Uncertainty of Measurement in the Medical Laboratory Florent Vanstapel MD PhD Clinical Chemistry – POCT Coordinator Laboratory Medicine UZ Leuven – KULeuven – Belgium CAT-Series Leuven April 23 2015

Upload: vandieu

Post on 26-Jun-2018

228 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Uncertainty of Measurementin the Medical Laboratory

Florent Vanstapel MD PhDClinical Chemistry – POCT Coordinator

Laboratory MedicineUZ Leuven – KULeuven – Belgium

CAT-Series – Leuven April 23 2015

Page 2: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 1

Everything you wanted to know about UM - norm-compliance? - clinical relevance?

But were afraid to ask- is it difficult? - do you get payed for it?

Page 3: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 2

Relevant Standards

- EU IVD Regulation - ISO 15189 - Technical standards

Page 4: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

EU DIRECTIVE 98/79/EC (IVD Directive)

F. Vanstapel - Measurement Uncertainty 3

What the industry is required to do

Art. 1. 3. For the purposes of this Directive, calibration and control materials refer to any substance, material or article intended by their manufacturer either to establish measurement relationshipsor to verify the performance characteristics of a device in conjunction with the intended use ...

Annex III. 3. adequate performance evaluation data showing the performances claimed … and supported by a reference measurement system, with information on the reference methods, the reference materials, the known reference values, the accuracy and measurement units used; such data should originate from studies in a clinical or other appropriate environment or result from relevant biographical references,

Page 5: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

ISO-15189:2014

F. Vanstapel - Measurement Uncertainty 4

What the laboratory is required to do

5.3.1.4 Equipment calibration and metrological traceabilityb) metrological traceability of the standard and calibration c) verifying measurement accuracy … at defined intervals;traceability shall be to a reference material or procedure

5.5.1.2 Verification of examination proceduresapplies to Validated (= established) examination procedures … independent verification by laboratory … through objective evidence 5.5.1.3 Validation of examination proceduresvalidate a) non-standard methods; b) laboratory developed methods; c) standard methods used outside intended scope; d) validated methods subsequently modified. NOTE Performance characteristics include: trueness, accuracy, precision; uncertainty, analytical specificity, …

Page 6: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 5

ISO-15189:2014

5.5.1.4 Measurement uncertainty of measured quantity valuesshall … regularly review estimates of measurement uncertainty

NOTE 1 The relevant components … actual measurement process, commencing with the presentation of the sample to the measurement procedure and ending with the output of the measured value.

NOTE 2 Measurement uncertainties may be … obtained by measurement of quality control materials under intermediate conditions that include e.g. changes of reagent and calibrator batches, different operators, scheduled instrument maintenance.

Upon request, the lab shall make its estimates of measurement uncertainty available to laboratory users.

Internal Quality Control: ANOVA

Very narrowly defined

Page 7: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 6

Technical Standards

Eurachem: The Fitness for Purpose of Analytical MethodsLaboratory Guide to Method Validation & Related Topics

6.5 METHOD DEVELOPMENT 6.6 THE DIFFERENT PERFORMANCE CHARACTERISTICS OF A METHOD6.6.4 Trueness6.6.5 Precision6.6.6 Measurement uncertainty

TrainMic: EC JRC IRMM Training in Metrology in Chemistry

3. Single Laboratory Validation of Measurement Procedures4. Uncertainty of Measurement Principles5. Uncertainty of Measurement: Approaches to Evaluation

How things are done

Page 8: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Message #1

F. Vanstapel - Measurement Uncertainty 7/7

Standards don’t ask for much neither from us or from the industry

- Method verification vs validation- Trueness / Traceability / Commutability - Budget of Measurement Uncertaintycan be derived from internal Quality Control Data

- …

Page 9: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Casus:Typical Method Implementation

F. Vanstapel - Measurement Uncertainty 8

Validation File:Fit for the Intended Use?

Page 10: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

0

1

2

3

4

0 1 2 3 4

Alt

ern

ativ

e Te

st

Comparator Method Is this a good test?

Method Validation: Traceability to Comparator

Method Comparison

Allowed Total Error (dashed line)= 31% of Spread in Reference Population

(Tonks criterium)

Analytical Imprecision (shaded area) < 1/3 of Total Error ~ 1/2 of a 6 σ process

What was the Question: Does this make it a test Fit for the Intended Use?

