u.s. food and drug administration device establishment ... establishment registration... · ita-fda...
TRANSCRIPT
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 1
U.S. Food and Drug Administration Device Establishment Registration and Listing
Presenter: Nicole Taylor Smith, Assistant Country DirectorChina Office, U.S. Food and Drug Administration
1
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 2
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListWhat? What are the RequirementsUnited States AgentHow? How to Register and ListFeesResources and Contacts
2
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 3
Regulatory Requirements Section 510 of the Food, Drug, and Cosmetic Act (21 U.S.C. § 360)
Enacted in 1976 (since amended)Required medical device establishments to annually register and list with FDAMisbranded if devices manufactured, prepared, propagated, compounded, or processed in an establishment not duly registered or listed
Food and Drug Administration Amendments Act (FDAAA)Enacted in September 2007Mandated electronic registration and listingRequired user fees for many establishments
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 4
Regulatory RequirementsFood and Drug Administration Safety and Innovation Act (FDASIA)
Enacted July 2012Expanded user fee requirements to all establishment types
21 Code of Federal Regulations (CFR) Part 807, revisedPublished August 2, 2012Regulations promulgated to implement law
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 5
Regulatory Requirements21 CFR 807: Establishment Registration and Device Listing for Manufacturers and Initial Importers of Devices
Subpart A – General Provisions/Definitions
Subpart B – Procedures for Device Establishments
Subpart C – Procedures for Foreign Device Establishments
Subpart D -- Exemptions
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 6
Regulatory RequirementsFailure to Register and List
Section 502(o) of the Food, Drug, and Cosmetic Act (21 U.S.C. § 352(0))Devices are considered misbranded if manufactured, prepared, propagated, compounded, or processed in an establishment not duly registered under section 510 of the Act, 21 U.S.C. § 360, and are not included in a list required by section 510(j) of the Act, 21 U.S.C. § 360(j) (emphasis added).
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 7
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListWhat? What are the RequirementsUnited States AgentHow? How to Register and ListFeesResources and Contacts
7
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 8
Why? Purpose of Registration and Listing
Establishment registration and listing provides FDA with the location of medical device establishments and a list of devices manufactured at those establishments.Knowing where devices are made increases the ability to prepare for and respond to public health emergencies.
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 9
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListUnited States AgentWhat? What are the RequirementsHow? How to Register and ListFeesResources and Contacts
9
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 10
Who? Who Must Register and List
DomesticManufacturers/Remanufacturers
• Kit AssemblersRepackagers/RelabelersContract Manufacturers/SterilizersSpecification DevelopersReprocessors of Sing Use DevicesComplaint HandlersInitial Importers (Initial Distributors)
Note: domestic distributors need not register
ForeignManufacturers/Remanufacturers
• Kit AssemblersRepackagers/RelabelersContract Manufacturers/SterilizersSpecification DevelopersReprocessors of Sing Use DevicesComplaint HandlersForeign Exporters and Private Label Distributors
Note: applies only if devices are sold in the United States
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 11
Who? Who Must Register and List
Establishment Type Domestic ForeignManufacturers/Remanufacturers/ Kit Assemblers
Yes Yes
Specification Developer Yes Yes
Contract Manufacturer/Sterilizer Yes Yes
Repackagers/Relabelers Yes Yes
Reprocessors of Single Use Devices Yes Yes
Complaint Handlers Yes Yes
Initial Importers Yes No
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 12
Who? Who Must Register and List
Exemptions – Not Required to Register/ListA manufacturer of raw materials or components to be used in the manufacture or assembly of a deviceA manufacturer of devices used solely for veterinary purposes
• Regulated by FDA’s Center for Veterinary MedicineLicensed practitioners who manufacture or otherwise alter devices solely for use in their practice
• Exemption is void if distributed to other practitionersRetail establishments that provide devices directly to end users
• For example, pharmacies, surgical supply outlets, etc.• Note: establishments receiving goods from foreign establishments should make
sure their shipments are properly labeledPersons who manufacture, prepare, propagate, compound, or process devices solely for use in research, teaching, or analysis and do not introduce such devices into commercial distribution
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 13
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListWhat? What are the RequirementsUnited States AgentHow? How to Register and ListFeesResources and Contacts
13
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 14
When? When to register and listDomestic establishments
Within 30 days of putting a device into commercial distribution
Foreign establishmentsPrior to exporting to the United States for the first time
Initial ImportersPrior to importing to the United States for the first timeOnly need to register; do not need to listMust identify the manufacturer of each device imported
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 15
When? When to register and list
Annual Registration and Listing CertificationEstablishments must certify their registration information is complete and accurate
AnnuallyBetween October 1 through December 31 of each yearMake changes as necessary
Information may be updated at any timeStrongly recommend if plans change, i.e. plans to discontinue, resume or begin marketing/distributing a device
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 16
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListWhat? What are the Requirements United States AgentHow? How to Register and ListFeesResources and Contacts
16
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 17
What? What are the Requirements - Registration
All establishments (domestic/foreign) must:1. Pay the annual registration fee2. Register the establishment in the FDA Unified
Registration and Listing Systems (FURLS)3. Provide information about the establishment,
Owner Operator and Official Correspondent (name, address, phone, email, etc.)
