ultrasensitive carbon nanotube-based biosensors using antibody-binding fragments

6
Analytical Biochemistry 381 (2008) 193–198 0003-2697/$ - see front matter © 2008 Elsevier Inc. All rights reserved. doi:10.1016/j.ab.2008.06.040 Contents lists available at ScienceDirect Analytical Biochemistry journal homepage: www.elsevier.com/locate/yabio Most biological sensing techniques depend mainly on optical detection principles. These methods are highly sensitive and spe- cific, although they suffer from several disadvantages such as their inherent compleity and requirement for multiple reagents and steps, signal amplification, a relatively large sample size, and com- ple data analysis [1–3]. Recently, several researchers have stud- ied electronic biodetection methods to overcome these problems [4–6]. One eample is carbon nanotube (CNT) 1 -based biosensors that take advantage of the eotic properties of CNTs [7–9]. Because two-thirds of as-synthesized single-walled CNTs ehibited semi- conducting properties, one can build CNT-based field effect transis- tors (FETs) whose conductance changes sensitively by the charge transfer from molecules adsorbed onto the CNT surface [10–12]. In one of the biodetection schemes, the CNT–FET surface is func- tionalized with specific receptor molecules that bind to desired tar- get biomolecules. When the target molecules bind to the receptor molecules in solution, the charges of the target molecules affect the conductance of the CNT–FETs. Thus, one can detect specific tar- get molecules electrically in real time by monitoring the conduc- tance of CNT–FETs. In buffer solution, the double layer is formed within the range of Debye length (»3 nm in 10 mM buffer solution) [13,14] around the CNT–FETs, and the target and receptor biomol- ecules should fit in the double layer to change the conductance of CNT–FETs. However, the dimension of the antibodies (»10–12 nm) used as receptors is usually much larger than the Debye length, implying that the target molecules cannot approach the CNT–FETs within the double layer. One strategy to overcome this problem is using small receptors such as aptamers [15,16]. Although aptamers, which are artificial nucleic acid ligands, are very small (»2 nm) and show high selectivity, specificity, and affinity for amino acids, drugs, proteins, and other small molecules [17,18], they have not yet been developed for many diseases and they can also suffer from pleomorphism. Here we suggest a novel method of reducing the receptor size on CNT–FET biosensors using antibody-binding frag- ments [F(ab9) 2 , Fab] for the antigen–antibody immune reaction. Antibodies are immune system-related proteins called immuno- globulins [19]. Each antibody monomer has a molecular weight of approimately 150,000 Da (150 kDa) and is composed of two identi- cal heavy polypeptide chains and two identical light chains joined to form a Y-shaped molecule covalently bonded via interchain disulfide (S–S) linkages between cysteine residues. Moreover, these monomers are arranged in three discrete domains: two Fab fragments at the tips of the “Y” and one Fc at its pole. The enzyme papain can be used to cleave an immunoglobulin monomer into two Fab fragments and an Fc fragment [20,21]. The enzyme pepsin Ultrasensitive carbon nanotube-based biosensors using antibody-binding fragments Jun Pyo Kim a , Byung Yang Lee b , Seunghun Hong b , Sang Jun Sim a, * a Deparmen of Chemcal Engneerng, Sungkyunkwan Unversy, Changan-gu, Suwon 440-746, Souh Korea b School of Physcs and Asronomy, Seoul Naonal Unversy, Gwanak-gu, Seoul 151-742, Souh Korea article info abstract Arcle hsory: Received 10 March 2008 Available online 9 July 2008 We report a method to build ultrasensitive carbon nanotube-based biosensors using immune binding reaction. Here carbon nanotube–field effect transistors (CNT–FETs) were functionalized with antibody- binding fragments as a receptor, and the binding event of target immunoglobulin G (IgG) onto the frag- ments was detected by monitoring the gating effect caused by the charges of the target IgG. Because the biosensors were used in buffer solution, it was crucial to use small-size receptors so that the charged target IgG could approach the CNT surface within the Debye length distance to give a large gating effect. The results show that CNT–FET biosensors using whole antibody had very low sensitivity (detection limit »1000 ng/ml), whereas those based on small Fab fragments could detect 1 pg/ml (»7 fM level). Moreover, our Fab-modified CNT–FET could successfully block the nontarget proteins and could selectively detect the target protein in an environment similar to that of human serum electrolyte. Significantly, this strat- egy can be applied to general antibody-based detection schemes, and it should enable the production of label-free ultrasensitive electronic biosensors to detect clinically important biomarkers for disease diagnosis. © 2008 Elsevier Inc. All rights reserved. Keywords: Carbon nanotube Field effect transistor Biosensor Antibody-binding fragments [F(ab9) 2 , Fab] Immune reaction * Corresponding author. Fa: +82 31 290 7272. E-mal address: [email protected] (S.J. Sim). 1 Abbrevaons used: CNT, carbon nanotube; FET, field effect transistor; IgG, immunoglobulin G; SWNT, single-walled nanotube; BSA, bovine serum albumin; DI, deionized; OTS, octadecyltrichlorosilane; SA, streptavidin; SAM, self-assembled monolayer; PR, photoresist; PBS, phosphate-buffered saline.

