total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation,...

8
Received 23-02-2006; Accepted 07-04-2006 Author and address for correspondence: Rodica Anghel, MD, PhD Institute of Oncology Bucharest 252 Fundeni Street, Sector 2 PO 022328 Bucharest Romania E-mail: [email protected] Journal of BUON 11: 167-174, 2006 © 2006 Zerbinis Medical Publications. Printed in Greece ORIGINAL ARTICLE T otal body irradiation prior to bone marrow transplantation - the experience of the Institute of Oncology “Prof. Dr. Al. Trestioreanu” Bucharest R. Anghel 1,2 , G. Matache 2 , M. Vasile 2 , R.I. Matache 1 , L. Oprea 2 , R. Popa 2 , A. Sucitu 2 , N. Costandache 2 , I. Bărbulescu 2 , D. Colita 3 , Z. Varady 3 , A. Tanase 3 , A. Moicean 3 , C. Arion 4 , A. Colita 4 , L. Dumitrache 4 1 University of Medicine and Pharmacy “Carol Davila”, Bucharest; 2 Institute of Oncology “Prof. Dr. Al. Trestioreanu”, Department 2 of Radiotherapy, Bucharest; 3 Fundeni Clinical Institute, Department of Hematology and Bone Marrow Transplantation, Bucharest; 4 Fundeni Clinical Institute, Department of Pediatrics and Department of Bone Marrow Transplantation Department, Bucharest, 4 Romania Summary Purpose: T o present the technique of total body ir- radiation (TBI), applied for the rst time in Romania, at the Institute of Oncology Bucharest, as part of stem cell transplantation for hematological malignancies. Patients and methods: The total dose administered was 12 Gy at the reference point, 2 Gy/fraction, one fraction per day , 6 consecutive days, with a total dose of 8 - 11.4 Gy delivered to the lung, using Mevatron Primus linear ac- celerator (6 MV & 15 MV, 200-300 cGy/min in isocenter), in vivo dosimetry detectors and equipment for the reference dosimetry, personalized blocks for lung shielding sustained by polymethylmethaacrylate (PPMA) plate, Simulix HP simulator, and computer tomographic (CT) scans. Tech- niques used were: a) two parallel opposed anteroposterior / posteroanterior (AP/PA) elds with the patient in prone and supine position; b) two parallel opposed lateral elds with the patient placed on a lateral table, at 320 cm from the source. The percentage depth dose, tissue maximum ratio (TMR), off axis ratio (OAR) and the reference dose rate were measured for every patient’s geometrical characteristics, with an uncertainty of ± 2.2% and were used to calculate monitor units and to evaluate the dose in organs at risk (lungs, gonads, eyes etc). Results: 5 patients (3 with the AP/PA technique and 2 with the lateral technique) were irradiated. All patients completed their irradiation in good clinical condition. The acute side effects were minimal (WHO grade 1: nausea/ vomiting – all patients; diarrhea – 1 patient; headache – 2 patients; photophobia and diplopia – 1 patient; head and neck skin erythema – all patients). Because of the short fol- low-up period no safe evaluation of late side effects can be done. However, during this period one patient developed a non-aggressive form of chronic liver graft vs. host disease (GVHD) and one patient died due to acute GVHD. Conclusion: TBI as part of stem cell transplantation for hematological malignancies was successfully realized at our Institute, with favorable clinical results. This technique is easy to carry out and reproducible. Key words: hematological malignancies, in vivo dosim- etry , radiotherapy , stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy for hematological malignancies with rapid expansion during the last 30 years (more than 23,000 transplan- tations in 2001 all over the world), with remarkable results [1]. Nowadays the main indications for stem cell transplantation are acute lymphoblastic leukemia (ALL), acute myeloid leukemia (AML), chronic my- elogenous leukemia (CML), multiple myeloma, some malignant lymphomas, and some solid tumors.

