Mistreatment may cause the pigtail to burst.
• Do not apply heat. Oil and grease may react with explosive force when ignited while in contact with some gases – particularly oxygen and nitrous oxide.
• Cylinders should always be secured with racks, chains, or straps. Unrestrained cylinders may fall over and damage or break off the cylinder valve which may propel the cylinder from its current position.
• Oxygen manifolds and cylinders should be grounded. Static discharges and lightning may ignite materials in an oxygen atmosphere, creating a fire or explosive force.
• Welding should not be performed near nitrous oxide piping. Excessive heat may cause the gas to dissociate, creating an explosive force.
• Remove all protective caps prior to assembly. The protective cap may ignite due to heat of recompression in an oxygen system.
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 5 of 28
Control Cabinet Installation• Determineandmarktheverticalcenterlineforinstallationofthemanifoldcontrolcabinet.
• If you wish to place cylinders under themanifold cabinet,measure from thefloor to a point 79” in height above thefinished floor of this vertical line. Using a level, mark ahorizontallineatthispointextendingapproximately7”totheleftand7”totherightofcenter.ThislineindicatesthelocationforthebottomtwomountingboltsoftheZmountingbracket.MountingtheZbracket@79”aff*willresultinthebottomofthemanifoldcabinetbeing60”aff–allowingcylinderstobeplacedunderthemanifoldcabinet.Ifyoudonotwishtoplacecylindersunderthemanifoldcabinet,measurefromthefloortoapoint58½”affandfollowthesamestepsabove.
•Mount the Zmounting bracket to the wall using fastenerssuitableforthetypeofwallconstruction.
• TemporarilyhangthemanifoldcabinetontheZbracketjustinstalled, mating it with the Z bracket on the back of themanifoldcabinet.Markthelocationsofthetwolowercabinetmountingholesonthewall.
• Remove themanifold cabinet and install female portion ofsuitablefastenersforlowercabinetmountingholes.
• Re-hangthemanifoldcabinetandinstallsuitablefastenersinthelowercabinetmountingholes.
*aff=abovefinishedfloor
Header Installation• Attachtheheaderstotheuniononeachsideofthemanifoldcontrol cabinet. Using a level, mark the placement ofmountingbracketswhilekeepingtheheaderonahorizontalplane.
• Remove the U-bolt assemblies from the header mountingbrackets.Positionthebracketssothatthetopofthebracketis alignedwith the bottom of the headers and is centeredbetweenthecylinderconnections.Theendbracketshouldbeplacedasclosetothelastcylinderaspossibletoprovidethemostsupportandstability.
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 6 of 28
•Markthemountingholeandinstallfastenerssuitablefortypeofwallconstruction.
• Fit the U-bolt over the header piping and tighten the twomountingnuts.
Plumbing• Theoutletofthemanifoldislocatedatthetopcenteroftheunitasshownhere.Theoutletconnectionis½NPTfemale.A½NPTmaleunionshouldbeinstalledbetweentheoutletofthemanifoldcabinetandthepipelinesystem.ThisunionisavailableaspartofanaccessorykitfromPowerex(part#PX-88-1075).
• NPC models include a PX-48-0023 ball valve ½ M NPT x ¾ extensiontube.
• TheintermediatereliefvalvesontheNPCmodelhavebeenplumbedtogether.The88-1075SinglePointVentKitshouldbe installed between the intermediate relief valves andthe relief (vent) pipeline system(s) for all indoor medicalapplicationsinordertobecompliantwithNFPA99.
• The TMC models do not include the pre-plumbing of theintermediate relief valves. Thisplumbingwouldneed tobeaddedbeforethe88-1075kitcouldbeused.
• PlumbingforTMCmodelsshownatright.Ventlines,(providedbyplumber),shownindottedlines.
• PlumbingforTMLmodelsshownbelow.Ventlines,(providedbyplumber),shownindottedlines.
Line relief valve ½ F npt
Intermediatereliefconnectionsfrombothintermediatereliefvalves(locatedontheheaders)maybeconnectedtoacommonventlinewiththelinereliefvalve.
Highpressurereservemaybeplumbedintotheintermediatereliefvalvefromeithertherightorleftsideofthecabinet.
Line relief valve ½ F nptOutlet½ F npt
Intermediatereliefconnectionsfrombothprimaryregulatorsconnected to a common vent line withthelinereliefvalve.88-1075SinglePointVentkitshownasdottedlines.
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 7 of 28
• In addition to connecting the left & right primary andsecondaryportablebulkvesselsupplybanks,theTMLmodelsmust also have a high pressure reserve bank of cylindersconnected to the cabinet, a check valve, an “EmergencyReserve inUse”andan“EmergencyReserveLow”pressureswitchtobeincompliancewithNFPA99guidelines.
• Slotshavebeenprovidedonboththeleftandrightsidesofthe control cabinet to allow for the high pressure reservepiping.
• A check valve (part # PX-CV-050F shown here) must beinstalled between the emergency reserve in use pressureswitchandthehighpressurereserveregulator.
• Part#PX-17-0169union(soldseparately)shownhere.
Emergency reserve in usepressureswitch(#PX-PS-2-150)
NFPA 99 compliant TML(shownwithoutventpiping) Emergency reserve
lowpressureswitch(#PX-PS-160-3200)
Highpressurereservemanifold(RWPseries)
Checkvalve(#PX-CV-050F)
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 8 of 28
Electrical• Useoneofthetwo½”conduitknock-outsprovidedlocatednearesttothetopleftcornerofthecabinettorouteconduittosupply120VACtothepowersupply.