Upload: jun-pyo-kim

Post on 26-Jun-2016

215 views

Category:

Documents


1 download

TRANSCRIPT

Analytical Biochemistry 381 (2008) 193–198

Contents lists available at ScienceDirect

Analytical Biochemistry

journal homepage: www.elsevier.com/ locate /yabio

Ultrasensitive carbon nanotube-based biosensors using antibody-binding

fragments

Jun Pyo Kim a, Byung Yang Lee b, Seunghun Hong b, Sang Jun Sim a,*

a Depart­ment­ of Chemi­cal Engi­neeri­ng, Sungkyunkwan Uni­versi­t­y, Changan-gu, Suwon 440-746, Sout­h Koreab School of Physi­cs and Ast­ronomy, Seoul Nat­i­onal Uni­versi­t­y, Gwanak-gu, Seoul 151-742, Sout­h Korea

a r t i c l e i n f o a b s t r a c t

Art­i­cle hi­st­ory:

Received 10 March 2008

Available online 9 July 2008

We report a method to build ultrasensitive carbon nanotube-based biosensors using immune binding

reaction. Here carbon nanotube–field effect transistors (CNT–FETs) were functionalized with antibody-

binding fragments as a receptor, and the binding event of target immunoglobulin G (IgG) onto the frag-

ments was detected by monitoring the gating effect caused by the charges of the target IgG. Because the

biosensors were used in buffer solution, it was crucial to use small-size receptors so that the charged

target IgG could approach the CNT surface within the Debye length distance to give a large gating effect.

The results show that CNT–FET biosensors using whole antibody had very low sensitivity (detection limit

»1000 ng/ml), whereas those based on small Fab fragments could detect 1 pg/ml (»7 fM level). Moreover,

our Fab-modified CNT–FET could successfully block the nontarget proteins and could selectively detect

the target protein in an environment similar to that of human serum electrolyte. Significantly, this strat-

egy can be applied to general antibody-based detection schemes, and it should enable the production

of label-free ultrasensitive electronic biosensors to detect clinically important biomarkers for disease

diagnosis.

© 2008 Elsevier Inc. All rights reserved.

Keywords:

Carbon nanotube

Field effect transistor

Biosensor

Antibody-binding fragments [F(ab9)2, Fab]

Immune reaction

00

do

de

ci

in

st

pl

ie

[4

th

tw

co

to

tr

ti

ge

m

th

ge

im

DI

m

Most biological sensing techniques depend mainly on optical

tection principles. These methods are highly sensitive and spe-

fic, although they suffer from several disadvantages such as their

herent complex­ity and requirement for multiple reagents and

eps, signal amplification, a relatively large sample size, and com-

ex­ data analysis [1–3]. Recently, several researchers have stud-

d electronic biodetection methods to overcome these problems

–6]. One ex­ample is carbon nanotube (CNT)1-based biosensors

at take advantage of the ex­otic properties of CNTs [7–9]. Because

o-thirds of as-synthesized single-walled CNTs ex­hibited semi-

nducting properties, one can build CNT-based field effect transis-

rs (FETs) whose conductance changes sensitively by the charge

ansfer from molecules adsorbed onto the CNT surface [10–12].

In one of the biodetection schemes, the CNT–FET surface is func-

onalized with specific receptor molecules that bind to desired tar-

t biomolecules. When the target molecules bind to the receptor

olecules in solution, the charges of the target molecules affect

e conductance of the CNT–FETs. Thus, one can detect specific tar-

t molecules electrically in real time by monitoring the conduc-

03-2697/$ - see front matter © 2008 Elsevier Inc. All rights reserved.

i:10.1016/j.ab.2008.06.040

* Corresponding author. Fax­: +82 31 290 7272.

E-mai­l address: [email protected] (S.J. Sim).1 Abbrevi­at­i­ons used: CNT, carbon nanotube; FET, field effect transistor; IgG,

munoglobulin G; SWNT, single-walled nanotube; BSA, bovine serum albumin;

, deionized; OTS, octadecyltrichlorosilane; SA, streptavidin; SAM, self-assembled

onolayer; PR, photoresist; PBS, phosphate-buffered saline.

tance of CNT–FETs. In buffer solution, the double layer is formed

within the range of Debye length (»3 nm in 10 mM buffer solution)

[13,14] around the CNT–FETs, and the target and receptor biomol-

ecules should fit in the double layer to change the conductance of

CNT–FETs. However, the dimension of the antibodies (»10–12 nm)

used as receptors is usually much larger than the Debye length,

implying that the target molecules cannot approach the CNT–FETs

within the double layer. One strategy to overcome this problem is

using small receptors such as aptamers [15,16]. Although aptamers,

which are artificial nucleic acid ligands, are very small (»2 nm)

and show high selectivity, specificity, and affinity for amino acids,

drugs, proteins, and other small molecules [17,18], they have not

yet been developed for many diseases and they can also suffer from

pleomorphism. Here we suggest a novel method of reducing the

receptor size on CNT–FET biosensors using antibody-binding frag-

ments [F(ab9)2, Fab] for the antigen–antibody immune reaction.

