the spa, a new miniature mri-compatible stepping piezo ......the spa, a new miniature mri-compatible...

16
The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen 1 , Christian Belly 1 , Hervé Mathieu 2,3,4 [email protected] 06/04/2011 1 Cedrat Technologies, Grenoble, France 2 INSERM, U836 , Grenoble, France 3 University Joseph Fourier , Grenoble, France 4 MRI facility, Grenoble, France FORUM INNOVATIONS FOR INDUSTRY – AT MICRONANOTEC.

Upload: others

Post on 14-Dec-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications

Frank Claeyssen1, Christian Belly1, Hervé Mathieu2,3,4

[email protected]

06/04/2011

1 Cedrat Technologies, Grenoble, France2 INSERM, U836 , Grenoble, France3 University Joseph Fourier , Grenoble, France4 MRI facility, Grenoble, France

FORUM INNOVATIONS FOR INDUSTRY – AT MICRONANOTEC.

Page 2: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 2

Outline

Stepping Piezo Actuator (SPA) concept

SPA30uXS

MRI compatibilityThree rulesMRI test benchExperimental results

Conclusions

Page 3: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 3

Piezoelectricity?

A link between voltage and mechanical stressDirect effect : SensorReverse effect : Actuator

Amplified Piezo Actuator

Page 4: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 4

Linear piezo motorLong stroke (several millimeters)High resolution (nanometer range)

Components:

Mass APA (piezo) Rod Clamp

SPA60SM based on APA60SMSPA Structure

What is the Stepping Piezo Actuator ?

APA60SM

Page 5: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 5

How does the SPA work?

2 working modesStepping mode Stick and slip

Slow contraction Fast expansion

Long stroke

Deformation mode Standard use of Amplified Actuator Nanometric resolution

Sticking

SlidingDis

plac

emen

t

Time

Volta

ge

Page 6: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 6

Prototypes

SPA30uXSMass: 2 gramsLess than 500 mm3

Stroke = 4mmResolution < 5nmSpeed : from 0 to 70 mm/sForce : Up to 0,2 N

SPA30uXS versus a 1 Euro Cent coin

Page 7: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 7

MRI compatibility

Three MRI compatibility rulesSafetyInvisibilityImperturbability

Page 8: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 8

MRI compatibility

Three MRI compatibility rulesSafety: Must not be attracted Non magnetic actuator Tests performed into an MRI 4,7 Tesla magnet

These tests have been performed on Small Animals platform, INSERM Grenoble, France

Page 9: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 9

MRI compatibility

Three MRI compatibility rulesSafetyInvisibility: Must not trouble MRI scanner and images Tests performed into an MRI 4,7 Tesla magnet

« Phantom »,Area tested

Reception antenna

MRI scanner

Page 10: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 10

MRI compatibility

Three MRI compatibility rulesSafetyInvisibility: Must not trouble MRI scanner and images Examples of image distortion:

Inox screw replaced by Titanium screw Antenna phenomenon caused by wires

Distortion caused by inox screw Antenna phenomenon with wires

Page 11: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 11

MRI compatibility

Three MRI compatibility rulesSafetyInvisibility: Must not trouble MRI scanner and images SNR measurement on phantom, realized on 20 images Without motor : 55.9 with standard variation of 2.5 With motor : 55.7 with standard variation of 2.5

Actuator position (dot) against measured position (dashed)

Page 12: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 12

MRI compatibility

Three MRI compatibility rulesSafetyInvisibility: Must not trouble MRI scanner and images SNR measurement on rat, realized on 10 images Without motor : 33.6 with standard variation of 1.8 With motor : 32.5 with standard variation of 3.1

Rat

Motor

Page 13: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 13

MRI compatibility

Three MRI compatibility rulesSafetyInvisibilityImperturbability : Must keep performances within MRI Scanner Adapted test bench Axial tool motion simulation No adapted sensor Use of Cine-Flash images MRI high vertical resolution (20µm) 1 Image/350ms

Page 14: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 14

MRI compatibility

Three MRI compatibility rulesSafetyInvisibilityImperturbability : Must keep its performances under magnetic field No measured perturbation from the high magnetic field

Reconstructed sequence

Page 15: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 15

Conclusion (1)

SPA30uXS quick facts

Large linear speed range (from 0 to 70 mm/s)Large linear force (up to 30 grams force)Holding force with no consumption (80gr force)

MRI-compatible Safety Invisibility Imperturbability

Bio Med Applications in progress Medical implants Robotic surgery

Page 16: The SPA, a new miniature MRI-compatible Stepping Piezo ......The SPA, a new miniature MRI-compatible Stepping Piezo Actuator for Bio Medical Applications Frank Claeyssen1, Christian

Hannover Messe 2011 / Innovation Forum / SPA : MRI-compatible actuator for Bio Med 16

Conclusion (2)

SPA30uXS commercial availabilityStand alone motorDeveloper Kit including Motor Position sensor Controller PC interface

Customised piezo motorsLarger-force linear motors SPA40SM Driving force = 20N

Rotational motors RSPA30uXS

Please visit CEDRAT booth at H2 D55