UNIQUE SELECTIVITY OF ZIRCHROM COLUMNS IN SEPARATIONS OF POLAR AND NON-POLAR MIXTURES Hanlin Li*, Shawn B. Jones, William J. Meyerhoffer Analytical Research & Development, Pfizer Global Research & Development, Groton/New London Laboratories, Pfizer Inc, Groton, CT 06340 © Pfizer 2004 • CS • GNL • 61410 ABSTRACT Efficient reversed-phase liquid chromatography (RP-HPLC) separations of pharmaceutical mixtures containing both highly polar and non-polar analytes are typically achieved by gradient elutions. In this work, we found that the unique selectivity observed in both ZirChrom-CARB and ZirChrom-MS columns enabled moderate retention of both polar acidic and non-polar compounds. Rapid isocratic methods were successfully developed for the separations of the highly polar acidic compound, namely amoxicillin, acidic β-lactam antibiotics, and the non-polar ester prodrug. More than 10 other polar columns such as polar-embedded columns and phenyl columns were also evaluated and compared. Possible separation mechanisms of ZirChrom-CARB and ZirChrom-MS columns were proposed and potential challenges were also discussed. β-lactamase inhibitors are weak antibotics per se, but they are often used in combination with a real antibiotic agent to protect it from bacterial hydrolysis, such as amoxicillin. An acidic β-lactamase inhibitor (BLI) was selected as the active pharmaceutical ingredient (API), and an ester prodrug was also made to improve bioavailability. Therefore, developing an analytical method capable of analyzing all three components is required. Both amoxicillin and the active BLI are acidic polar compounds, and they elute right at or very close to the solvent front on most reversed-phase HPLC columns. The European Pharmacopoeia (EP) method for amoxicillin uses a 25-minute linear gradient from 92% aqueous to 80% aqueous, due to its polar characteristics. However, this method is not suitable in our application, as the non-polar prodrug needs to be eluted within a reasonable time as well. Almost all separations of polar and non-polar mixtures rely on gradient elution. Recent advances in column technologies allow the gradient to start from 100% aqueous to increase polar components retentions. However, isocratic methods are preferred wherever it is possible because of the fast run time and robustness in method transfer. Various column types were evaluated in this study, including polar-embedded columns, phenyl columns and zirconia columns, etc. Unique selectivity was observed in both ZirChrom-MS and ZirChrom-CARB columns, where both polar and non-polar analytes could be eluted isocratically. O H NH 2 O N H O N S COOH Amoxicillin β-lactam antibotic N O R2 R3 R1 COOH β-lactamase inhibitor (Active BLI) N O R2 R3 R1 O O R4 β-lactamase inhibitor (Prodrug, ester) O H NH 2 O N H O N S COOH Amoxicillin β-lactam antibotic N O R2 R3 R1 COOH β-lactamase inhibitor (Active BLI) N O R2 R3 R1 O O R4 β-lactamase inhibitor (Prodrug, ester) INTRODUCTION β-lactam antibotics are recognized with their unique structural feature – the presence of the four-member 2-azetidinone ring. N O R2 R1 R3 Experiments Two gradient programs were performed on each column. With the data from gradient runs and the assistance of DryLab software, the possibility of running an isocratic method was predicted. Chromatographic conditions: System: Waters Alliance 2695 with 2996 PDA Mobile phases: A: 5/95/0.1 ACN/water/phosphoric acid (pH ~ 2) B: 70/30/0.1 ACN/water/phosphoric acid (pH ~ 2) Flow rate: 1.0 mL/min Column temp.: 35°C Detection: 230 nm DryLab 2000 plus: LC Resources Inc. Sample: Mixture of amoxicillin, active BLI and ester prodrug at 1.5 mg/mL, 1.0 mg/mL and 0.1 mg/mL, respectively. Sample diluent: 40/60/0.1 ACN/water/phosphoric acid (pH ~ 2) RESULTS AND DISCUSSION Columns Screened Column Functional Group Embedded Polar Group Bonded Phase Chemistry Mode of Interaction Manufacturer XTerra Phenyl Phenyl N/A Polar, aromatic Waters