Antibodies are immune system-related proteins called immuno-

globulins [19]. Each antibody monomer has a molecular weight of

approx­imately 150,000 Da (150 kDa) and is composed of two identi-

cal heavy polypeptide chains and two identical light chains joined

to form a Y-shaped molecule covalently bonded via interchain

disulfide (S–S) linkages between cysteine residues. Moreover,

these monomers are arranged in three discrete domains: two Fab

fragments at the tips of the “Y” and one Fc at its pole. The enzyme

papain can be used to cleave an immunoglobulin monomer into

two Fab fragments and an Fc fragment [20,21]. The enzyme pepsin

194 Ult­rasensi­t­i­ve carbon nanot­ube-based bi­osensors / J.P. Ki­m et­ al. / Anal. Bi­ochem. 381 (2008) 193–198

cleaves below the hinge region, so an F(ab9)2 fragment and an Fc

fragment are formed. Because it is the Fab fragments that bind to

antigens, the size of an antibody can be reduced substantially by

removing the Fc fragment.

In this article, we report the first successful demonstration of a

CNT–FET biosensor using Fab fragments as receptors. We function-

alized CNT–FETs with three types of anti-human immunoglobulin

G (IgG): (i) whole antibody (Fig. 1A), (ii) F(ab9)2 (Fig. 1B), and (iii)

Fab (Fig. 1C). Nex­t, the response of each CNT–FET was monitored

in real time after the introduction of human IgG at various concen-

trations (100 fg/ml–1000 ng/ml).

Mate­ri­als and me­th­ods

Mat­eri­als

Purified single-walled nanotubes (SWNTs) were purchased from

Carbon Nanotechnologies (USA), and 1-pyrenebutanoic acid succ-

inimidyl ester was obtained from Molecular Probes (USA). Whole

antibody, F(ab9)2 fragments and Fab fragments of goat anti-human

IgG as receptors and human IgG as target biomolecules were pur-

chased from Jackson Immuno Research Laboratories (USA). Bovine

serum albumin (BSA), fibrinogen from human plasma, streptavidin

(SA), and other chemical reagents were supplied by Sigma. Deion-

ized (DI) water, obtained from a water purification system (Human

Corporation, Korea), was used for the preparation of the washing

and buffer solutions.

Fabri­cat­i­on of CNT–FET devi­ces

Methyl-terminated octadecyltrichlorosilane (OTS) self-assem-

bled monolayer (SAM) patterns on silicon ox­ide wafer were gen-

erated by first patterning AZ 5214 photoresist (PR) via standard

photolithography, dipping the wafer in OTS solution (1:500 [v/v] in

hex­ane) for 3 min, and finally removing the PR patterns using ace-

tone. SWNT solution was prepared by dispersing purified SWNTs

in 1,2-dichlorobenzene with ultrasonication for 1 h (concentration

»0.1 mg/ml). Nex­t, the patterned silicon ox­ide wafer was dipped

in the SWNT solution for 10 s, rinsed thoroughly with 1,2-dichloro-

benzene, and then dried with nitrogen gas [22]. This step allowed

SWNTs to be adsorbed selectively onto bare SiO2 regions on the

wafer, while methyl-terminated OTS SAMs blocked nonspecific

adsorption of CNTs. After the assembly of the SWNTs, electrodes

(30-nm Au layer on 10 nm Pd) were fabricated via a standard pho-

tolithography and lift-off process [23]. Finally, we also performed

an additional photolithography process to pattern PR (AZ 5214) to

cover up the electrodes and avoid leakage current from electrodes

in buffer solution.

Preparat­i­on of t­he CNT surface for sensi­ng i­n solut­i­on

For the noncovalent functionalization of the CNT surface, CNT–

FET devices were incubated with 1 mM 1-pyrenebutanoic acid

succinimidyl ester in pure methanol for 1 h at room temperature

followed by rinsing with pure methanol to wash away any ex­cess

reagent [24]. For the covalent immobilization of the receptor pro-

teins on the CNT surface, each CNT–FET device was ex­posed to

20 nM anti-human IgG [whole antibody, F(ab9)2 and Fab fragments]

in phosphate-buffered saline (PBS, pH 7.4) overnight at room tem-

perature, rinsed thoroughly in DI water for 6 h, and then dried with

nitrogen gas. To deactivate and block the ex­cess reactive groups

remaining on the CNT surface, 100 mM ethanolamine was added

onto the channel region of the CNT–FET device and incubated for

30 min [16]. Then the CNT–FET device was rinsed with PBS.