F. Vanstapel - Measurement Uncertainty 9

Intercept = 0; Slope = 1

Page 11: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

The Raw Data

Clinical Validation: ROC-curve

0.00.10.20.30.40.50.60.70.80.91.0

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

P(x

<X|D

is-)

P(x<X|Dis+)

0 1 2 3 4

Measurand (level)

AvgDis+ at 3 SDDis- = Definition of Limit of Detection

AvgDis+ - AvgDis-

= 3 SDDis-

F. Vanstapel - Measurement Uncertainty 10

Does this ROC-curve make it a test fit for purpose?Does this make it a test unfit for purpose?

Page 12: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Black – Gamble – White

0 1 2 3 4

Measurand (level)

F. Vanstapel - Measurement Uncertainty 11

Loss of Truthdue to misclassification

Dichotomic Analysis

T-

T+

1

-1

D-

-1

1

D+

Kappa-weights

Page 13: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Black – Gamble – White

0 1 2 3 4

Measurand (level)

F. Vanstapel - Measurement Uncertainty 12

Dichotomic Analysis

Loss of Certainty due toUM

Biological

AnalyticalTotal

Uncertainty Budget

Biological Spread

Gai

ns

Loss

es

Simulator

Page 14: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

How a Test Result is Interpreted

F. Vanstapel - Measurement Uncertainty 13

Black – Grey – White

0 1 2 3 4

Measurand (level)

Differential Diagnosis

- Don’t miss what needs urgent intervention

You need not to confirm doubt but to exclude Dis- or Dis+

- Workup list of common causes

- But also exclude rare causes

Page 15: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Black – Undecided – White= losses

0 1 2 3 4

Measurand (level)

F. Vanstapel - Measurement Uncertainty 14

Dichotomic Analysis

If you refuse to Gamble

predictive value increases trade-in: sensitivity

& specificity

CAVE: grey zone no decision = also a decision

Page 16: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Where can the validation exercise create value ?

Cost of Test

Unavoidable

Quality Failure

(Grey Zone)

Avoidable Failures

Created Value

Cost of Quality Failure

Benefit

Normal

Production

Costs

Actual

Production

Costs

Substandard UM

Cause-Effect Analysis

Test Utilization

Biological Overlap

Unavoidable Uncert. of Measurement

F. Vanstapel - Measurement Uncertainty 15

Page 17: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Message #2

F. Vanstapel - Measurement Uncertainty 16/16

Typically, Analytical Uncertainty has little effect on Diagnostic Power

UM doesn’t ruin a good test & can’t rescue a bad test

Diagnostic Value depends on whether you ordered a test with Power of Exclusion

Page 18: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 17

Everything you wanted to know about UM - norm-compliance? - clinical relevance?

But were afraid to ask- is it difficult? - do you get payed for it?

Page 19: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 18

Casus: UM properly

The Nuts and Bolts of UM Evaluation

Page 20: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 19

Inventory of Sources of Error

UM Evaluation starts with

Page 21: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Risk Analysis Generic Laboratory Process

F. Vanstapel - Measurement Uncertainty 20

Sample

Reagents

Instruments

Controls

Calibrators

collect

condition

obtain

obtain

obtain

install

transport

store

store

store

maintain

condition

aliquot

condition

formulate

condition

formulate

condition

formulate

setup

calibrate measure

Operators train

report

transmission

generation

transcription

USETest

order

LIS

After ISO/TS 22367:2008 = Workshop Approach= Fishbone analysis

Page 22: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 21

Risk Analysis: Failure-Mode analysis: What can cause relevant failures?

Right Test at the Right Time … for the Right Patient …

Pre-analytical

Biological Variation

Pulsitility, Diurnal & Seasonal Variation

Physiological (Starvation, Exercise, …) Variation, TimeStamp

Specimen Collection

Posture

Stasis, Hemolysis, Filling of the tube

Right patient, Correct labelling, sample, recipient, …

Analytical = Measurement ProcessSample reception & processing

Right identification of primary and secondary sample

Completeness of coagulation, …

Micro cloths: obstruction needle, light scattering, …

Membrane ghosts and fragments, …

Uncertainty of Measurement

Bias, Specificity, …

Calibration, Linearity, …

Inter-batch random error, …

Other sources of (random) imprecision

Equipment faults (aspiration, carry-over, sporadic faults, ..)

Post-analytical

Reporting for the right patient

Transcription errors , …, …

Data transfer errors, ….

Test Ordering

Adequacy, sampling details , specimen quality, …, …

Adequacy cut-offs, reference ranges, …

Adequacy interpretation support

Interpretation

Clasical Focus of

Method ValidationIs this where

most relevant

failures originate?