FluoroSep RP Phenyl Phenyl N/A Polar, aromatic, pi-pi electron interaction ES Industries SymmetryShield RP18 ODS, C18 Carbamate Hydrophobic, polar Waters SymmetryShield RP8 Octyl, C8 Carbamate Medium hydrophobic, polar Waters Polaris C18-A ODS, C18 Amide Hydrophobic, polar MetaChem Discovery RP Amide 16 C16 Amide Hydrophobic, polar Supelco XTerra RP8 Octyl, C8 Carbamate Medium hydrophobic, polar Waters AquaSep Octyl, C8 Ether Medium hydrophobic, polar ES Industries YMC Basic Octyl, C8 N/A Medium hydrophobic Waters ProTec RP18 ODS, C18 Amide* Hydrophobic, polar, with little anion- exchange ES Industries Polarity RP18 Unknown Carbamate Hydrophobic, polar Waters Discovery Zr- CARB Zirconia-Carbon N/A Dispersive interactions (spatial resolution); charge induced interaction (polar) Supelco ZirChrom-MS Zirconia-EDTPA-PBD N/A Mixed mode (RP and ion-exchange) ZirChrom Separations Si O R1 R2 F F F F F Si O R1 R2 O Si R1 R2 O C O H N (CH2)17CH3 O Si R1 R2 O C O H N (CH2)7CH3 O Si R1 R2 H N C O (CH2)17CH3 O Si R1 R2 H N C O (CH2)14CH3 O Si R1 R2 O C O H N (CH2)7CH3 O Si R1 R2 O (CH2)7CH3 O Si R1 (CH2)7CH3 R2 O Si R1 R2 H N C O (CH2)17CH3 O Si R1 R2 O C O H N (CH2)17CH3 Typical Chromatograms with Polar-Embedded Column Amoxicillin appeared to be more polar than active BLI on typical polar-embedded columns. Isocratic elution is not feasible based on DryLab prediction Observations: Both amoxicillin and active BLI were hardly retained under isocratic condition, while the non-polar drug had a lot of retention. It was confirmed that an isocratic condition is not applicable on typical polar-embedded columns. Symmetry Shield RP18, 4.6 x 150 mm, 3.5 um Gradient elution: 0-100% B in 15 min 0 10 20 30 40 60 70 80 90 100 %B Rs 26.6 Amoxicillin Active BLI Ester Prodrug Resolution map predicted by DryLab 1500 1000 500 0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 Symmetry Shield RP18, 4.6 x 150 mm, 3.5 um Gradient elution: 0-100% B in 15 min 0 10 20 30 40 60 70 80 90 100 %B Rs 26.6 Amoxicillin Active BLI Ester Prodrug Resolution map predicted by DryLab 1500 1000 500 0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 Amoxicillin Active BLI Ester Prodrug SymmetryShield RP18, Isocratic elution: 35/65/0.1 ACN/water/phosphoric acid 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 1000 0 Amoxicillin Active BLI Ester Prodrug SymmetryShield RP18, Isocratic elution: 35/65/0.1 ACN/water/phosphoric acid 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 1000 0 ZirChrom CARB Column Note the behavior change of amoxicillin on ZirChrom CARB column. Isocratic elution is possible based on DryLab prediction. Note: The elution order of the three components is different from typical polar-embedded columns, where amoxicillin had even more retention than the ester prodrug. Active BLI Amoxicillin Ester Prodrug 0 10 20 30 40 60 70 80 90 100 %B Rs 33.9 Typical gradient chromatogram: 0-100% B in 15 min Resolution map predicted by DryLab 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 200 150 100 50 Active BLI Amoxicillin Ester Prodrug 0 10 20 30 40 60 70 80 90 100 %B Rs 33.9 Rs 33.9 Typical gradient chromatogram: 0-100% B in 15 min Resolution map predicted by DryLab 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 200 150 100 50 0 100 200 300 400 500 600 700 800 900 1000 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Active BLI Ester Prodrg Amoxicillin Isocratic elution: 70/30/0.1 (v/v/v) Acetonitrile/water/phosphoric acid 0 100 200 300 400 500 600 700 800 900 1000 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Active BLI Ester Prodrg Amoxicillin 0 100 200 300 400 500 600 700 800 900 1000 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Active BLI Ester Prodrg Amoxicillin Isocratic elution: 70/30/0.1 (v/v/v) Acetonitrile/water/phosphoric acid ZirChrom MS Column Observations: Elution order different from both ZirChrom-CARB and typical polar-embedded columns was observed. Isocratic elution is