Upload: others

Post on 05-Oct-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy

Received 23-02-2006; Accepted 07-04-2006

Author and address for correspondence:

Rodica Anghel, MD, PhDInstitute of Oncology Bucharest252 Fundeni Street, Sector 2PO 022328BucharestRomaniaE-mail: [email protected]

Journal of BUON 11: 167-174, 2006© 2006 Zerbinis Medical Publications. Printed in Greece

ORIGINAL ARTICLE

Total body irradiation prior to bone marrow transplantation - the experience of the Institute of Oncology “Prof. Dr. Al. Trestioreanu” Bucharest

R. Anghel1,2, G. Matache2, M. Vasile2, R.I. Matache1, L. Oprea2, R. Popa2, A. Sucitu2, N. Costandache2, I. Bărbulescu2, D. Colita3, Z. Varady3, A. Tanase3, A. Moicean3, C. Arion4,A. Colita4, L. Dumitrache41 University of Medicine and Pharmacy “Carol Davila”, Bucharest; 2Institute of Oncology “Prof. Dr. Al. Trestioreanu”, Department 2

of Radiotherapy, Bucharest; 3Fundeni Clinical Institute, Department of Hematology and Bone Marrow Transplantation, Bucharest;4Fundeni Clinical Institute, Department of Pediatrics and Department of Bone Marrow Transplantation Department, Bucharest,4

Romania

Summary

Purpose: To present the technique of total body ir-radiation (TBI), applied for the fi rst time in Romania, at the Institute of Oncology Bucharest, as part of stem cell transplantation for hematological malignancies.

Patients and methods: The total dose administered was 12 Gy at the reference point, 2 Gy/fraction, one fraction per day, 6 consecutive days, with a total dose of 8 - 11.4 Gy delivered to the lung, using Mevatron Primus linear ac-celerator (6 MV & 15 MV, 200-300 cGy/min in isocenter), in vivo dosimetry detectors and equipment for the reference dosimetry, personalized blocks for lung shielding sustained by polymethylmethaacrylate (PPMA) plate, Simulix HP simulator, and computer tomographic (CT) scans. Tech-niques used were: a) two parallel opposed anteroposterior / posteroanterior (AP/PA) fi elds with the patient in prone and supine position; b) two parallel opposed lateral fi elds with the patient placed on a lateral table, at 320 cm from the source. The percentage depth dose, tissue maximum ratio (TMR), off axis ratio (OAR) and the reference dose rate were measured for every patient’s geometrical characteristics,

with an uncertainty of ± 2.2% and were used to calculatemonitor units and to evaluate the dose in organs at risk (lungs, gonads, eyes etc).

Results: 5 patients (3 with the AP/PA technique and 2 with the lateral technique) were irradiated. All patientscompleted their irradiation in good clinical condition. Theacute side effects were minimal (WHO grade 1: nausea/vomiting – all patients; diarrhea – 1 patient; headache – 2patients; photophobia and diplopia – 1 patient; head and neck skin erythema – all patients). Because of the short fol-low-up period no safe evaluation of late side effects can bedone. However, during this period one patient developed anon-aggressive form of chronic liver graft vs. host disease(GVHD) and one patient died due to acute GVHD.

Conclusion: TBI as part of stem cell transplantationfor hematological malignancies was successfully realized at our Institute, with favorable clinical results. This techniqueis easy to carry out and reproducible.

Key words: hematological malignancies, in vivo dosim-etry, radiotherapy, stem cell transplantation, total bodyirradiation

Introduction

Stem cell transplantation is a systemic therapyfor hematological malignancies with rapid expansionduring the last 30 years (more than 23,000 transplan-tations in 2001 all over the world), with remarkableresults [1]. Nowadays the main indications for stemcell transplantation are acute lymphoblastic leukemia(ALL), acute myeloid leukemia (AML), chronic my-elogenous leukemia (CML), multiple myeloma, somemalignant lymphomas, and some solid tumors.

Page 2: Total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy

168

TBI is an important technique, an integral part of the protocols of allogenous or autologous stem cell transplantation and has 3 main purposes:

1. “Cleaning” the host marrow to allow repopu-lation with donor marrow cells;

2. Providing suffi cient degree of immunosup-pression to avoid allograft rejection by immunologi-cally active cells of the host;

3. Destroying residual neoplastic cells; and these goals must be achieved with minimum side effects for the patients.

The selection of an optimal value of risk – benefi t ratio, associated with local technical possibilities de-termined the development of a great variety of irradia-tion techniques and treatment schedules [2,3].

There is no ideal technique to irradiate the whole body homogeneously and precisely; AP/PA techniques are mostly used, with the patient sitting on the table or standing, or on a mobile table with speed controlled by computer, with adjacent fi elds or latero-lateral fi elds; each technique has its advantages and disadvantages re-garding the dose distribution accuracy and the patient’s tolerance to the treatment as well [3].