Note: Separateconduitshouldbeusedforlowvoltagewires(useknockoutsprovidedontheleftsideofthebox).
120-240/1/50–60Hzmaybeusedwithallunits.Iftheunithasheaters,theheatersonlymaybewiredto120/1/50-60.Anadditionaltransformer(soldseparatelypart#35-3004)isrequiredtoconnectheatersto240VAC.
• Remove the power supply cover by loosening the screwlocatedat the topof the coverand then sliding thepowersupplycovertotherightuntilthescrewisinthecenterofthetear-dropshapedcutout.Next,pullthecoverforwarduntilitclearsthescrewheadandthefuse.Note:thebottomofthecoverinsertsintoaslotinthebackplate.Allowthecovertorestontheduallineregulatorassemblyplumbingjustbelowthepowersupply.
• Route wires of proper gauge (per local building coderequirement) through the power supply conduit, thru thegrommetonthepowersupplybracketandintotheterminalstrip.
• Connect the 120 VAC facility emergency power sourceelectricalwiringtotheterminalstripprovidedonthefrontofthepowersupplymountingbracket(perphotosright).
(N=neutral,L=load,FG=fieldground). Note: Thegroundmustbeasolidearthgroundwithlittleornoresistance.A“noisy”earthgroundmayaffectthedigitaldisplayofthemanifold.
Conduit knockouts for 120 VAC Conduit knockout for low voltage alarm signals
Neutral
Load
FieldGround
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 9 of 28
Caution: Never connect or disconnect any electrical components with the power on. This may result in damage to components and is not covered under warranty.
•Wires for remote alarms should be brought into thecabinet thru conduit or shielded cables (check local coderequirements) thru the knockouts on the left side of thecabinetshownhere.
Note: Separate conduit should be used for high voltagewires–neverrun lowvoltagewires inthesameconduitashighvoltagewires.
• IfyouareinstallingamodelNPC(cylinderxcylinder)cabinettomeetNFPA99therearethreealarmsignalsrequiredperNFPA99,HighLinePressure,LowLinePressureandReserveinUse.TheNPCcircuitboardwilltriggerthe“SecondaryinUse” alarm. Anoptional hi/lowpressure switchPart # PX-PS-1-80maybeinstalledinordertoreceivehigh&lowlinepressurealarmsignals.
Note:Thehi/lowpressureswitchwouldbewireddirectlytothemasteralarmpanels–nottothemanifoldcircuitboard.
• 3.Remotealarmwiresforthe“SecondaryinUse”alarmmaybeconnectedtoaterminalstrip locatedjusttotherightofthecircuitboard.Thewires(labeledA1&A2)connectingthisterminalstriptothecircuitareconnectedtodrycontactsonthecircuitboard.Novoltagehigherthan24VDCshouldbeconnectedtotheseterminals.
Note:Theremotealarmterminalsarenormallyclosedwhenthegaspressureisinthenormalrange.Thehi/lowsetpointspre-setinthemanifoldareasperthefollowingcharts:
Knockout for low voltage remote alarm wiring
Secondary in use “changeover”remotedrycontactconnections
Maximum Inlet Pressure Normal Delivery Press Line Relief Setting Secondary In Use Set Point
3,000 50 75 135leftbank/110rightbank
3,000 80 150 235leftbank/220rightbank
3,000 170 250 235leftbank/220rightbank
Maximum Inlet Pressure Normal Delivery Press Line Relief Setting Secondary In Use Set Point
400 50 75 60bothbanks
400 80 150 135bothbanks
400 170 250 190bothbanks
Pressure Settings for NPC & TMC Models(all pressures in psig)
Pressure Settings for TML Models(all pressures in psig)
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 10 of 28
Installing Headers & Attaching CylindersNPC & TMC Models•Matchthegasserviceindicatedbythelabelsonthecontrolcabinet with the gas service indicated by the label on theheadersandtheCGAfittingsontheheaderandthepigtails.Itiscriticalthattheheaders,pigtailsandcontrolcabinetgasservice/CGAmatch.
CGA320=CO2CarbonDioxide CGA326=N2ONitrousOxide CGA346=MedicalAir CGA540=Oxygen CGA580=N2Nitrogen or Argon
• Usingtwo1½”wrenches,oneontheinletblock insidethecontrolcabinetandtheotherontheunionnut,tightentheheaderstothecontrolcabinet.
• Check themaster valves to be certain they are open (turncounter-clockwisetoopen).(Note:themastervalveshouldalwaysbeleftopen.Itistobeusedonlyintheeventofanemergency).
• Attachthepigtailstotheheadercheckvalveoutletsusingtwo11/8”openendwrenches.
• SLOWLY open all cylinder valves (turn counter-clockwise toopen). Check all cylinder and pigtail connections for leaksusinganoxygensafeleaktestsolution(anybubblesformingaroundconnectionsindicateleakage).
• Checkallcylinderandpigtailconnectionsfor leaksusinganoxygensafeleaktestsolution(anybubblesorfoamformingaroundconnectionsindicateleakage).
Master Valve
CGA#saremarkedhere
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 11 of 28
Installing Pigtails & Attaching CylindersTML Models• AtypicalheaderforamodelTMLportablebulkvesselvaporwithdrawal manifold is shown here. CGA connections arestampedonboththeheaderfittingsandonthepigtailfittings.It is critical that thesefittingsmatch the gas serviceof thecontrolcabinet.
