Ultrasound Physics
Reflections &
Attenuation
‘97
![Page 2: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/2.jpg)
Perpendicular Incidence
Sound beam travels perpendicular to boundary between two media 90o
IncidentAngle
1
2Boundarybetweenmedia
![Page 3: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/3.jpg)
Oblique Incidence
Sound beam travel not perpendicular to boundary
ObliqueIncidentAngle
(not equal to 90o)
1
2
Boundarybetweenmedia
![Page 4: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/4.jpg)
Perpendicular Incidence
What happens to sound at boundary?reflected
sound returns toward source
transmitted sound continues in
same direction
1
2
![Page 5: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/5.jpg)
Perpendicular Incidence
Fraction of intensity reflected depends on acoustic impedances of two media
1
2
Acoustic Impedance =Density X Speed of Sound
![Page 6: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/6.jpg)
Intensity Reflection Coefficient (IRC)&Intensity Transmission Coefficient (ITC)
IRCFraction of sound intensity
reflected at interface<1
ITCFraction of sound intensity
transmitted through interface<1
Medium 1
Medium 2IRC + ITC = 1
![Page 7: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/7.jpg)
IRC Equation
Z1 is acoustic impedance of medium #1
Z2 is acoustic impedance of medium #2
2 reflected intensity z2 - z1
IRC = ------------------------ = ----------
incident intensity z2 + z1
For perpendicular incidence
Medium 1
Medium 2
![Page 8: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/8.jpg)
Reflections
Impedances equal no reflection
Impedances similar little reflected
Impedances very different virtually all reflected
2 reflected intensity z2 - z1
Fraction Reflected = ------------------------ = ----------
incident intensity z2 + z1
![Page 9: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/9.jpg)
Why Use Gel?
Acoustic Impedance of air & soft tissue very different
Without gel virtually no sound penetrates skin
2 reflected intensity z2 - z1
IRC = ------------------------ = ----------
incident intensity z2 + z1
Acoustic Impedance
(rayls)
Air 400Soft Tissue 1,630,000
Fraction Reflected: 0.9995
![Page 10: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/10.jpg)
Rayleigh Scattering
redirection of sound in many directionscaused by rough surface with respect to
wavelength of sound
![Page 11: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/11.jpg)
Diffuse Scattering & Rough Surfaces
heterogeneous mediacellular tissueparticle suspension
blood, for example
![Page 12: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/12.jpg)
ScatteringOccurs if
boundary not smoothRoughness related to frequency
frequency changes wavelength higher frequency shortens wavelength shorter wavelength “roughens” surface
![Page 13: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/13.jpg)
Specular Reflections
Un-scattered soundoccurs with smooth
boundariessimilar to light reflection
from mirroropposite of scatter from
rough surfacewall is example of rough
surface
![Page 14: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/14.jpg)
Backscatter
sound scattered back in the direction of source
![Page 15: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/15.jpg)
Backscatter Comments
Caused byrough surfacesheterogeneous media
Depends on scatterer’ssizeroughnessshapeorientation
Depends on sound frequencyaffects wavelength
![Page 16: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/16.jpg)
Backscatter Intensity
normally << than specular reflections
angle dependancespecular reflection very angle dependentbackscatter not angle dependent
echo reception not dependent on incident angle
increasing frequency effectively roughens surfacehigher frequency results in more backscatter
![Page 17: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/17.jpg)
PZT is Most Common Piezoelectric Material
Lead Zirconate TitanateAdvantages
Efficient More electrical energy transferred to sound & vice-
versaHigh natural resonance frequencyRepeatable characteristics
Stable designDisadvantages
High acoustic impedance Can cause poor acoustic coupling Requires matching layer to compensate
![Page 18: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/18.jpg)
Resonant FrequencyFrequency of Highest Sustained
IntensityTransducer’s “preferred” or resonantresonant
frequencyExamples
Guitar StringBell
![Page 19: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/19.jpg)
Operating Frequency
Determined bypropagation speed of transducer
material typically 4-6 mm/sec
thickness of element
prop. speed of element (mm /sec)oper. freq. (MHz) = ------------------------------------------------ 2 X thickness (mm)
![Page 20: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/20.jpg)
Pulse Mode Ultrasoundtransducer driven by short voltage
pulsesshort sound pulses producedLike plucking guitar string
Pulse repetition frequency same as frequency of applied voltage pulsesdetermined by the instrument (scanner)
![Page 21: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/21.jpg)
Pulse Duration Review
typically 2-3 cycles per pulseTransducer tends to continue ringing
minimized by dampeningdampening transducer element
Pulse Duration = Period X Cycles / Pulse
![Page 22: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/22.jpg)
Damping MaterialGoal:
reduce cycles / pulseMethod:
dampen out vibrations after voltage pulse
Constructionmixture of powder & plastic or
epoxyattached to near face of
piezoelectric element (away from patient)
DampingMaterial
PiezoelectricElement
![Page 23: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/23.jpg)
