the service and support of recordum and related topics aleksandar tomic, june 2012
TRANSCRIPT
Service InterfaceGet to your system’s service interface LinSens
See all diagnostic data in real-time
Ticket Systemsupport.recordum.com
Open a new ticket
Check ticket status
For internal use onlyDON’T give this to customers
Ticket SystemType in your contact details
Select the productPlease also indicate serial number, software version and remote accessibility as far as known
Select a topic an send describe your issue and attach a file if needed
e-mail notification
Ticket SystemCheck your ticket status by logging in with your e-mail and a ticket number
Select if you want to see all/open/closed tickets
Post replies to our suggestions
Ticket SystemE-Mail Alert to administrator that a new ticket is opened
Notifies last respondent when a new post is added
TestbenchWorkstation PC
airpointer PC
AC and DC power connectors
Connectors for analysis module
Connect board
TestbenchEasily connect the airpointer gas analysis module to the test bench
Check all electronics of the module
TestbenchSame user interface as the airpointer
Select the connected analysis module
Work in LinSens
When finished shut the airpointer PC down (Service Manager) and turn the workstation PC off
Leak CheckerCheck if airpointer’s analysis module (or any other gas analyzer) is tight
Plug the inlet of the module and connect the Leak Checker on the other end
If the vacuum of the pump doesn’t stay stable after it is switched off, a leak is present