CGA320=CO2CarbonDioxide CGA326=N2ONitrousOxide CGA346=MedicalAir CGA540=Oxygen CGA580=N2Nitrogen or Argon
• ConnecttheendofthepigtailwiththeCGAfittingtothe“Use”valvemating fitting on the portable bulk vessel. Open theusevalve.Thepressurebuildingvalveorregulatorshouldbeturnedonoropenedforallvesselsconnectedtothemanifold(bothserviceandreservebanks).Allowapproximately1hourfortheportablebulkvessel(s)tobuildpressure.
• Checkallcylinderandpigtailconnectionsfor leaksusinganoxygensafeleaktestsolution(anybubblesorfoamformingaroundconnectionsindicateleakage).
• Verify that the pressure being supplied to the manifoldcabinetexceedstheminimuminletpressurerequirementsinthetablebelow.
CGAconnectionsforpigtailswithgastightcap,chain,andhookrings
Intermediate pressure relief valvex⅜Fnpt
Manifold Delivery Pressure
Minimum Inlet Pressure Relief Valve Setting on Vessel
50 150 235
80 150 235
170 250 350
Minimum Inlet Pressure Requirements for TML Manifold(all pressures in psig)
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 12 of 28
Start Up & Testing Procedures• Turnon thepower to theunit. Both the left& right bankRed(Depleted)LED’sshouldbeilluminatedandtheleftandright bank Green (In Use) and Yellow LED’s (Ready) shouldbeextinguished.Ifthemanifoldiswiredtoaremotealarm(masteralarm)–theSecondaryinUsealarmwillbeactivated.
• SLOWLYopenonecylindervalveon the leftbank. The leftbankpressuregauge(insidethecabinet)shoulddisplaythefull pressure of the cylinder. The Red (Empty) LED for theleft bank should have extinguished and theGreen (InUse)LEDshouldbeilluminated.TherightbankRed(Empty)LEDshouldremain illuminatedand if themanifold iswiredtoaremotealarm(masteralarm)–theSecondary inUsealarmwillbeactivated.
• SLOWLYopenonecylindervalveontherightbank.Therightbankpressuregauge(insidethecabinet)shoulddisplaythefull pressure of the cylinder. The Red (Empty) LED for therightbankshouldhaveextinguishedandtheYellow(Ready)LED should have illuminated. If themanifold iswired to aremotealarm(masteralarm)–theSecondary inUsealarmwillnotbeactivated.
• Turnoffallopenleftbankcylindervalves.Createaslightflowofgasinthedeliverypipelinesystem.Theleftbankpressuregaugeshouldfallandthecontrolautomaticallyswitchesovertotherightbank.Deliverypressureremainsconstant. TheleftbankRed(Empty)LEDwillilluminate.TheSecondaryinUsealarmshouldactivateontheremoteormasteralarm(s).
• SLOWLYreopenthecylindersontheleftbank.Theleftbankpressuregaugeshouldreturntofullpressure.Theleftbankyellow (Ready) LEDwill illuminate. Simultaneously the leftbankred(Empty)willextinguish.AllremoteSecondaryinUsealarmswillbecanceled.Repeatsteps2–4substitutingrightforlefttosimulateachangeoverfromtherightbanktotheleftbank.
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 13 of 28
Model TML – Emergency Reserve Pressure Settings & Alarm Set Points• NFPA99requiresfivealarmsforportablebulkmanifoldsystemswithhighpressurereserve;highlinepressure,lowlinepressure,secondaryinuse,emergencyreserveinuseandemergencyreservelow.
• Thehighandlowlinepressurealarmsaregeneratedbyapressureswitchwhichismountedonthedownstreamsideofthesourcevalve.NFPA99alsorequiresagaugeandaDISSunionassemblyonthedownstreamsideofthesourcevalve.Thisassembly(high/lowswitch,gaugeandDISSunionisavailableasanassembly.Thepart#’sareasfollows:
Gas Service Part #
Oxygen(O2) PX-PST-24
Medical Air PX-PST-16
NitrousOxide(N2) PX-PST-04
CarbonDioxide(CO2) PX-PST-08
Nitrogen(N2) PX-PSX-12
• Theemergencyreservelowbankpressureswitch(part#PX-PS-160-3200)shouldbeinstalledontheextraportontheRWPseriesmanifold.Thisportislocatedprior(upstream)tothemastervalveandtheregulator.
• Theemergencyreserveinusepressureswitch(part#PX-PS-2-150)shouldbeinstalledoncoppertubing(providedbyplumbingcontractor)after(downstream)thecheckvalve(part#PX-CV-050F).
• Refertothechartbelowforinformationonsettingthedelivery(outlet)pressureoftheemergencyreserveregulatorandemergencyreserveinuseandemergencyreservelowpressureswitches.
Manifold Delivery PressureRecommended Emergency Reserve Regulator Delivery
Pressure Setting
Recommended Emergency Reserve in Use alarm set
point
Pre-set* Emergency Reserve Low alarm set point
50 psig 65 psig 75 psig 1200 psig
80 psig 70 psig 80 psig 1200 psig
200 psig 170 psig 180 psig 1200 psig
*Emergencyreservelowpressureswitch(#PX-PS-160-3200)isshippedpre-setatapproximately1,200psig.Ifdesired,thisswitchmaybere-setinthefieldatadifferentpressuresetting.
Emergency reserve in usepressureswitch(#PX-PS-2-150)
NFPA 99 compliant TML(shownwithoutventpiping) Emergency reserve
lowpressureswitch(#PX-PS-160-3200)
Highpressurereservemanifold(RWPseries)
Checkvalve(#PX-CV-050F)
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 14 of 28
Cylinder Replacement & Handling• Closeallcylindervalvesonthedepletedbank.