Disadvantages of Damping
reduces beam intensityproduces less pure frequency (tone)
![Page 24: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/24.jpg)
Bandwidth
Damping shortens pulsesthe shorter the pulse, the higher the range
of frequencies Range of frequencies produced called
bandwidthbandwidth
![Page 25: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/25.jpg)
Bandwidthrange of frequencies present in an
ultrasound pulse
Frequency
Intensity
Ideal
Frequency
Intensity
Actual
Bandwidth
OperatingFrequency
![Page 26: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/26.jpg)
operating frequencyQuality Factor = ----------------------------- bandwidth
Quality Factor (“Q”)
UnitlessQuantitative Measure of
“Spectral Purity”
Frequency
Intensity
Actual
Bandwidth
![Page 27: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/27.jpg)
Damping
More damping results inshorter pulsesmore frequencieshigher bandwidth lower quality factor lower intensity
Rule of thumb for short pulses (2 - 3 cycles)
quality factor ~ number of cycles per pulse
![Page 28: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/28.jpg)
Transducer Matching LayerTransducer element has different acoustic
impedance than skinMatching layer reduces reflections at surface
of piezoelectric elementIncreases sound energy transmitted into body
Transducer – skin interface
![Page 29: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/29.jpg)
Transducer Matching Layerplaced on face of transducerimpedance between that of
transducer & tissuereduces reflections at surface of
piezoelectric elementCreates several small transitions in acoustic
impedance rather than one large one
reflected intensity z2 - z1
IRC = ------------------------ = ----------
incident intensity z2 + z1
( )2 Matching
Layer
![Page 30: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/30.jpg)
Transducer ArraysVirtually all commercial transducers
are arraysMultiple small elements in single
housingAllows sound beam to be electronically
FocusedSteeredShaped
![Page 31: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/31.jpg)
Electronic Scanning
Transducer ArraysMultiple small transducersActivated in groups
![Page 32: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/32.jpg)
Electrical ScanningPerformed with transducer arraysarrays
multiple elements inside transducer assembly arranged in either a line (linear array)
concentric circles (annular array)
Curvilinear Array Linear Array
![Page 33: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/33.jpg)
Linear Array Scanning
Two techniques for activating groups of linear transducers Switched ArraysSwitched Arrays
activate all elements in group at same time Phased ArraysPhased Arrays
Activate group elements at slightly different times impose timing delays between activations of elements in
group
![Page 34: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/34.jpg)
Linear Switched ArraysElements energized as
groupsgroup acts like one large
transducerGroups moved up & down
through elementssame effect as manually
translatingvery fast scanning possible
(several times per second) results in real time image
![Page 35: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/35.jpg)
Linear Switched Arrays
![Page 36: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/36.jpg)
Linear Phased ArrayGroups of elements energized
same as with switched arrays
voltage pulse applied to all elements of a group
BUTelements not all pulsed at
same time
1
2
![Page 37: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/37.jpg)
Linear Phased Arraytiming variations allow beam
to be shapedsteeredfocused
Above arrows indicate timing variations.By activating bottom element first & top last, beam directed upward
Beam steered upward
![Page 38: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/38.jpg)
Linear Phased Array
Above arrows indicate timing variations.By activating top element first & bottom last, beam directed downward
Beam steered downward
By changing timing variations between pulses, beam can be scanned from top to bottom
![Page 39: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/39.jpg)
Linear Phased Array
Above arrows indicate timing variations.By activating top & bottom elements earlier than center ones, beam is focused
Beam is focused
Focus
![Page 40: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/40.jpg)
Linear Phased ArrayFocus
Focal point can be moved toward or away from transducer by altering timing variations between outer elements & center
![Page 41: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/41.jpg)
Linear Phased ArrayFocus
Multiple focal zones accomplished by changing timing variations between pulses•Multiple pulses required•slows frame rate
![Page 42: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/42.jpg)
Listening ModeListening direction can be steered
& focused similarly to beam generationappropriate timing variations
applied to echoes received by various elements of a group
Dynamic Focusinglistening focus depth can be
changed electronically between pulses by applying timing variations as above
2
![Page 43: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/43.jpg)
1.5 Transducer~3 elements in elevation directionAll 3 elements can be combined for thick slice1 element can be selected for thin slice
Elevation
Direction
![Page 44: Ultrasound Physics Reflections & Attenuation ‘97](https://reader033.vdocuments.mx/reader033/viewer/2022061611/56649f465503460f94c68bed/html5/thumbnails/44.jpg)
1.5 & 2D TransducersMultiple elements in 2 directionsCan be steered & focused anywhere in 3D
volume