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 18
What? What are the Requirements - Registration
All establishments (domestic/foreign) must:4. Create at least one listing at time of initial
registration5. Identify all proprietary names under which
the product is marketed in the United States
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 19
What? What are the Requirements - Registration
Foreign Establishments must:1. Identify all persons you know of who
import or who offer to import your product into the United States
2. Identify a United States Agent
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 20
What? What are the Requirements - Registration
Initial Importers must:1. Identify manufactures of the product you
import by establishment name, address, establishment registration number or device listing number
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 21
What? What are the Requirements – Listing
Device ListingMost establishments that are required to register with FDA are also required to list devices made at the establishment and the activities performed on the devices (Exception: Initial Importers/Distributors are not required to list)Foreign establishments must list device before it can be imported into the United StatesManufacturer or Specification Developer must list device first; then contract manufacturer or sterilizer may listFor combination products, identify the type of the combination (i.e., drug/device, etc.)
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 22
What? What are the Requirements – ListingInformation needed to list
For exempt/pre-amendment devices:• Product code• Activities performed at establishments for the device (i.e., manufacturing,
relabel, etc.)• Proprietary or Brand name of the marketed device (this information may be
marked confidential)
For devices that required FDA premarket submission:• Submission number from your clearance/approval letter (KXXXXXX; or
PXXXXXX)• Activities performed at establishment for the device (i.e., manufacturing,
relabel, etc.)• Proprietary or Brand name of the marketed device (this information may be
marked confidential)
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 23
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListWhat? What are the RequirementsUnited States AgentHow? How to Register and ListFeesResources and Contacts
23
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 24
United States AgentAll Foreign Establishments must have a United States AgentRoles of Agent:
Assist in communication with FDAMay be granted authority by foreign establishment to act as official correspondentReceive official FDA information or documentsRespond to questions concerning products being imported or offered for import
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 25
United States AgentResponsibilities
Information about a foreign establishment’s U.S. agent is submitted electronically using the FURLS system and is part of the establishment registration process Each foreign establishment may designate only one U.S. agent Agent must reside or have a place of business in the United States with someone to answer the phone during normal business hours; Post Office Box (P.O. Box) is not sufficient; and cannot just have an answering service Agent does not have responsibility to report adverse events or submit 510(k)s
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 26
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListWhat? What are the Requirements United States AgentHow? How to Register and ListFeesResources and Contacts
26
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 27
How? How to Register and List
How to register and listFor all establishments (domestic/foreign) you must register electronically using FDA’s web-based Registration and Listing System (FURLS) (unless waiver is granted by FDA)
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 28
How? How to Register and List
FURLSTwo types of accounts:
• Owner Operator; • Official Correspondent
Passwords: must be reset every 90 daysNote: do not create a new FURLS account if the establishment was registered with FDA in the past, unless directed by FDA
FDA Industry Systems / FDA Unified Registration and Listing Systems (FURLS)
https://www.access.fda.gov/oaa/logonFlow.htm?execution=e3s1
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 29
How? How to Register and List
FURLSOwner Operator Account
• “Master,” “Enterprise,” or “Primary Account”• Assigned to corporation, subsidiary, etc.• Creates and updates all FURLS accounts• Creates, updates and deactivates registrations and listings
Official Correspondent Account• Assigned by Owner Operator• Responsible for annual registration and device listing• Creates new registration and listings• Makes changes, updates• Cannot change the Owner Operator or Official Correspondent
information (i.e., name, address, telephone, email, etc.)