• SLOWLYloosenandremovethepigtailconnectionfromthedepletedcylinders.
• Removedepletedcylindersandreplaceprotectivecaps.
• Placeandsecurefullcylindersintopositionusingchains,beltsorcylinderstands.
• Remove protective cylinder caps from full replacement cylinders.Withthevalveoutletpointedawayfromallpeopleinthearea,slowlyopeneachcylindervalveslightlytoblowoutanydirtorcontaminantsthatmayhavebecomelodgedinthecylindervalve.
• Connectpigtailstocylindervalvesandtightenwithwrench.
• SLOWLYturneachcylindervalveuntileachcylinderisfullyon.
• Observe the following conditions: The red (Empty) LED isextinguishedand theyellow (Ready)LED is illuminatedandtheremotereserveinusealarmiscancelled.
• The manifold supply bank is now replenished andautomaticallyplacedin“reserve”.
Line Delivery Pressure Adjustment• Leavethemanifoldinfulloperationalstatus.
• Createaflowconditioninthedeliverypipingsystem.ADISSdemandvalvehasbeenprovidedontheoutletblock.MatingDISSfittingsmaybeusedtocreateaflowofgaswithinthemanifoldcabinet.
• The line pressure regulator(s) is located at the top of themanifold cabinet. Ball valves are providedonly onmodelswithduallineregulatorsontheinletandoutletsidesofeachregulatordeterminewhichregulatoris“online”andwhichis“offline”.Note:whentheballvalvehandleisperpendiculartothepipeline,theballvalveisclosed.
• Turn the T-bar handle clockwise to increase line pressureor counter-clockwise to decrease line pressure. Itmay benecessarytousea3/4”open-endwrench,loosenthelocknutontheadjustingscrew(onhighflowmodelsonly).
• Afteradjustment,retightenthelocknut(onhighflowmodelsonly)ontheadjustingscrewandclosethecabinetdoor.
DISSDemandValve
Line Regulator
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 15 of 28
Item Part # Description
Primary Regulator & Repair Kits PX-68-0003R 0–300psigdelivery,3,000psiginlet,6port
PX-68-0003RK Rebuild kit for primary regulator 68-0003
Line Regulators & Repair Kits PX-68-0017R 5–125psigstandardflowlineregulator
PX-68-0017RK Rebuildkitfor5-125standardflowlineregulator68-0017
PX-68-0002R 5–125psighighflowlineregulator
PX-68-0002RK Rebuildkitfor5-125highflowlineregulator68-0002
PX-68-0001R 10–200psighighflowlineregulator
PX-68-0001RK Rebuildkitfor0–200highflowlineregulator68-0001
CircuitBoard PX-35-1007R UniversalNPC,TMC&TMLManifoldCircuitBoard
PressureSwitches(incontrolcabinet)
PX-14-2013 NPC,TMC&TMLLeftBankPressureSwitch
PX-14-2014 NPC,TMC&TMLRightBankPressureSwitch
PressureSwitches(outsidecabinet)
PX-PS-1-80 High/Lowlinepressureswitchrange1–80psigpreset@40/60
PX-PS-2-150 Emergencyreserveinuseswitchrange2–150psig
PX-PS-10-250L N2Lowlinepressureswitchrange10-250psigpreset@160onfall
PX-PS-10-250H N2Highlinepressureswitchrange10-250psigpreset@200onrise
PX-PS-160-3200 Emergencyreservelowswitchrange160–3200psigpreset@1200
PressureSwitch,Gauge,DISSAssemblies(includesaboveappropriatepressureswitch(switches2forN2)pre-assembledinaTeeorCrosswitha4”gauge&DISSunioncheckvalve
PX-PST-04 N2O
PX-PST-08 CO2
PX-PSX-12 N2
PX-PST-16 Medical Air
PX-PST-24 Oxygen
DISSDemandCheckUnionAssemblies
PX-PS-04 NitrousOxide¼Mnptdemandvalvex1/8Mnptnipple
PX-PS-08 CO2¼Mnptdemandvalvex1/8Mnptnipple
PX-PS-12 N2¼Mnptdemandvalvex1/8Mnptnipple
PX-PS-16 MedicalAir¼Mnptdemandvalvex1/8Mnptnipple
PX-PS-24 Oxygen¼Mnptdemandvalvex1/8Mnptnipple
Solenoid Valves PX-48-1007R Used on TMC models for all gas services
PX-48-1008R UsedonTMLmodelsforallgasservices-leftside
PX-48-1009R UsedonTMLmodelsforallgasservices-rightside
Power Supply PX-35-2013R 110VAC/24&5VDC
Gauges(insidecabinet) PX-14-1018 0–4,000psig1½”x1/8Mnptcenterback
PX-14-1016 0–400psig2”x¼Mnptbottomport
PX-14-1017 0–400psig1½”x1/8Mnptcenterback
PX-14-1009 0–300psig1½”x1/8Mnptcenterback
PX-14-1008 0–100psig1½”x1/8Mnptcenterback
Gauges(outsidecabinet) PX-ZV300-5 0–100psig4”x¼Mnptbottomport
PX-ZV300-6 0–300psig4”x¼Mnptbottomport
Relief Valves PX-RV-22-075 75 psig x ½ M npt inlet
PX-RV-22-150 150 psig x ½ M npt inlet
PX-RV-22-250 250 psig x ½ M npt inlet
PX-RV-11-400 400psigx¼Mnptinlet
Replacement Parts & Accessories
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 16 of 28
Item Part # Description
IntermediateCheckValve PX-17-4003R ½Mnptx½ODtubecompression
Pigtails for cylinders PX-20-1001 24”singlelooprigidcopperO2–CGA540