possible on ZirChrom-MS column. Active BLI Amoxicillin Ester Prodrug 0 10 20 30 40 60 70 80 90 100 %B Rs 49.8 Typical gradient chromatogram: 0-100% B in 15 min Resolution map predicted by DryLab 0.0 1.0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 1000 500 Active BLI Amoxicillin Ester Prodrug 0 10 20 30 40 60 70 80 90 100 %B Rs 49.8 Typical gradient chromatogram: 0-100% B in 15 min Resolution map predicted by DryLab 0.0 1.0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 1000 500 0 500 1000 1500 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5 Active BLI Amoxicillin Ester Prodrug Isocratic elution: 30/70/0.1 (v/v/v) ACN/water/phosphoric acid 0 500 1000 1500 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5 Active BLI Amoxicillin Ester Prodrug 0 500 1000 1500 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5 Active BLI Amoxicillin Ester Prodrug Isocratic elution: 30/70/0.1 (v/v/v) ACN/water/phosphoric acid Possible Separation Mechanisms ZirChrom-CARB: ZirChrom-CARB column uses zirconia as the substrate. The non-polar stationary phase is pyrolitic carbon coated onto the solid support. At low pHs (<4), the column shows net positive charge; therefore, mix-mode separation mechanism-hydrophobic interaction and anion-exchange was in effect. In addition, the column also retains analytes through electronic (π-π) interactions*, which might account for the strong retention observed on amoxicillin over the active BLI. ZirChrom-MS: ZirChrom-MS phase is Lewis acid site deactivated. It shows net zero charge at low pHs and appears to have net-negative charge at pH 5 and above. Therefore, it offers mix-mode separation mechanism and unique selectivity for basic compounds. Interestingly, it also shows different selectivity from both ODS and ZirChrom-CARB columns. It is much less retentive for the non-polar compound compared to ODS column in this application, but also provides reasonable retention for some acidic analyte (e.g., amoxicllin). * T.P. Weber, P.T. Jackson and P.W. Carr, Analytical Chemistry, 1995 (67), 3042-3050 Potential Challenges ZirChrom-CARB: Method robustness is often a challenge for mix-mode columns. The same issue is also revealed from the DryLab resolution map, as small changes in B% would generate large changes in both resolution and run time. Reverse of elution order was also predicted by DryLab, which was later confirmed with experiments (as seen below). Similar phenomena were also observed on another mix-mode column – ProTec RP18. ZirChrom-MS: ZirChrom-MS column exhibits better method robustness for this particular application compared with the CARB column, which was consistent with the DryLab prediction. As for neutral and acidic compounds at low pHs, no ion-exchange mechanism was present. Active BLI Amoxicillin Ester Prodrug 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Active BLI Ester Prodrug Amoxicillin Isocratic elution: 70/30/0.1 ACN/water/H 3 PO 4 Isocratic elution: 60/40/0.1 ACN/water/H 3 PO 4 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 1000 800 600 400 200 0 1000 500 0 Active BLI Amoxicillin Ester Prodrug 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Active BLI Ester Prodrug Amoxicillin Isocratic elution: 70/30/0.1 ACN/water/H 3 PO 4 Isocratic elution: 60/40/0.1 ACN/water/H 3 PO 4 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 1000 800 600 400 200 0 1000 500 0 CONCLUSION Rapid isocratic methods were developed for a mixture of polar and non-polar compounds on both ZirChrom-CARB and ZirChrom-MS columns, where similar separation could only be achieved by gradient elution on conventional ODS or polar-embedded columns. Different retention mechanisms were exhibited for the two columns. At mix-mode separation, retention of amoxicillin was very sensitive to changes in mobile phase composition. Better method robustness was observed on ZirChrom-MS column. ACKNOWLEDGMENT ZirChrom Separations Inc. for technical assistance.