As for the treatment schedule, reference doses of 5 to 16 Gy are delivered (6 - 15 Gy to the lung) in 1-13 fractions on days 1-6.

In this paper we present the experience and the results of the Institute of Oncology “Prof.Dr.Al.Tres-tioreanu” Bucharest with TBI as part of conditioning regimen to stem cell transplantation, a technique that was used for the fi rst time in our country since the beginning of November 2003.

Patients and methodsPatients

Five patients (4 males and 1 female) were se-

lected for TBI, aged between 13 and 48 years. In 4cases allogenous and in one patient autologous stemcell transplantation were carried out.

The patients were diagnosed with AML-M4 withcentral nervous system relapse, ALL also with centralnervous system relapse, ALL and non-Hodgkin’s lym-phoma.

Patients’ selection (Table 1) for allogenous or autologous stem cell transplantation was performed inthe Bone Marrow Transplantation Department of theFundeni Clinical Institute, according to the protocol’scriteria. The current indications are high risk patientswith leukemia or lymphoma in 1st complete remissionand patients in 2nd complete remission. TBI + cyclo-phosphamide is the standard conditioning protocol for patients with acute leukemia below 50 years of age,without comorbidities. After selection, the patientswere treated according to the chosen protocol.

Treatment protocol

We use the following treatment protocol:Day -9 → - 4: TBI, total dose 12 Gy.Day -3 → -2: high dose chemotherapy - cyclo-

phosphamide 120 mg/kg/day.Day 0: stem cell transplantation.GVHD prophylaxis: Day -1 → day +180: cyclo-

sporine 1 mg/kg/day; and short term methotrexate p.o.:15 mg/m2 day+1, and 10 mg/m2 day +3 and +6.

Irradiation technique

The radiotherapy equipment used was linear accelerator MEVATRON PRIMUS; pho ton-beamwith nominal energy of 6 MV, with a dose rate in theisocentre of 200 cGy/min.

Initially, taking into account the need for a pre-cise dose distribution correlated with minimal side

Table 1. Patient characteristics

Age(years) Diagnosis Total dose

(Gy)Median dose to the

lung (Gy) Current state

13 Acute myeloid leukemia – AML-M4 12 8.0 Central nervous system relapse.Without signs of GVHD.

18 Acute lymphoblastic leukemia 12 9.4 Complete remission.Chronic liver GVHD – non-aggressiveform.

48 Acute lymphoblastic leukemia 12 9.6 Died on day +72 of acute GVHD

13 Non-Hodgkin’s lymphoma 12 9.8 Complete remission.

20 Acute lymphoblastic leukemia 12 11.4 Complete remission.Without signs of GVHD.

GVHD: graft versus host disease

Page 3: Total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy

169

effects (especially interstitial pneumonitis), we chose an irradiation technique that allows the best control of the delivered dose to the entire target volume and especially to the lungs by using accurate positioning individualized blocks (Figures 1, 2).

The patients were placed in prone and supine position in a wooden case and covered with rice, in-cluding the lateral sides to obtain an equilibrium layer of 1 – 2 cm; with this approach we intended to move the maximal dose to the skin surface and eliminate dose inhomogeneities resulting from patient thickness variation (excepting the head), obtaining an inhomo-geneity of maximum ±7 % (Figure 3) [4].

The source-skin distance (SSD) was 190 cm, and the beam orientation was vertical down.

The absorbed dose rate in the prescription point was approximately 50 cGy/min, depending on the AP diameter of the patient.

We used CT scans to calculate the lung correc-

tion factor and also to define the lung volume and position with regard to bony marks.

For the designing of the lung blocks we used SIMULIX simulator, with the same irradiation geom-etry as during the treatment.

For the last 2 patients we modifi ed the irradiationtechnique and used two parallel opposed lateral fi elds,with the patient in supine position on a lateral table, withthe knees immobilized; the accelerator’s head was rotated at 900, collimator at 450 and the fi eld size was 40 cm × 40cm at 100 cm distance; in this manner the SSD was 320cm and the variation of the dose was less than 10%. Thedose rate at the treatment distance was 20.5 cGy/min.

Irradiation schedule

The total dose in the reference point was 12 Gy,delivered in 6 days, one fraction / day, 2 Gy/fraction (1Gy on the AP and the PA fi eld each) and right and left fi eld respectively (for the last 2 patients, with lateraltechnique). The irradiation was performed during days-9 → - 4. The prescription point was on the middleline, at the umbilicus.