PX-20-0001 24”FlexiblestainlessbraidedO2-CGA540
PX-20-1002 24”singlelooprigidcopperN2O–CGA326
PX-20-0002 24”FlexiblestainlessbraidedN2O-CGA326
PX-20-0003 24”FlexiblestainlessbraidedCO2–CGA320
PX-20-0004 24”FlexiblestainlessbraidedAIR–CGA346
PX-20-0005 24”FlexiblestainlessbraidedN2orAr–CGA580
Pigtails for portable bulk vessels PX-20-2001 72”Flexiblewithcheckvalve–O2–CGA540
PX-20-2002 72”Flexiblewithcheckvalve–N2orAr–CGA580
PX-20-2003 72”Flexiblewithcheckvalve–CO2–CGA320
PX-20-2004 72”Flexiblewithcheckvalve–N2O–CGA326
Union PX-17-0169 ½”Mnptx½”Mnpt3pieceunion
WallBrackets PX-WB-1 Singlecylinderwallbracketwithchain
PX-WB-2 Dualcylinderwallbracketwithchain
Source Valve PX-48-0023 Ballvalve-1/2MNPTx3/4tubeextension
Replacement Parts & Accessories (Continued)
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 17 of 28
Symptom Probable Cause Remedy or Check
Cabinet Indicator Lights
Noindicatorlightsonfrontpanelilluminatewhenpowerisconnected.
Power input Checkelectricalpowersupply
Blownfuse Checkfusesonpowersupply
Internal wiring disconnected Checkallwiringconnections
Redindicatorlightsareonbutbothbanksarefull
Master valve or cylinder valves on bank are closed
OpenvalvesSLOWLY
Pigtailsareinstalledwithcheckvalvesinwrongdirection
Closecylindersandre-installpigtailsinproperflowdirection
Bankpressureisnotsufficientforlogicboardtoplaceitin“InUse”or“Ready”status
Replacecylinderswithfullcylinders.Or,ifusingportablebulkvessels,open pressure building valve on vessel or replace portable bulk vessel withhigherdeliverypressureportablebulkvessel
Primary regulators are out of adjustment
Adjustprimaryregulators–permanual
Circuitboarddefective Replace circuit board
Loss of Cylinder Contents
Audible or inaudible gas leakage (originunknown)
Leakage at manifold piping connections
Tighten,resealorreplace
Leakagethrumanifoldsolenoidvent / relief
Replace solenoid valve
Regulatorwithbadseat Rebuild or replace regulator
Leaking gauge Replace gauge
Ventingatreliefvalve Regulatorsettoohigh Setdeliverypressureperspecifications
Overpressureduetofailedregulator seat
Rebuild or replace regulator
Regulatorfreeze-up(N2OorCO2)/heaterfailure
Repairheateroraddheaterandconsideraddingadditionalcylinders
Leftbankwillnotbeplacedintoready(Red/Emptystatusstaysilluminated)whenbankpressureis full
Leftprimaryregulatorisoutofadjustment–settoolow
Adjustleftprimaryregulator–permanual
Rightbankwillnotbeplacedintoready(Red/Emptystatusstaysilluminated)whenbankpressureis full
Rightprimaryregulatorisoutofadjustment–settoolow
Adjustrightprimaryregulator–permanual
Bothbanksfeeding Leakingheader/pigtailconnection Tightenfittingorre-tapewithOxygensafeTeflontape(ifnptfitting)andtighten.
Leakingintermediatecheckvalve Replacecheckvalve
Leaking solenoid valve Replace solenoid valve
ModelNPC&TMC–primaryregulator(s)outofadjustment
Setdeliverypressuretospecificationsperchartonpage20
ModelTML–inletpressuretocontrol cabinet is too low
Verifythatminimuminletpressurerequirementsaremet
ModelTML–portablebulkventing Gasusagenothighenoughtojustifyportablebulkreserve
Troubleshooting GuideNOTE: Troubleshootingandrepairsshouldbedonebyqualifiedpersonnelonly.
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 18 of 28
Symptom Probable Cause Remedy or Check
Bothbanksfeeding(continued) ModelTML–gasflowingthrueconomizercircuit
Thisisnormalwhenreservebankpressureis50psiggreaterthanservicebankpressure–nocorrectionneededtomanifoldcontrolcabinet.Mayconsiderreducingsizeofbanksifreservebankismorethan35%depletedattimeitisplacedinservice“InUse”
Ifgasisflowingthrueconomizerwhenthereservebankpressureisnot50psiggreaterthanservicebankpressure–theeconomizercheckvalveneedstobereplaced.
Changeoveroccurs,reserveinusealarmistriggeredandthenclears
Manifold is unable to support requiredflow
Increasemanifoldflowcapacity(callfactoryforassistance)
ModelTML–portablebulkvessel(s)areunabletosupportrequiredflow
Increasebanksize.
If using two or more portable bulk vessels per bank currently, connect pigtail(s)(nocheckvalve)fromventtoventofallvesselsonthesamebankandopentheventvalves.Thiswillequalizetheheadpressureofthevesselsandutilizethecombinedvaporizationcapacity–notjustthecapacityofthevesselwiththehighestdeliverypressuresetpoint.