Upload: others

Post on 30-Sep-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: UNIQUE SELECTIVITY OF ZIRCHROM COLUMNS IN … · UNIQUE SELECTIVITY OF ZIRCHROM COLUMNS IN SEPARATIONS OF POLAR AND NON-POLAR MIXTURES Hanlin Li*, Shawn B. Jones, William J. Meyerhoffer

UNIQUE SELECTIVITY OF ZIRCHROM COLUMNS IN SEPARATIONS OF POLAR AND NON-POLAR MIXTURESHanlin Li*, Shawn B. Jones, William J. MeyerhofferAnalytical Research & Development, Pfizer Global Research & Development, Groton/New London Laboratories, Pfizer Inc, Groton, CT 06340

© Pfizer 2004 • CS • GNL • 61410

ABSTRACTEfficient reversed-phase liquid chromatography (RP-HPLC) separations of pharmaceutical mixtures containing both highly polar and non-polar analytes are typically achieved by gradient elutions. In this work, we found that the unique selectivity observed in both ZirChrom-CARB and ZirChrom-MS columns enabled moderate retention of both polar acidic and non-polar compounds. Rapid isocratic methods were successfully developed for the separations of the highly polar acidic compound, namely amoxicillin, acidic β-lactam antibiotics, and the non-polar ester prodrug. More than 10 other polar columns such as polar-embedded columns and phenyl columns were also evaluated and compared. Possible separation mechanisms of ZirChrom-CARB and ZirChrom-MS columns were proposed and potential challenges were also discussed.

β-lactamase inhibitors are weak antibotics per se, but they are often used in combination with a real antibiotic agent to protect it from bacterial hydrolysis, such as amoxicillin. An acidic β-lactamase inhibitor (BLI) was selected as the active pharmaceutical ingredient (API), and an ester prodrug was also made to improve bioavailability. Therefore, developing an analytical method capable of analyzing all three components is required.

Both amoxicillin and the active BLI are acidic polar compounds, and they elute right at or very close to the solvent front on most reversed-phase HPLC columns.

The European Pharmacopoeia (EP) method for amoxicillin uses a 25-minute linear gradient from 92% aqueous to 80% aqueous, due to its polar characteristics. However, this method is not suitable in our application, as the non-polar prodrug needs to be eluted within a reasonable time as well.

Almost all separations of polar and non-polar mixtures rely on gradient elution. Recent advances in column technologies allow the gradient to start from 100% aqueous to increase polar components retentions. However, isocratic methods are preferred wherever it is possible because of the fast run time and robustness in method transfer.

Various column types were evaluated in this study, including polar-embedded columns, phenyl columns and zirconia columns, etc. Unique selectivity was observed in both ZirChrom-MS and ZirChrom-CARB columns, where both polar and non-polar analytes could be eluted isocratically.

OH

NH2

O

NH

ON

S

COOH

Amoxicillin

β-lactam antibotic

NO

R2R3

R1

COOH

β-lactamase inhibitor

(Active BLI)

NO

R2R3

R1

O OR4

β-lactamase inhibitor

(Prodrug, ester)

OH

NH2

O

NH

ON

S

COOH

Amoxicillin

β-lactam antibotic

NO

R2R3

R1

COOH

β-lactamase inhibitor

(Active BLI)

NO

R2R3

R1

O OR4

β-lactamase inhibitor

(Prodrug, ester)

INTRODUCTIONβ-lactam antibotics are recognized with their unique structural feature – the presence of the four-member 2-azetidinone ring. N

O

R2

R1

R3

Experiments

Two gradient programs were performed on each column. With the data from gradient runs and the assistance of DryLab software, the possibility of running an isocratic method was predicted.

Chromatographic conditions:

Ø System: Waters Alliance 2695 with 2996 PDA

Ø Mobile phases: A: 5/95/0.1 ACN/water/phosphoric acid (pH ~ 2)

B: 70/30/0.1 ACN/water/phosphoric acid (pH ~ 2)

Ø Flow rate: 1.0 mL/min

Ø Column temp.: 35°C

Ø Detection: 230 nm

DryLab 2000 plus: LC Resources Inc.

Sample: Mixture of amoxicillin, active BLI and ester prodrug at 1.5 mg/mL, 1.0 mg/mL and 0.1 mg/mL, respectively.

Sample diluent: 40/60/0.1 ACN/water/phosphoric acid (pH ~ 2)

RESULTS AND DISCUSSION

Columns Screened

Column Functional Group Embedded Polar Group

Bonded Phase Chemistry Mode of Interaction Manufacturer

XTerra Phenyl Phenyl N/A

Polar, aromatic Waters

FluoroSep RP Phenyl

Phenyl N/A

Polar, aromatic, pi-pi electron interaction ES Industries

SymmetryShield RP18

ODS, C18 Carbamate

Hydrophobic, polar Waters

SymmetryShield RP8

Octyl, C8 Carbamate

Medium hydrophobic, polar Waters

Polaris C18-A ODS, C18 Amide Hydrophobic, polar MetaChem

Discovery RP Amide 16

C16 Amide Hydrophobic, polar Supelco

XTerra RP8 Octyl, C8 Carbamate

Medium hydrophobic, polar Waters

AquaSep Octyl, C8 Ether Medium hydrophobic, polar ES Industries

YMC Basic Octyl, C8 N/A Medium hydrophobic Waters

ProTec RP18 ODS, C18 Amide* Hydrophobic, polar, with little anion-exchange

ES Industries

Polarity RP18 Unknown Carbamate Hydrophobic, polar Waters

Discovery Zr-CARB

Zirconia-Carbon N/A Dispersive interactions (spatial resolution); charge induced interaction (polar)