Process Approachof Diagnostic Cycle

Page 23: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Message #3

F. Vanstapel - Measurement Uncertainty 22/22

The Core Analytical Processand the associated UMshould not be your only focus, maybe not even your primary focus?

Page 24: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 23

Approach #1Inventory of Sources of Analytical Error

UM Evaluation (in the narrow sense)

Page 25: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Fishbone Analysis of an Analytical Assay

F. Vanstapel - Measurement Uncertainty 24

e.g. Calibrator formulation

Linearity Linearity

Volume

Temperature

Calibration

Repeatability

Purity

Massa

C(compound)

Readability

Repeatability

M(tare)

Calibration

M(gross)

Calibration

Readability

RepeatabilitySensitivity Sensitivity

Barometric PressureTemperature Humidity

Page 26: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Message #4

F. Vanstapel - Measurement Uncertainty 25/25

Fishbone Analysis identifies- Input Variables

= Model Equation (see next section) - & Influence Variables

typically not represented in the modeldid your supplier develop a rugged test?

Page 27: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Approach #2

Model Equation -> Error Propagation: Sensitivity or Ruggedness Analysis

F. Vanstapel - Measurement Uncertainty 26

Page 28: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

R

f

m

V

b

aAw di

What are the critical issues ? m/V: Unfavourable? fdi: Unfavourable dilution in the assay?R: Incomplete and haphazard extraction? (A-a)/b: Unfavourable dynamic Range A vs a; Low response factor A/b? a: Constructed Blank or Extrapolated from Calibration Line ?b: Historical Colour Yield or In Run Calibration ?

Sample Mass extracted in Volume

Dilution in Assay/ RecoveryBlank

Response/ Sensitivity

Result

Measurement

Example: Extraction & Calibration

F. Vanstapel - Measurement Uncertainty 27

Page 29: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Message #5

F. Vanstapel - Measurement Uncertainty 28

Writing down the Model Equation & critically inspecting it tells you a lot

Page 30: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

For independent parameters, the second term is absent

Covariance: uiujrijSensitivity AnalysisConcentrate on what matters = ruggedness

Partial derivatives ? Don’t panick : mostly additions/subtractions & multiplications/divisions

Pythagoras : destructive accumulation of noise

F. Vanstapel - Measurement Uncertainty 29

The combined standard uncertainty = reasonable dispersion of measurement results = function of the uncertainty of the component quantities

N N-1 N

u2(y(x1,x2,…)) = ∑ δy/δxi ui2 +2 Σ Σ (δy/δxi) (δy/δxj) ci,j

i =1 i=1 j=i+1

First order Taylor Series approximation of the process

Page 31: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 30

Kragten (or spreadsheat) Method:

X = A*B/C

The simplest of cases

X: concentration resulting from A: dilution of the sampleB: measurementC: calibration factor

Page 32: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 31

Kragten Method: Propagation of error

A

B

C

a

b

c

b

c

a + ua

b + ub

c

a

b

c + uc

a

(xa – x)2 (xb – x)2 (xc – x)2ux,n2

ux,n2∑ u2

Evaluation

X = A*B/C

Model Effect of partial disturbances

ua ub uc

x xa xb xc

Page 33: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 32

Propagation of error: An ExampleA sample is diluted in a reagent mix. At the end point of the reaction the color yield is read, and divided by a calibration factor. That result is multiplied with the dilution factor.

= (x3/x4)(x1+x2)/x1

Sample volumeReagens volumeResponse ReadingResponse Factor

x1

x2

x3

x4

52000.7

2000

359µLµL

cm-1

106.L.

mol-1

.cm-1

mMSI

4%2%

0.05%0.02%

0.1%ui/xi

from pipet control programfrom pipet control programfrom equipment specsEpistemic variable (literature)

yValue

66%16%

16%

2%contribution to u2

Sample Volume

Reagens Volume

Response Reading

Response FactorSimulator

Page 34: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 33

The analysis tells you what to focus on

Message #6

Don’t be scared of a Taylor Series

Page 35: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Approach #3a

Measurement Uncertainty From Internal Quality Data

F. Vanstapel - Measurement Uncertainty 34

Page 36: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Internal Quality Control Data

F. Vanstapel - Measurement Uncertainty 35

-6

-5

-4

-3

-2

-1

0

1

2

3

4

5

6

0

32

64

96

12

8

16

0

19

2

22

4

25

6

28

8

32

0

35

2

38

4

41

6

44

8

48

0

51

2

54

4

57

6

60

8

64

0

67

2

70

4

73

6

76

8

80

0

83

2

86

4

89

6

92

8

96

0

99

2

10

24

z-s

co

re

Do we have a problem?