The dose delivered to the lungs was carefullyevaluated; in order to reduce the risk of interstitialpneumonitis the average dose to the lung was 8 Gy inthe fi rst patient; for the remaining patients the dose tothe lung was raised to 9.0 – 9.6 Gy and even to 11.4 Gy[5]. In the lateral technique the protection of the lungwas realized by arms tightly fl exed and immobilized with arm holder; the correct and the best position of the arms was verifi ed at the simulator.

Two patients had central nervous system relapseand had already received whole brain irradiation witha total dose of 20 Gy; this previous dose did not cause

Figure 1. Patient’s alignment.

Figure 2. Marking the protection limits on the skin.

Figure 3. Irradiation of the patient in AP position.

Page 4: Total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy

170

a reduction of the dose delivered to the brain during the total body irradiation.

During the 6 days of irradiation the patients re-ceived prophylactic medication to reduce the incidence of acute side effects – for acute gastrointestinal toxicity (nausea, vomiting) they received granisetron 3 mg and dexamethasone 8 mg i.v. half an hour before and after irradiation; for increased intracranial pressure they received mannitol 100 ml half an hour before and after irradiation [6].

The patients were monitored for acute side ef-fects and a blood count was performed every day dur-ing the 6-day duration of irradiation.

Dosimetry

Because of atypical irradiation geometry, to en-sure an optimal control of the relative dose distribution into the body and especially of the absorbed dose at the prescription point and in other important points as well, our team performed specifi c dosimetric verifi cation in the same geometry for both techniques (parallel op-posed AP/PA and parallel opposed lateral technique).

Tissue maximum ratio measurement (Figure 4)

TMR was measured in a water phantom and in PMMA plates; we proved that this ratio does not depend on the distance source – measurement point and is dependent on the fi eld size in the measurement point. The percentage depth dose in irradiation condi-tions was also measured.

Off Axis Ratio measurements (Figure 5)

OAR measurements were performed in water phantom and also in PMMA plates, for different depths.

Absolute dosimetry

The absolute dosimetry that allowed the monitor units calculation was performed with ionizing cham-ber type M 30001 PTW, connected to the electrometer UNIDOS; the chamber had been calibrated beforehand in absorbed dose in water with a gamma 60Co beam,according to the Secondary Standard Laboratory Do-simeter (SSLD Bucharest). The measurements wereperformed in water phantom and in PMMA platesunder the exact irradiation conditions; the IAEA TRS398 Code of Practices was applied [7].

Dosimeters’ calibration for in vivo dosimetry

For in vivo verifi cation of the delivered dose to thepatients we used Metal Oxide Semiconductor Field Ef-fect Transistor (MOSFET) and semiconductor detectors(PTW, connected to a MULTIDOS electrometer) cali-brated in operating conditions (SSD, fi eld dimensions at the prescription point, beam energy); the detectors werecalibrated with electronic equilibrium plate, accordingto the information given by the ionizing chamber placed in water, at the maximum dose point. The externalinfl uences (temperature at the detector’s level, fi eld dimensions) were analyzed, and after each irradiationfraction the doses were measured in at least two points(one being the prescription point).

With the semiconductor detectors we were ableto simultaneously determine in vivo doses in 12 points,including the doses at the entrance and exit points(Figure 6) [8].

Measurement of the individualized blocks coefficient

The correction coeffi cient for the absorbed doseto the lung was calculated based on the information

Figure 4. Typical tissue maximum ratio curves.Figure 5. Typical percentage depth dose curves, horizontal ge-ometry.

Page 5: Total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy

171

obtained from the CT scans. One of the most important side effects, with serious consequences for the patient, is interstitial pneumonitis after large doses. In order to reduce the dose to the lung at a desired value individu-alized blocks using the SIMULIX HP simulator were manufactured, with the patient in treatment position (the blocks were separately manufactured for AP and PA fi elds); the thickness of the block was relevant to the desired dose to the lung. For every block the attenuation coeffi cient in treatment position was measured using the phantom and the absolute dosimetry equipment. The median value of the attenuation coeffi cient was used to evaluate the real dose delivered to the lung.

Elaboration of the treatment planning

The treatment planning was individualized for every patient, using patient’s dimensions and the information acquired from the phantom under operating conditions.