Elect­ri­cal measurement­

The electrical properties of the CNT–FET devices during the

introduction of the target proteins were measured by a source

meter (Keithley 2400, USA) after the devices were immobilized by

a probe station that was able to connect with each source and drain

electrode. A source drain bias of 10 mV was maintained through-

out the measurements of the electrical signal, and the pulse width

was 1 s. The sample solutions containing the target proteins with

increasing concentrations were introduced sequentially into the

channel regions of the CNT–FET devices using sample volumes

of 10 ll. After the measurement of the electrical signal was com-

pleted for each sample, the devices were washed thoroughly with

DI water and dried with nitrogen gas. Then they were reloaded in

the source meter.

Re­sults and di­scussi­on

Fig. 2 shows the CNT–FET devices fabricated via the linker-free

directed assembly method [22]. In this method, the CNT network

patterns were formed directly on a bare silicon ox­ide surface with-

out any linker molecules, and they were used as the channel for

FETs. These devices ex­hibited typical “p-type” characteristics, with

decreased source drain current with positive gate bias. The on/off

ratio was low (»3), just like other network transistors, because the

SWNT network was composed of both semiconducting and metal-

lic CNTs. However, our CNT–FET had significant advantages for sen-

sor applications. First, because our fabrication method did not use

any linker molecules, we could minimize possible signal contam-

ination by linker molecules. Furthermore, during this process, as

SWNTs were adsorbed onto bare silicon ox­ide surface, the surface

became nonpolar and blocked the formation of multiple SWNT

a b c

Whole antibody

F(ab’)2 Fab

Debyelength

Ethanolamine1-pyrenebutanoic acid succinimidyl ester

Fi­g. 1. Schematic diagram of CNT–FETs modified with three types of receptors on CNT surface: (A) immobilization of whole antibody; (B) immobilization of F(ab9)2; (C)

immobilization of Fab.

Ult­rasensi­t­i­ve carbon nanot­ube-based bi­osensors / J.P. Ki­m et­ al. / Anal. Bi­ochem. 381 (2008) 193–198 195

Fi­g. 2. Photographs of CNT–FET device: (A) optical image of a CNT–FET chip with patterned gold electrodes; (B) optical micrograph of a CNT–FET device with gold electrodes

passivated with PR; (C) atomic force microscopy image of channel of CNT–FET device.

layers. This self-limiting mechanism warranted the reproducibility

of our CNT–FET fabrication process [23].

Using the prepared devices, we performed the systematic

study regarding the effect of the receptor size on the sensitivity

of CNT–FET biosensors. First, CNT–FETs were functionalized with

three types of receptors [the whole antibody, F(ab9)2, and Fab from

anti-human IgG] after the formation of an SAM using 1-pyreneb-

utanoic acid succinimidyl ester on the CNT surface (Fig. 1). Nex­t,

the electrical conductance change (G/G0) of the devices on the addi-

tion of human IgG solution with various concentrations (1 fg/ml

to 1000 ng/ml) was monitored. The ex­periments in each condition

were repeated three times.

In any biosensor for detection of target proteins, interaction

between nontarget and target proteins could be a source of noise

leading to aberrances in the final result. Therefore, to investigate

the specificity of the device between the surface materials, nontar-

get proteins such as BSA, fibrinogen, and SA were used on Fab-mod-

ified CNT–FETs. Fig. 3 shows the electrical responses of CNT–FETs

after the introduction of PBS solution, nontarget proteins (BSA,

fibrinogen, SA, each 10 mg/ml), and target protein (10 pg/ml IgG)

onto the IgG Fab-modified CNT–FET. When target protein was intro-

duced on the CNT channel, the electrical signal rapidly decreased.

In contrast, upon addition of PBS buffer as control or nontarget pro-

teins, the devices showed a slight increase in electrical conductance.

The addition of PBS liquid droplets created an additional electrical

path around the CNT network channels and could increase conduc-

tance. Moreover, after the introduction of nontarget proteins into

the CNT–FET channel, electrical conduction increased slightly as

in the case of PBS. These phenomena were also observed in cases

of whole antibody and F(ab)2-modified CNT–FETs. Therefore, these

results indicate that the nonspecific binding of nontarget proteins

was successfully suppressed in the CNT–FET device with the treat-

ment of 100 mM ethanolamine.

Fig. 4 shows the response of the CNT–FETs modified with

whole anti-human IgG when droplets of human IgG (10 ll) at var-

ious concentrations (10–1000 ng/ml) were placed on the devices.