Typical immunometric assay

Simulator

There are episodes where results consistentlydiffer from the majority of results?

Page 37: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

ANOVA of Internal Quality Control Data

F. Vanstapel - Measurement Uncertainty 36

#Componentof variation CV Comments: Typical sources of variability

1 Cal. Nominal Uncert. 1.0% Applies to each new calibrator batch2 Cal Manipulations 1.0% Applies to manipulations and conditioning 3 Between Reagent 1 1.4% Applies to start-up of new reagent batch4 Within Reagent 1 0.7% Applies to conditioning of individual reagent packs5 Between Reagent 2 1.4% Idem as 16 Within Reagent 2 0.7% Idem as 17 Maintenance 1.0% Maintenance 8 Start-Up procedure 0.0% Start-Up procedure9 Operator 0.0% Operator dependent variability

10 Between Run 5.0% Applies to long-term variability of the analytical system 10 Within Run 5.0% Applies to short-term variabilityof the analytical system

12 iQC Manipulations 1.0% Applies to manipulations and conditioning

Total UM 10% Uncertainty of measurement

- The effects of calibrations, lots, ... persist- Certain events are synchronous

Page 38: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Message #7

F. Vanstapel - Measurement Uncertainty 37

The example was a simulation:nothing abnormal was happeningrandom variation within expectations

ANOVA of iQC data = UM Budget helps you to avoid panick & futile interventions

Page 39: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Approach #3b

Measurement AccuracyFrom Proficiency Testing Data

F. Vanstapel - Measurement Uncertainty 38

Page 40: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

EQAS commutabilitysingle manufacturer? not OK

Belgian Proficiency Testing Scheme: Chemistry Calcium

F. Vanstapel - Measurement Uncertainty 39

Page 41: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

ISO 17511: Hierarchical scheme of traceability

SI-UnitDefinition of the measurand

Certification of RM purity

Primary Calibrator

Primary reference measurement procedure

Secondary CalibratorSecondary reference

measurement procedure

Manufaturer’sMaster Calibrator

Manufacturer’s Standing Procedure

Bureau International de Poids et de MesuresGeneral Conference on Weights and Measures

National Metrology Institute

Accredited Reference Measurement Lab

Manufacturers Laboratory

Manufaturer’sProduct Calibrator

End-user’sRoutine Procedure

Do you feel traceable?Do you feel certain? IFCC - Clin Chem Lab Med 2012

Gravimetrically

IDMS

Reference method

Routine method

MaybeRather uncertain

End User’s Laboratory

F. Vanstapel - Measurement Uncertainty 40

Page 42: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

Message #8

F. Vanstapel - Measurement Uncertainty 41

Hence “Proficiency” Schemes:- document non-commutability

of methods or of materials?

The design of proficiency testing is flawed:- at best they measure “relative” bias - differences may or may not be

attributable to matrix effects

The perversion of “Matrix Effects”: - makes it harder for the lab to switch methods

Page 43: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

F. Vanstapel - Measurement Uncertainty 42

Who is paying for this?Reaping the Fruits

Summary

Page 44: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

The Economy of your Quality System

F. Vanstapel - Measurement Uncertainty 43

Risk

Scrap what is redundantRetain what is relevantManage what is critical

Process Care

UM = Failure Mode Analysis

Process

Reduction inQuality Failures

Reduction in Production Costs

Increased Quality / Cost Ratio

mitigate risk secure supply chain

Clean Lean Process

Page 45: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

CONCLUSIONS

F. Vanstapel - Measurement Uncertainty 44

Risk-Mitigation = Process-Care Tool

UM = Failure-Mode Analysis= Risk Analysis Tool

Process-Care = Quality at No Cost

Page 46: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

CONCLUSIONS

F. Vanstapel - Measurement Uncertainty 45

Production Department: - to focus on what matters- not to spend resources on futile projects

Understanding UM allows

Interfacing with Prescribers: - analytical precision suffices- transmural commutability is a problem

- invest recuperated opportunity costs intest ordering and interpretation support

Page 47: Uncertainty of Measurement in the Medical Laboratory · Uncertainty of Measurement in the Medical Laboratory ... 5.5.1.4 Measurement uncertainty of measured quantity values ... Temperature

TO DO

F. Vanstapel - Measurement Uncertainty 46

Commercialize Commutable Calibrators:to realize longitudinal coordinated transmural care

Develop Rugged Methods:to reduce waste activities (iQC, etc.)