Supelco

ZirChrom-MS Zirconia-EDTPA-PBD N/A Mixed mode (RP and ion-exchange) ZirChrom Separations

SiO

R1

R2

FF

F F

FSiO

R1

R2

O Si

R1

R2

O

C

O

HN

(CH2)17CH3

O Si

R1

R2

O

C

O

HN

(CH2)7CH3

O Si

R1

R2

HN

C

O

(CH2)17CH3

O Si

R1

R2

HN

C

O

(CH2)14CH3

O Si

R1

R2

O

C

O

HN

(CH2)7CH3

O Si

R1

R2

O

(CH2)7CH3

O Si

R1

(CH2)7CH3

R2

O Si

R1

R2

HN

C

O

(CH2)17CH3

O Si

R1

R2

O

C

O

HN

(CH2)17CH3

Typical Chromatograms with Polar-Embedded Column

Ø Amoxicillin appeared to be more polar than active BLI on typical polar-embedded columns.

Ø Isocratic elution is not feasible based on DryLab prediction

Observations: Both amoxicillin and active BLI were hardly retained under isocratic condition, while the non-polar drug had a lot of retention. It was confirmed that an isocratic condition is not applicable on typical polar-embedded columns.

Symmetry Shield RP18, 4.6 x 150 mm, 3.5 umGradient elution: 0-100% B in 15 min

0 10 20 30 40 60 70 80 90 100%B

Rs

26.6Amoxicillin

Active BLIEster Prodrug

Resolution map predicted by DryLab1500

1000

500

0

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0

Symmetry Shield RP18, 4.6 x 150 mm, 3.5 umGradient elution: 0-100% B in 15 min

0 10 20 30 40 60 70 80 90 100%B

Rs

26.6Amoxicillin

Active BLIEster Prodrug

Resolution map predicted by DryLab1500

1000

500

0

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0

Amoxicillin

Active BLIEster Prodrug

SymmetryShield RP18, Isocratic elution: 35/65/0.1 ACN/water/phosphoric acid

0 2 4 6 8 10 12 14 16 18 20 22 24 26 28

1000

0

Amoxicillin

Active BLIEster Prodrug

SymmetryShield RP18, Isocratic elution: 35/65/0.1 ACN/water/phosphoric acid

0 2 4 6 8 10 12 14 16 18 20 22 24 26 28

1000

0

ZirChrom CARB Column

Ø Note the behavior change of amoxicillin on ZirChrom CARB column.

Ø Isocratic elution is possible based on DryLab prediction.

Note: The elution order of the three components is different from typical polar-embedded columns, where amoxicillin had even more retention than the ester prodrug.

Active BLI Amoxicillin

Ester Prodrug

0 10 20 30 40 60 70 80 90 100%B

Rs

33.9Typical gradient chromatogram: 0-100% B in 15 min

Resolution map predicted by DryLab

1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0

200

150

100

50

Active BLI Amoxicillin

Ester Prodrug

0 10 20 30 40 60 70 80 90 100%B

Rs

33.9

Rs

33.9Typical gradient chromatogram: 0-100% B in 15 min

Resolution map predicted by DryLab

1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0

200

150

100

50

0

1 0 0

2 0 0

3 0 0

4 0 0

5 0 0

6 0 0

7 0 0

8 0 0

9 0 0

1 0 0 0

0 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 1 7 1 8 1 9

Active BLI

Ester Prodrg

Amoxicillin

Isocratic elution: 70/30/0.1 (v/v/v) Acetonitrile/water/phosphoric acid

0

1 0 0

2 0 0

3 0 0

4 0 0

5 0 0

6 0 0

7 0 0

8 0 0

9 0 0

1 0 0 0

0 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 1 7 1 8 1 9

Active BLI

Ester Prodrg

Amoxicillin

0

1 0 0

2 0 0

3 0 0

4 0 0

5 0 0

6 0 0

7 0 0

8 0 0

9 0 0

1 0 0 0

0 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 1 7 1 8 1 9

Active BLI

Ester Prodrg

Amoxicillin

Isocratic elution: 70/30/0.1 (v/v/v) Acetonitrile/water/phosphoric acid

ZirChrom MS Column

Observations: Elution order different from both ZirChrom-CARB and typical polar-embedded columns was observed. Isocratic elution is possible on ZirChrom-MS column.