On the addition of 10 to 100 ng/ml human IgG, the devices ex­hib-

ited increased electrical conductance as in the case of PBS. This

result means that the CNT–FETs modified with whole anti-human

IgG could not detect concentrations below 100 ng/ml IgG. How-

ever, the addition of a high-concentration solution (1000 ng/ml

human IgG in PBS) reduced the device conductance. These results

were quite consistent throughout three replicates. Human IgG had

positive charges because its isoelectric point (pI = 8.6 ± 04) [25,26]

was higher than the pH of the PBS solution (pH 7.4). Thus, human

IgG adsorbed onto the CNT surface should decrease the CNT–FET

conductance because it is equivalent to applying positive gate volt-

ages [27–30]. When the CNT–FETs were functionalized with whole

Sample injection

Time (s)0 50 100 150 200

Ele

ctri

cal s

ign

al G

/G0

0.6

0.8

1.0

1.2

BlankPBS bufferBSAFibrinogenStreptavidinIgG

Fi­g. 3. Electrical conductance of CNT–FETs after the introduction of PBS buffer solu-

tion, nontarget proteins (BSA, fibrinogen, streptavidin), and target protein onto the

IgG Fab-modified CNT–FET. “Blank” means electrical conductance where nothing

was introduced on the CNT–FET device. (For interpretation of the references to

color in the legend to this figure, the reader is referred to the Web version of this

article.)

Time (s)

0 50 100 150 200 250 300

Ele

ctri

cal s

ign

al G

/G0

0.8

1.0

1.2

1.4 Blankhuman IgG 10 ng/mlhuman IgG 100 ng/mlhuman IgG 1000 ng/ml

Sampleinjection

Fi­g. 4. Electrical conductance change of the CNT–FETs modified with whole antibod-

ies on the addition of human IgG solution with various concentrations (10–1000 ng/

ml). The arrow indicates the point of IgG injection. (For interpretation of the refer-

ences to color in the legend on this figure, the reader is referred to the Web version

of this article.)

196 Ult­rasensi­t­i­ve carbon nanot­ube-based bi­osensors / J.P. Ki­m et­ al. / Anal. Bi­ochem. 381 (2008) 193–198

anti-human IgG that was much larger than the Debye length, most

of the human IgG molecules could not approach the CNT surface

within the distance of the Debye length to give a gating effect. As

a result, the conductance of CNT–FET increased on the addition of

the solution due to the additional current path unless very high-

concentration human IgG solution (»100 ng/ml) was added. There-

fore, whole antibodies with the size of 10–12 nm are not suitable as

receptors for CNT–FET-based biosensors.

To resolve this problem, we introduced a method using Fab

fragments of the antibody. The immune reactions on CNT–FETs

modified with F(ab9)2 were performed at various concentration of

human IgG ranging from 1 to 5000 ng/ml. Fig. 5A shows the changes

in the conductance due to the immune reaction when F(ab9)2 was

immobilized on the CNT surface. The conductance decreased step-

wise with ex­posure to human IgG at concentrations increasing

from 10 to 1000 ng/ml. However, the analyte could not be deter-

mined at low concentration of 1 ng/ml IgG. At the concentration of

5000 ng/ml IgG, the conductance signal was not much higher than

that of the 1000 ng/ml IgG sample. This could be ex­plained because

almost all binding sites of the anti-human IgG F(ab9)2 on CNT sur-

face were already occupied by human IgG molecules at the concen-

tration of 1000 ng/ml human IgG. Thus, as shown in the inset to

Fig. 5B, the linear dynamic range was from 10 to 1000 ng/ml and

the detectable minimum concentration (10 ng/ml) in the case of

the CNT–FETs modified with F(ab9)2 was ameliorated; it was lower

than that of the CNT–FETs modified with whole antibodies.

The hinge region between the Fc and two Fab domains in

antibodies is ex­tremely flex­ible [31]. As a result, it is difficult

to predict the ex­act conformation of the IgG and its size when

the antibody is adsorbed on a surface. The approx­imate size of

human IgG was determined by computational simulation [32]

and small-angle X-ray scattering in solution [33]. The F(ab9)2 of

human IgG consisted of two Fabs, with the unavailable Fc region

(elliptical cylinder shape, height » 7.0 nm) [33] in the antibody

being cut off by pepsin; the vertical length of F(ab9)2 was approx­-

imately 5 nm, which is half the size of the whole antibody. As the

size of the receptors is reduced, the positively charged proteins

can more easily approach the CNT surface within the distance of

the Debye length and affect the conductance of the CNT–FETs.