For the reference prescribed dose (12 Gy) the absorbed doses in different points of interest (lung, mediastinum, head, neck, eye’s back, pelvis, legs etc.) were measured. Also the expected dose in the points of the in vivo detectors and the maximum dose in these points was calculated.

Verification with radiological films

KODAK X-omat radiological fi lms placed under the patient during the irradiation session were used to confi rm the correct positioning of the lung protection (in AP/PA technique) and to verify the correct posi-tioning of the adjacent fi elds.

Reference dose verification

After each fraction of irradiation the reference dose rate and the dose by monitor unit were measured.

The measurement was accomplished with the ionizingchamber placed in PMMA plates in the prescriptionpoint of the patient.

Results

Five patients (3 with the AP/PA technique and 2with the lateral technique) were irradiated. The totaldose delivered to the prescription point was 12 Gy and to the lung 8 - 11.4 Gy.

The absorbed doses calculated in other points thanthe prescription point were between 11.1 Gy and 13.0Gy, leading to an inhomogeneity of±8 % for the AP/PA technique, with rice as compensator material, and be-tween 10.1 Gy to 15 Gy for the lateral technique.

The values measured with semiconductor detec-tors were about ±5 % from the calculated values.

From a dosimetric point of view, in Figures 4-7represented are typical tissue – maximum dose curvesand profi les that helped us achieve an individualized planning treatment.

All patients completed their irradiation in good clinical condition.

The acute side effects were minimal (WHO gra-de 1) and consisted in: nausea, vomiting - all patients,controlled with antiemetic agents; diarrhea - 1 patient;headache - 2 patients; controlled with corticosteroidsand mannitol; photophobia and diplopia - 1 patient;controlled with corticosteroids and mannitol; skin ery-thema, especially to the head and neck - all patients.

As for late side effects it is too soon to have anevaluation because of the short period of follow-up(18 months, 11 months, and 10 months and 2 weeks re-spectively). During this period one patient developed anon-aggressive form of chronic liver GVHD and onepatient died on day 72 due to acute GVHD.

Figure 6. Typical off axis ratio curves for vertical irradiation. Figure 7. Typical off axis ratio curves for lateral technique.

Page 6: Total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy

172

Discussion

TBI is an important component of the stem cell transplantation protocols. Some trials showed an im-provement of disease-free survival rates for the patients who underwent TBI for stem cell transplantation [9].

TBI offers several advantages: the delivered dose is homogenous to the entire body, independent of the blood supply; all cells in the body will be treated, in-cluding sanctuary sites (body compartments where ef-fective anticancer drug levels cannot be achieved, such as the central nervous system or the testes); irradiation is not cross-reactive with any chemotherapeutic agent, the dose distribution can be modifi ed by using blocks for organs at risk (e.g. lung) or the dose can be boosted to organs with high resistance to treatment (e.g. testes); TBI offers also a better immunosuppression.

There is no ideal technique for TBI. Litera-ture review reveals that treatment methods for TBI are heterogeneous, with a large variety concerning fractionation (single dose, multiple fractions, highly fractionated regimens) and treatment planning (paral-lel opposed AP/PA, parallel opposed lateral fi elds or a combination of these techniques) [9,10].

Why did we choose a multiple fractions ap-proach? Studies [11,12] suggest that fractionated TBI is better than single dose in reducing toxicity to normal tissues and the incidence of interstitial pneumonitis [13], without increasing the relapse rate. Interstitial pneumonitis is of great concern following TBI. It usu-ally occurs within 90 days after transplantation and accounts for approximately 40% of transplantation-related deaths. Its incidence depends on total dose and fractionation, with increased incidence for single dose TBI rather than fractionated regimens [9].

Today the majority of the centers use fractionated regimens [3].

Vriesendorp et al. [14] created tables of surviv-ing target cells (e.g. lung, intestine, bone marrow); they showed a dramatic sparing of lung and intestine with more fractionated regimens, probably due to their large shoulder on the survival curve and ability to repair sublethal damage.

Another advantage of fractionated regimens is the possibility to administer a larger total dose, with a bet-ter immunosuppressive effect (numerous experimental studies have demonstrated that the immunosuppressive effect is highly fractionation-sensitive) [14,15].

Why did we initially choose the parallel opposed AP/PA technique? Many centers prefer this technique to that with parallel opposed lateral fi elds because of better homogeneity of dose distribution (the dose variation to the various parts along the axis of the body

ranges from ±7 - ±10%) [15,16]. Our patients had adose inhomogeneity of ±8%.