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 30
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListWhat? What are the Requirements United States AgentHow? How to Register and ListFeesResources and Contacts
30
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 31
FeesAll Establishments required to register must first pay the annual Establishment Registration Fee
Pay fee by accessing your user fee account on the Device Facility User Fee Website(DFUF) (not FURLS)
Must complete payment of user fees on DFUF first before completing electronic registration and listing on FURLS
Device Facility User FeeWebsite (DFUF)
https://userfees.fda.gov/OA_HTML/furls.jsp
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 32
FeesAfter you pay the annual registration user fee on DFUF, you will receive by e-mail:
PIN (Payment Identification Number)PCN (Payment Confirmation Number)
You will use your PIN and PCN to complete your FURLS registration
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 33
FeesAnnual Registration User Fees
Year FY 2013 FY 2014 FY 2015 FY 2016 FY 2017
Fee $2,575 $3,200 $3,750 $3,872 $3,872
Payment MethodsElectronic Payments (i.e. credit card, ACH electronic checks)By mail – paper check drawn on U.S. bank check in U.S. currencyWire Transfers
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 34
OverviewRegulatory Requirements for Device Establishment Registration and ListingWhy? Purpose of Registration and ListingWho? Who Must Register and ListWhen? When to Register and ListWhat? What are the Requirements United States AgentHow? How to Register and ListFeesResources and Contacts
34
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 35
Resources and ContactsReleasable establishment registration and listing information under the Freedom of Information Act is available by searching the Establishment Registration and Listing databaseThis database includes:
medical device establishments registered with FDA, andmedical devices listed with FDA
Updated weekly, usually every Mondayhttp://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRL/rl.cfm
Reminder: Registration of a device establishment, assignment of a registration number, or listing of a medical device does not in any way denote approval of the of the establishment or its products by FDA
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 36
Resources and ContactsFDA’s website:
Registration and Listing: www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/RegistrationandListing/default.htmWho Must Register, List and Pay Fee: www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/RegistrationandListing/ucm053165.htm
Contact Information: CDRH Registration and Listing Helpdesk (including FURLS/DRLM): at [email protected], 301-796-7400Assistance with policy questions and import detention issues: [email protected]
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 37
Quick RecapAnnual Registration
Registration information must be submitted each year between October 1 and December 31, even if no changes have occurred. Listing information must be reviewed each year between October 1 and December 31, at the same time you review your registration information. Submit any updates at that time.
Initial RegistrationSubmit registration and/or listing information within 30 days of an establishment beginning an activity or putting a device into commercial distribution; foreign establishments must register before exporting products to the United States, and domestic importers must register before importing products. Reminder that if your device requires premarket notification clearance or approval, you will have to wait until your premarket submission [510(k), PMA, etc.] is cleared or approved to register your establishment and list the device.
Update Registration & Listing InformationCan access FURLS at any time throughout the year to update changes to registration and listing informationEstablishment registrations are based on FDA’s fiscal year which runs from October 1 to September 30. However, FDA will continue to consider an establishment’s registration active through the end of each calendar year.
Reminders:→ Complete of annual establishment registration fee on Device Facility User Fee Website (DFUF) first before completing
electronic registration and listing on FURLS→ Additional information is available at: www.fda.gov
ITA-FDA Medical Devices Regulatory Capacity Building Training Program for International Medical Devices Regulators March 27 - 28, 2014; San Francisco, California 38
Questions??