Thus, the sensitivity in the CNT–FETs modified with F(ab9)2 was

100 times greater than that in the CNT–FETs modified with whole

antibodies. Although the height of F(ab9)2 was only half that of

the antibody, the detectable minimum concentration was still

very high because the horizontal length of F(ab9)2, which con-

[Human IgG] (pg/ml)

0 200 400 600 800 1000 1200

Ele

ctri

cal s

ign

al (

G/G

0)

0.6

0.8

1.0

1.2

B

Log10 [Human IgG concentration] (pg/ml)

1 10 100

Ele

ctri

cal s

ign

al (

G/G

0)

0.5

0.6

0.7

0.8

0.9

1.0

A

Time (s)0 50 100 150 200 250 300

Ele

ctri

cal s

ign

al (

G/G

0)

0.6

0.8

1.0

1.2

1.4 Blankhuman IgG 100 fg/mlhuman IgG 1 pg/mlhuman IgG 10 pg/mlhuman IgG 100 pg/ml

Sampleinjection

Fi­g. 6. (A) Electrical conductance change of the CNT–FETs modified with Fab on

the addition of human IgG solution with various concentrations (100 fg/ml–100 pg/

ml). (B) The calibration curve of IgG detection on the CNT–FETs modified with Fab.

The error bars illustrate the relative standard deviation (RSD) of three replicates.

The inset shows the logarithmic linear range between the electrical signal and the

human IgG concentrations. The arrow indicates the point of IgG injection. (For

interpretation of the references to color in the legend on this figure, the reader is

referred to the Web version of this article.)

A

Time (s)0 50 100 150 200 250 300

Ele

ctri

cal s

ign

al (

G/G

0)

0.8

0.9

1.0

1.1

1.2

1.3

Blankhuman IgG 1 ng/mlhuman IgG 10 ng/mlhuman IgG 100 ng/mlhuman IgG 1000 ng/mlSample

injection

[Human IgG] (ng/ml)0 1000 2000 3000 4000 5000 6000

Ele

ctri

cal s

ign

al (

G/G

0)

0.8

0.9

1.0

1.1

1.2

B

Log10 [Human IgG concentration] (ng/ml)10 100 1000

Ele

ctri

cal s

ign

al (

G/G

0)

0.7

0.8

0.9

1.0

1.1

Fi­g. 5. (A) Electrical conductance change of the CNT–FETs modified with F(ab9)2

upon addition of human IgG solution at various concentrations (1–1000 ng/ml).

(B) The calibration curve of IgG detection on the CNT–FETs modified with F(ab9)2.

The error bars illustrate the relative standard deviation (RSD) of three replicates.

The inset shows the logarithmic linear range between the electrical signal and the

human IgG concentrations. The arrow indicates the point of IgG injection. (For

interpretation of the references to color in the legend on this figure, the reader is

referred to the Web version of this article.)

Ult­rasensi­t­i­ve carbon nanot­ube-based bi­osensors / J.P. Ki­m et­ al. / Anal. Bi­ochem. 381 (2008) 193–198 197

tained a disulfide bond forming a uniform angle (115–130°) [32]

between the Fabs, was very large (»10–11 nm). Therefore, the

size of the receptors should be smaller than F(ab9)2; the CNT–FET

biosensor modified with F(ab9)2 was not applicable in the diagno-

sis of disease.

To obtain better sensitivity for the CNT–FET-based detection,

CNT surface was functionalized with Fab, the smallest unit con-

taining an epitope in the antibody, and their conductance change

was monitored on the addition of human IgG solution with var-

ious concentrations in the range of 100 fg/ml–1000 pg/ml. As

shown in (Fig. 6)A, the conductance value decreased with increas-

ing IgG concentrations. However, the target analyte could not be

detected at concentrations lower than 100 fg/ml or higher than

1000 pg/ml. Therefore, the linear dynamic range of the Fab-mod-

ified CNT–FET biosensor was shown to be from 1 to 100 pg/ml

and the detectable minimum concentration was as low as 1 pg/

ml (7 fM level) of human IgG (Fig. 6B). The improvement in sensi-

tivity can be ex­plained by the small size of Fab. Because the size

of Fab (height »3–5 nm) [33,34] was much smaller than that of

whole antibody or F(ab9)2, it is more likely that the binding event

between the Fab fragments and the target proteins occurred

inside an electrical double layer close to the CNT surface so that

the charges of human IgG protein had a large effect on the con-

ductance of CNT–FETs.

Finally, the CNT–FET device was ex­amined in a mix­ed environ-

ment containing BSA, fibrinogen, and SA to investigate nonspe-

cific absorption and selectivity of the biosensor. Aliquots of these

undesired proteins (1 mg/ml) and PBS buffer (pH 7.4) were mix­ed

together, and then 10 ll of this mix­ture was introduced into the

channel regions to check the nonspecific binding on the CNT–FET

devices. To investigate the selectivity of the assay, the ex­periment

was performed by replacing the volume of PBS buffer with an

equal volume of the target protein (10 pg/ml) in the preparation

described above. As shown in Fig. 7, when the mix­ture solution

without target proteins was injected into the channel of the device,

the electrical signal increased slightly as with only PBS buffer. This

shows that nothing was detected in the mix­ture without target pro-

tein, and the nonspecific adsorption of these proteins was negli-

gible even at very high concentration. Interestingly, the electrical

signal was rapidly reduced in the solution with human IgG. This

result indicates that target protein can be selectively detected at

low levels in the presence of high concentrations of nontarget pro-

teins by the CNT–FET devices.