For the parallel opposed lateral technique dosevariation can range up to 50% in the head and neck and 10% in other parts of the body.

It must be mentioned that for the lateral tech-nique it is easier to do individualized protection toorgans at risk.

The lateral technique allows the irradiation of thewhole body with two opposed fi elds, lowering the risk of overdosage by overlapping fi elds and offers better convenience for the patient during radiotherapy. It alsomakes things easier for the working team, by dimin-ishing the treatment time up to 50% versus the AP/PA technique [15,17].

As for the results and side effects it is too soon tocome to conclusions. Still, from comparative analysisof literature data it can be stated that a single dose of 9.9 Gy, which is equivalent with fractionated 12-13.5Gy, can result in 7-year survival rate of 74% and in on-ly 38% if the dose is lower than 9.9 Gy; increasing thesingle dose over 10 Gy (and of fractionated total doseover 12-13.5 Gy) it is not necessary correlated with a7-year increased probability of survival [18,19].

As for the immunosuppressive effect, studiesshow that patients treated with fractionated irradiationwith 12 Gy total dose have a higher probability of de-veloping chronic GVHD than those treated with doseslower than 12 Gy [17,18]. In univariate and multivari-ate analysis GVHD was the main indicator of relapseprevention after stem cell transplantation for acuteleukemia and it favors overall survival [20].

Conclusions

Four out of 5 patients that received TBI are aliveat 18, 11, 10 months and 2 weeks, respectively, fromthe stem cell transplantation. One patient developed chronic and one acute GHVD.

Total doses and fractionation schemes were ac-cording to internationally accepted protocols.

The performance status of the patients duringand after treatment makes us consider this protocolacceptable, with good results.

The treatment technique has been chosen in amanner to achieve some important goals of TBI, suchas the use of rice.

Homogenous dose distribution: Using rice ascompensator material leads to irradiation geometrysimilar to phantom (used for absolute and relativedosimetry), minimizing the infl uence of patient thick-ness variation.

Page 7: Total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy

173

Accurate checking of personalized blocks posi-tioning and of doses administered to the lung: Using this technique the protection blocks can be positioned with accuracy for every patient by using the same coordinates at the simulator and during irradiation; individualized measures of transmission for the personalized blocks offer a better control of the dose administered to the lung.

The main disadvantage of AP/PA technique is represented by the overlapping of adjacent fi elds, but with rigorous calculation of adjacent fi elds’ position the over- and underdosage can be eliminated. Use of radiologic fi lms for high doses (X-omat) allows bet-ter visualization of adjacent fi elds; the calculus and the measurements show that dose inhomogeneities in adjacent areas are less than ±10 %.

To avoid overlapping of adjacent fi elds we intro-

duced the lateral technique, with 2 parallel opposed fi elds, with the patient in prone position and with theaccelerator’s head rotated at 900, collimator at 450; toassure an homogenous irradiation to the skin we used a PMMA plate placed between the patient and the ra-diation source, as close as possible to the patient; theSSD was 320 cm (Figures 7-9). The method has theadvantage that the whole body is irradiated at the sametime, without the risk of over- or underdosage; it is alsobetter tolerated by the patient and the total time for totaldose delivery is shorter than with AP/PA technique.

The disadvantage of the lateral technique is thehigh dose inhomogeneity. But the method has theessential advantage of in vivo dose measurement in12 points simultaneously. The calibration of these12 detectors was done simultaneously and they weregiven numbers – even numbers for exit points and odd numbers for enter points. Each pair (odd-even)

Figure 8. MOSFET detectors calibration.

Figure 9. Patient during irradiation, lateral technique.

Figure 10. Dose measurement at the points of interest by in vivodosimetry.

Page 8: Total body irradiation prior to bone marrow ... · etry, radiotherapy, stem cell transplantation, total body irradiation Introduction Stem cell transplantation is a systemic therapy

174

offers dosimetric information for an organ at risk or from the prescription point (umbilicus) (Figure 10). There are 6 points of interest for in vivo dosimetry, and the values indicated by the detectors are noted in a Table and compared with the expected values. The reproducibility of the detectors’ results over time as well as the accuracy of the measurements were according to international standards.

References

1. Hoffstetter S, Marchal C, Bordigoni P et al. Total body irra-diation in France in the past twenty years. Cancer Radiother 2003; 7: 166-171.