Troubleshooting Guide (Continued)NOTE: Troubleshootingandrepairsshouldbedonebyqualifiedpersonnelonly.
General Maintenance
Interval Control Cabinet Headers & Pigtails
Daily Record line and bank pressures Observenitrousoxideandcarbondioxidesystemsforcylinderfrostingorsurfacecondensation.Shouldexcessivecondensationorfrostingoccuritmaybenecessarytoincreasemanifoldcapacity.
Monthly Checkregulators,compressionfittingsandvalvesforexternalleakage.Checkvalvesforclosureability.Alternatelineregulatorinuse(ifdual).Checkregulatorsforstabledeliverypressure.
Inspectvalvesforproperclosure.Checkpigtailsforcleanliness,flexibility,wear,leakage,kinked,pinchedortwistedandthreaddamage.Replacedamagedpigtailsimmediately.Inspectheadercheckvalveoutletsforclosureability.
Annually CheckreliefvalvepressuresCheckregulatorseats.
Checkpigtailsforcleanliness,flexibility,wear,leakage,kinked,pinchedortwistedandthreaddamage.Replacedamagedorwornpigtailsimmediately.
Tightenregulatorbonnetsapproximately1to2degrees(outof360).
Every 4 years Replace all pigtails
Every 10 years Rebuildorreplace:allregulators,PX-17-4003intermediatecheckvalves,PX-48-1007solenoidvalve,andDISSdemandcheckvalves.
Rebuildorreplacemastervalves.Replaceallcheckvalveoutlets.
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 19 of 28
Appendix A
Manifold SpecificationsMinimum Inlet Pressure: ModelNPC&TMC: 3000psig
ModelTML: 400psig
Operating Ambient Temperature range: ModelNPC&TMC: 0°F(-18°C)to130°F(54.4°C)allgasesexceptN20&CO2
20°F(-7°C)to130°F(54.4°C)N2O&CO2
ModelTML: -20°F(-29°C)to130°F(54.4°C)
Storage Temperature:-20°C(-4°F)to+85°C(185°F)
AC Input:120voltsAC–50-60Hz
Input Fuse: 5ampinputAClinefuseprotectstheinputwiringtopowersupply
Power Consumption: 45W(0.4ampsusing120VAC)maximumwithoutheaters
50W(2.1ampsusing120VAC)maximumwithheaters
Pressure Measurement Accuracy: Allgaugesare3/2/3design.Theyare±3%offullscaleinthefirstandlast1/3oftheirrange. Theyare±2%offullscaleinthemiddle1/3rdoftheirrange.
Solenoid Valve: 24VDC–NormallyOpen(Valveopenswhende-energized)
Control Cabinet Dimensions: Dimensionsexcludinginlet&outletfittings
15¾”Wx24¾”Hx9¼”D
Dimensionsincludinginlet&outletfittings
17¼”Wx27”Hx9¼”D
Primary Regulator Settings (NPC & TMC Models)
NOTE: Allsettingsdonewith2,200psigcylinderpressureandwithslightgasflowthruthemanifold.Primaryregulatoroutletpressurewillvarywithvaryinginletpressures.(Theoutletpressurewillriseasthecylinderpressuredecreasesasmuchas40–50psig).Allpressuresshowninpsig.
Normal Delivery Pressure (factory delivery pressure) Left Primary Regulator Set Point Right Primary Regulator Set Point
50 160 120
80 260 220
170 260 220
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 20 of 28
Appendix B
NPC & TMC Piping Schematic
120VACconnectionterminals,fuses and power supply
(transformer)
Possible line pressure regulator(iforderedwith
duallineregulators)
Checkvalve
Pressureswitch(seepage9forproper
settings)
Rightbankprimaryregulator(seepage18for
propersettings)
Rightbankprimaryregulator outlet pressure gauge
Inlet block
Inlet pressure gauge
Checkvalve
Intermediate block
Line pressure regulatorSeepage14forsetting
instructions
DISSdemandvalve–maybe used to bleed line
pressure during service and as an emergency feed port
forthefacility
Line pressure relief valve ½ F npt(notvisibleinthisphoto-outsidecabinet)
Outlet–½Fnpt(notvisibleinthisphoto–outsidecabinet)Line pressure gauge
Intermediate pressure relief valves(behindcircuitboardonbottomportofprimaryregulators
Solenoid valve
Pressureswitch(seepage9forproper
settings)
Leftbankprimaryregulator(seepage18for
propersettings)
Leftbankprimaryregulator outlet pressure gauge
Inlet pressure block
Inlet pressure gauge
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 21 of 28
Appendix C
TML Piping Schematic
120VACconnectionterminals,fuses and power supply
(transformer)
Possible line pressure regulator(iforderedwith
duallineregulators)
Checkvalve
Intermediate block
Line pressure regulatorSeepage14forsetting
instructions
DISSdemandvalve–maybe used to bleed line
pressure during service and as an emergency feed port
forthefacility
Line pressure relief valve ½ F npt(notvisibleinthisphoto-outsidecabinet)
Outlet–½Fnpt(notvisibleinthisphoto–outsidecabinet)Line pressure gauge
Circuit board
Solenoid valve
Inlet pressure gauge
Highpressureswitch
Lowpressureswitch
EconomizercircuitCheckvalve
Solenoid valve
Inlet pressure gauge
Highpressureswitch
Lowpressureswitch
Economizercircuit
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 22 of 28
Appendix D
NPC & TMC Wiring Schematic
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 23 of 28
Appendix E
TML Wiring Schematic
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 24 of 28
Appendix F
TMC Heater Wiring Schematic
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 25 of 28
Powerex Limited WarrantyWarranty and Remedies. (a) Standard Period of Warranty – Parts and Labor. PowerexwarrantsandrepresentsallProductsshallbefreefromDefectsforthefirsttwenty-four(24)monthsfromthedateofshipmentbyPowerex.Duringsuchwarrantyperiod,PowerexshallbefullyliableforallDefectsintheProducts(the“ProductDefects”),i.e.,allcostsofrepairorreplacement,whichmayinclude“inandout”charges,solongastheProductsarelocatedintheUnitedStatesorCanada,andtheProductsarereasonablylocatedandaccessiblebyservicepersonnelforremoval.“Inandout”chargesincludethecostsofremovingaProductfrombuyer’sequipmentforrepairorreplacement.