Conclusi­ons

The Debye length is known to be one of the most important

factors determining the detection limits of CNT–FET biosensors

in solution. Here we devised CNT–FETs that were modified with

Fab fragments as receptors so that the immune-binding reaction

occurred within the Debye length from the CNT surface. In this

way, we were able to lower the detection limit to a protein concen-

tration of 1 pg/ml (»7 fM level) with no labeling of target proteins.

Moreover, our Fab-modified CNT–FET could successfully suppress

the nontarget proteins and could selectively detect the target pro-

tein in a mix­ed environment containing BSA, fibrinogen, SA at high

concentration. This strategy allowed us to build a CNT–FET biosen-

sor system based on the well-established immune reaction of anti-

gen–antibody. The use of small receptors such as Fab fragments

has several advantages in biosensors based on CNT–FET. First, all of

the antibodies developed for disease diagnosis can be used to build

CNT–FET biosensors. In addition, the reduced size of receptors sig-

nificantly improves the detection limit of CNT–FET biosensors by

binding target protein close to the CNT surface. Thus, the strategy

of using Fab fragments in CNT–FET biosensors can be a major break-

through that enables various important applications, for ex­ample,

in proteomics and medical diagnostics.

Acknowle­dgme­nts

This work was supported by Grant 10017190 from the Nex­t Gen-

eration New Technology Development Program of the Ministry of

Knowledge Economy (MKE).

Re­f­e­re­nce­s

[1] P.B. Luppa, L.J. Sokoll, D.W. Chan, Immunosensors: principles and applications to clinical chemistry, Clin. Chim. Acta 314 (2001) 1–26.

[2] M. Wells, Advances in optical detection strategies for reporter signal measure-ments, Curr. Opin. Biotechnol. 17 (2006) 28–33.

[3] M. Seydack, Nanoparticle labels in immunosensing using optical detection methods, Biosens. Bioelectron. 20 (2005) 2454–2469.

[4] Y. Wang, Z. Tang, N.A. Kotov, Bioapplication of nanosemiconductors, Mater. Today 8 (2005) 20–31.

[5] G. Gruner, Carbon nanotube transistors for biosensing applications, Anal. Bio-anal. Chem. 384 (2006) 322–335.

[6] A. Star, J.C.P. Gabriel, K. Bradley, G. Gru1ner, Electronic detection of specific protein binding using nanotube FET devices, Nano Lett. 3 (2003) 459–463.

[7] A. Merkoci, M. Pumera, X. Llopis, B. Perez, M.D. Valle, S. Alegret, New materi-als for electrochemical sensing: VI. Carbon nanotubes, Trends Anal. Chem. 24 (2005) 826–838.

[8] P. Avouris, J. Chen, Nanotube electronics and optoelectronics, Mater. Today 9 (2006) 46–54.

[9] H. Dai, Carbon nanotubes: synthesis, integration, and properties, Acc. Chem. Res. 35 (2002) 1035–1044.

[10] J. Robertson, Realistic applications of CNTs, Mater. Today 7 (2004) 46–52.[11] M. Trojanowicz, Analytical applications of carbon nanotubes: a review, Trends

Anal. Chem. 25 (2006) 480–489.[12] P. Qi, O. Vermesh, M. Grecu, A. Javey, Q. Wang, H. Dai, S. Peng, K.J. Cho, Toward

large arrays of multiplex­ functionalized carbon nanotube sensors for highly sensitive and selective molecular detection, Nano Lett. 3 (2003) 347–351.

[13] R.B.M. Schasfoort, P. Bergveld, Possibilities and limitations of direct detection of protein charges by means of an immunological field-effect transistor, Anal. Chim. Acta 238 (1990) 323–329.

[14] R.B.M. Schasfoort, R.P.H. Kooyman, P. Bergveld, J. Greve, A new approach to immunoFET operation, Biosens. Bioelectron. 5 (1990) 103–124.

[15] H. So, K. Won, Y.H. Kim, B. Kim, B.H. Ryu, P.S. Na, H. Kim, J. Lee, Single-walled carbon nanotube biosensors using aptamers as molecular recognition ele-ments, J. Am. Chem. Soc. 127 (2005) 11906–11907.

[16] K. Maehashi, T. Katsura, K. Kerman, Y. Takamura, K. Matsumoto, E. Tamiya, Lable-free protein biosensor based on aptamer-modified carbon nanotube field-effect transistors, Anal. Chem. 79 (2007) 782–787.

[17] J.F. Lee, G.M. Stovall, A.D. Ellington, Aptamer therapeutics advance, Curr. Opin. Chem. Biol. 10 (2006) 282–289.

[18] W. James, Nucleic acid and polypeptide aptamers: a powerful approach to ligand discovery, Curr. Opin. Pharm. 1 (2001) 540–546.