2. Dobbs J, Barrett A, Ash D. Systemic Irradiation. In: Dobbs J, Barrett A, Ash D (eds): Practical Radiotherapy Planning (3rd edn). Arnold Publ, London, New York, 1999, Ch 37, pp 362-368.

3. Van Dyk J. Special Techniques in Radiotherapy; Part A: Total body irradiation with photon beams. In: Van Dyk J (ed): The Modern Technology of Radiation Oncology. Medical Physics Publishing, Madison, Wisconsin 1999, Ch 37, pp 641-663.

4. Mangili P, Fiorino C, Rosso A et al. In vivo dosimetry by diode semiconductors in combination with portal fi lms during TBI: reporting a 5 year clinical experience. Radiother Oncol 1999; 52: 269-276.

5. Volpe AD, Ferreri AJS, Annaloro C et al. Lethal pulmonary complications signifi cantly correlate with individually as-sessed mean lung dose in patients with hematologic malig-nancies treated with total body irradiation. Int J Radiat Oncol Biol Phys 2002; 52: 483-488.

6. Belkacémi Y, Ozsahin M, Pène F et al. Total body irradiation prior to bone marrow transplantation: effi cacy and safety of granisetron in the prophylaxis and control of radiation-induced emesis. Int J Radiat Oncol Biol Phys 1996; 36: 77-82.

7. Fiorino C, Mangili P, Cattaneo GM et al. Polarity of ioniza-tion chambers used in TBI dosimetry due to cable irradiation. Med Dosimetry 2000; 25: 121-126.

8. Duch MA, Ginjaume M, Chakkor H et al. Thermoluminescence dosimetry applied to in vivo dose measurements for total

body irradiation techniques. Radiother Oncol 1998; 47:319-324.

9. Wheldon TE, Barrett A. Radiobiological modelling of thetreatment of leukaemia by total body irradiation. Radiother Oncol 2001; 58: 227-233.

10. Abraham D, Colussi V, Shina D et al. TBI treatment planning using the ADAC Pinnacle Treatment Planning System. Med Dosimetry 2000; 25: 219-224.

11. Girinsky T. Benhamov E, Bourhis JH et al. Prospective ran-domized comparison of single-dose versus hyperfractionated total body irradiation in patients with hematologic malignan-cies. J Clin Oncol 2000; 18: 981-986.

12. Sagus Sampath BA, Timothy E, Schultheiss JW. Dose re-sponse and factors related to interstitial pneumonitis after bone marrow transplant. Int J Radiat Oncol Biol Phys 2005;63: 876-884.

13. Thomas L, Falk PM, Kogut N, Kayur HS, Tome M, KaganR. A comparison of single-dose and fractionated total bodyirradiation on the development of pneumonitis followingbone marrow transplantation. Int J Radiat Oncol Biol Phys1996; 36: 61-66.

14. Vriesendorp HM, Chu H, Ochran TG et al. Radiobiology of total body radiation. Bone Marrow Transpl 1994; 14(Suppl):54-58.

15. Hsiu-San Lin, Drzymala R. Total body and hemibody irradia-tion. In: Perez CA, Brady LW (eds): Principles and Practice of Radiation Oncology (3rd edn). Lippincott-Raven Publishers,Philadelphia, 1997, pp 333-340.

16. Gerrard GE, Vail A, Taylor RE et al. Toxicity and dosimetryof fractionated total body irradiation prior to allogeneic bonemarrow transplantation using a straightforward radiotherapytechnique. Clin Oncol 1998; 10: 379-383.

17. Lawton CA. Total Body Irradiation for Stem Cell Transplant Patients. 42nd Ann Sci Meet Am Soc Ther Radiol Radiat Oncol, October 2000 (refresher course no. 215).

18. Gocheva L. Total body irradiation prior to bone marrowtransplantation; some aspects of fi fty year experience. JBUON 2004; 9: 147-160.

19. Pape H. Knochenmark transplantation: Die Rolle der Strahlen-therapie. Schweiz Rundsch Med Prax 1996; 85: 749-752.

20. Panigrahi S, Morecki S, Yakovlev E et al. A novel approachfor prevention of lethal GVHD by selective elimination of alloreactive donor lymphocytes prior to stem cell transplanta-tion. Exper Hematol 2004; 32: 756-764.