Active BLI

Amoxicillin

Ester Prodrug

0 10 20 30 40 60 70 80 90 100%B

Rs

49.8

Typical gradient chromatogram: 0-100% B in 15 min

Resolution map predicted by DryLab

0.0 1.0

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0

1000

500

Active BLI

Amoxicillin

Ester Prodrug

0 10 20 30 40 60 70 80 90 100%B

Rs

49.8

Typical gradient chromatogram: 0-100% B in 15 min

Resolution map predicted by DryLab

0.0 1.0

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0

1000

500

0

500

1000

1500

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5

Active BLI

Amoxicillin

Ester Prodrug

Isocratic elution: 30/70/0.1 (v/v/v) ACN/water/phosphoric acid

0

500

1000

1500

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5

Active BLI

Amoxicillin

Ester Prodrug

0

500

1000

1500

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5

Active BLI

Amoxicillin

Ester Prodrug

Isocratic elution: 30/70/0.1 (v/v/v) ACN/water/phosphoric acid

Possible Separation Mechanisms

Ø ZirChrom-CARB:

ZirChrom-CARB column uses zirconia as the substrate. The non-polar stationary phase is pyrolitic carbon coated onto the solid support. At low pHs (<4), the column shows net positive charge; therefore, mix-mode separationmechanism-hydrophobic interaction and anion-exchange was in effect. In addition, the column also retains analytes through electronic (π-π) interactions*, which might account for the strong retention observed on amoxicillin over the active BLI.

Ø ZirChrom-MS:

ZirChrom-MS phase is Lewis acid site deactivated. It shows net zero charge at low pHs and appears to have net-negative charge at pH 5 and above. Therefore, it offers mix-mode separation mechanism and unique selectivity for basic compounds. Interestingly, it also shows different selectivity from both ODS and ZirChrom-CARB columns. It is much less retentive for the non-polar compound compared to ODS column in this application, but also provides reasonable retention for some acidic analyte (e.g., amoxicllin).

* T.P. Weber, P.T. Jackson and P.W. Carr, Analytical Chemistry, 1995 (67), 3042-3050

Potential Challenges

Ø ZirChrom-CARB:

Method robustness is often a challenge for mix-mode columns. The same issue is also revealed from the DryLab resolution map, as small changes in B% would generate large changes in both resolution and run time. Reverse of elution order was also predicted by DryLab, which was later confirmed with experiments (as seen below). Similar phenomena were also observed on another mix-mode column – ProTec RP18.

Ø ZirChrom-MS:

ZirChrom-MS column exhibits better method robustness for this particular application compared with the CARB column, which was consistent with the DryLab prediction. As for neutral and acidic compounds at low pHs, no ion-exchange mechanism was present.

Active BLI

Amoxicillin

Ester Prodrug

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19

Active BLI

Ester Prodrug

AmoxicillinIsocratic elution: 70/30/0.1 ACN/water/H3PO4

Isocratic elution: 60/40/0.1 ACN/water/H3PO4

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19

1000

800

600

400

200

0

1000

500

0

Active BLI

Amoxicillin

Ester Prodrug

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19

Active BLI

Ester Prodrug

AmoxicillinIsocratic elution: 70/30/0.1 ACN/water/H3PO4

Isocratic elution: 60/40/0.1 ACN/water/H3PO4

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19

1000

800

600

400

200

0

1000

500

0

CONCLUSIONRapid isocratic methods were developed for a mixture of polar and non-polar compounds on both ZirChrom-CARB and ZirChrom-MS columns, where similar separation could only be achieved by gradient elution on conventional ODS or polar-embedded columns. Different retention mechanisms were exhibited for the two columns. At mix-mode separation, retention of amoxicillin was very sensitive to changes in mobile phase composition. Better method robustness was observed on ZirChrom-MS column.

ACKNOWLEDGMENTZirChrom Separations Inc. for technical assistance.