[19] R.G. Hamilton, D. Abmli, The Human IgG Subclasses, CalBioChem–NovaBio-Chem, Darmstadt, Germany, 2001.

Time (sec)

0 50 100 150 200 250

Ele

ctri

cal s

ign

al G

/G0

0.6

0.8

1.0

1.2

BlankMixtureIgG + Mixture

Sampleinjection

Fi­g. 7. The selectivity of target proteins (human IgG) in mix­ture solution (BSA, fibrin-

ogen, streptavidin) on CNT–FETs modified with Fab.

198 Ult­rasensi­t­i­ve carbon nanot­ube-based bi­osensors / J.P. Ki­m et­ al. / Anal. Bi­ochem. 381 (2008) 193–198

[20] P.C. Ng, Y. Osawa, Preparation and characterization of the Fab and F(ab9)2 frag-ments of an aromatase activity-suppressing monoclonal antibody, Steroids 62 (1997) 776–781.

[21] L.J. Harris, S.B. Larson, K.W. Hasel, J. Day, A. Greenwod, A. McPherson, The three-dimensional structure of an intact monoclonal antibody for canine lym-phoma, Nature 360 (1992) 369–372.

[22] S.G. Rao, L. Huang, W. Setyawan, S. Hong, Large-scale assembly of carbon nano-tubes, Nature 425 (2003) 36–37.

[23] M. Lee, J. Im, B.Y. Lee, S. Myung, J. Kang, L. Huang, Y.K. Kwon, S. Hong, Linker-free directed assembly of high-performance integrated devices based on nano-tubes and nanowires, Nat. Nanotechnol. 1 (2006) 66–71.

[24] R.J. Chen, Y. Zhang, D. Wang, H. Dai, Noncovalent sidewall functionalization of single-walled carbon nanotubes for protein immobilization, J. Am. Chem. Soc. 123 (2001) 3838–3839.

[25] P.A.J. Rosa, A.M. Azevedo, M.R. Aires-Barros, Application of central composite design to the optimisation of aqueous two-phase ex­traction of human antibod-ies, J. Chromatogr. 1141 (2007) 50–60.

[26] R.P. Tracy, R.M. Currie, R.A. Kyle, D.S. Young, Two-dimensional gel electropho-resis of serum specimens from patients with monoclonal gammopathies, Clin. Chem. 28 (1982) 900–907.

[27] R.J. Chen, S. Bangsaruntip, K.A. Drouvalakis, N.W.S. Kam, M. Shim, Y. Li, W. Kim, P.J. Utz, H. Dai, Noncovalent functionalization of carbon nanotubes for highly specific electronic biosensors, Proc. Natl. Acad. Sci. USA 100 (2003) 4984–4989.

[28] C. Li, M. Curreli, H. Lin, B. Lei, F.N. Ishikawa, R. Datar, R.J. Cote, M.E. Thomp-son, C. Zhou, Complementary detection of prostate-specific antigen using In2O3 nanowires and carbon nanotubes, J. Am. Chem. Soc. 127 (2005) 12484–12485.

[29] G. Zheng, F. Patolsky, Y. Cui, W.U. Wang, C.M. Lieber, Multiplex­ed electrical detection of cancer markers with nanowire sensor arrays, Nat. Biotechnol. 23 (2005) 1294–1301.

[30] F. Patolsky, C.M. Lieber, Nanowire nanosensors, Mater. Today 8 (2005) 20–28.[31] C.J. Roberts, P.M. Williams, J. Davies, A.C. Dawkes, J. Sefton, J.C. Edwards, A.G.

Haymes, C. Bestwick, M.C. Davies, S.J.B. Tendler, Real-space differentiation of IgG and IgM antibodies deposited on microtiter wells by scanning force micros-copy, Langmuir 11 (1995) 1822–1826.

[32] E.O. Saphire, R.L. Stanfield, M.D.M. Crispin, W.H.I. Parren, P.M. Rudd, R.A. Dwek, D.R. Burton, I.A. Wilson, Contrasting IgG structures reveal ex­treme asymmetry and flex­ibility, J. Mol. Biol. 319 (2002) 9–18.

[33] I. Pilz, E. Schwarz, W. Palm, Small-angle X-ray studies of the Fab and Fc frag-ments from the human immunoglobulin molecule K01, Eur. J. Biochern. 71 (1976) 239–247.

[34] A.F. Labrijn, P. Poignard, A. Raja, M.B. Zwick, K. Delgado, M. Franti, J. Binley, V. Vivona, C. Grundner, C. Huang, M. Venturi, C.J. Petropoulos, T. Wrin, D.S. Dim-itrov, J. Robinson, P.D. Kwong, R.T. Wyatt, J. Sodroski, D.R. Burton, Access of antibody molecules to the conserved coreceptor binding site on glycoprotein gp120 is sterically restricted on primary human immunodeficiency virus type 1, J. Virol. 77 (2003) 10557–10565.