(b) Additional Period of Warranty – Parts Only (No Labor). Inadditiontotheabove,Powerexwarrantstheproductsdescribedhereintobefreefromdefectsinmaterialandworkmanshipforsixty(60)monthsfromendofStandardPeriodofWarrantywiththeexceptionofanycomponentswhicharerecommendedtobereplacedinlessthansixtymonthsinourInstallation/Operationmanuals.WithinsaidperiodPowerexwillrepairorreplaceanypartorcomponentwhichisproventobedefectiveineithermaterialorworkmanship.Thiswarrantycoverspartsonly.Laborisnotincluded.ThiswarrantyisvalidonlywhentheproducthasbeenproperlyinstalledaccordingtoPowerex specifications,used in anormalmannerand servicedaccording to factory recommendations. Thiswarrantydoesnotcoverfailuresduetodamagewhichoccursinshipmentorfailureswhichresultfromaccidents,misuse,abuse,neglect,mishandling,alteration,misapplicationordamageduetoactsofnature.
(c) General. PowerexwarrantseachPowerexbrandedPipelineAccessory(collectively“Products”,individuallyeacha“Product”)tobe free fromdefects inmaterialandworkmanship (“Defects”)at thedateofshipment.EXCEPTASSETFORTHBELOW,NOOTHERWARRANTY,WHETHER EXPRESSOR IMPLIED, INCLUDINGANYWARRANTYOFMERCHANTABILITYOR FITNESS FOR A PARTICULARPURPOSE,SHALLEXISTINCONNECTIONWITHTHESALEORUSEOFSUCHPRODUCTS.TOTHEEXTENTPERMITTEDBYLAW,ANYANDALLIMPLIEDWARRANTIESAREEXCLUDED.AllwarrantyclaimsmustbemadeinwritinganddeliveredtoPowerexinaccordancewiththeproceduressetforthonitswebsite(www.powerexinc.com),orsuchclaimshallbebarred.Upontimelyreceiptofawarrantyclaim,PowerexshallinspecttheProductclaimedtohaveaDefect,andPowerexshallreplaceanyProductwhichitdeterminestohavehadaDefect;provided,however,thatPowerexmayelect,uponreturnoftheProduct,torefundtobuyeranypartofthepurchasepriceofsuchProductspaidtoPowerex.FreightforreturningProductstoPowerexforinspectionorforshippingwarrantypartsshallbepaidbybuyerwherepermittedbyapplicablelaw.Powerexisnotresponsibleforanyimportfees,taxes,duties,licensesorotherfeesimposedbyanygovernmentalauthorityupontheproduction,sale,shipmentand/oruseofProductscoveredhereunder.ThewarrantiesandremedieshereinarethesoleandexclusiveremedyforanybreachofwarrantyorforanyotherclaimbasedonanyDefect,ornon-performanceoftheProducts,whetherbaseduponcontract,warrantyornegligence.
(d) Coverage. ThewarrantyprovidedhereinappliestoPowerexpipelineproductsonly.
(e) Exceptions. Notwithstandinganythingtothecontraryherein,PowerexshallhavenowarrantyobligationswithrespecttoProducts:
(i)ThathavenotbeeninstalledinaccordancewithPowerex’swrittenspecificationsandinstructions;
(ii)ThathavenotbeenmaintainedinaccordancewithPowerex’swritteninstructions;
(iii) ThathavebeenmateriallymodifiedwithoutthepriorwrittenapprovalofPowerex;or
(iv)Thatexperiencefailuresresultingfromoperation,eitherintentionalorotherwise,inexcessofratedcapacitiesorinanotherwiseimpropermanner.
Thewarranty provided herein shall not apply to: (i) any defects arising from corrosion, abrasion, use of insoluble lubricants, or negligentattendancetoorfaultyoperationoftheProducts;(ii)ordinarywearandtearoftheProducts;or(iii) defects arising from abnormalconditionsoftemperature,dirtorcorrosivematter;(iv)anyOEMcomponentwhichisshippedbyPowerexwiththeoriginalmanufacturer’swarranty,whichshallbethesoleapplicablewarrantyforsuchcomponent.
