troubleshooting laser diffraction particle size data
TRANSCRIPT
![Page 1: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/1.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Troubleshooting Laser Diffraction Particle Size Data
Ian [email protected]/us/particle
![Page 2: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/2.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Typical workflow
Calculation optimization
Hardware optimization
LA-950 data analysis tools
Q&A
Plan of Attack
![Page 3: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/3.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Typical Workflow
Compatible technologies?YES Compare ConditionsNO How are they different
Compare ConditionsSAME Investigate scattering patternDIFFERENT Are differences relevant for
this material?– YES Re-test with new conditions– NO Investigate scattering pattern
When you hit a wall Ask the experts!
![Page 4: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/4.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Compatible Technologies
Different technologies measure different material propertiesIdeally the comparison is apples to apples
– Different instruments of same technology are close enough (Gala to Fuji)
Diffraction to SEM (or similar) is like apple to oranges and must be approached differently
– Different technologies gives you moreinformation, not necessarily bad
![Page 5: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/5.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
SEM Image
![Page 6: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/6.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
LA-950 Result
SEM data alone makes us think this is the size range
![Page 7: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/7.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Typical workflow
Calculation optimization
Hardware optimization
LA-950 data analysis tools
Q&A
Plan of Attack
![Page 8: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/8.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Compare Conditions
First, check calculation conditionsRefractive Index: real and imaginaryDistribution Base: volume is bestIterations: wide or narrow size range
Second, check hardware conditionsConcentration: transmittanceParticle support: pumpingDispersion: US for wet, air pressure for dryDuration: wide distributions
![Page 9: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/9.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Refractive Index
Seemingly minor differences calculate different distributions
![Page 10: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/10.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Volume basis by defaultExcellent for mass balancingNumber basis recalc significant error
VolumeArea
Number
Distribution Base
![Page 11: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/11.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Typical workflow
Calculation optimization
Hardware optimization
LA-950 data analysis tools
Q&A
Plan of Attack
![Page 12: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/12.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Compare Conditions
First, check calculation conditionsRefractive Index: real and imaginaryDistribution Base: volume is bestIterations
Second, check hardware conditionsConcentration: transmittanceParticle support: pumpingDispersion: US for wet, air pressure for dryDuration: wide distributions
![Page 13: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/13.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
High enough for good S/N ratio Low enough to avoid multiple
scattering Typically 95 – 80 %T Measure at different T%, look at
Chi Square calculation
d50
Chi2
Concentration
![Page 14: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/14.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Must be high enough to suspend & circulate heavy particles
Not so high that bubbles are introduced
Adding energy – can disperse loose agglomerates
Measure at several settings & select optimum
Can be automated in software (see right)
Exp # Agitation Circulation Dmean (nm) D10 (nm) D90 (nm)
1 1 1 187.03 137.5 245.7
2 1 3 184.23 135.9 242.1
3 3 1 187.28 137.8 245.8
4 3 3 184.61 136.1 242.5
5 1 1 185.32 136.3 243.7
6 1 3 184.04 135.8 241.8
7 3 1 184.13 135.8 241.9
8 3 3 184.98 136.4 242.9
Parameters Selected: Agitation: 2 Circulation: 2
Pump & Stirrer
![Page 15: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/15.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Adding energy to break up agglomerates – disperse to primary particles, without breaking particles
Similar to changing air pressure on dry powder feeder Typically set to 100% energy, vary time (sec) on Investigate tails of distribution
High end to see if agglomerates removed Small end to see if new, smaller particles appear (breakage)
Test reproducibility, consider robustness Note:
Do not use on emulsions Can cause thermal mixing trouble w/solvents - wait Use external probe if t> 2-5 minutes
Ultrasonic Dispersion
![Page 16: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/16.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Long enough for reproducibility
Typically 5 sec, up to several minutes
Longer time for large, broad distributions
Can be automated in software
Measurement Duration
![Page 17: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/17.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Examples
![Page 18: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/18.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Typical workflow
Calculation optimization
Hardware optimization
LA-950 data analysis tools
Q&A
Plan of Attack
![Page 19: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/19.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
LA-950 Method Expert
![Page 20: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/20.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Method Expert Hardware
There are four important tests…
CirculationConcentration
DispersionDuration
![Page 21: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/21.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
There are two important tests…
Real RIImaginary RI
Method Expert Calculation
![Page 22: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/22.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Why is the test important?What does the test do?How will the results be displayed?What is the best value?
User selects up to 5 values for testing
LA-950 Method Expert
![Page 23: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/23.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Method Expert guides user to prepare the LA-950 for each test
Results displayed in multiple formats: PSD, D50, R-parameter
LA-950 Method Expert
![Page 24: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/24.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
LA-950 Method Expert
![Page 25: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/25.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
LA-950 Method Expert
![Page 26: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/26.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Intensity Graph
Diffraction analyzer measures light scattering pattern, algorithm transforms this into a particle size distribution
![Page 27: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/27.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
LARGE PARTICLE: Low angle scatter Large signal
SMALL PARTICLE: High Angle Scatter Small Signal
Wide Pattern - Low intensity
Narrow Pattern - High intensity
Size affects intensity
![Page 28: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/28.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Intensity Graph
![Page 29: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/29.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
For More InformationVisit www.horiba.com/us/particle
Previously recorded webinars:-- LA-950 Method Expert Software-- Understanding and Interpreting Your PSA Results-- Setting Attainable Size Specifications-- “Boot Camp” Laser Diffraction 6-part Training Course
Many application and technical notes
Ask a question at [email protected]
Keep reading the monthly HORIBA Particle e-mail newsletter!
Visit the Download Center to find the video from this webinar.
![Page 30: Troubleshooting Laser Diffraction Particle Size Data](https://reader036.vdocuments.mx/reader036/viewer/2022081512/5568f2ffd8b42aff2e8b4940/html5/thumbnails/30.jpg)
© 2011 HORIBA, Ltd. All rights reserved.
Typical workflow
Calculation optimization
Hardware optimization
LA-950 data analysis tools
Q&A
Plan of Attack