Limitation of Liability. NOTWITHSTANDINGANYTHINGTOTHECONTRARYHEREIN,TOTHEEXTENTALLOWABLEUNDERAPPLICABLELAW,UNDERNOCIRCUMSTANCESSHALLPOWEREXBELIABLEFORANY INCIDENTAL,CONSEQUENTAL,PUNITIVE,SPECULATIVEORINDIRECTLOSSESORDAMAGESWHATSOEVERARISINGOUTOFORINANYWAYRELATEDTOANYOFTHEPRODUCTSORGOODSSOLDORAGREEDTOBESOLDBYPOWEREXTOBUYER. TOTHEEXTENTALLOWABLEUNDERAPPLICABLELAW,POWEREX’SLIABILITY INALLEVENTSISLIMITEDTO,ANDSHALLNOTEXCEED,THEPURCHASEPRICEPAID.Intheeventofbreachofanywarrantyhereunder,Powerex’ssoleandexclusiveliabilityshallbeatitsoptioneithertorepairortoreplaceanydefectiveproduct,ortoacceptreturn,transportationprepaid,of suchproductand refund thepurchaseprice; ineithercaseprovided thatwrittennoticeof suchdefectisgiventoPowerexwithintwelve(12)monthsfromdateofshipmenttoBuyer,thattheproductisfoundbyPowerextohavebeendefectiveatthetimeofsuchshipment,thattheproducthasbeeninstalledand/oroperatedinaccordancewithPowerex’sinstructions,
Buyer’s purchase order is subject to the following conditions of sale:
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 26 of 28
thatnorepairs,alterationsorreplacementshavebeenmadebyotherswithoutPowerex’swrittenapproval,andthatBuyernotifiesPowerexinwritingwithinfifteen(15)daysafterthedefectbecomesapparentandpromptlyfurnishesfullparticularsinconnectiontherewith;andprovidedfurtherthatinnoeventshalltheaggregateliabilityofPowerexinconnectionwithbreachofanywarrantyorwarrantiesexceedthepurchasepricepaidfortheproductpurchasedhereunder.Powerexmay,atitsoption,requirethereturnofanyproduct,transportationanddutiesprepaid,toestablishanyclaimofdefectmadebyBuyer.Unlessotherwiseagreedinwriting(a) Powerexwillnotacceptandshallhavenoresponsibilityforproductsreturnedwithoutitspriorwrittenconsent,and(b) Powerex willnotassumeanyexpenseorliabilityforrepairstoproductsmadeoutsideofitsplantbythirdparties.IntheeventPowerexelectstoreplaceadefectiveproduct,costsofinstallation,labor,service,andallothercoststoreplacetheproductshallbetheresponsibilityofBuyer.
Powerexshallnot,exceptassetforthabove,beotherwiseliabletoBuyerortoanypersonwhoshallpurchasefromBuyer,oruse,anyproductssuppliedhereunderfordamagesofanykind,including,butnotlimitedto,indirect,specialorconsequentialdamagesorlossofproductionoflossofprofitsresultingfromanycausewhatsoever,including,butnotlimitedto,anydelay,act,errororomissionofPowerex.Supplier’srepairorreplacementofanyProductshallnotextendtheperiodofanywarrantyofanyProduct.
Warranty Disclaimer.PowerexhasmadeadiligentefforttoillustrateanddescribetheProductsinitsliterature,includingitsPriceBook,accurately;however,suchillustrationsanddescriptionsareforthesolepurposeofidentification,anddonotexpressorimplyawarrantythattheProductsaremerchantable,orfitforaparticularpurpose,orthattheProductswillnecessarilyconformtotheillustrationsordescriptions.
Product Suitability.Manyjurisdictionshavecodesandregulationsgoverningsales,construction,installation,and/oruseofProductsforcertainpurposes,whichmayvaryfromthoseinneighboringareas.WhilePowerexattemptstoassurethatitsProductscomplywithsuchcodes,itcannotguaranteecompliance,andcannotberesponsibleforhowtheproductisinstalledorused.BeforepurchaseanduseofaProduct,pleasereviewtheProductapplications,andnationalandlocalcodesandregulations,andbesurethattheProduct,installation,andusewillcomplywiththem.
Claims.Anynon-warrantyclaimspertainingtotheProductsmustbefiledwithPowerexwithin(6)monthsoftheinvoicedate,ortheywillnotbehonored.Prices,discounts,andtermsaresubjecttochangewithoutnoticeorasstipulatedinspecificProductquotations.Powerexshallnotbeliableforanydelayorfailurearisingoutofactsofthepublicenemy,fire,flood,oranydisaster,labortrouble,riotordisorder,delayinthesupplyofmaterialsoranyothercause,whethersimilarordissimilar,beyondthecontrolofCompany.Allshipmentsarecarefullyinspectedandcountedbeforeleavingthefactory.PleaseinspectcarefullyanyreceiptofProductsnotinganydiscrepancyordamageonthecarrier’sfreightbillatthetimeofdelivery.Discrepanciesordamagewhichobviouslyoccurredintransitarethecarrier’sresponsibilityandrelatedclaimsshouldbemadepromptlydirectlytothecarrier.ReturnedProductswillnotbeacceptedwithoutpriorwrittenauthorizationbyPowerexanddeductionsfrominvoicesforshortageordamageclaimswillnotbeallowed.UNLESS OTHERWISE AGREED TO IN WRITING, THE TERMS AND CONDITIONS CONTAINED IN THIS LIMITED WARRANTY WILL CONTROL IN ANY TRANSACTION WITH POWEREX.Anydifferentorconflictingtermsasmayappearonanyorderformnoworlatersubmittedbythebuyerwillnotcontrol.AllordersaresubjecttoacceptancebyPowerex.
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 27 of 28
Notes
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
Installation & Operating Instructions forGenesys® Analog Series Manifolds
Powerex • 150 Production Drive • Harrison, OH 45030 • USA(T) 1.888.769.7979 • (F) 513-367-3125 • www.powerexinc.com
IN595900AV • 02/2017Page 28 of 28
Notes
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________
____________________________________________________________________________________________________________